The Wrist Watch Electronic Blood Pressure Monitor uses the oscillometric method to measure diastolic blood pressure, systolic blood pressure. It is intended for use in adults with a wrist circumference between 15-21 cm who do not have arrhythmias (such as atrial fibrillation) and who are not pregnant. The values are for reference purposes and do not constitute a diagnosis.
Device Story
Wrist-worn electronic blood pressure monitor; utilizes oscillometric method via integrated airbag/cuff; performs automatic inflation/deflation; measures systolic and diastolic blood pressure. Intended for home/OTC use by adult patients. Device consists of main unit, watch strap with airbag, and charging cable. Output displayed to user for reference; does not provide clinical diagnosis. Includes irregular heartbeat detection (threshold >±25% mean pulse interval).
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018+AMD1:2020+AMD2:2024. Study included 103 subjects; auscultation with calibrated mercurial sphygmomanometer used as gold standard. Results: Systolic mean deviation 1.3 mmHg (SD 6.2 mmHg); Diastolic mean deviation -1.5 mmHg (SD 4.9 mmHg). Results met acceptance criteria (mean ≤ ±5.0 mmHg, SD ≤ 8.0 mmHg).
Technological Characteristics
Oscillometric measurement; lithium battery (3.85V) power; wrist-worn form factor. Biocompatible materials (cytotoxicity, irritation, sensitization testing passed). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30, ISO 81060-2. Connectivity includes 2.4 GHz ISM band radio (EN 300328).
Indications for Use
Indicated for adult patients (wrist circumference 15-21 cm) for non-invasive oscillometric blood pressure measurement. Contraindicated for pregnant patients and patients with arrhythmias (e.g., atrial fibrillation).
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 14, 2026
Shenzhen Urion Technology Co., Ltd.
Fraser Liu
Regulation Assistant
Floor 4-6th Of Bldg. D, Jiale Science & Technology
Industrial Zone, #3, Chuang Wei Rd., Heshuikou Community
Shenzhen, Guangdong 518106
China
Re: K254048
Trade/Device Name: Wrist Watch Electronic Blood Pressure Monitor (U16H, U16L, U16P, U16W, U19M, U19R, U19S, and U19T)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: May 3, 2026
Received: May 4, 2026
Dear Fraser Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K254048 - Fraser Liu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}
K254048 - Fraser Liu
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**STEPHEN C. BROWNING -S**
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254048 | |
| Device Name Wrist Watch Electronic Blood Pressure Monitor (U16H, U16L, U16P, U16W, U19M, U19R, U19S, and U19T) | |
| Indications for Use (Describe) The Wrist Watch Electronic Blood Pressure Monitor uses the oscillometric method to measure diastolic blood pressure, systolic blood pressure. It is intended for use in adults with a wrist circumference between 15-21 cm who do not have arrhythmias (such as atrial fibrillation) and who are not pregnant. The values are for reference purposes and do not constitute a diagnosis. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
Wrist Watch Electronic Blood Pressure Monitor
K254048
# 510(k) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
1. Submitter: Shenzhen Urion Technology Co., Ltd.
Floor 4-6th of Building D, Jiale Science&Technology Industrial Zone, No.3, ChuangWei Road, Heshuikou Community, MaTian Street, GuangMing New District, 518106 Shenzhen, PEOPLE'S REPUBLIC OF CHINA
Tel: +86-755-29231308
Fax: +86-755-27493959
Contact Person: Xiaona Guo
Prepare date: 2025-09-10
2. Device name and classification:
Device Name: Wrist Watch Electronic Blood Pressure Monitor
Model: U16H, U16L, U16P, U16W, U19M, U19R, U19S, and U19T
Classification Name: Noninvasive Blood Pressure Measurement System
21 CFR 870.1130 System, measurement, blood-pressure, non-invasive
Product code: DXN
Regulatory Class: Class II
3. Predicate Device(s): Wrist Electronic Blood Pressure Monitor
K231367/Shenzhen Finicare Co., Ltd.
4. Device Description: This device is a digital self-monitor intended for use in measuring blood pressure in adult population with wrist circumference ranging from 15.0 cm to 21.0 cm. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure of adult person at wrist within its claimed range and accuracy via the oscillometric technique.
The Wrist Watch Electronic Blood Pressure Monitor consists of a main unit, watch strap(including an airbag), and a charging cable.
Page 1 of 6
{6}
Wrist Watch Electronic Blood Pressure Monitor
K254048
## 5. Indications for Use:
The Wrist Watch Electronic Blood Pressure Monitor uses the oscillometric method to measure diastolic blood pressure, systolic blood pressure. It is intended for use in adults with a wrist circumference between 15-21 cm who do not have arrhythmias (such as atrial fibrillation) and who are not pregnant. The values are for reference purposes and do not constitute a diagnosis.
## 6. Predicate Device Comparison
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device.
The differences between the subject device and predicate device include physical specifications, and color. All above differences do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
## 7. Effectiveness and Safety Considerations:
### Clinical test:
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intended use.
The clinical trials for the Wrist Watch Electronic Blood Pressure Monitor were performed according the standard of ISO 81060-2:2018+AMD1:2020+AMD2:2024, Non-Invasive Sphygmomanometers -Part 2: Clinical Validation of Automated Measurement Type, and relevant volunteers were collected to conduct actual clinical trial of blood pressure measurement.
There were 103 subjects been selected to participate in the trial, and Auscultation was applied as gold standard with the qualified calibrated mercurial sphygmomanometer used as control group for comparison with the subject device.
The results demonstrated that the clinical performance of the proposed device met the requirements of ISO 81060-2:2018+AMD1:2020+AMD2:2024. The mean deviation for systolic blood pressure was 1.3 mmHg with a standard deviation of 6.2 mmHg, and the mean deviation for diastolic blood pressure was -1.5 mmHg with a standard deviation of 4.9 mmHg. These results are well within the standard's acceptance criteria, which require a mean deviation ≤ ±5.0 mmHg and a standard deviation ≤ 8.0 mmHg. Consequently, the subject device is confirmed to have clinical accuracy equivalent to the reference manual auscultation method.
### Non-clinical test:
The following safety standards are conducted on the subject device:
The proposed devices were tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Wrist Watch Electronic Blood Pressure Monitor – Premarket Notification [510(k)] Submission issued on March 5, 2004
The test results demonstrated that the proposed device complies with the following standards:
ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R) 2012(Cons. Text) [Incl. AMD2:2021] - Medical electrical equipment - Part 1: General
Page 2 of 6
{7}
Wrist Watch Electronic Blood Pressure Monitor
K254048
requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
IEC 60601-1-2:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
IEC 60601-1-11:2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 80601-2-30:2018, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non Invasive Sphygmomanometers.
EN 300328:2019 Electromagnetic compatibility and Radio spectrum Matters (ERM);Wideband transmission systems; Data transmission equipment operating in the 2,4 GHz ISM band and using wide band modulation techniques.
EN 301489-1:2019 ElectroMagnetic Compatibility (EMC)standard for radio equipment and services;Part 1:Common technical requirements.
EN 301489-17:2020 Electromagnetic compatibility and Radio spectrum Matters (ERM);ElectroMagnetic Compatibility (EMC) standard for radio equipment; Part 17:Specific conditions or Broadband Data Transmission Systems.
EN 62479:2010 Assessment of the compliance of low power electronic and electrical equipment with the basic restrictions related to human exposure to electromagnetic fields (10 MHZ to 300 GHZ)
IEC 62304:2015 standard and FDA Guidance for the Content of Pre-MarketSubmission for Software Contained in Medical Devices standard.
Table 1 Performance Comparison
| Device | Proposed Device | Predicate Device K231367 | Result |
| --- | --- | --- | --- |
| Max Cuff pressure | 300 mmHg | 300 mmHg | Same |
| BP Range | 0-299 mmHg | 0-299 mmHg | Same |
| BP Accuracy | ±3 mmHg | ±3 mmHg | Same |
| Irregular heartbeat detection | More than ±25% to the mean interval of pulse intervals | More than ±25% to the mean interval of pulse intervals | Same |
| Inflation Method | Automatic inflation by pump | Automatic inflation by pump | Same |
Page 3 of 6
{8}
Wrist Watch Electronic Blood Pressure Monitor
K254048
| Deflation Method | Automatic rapid deflation | Automatic rapid deflation | Same |
| --- | --- | --- | --- |
| Memory Size | U16H:4x50 set of data U16W:1x25 set of data U16P:2x25 set of data U16L:3x25 set of data U19M:4x25 set of data U19R:1x50 set of data U19S:2x50 set of data U19T:3x50 set of data | 2x120 set of data | Similar(1) |
| Operation Condition | Temperature: 5°C ~ 40°C Relative Humidity: 15% RH ~ 85% RH (no condensation) Atmospheric Pressure: 80.0 kPa ~ 106.0 kPa | Temperature: 10~40°C Humidity: 15~85%RH Atmospheric pressure: 70 -106kPa | Similar(2) |
| Storage Condition | Temperature: -20°C ~ 55°C Humidity: 10% RH ~ 93% RH (no condensation) Atmospheric Pressure: 70.0 kPa ~ 106.0 kPa | Temperature: -20~60 °C Humidity: 10 to 95% RH | Similar(3) |
| Power Supply | lithium battery DC 3.85V | 2 AAA alkaline batteries, DC3V | Similar(4) |
| Performance Standard | Comply with IEC 80601-2-30 | Comply with IEC 80601-2-30 | Same |
| Blood pressure classification | Accordance with the AHA hypertension guidelines | / | Slight different(1) |
## Similarities
(1) Memory Size: Both the subject device and the predicate device have been defined the memory size, The difference is some modes for the subject device and the predicate device has only one memory size. The subject device comply with the requirement of standard IEC
Page 4 of 6
{9}
Wrist Watch Electronic Blood Pressure Monitor
K254048
60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 80601-2-30. This difference will not cause any safety or performance issue.
(2) Operation Condition: Both the subject device and the predicate device have been defined the similar operation condition. The subject device comply with the requirement of standard IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 80601-2-30. This difference will not cause any safety or performance issue.
(3) Storage Condition: Both the subject device and the predicate device have been defined the similar storage condition. The subject device comply with the requirement of standard ASTM D4169-23. This difference will not cause any safety or performance issue.
(4) Power Supply: Both the subject device and the predicate device are powered by batteries. The subject device comply with the requirement of standard IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 80601-2-30. This difference will not cause any safety or performance issue.
## Difference
Blood pressure classification of subject device is determined accordance with the AHA hypertension guidelines. But the predicate device have no this classification.
Table 2 Performance testing
| Item | Proposed device | Result |
| --- | --- | --- |
| IEC 60601-1:2005/AMD2: 2020 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| IEC 60601-1-2:2014 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| IEC 60601-1-11:2015/ AMD1:2020 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| IEC 80601-2-30:2019 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| IEC 62368-1:2024 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| FCC CFR Title 47 Part 15 Subpart C § 15.247 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| ANSI/USEMCSC C63.27 2021 AAMI TIR69:2017(R2020) | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| ISO81060-2:2018+AMD1:2020+AM D2:2024 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| EN 300328:2019 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| EN 301489-17:2020 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| EN 62479:2010 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
Page 5 of 6
{10}
Wrist Watch Electronic Blood Pressure Monitor
K254048
| IEC 60601-1-11:2015 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
| --- | --- | --- |
| IEC 62133-2:2017 IEC 62133-2:2017/AMD1:2021 | Wrist Watch Electronic Blood Pressure Monitor | Pass |
Table 3 Biocompatibility testing
| Item | Proposed device | Result |
| --- | --- | --- |
| Cytotoxicity | Under the conditions of the study, the device is noncytotoxic. | Pass |
| Irritation | Under the conditions of the study, the device is nonirritating. | Pass |
| Sensitization | Under the conditions of the study, the device is nonsensitizing | Pass |
## 8. Substantially Equivalent Determination
Verification and validation testing was conducted on the Wrist Watch Electronic Blood Pressure Monitor. This premarket notification submission demonstrates that Wrist Watch Electronic Blood Pressure Monitor is substantially equivalent to the predicate devices.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.