This blood pressure monitor (Model: FC-BP200, FC-BP201, FC-BP210, FC-BP211, FC-BP220, FC-BP221) is intended to be used to measure blood pressure (systolic and diastolic) and heart rate from the wrist by using the oscillometric method. The device detects the appearance of irregular heartbeats during measurement and gives a symbol with reading. The device is intended for using in only adult population, not applied to the other populations such as neonatal baby. It can't be used while the wrist (arm) has bleeding or wound to avoid the blood flowing from the wound in pressurizing. The wrist circumference is limited to 13.5cm - 21.5cm.
Device Story
Wrist-worn automatic blood pressure monitor; uses oscillometric method to detect brachial artery movement; converts signals to digital systolic/diastolic blood pressure and pulse rate readings. Operated by adult users in home settings. Features include irregular heartbeat detection (symbol display if >25% deviation from mean), WHO-based blood pressure classification, voice broadcast, and memory storage (2x120 sets). Models FC-BP200, FC-BP210, FC-BP220, and FC-BP221 include Bluetooth LE 4.2 for data transmission to paired devices. Safety features include 300 mmHg pressure limit with auto-exhaust and 2-minute auto-shutdown. Healthcare providers use stored data to monitor patient blood pressure trends; benefits include convenient, non-invasive home monitoring.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018. Study included 130 subjects. Auscultation using a calibrated mercurial sphygmomanometer served as the gold standard control. Results confirmed device accuracy within ±5 mmHg, meeting declared performance specifications.
Technological Characteristics
Oscillometric measurement; powered by 2x AAA batteries (DC3V); wrist cuff (brushed fabric, ISO 10993 compliant); Bluetooth LE 4.2 connectivity (select models); LCD display. Safety standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30. Software compliant with IEC 62304.
Indications for Use
Indicated for adult population for noninvasive measurement of systolic/diastolic blood pressure and heart rate at the wrist via oscillometric method. Contraindicated for neonates and individuals with wrist bleeding or wounds. Wrist circumference range: 13.5cm - 21.5cm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
September 18, 2023
Shenzhen Finicare Co., Ltd. % Boyle Wang, Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161,East Lujiazui Rd.,Pudong Shanghai, Shanghai 200120 China
Re: K231367
Trade/Device Name: Wrist Electronic Blood Pressure Monitor (Model FC-BP200, FC-BP201, FC-BP210,FC-BP211, FC-BP220, FC-BP221) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 21, 2023 Received: August 25, 2023
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231367
### Device Name
Wrist Electronic Blood Pressure Monitor (Model:FC-BP200, FC-BP210,FC-BP211,FC-BP211,FC-BP221)
### Indications for Use (Describe)
This blood pressure monitor (Model: FC-BP20), FC-BP201, FC-BP211, FC-BP220, FC-BP221) is intended to be used to measure blood pressure (systolic and diastolic) and heart rate from the wrist by using the oscillometric method. The device detects the appearance of irregular heartbeats during measurement and gives a symbol with reading. The device is intended for using in only adult population, not applied to the other populations such as neonatal baby. It can't be used while the wrist (arm) has bleeding or wound to avoid the blood flowing from the wound in pressurizing. The wrist circumference is limited to 13.5cm - 21.5cm.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <span style="white-space: nowrap;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="white-space: nowrap;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# K231367
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
# 1.0 Submitter's Information
| Name: | Shenzhen Finicare Co., Ltd. |
|----------|-------------------------------------------------------------------------------------------------------------------|
| Address: | 201, No.50, the 3rd Industrial Park, Houting Community, Shajing<br>Street, Bao'an District, Shenzhen 518104 China |
| Tel: | 86-755-23013503 |
| Contact: | Chao Li |
# Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------|-------------------------------------------------------------------------|
| Name: | Shanghai Truthful Information Technology Co., Ltd. |
| Address: | Room 1801, No. 161 East Lu Jiazui Rd., Pudong,Shanghai,<br>200120 China |
| Tel: | +86-21-50313932 |
| Email: | Info@truthful.com.cn |
Date of Preparation: Aug.23,2023
# 2.0 Device Information
| Trade name: | Wrist Electronic Blood Pressure Monitor |
|----------------------|---------------------------------------------------------------|
| Common name: | Noninvasive Blood Pressure Measurement System |
| Classification name: | Noninvasive Blood Pressure Measurement System |
| Model(s): | FC-BP200, FC-BP201, FC-BP210,<br>FC-BP211, FC-BP220, FC-BP221 |
| Production code: | DXN |
| Regulation number: | 21 CFR 870.1130 |
| Classification: | Class II |
| Panel: | Cardiovascular |
# 3.0 Predicate Device Information
| Manufacturer: | Shenzhen Kingyield Technology Co., Ltd. |
|--------------------|-----------------------------------------|
| Trade/Device name: | Bluetooth Blood Pressure Monitor, BPW1 |
{5}------------------------------------------------
#### 510(k) number: K182018
## 4.0 Indication for Use Statement
This blood pressure monitor (Model: FC-BP200, FC-BP201, FC-BP210, FC-BP211, FC-BP220, FC-BP221) is intended to be used to measure blood pressure (systolic and diastolic) and heart rate from the wrist by using the oscillometric method. The device detects the appearance of irregular heartbeats during measurement and gives a symbol with reading. The device is intended for using in only adult population, not applied to the other populations such as neonatal baby. It can't be used while the wrist (arm) has bleeding or wound to avoid the blood flowing from the wound in pressurizing. The wrist circumference is limited to 13.5cm - 21.5cm.
## 5.0 Device Description
The subject device, Wrist Electronic Blood Pressure Monitor, is an automatic noninvasive blood pressure monitor which can be driven by AAA alkaline battery, DC3V. It uses an inflatable cuff which is wrapped around the patient's wrist to measure the systolic and diastolic blood pressure as well as the pulse rate of adult at household, not for neonate or pregnancy.
The unit uses the oscillometric method of blood pressure measurement. It means the unit detects the movement of your blood through your brachial artery, and converts your blood pressure into a digital reading. The unit is simple to use because a stethoscope is not needed while using an oscillometric monitor.
The unit stores the blood pressure and pulse rate in the memory after completing a measurement each time. 2x120 sets of measurement values can be stored automatically. The unit also calculates an average reading based on the values of the latest 3 times measurement.
This blood pressure monitor has the function of blood pressure classification according to WHO recommendation, which is convenient for you to judge whether your blood pressure is normal or not.
This blood pressure monitor has voice broadcast function. During measurement and recall the memory, there will be voice operation tips.
The device detects an Irregular Heart Beat (IHB) (a Heartbeat that is more than 25% slower or 25% faster from the average Heartbeat) two or more times during the measurement, the irregular heartbeat Symbol will appear on the display with the measurement values.
The device features a built-in "Bluetooth Data Transmission" function, which enables
{6}------------------------------------------------
the device automatically transmit measuring results through Bluetooth LE4.2 to paired Bluetooth-enabled device (FC-BP200, FC-BP210, FC-BP220 and FC-BP221 applied).
There is a maximum pressure safety setting at 300 mmHg, when the pressure is more than 300mmHg, the device will exhaust fast automatically.
No operation for 2 minute the device will shut down automatically.
The device includes model FC-BP200, FC-BP201, FC-BP210, FC-BP211, FC-BP220, FC-BP221. Six models are identical in terms of software design, cuff type and measurement range. The schematic circuit diagram are identical in all models except for model FC-BP220. Six models PCB layout are different because of different appearance (such as structure, buttons layout, etc.) and function.
# 6.0 Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 &
A2:2010/(R)2012(Cons. Text) [Incl. AMD2:2021] - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-
1:2005, MOD) [Including Amendment 2 (2021)]
IEC 60601-1-2:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
IEC 60601-1-11:2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 80601-2-30:2018, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers.
EN 300328:2019 Electromagnetic compatibility and Radio spectrum Matters (ERM); Wideband transmission systems; Data transmission equipment operating in the 2,4 GHz ISM band and using wide band modulation techniques.
301489-1:2019 ElectroMagnetic Compatibility (EMC)standard for radio EN equipment and services;Part 1:Common technical requirements.
EN 301489-17:2020 Electromagnetic compatibility and Radio spectrum Matters
(ERM);ElectroMagnetic Compatibility (EMC) standard for radio equipment; Part 17: Specific conditions or Broadband Data Transmission Systems.
EN 62479:2010 Assessment of the compliance of low power electronic and electrical equipment with the basic restrictions related to human exposure to electromagnetic
{7}------------------------------------------------
### fields (10 MHZ to 300 GHZ )
IEC 62304:2015 standard and FDA Guidance for the Content of Pre-Market Submission for Software Contained in Medical Devices standard.
# 7.0 Clinical Test Conclusion
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intended use.
The clinical trials for the Wrist Electronic Blood Pressure Monitor were performed according the standard of ISO 81060-2:2018, Non-Invasive Sphygmomanometers -Part 2: Clinical Validation of Automated Measurement Type, and relevant volunteers were collected to conduct actual clinical trial of blood pressure measurement.
There were 130 subjects been selected to participate in the trial, and Auscultation applied as gold standard with the qualified calibrated mercurial was was was and sphygmomanometer used as control group for comparison with the subject device. The results shown that the accuracy of proposed device meet the requirements of ISO 81060-2:2018 within the ±5mmHg.And the subject device comply with the standard requirements and the accuracy the manufacture declared.
# 8.0 Technological Characteristic Comparison Table
{8}------------------------------------------------
| Item | Subject Device<br>K231367 | Predicate Device<br>K182018 | Remark |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Manufacturer | Shenzhen Finicare Co., Ltd. | Shenzhen Kingyield<br>technology Co., Ltd | / |
| Product Name | Wrist Electronic Blood Pressure<br>Monitor<br>FC-BP200, FC-BP201, FC-<br>BP210,FC-BP211, FC-BP220,<br>FC-BP221 | Bluetooth blood pressure<br>monitor, BPW1 | / |
| Product Code | DXN | DXN | Same |
| Regulation No. | 21 CFR 870.1130 | 21 CFR 870.1130 | Same |
| Class | II | II | Same |
| Intended<br>Use/Indication<br>for Use | This blood pressure monitor<br>(Model: FC-BP200, FC-BP201,<br>FC-BP210, FC-BP211, FC-<br>BP220, FC-BP221) is intended to<br>be used to measure blood<br>pressure (systolic and diastolic)<br>and heart rate from the wrist by<br>using the oscillometric method.<br>The device detects the<br>appearance of irregular<br>heartbeats during measurement<br>and gives a symbol with reading.<br>The device is intended for using<br>in only adult population, not<br>applied to the other populations<br>such as neonatal baby. It can't be<br>used while the wrist (arm) has<br>bleeding or wound to avoid the<br>blood flowing from the wound in<br>pressurizing. The wrist<br>circumference is limited to 13.5cm<br>- 21.5cm. | This blood pressure monitor is<br>intended to be used to<br>measure blood pressure<br>(systolic and diastolic) and<br>heart rate from the wrist by<br>using the oscillometric method.<br>The device detects the<br>appearance of irregular<br>heartbeats during<br>measurement and gives a<br>symbol with reading. The<br>device is intended for using in<br>only adult population, not<br>applied to the other<br>populations such as neonatal<br>baby. It can't be used while the<br>wrist (arm) has bleeding or<br>wound to avoid the blood<br>flowing from the wound in<br>pressurizing. The wrist<br>circumference is limited to<br>13.5cm - 21.5cm | Same |
| Application Site | Wrist | Wrist | Same |
| Wrist<br>Circumference | 13.5cm - 21.5cm | 13.5cm - 21.5cm | Same |
| Patients<br>Contacting<br>Materials | Patient contact materials of the<br>cuff: Brushed Fabric<br>According to ISO-10993 | Zinc alloy for the case of the<br>device and ABS for Button.<br>Biocompatible materials are<br>used for the applied parts<br>(Bladder + wristband)<br>According to ISO-10993 | Different |
| Patient<br>Population | Adult | Adult | Same |
| Measurements<br>Item | SYS,DYS,Pulse | SYS,DYS,Pulse | Same |
| Display<br> | LCD Digital Display | LCD | Same |
| Item | Subject Device | Predicate Device<br>K182018 | Remark |
| Max Cuff<br>pressure | 300 mmHg | 300mmHg | Same |
| BP Range | 0-299 mmHg | 0 ~ 299 mmHg | Same |
| BP Accuracy | ±3 mmHg | ±3 mmHg | Same |
| PR Range | 40-180 beats/min | 40 ~ 180 beats/min | Same |
| Pulse Accuracy | ±5% of reading<br>value | ±5% of reading<br>value | Same |
| Irregular<br>heartbeat<br>detection | More than ±25% to the<br>mean interval of pulse<br>intervals. | More than ±25% to the<br>mean interval of pulse<br>intervals. | Same |
| Inflation<br>Method | Automatic inflation by pump | By air pump | Same |
| Deflation<br>Method | Automatic rapid deflation | Not apply | Different |
| Memory Size | 2x120 set of data | 30 set | Different |
| Operation<br>Condition | 10~40°C<br>Humidity: 15~85%RH<br>Atmospheric pressure: 70 -<br>106kPa | 10~40℃,<br>Humidity:15~90%RH<br>Atmospheric: 106kPa~80kPa | Different |
| Storage<br>Condition | -20~60 °C<br>Humidity: 10 to 95% RH | -20°C~ 55°C,<br>0~95%RH (noncondensing ) | Different |
| Power Supply | 2 AAA alkaline batteries, DC3V | Li-ion Rechargeable battery<br>with 250 charging cycles | Different |
| Performance<br>Standard | Comply with IEC 80601-2-30 | Comply with IEC 80601-2-30 | Same |
# Table1-General Comparison
Analysis:
{9}------------------------------------------------
The material of the wrist cuff of the subject device is different with that of the predicate device. Biocompatibility testing was carried out and the test results shown the materials of both devices are complied with the requirements of ISO 10993-1.
Table 2 Performance Comparison
# Analysis:
The subject and predicate device have same general intended use, similar design features and performance specifications.
The differences in Memory Size do not raise different questions of safety or effectiveness since the device complied with the requirement of IEC 80601-2-30. The differences in Operation Condition and Storage Condition between both devices are insignificant in terms of safety and effectiveness. Also, the little difference in power supply according to the Electrical Safety and EMC test result, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device. So we think the subject device is substantially equivalent to the predicate device.
# Table 3 Safety Comparison
| Item | Proposed Device | Predicate Device<br>K182018 | Remark |
|------|-----------------|-----------------------------|--------|
|------|-----------------|-----------------------------|--------|
{10}------------------------------------------------
| Electrical Safety | Comply with ANSI AAMI<br>ES60601-1 | Comply with IEC 60601-1 | Same |
|-------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------|
| Home Use | Comply with IEC 60601-1-<br>11 | Comply with IEC 60601-1-11 | Same |
| EMC | Comply with IEC 60601-1-<br>2 | Comply with IEC 60601-1-2 | Same |
| FCC conformity | FCC 47 part 15 subject B<br>FCC 47 CFR Part 15,<br>Subpart C | FCC 47 CFR part 15,Subject B<br>FCC 47 CFR Part 15, Subpart C<br>FCC 47 CFR Part 1.1307<br>FCC 47 CFR Part 2.1093 | Same |
| ERM conformity | EN 301489-1:2017;<br>EN301489-17:2017 | EN 301489-1:2017;<br>EN 301489-17:2017 | Same |
| RF conformity | EN300328:2016 | EN300328:2016 | Same |
| Health | EN62479:2010 | EN62479:2010 | Same |
| Biocompatibility | Comply with ISO 10993-1,<br>FDA Guidance | Comply with ISO 10993-1, FDA<br>Guidance | Same |
# 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.