The X-ray system creates data for digital exposures in the maxillofacial area and in sub-areas for dentistry and pediatric dentistry and for hard tissue diagnostics within ENT medicine.
Device Story
Primevision 3D is a digital panoramic and cone beam computed tomography (CBCT) X-ray system; captures maxillofacial images using a single imaging detector for both 2D and 3D modes. Device operates via hardware interface and cloud-based acquisition software accessed through standard internet browsers. System performs filtered back projection (FBP) for 3D reconstruction; includes automated motion and spine artifact compensation features. Used in dental and ENT clinical settings by trained professionals. Output consists of digital cross-sectional and panoramic images for hard tissue diagnostics. Cloud-based workflow replaces traditional on-premise software for program selection and acquisition management. Clinical benefits include improved image homogeneity, diagnostic interpretability, and workflow efficiency through artifact compensation and simplified HMI.
Clinical Evidence
Clinical image validation report provided. Expert clinical evaluators assessed images using a Likert scale to verify performance of motion and spine artifact compensation. Results demonstrated improved 3D volume quality, panoramic homogeneity, and diagnostic interpretability. Bench testing confirmed compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-63, 61223-3-4, and 61223-3-7. Imaging performance metrics (spatial resolution, homogeneity, geometric accuracy) were equal to or better than the predicate.
Technological Characteristics
Digital CBCT/panoramic X-ray system. Single sensor for 2D/3D. FBP reconstruction. Cloud-based acquisition software. Hardware includes simplified HMI and hand switch. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-63 (dental X-ray), and IEC 61223-3-7 (CBCT performance).
Indications for Use
Indicated for digital X-ray imaging of the maxillofacial area, including dentistry, pediatric dentistry, and hard tissue diagnostics in ENT medicine.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 5, 2026
Dentsply Sirona
% Deepthi Paknikar
Sr Manager Regulatory Affairs
221 W. Philadelphia St.
Suite 60w
YORK, PA 17401
Re: K253959
Trade/Device Name: Primevision 3D
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed tomography x-ray system
Regulatory Class: Class II
Product Code: OAS
Dated: December 10, 2025
Received: December 10, 2025
Dear Deepthi Paknikar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253959 - Deepthi Paknikar
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253959 - Deepthi Paknikar
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253959 | |
| Device Name Primevision 3D | |
| Indications for Use (Describe) | |
| The X-ray system creates data for digital exposures in the maxillofacial area and in sub-areas for dentistry and pediatric dentistry and for hard tissue diagnostics within ENT medicine. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
510(k) #: K253959
Applicant Name: Dentsply Sirona Inc
Applicant Address: 221 West Philadelphia St. York PA 17401 United States
Applicant Contact Telephone: 630-201-1612
Applicant Contact: Dr. Deepthi Paknikar
Applicant Contact Email: Deepthi.Paknikar@dentsplysirona.com
Device Trade Name: Primevision 3D
Common Name: Computed tomography x-ray system
Classification Name: Computed tomography x-ray system
Regulation Number: 21 CFR 892.1750
Product Code(s): OAS
Device Class: 2
Predicate Device: Axeos (K201140)
Common Name: Computed tomography x-ray system
Classification Name: Computed tomography x-ray system
Regulation Number: 21 CFR 892.1750
Product Code(s): OAS
Device Class: 2
Reference Device(s): Planmeca Viso (K230985), Planmeca Viso (K181576), Trophy CS 9600 (K181136), Planmeca ProMax 3D Max (K160506), Orthopantomograph™ OP 3D LX (K230505)
Common Name: Computed tomography x-ray system
Classification Name: Computed tomography x-ray system
Regulation Number: 21 CFR 892.1750
Product Code(s): OAS
Device Class: 2
1. Device Description Summary
Primevision 3D is a digital panoramic and cone beam computed tomography (CBCT) x-ray device. A Computed tomography x-ray system (21 CFR 892.1750) is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. The subject Primevision 3D uses just one imaging detector to collect data for both 2D and 3D modes of operation. Additionally, the Primevision 3D does not offer a cephalometric mode.
1.1 Intended Use/Indications for Use
The X-ray system creates data for digital exposures in the maxillofacial area and in sub-areas for dentistry and pediatric dentistry and for hard tissue diagnostics within ENT medicine.
1.2 Indications for Use Comparison
The subject device is substantially equivalent to the predicate device in terms of intended use, fundamental design, and core functionality. Both devices are intended for dental imaging
{5}
applications and support panoramic and 3D imaging. Although the subject device does not include cephalometric imaging, this omission does not alter the intended use or raise new questions of safety or effectiveness. Performance testing and documentation provided in this submission demonstrate that the subject device meets applicable standards and performs as safely and effectively as the predicate.
## 2. Technological Comparison
The subject device introduces several technological updates compared to the predicate. It utilizes a single sensor for both 2D and 3D imaging, whereas the predicate used separate sensors. The subject device also incorporates cloud-based acquisition software accessed via a standard internet browser, replacing the predicate's on-premise software. Additional enhancements include motion and spine artifact compensation, and a revised hardware interface with simplified controls. These differences reflect advancements in technology and workflow efficiency, and do not impact the device's safety or effectiveness.
## 3. Non-Clinical and/or Clinical Tests
Testing to verify the performance requirements of the subject device was conducted and included in this premarket notification. The results of the performance testing support substantial equivalence. Tests included in this premarket notification verify the conformity of the proposed device with the requirements of:
- IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020: Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60825-1:2014: Safety of laser products – Part 1: Equipment classification and requirements.
- IEC 60601-1-3:2008, IEC 60601-1-3:2008/AMD1:2013, IEC 60601-1-3:2008/AMD2:2021 for use in conjunction with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020: Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment.
- IEC 62366-1:2015/AMD1:2020: Medical devices – Application of usability engineering to medical devices.
- IEC 62304:2006 (First Edition) + A1:2015 (or IEC 62304:2015 CSV): Medical device software – Software life cycle processes.
- IEC 60601-2-63:2012; IEC60601-2-63:2012/AMD1:2017, IEC 60601-2-63:2012/AMD2:2021 for use in conjunction with IEC 60601-1:2005: Medical electrical equipment - Part 2-63: Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment
{6}
- IEC 60601-1-6:2010, IEC 60601-1-6:2010/AMD1:2013, IEC 60601-1-6:2010/AMD2:2020 for use in conjunction with IEC 62366-1:2015: Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability, (with third edition of 60601-1)
- IEC 61223-3-4:2000: Evaluation and routine testing in medical imaging departments - Part 3-4: Acceptance tests - Imaging performance of dental X-ray equipment
- IEC 61223-3-7:2021: Evaluation and routine testing in medical imaging departments - Part 3-7: Acceptance and constancy tests - Imaging performance of X-ray equipment for dental cone beam computed tomography
Verification activities for confirmation of the image quality of the proposed device have been performed. The results of the image quality review have demonstrated that the device is substantially equivalent to the predicate device.
The subject system Primevision 3D has fundamentally the same hardware as the predicate device, without the inclusion of a Cephalometric imaging feature for the subject device. The remote control has been updated for the subject device to include a hand switch, and the new device HMI (referred to as the "easypad" or "multipad" for prior devices) has a different appearance, yet the same core functionality, as the predicate device. The subject device HMI contains less functionality than the predicate, because some of the functionality has been moved to cloud-based software (e.g. program selection) for the subject device. The EMC testing generated substantially equivalent results results as the predicate. Bench testing for image quality has been performed and the results indicate substantial equivalence to the predicate Axeos, for both 2D and 3D imaging modes (See the tabulated values below for details).
Animal testing was not necessary for this premarket submission. Clinical testing was provided in this submission, and extraoral images were submitted along with an image evaluation report. The Clinical Testing reports indicate device effectiveness.
Software Verification and Validation testing was provided as part of this submission. The predicate device had the option of cephalometric exposures along with panoramic and 3D exposure; the subject device does not have the option of cephalometric exposure. For the subject device, data processing is distributed on the device and the cloud. This is not the case with the predicate; the predicate does not have any cloud-based software component. The 3D reconstruction method (CBCT mode of operation) is the same as the predicate device and is filtered back projection (FBP). The fundamental steps for the user to take an image remain the same, the difference is that the user can use their internet browser with the subject device compared to a downloadable on-premises software with the predicate. All software functions are strongly based on the predicate device.
All cybersecurity testing has been successfully performed per FDA requirements and guidance.
{7}
The subject device software includes new features not present in the predicate device, specifically motion artifact compensation and spine artifact compensation. These were evaluated through comprehensive performance testing, including expert clinical evaluation and software verification testing. The metal artifact compensation feature remains the same as in the predicate device and has not been changed. For the motion and spine compensation features, the results of the clinical evaluation testing met prespecified success criteria via a Likert scale-based evaluation. The expert clinical evaluators reported improvements in 3D volume image quality, panoramic image homogeneity, and diagnostic interpretability when reviewing images with the compensation applied.
In summary, a comprehensive set of performance-supporting documents was included in this submission in addition to the software verification and validation reports provided. These materials include information outlined in the FDA guidance for Solid State X-ray Imaging Devices, an image quality assessment report, qualified expert evaluations conducted in accordance with the SSXI FDA guidance, a clinical image validation report, and an overview comparing the subject device's sensor performance to legally marketed predicate devices.
As mentioned above, the imaging performance of the Primevision 3D, subject device has been tested against FDA Recognized Consensus Standards. The standards used are the same with the predicate device Axeos. An additional table has been provided below for information on imaging performance results of the subject and predicate devices:
| | Imaging Performance | CCBCT R1 | Axeos | Comparison |
| --- | --- | --- | --- | --- |
| Panoramic (IEC 61223-3-4) | Low Contrast | At least 1mm low contrast elements | At least 1mm low contrast elements | Same |
| | Spatial Resolution | At least 2.5 lp/mm | At least 2.5 lp/mm | Same |
| CBCT (IEC 61223-3-7) | Spatial Resolution | ≥1 lp/mm (MTF10) | ≥1 lp/mm (MTF10) | Same |
| | Homogeneity | ≥5 | ≥5 | Same |
| | Geometric accuracy | 1% | 1% | Same |
| | Acceptance Indicator | ≥100 / (mGycm2) | ≥100 / (mGycm2) | Same |
| CBCT (1020.33) | Spatial Resolution | ≥2.6 lp/mm (MTF10) ≥1.3 lp/mm (MTF50) | ≥1.0 lp/mm (MTF10) ≥0.7 lp/mm (MTF50) | Improved |
| | Noise | ≤2.1 | ≤3 | Improved |
# 4. Conclusion
The information included in this premarket notification supports the substantial equivalence of the subject device with the identified predicate. The subject device has the identical intended use as the legally marketed predicate device. The subject device also has the identical indications for use and incorporates the same fundamental technology as the predicate device.
{8}
Performance data are included in this premarket notification to demonstrate the performance of the subject device against its design, functional, and safety requirements. The results of the testing included in this premarket notification support a determination of substantial equivalence.
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.