K230505 · Palodex Group OY · OAS · Jun 9, 2023 · Radiology
Device Facts
Record ID
K230505
Device Name
Orthopantomograph™ OP 3D LX
Applicant
Palodex Group OY
Product Code
OAS · Radiology
Decision Date
Jun 9, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ORTHOPANTOMOGRAPH™ OP 3D LX is an X-ray device that is intended to be used for imaging of adult and pediatric patients. The device can be configured to take panoramic, cephalometric or 3D images of cranio-maxillofacial complex including the ear, nose and throat (ENT) and airway regions, and cervical spine. The device can be configured to take carpus images. The device is operated and used by qualified healthcare professionals.
Device Story
ORTHOPANTOMOGRAPH™ OP 3D LX is a dental X-ray system for 2D and 3D imaging of the cranio-maxillofacial complex. It captures panoramic, cephalometric, and cone beam CT (CBCT) images. In panoramic/cephalometric modes, a rotating unit scans the patient with a narrow X-ray beam. In 3D CBCT mode, the unit rotates around the head using a rectangular beam to record 2D projections, which are then reconstructed into 3D data. The device is operated by qualified healthcare professionals in a clinical setting. Output consists of digital image data used for diagnosis and treatment planning (e.g., implants, orthodontics, prosthetics). The device aids in early diagnosis and less invasive treatment. It supports standing and wheelchair-bound patients.
Clinical Evidence
Clinical image review performed to support substantial equivalence. No specific clinical trial metrics (e.g., sensitivity/specificity) provided. Bench testing conducted per IEC/ISO standards (e.g., IEC 60601-1, IEC 60601-2-63, IEC 61223-3-7) confirmed imaging performance and safety.
Technological Characteristics
X-ray system with TFT flat panel detector (CsI scintillator). Imaging modes: Panoramic, TMJ, BW, 3D (CBCT), Cephalometric. 3D reconstruction via Filtered Back Projection (FBP). Materials: Cycoloy C2800, Grivory GC-4H, Polycasa PETG, Grilamid TR 90+, PLA MP BIO. Standards: IEC 60601-1, IEC 60601-2-63, ISO 10993-1, IEC 61223-3-7. Connectivity: Digital dental workflow integration.
Indications for Use
Indicated for imaging of adult and pediatric patients to capture panoramic, cephalometric, or 3D images of the cranio-maxillofacial complex, including ear, nose, throat (ENT), airway regions, cervical spine, and carpus.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
June 9, 2023
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Palodex Group OY % Frank Ray Director Regulatory Affairs Nahkelantie 160 Tuusula, Etela-Karjala 04300 FINLAND
Re: K230505
Trade/Device Name: Orthopantomograph™ OP 3D LX Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: May 12, 2023 Received: May 12, 2023
Dear Frank Ray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K230505
Device Name ORTHOPANTOMOGRAPH™ OP 3D LX
### Indications for Use (Describe)
ORTHOPANTOMOGRAPH™ OP 3D LX is an X-ray device that is intended to be used for imaging of adult and pediatric patients. The device can be configured to take panoramic, cephalometric or 3D images of cranio-maxillofacial complex including the ear, nose and throat (ENT) and airway regions, and cervical spine. The device can be configured to take carpus images.
The device is operated and used by qualified healthcare professionals.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PSC Publishing Services (301) 443-6740 El
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Palodex Group. The word "Palodex" is written in a large, bold, sans-serif font. The word "Group" is written in a smaller, sans-serif font and is located to the lower right of the word "Palodex."
# Section V - 510(k) Summary for ORTHOPANTOMOGRAPH™ OP 3D LX
- 1. Submitter Information:
Palodex Group Oy Nahkelantie 160 FI-04300, Tuusula Finland
Contact Person: Frank Ray Telephone Number: (+358)10 270 2000 E-mail: frank.ray@envistaco.com
Date Prepared: June 9, 2023
#### 2. Device Name:
- ORTHOPANTOMOGRAPH™ OP 3D LX Proprietary Name: ●
- Manufacturer: .
- Common Name: ●
- Classification Name: ●
- CFR Number:
- Device Class: ●
- Product Code:
#### Predicate Device (Primary): 3.
- ORTHOPANTOMOGRAPH™ OP 3D (K180947) Proprietary Name: ●
Palodex Group OY
Palodex Group Oy
21 CFR 892.1750
II
OAS
X-Ray, Tomography, Computed, Dental
X-Ray, Tomography, Computed, Dental
X-Ray, Tomography, Computed, Dental
Computed tomography x-ray system
Computed tomography x-ray system
- Manufacturer: ●
- Common Name:
- Classification Name: Computed tomography x-ray system ●
- CFR Number: ●
- Device Class: ●
- 21 CFR 892.1750 II OAS Product Code:
# Predicate Device #2:
●
.
●
- Proprietary Name: Orthopantomograph OP300 (K163423) ●
Palodex Group OY
- Manufacturer: .
- Common Name: ●
- Classification Name:
- CFR Number: ● Device Class:
- CFR 892.1750 II
- OAS Product Code:
{4}------------------------------------------------
#### 4. Description of Device:
ORTHOPANTOMOGRAPH™ OP 3D LX manufactured by PaloDEx Group Oy is a panoramic, cephalometric and cone beam CT (CBCT) x-ray device for 2D and 3D imaging of patient's cranio-maxillofacial complex. It can be used to capture 2D and 3D x-ray images from pediatric and adult patients. The Primary Predicate Device ORTHOPANTOMOGRAPH™ OP 3D has already been cleared for panoramic and cone beam CT usage (K180947).
The ORTHOPANTOMOGRAPH™ OP 3D LX is an X-ray device that can be configured to take panoramic, cephalometric or 3D images of craniomaxillofacial complex, and neck areas including the ear, nose and throat (ENT). The device can be configured to take carpus images. The device is operated and used by qualified healthcare professionals. The ORTHOPANTOMOGRAPH™ OP 3D LX is intended to be used for imaging of adult and pediatric patients.
The ORTHOPANTOMOGRAPH™ OP 3D LX is part of digital dental workflow providing image data for diagnosis and treatment planning for the healthcare professionals. Xray images reveal the targeted craniofacial anatomy, and condition and the position of anatomical structures inside FOV, such as teeth, mandibular joints, and oral and nasal cavities. This helps dentists to prepare for various dental procedures such as implant placement, orthodontics and dental prosthetics, providing possible early diagnosis, which enables early and less invasive treatment.
Per the Guidance for Industry and FDA Staff entitled, "Bundling Multiple Devices or Multiple Indications in a Single Submission," dated June 22, 2007, PaloDEx Group Oy is bundling the ORTHOPANTOMOGRAPH™ OP 3D LX models below as they do not differ significantly in purpose, design, materials, energy source, function or any other feature related to substantial equivalence. The device description and intended use are identical for all models listed below. Also the supporting data is relevant to the ORTHOPANTOMOGRAPH™ OP 3D LX as a whole, the bundled device components and accessories are used together during therapeutic procedures, and one review group will be involved.
| Model Type | Model Part Numbers | Description |
|------------|----------------------------------------|-----------------------|
| PCX-1 | 0.805.5753 (SAP) / 901265-PTU (Oracle) | OP 3D LX (PAN 3D) |
| PCX-1 | 0.805.5754 (SAP) / 901266-PTU (Oracle) | OP 3D LX (PAN CEPH 3D |
Table 5.1: ORTHOPANTOMOGRAPH™ OP 3D LX models
## Principle of Operation / Mechanism of Action:
Panoramic imaging is used for acquiring 2D images. In the panoramic mode the rotating unit rotates around patient head and scans the patient with a narrow x-rav beam. Generation of exposure is turned off and turned on during the exposure to produce partial panoramic, bitewing (BW) and TMJ (Temporo Mandibular Joint) images.
Cephalometric imaging is performed by utilizing a second x-ray source. System is using same sensor for all imaging modalities. Both the x-ray source and the rotating unit make a scanning movement to produce either lateral cephalometric or PA image of the patient's head by using a narrow x-ray beam. Patient head is positioned in the ceph
{5}------------------------------------------------
head support. For carpus imaging there is a dedicated carpus holder to be attached to the ceph head support for supporting the object during exposure.
In 3D CBCT mode the rotating unit rotates around patient's head and scans the patient with a rectangular beam, where the beam size is adjusted depending on the selected FOV (Field-of-View) size. Several 2D projection images are recorded during this rotational movement. These 2D images are saved, and based on the data, the 3D data is generated by reconstruction. Once images are reconstructed, they can be viewed in an image viewing software (dental application). The image viewing software is not part of this proposed device.
## Accessories:
The Accessories for the ORTHOPANTOMOGRAPH™ OP 3D LX are listed in Table 5.1A below. All Accessories are the same/identical to the cleared Primary Predicate Device ORTHOPANTOMOGRAPH OP 3D (K180947).
| Device /<br>Accessory | Description | Manufacturer |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Disposable<br>covers | Intended to serve as a disposable barrier for dental<br>instruments to reduce the risk of cross contamination<br>between dental patients. | Palodex Group Oy |
| Chin rest | Used with bite block or lip support in standard and<br>pediatric imaging to provide patient support | Palodex Group Oy |
| Bite block | Used with chin rest in standard and pediatric imaging for<br>dentate patients to provide patient support | Palodex Group Oy |
| Lip support | Used with chin rest in standard and pediatric imaging for<br>edentulous patients and for TMJ imaging to provide<br>patient support | Palodex Group Oy |
| Head support | Used to provide patient support | Palodex Group Oy |
| Carpus holder | Used in carpus imaging to provide patient support | Palodex Group Oy |
## Table 5.1A: Accessories
#### ട. Indications for Use:
ORTHOPANTOMOGRAPH™ OP 3D LX is an X-ray device that is intended to be used for imaging of adult and pediatric patients. The device can be configured to take panoramic, cephalometric or 3D images of cranio-maxillofacial complex including the ear, nose and throat (ENT) and airway regions, and cervical spine. The device can be configured to take carpus images.
The device is operated and used by qualified healthcare professionals.
{6}------------------------------------------------
#### 6. Substantial Equivalence:
#### Details of the similarities between subject and predicate devices: A
The similarities between Subject Device, ORTHOPANTOMOGRAPH™ OP 3D LX, and the Primary Predicate Device, ORTHOPANTOMOGRAPH™ OP 3D (K180947), are listed in Table 12.1 below for Technological Characteristics and Performance Testing.
The proposed device ORTHOPANTOMOGRAPH™ OP 3D LX shares the similar architectural components and utilize as ORTHOPANTOMOGRAPH™ OP 3D (K180947). Both devices utilize cone beam x-ray technology to acquire volumetric data. The Conical x-ray beam rotates around the patient's head and incidents upon the receptor and the system generates volumetric data from the 2D projection images, similarly to the predicate device. Image modes, 3D resolutions, exposure times and 2D imaging programs are identical. The reconstruction of 3D images from 2D image data in the proposed device is made using the same reconstruction techniques as in the predicate device.
Performance testing were conducted on ORTHOPANTOMOGRAPH™ OP 3D LX. Image quality testing was performed as part of bench testing, and the result indicate that ORTHOPANTOMOGRAPH™ OP 3D LX is substantially equivalent to the predicate devices.
Biocompatibility evaluation was conducted on patient contacting parts and their material and found to be in conformance with ISO 10993-1 for both proposed and the predicate device.
Software on ORTHOPANTOMOGRAPH™ OP 3D LX is designed according to the manufacturer´s software development quidelines complying with IEC 62304. Testing was performed as part of the electrical safety testing.
- A Details of the differences between subject and predicate devices:
There are no major differences between the subject device (ORTHOPANTOMOGRAPH™ OP 3D LX) and predicate devices (ORTHOPANTOMOGRAPH™ OP 3D (K180947) and ORTHOPANTOMOGRAPH™ OP300 (K163423)), however there are some minor differences as described below.
## Intended Use/Indications of use:
"Ear, nose and throat (ENT) and airway regions, and cervical spine." is not part of the primary predicate intended use/indications of use. Secondary predicate ORTHOPANTOMOGRAPH™ OP300 (K163423) has also "neck area, ear, nose and throat (ENT) regions" included in the intended use/indications of use. Performance testing demonstrates the proposed ORTHOPANTOMOGRAPH™ OP 3D LX and both predicate devices can image the same anatomical areas, that is the dentomaxillofacial complex and neck area which include the airway regions and cervical spine.
As a summary ORTHOPANTOMOGRAPH™ OP 3D LX Indications for Use statement is utilizing more descriptive wording to communicate anatomical structures inside the scanned volume.
{7}------------------------------------------------
### X-ray source:
The proposed and the predicate devices utilize the same x-ray tube. The chosen technique factors are slightly different. The small difference in the device's tube current is due to the proposed device having longer SID requiring a bit larger current for imaging.
### Image Detectors:
The proposed device has only one TFT (Thin Film Transistor) sensor capable of capturing all modalities (panoramic, cephalometric and CBCT image data).
Using only one sensor is possible due to the bigger image detector size at the proposed device (Effective detector area 240 x168 mm) compared to the Primary Predicate (Effective detector area 125 x125 mm).
TFT imaging performance is similar to CMOS sensors used with the Predicate device and equivalent clinical performance can be achieved. Measured DQE performance is substantially equivalent to the predicate device.
### 3D Field of View:
The larger image detector in the proposed device enables bigger field of view options. Field of view options in the proposed device are optimized for clinical needs to achieve intended use of the device.
### Pixel size:
The proposed and the primary predicate device have almost identical pixel sizes. 5 um /10 um difference in pixel sizes is not significant in clinical practice.
## System footprint:
Minor difference in the adjustable height of the device does not impact the safety or effectiveness of the device.
## Materials (Patient contacting):
The proposed device uses the same materials for the patient contacting parts with exception for the head support. ORTHOPANTOMOGRAPH™ OP 3D LX uses "Grilamid TR 90+ EMD- 802744 MB Black" also for the head support and "Ultradur B2550 FC+ EMD-802744 MB Black" is not used.
However, these minor differences do not raise new concerns of substantial equivalence. The comparison below (Table 5-2) for the Subject Device and Predicate Devices demonstrates that the Subject device is substantially equivalent to the Primary Predicate with regards to their Indications for use, technology and performance specifications.
Furthermore the subject device does not introduce a fundamentally new scientific technology, and nonclinical performance testing demonstrates that the device is substantially equivalent as the Subject device performs as well as the Primary Predicate device for its intended use.
{8}------------------------------------------------
# Device Comparison Table:
| Table 5.2: SE Table | | | | | |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|--|
| Descriptive<br>Information | Subject Device<br>ORTHOPANTOMOGRAPH<br>TM OP 3D LX | Primary Predicate Device<br>ORTHOPANTOMOGRAPH<br>TM OP 3D (K180947) | Predicate Device #2<br>Orthopantomograph<br>OP300<br>(K163423) | Comparison | |
| Regulatory Information | | | | | |
| Manufacturer | Palodex Group OY | Palodex Group OY | Palodex Group OY | Same | |
| Regulation # | 21 CFR 892.1750 | 21 CFR 892.1750 | 21 CFR 892.1750 | Same | |
| Regulation<br>Title | X-Ray, Tomography,<br>Computed, Dental | X-Ray, Tomography,<br>Computed, Dental | X-Ray, Tomography,<br>Computed, Dental | Same | |
| Regulation<br>Class | Class II | Class II | Class II | Same | |
| Product Code | OAS | OAS | OAS | Same | |
| Indications for Use/Intended Use | | | | | |
| Indications for<br>Use | ORTHOPANTOMOGRAPH<br>PHTM OP 3D LX is an X-<br>ray device that is<br>intended to be used for<br>imaging of adult and<br>pediatric patients. The<br>device can be<br>configured to take<br>panoramic,<br>cephalometric or 3D<br>images of cranio-<br>maxillofacial complex<br>including the ear, nose<br>and throat (ENT) and<br>airway regions, and<br>cervical spine. The<br>device can be<br>configured to take<br>carpus images.<br><br>The device is operated<br>and used by qualified<br>healthcare<br>professionals. | ORTHOPANTOMOGRAPH<br>PHTM OP 3D is an x-ray<br>device that is intended to<br>be used for imaging of<br>adult and pediatric<br>patients. The device can<br>be configured to take<br>panoramic,<br>cephalometric, or 3D<br>images of the cranio-<br>maxillofacial complex for<br>use in diagnostic<br>support.<br>The device can also be<br>configured to take<br>carpus images.<br><br>ORTHOPANTOMOGRAPH<br>PHTM OP 3D must only<br>be used and operated by<br>dentist and other<br>qualified professionals. | ORTHOPANTOMOGR<br>APHTM OP300<br>panoramic,<br>cephalometric and<br>cone beam computed<br>tomography x-ray<br>device is intended to<br>image the head and<br>neck areas for<br>diagnostic support.<br>This includes<br>temporomandibular<br>Joints (TMJs) and<br>dentomaxillofacial<br>areas, and with the<br>13x15 cm field of view<br>(FOV) this additionally<br>includes the ear, nose<br>and throat (ENT)<br>regions. The x-ray<br>device produces<br>conventional 2D x-ray<br>images and x-ray<br>projection images for<br>the reconstruction of a<br>3D view. The device is<br>operated and used by<br>qualified healthcare<br>professionals. | Same<br>Indication for<br>Use as<br>Primary<br>Predicate and<br>Predicate #2<br>and expressed<br>through a<br>similar choice<br>of words. | |
| Technological Characteristics | | | | | |
| Imaging<br>modes | Panoramic, TMJ, BW,<br>3D (CBCT),<br>cephalometric | Panoramic, TMJ, BW,<br>3D (CBCT),<br>cephalometric | Panoramic, TMJ,<br>3D(CBCT),<br>cephalometric | Identical to<br>Primary<br>Predicate | |
| X-ray source | 3D mode: 95kV<br>Pan mode: 60-90 kV<br>Ceph mode :60-95kV | 3D mode: 95kV<br>Pan mode: 60-90 kV<br>Ceph mode :60-95kV | 3D mode: 90kV<br>Pan mode: 57-90 kV<br>Ceph mode: 60-90 kV | Similar to<br>Primary<br>Predicate | |
| | kV accuracy: +/-5kV | kV accuracy: +/-5kV | kV accuracy: +/-5kV | | |
| Discriptive<br>Information | Subject Device<br>ORTHOPANTOMOGRAPH<br>TM OP 3D LX | Primary Predicate Device<br>ORTHOPANTOMOGRAPH<br>TM OP 3D (K180947) | Predicate Device #2<br>Orthopantomograph<br>OP300<br>(K163423) | Comparison | |
| | mA range: 2-16 mA<br>3D power mode:<br>continuous/pulsed | mA range: 1-16 mA<br>3D power mode:<br>continuous/pulsed | mA range: 3.2-16 mA<br>3D power mode:<br>pulsed | | |
| Focal spot | 0.5 mm | 0.5 mm | 0.5 mm | Identical to<br>Primary<br>Predicate | |
| Image<br>detector(s) | TFT Flat Panel Detector<br>FXDD-1724RA (Vivix-D<br>1724R)<br>(2D and 3D) | CMOS Flat Panel<br>CD41M115 (3D)<br>CMOS detector<br>CD51M114/CD52M21X<br>(2D) | CMOS Flat Panel<br>C10900D-05 (3D)<br>CMOS detector<br>C10500D/C10502D<br>(2D) | Similar to<br>Primary<br>Predicate | |
| Image<br>detector<br>scintillator | CsI | CsI | CsI | Same | |
| 2D imaging<br>performance<br>- DQE<br>70kV, RQA5 | 55% @ 1lp/mm<br>40% @ 2lp/mm | 45% @ 1lp/mm<br>30% @ 2lp/mm | N/A | Similar to<br>Primary<br>Predicate | |
| 2D imaging<br>performance<br>- MTF<br>70kV, RQA5 | 60% @ 1lp/mm<br>30% @ 2lp/mm | 60% @ 1lp/mm<br>30% @ 2lp/mm | N/A | Identical to<br>Primary<br>Predicate | |
| 3D imaging<br>technique | Reconstruction from 2D<br>images | Reconstruction from 2D<br>images | Reconstruction from<br>2D images | Identical to<br>Primary<br>Predicate | |
| 3D's Field Of<br>View (FOV) | 5 x 5 cm<br>6 x 9 cm<br>8 x 8 cm<br>10 x 10 cm<br>12 x 15 cm (optional)<br>15 x 20 cm (optional) | 5 x 5 cm<br>6 x 9 cm<br>9 x 11 cm<br>9 x 14 cm (optional) | 4 x 6 cm<br>5 x 5 cm<br>6.1 x 7.8 cm<br>7.8 x 7.8 cm<br>7.8 x 15 cm<br>13 x 15 cm | Similar to<br>Primary<br>Predicate<br>Similar to<br>Predicate #2<br>(biggest FOV) | |
| 3D's total<br>viewing angle | 360° | 360° | 360° | Identical to<br>Primary<br>Predicate | |
| Pixel size | Flat Panel Detector<br>95μm (2D)<br>190 μm (3D) | CMOS PAN & 3D<br>combo flat panel:<br>3D pixel size: 198 μm<br>pan/ceph pixel size:<br>99μm | CMOS flat panel for<br>3D: 200 μm<br>CMOS for panoramic<br>imaging: 100 μm | Similar to<br>Primary<br>Predicate | |
| Voxel size | 80–400 μm | 80 - 400 μm | 85-420 μm | Identical to<br>Primary<br>Predicate | |
| Reconstructio<br>n Software | Filtered Back Projection<br>(FBP) | Filtered Back Projection<br>(FBP) | Filtered Back<br>Projection (FBP) | Identical to<br>Primary<br>Predicate | |
| Discriptive<br>Information | Subject Device<br>ORTHOPANTOMOGRAPH<br>TM OP 3D LX | Primary Predicate Device<br>ORTHOPANTOMOGRAPH<br>TM OP 3D (K180947) | Predicate Device #2<br>Orthopantomograph<br>OP300<br>(K163423) | Comparison | |
| 3D's effective<br>exposure<br>time | 1,7-20 sec | 1,7-20 sec | 1.2 - 12.6 sec | Identical to<br>Primary<br>Predicate | |
| Ceph<br>exposure<br>time | 8.1-10.5 s | 8.1–10.5 s | 6,5-20s | Identical to<br>Primary<br>Predicate | |
| Patient's<br>Position | Standing and wheelchair | Standing and wheelchair | Standing and<br>wheelchair | Identical to<br>Primary<br>Predicate | |
| System<br>footprint<br>(includes the<br>operator) | H169-244cm x<br>D 124-149cm x 167cm | H167-247cm x D124-<br>149cm x 167 cm | H161-241cm x<br>D1390cm x W97-193<br>cm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.