K253935 · medXion Technologies, Inc. · OAS · May 15, 2026 · Radiology
Device Facts
Record ID
K253935
Device Name
medXion NEXUS
Applicant
medXion Technologies, Inc.
Product Code
OAS · Radiology
Decision Date
May 15, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
MedXion NEXUS is intended to produce three-dimensional digital X-ray images of the dental, oral, maxillofacial region, and ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients.
Device Story
medXion NEXUS is a computed tomography (CT) X-ray system for dental, oral, maxillofacial, and ENT imaging. The system comprises a scanner and a console. During operation, a rotation arm moves around the patient's head; an X-ray generator emits radiation, which is captured by a flat panel detector (FPD). The console receives raw absorption data, performs image reconstruction to generate 3D CT and tomographic images, and provides image analysis tools (e.g., airway measurement). The device is operated by healthcare professionals in a clinical setting. Patient movement is managed via neck supports and head stabilizers; no patient restraints are used. The system utilizes continuous exposure with high-frame-rate acquisition to maintain diagnostic quality at lower radiation doses compared to pulsed systems. Output images are used by clinicians for diagnostic support. The device benefits patients by providing high-resolution diagnostic imaging with optimized radiation exposure.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1 series), electromagnetic compatibility (IEC 60601-1-2), and image quality performance (IEC 61223-3-7). Metal artifact reduction (MAR) was verified using phantoms.
Technological Characteristics
Computed tomography X-ray system; 90-110kV tube voltage; 1-3mA tube current; 0.3mm x 0.3mm focal spot. Detector: IGZO TFT flat panel. Connectivity: DICOM (NEMA PS 3.1-3.20). Non-sterile; no patient-contacting materials. Software: Image analysis system for 3D reconstruction and airway measurement. Standards: ANSI/AAMI 60601-1, IEC 60601-1-2 Ed 4.1, IEC 60601-1-3 Ed 2.2, IEC 60601-1-6 Ed 3.2, IEC 60601-2-63 Ed 1.2, IEC 62304 Ed 1.1, ISO 14971.
Indications for Use
Indicated for adult and pediatric patients (≥21 kg, ≥113 cm) for 3D digital X-ray imaging of dental, oral, maxillofacial, ENT, and neck regions as diagnostic support.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 15, 2026
medXion Technologies, Inc.
% Kazuhito Tomii
CEO
2-12-27 Kodo
Adachi-Ku
TOKYO, 1200013
JAPAN
Re: K253935
Trade/Device Name: medXion NEXUS
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: OAS
Dated: December 1, 2025
Received: April 13, 2026
Dear Kazuhito Tomii:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253935 - Kazuhito Tomii
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253935 - Kazuhito Tomii
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253935 | |
| Device Name medXion NEXUS | |
| Indications for Use (Describe) MedXion NEXUS is intended to produce three-dimensional digital X-ray images of the dental, oral, maxillofacial region, and ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
# 510(k) Summary
Device Name: medXion NEXUS
# K253935
1. Submission Sponsor
medXion Technologies Inc.
2-12-27 Kodo,
Adachi-ku Tokyo 1200013
Japan
Kazuhito Tomii
CEO
Email: tomii@medxion.co.jp
Office number: +81-5010557778
2. Submission Correspondent
Same as above
3. Date Prepared
2026-5-13
4. Device Identification
Product/Trade Name: medXion NEXUS
Common Name: Computed Tomography X-ray System
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: OAS
{5}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
## 5. Legally Marketed Predicate Device(s) and Reference Device
Predicate Device:
Product/Trade Name: PreXion3D Explorer
EX
510(K) Number: K193329
Common Name: Computed Tomography X-ray System
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: OAS
## 6. Indication for Use Statement
MedXion NEXUS is intended to produce three-dimensional digital X-ray images of the dental, oral, maxillofacial region, and ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients.
## 7. Device Description
medXion NEXUS consists of a scanner, which is used for generating X-ray and detecting image data, and a Console, which is used for operating the scanner and managing the data. The scan data acquired by the scanner will be transferred to the Console. medXion NEXUS Image Analysis System will then perform the image analysis or image edition (creating cross-section diagram, etc.), and output the image to a printer.
X-ray image data is acquired while the rotation arm is rotating around the secured “patient’s head” at a constant speed. X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm). The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan) and the tomographic image.
This device does not use patient restraints such as chin rests or headrests, as requested by the customer. Neck supports and head stabilizers are used to suppress patient movement.
As “Table-6A: Comparison of Characteristics” shows that medXion NEXUS (Device in concern) is the equivalent with the predicate device, PreXion3D Explorer EX (K193329).
{6}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
# 8. Substantial Equivalence Discussion
The selected predicate device (K193329) has identical indications for use and utilizes the same fundamental technology as the subject device. Any minor technological differences do not raise new questions of safety or effectiveness, as demonstrated by the bench performance testing provided in this submission.
Table-6A: Comparison of Characteristics
| | Subject Device | Predicate Device | Comparison to predicate |
| --- | --- | --- | --- |
| Manufacturer | medXion Technologies Inc. | PreXion Corporation | |
| Trade Name | medXion NEXUS | PreXion3D Explorer EX | |
| 510(k) Number | | K193329 | |
| Product Code | OAS | OAS | Same |
| Regulation Number | OAS: 21 CFR 892.1750 | OAS: 21 CFR 892.1750 | Same |
| Regulation Name | OAS: Computed tomography x-ray system | OAS: Computed tomography x-ray system | Same |
| Device Classification Name | X-Ray, Tomography, Computed, Dental | X-Ray, Tomography, Computed, Dental | Same |
| Indications for use: | medXion NEXUS is intended to produce three-dimensional digital X-ray images of the dental, oral, maxillofacial region, and ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year-old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing radiation is higher), and so unnecessary radiation exposure | PreXion3D Explorer EX is intended to produce two-dimensional digital x-ray images including panoramic image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year-old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more radiosensitive than adults (i.e., the | The differences in pediatric-related wording do not raise new questions of safety or effectiveness, For details, please refer to 1. |
{7}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
| | is of particular concern for pediatric patients. | cancer risk per unit dose of ionizing radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients. | |
| --- | --- | --- | --- |
| Patient/User Characteristics | | | |
| Target Population | Children aged 6 (except infants) to elderly | Children aged 6 (except infants) to elderly | Same |
| Anatomical Site | The dental, oral, maxillofacial region ENT (Ear, Nose and Throat) and neck region | The dental, oral, maxillofacial region ENT (Ear, Nose and Throat) and neck region | Same |
| Users | Health care professionals | Health care professionals | Same |
| Technological Characteristics and Performance | | | |
| Patient Contact Material | None Patient stabilization approach without rigid fixation | CHIN REST: polycarbonate Forehead Holder: silicone rubber HANDLE GRIP: silicone rubber | Subject device has no parts come into contact with the patient. For details, please refer to 2. |
| Sterility | Non-sterile | Non-sterile | Same |
| X-ray Generator | | | |
| Tube Voltage | 90-110KV | 90-110KV | Same |
| Exposure function | Continuous | Pulsed | Lower total radiation dose compared to predicate. For details, please refer to 3. |
| Tube Current | 1-3mA | 1-5mA | Reducing the tube current reduces the radiation dose. |
| Focal Spot Size | 0.3mm x 0.3mm | 0.3mm x 0.3mm | Same |
| FPD Detector | | | |
| Detector | FPD (IGZO TFT) | FPD (TFT) | The sensitivity has been improved. |
| Pixel Size | 560 μm x 560μm (With 4x4binning) 280 μm x 280μm (With 2x2binning) 140 μm x140μm (Without binning) | 248 μm x248μm (With 2x2 binning) 124 μm x124μm (Without binning) | The Subject Device is substantially equivalent to the Predicate Device with respect to pixel size and image resolution. For details, please refer to 4. |
| Pixel Number | 756x756 (With 4x4 binning) 1536x1536 (With 2x2 binning) 3072x3072 (Without binning) | 1024x1280(With 2x2 binning) 2560x2048(Without binning) | The Subject Device is substantially equivalent to the |
{8}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
| | | | Predicate Device with respect to pixel size and image resolution. For details, please refer to 4. |
| --- | --- | --- | --- |
| **Size of Active Area** | 430.08mm x 430.08mm | 253.95mm x 317.44mm 230mm x 15mm (Panoramic) | The larger active area gives the advantage. For details, please refer to 5. |
| **Number of Bits** | 16bits | 16bits | Same |
| **Scanner** | | | |
| **SID/SOD** | 800mm/ 520mm | 700mm/ 420mm | The difference in SID is due to geometric balancing and does not result in differences in safety or performance. For details, please refer to 6. |
| **Dimension (WxDxH)** | 1210 mm x 1247 mm x 2271 mm | 880 mm x 1237 mm x 2268 mm | Differences in device shape do not affect safety or performance |
| **Voltage** | 120VAC | 120VAC | Same |
| **Input** | 1.0 kVA | 1.0 kVA | Same |
| **3D Imaging** | | | |
| **Scan Time** | 6-16.4sec | 10-20sec | The exposure time is shortened to keep the dose low. |
| **FOV (Voxel Size)** | φ 300mm x H300mm (0.300 - 0.500mm) φ 150mm x H150mm (0.150 - 0.200mm) | φ 150mm x H100mm (0.100 - 0.200mm) φ 100mm x H100mm (0.100 - 0.200mm) φ 50mm x H50mm(0.100 - 0.200mm) | The larger FOV than the predicate allows for greater coverage of a single scan, providing an advantage. For details, please refer to 7. |
| **Dose Level (CTDIw)** | Standard mode Conditions: 110KV/2mA/6s 30x30cm: 2.19mGy | Standard mode Conditions: 110KV/2mA/10s 15x15cm: 8.7mGy 5x5 cm: 0.9mGy | Compared to predicates, it has advantages due to its low dose. For details, please refer to 8. |
5
{9}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
| | High Definition Mode Conditions: 110KV/2mA/12s 30x30cm: 4.51mGy Ultra-High Definition Conditions: 110KV/2mA/16s 30x30cm: 5.95mGy | High Definition Mode Conditions: 110KV/2mA/18s 15x15cm: 15.3mGy 5x5 cm: 1.6mGy Ultra-High Definition Conditions: 110KV/2mA/20s 15x15cm: 23.2mGy 5x5 cm: 4mGy | |
| --- | --- | --- | --- |
| Spatial Resolution | Standard mode: 50%MTF: 0.95 LP/mm 20%MTF: 2.0 LP/mm High definition mode: 50%MTF: 0.95LP/mm 20%MTF: 2.0 LP/mm Ultra High definition mode: 50%MTF: 0.95LP/mm 20%MTF: 2.0LP/mm | Standard mode: 50%MTF: 0.95LP/mm 20%MTF:2.0LP/mm High Resolution mode: 50%MTF: 0.95LP/mm 20%MTF:2.0LP/mm High contrast mode: 50%MTF: 0.95LP/mm 20%MTF:2.0LP/mm | It has the same performance as predicate. |
| Viewer Software (Image Analysis System Software) | Display High-resolution 2D and 3D Images Function Image Processing Function includes Airway measurement Image Operation Function Output Function | Display High-resolution 2D and 3D Images Function Image Processing Function include Airway measurement Image Operation Function Output Function | Same |
| Console Software System Settings | CT Scan | CT Scan include CT-Panoramic mode | Same |
| Applied Standard | | | |
| Electrical Safety Standard | ANSI/AAMI 60601-1 IEC60601-1-2 Ed 4.1 IEC60601-1-3 Ed 2.2 IEC60601-1-6 Ed 3.2 IEC60601-2-63 Ed 1.2 IEC62304 Ed 1.1 IEC 62366-1:2025+A1:2020 IEC61223-3-4 IEC61223-3-7 ISO 14971 | ANSI/AAMI ES60601-1 IEC60601-1-2 IEC60601-1-3 IEC60601-1-6 IEC60601-2-63 IEC62304 IEC 62366 IEC61223-3-4 IEC61223-3-5 IEC60825-1 ISO 14971 | Because the subject device does not use laser units, it is not subject to IEC 60825-1. |
{10}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
| Laser Safety Standard | None | IEC 60825-1 | This device does not use lasers for positioning. |
| --- | --- | --- | --- |
| DICOM Standard | NEMA PS 3.1-3.20 | NEMA PS 3.1 - 3.20 | Same |
| Biocompatibility | None | ISO 10993-1 ISO 10993-5 ISO 10993-10 | No parts come into contact with the patient |
The subject and predicate devices also have differences in technological characteristics, summarized as follows:
1. Indications for use
The Subject Device, medXion NEXUS, and the Predicate Device, PreXion3D Explorer EX, have the same intended use. Both devices are intended to produce three-dimensional digital X-ray images of the dental, oral, maxillofacial, ENT (ear, nose, and throat), and neck regions. Both devices are used at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
The Predicate Device includes additional descriptive language regarding minimum patient height and weight and warnings related to radiation exposure in smaller pediatric patients. The Subject Device does not include explicit numerical height or weight limitations in its Indications for Use. This difference does not represent a change in intended use or target patient population. Rather, the additional language in the Predicate Device provides precautionary and explanatory information related to known radiation safety considerations for pediatric patients.
Therefore, the differences in pediatric-related wording do not raise new questions of safety or effectiveness, and the Subject Device is substantially equivalent to the Predicate Device with respect to Indications for Use.
2. Patient Contact Material:
The Predicate Device, PreXion3D Explorer EX, includes a chin rest that may contact the patient's skin during use. The Subject Device, medXion NEXUS, does not include a chin rest and therefore does not have any direct or indirect patient-contacting components.
This difference does not affect the intended use, clinical application, or safety and effectiveness of the device. The Subject Device eliminates patient skin contact associated with a chin rest by using alternative non-contact or repositioned patient support and alignment methods. (e.g., using a chair with a backrest or headrest) As a result, the Subject Device presents an equal or lower biocompatibility risk compared to the Predicate Device.
{11}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
Because the Subject Device has no patient-contacting materials, biocompatibility evaluation in accordance with ISO 10993 is not required. The absence of patient-contacting components does not introduce new risks and does not raise new questions of safety or effectiveness.
Therefore, despite the presence of a chin rest in the Predicate Device and its absence in the Subject Device, the devices are substantially equivalent with respect to patient contact materials.
## 3. Exposure Function
The Predicate Device utilizes a pulsed exposure function, while the Subject Device, medXion NEXUS, employs a continuous exposure mode during image acquisition. This difference represents an implementation choice in exposure control and does not change the intended use or diagnostic purpose of the device.
The Subject Device adopts continuous exposure in combination with a high-speed, high frame-rate image acquisition system. This design allows sufficient image data to be acquired within a shorter total scan time, thereby enabling the use of lower exposure per frame and reducing the need for repeated or prolonged exposures. As a result, the overall patient radiation dose can be equal to or lower than that of pulsed exposure systems designed for comparable diagnostic performance.
Therefore, the use of continuous exposure with high frame-rate imaging in the Subject Device does not raise new questions of safety or effectiveness and supports substantial equivalence to the Predicate Device with respect to the exposure function.
## 4. Pixel Size / Pixel Number
The Subject Device, medXion NEXUS, and the Predicate Device, PreXion3D Explorer EX, utilize digital X-ray detectors with different nominal pixel sizes and numbers. This difference reflects a design choice in detector technology and does not affect the intended use, clinical application, or diagnostic purpose of the devices.
Both devices are designed to produce diagnostically acceptable three-dimensional images of the dental, oral, maxillofacial, ENT, and neck regions. The effective spatial resolution of each system is determined by the overall imaging chain, including detector characteristics, focal spot size, system geometry, reconstruction algorithms, and image processing, rather than pixel size alone.
{12}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
The Subject Device compensates for pixel size differences through system-level optimization, including detector performance, acquisition parameters, and image reconstruction methods, to achieve image quality suitable for the same diagnostic tasks as the Predicate Device. Image quality performance has been verified through appropriate bench testing and image quality evaluations, demonstrating that the Subject Device provides comparable diagnostic information.
Therefore, the difference in pixel size does not introduce new clinical indications, does not impact diagnostic performance, and does not raise new questions of safety or effectiveness. The Subject Device is substantially equivalent to the Predicate Device with respect to pixel size and image resolution.
# 5. Size of Active Area
The Subject Device, medXion NEXUS, utilizes a digital X-ray detector with a larger active area compared to the Predicate Device, PreXion3D Explorer EX. This difference represents a design enhancement and does not alter the intended use, diagnostic purpose, or target patient population of the device.
The larger active area of the Subject Device provides improved anatomical coverage per acquisition, which reduces the need for repositioning or repeat scans and supports efficient image acquisition. This design characteristic may improve workflow and can contribute to dose efficiency by minimizing the likelihood of additional exposures due to incomplete coverage.
Despite this difference, both devices are intended to acquire diagnostically acceptable two-dimensional and three-dimensional images of the dental, oral, maxillofacial, ENT, and neck regions. The fundamental imaging principles, X-ray generation, image acquisition, reconstruction, and display functions are comparable between the two devices.
The increased active area does not introduce new clinical indications, new imaging modes, or new user interactions. Image quality and radiation dose performance of the Subject Device have been verified through appropriate bench testing to demonstrate that it meets applicable performance and safety requirements and provides diagnostic information comparable to the Predicate Device.
Therefore, the difference in detector active area does not raise new questions of safety or effectiveness. The Subject Device is substantially equivalent to the Predicate Device with respect to detector active area.
9
{13}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
## 6. SID
The Subject Device, medXion NEXUS, and the Predicate Device, PreXion3D Explorer EX, have different nominal source-to-image distances (SID). This difference reflects a system geometry design choice and does not affect the intended use, clinical application, or diagnostic purpose of the devices.
Both devices are designed to produce diagnostically acceptable three-dimensional X-ray images of the dental, oral, maxillofacial, ENT, and neck regions. The overall image quality and radiation dose performance are determined by the complete imaging system, including SID, focal spot size, detector characteristics, acquisition parameters, and image reconstruction algorithms, rather than SID alone.
The Subject Device optimizes its system geometry to balance image magnification, spatial resolution, and radiation dose. Any differences in SID are compensated through detector design, acquisition settings, and reconstruction processing to achieve image quality suitable for the same diagnostic tasks as the Predicate Device. Performance verification testing demonstrates that the Subject Device provides comparable diagnostic information.
The difference in SID does not introduce new clinical indications, new imaging modes, or new user interactions, and does not raise new questions of safety or effectiveness.
Therefore, despite differences in SID, the Subject Device is substantially equivalent to the Predicate Device with respect to imaging geometry and diagnostic performance.
## 7. FOV
The Subject Device, medXion NEXUS, provides a larger maximum field of view (FOV) compared to the Predicate Device, PreXion3D Explorer EX. This difference represents a design enhancement and does not change the intended use, diagnostic purpose, or target patient population of the device.
The larger FOV of the Subject Device allows broader anatomical coverage in a single acquisition, which may reduce the need for multiple scans or patient repositioning and can improve clinical workflow. This capability supports comprehensive imaging of the dental, oral, maxillofacial, ENT, and neck regions for the same diagnostic applications as the Predicate Device.
10
{14}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
Despite the difference in maximum FOV size, both devices operate based on the same imaging principles and are intended to produce diagnostically acceptable three-dimensional X-ray images for identical clinical indications. The Subject Device does not introduce new imaging indications, new clinical applications, or new use environments as a result of the larger FOV.
Therefore, the larger FOV of the Subject Device does not raise new questions of safety or effectiveness. The Subject Device is substantially equivalent to the Predicate Device with respect to field of view.
## 8. Dose Level:
The Subject Device, medXion NEXUS, utilizes a higher-performance flat panel detector (FPD) compared to the detector used in the Predicate Device, PreXion3D Explorer EX. The enhanced detector performance represents a design improvement and does not change the intended use, diagnostic purpose, or target patient population of the device.
The higher-performance FPD of the Subject Device provides improved X-ray detection efficiency and signal-to-noise performance. As a result, diagnostically acceptable image quality can be achieved at lower X-ray exposure levels, enabling a reduction in patient radiation dose for equivalent imaging tasks. This improvement supports dose efficiency without compromising diagnostic performance.
Both devices are intended to produce two-dimensional and three-dimensional digital X-ray images of the dental, oral, maxillofacial, ENT, and neck regions and are operated in accordance with established radiation safety principles, including the ALARA principle. The Subject Device does not introduce new imaging modes, new clinical indications, or new user interactions as a result of the higher-performance detector.
Performance verification testing demonstrates that the Subject Device achieves image quality suitable for the same diagnostic applications as the Predicate Device while operating at comparable or lower radiation dose levels. Therefore, the use of a higher-performance FPD does not raise new questions of safety or effectiveness.
Accordingly, the Subject Device is substantially equivalent to the Predicate Device with respect to detector technology and radiation dose performance.
11
{15}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
## 8. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of medXion NEXUS and in showing substantial equivalence to the predicate device, medXion Technologies Inc. completed a number of non-clinical performance tests. The medXion NEXUS meets all the requirements for overall design, biocompatibility, performance, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
The medXion NEXUS passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
### Electrical safety and electromagnetic compatibility (EMC):
- Electrical safety testing per ABSI/AAMI ES 60601-1, IEC 60601-1-3 Ed 2.2 and IEC 60601-1-6 Ed 3.2
- Electromagnetic Safety testing per IEC 60601-1-2 Ed 4.1
- Dental extra-oral X-ray equipment testing per IEC 60601-2-63 Ed 1.2
- Acceptance testing of X-ray equipment per IEC 61223-3-4
### Software Verification and Validation Testing:
- DICOM testing per NEMA PS 3.1 - 3.20
- Usability testing per IEC 62366-1:2025+A1:2020
- Software verification and validation per IEC 62304 Ed 1.1
- Software Documentation per: “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices”
### Bench Testing:
- IEC 61223-3-7:2021 for CBCT image quality and acceptance testing
- Applicable portions of IEC 60601-2-63 related to dental CBCT system performance and image quality evaluation
- Regarding the metal artifact reduction (MAR) function, we conducted verification using dummy implants and bone phantoms and confirmed that diagnostic image quality is maintained without generating new artifacts.
12
{16}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
- Performance testing in laboratory environment was performed with PreXion3D Explorer EX (primary predicate device) of Bone Phantoms. The data was compared in order to see that the performance of the device remains substantially similar to that of the primary predicate device.
## Solid State X-ray Imagers (SSXI):
- Establish the substantial equivalence of an SSXI to a previously cleared conventional radiographic SSXI per: "Guidance for the Submission of 510(k)" for Solid State X-ray Imaging Devices"
## Cyber Security:
- Cybersecurity Activities per: "Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions" issued on February 3, 2026
The non-clinical testing outcomes for this device are satisfactory and demonstrate similarity to those of the predicate, supporting a finding of substantial equivalence with respect to technological characteristics and performance.
## 9. Clinical Performance Data
There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
## 10. Cybersecurity Activities
We followed the guidance and performed the following activities:
- Cybersecurity risk assessment and threat modeling
- Software Bill of Materials (SBOM) generation and review
- Vulnerability assessment of software components and operating environment
- Evaluation of network interfaces, communication pathways, and access control mechanisms
- Review of update and patch management procedures
- Assessment of logging and monitoring capabilities
- Verification of secure configuration and deployment requirements
- Vulnerability Testing
- Penetration testing
{17}
medXion Technologies Inc.
Traditional 510(k) Premarket Submission
medXion NEXUS
We make post-market Cybersecurity actions as follows:
1. Collect cybersecurity information from the link described in Post-market Plan
As of April, 2026, there is no cyber security thread and penetration by medXion’s investigation.
2. Collect information if there is any cyber security breach at customer site.
## 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device(s).
The medXion NEXUS, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.