PreXion3D Explorer EX is intended to produce two-dimensional digital x-ray images including panoramic image and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatic patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients.
Device Story
PreXion3D Explorer EX is a computed tomography X-ray system for dental/maxillofacial imaging. It uses a rotating arm with an X-ray generator and flat panel detector to capture absorption data while rotating around a patient's head. The system reconstructs 2D panoramic and 3D CT images on a console. Operated by healthcare professionals in clinical settings, the device provides diagnostic images for clinical decision-making. Modifications from the predicate include adding a panoramic scan function, new FOV options, and updated user interface components (chinrest, touch panel, mirror) derived from the PreXion3D Excelsior. The device is intended to support diagnosis in dental and ENT applications.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and imaging performance (IEC 61223-3-4/5). An American board-certified radiologist confirmed that images produced with the modified detector effective area are of diagnostic quality.
Indicated for adult and pediatric patients (≥21 kg, ≥113 cm) for 2D (panoramic) and 3D digital X-ray imaging of dental, oral, maxillofacial, ENT, and neck regions as diagnostic support.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
PreXion Corporation % Hiroaki Takahashi General Manager, Quality Assurance & Regulatory Division 1-14-1, Kanda Suda-cho Chiyoda-ku. Tokyo 101-0041 JAPAN
April 9, 2020
### Re: K193329
Trade/Device Name: PreXion3D Explorer EX Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: March 16, 2020 Received: March 19, 2020
#### Dear Hiroaki Takahashi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K193329
Device Name PreXion3D Explorer EX
#### Indications for Use (Describe)
PreXion3D Explorer EX is intended to produce two-dimensional digital x-ray images including panoramic image and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatic patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients.
Type of Use (Select one or both, as applicable)
| <div style="display:inline-block;"><div><span style="font-size: 20px;">☑</span></div>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:inline-block;"><div><span style="font-size: 20px;">☐</span></div>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Prexion. The logo features the word "PreXion" in a light gray, elegant font. Below the name is the tagline "make IT visible" with the "IT" in blue. Above and to the right of the name is a blue swoosh design with a gray shadow, adding a modern and dynamic element to the logo.
# 510(k) Summary
PreXion3D Explorer EX Device Name:
## K193329
#### 1. Submission Sponsor
PreXion Corporation 1-14-1, Kandasuda-cho, Chiyoda-ku, Tokyo 101-0041 Japan Hiroaki Takahashi General Manager, Quality Assurance & Regulatory Division Email: px-ra@prexion.co.jp Office number: +81-3-5297-7551
#### 2. Submission Correspondent
Same as above
#### 3. Date Prepared
November 29th, 2019
#### 4. Device Identification
| Product/Trade Name: | PreXion3D Explorer EX |
|-----------------------|----------------------------------|
| Common Name: | Computed Tomography X-ray System |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | OAS |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for PreXion. The logo features the word "PreXion" in a stylized font, with the "X" being slightly larger than the other letters. Below the company name is the tagline "make IT visible" in a smaller font. Above and to the right of the company name is a blue and white graphic design.
#### 5. Legally Marketed Predicate Device(s) and Reference Device
| Predicate Device: | |
|-----------------------|----------------------------------|
| Product/Trade Name: | PreXion3D Explorer |
| 510(K) Number: | K190320 |
| Common Name: | Computed Tomography X-ray System |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | OAS |
| Reference Device: | |
| Product/Trade Name: | PreXion3D Excelsior |
| 510(K) Number: | K181983 |
| Common Name: | Computed Tomography X-ray System |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR §892.1750 |
| Device Class: | Class II |
#### 6. Indication for Use Statement
PreXion3D Explorer EX is intended to produce two-dimensional digital x-ray images including panoramic image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year-old. Use of equipment and exposure settings designed for adults of average size can result in excessive radiation exposure for a smaller patient. Studies have shown that pediatric patients may be more radiosensitive than adults (i.e., the cancer risk per unit dose of ionizing radiation is higher), and so unnecessary radiation exposure is of particular concern for pediatric patients.
#### 7. Device Description
<Summary of Predicate Device (PreXion3D Explorer)>
PreXion3D Explorer (Predicate Device) consists of a scanner, which is used for generating X-ray and detecting image data, and a Console, which is used for operating the scanner and managing the data. The scan data acquired by the scanner will be transferred to the Console. PreXion3D Explorer Image Analysis System will then perform the image analysis (2D/3D) or image edition (creating cross-section diagram, etc.), and output the image to a printer.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for PreXion. The logo features the word "PreXion" in a stylized font, with the "Pre" in a lighter gray and the "Xion" in a darker gray. Below the word "PreXion" is the tagline "make IT visible" in a smaller font. Above the word "PreXion" is a blue swoosh design, with a gray shadow behind it.
X-ray image data is acquired while the rotation arm is rotating around the secured "patient's head" at a constant speed. X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm). The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan), the tomographic image (CT scan, CT-Panoramic scan) and Cephalometric Scan
<Summary of Modification>
The Primary intention of this Traditional 510(k) submission is to reflect the customers' voice (which does not affect to product safety and performance) to the predicate device and to expand the line-up of PreXion3D Explorer Family.
#### Modifications:
- 1. To apply Chinrest/Forehead Holder which are used in PreXion3D Excelsior (K181983)
- To add a touch panel and a mirror for ease of use which is used in PreXion3D Excelsior 2. (K181983)
- 3. To modify middle size FOV: From "15 x 8" to "15 x 10"
- 4. To add middle size FOV: 10 x 10
- 5. To add Panoramic Scan Function which is used in PreXion3D Excelsior (K181983)
- To change detector's effective area (smaller by 1.5mm due to vendor design change) 6.
- 7. To disable Cephalometric Scan Function by replacing the console software. The console software is modified to do so.
As "Table-6A: Comparison of Characteristics" shows that PreXion3D Explorer EX (Device in concern) is the equivalent with the predicate device, PreXion3D Explorer (K190320), and it appears to be PreXion3D Explorer without Cephalometric Scan Function.
<Summary of Device Description in Concern>
PreXion3D Explorer EX consists of a scanner, which is used for generating X-ray and detecting image data, and a Console, which is used for operating the scanner and managing the data. The scan data acquired by the scanner will be transferred to the Console. PreXion3D Explorer EX Image Analysis System will then perform the image analysis (2D/3D) or image edition (creating cross-section diagram, etc.), and output the image to a printer.
X-ray image data is acquired while the rotation arm is rotating around the secured "patient's head" at a constant speed. X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm). The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan) and the tomographic image (CT scan, CT-Panoramic scan, Panoramic Scan).
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Prexion. The word "Prexion" is written in a cursive font, with the "P" being larger than the other letters. Below the word "Prexion" is the phrase "make IT visible" in a smaller, sans-serif font. Above and to the right of the word "Prexion" is a blue and white swoosh design.
#### Substantial Equivalence Discussion 8.
The following table compares the PreXion3D Explorer EX to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| | Subject Device | Predicate Device | Reference Device | |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Manufacturer | PreXion Corporation | PreXion Corporation | PreXion Corporation | Comparis |
| Trade Name | PreXion3D Explorer<br>EX | PreXion3D Explorer | PreXion3D Excelsior | on to<br>Predicate |
| 510(k) Number | K193329/S001 | K190320 | K181983 | N/A |
| Product Code | OAS | OAS | OAS | Same |
| Regulation Number | OAS: 21 CFR<br>892.1750 | OAS: 21 CFR 892.1750 | OAS: 21 CFR 892.1750 | Same |
| Regulation Name | OAS: Computed<br>tomography x-ray<br>system | OAS: Computed<br>tomography x-ray system | OAS: Computed<br>tomography x-ray<br>system | Same |
| Device Classification<br>Name | X-Ray, Tomography,<br>Computed, Dental | X-Ray, Tomography,<br>Computed, Dental | X-Ray, Tomography,<br>Computed, Dental | Same |
| Indications for use: | PreXion3D Explorer<br>EX is intended to<br>produce two-<br>dimensional digital x-<br>ray images including<br>panoramic image, and<br>three-dimensional<br>digital x-ray images of<br>the dental, oral,<br>maxillofacial region,<br>ENT (Ear, Nose and<br>Throat) and neck region<br>at the direction of<br>healthcare professionals<br>as diagnostic support<br>for adult and pediatric<br>patients.<br>This device is not<br>intended for use on<br>patients less than<br>approximately 21 kg<br>(46 lb) in weight and<br>113 cm (44.5 in) in<br>height; these height and | PreXion3D Explorer is<br>intended to produce two-<br>dimensional digital x-ray<br>images including<br>panoramic and<br>cephalometric image, and<br>three-dimensional digital<br>x-ray images of the dental,<br>oral, maxillofacial region,<br>ENT (Ear, Nose and<br>Throat) and neck region at<br>the direction of healthcare<br>professionals as diagnostic<br>support for adult and<br>pediatric patients.<br>Cephalometric imaging<br>also includes the hand and<br>wrist to obtain carpus<br>images for growth and<br>maturity assessment. | PreXion3D Excelsior is<br>intended to produce two<br>dimensional digital<br>panoramic and<br>cephalometric images,<br>and three-dimensional<br>digital x-ray images of<br>the dental(oral).<br>maxillofacial, and ENT<br>(Ear, Nose and Throat)<br>region at the direction<br>of healthcare<br>professionals as<br>diagnostic support for<br>adult and pediatric<br>patients. Cephalometric<br>imaging also includes<br>the hand and wrist to<br>obtain carpus images<br>for growth and maturity<br>assessment | Cephalome<br>tric<br>Scan<br>Function is<br>disabled by<br>the console<br>software.<br>Otherwise<br>equivalent |
#### Table-6A: Comparison of Characteristics
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for PreXion. The logo features the name "PreXion" in a light gray, elegant font. Below the name is the tagline "make IT visible" in a smaller, sans-serif font, with "IT" in a slightly darker shade. Above and to the right of the name is a stylized graphic consisting of two curved shapes in shades of blue and gray, creating a sense of depth and movement.
| | | | Patient/User Characteristics | | | |
|----------------------------------------------|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Target Population | | | Children aged 6 (except infants) to elderly | Children aged 6 (except infants) to elderly | Children aged 6 (except infants) to elderly | Same |
| Anatomical Site | | | The dental, oral, maxillofacial region ENT (Ear, Nose and Throat) and neck region | The dental, oral, maxillofacial region ENT (Ear, Nose and Throat) and neck region | The dental, oral, maxillofacial region ENT (Ear, Nose and Throat) and neck region | Same |
| | | | | | | |
| Users | | | Health care professionals | Health care professionals | Health care professionals | Same |
| | | | Technological Characteristics and Performance | | | |
| Patient Contact | | | CHIN REST:<br>polycarbonate<br>Forehead Holder:<br>silicone rubber<br>HANDLE GRIP:<br>silicone rubber | CHINREST:<br>polycarbonate<br>HEADREST: carbon<br>BELT: polyvinyl chloride | CHIN REST:<br>polycarbonate<br>Forehead Holder:<br>silicone rubber<br>HANDLE GRIP:<br>silicone rubber | Same materials are used |
| Material | | | | | | |
| Sterility | | | Non-sterile | Non-sterile | Non-sterile | Same |
| X-ray<br>Generation<br>Device | Tube<br>Voltage | | 90-110KV | 90-110KV | 60-110KV | Same or Equivalent |
| | Pulse<br>Exposure<br>function | | Yes | Yes | Yes | Same |
| | Tube<br>Current | | 1-5mA | 1-3mA | 1-6mA | Tube Current is similar.<br>Added according with Panoramic |
| | Focal<br>Spot Size | 0.3mm x 0.3mm | 0.3mm x 0.3mm | 0.3mm x 0.3mm | Same | |
| | Detector | FPD (TFT) | FPD (TFT) | FPD (TFT) | Same | |
| | Pixel<br>Size | 248 µm x248µm (With<br>binning)<br>(CT, CT-Panoramic,<br>Panoramic)<br>124 µm x124µm<br>(Without binning)<br>(CT, CT-Panoramic,<br>Panoramic) | 248 µm x248µm (With<br>binning)<br>(CT, CT-Panoramic,<br>Ceph)<br>124 µm x124µm (Without<br>binning)<br>(CT, CT-Panoramic,<br>Ceph) | 250 µm x250µm (With<br>binning)<br>(CT, CT-Panoramic.<br>Panoramic)<br>125 µm x125µm<br>(Without binning)<br>(CT, CT-Panoramic)<br>125µm x 125µm<br>(Panoramic)<br>140 µm x 140µm<br>(Ceph) | Equivalent | |
| X-ray<br>Image<br>Capturin<br>g Device | Pixel<br>Number | 1024x1280(With<br>binning)<br>(CT, CT-Panoramic)<br>2560x2048(Without<br>binning)<br>(CT, CT-Panoramic)<br>1900 x 120<br>(Panoramic) | 1024x1280(With binning)<br>(CT, CT-Panoramic)<br>2560x2048(Without<br>binning)<br>(CT, CT-Panoramic,<br>Ceph)<br>1024x1280(With binning) | 1024x1280 (CT)<br>128x128 (Panoramic)<br>2112x1754<br>(Cephalometric) | Comparing<br>to<br>K190320,<br>Device<br>is | |
| | Size of<br>Area<br>Receivin<br>g X-ray | 253.95mm x 317.44mm<br>(CT, CT-Panoramic)<br>230mm x 15mm<br>(Panoramic) | 253.95mm x 317.44mm<br>(CT, CT-Panoramic,<br>Ceph) | 160mm x 128mm (CT)<br>160mm x 12.5mm<br>(Panoramic)<br>295.68 x 245.56mm<br>(Ceph) | same, just<br>the use is<br>changed so<br>that Ceph is<br>disabled<br>and<br>Panoramic<br>Scan<br>is<br>added. | |
| | Number<br>of Bits | 16bits (CT, CT-<br>Panoramic, Panoramic) | 16bits (CT, CT-<br>Panoramic, Ceph) | 16bits (CT, CT-<br>Panoramic, Panoramic)<br>14bits (Ceph) | | |
| Scanner | SID/SO<br>D | 700mm/ 420mm (CT,<br>CT-Panoramic.<br>Panoramic) | 700mm/ 420mm (CT, CT-<br>Panoramic, Ceph) | 700mm/ 470mm (CT,<br>Panoramic)<br>1735mm / 1500mm<br>(Ceph) | Comparing<br>to<br>K190320, Device is<br>same, just<br>the use is<br>changed so<br>that Ceph is<br>disabled<br>and<br>Panoramic<br>Scan<br>is<br>added. | |
| | Dimensi<br>on<br>(WxDxH) | 880 mm x 1237 mm x<br>2268 mm (CT, CT-<br>Panoramic, Panoramic) | 880 mm x 1237 mm x<br>2268 mm (CT, CT-<br>Panoramic, Ceph) | 930 mm x 1230 mm x<br>2220 mm (CT, CT-<br>Panoramic, Panoramic)<br>1816 mm x 1230 mm x | K190320, geometry<br>is same | |
| | | | | 2220 mm (with Ceph) | | |
| | Weight | 185 kg (CT, CT-<br>Panoramic, Panoramic) | 165 kg (CT, Panoramic,<br>Ceph) | 165 kg (CT, Panoramic)<br>200kg (Ceph) | Similar | |
| Imaging Mode | | CT scan, CT-<br>Panoramic, Panoramic<br>scan | CT scan, CT-Panoramic,<br>Cephalometric<br>radiography | CT scan, CT-<br>Panoramic, Panoramic<br>scan,<br>Cephalometric<br>radiography | No<br>Cephalome<br>tric<br>Radiograph<br>y<br>Added<br>Panoramic<br>scan<br>(Identical<br>to<br>K181983) | |
| Panoramic Scan<br>Performance (Scan<br>Time) | | 8-16sec | 18sec | 8-16sec | Scan Time<br>is<br>improved<br>or same. | |
| Cephalometric<br>Radiography (Scan<br>Time) | | N/A | 0.16 sec | 0.5-0.8 sec | No<br>Cephalome<br>tric<br>Radiograph<br>y | |
| CT Scan<br>Performa<br>nce | Scan<br>Time | 10-20sec | 10-20sec | 5.2-23.6sec | Same or<br>Similar | |
| | FOV<br>(Voxel<br>Size) | Diameter 150mm x<br>H156mm (0.100 -<br>0.200mm)<br><br>Diameter 150mm x<br>H100mm (0.100 -<br>0.200mm) | Diameter 150mm x<br>H156mm (0.100 -<br>0.200mm)<br><br>Diameter 150mm x<br>H78mm (0.100 -<br>0.200mm) | Diameter 150mm x<br>H130mm<br>(0.100 - 0.200mm)<br><br>Diameter 150mm x H81<br>(0.200mm)<br><br>Diameter 100mm x<br>H81mm (0.100 -<br>0.200mm)<br> | Added one<br>middle size<br>FOV and<br>gets bigger<br>in middle<br>size FOV<br>for user<br>requiremen<br>t. | |
| | | Diameter 100mm x<br>H100mm (0.100 -<br>0.200mm)<br><br>Diameter 50mm x<br>H50mm<br>(0.100 - 0.200mm) | Diameter 50mm x H50mm<br>(0.100 - 0.200mm) | Diameter 100mm x<br>H50mm (0.100 -<br>0.200mm)<br><br>Diameter 50mm x<br>H50mm (0.100 -<br>0.200mm) | | |
| | | | | | | |
| Spatial Resolution | | Standard mode:<br>50%MTF: 0.95LP/mm<br>20%MTF:2.0LP/mm<br><br>Rapid Mode:<br>50%MTF: 0.95LP/mm<br>20%MTF:2.0LP/mm | Standard mode:<br>50%MTF: 0.95LP/mm<br>20%MTF:2.0LP/mm<br><br>Rapid Mode:<br>50%MTF: 0.95LP/mm<br>20%MTF:2.0LP/mm | Standard mode<br>50%MTF: 1.2LP/mm<br>20%MTF:2.3LP/mm<br><br>Rapid Mode<br>50%MTF: 1.2LP/mm<br>20%MTF:2.3LP/mm | Same or<br>equivalent | |
| | | | | | | |
| | | High Resolution mode: | High Resolution mode: | High Resolution mode: | | |
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for PreXion. The logo features the name "PreXion" in a light gray, sans-serif font. Below the name is the tagline "make IT visible" in a smaller font, with the "IT" in blue. Above and to the right of the name is a blue and gray swoosh design, adding a modern and dynamic element to the logo.
{10}------------------------------------------------
Image /page/9/Picture/1 description: The image contains the logo for PreXion. The logo features the company name in a stylized font, with the "IT" in "make IT visible" in blue. Above and to the right of the name is a blue and white swoosh design, adding a modern and dynamic element to the overall branding.
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for PreXion. The logo features the name "PreXion" in a light gray, sans-serif font. Below the name is the tagline "make IT visible" in a smaller font. Above and to the right of the name is a blue and gray swoosh design.
| | 50%MTF: 0.95LP/mm<br>20%MTF:2.0LP/mm | 50%MTF: 0.95LP/mm<br>20%MTF:2.0LP/mm | 50%MTF: 1.2LP/mm<br>20%MTF:2.3LP/mm | |
|--------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| | High contrast mode:<br>50%MTF: 0.95LP/mm<br>20%MTF:2.0LP/mm | High contrast mode:<br>50%MTF: 0.95LP/mm<br>20%MTF:2.0LP/mm | High contrast mode:<br>50%MTF: 1.2LP/mm<br>20%MTF:2.3LP/mm | |
| Viewer Software<br>(Image Analysis<br>System Software) | Display High-<br>resolution 2D and 3D<br>Images Function<br><br>Image Processing<br>Function includes<br>Airway measurement<br><br>Image Operation<br>Function<br><br>Output Function | Display High-resolution<br>2D and 3D Images<br>Function<br><br>Image Processing Function<br>include Airway<br>measurement<br><br>Image Operation Function<br><br>Output Function | Display High-resolution<br>2D and 3D Images<br>Function<br><br>Image Processing<br>Function include Airway<br>measurement<br><br>Image Operation<br>Function<br><br>Output Function | Same |
| Console Software<br>System Settings | CT Scan<br>include CT-Panoramic<br>mode | CT Scan<br>include CT-Panoramic<br>mode | CT Scan<br>include CT-Panoramic<br>mode | Same |
| Applied Standard | | | | |
| Electrical Safety<br>Standard | ANSI/AAMI ES60601-1 | ANSI/AAMI ES60601-1 | ANSI/AAMI ES60601-1 | |
| Electromagnetic<br>Compatibility<br>Standard | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | |
| Radiation Safety<br>Standard | IEC 60601-1-3 | IEC 60601-1-3 | IEC 60601-1-3 | |
| Electrical<br>Equipment Usability<br>Safety Standard | IEC 60601-1-6 | IEC 60601-1-6 | IEC 60601-1-6 | |
| Usability<br>Engineering<br>Standard | IEC 62366 -1 | IEC 62366 | IEC 62366 | |
| Software Lifecycle<br>Process Standard | IEC 62304 | IEC 62304 | IEC 62304 | Same |
| Essential<br>performance of<br>dental extra-oral X-<br>ray equipment<br>Standard | IEC 60601-2-63 | IEC 60601-2-63 | IEC 60601-2-63 | |
| Acceptance tests of<br>Imaging<br>performance of<br>dental X-ray<br>equipment Standard | IEC 61223-3-4 | IEC 61223-3-4 | IEC 61223-3-4 | |
| Acceptance tests of<br>Imaging<br>performance of<br>computed<br>tomography X-ray<br>equipment Standard | IEC 61223-3-5 | IEC 61223-3-5 | IEC 61223-3-5 | |
| Laser Safety<br>Standard | IEC 60825-1 | IEC 60825-1 | IEC 60825-1 | |
| Risk Management<br>Standard | ISO 14971 | ISO 14971 | ISO 14971 | |
| DICOM Standard | NEMA PS 3.1 - 3.20 | NEMA PS 3.1 - 3.20 | NEMA PS 3.1 - 3.20 | |
| Biocompatibility<br>Standard<br>Compliance | ISO 10993-1 | ISO 10993-1 | ISO 10993-1 | |
| Biocompatibility<br>Standard<br>Compliance | ISO 10993-5 | ISO 10993-5 | ISO 10993-5 | |
| Biocompatibility<br>Standard<br>Compliance | ISO 10993-10 | ISO 10993-10 | ISO 10993-10 | |
{12}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Prexion, a company that aims to make IT visible. The logo features the company name in a stylized font, with "Pre" in a cursive-like font and "Xion" in a more standard font. Below the name is the tagline "make IT visible" in a smaller font. A blue swoosh design arches over the "Xion" part of the name, adding a modern and technological feel to the logo.
#### 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of PreXion3D Explorer EX and in showing substantial equivalence to the predicate device, PreXion Corporation completed a number of non-clinical performance tests. The PreXion3D Explorer EX meets all the requirements for overall design, biocompatibility, performance, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
Summary of performance testing performed for each modification stated in Section 7:
- 1. To apply Chinrest/Forehead Holder which are used in PreXion3D Excelsior (K181983)
The biocompatibility is safe because same materials as to marketed device are used.
- 2. To add a touch panel and a mirror for ease of use which is used in PreXion3D Excelsior (K181983)
IEC60601-1 test was performed by the outside test lab qualified with ISO17025. Regarding added mirror, risk management was performed and decided it's acceptable.
- 3. To modify middle size FOV: From "15 x 8" to "15 x 10"
System test was performed to confirm that 15 x 10 of middle size FOV is valid.
- 4. To add middle size FOV: 10 x 10
System test was performed to confirm that 10 x 10 of middle size FOV is valid.
- 5. To add Panoramic Scan Function which is used in PreXion3D Excelsior (K181983)
Imaging performance test was performed and has confirmed to comply with IEC61223-3-4.
- 6. To change detector's effective area (smaller by 1.5mm due to vendor design change)
An American board-certified radiologist has confirmed that images with the modification as a change to the detector's effective area is of diagnostic quality.
{13}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for PreXion, a company that appears to be in the IT sector. The logo features the company name in a stylized font, with the "Pre" in a lighter gray and "Xion" in a darker gray. Below the company name is the tagline "make IT visible" in a smaller, sans-serif font. Above and to the right of the company name is a swooping, curved graphic in shades of blue and white, adding a modern and dynamic element to the logo.
7. To disable Cephalometric Scan Function by replacing the console software. The console software is modified to do so.
System test was performed to confirm that system is valid including the function of disabling Cephalometric Scan Function.
The PreXion3D Explorer EX passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
- Biocompatibility testing per ISO 10993-1, ISO 10993-5 and ISO 10993-10 .
- Electrical safety testing per ABSI/AAMI ES 60601-1, IEC 60601-1-3 and IEC 60601-1-6
- . Electromagnetic Safety testing per IEC 60601-1-2
- . Dental extra-oral X-ray equipment testing per IEC 60601-2-63
- . Acceptance testing of X-ray equipment per IEC 61223-3-4 and IEC 61223-3-5
- Laser safety testing per IEC 60825-1 ●
- . DICOM testing per NEMA PS 3.1 - 3.20
- . Usability testing per IEC 62366-1
- . Software verification and validation per IEC 62304
- Storage and Transport Testing per ISO 4180
- Software Documentation per: "Guidance for the Content of Premarket Submissions for Software . Contained in Medical Devices
- Establish the substantial equivalence of an SSXI to a previously cleared conventional radiographic ● SSXI per: "Guidance for the Submission of 510(k) s for Solid State X-ray Imaging Devices"
- Cybersecurity Activities per: "Cybersecurity-for-Networked-Medical-Devices-Containing-Off-the-Shelf-(OTS)-Software---Guidance-for-Industry", "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"
#### 10. Clinical Performance Data
There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
{14}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the logo for PreXion. The logo features the name "PreXion" in a stylized font, with "Pre" in a darker gray and "Xion" in a lighter gray. Below the name is the tagline "make IT visible" in a smaller, sans-serif font. Above and to the right of the name is a curved, blue graphic element with a subtle shadow effect, adding a sense of depth and dimension to the logo.
#### 11. Cybersecurity Activities
We make post-market Cybersecurity actions as follows:
- 1…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.