The Disposable Radiofrequency Cannulas, in combination with approved compatible generators and electrodes, are indicated as an aid in the management of pain in the peripheral nervous system. Examples include facet denervation, trigeminal rhizotomy, and related functional neurosurgical procedures. These are single-use-only products.
Device Story
Disposable RF Cannula (needle) used in radiofrequency ablation (RFA) to treat peripheral nerve pain. Device consists of hollow needle shaft, hub, and stylet. Clinician inserts cannula parallel to target nerve; removes stylet; inserts RF electrode through cannula. RF generator delivers energy through active tip to create heat lesions on nerves, interrupting pain signals. Dispersive grounding pad provides return path. Used in clinical settings for functional neurosurgical procedures. Benefits include targeted pain relief via minimally invasive access. Subject device is a product family (Classic, Select, SMK) varying in materials, dimensions, and gauge.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601), performance verification (ISO 7864, 9626, 80369-7, 6009), sterilization (ISO 11135), and packaging/shelf-life (ASTM F2825, D4169).
Technological Characteristics
Hollow needle with insulation and exposed active tip. Materials vary by model (Classic, Select, SMK). Standards: ISO 7864, ISO 9626, ISO 80369-7, ISO 6009. Electrical safety per IEC 60601-1, 60601-1-2, 60601-2-2. Sterile (ethylene oxide, SAL 10^-6). Single-use.
Indications for Use
Indicated for patients requiring management of peripheral nervous system pain, including procedures such as facet denervation and trigeminal rhizotomy. For use in combination with compatible RF generators and electrodes.
Regulatory Classification
Identification
A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 16, 2026
Abbott Medical
Katee Perez Anderson
Regulatory Affairs Specialist
6901 Preston Rd.
Plano, Texas 75024
Re: K253907
Trade/Device Name: Disposable Radiofrequency Cannula
Regulation Number: 21 CFR 882.4725
Regulation Name: Radiofrequency Lesion Probe
Regulatory Class: Class II
Product Code: GXI
Dated: December 5, 2025
Received: December 5, 2025
Dear Katee Perez Anderson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K253907 - Katee Perez Anderson
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253907 - Katee Perez Anderson
Page 3
Sincerely,
JULIA E. SLOCOMB -S
Digitally signed by JULIA E. SLOCOMB -S
Date: 2026.04.16 14:37:43 -04'00"
for Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Disposable Radiofrequency Cannula
Page 12 of 43
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253907 | ? |
| Please provide the device trade name(s). | | ? |
| Disposable Radiofrequency Cannula | | |
| Please provide your Indications for Use below. | | ? |
| The Disposable Radiofrequency Cannulas, in combination with approved compatible generators and electrodes, are indicated as an aid in the management of pain in the peripheral nervous system. Examples include facet denervation, trigeminal rhizotomy, and related functional neurosurgical procedures. These are single-use-only products. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253907
# 510(k) Summary
510(k) Number: K253907
510(k) Type: Traditional 510(k)
Date Prepared: December 5, 2025
Manufacturer Name & Address: Abbott Medical
5050 Nathan Lane North
Plymouth, MN, 55442 USA
Primary Correspondent: Katee Perez Anderson
Regulatory Affairs Specialist
(945) 256 9915
katee.perez@abbott.com
Secondary Correspondent: Jenna Roycraft
Associate Director Regulatory Affairs
(972) 526 4645
jenna.roycraft@abbott.com
## Device Information
Trade Name: Disposable Radiofrequency Cannula
Common Name: Radiofrequency lesion probe
Class: II
Classification Name: 882.4725 Radiofrequency lesion probe
Product Code: GXI
Predicate Device 510(k)s: K090955 (Smith & Nephew RF Cannulae) (Primary), K000073 (Mercury Medical Disposable Radiofrequency Cannula), K042375 ([Technomed] Disposable Cannula),
Reference Device 510(k): K242841 (Disposable RF Electrode)
Device Description: Abbott Medical Radiofrequency (RF) Cannulas, also called needles, are part of the Radiofrequency Ablation (RFA) system. The RFA System is used to create heat lesions on nerves by interrupting the pain signals generated from the target nerve. The Disposable RF Cannula is a small hollow needle, with a layer of insulation and an exposed, sharp tip. During clinical use, the Cannula is inserted parallel to the targeted nerve, the stylet
Page 1 of 3
{5}
K253907
is removed and discarded. An electrode, or probe, connected to the RF generator is inserted through the Cannula, which sends radiofrequency waves through the uninsulated portion, or active tip, to the targeted nerve which lesions the targeted nerve. A dispersive grounding pad provides a return path for the applied RF current.
The RF Cannulas consists of a stylet, stylet cap, hub, and needle shaft. The subject device is a product family made of three model types; the Classic, Select and SMK Cannulas; which differ in hub material, hub design and insulation material, and vary in shaft length, active tip length, and gauge size. The RF Cannula is provided as sterile (using ethylene oxide) and is a single-use product.
**Indications for Use:**
The Disposable Radiofrequency Cannulas, in combination with approved compatible generators and electrodes, are indicated as an aid in the management of pain in the peripheral nervous system. Examples include facet denervation, trigeminal rhizotomy, and related functional neurosurgical procedures. These are single-use-only products.
**Comparison of Characteristics with the Predicate Devices**
Abbott Medical is submitting this traditional 510(k) seeking market clearance for the Disposable RF Cannula including cumulative design changes that have been made since latest clearances (K090955, K000073, K042375) and proposed changes that are being made to the device. The subject device and predicate devices have different indication for use but the same intended use, which is for lesioning of neural tissue in the nervous system by allowing access to the targeted nerve site and directing RF energy to the targeted nerve site. The difference in indications for use does not change the intended use. The indications for use has been updated to align with the indications for use statement of the compatible IonicRF Generator and the Disposable RF Electrode, since the subject device must be used together with these devices in order to treat patients with neural pain.
The subject device and predicate device have similar technological characteristics. Any differences between the subject device and predicate devices do not raise different questions of safety and effectiveness as confirmed by performance testing, biocompatibility testing, sterilization validation, packaging validation, and shelf-life validation, which is also comparable to the reference device. Therefore, Abbott has determined the Disposable RF Cannula to be substantially equivalent to the predicate devices.
**Non-Clinical Testing Summary**
Non-clinical testing was performed to confirm that the physical, electrical, and biological properties of the Disposable RF Cannulas met all the performance requirements and specifications. The following testing was provided in support of the substantial equivalence determination:
- Biocompatibility: The Disposable RF Cannula was evaluated per ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 10993-12, ISO 10993-18, and ISO 10993-23. Testing included cytotoxicity, sensitization, irritation, and acute systemic toxicity.
- EMC and Electrical Safety: The Disposable RF Cannula was evaluated per IEC 60601-1, 60601-1-2, and 60601-2-2. Testing included assessment of the electrical safety of the device.
Page 2 of 3
{6}
K253907
- Performance Testing: The Disposable RF Cannula was tested to verify the subject device meets product requirements for performance. Applicable standards include ISO 7864, ISO 9626, ISO 80369-7, and ISO 6009.
- Sterilization: The Disposable RF Cannula was tested to verify the subject device meets SAL 10⁻⁶ according to ISO 11135.
- Packaging and Shelf-Life: The Disposable RF Cannula packaging was tested to verify the packaging meets product requirements and maintains its integrity during conditions normally encountered during sterilization, handling, shipping, and storage. To verify the shelf-life and package integrity, the effects of aging on the device packaging were evaluated through accelerated aging testing. Applicable standards include ASTM F2825, and ASTM D4169.
## Clinical Testing
Based on substantial equivalence to the predicate device and results of testing, clinical studies were not required to establish substantial equivalence of the Disposable RF Cannula.
## Substantial Equivalence Conclusion
Based on comparison of intended use and technical characteristics, the Disposable RF Cannula is substantially equivalent to the predicate devices. The results of testing confirm that the subject device meets all requirements and specifications. Any differences in the subject and predicate devices do not raise new types of questions of safety and effectiveness. Therefore, the Disposable RF Cannula is substantially equivalent to the predicate devices.
Page 3 of 3
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.