Powder Free Nitrile Examination Gloves Non-Sterile Black
K253824 · Kanglongda Vietnam Protection Technology Company Limited · LZA · Jun 18, 2026 · General Hospital
Device Facts
Record ID
K253824
Device Name
Powder Free Nitrile Examination Gloves Non-Sterile Black
Applicant
Kanglongda Vietnam Protection Technology Company Limited
Product Code
LZA · General Hospital
Decision Date
Jun 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Powder Free Nitrile Examination Gloves Non-Sterile Black is a disposable device intended for medical purposes that is worn on an examiner's hand to prevent the contamination between patient and examiner
Device Story
Disposable nitrile examination glove; provides single-use barrier protection for wearer. Used in clinical settings by healthcare examiners. Device acts as physical barrier to prevent cross-contamination between patient and examiner. Manufactured from nitrile compound; black color. Complies with ASTM D6319-19 standards for physical properties, dimensions, and barrier integrity. Device is non-sterile and powder-free.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM D6319-19 (dimensions, tensile strength, elongation), ASTM D5151-19 (freedom from holes), and ASTM D6124-06 (residual powder). Biocompatibility testing performed per ISO 10993-23 (skin irritation), ISO 10993-10 (skin sensitization), and ISO 10993-11 (acute systemic toxicity). All tests passed.
Technological Characteristics
Nitrile compound; powder-free; non-sterile. Dimensions and physical properties conform to ASTM D6319-19. Biocompatibility per ISO 10993-23, ISO 10993-10, and ISO 10993-11. Available in sizes XS through XXL.
Indications for Use
Indicated for use as a nonsterile, disposable, patient examination glove worn on the examiner's hands to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Disposable Medical Examination Nitrile Gloves (K232008)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
Kanglongda Vietnam Protection Technology Company Limited
Thamira Ranasinghe
Vice President - Technical and Sales
Lot Cn 05, Viglacera - Phong Dien Industrial Park
Phong Dinh Ward
Hue City, 530000
Vietnam
Re: K253824
Trade/Device Name: Powder Free Nitrile Examination Gloves Non-Sterile Black
Regulation Number: 21 CFR 880.6250
Regulation Name: Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved
Product Code: LZA
Dated: May 19, 2026
Received: May 19, 2026
Dear Thamira Ranasinghe:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253824 - Thamira Ranasinghe
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253824 - Thamira Ranasinghe
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection
Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253824
Device Name
Powder Free Nitrile Examination Gloves Non-Sterile Black
Indications for Use (Describe)
Powder Free Nitrile Examination Gloves Non-Sterile Black is a disposable device intended for medical purposes that is worn on an examiner's hand to prevent the contamination between patient and examiner
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
康隆达防护
KANG LONG DA PROTECTION
# 510(k) SUMMARY (K253824)
**0.0 Summary Preparation Date:** 2026-06-18
**1.0 Applicant Information**
- Name: Kanglongda Vietnam Protection Technology Company Limited
- Address: Lot CN 05, Viglacera - Phong Dien Industrial Park, Phong Dinh Ward, Hue City, Vietnam
- Establishment Registration Number: 3024897455
**2.0 CORRESPONDENT**
- Contact Name: Thamira Ranasinghe
- Contact Title: Vice President -Technical
- Phone Number: +84788611640
- Email: thamira@kanglongda.com.cn
**3.0 Name of the Device**
- Proprietary/Trade name: Powder Free Nitrile Examination Gloves Non-Sterile Black
- Common Name: Nitrile Examination Gloves
- Classification Name: Non-powdered Patient Examination Glove
- Device Classification: Class I
- Regulation Number: 21 CFR 880.6250
- Product Code: LZA
- 510(K) Number: K253824
**4.0 Predicate**
- Device Name: Disposable Medical Examination Nitrile Gloves
- Company name: Raxwell Industries LLC
- Device Classification: Class I
- Regulation Number: 21 CFR 880.6250
- Product Code: LZA
- 510(K) Number: K232008
**5.0 Device Description:**
"Powder Free Nitrile Examination Gloves Non-Sterile Black" is a patient examination glove made from nitrile compound, nonsterile (as per 21 CFR 880.6250, Class I). The principal operation of this medical device is to provide single use barrier protection for the wearer and the device meets the specifications for barrier protection and physical
1
{5}
**康隆达防护**^{}[] KANG LONG DA PROTECTION
properties as defined in ASTM D6319-19, Standard specification for Nitrile Examination Gloves.
### 6.0 Device Indications for Use:
“Powder Free Nitrile Examination Gloves Non-Sterile Black” is a nonsterile disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner.
### 7.0 Comparison of Device Technological Characteristics:
| Device Characteristic | Predicate Device | Subject Device | Comparison |
| --- | --- | --- | --- |
| Product Name | Disposable Medical Examination Nitrile Gloves | Powder Free Nitrile Examination Gloves Non-Sterile Black | similar |
| 510(K) No. | K232008 | K253824 | n/a |
| Applicant | Raxwell Industries LLC | Kanglongda Vietnam Protection Technology Company Limited | Different |
| Product Code | LZA | LZA | Same |
| Regulation | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Class | 1 | 1 | Same |
2
{6}
**康隆达防护**^{}[] KANG LONG DA PROTECTION
| Indications for Use | The Disposable Medical Examination Nitrile Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | 'Powder Free Nitrile Examination Gloves Non-Sterile Black' is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same |
| --- | --- | --- | --- |
| Powder free | Yes | Yes | Same |
| Material | Nitrile | Nitrile | Same |
| Colorant | Black | Black | Same |
| Size | XS/S/M/L/XL/XXL | XS/S/M/L/XL/XXL | Same |
| Single Use | Yes | Yes | Same |
| Non-Sterile | Yes | Yes | Same |
| Dimensions - Length | Complies with ASTM D6319-19 XS/S: ≥220; M/L/XL/XXL: ≥230; | Complies with ASTM D6319-19 XS/S: ≥220; M/L/XL/XXL: ≥230; | Same |
3
{7}
**康隆达防护**^{}[] KANG LONG DA PROTECTION
| Dimensions - Palm Width | Complies with ASTM D6319-19 XS: 70 ± 10 mm; S: 80 ± 10 mm; M: 95 ± 10 mm; L: 110 ± 10 mm; XL: 120 ± 10 mm; XXL:130± 10 mm; | Complies with ASTM D6319-19 XS: 70 ± 10 mm; S: 80 ± 10 mm; M: 95 ± 10 mm; L: 110 ± 10 mm; XL: 120 ± 10 mm; XXL:130± 10 mm; | Same |
| --- | --- | --- | --- |
| Dimensions - Thickness | Complies with ASTM D6319-19 Palm ≥0.05 mm; Finger ≥0.05 mm | Complies with ASTM D6319-19 Palm ≥0.05 mm; Finger ≥0.05 mm | Same |
| Physical Properties - Tensile Strength | Complies with ASTM D6319-19 Before aging ≥14MPa After aging ≥14MPa | Complies with ASTM D6319-19 Before aging ≥14MPa After aging ≥14MPa | Same |
| Physical Properties - Elongation | Complies with ASTM D6319-19 Elongation: Before Aging: ≥ 500% After Aging: ≥400% | Complies with ASTM D6319-19 Elongation: Before Aging: ≥ 500% After Aging: ≥400% | Same |
| Residual powder | Complies with ASTM D6319-19 < 2.0mg per glove | Complies with ASTM D6319-19 < 2.0mg per glove | Same |
| Freedom from Holes | In accordance with ASTM D6319-19 (G-1 with AQL 2.5) | In accordance with ASTM D6319-19 (G-1 with AQL 2.5) | Same |
4
{8}
**康隆达防护**^{}[] KANG LONG DA PROTECTION
| Biocompatibility | ISO 10993-10, under the conditions of the study, the device is not a sensitizer | ISO 10993-10, under the conditions of the study, the device is not a sensitizer | Same |
| --- | --- | --- | --- |
| | ISO 10993-10, under the conditions of the study, the device is not an irritant | ISO 10993-23, under the conditions of the study, the device is not an irritant | Similar* |
| | ISO 10993-5, under conditions of the study, device extract is cytotoxic | Not test, addressed by alternative testing | Different – Evaluation results demonstrate conformity with the requirements for acute systemic toxicity. |
| | ISO 10993-11, under the conditions of the study, the device showed no evidence of acute systemic toxicity | ISO 10993-11, under the conditions of the study, the device showed no evidence of acute systemic toxicity | Same |
*Note 1: The skin irritation test was conducted in accordance with the latest applicable standard, ISO 10993-23, which supersedes and refines the methodology previously found in ISO 10993-10. The test is equivalent in its objective and acceptance criteria, and the passing result confirms the device meets the required safety profile for skin irritation.
5
{9}
**康隆达防护**^{}[] KANG LONG DA PROTECTION
## 8.0 Assessment of Non-Clinical Performance Data:
Non-clinical testing was performed to verify that the subject devices meet the acceptance criteria of the performance test and all design specifications. The test results demonstrated that the subject device complies with the following standards as shown
| Test | Purpose | Criteria | Result |
| --- | --- | --- | --- |
| Standard Test Method for Detection of Holes in Medical Gloves ASTM D5151-19 (2023) | To demonstrate glove integrity. | Freedom from holes. AQL 2.5. | Pass |
| Standard Test Method for Residual Powder on Medical Gloves ASTM D6124-06 (2022) | To demonstrate the gloves are 'powder free'. | Average less than 2 mg per glove. | Pass |
| Dimensional Conformance ASTM D6319-19(2023) | To demonstrate appropriate dimensions for labeled sizes. | Conforms to ASTM D6319 width, thickness, and length requirements for sizes XS, S, M, L, XL and XXL AQL 4 | Pass |
| Tensile Performance ASTM D6319-19(2023) | To demonstrate adequate tensile and elongation properties. | Conforms to ASTM D6319 requirements: - Tensile Strength (Before & After Aging): ≥ 14 MPa - Ultimate Elongation (Before Aging): ≥ 500% - Ultimate Elongation (After Aging): ≥ 400% AQL 4 | Pass |
6
{10}
**康隆达防护**^{}[] KANG LONG DA PROTECTION
| Biocompatibility: Skin Irritation ISO 10993-23 | To demonstrate low potential for skin irritation. | Under the conditions of the study, the device is not an irritant. | Pass |
| --- | --- | --- | --- |
| Biocompatibility: Skin Sensitization ISO 10993-10 | To demonstrate low potential for skin sensitization. | Under the conditions of the study, the device is not a sensitizer. | Pass |
| Biocompatibility: Acute Systemic Toxicity ISO 10993-11 | To demonstrate low acute systemic toxicity. | Under conditions of the study, the device showed no evidence of acute systemic toxicity | Pass |
### **9.0 Assessment of Clinical Performance Data:**
Clinical data is not needed for this type of device.
### **10.0 Conclusion:**
The conclusions drawn from the nonclinical tests demonstrate that the subject device “Powder Free Nitrile Examination Gloves Non-Sterile Black” is as safe, as effective, and performs as well as or better than the legally marketed device, K232008.
7
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.