K253780 · Ge Medical Systems, LLC · LNH · Feb 6, 2026 · Radiology
Device Facts
Record ID
K253780
Device Name
SIGNA™ Bolt
Applicant
Ge Medical Systems, LLC
Product Code
LNH · Radiology
Decision Date
Feb 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Anatomical landmark detection
Deep learning model
Accuracy > 90%
Up to 99%
—
—
80 volunteers in the US and China
—
Spine scan plane prescription
Deep learning models
—
—
—
—
376 subjects from multiple clinical and internal sites
—
Prostate scan plane prescription
Deep learning models
—
—
—
—
785 exams from clinical sites in the US and Europe
—
Indications for Use
The SIGNA™ Bolt system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA™ Bolt system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Device Story
3.0T superconducting MR scanner; 70 cm bore; utilizes pulse sequences and reconstruction algorithms to produce diagnostic images/spectra. Operated by MR technicians/radiologists in clinical settings. Features AI-based tools: AIRx for automated scan plane prescription (spine/prostate) and SIGNA One Camera for automated patient positioning. Contactless physiological gating (cardiac/respiratory) uses sensors to trigger acquisitions without physical attachments. Output images interpreted by physicians for diagnosis. Benefits include reduced scan times, improved workflow, and minimized inter-operator variability via automated landmarking and prescription.
Clinical Evidence
No clinical studies required. Bench testing included verification of software, cybersecurity, and hardware performance. AI models (AIRx, SIGNA One Camera) validated using independent datasets (n=80 for camera; n=376 for spine; n=785 for prostate) against ground truth. Radiologist review of sample images confirmed diagnostic quality.
Technological Characteristics
3.0T superconducting magnet; 70 cm bore; ZRMW water-cooled gradient coil; In-Scan-Room RF transmit architecture (36 kW peak power); 130-162 channel RF receive; DICOM compliant. Standards: IEC 60601-1, IEC 60601-2-33, ISO 10993-1, NEMA MS series. Software includes deep learning models for landmarking and scan prescription.
Indications for Use
Indicated for whole-body diagnostic MR imaging (head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, musculoskeletal) in patients requiring axial, sagittal, coronal, oblique, spectroscopic, or dynamic imaging. Contrast agents may be used.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 6, 2026
GE Medical Systems, LLC
Emily Fang
Regulatory Affairs Leader
3200 N. Grandview Blvd.
Waukesha, Wisconsin 53188
Re: K253780
Trade/Device Name: SIGNA™ Bolt
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH, LNI, MOS
Dated: November 26, 2025
Received: November 26, 2025
Dear Emily Fang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253780 - Emily Fang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253780 - Emily Fang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K253780
Device Name
SIGNA™ Bolt
Indications for Use (Describe)
The SIGNA™ Bolt system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.
The images produced by the SIGNA™ Bolt system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
GE HealthCare
SIGNA™ Bolt
Traditional 510(k) Premarket Notification
K253780
# 510(k) Summary
In according with 21 CFR 807.92 the following summary of information is provided:
| Date | November 26, 2025 |
| --- | --- |
| Submitter | GE Medical Systems, LLC
3200 N. Grandview Blvd.
Waukesha, WI 53188 |
| Primary Contact | Emily Fang
Regulatory Affairs Leader
609-285-9878
signa.bolt.premarket@gehealthcare.com |
| Secondary Contact | Sandra Bahling
Senior Regulatory Affairs Leader
262-720-8872
Sandra.Westphal@gehealthcare.com
Glen Sabin
Regulatory Affairs Director
262-894-4968
Glen.Sabin@gehealthcare.com |
| Device Trade Name | SIGNA™ Bolt |
| Common/Usual Name | MR System |
| Classification Name | Magnetic Resonance Diagnostic Device |
| Regulation Number | 21 CFR 892.1000 |
| Product Code | LNH, LNI, MOS |
| Predicate Device(s) | SIGNA™ Premier (K193282) |
| Reference Device(s) | SIGNA™ Victor (K223439) |
# Device Description
SIGNA™ Bolt is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times, and is designed for improved patient comfort and workflow. The system features a 3.0T superconducting magnet with a 70 cm bore size and can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. SIGNA™ Bolt is designed to conform to NEMA DICOM standards.
The SIGNA™ Bolt system will be offered as two commercial configurations with the following features:
| Specification | SIGNA™ Bolt – SuperXG Gradient | SIGNA™ Bolt – SuperXF Gradient |
| --- | --- | --- |
| Magnet | 3.0T superconducting magnet with a wide (70 cm) bore size and active shielding | |
| Maximum Gradient Strength | 80 mT/m | 65 mT/m |
| Maximum Slew Rate | 200 T/m/s | 200 T/m/s |
{5}
GE HealthCare
SIGNA™ Bolt
Traditional 510(k) Premarket Notification
| Specification | SIGNA™ Bolt – SuperXG Gradient | SIGNA™ Bolt – SuperXF Gradient |
| --- | --- | --- |
| RF Transmit | A liquid cooled In-Scan-Room RF transmit architecture with a peak power capability of 36 kW and 3.0T Platform Body Coil | |
| RF Receive Chain | 162 Ch available | 130 Ch available |
| Patient Table | Detachable SIGNA One Patient Table with embedded 3.0T AIR PA XL coil and up to four 32-channel high density auto-coil sensing connection ports | |
| Power Rating | 113 kVA | 90 kVA |
| Software | Software platform featuring various productivity enhancement features, designed to improve workflow and reduce scan time
- AIRx (previously cleared in K183231) – AI-based automated slice prescription tool now extended with new deep learning models for spine and prostate imaging
- SIGNA One Camera – Real-time AI-enabled image guidance that assists with automated patient positioning | |
| Gating Options | Wired, wireless, and contactless physiological gating options | |
# Comparison of Indications for Use
The indications for use have not been changed compared to the predicate, other than to reflect the SIGNA™ Bolt product name:
The SIGNA™ Bolt system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.
The images produced by the SIGNA™ Bolt system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Therefore, the intended use is the same as the predicate device in accordance with the FDA guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]", issued on July 28, 2014.
# Comparison of Technological Characteristics
Overall, SIGNA™ Bolt employs the same fundamental scientific technology and operating principles as the predicate device and reference device.
There are some differences in characteristics between the proposed device and the predicate/reference devices, as summarized below:
{6}
G E HealthCare
SIGNA™ Bolt
Traditional 510(k) Premarket Notification
| Subsystem or Component | Predicate Device SIGNA™ Premier (K193282) | Proposed Device SIGNA™ Bolt | Comments |
| --- | --- | --- | --- |
| Magnet | 3.0T superconducting magnet with active shielding. | | Identical. SIGNA™ Bolt uses the same type of magnet as used in the predicate device. |
| Gradient Coil | HRMW gradient coil with water-cooled, active-shielded design. | ZRMW gradient coil with water-cooled, active-shielded design. | Substantially Equivalent. The ZRMW gradient coil decreases heat generation and lowers the temperature of the inner bore. These differences do not impact safety or efficacy. |
| RF Transmit Subsystem | A liquid cooled RF transmit architecture with a peak power capability of 30 kW and 3.0T RF body coil. | A liquid cooled In-Scan-Room RF transmit architecture with a peak power capability of 36 kW and 3.0T Platform Body Coil. | Substantially Equivalent. The Platform Body Coil used in SIGNA™ Bolt is the same as the predicate device with design modifications to reduce heat dissipation. SIGNA™ Bolt also uses a newly designed In-Scan-Room (ISR) Transmit Driver which moves all the transmit driver components except the ISR Support Unit inside the scan room as compared to the equipment room for the predicate device. These modifications do not raise any new concern of safety or effectiveness. |
| RF Receive Subsystem | 3.0T Digitize-Per-Pin (DPP) receive chain architecture. | | Identical. SIGNA™ Bolt uses the same type of RF receive subsystem as used in the predicate device. |
| Gating Options | Respiratory peripheral and cardiac gating with wired and wireless connection. | Respiratory peripheral and cardiac gating with wired and wireless connection. Contactless peripheral cardiac and respiratory gating. | Substantially Equivalent. SIGNA™ Bolt offers the same wired and wireless gating capabilities as the predicate device. SIGNA™ Bolt introduces contactless gating, enabling peripheral cardiac and respiratory triggered MR exams without the need for physical attachment of monitoring accessories. This modification does not raise any new concern of safety or effectiveness. |
| Software Features | Comprehensive suite of software features, pulse sequences, and image processing applications to support MR imaging of all anatomies. | | Substantially Equivalent. SIGNA™ Bolt software is based off that of reference device SIGNA™ Victor, bringing additional parity with features available on the predicate device software platform along with modifications and enhancements to existing features. The modifications do not raise any new concern of safety or effectiveness. |
Page 3 of 6
{7}
GE HealthCare
SIGNA™ Bolt
Traditional 510(k) Premarket Notification
# Summary of Nonclinical Testing
The SIGNA™ Bolt system was developed and tested under the same design controls, risk management processes, and GE HealthCare quality system as the predicate device SIGNA™ Premier (K193282) to demonstrate substantial equivalence in safety and performance.
The following quality assurance measures were implemented during the development of SIGNA™ Bolt, consistent with those applied to the predicate device:
- Risk Analysis
- Requirements Reviews
- Design Reviews
- Unit-Level Testing (Verification)
- System Integration Testing (Verification)
- Performance Testing (Verification)
- Simulated Use Testing (Validation)
Testing was conducted in accordance with the following voluntary consensus standards:
- ANSI/AAMI/IEC 60601-1
- IEC 60601-1-2
- IEC 60601-2-33
- IEC TS 60601-4-2
- IEC 62304
- ISO 10993-1
- IEC 62464-1
- NEMA MS 1
- NEMA MS 3
- NEMA MS 4
- NEMA MS 6
- NEMA MS 8
- NEMA MS 9
- NEMA MS 14
- NEMA PS 3.1 - 3.20
Software verification and validation testing was conducted and documented in accordance with the FDA guidance “Content of Premarket Submissions for Device Software Functions” issued on June 14, 2023.
Cybersecurity testing was completed and documented in accordance with the FDA guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” issued on June 27, 2025.
# Contactless Gating
Verification and validation testing confirmed that the contactless gating feature meets its performance specifications by accurately detecting and displaying respiratory and peripheral cardiac waveforms without the need for physical accessories. Contactless gating also supports use of these waveforms for triggering MR acquisitions across multiple anatomical regions.
Page 4 of 6
{8}
GE HealthCare
SIGNA™ Bolt
Traditional 510(k) Premarket Notification
# SIGNA One Camera
The SIGNA One Camera is intended solely for use during the exam setup process prior to the initiation of the localizer scan and does not perform diagnostic imaging or evaluation. Accordingly, performance testing focused on assessing the deep learning model landmark inference accuracy using diverse datasets collected from 80 volunteers in the US and China. To ensure unbiased evaluation, the test dataset was entirely separate from the training and validation datasets. Accuracy was evaluated by comparing the MR system coordinates of the camera-predicted anatomical landmarks against ground truth annotations. The SIGNA One Camera met the predefined acceptance criterion of 90%, achieving successful detection in up to 99% of cases across all evaluated anatomical regions.
A time on task study was also conducted with 11 MR Scan Operators as participants to assess the efficiency and reliability of the SIGNA One Camera landmarking workflow compared to conventional laser landmarking. Results demonstrated that the camera workflow consistently enabled faster setup times for landmarking. Additionally, the camera achieved a 97% success rate in landmark acceptance with the addition of anatomical obstructions, exceeding the predefined acceptance criterion of 95%.
# AIRx
The AIRx feature is intended to automatically determine scan planes aligned to specific anatomical structures before the initiation of a scan. It was initially cleared for brain imaging under K183231 and subsequently extended to include knee imaging. In this submission, AIRx functionality has been further expanded to support imaging of the spine and prostate.
Performance testing was conducted on the AIRx Spine deep learning models using a dedicated dataset derived from 376 subjects. Data were sourced from multiple clinical sites and internal GE HealthCare sites to ensure diversity in anatomy, field strength, imaging protocols, pathology, and presence of implants. The test dataset was held separate from training and validation data. All models met their predefined acceptance criteria.
Comparative studies demonstrated that AIRx Spine reduced scan prescription times compared to manual workflows and minimized inter-operator variability, as confirmed by Structural Similarity Index (SSIM) analysis and visual comparisons. AIRx Spine successfully adapted prescriptions to patient-specific spinal curvatures, enabling accurate multi-slice, multi-angle prescriptions for cervical, thoracic, and lumbar stations. Additionally, AIRx Spine successfully automated the setup of Pars Interarticularis and Cervical Foramina scans, reducing the variability in manual prescription due to the complex angle and small size of these structures. All tests met predefined acceptance criteria with no deviations or failures, supporting the performance claims for AIRx Spine.
Performance testing was conducted on the six deep learning models that comprise the AIRx Prostate feature, evaluating automated prostate scan plane prescription using SSFSE localizer images across diverse datasets from clinical sites in the US and Europe. A total of 785 exams were used for the test dataset, which was kept separate from the training and validation data. All models met predefined acceptance criteria, confirming robustness to variations in anatomy, pathology, and presence of implants.
GE HealthCare intends to obtain clearance for AIRx Spine and AIRx Prostate on GE HealthCare’s 3.0T and 1.5T MR Systems.
Page 5 of 6
{9}
GE HealthCare
SIGNA™ Bolt
Traditional 510(k) Premarket Notification
## Summary of Clinical Testing
SIGNA™ Bolt did not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission in accordance with the FDA Guidance “Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices” issued on October 10, 2023.
A U.S. Board Certified radiologist reviewed these images and attested that images produced by SIGNA™ Bolt are of sufficient quality for diagnostic use.
## Conclusions Drawn from Performance Testing
In conclusion, GE HealthCare considers the SIGNA™ Bolt to be as safe, as effective, with performance that is substantially equivalent to the predicate device.
Page 6 of 6
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.