K253749 · Zimmer Switzerland Manufacturing GmbH · HSB · Dec 12, 2025 · Orthopedic

Device Facts

Record IDK253749
Device NameAffixus® Natural Nail® Proximal Humeral System
ApplicantZimmer Switzerland Manufacturing GmbH
Product CodeHSB · Orthopedic
Decision DateDec 12, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Affixus Natural Nail Humeral Nail System nails are intramedullary nails intended for temporary internal fixation and stabilization of humeral fractures or osteotomies. The Affixus Natural Nail Humeral Nail System is indicated for use in a variety of fractures, such as: • Proximal fractures (proximal short and long nails only) • Diaphyseal fractures (proximal long nails and antegrade/retrograde nails only) • Open and closed fractures • Comminuted fractures • Nonunions and malunions • Pathologic fractures

Device Story

Intramedullary fixation system for humeral fractures/osteotomies; consists of titanium alloy nails, bone screws, washers, and nail caps. Proximal humeral nails feature CoreLock™ locking mechanism to secure 4.0 mm blunt tip screws against migration. Device implanted by surgeons in clinical settings to restore bone shape and provide temporary stabilization. Modification replaces previous teeth-profile locking mechanism with multiple thread pitch design; maintains friction-fit principle while adding form-fit interaction for enhanced locking performance. Output is mechanical stabilization of bone segments; aids healing by maintaining alignment. Benefits include improved screw fixation and reduced risk of migration.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including mechanical integrity and engagement strength evaluations of the modified CoreLock™ interface, which met all predetermined acceptance criteria.

Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V, ISO 5832-3/ASTM F136) and C.P. Titanium (ISO 5832-2/ASTM F67). Principle: Intramedullary fixation with friction/form-fit screw locking mechanism (CoreLock™). Form factor: Variety of lengths/diameters. Sterilization: Same as predicate.

Indications for Use

Indicated for patients requiring temporary internal fixation and stabilization of humeral fractures or osteotomies, including proximal and diaphyseal fractures, open/closed fractures, comminuted fractures, nonunions, malunions, and pathologic fractures.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION December 12, 2025 Zimmer Switzerland Manufacturing GmbH Melanie Mitrov Regulatory Affairs Sr. Specialist II Sulzerallee 8 Winterthur, ZH 8404 Switzerland Re: K253749 Trade/Device Name: Affixus® Natural Nail® Proximal Humeral System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: November 25, 2025 Received: November 25, 2025 Dear Melanie Mitrov: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253749 - Melanie Mitrov Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K253749 - Melanie Mitrov Page 3 Sincerely, # Farzana Sharmin -S Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253749 | ? | | Please provide the device trade name(s). | | ? | | Affixus® Natural Nail® Proximal Humeral System | | | | Please provide your Indications for Use below. | | ? | | The Affixus Natural Nail Humeral Nail System nails are intramedullary nails intended for temporary internal fixation and stabilization of humeral fractures or osteotomies. The Affixus Natural Nail Humeral Nail System is indicated for use in a variety of fractures, such as: • Proximal fractures (proximal short and long nails only) • Diaphyseal fractures (proximal long nails and antegrade/retrograde nails only) • Open and closed fractures • Comminuted fractures • Nonunions and malunions • Pathologic fractures | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K253749 - Page 1 of 2 | 510(k) #: | 510(k) Summary | Prepared on: 2025-12-12 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | Zimmer Switzerland Manufacturing GmbH | | | Applicant Address | Sulzerallee 8, 8404 Winterthur, Switzerland Winterthur ZH 8404 Switzerland | | | Applicant Contact Telephone | +41791530853 | | | Applicant Contact | Ms. Melanie Mitrov | | | Applicant Contact Email | melanie.mitrov@zimmerbiomet.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | Affixus® Natural Nail® Proximal Humeral System | | | Common Name | Intramedullary fixation rod | | | Classification Name | Rod, Fixation, Intramedullary And Accessories | | | Regulation Number | 888.3020 | | | Product Code(s) | HSB | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K231114 | Affixus® Natural Nail® Humeral Nail System | HSB | | K200814 | Affixus® Natural Nail® System Humeral Nail | HSB | | K181827 | Affixus® Natural Nail® System Humeral Nail | HSB | | K033806 | ACE PROXIMAL HUMERAL NAIL SYSTEM | HSB | | K033878 | ACE ANTEGRADE RETROGRADE HUMERAL NAIL SYSTEM | HSB | | Device Description Summary | | 21 CFR 807.92(a)(4) | | The Affixus® Natural Nail® Humeral Nail System consists of temporary fixation intramedullary nails designed for fixation and stabilization of fractures or osteotomies of the humerus. The nails restore the shape of preinjured bone, and they are available in a variety of lengths and diameters to meet assorted patient needs. The humeral nails are used together with Bone Screws, Washers, and Nail Caps. Nail caps are available to protect the nail threads from tissue ingrowth and extend the nail length if necessary. Each of the intramedullary nails is secured by a series of screws that pass-through holes in the nail. A locking mechanism feature (CoreLock™) in the proximal segment of the Proximal Humeral Nails provides the option to lock 4.0 mm diameter Blunt Tip Screws within the nail. CoreLock™ secures the screws against medial or lateral migration. The current design locks on the outer thread diameter of the screw, relying on friction. The Proximal Humerus Nails are made of Titanium alloy [Protasul®-64WF (Ti-6Al-4V) ISO 5832-3/ASTM F136] and C.P Titanium [Protasul®-Ti ISO 5832-2/ASTM F67]. The reason for this submission is to introduce a design change of the CoreLock™ locking system in the Affixus® Natural Nail® Humeral Nails by replacing the teeth profile with a multiple thread pitch. The new design preserves the friction fit mechanism and adds a form-fit | | | {5} K253749 - Page 2 of 2 interaction for enhanced locking performance. ## Intended Use/Indications for Use 21 CFR 807.92(a)(5) The Affixus Natural Nail Humeral Nail System nails are intramedullary nails intended for temporary internal fixation and stabilization of humeral fractures or osteotomies. The Affixus Natural Nail Humeral Nail System is indicated for use in a variety of fractures, such as: - Proximal fractures (proximal short and long nails only) - Diaphyseal fractures (proximal long nails and antegrade/retrograde nails only) - Open and closed fractures - Comminuted fractures - Nonunions and malunions - Pathologic fractures ## Indications for Use Comparison 21 CFR 807.92(a)(5) The indications between the predicate and the subject device are the same. ## Technological Comparison 21 CFR 807.92(a)(6) The subject device, Affixus® Natural Nail® Humeral Nail System, maintains the same fundamental technological characteristics, intended use, and indications for use as the predicate device (Affixus® Natural Nail® Humeral Nail System, K231114). The only modification is an update to the CoreLock™ interface, changing the engagement mechanism from a teeth profile in the predicate device to a multiple thread pitch design in the subject device. This Special 510(k) is being submitted solely to address this minor design difference. The updated CoreLock™ interface preserves the same friction-fit principle as the predicate and does not introduce different risks, alter device function, or change clinical use. Verification and performance testing, including repeat biomechanical CoreLock™ Strength Evaluations conducted using well-established, standardized methods, demonstrate that the modified interface meets all performance requirements and does not raise different questions of safety or effectiveness. Overall, the data provided support that the subject device is at least as safe and effective as the legally marketed predicate device and is substantially equivalent in technological characteristics and performance. ## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non-Clinical Performance: Verification testing was performed to assess the updated CoreLock™ interface, including mechanical integrity and engagement strength. All testing met predetermined acceptance criteria, were compared to the predicate, and confirmed that the design modification does not raise different questions of safety or effectiveness. Clinical Performance: Clinical data were not required. Non-clinical testing was sufficient to support substantial equivalence. Conclusion: The subject device has the same intended use and same indications for use as the predicate device. The subject device uses the same operating principle, incorporate the same basic design and labeling and are manufactured and sterilized using the same materials and processes as the predicate device. Except for the modifications to the CoreLock™ interface described in this submission the subject device is identical to the predicate device, and the performance data and analyses demonstrate that: - any differences do not raise different questions of safety and effectiveness as established with performance testing; and - the subject devices are at least as safe and effective as the legally marketed predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...