The AFFIXUS Natural Nail System Humeral nails are temporary fixation intramedullary nails designed for fixation and stabilization of fractures or osteotomies of the humerus. They restore the shape of preinjured bone and are available in a variety of lengths and diameters to meet assorted anatomical needs. Nail caps are available to protect the nail threads from tissue ingrowth and extend the nail length if necessary. Each of the intramedullary nails is secured by a series of screws that pass through holes in the nail.
Device Story
Affixus Natural Nail System Humeral Nail is an intramedullary fixation device for humeral fractures/osteotomies. Device consists of titanium alloy (Tivanium) nails, screws, and nail caps. Surgeon inserts nail into humeral canal; secures with locking screws through pre-drilled holes. Long variants include axial compression mechanism. Device restores bone shape; provides temporary stabilization. Used in clinical/OR settings by orthopedic surgeons. Output is mechanical fixation of bone fragments; aids healing by maintaining alignment and stability.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing including bioburden, biocompatibility, torsional stiffness, static four-point bend, fatigue strength (nails and screws), self-tapping performance, driving torque, axial pull-out strength, and targeting guide functionality. Testing performed per ISO 11137-2, ISO 10993-1, ASTM F136-13, ASTM F543-17, ASTM F1264-16, and ISO 6475.
Technological Characteristics
Materials: Tivanium (Ti-6Al-4V alloy). Components: Intramedullary nails, screws, washers, nail caps. Features: Color-coded components, axial compression mechanism (long variants). Standards: ISO 11137-2, ISO 10993-1, ASTM F136-13, ASTM F543-17, ASTM F1264-16, ISO 6475. Mechanical fixation via intramedullary rod and locking screws.
Indications for Use
Indicated for patients with humeral fractures or osteotomies, including proximal fractures (proximal short/long nails), diaphyseal fractures (proximal long, antegrade/retrograde nails), open and closed fractures, comminuted fractures, nonunions, malunions, and pathologic fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 14, 2018
Zimmer GmbH Danielle Madureira Senior Specialist, Regulatory Affairs Sulzerallee 8 Winterthur, 8404 Switzerland
Re: K181827
Trade/Device Name: Affixus Natural Nail System Humeral Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: November 8, 2018 Received: November 13, 2018
Dear Danielle Madureira:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vesa
Vuniqi -S
Digitally signed
by Vesa Vuniqi -S
Date: 2018.12.14
16:07:07 -05'00'
For:
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K181827
### Device Name
AFFIXUS® NATURAL NAIL® SYSTEM HUMERAL NAIL
Indications for Use (Describe)
The AFFIXUS Natural Nail System Humeral nails are temporary fixation intramedullary nails designed for fixation and stabilization of fractures or osteotomies of the humerus. They restore the shape of preinjured bone and are available in a variety of lengths and diameters to meet assorted anatomical needs. Nail caps are available to protect the nail threads from tissue ingrowth and extend the nail length if necessary. Each of the intramedullary nails is secured by a series of screws that pass through holes in the nail.
The AFFIXUS Humeral Nails are indicated for use in a variety of fractures, such as:
- · Proximal fractures (proximal short and long nails only)
- · Diaphyseal fractures (proximal long, antegrade/retrograde nails)
- Open and closed fractures
- · Comminuted fractures
- Nonunions and malunions
- · Pathologic fractures .
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020 See PRA Statement below.
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# mise." Your prog
510(K) SUMMARY
SAP Title: 510(k) Summary
SAP Document: RA2-380737-E00-03
WT-FO 386809 Revision 01 Page 1 of 3
# 510(k) SUMMARY
| Sponsor: | Zimmer GmbH<br>Sulzerallee 8, P.O. Box<br>8404 Winterthur, Switzerland |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Danielle Jannuzzi Madureira<br>Senior Specialist, Regulatory Affairs<br>Telephone: +41 58 854 82 60<br>Fax: + 41 52 244 86 58 |
| Date: | June 29th, 2018 |
| Trade Name: | Affixus® Natural Nail® System Humeral Nail |
| Classification Product Code : | HSB - Intramedullary Fixation Rod |
| Device Classification Name: | Rod, fixation, intramedullary and accessories |
| Regulation Number /<br>Description: | 21 CFR § 888.3020 number - Intramedullary fixation rod |
| Predicate Device: | Versanail® Nailing System, manufactured by Biomet Trauma Inc, K033878<br>(Ace® Antegrade Retrograde Humeral Nail System, primary predicate) and<br>K033806 (Ace® Proximal Humeral Nail System) both cleared in 20 February<br>2004 and M/DN® Intramedullary Fixation System, manufactured by Zimmer<br>Inc, K142281 cleared 22 October 2014, K965098 cleared in 28 February 1997. |
| Device Description: | The Affixus Natural Nail System Humeral nails are temporary fixation<br>intramedullary nails designed for fixation and stabilization of fractures or<br>osteotomies of the humerus. They restore the shape of preinjured bone and<br>are available in a variety of lengths and diameters to meet assorted anatomical<br>needs.<br>Nail caps are available to protect the nail threads from tissue ingrowth and<br>extend the nail length if necessary. Each of the intramedullary nails is secured<br>by a series of screws that pass through holes in the nail. Nails and nail caps<br>are made from Tivanium® Ti-6Al-4V alloy. Screws are also made from<br>Tivanium alloy. Package labels indicate the material of each component.<br>Selected components of the Affixus Natural Nail System are color coded to aid<br>in identifying which components should be used together. Refer to the color<br>coding chart and/or surgical technique for more detailed instructions on the use<br>of Affixus Natural Nail System components. |
| Intended Use and<br>Indications for Use: | The Affixus® Natural Nail® System Humeral nails are temporary fixation<br>intramedullary nails designed for fixation and stabilization of fractures or<br>osteotomies of the humerus. They restore the shape of preinjured bone and<br>are available in a variety of lengths and diameters to meet assorted anatomical<br>needs. Nail caps are available to protect the nail threads from tissue ingrowth |
{4}------------------------------------------------
WT-FO 386809
| | SAP Title: 510(k) Summary | WT-FO 386809 |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| | SAP Document: RA2-380737-E00-03 | Revision 01 |
| | | Page 2 of 3 |
| | and extend the nail length if necessary. Each of the intramedullary nails is<br>secured by a series of screws that pass through holes in the nail. | |
| | The Affixus® Humeral Nails are indicated for use in a variety of fractures, such<br>as, proximal fractures (proximal short and long nails only), diaphyseal fractures<br>(proximal long, antegrade/retrograde nails), open and closed fractures,<br>comminuted fractures, nonunions and malunions and pathologic fractures.. | |
| Comparison to Predicate<br>Device: | The Affixus® Natural Nail® System Humeral nails, screws, washers, nail caps<br>covered by this submission are similar in intended use, design, materials, and<br>performance characteristics to the predicate devices. The Differences in the<br>technological characteristics between the primary predicate VersaNail and the<br>subject device Affixus® Natural Nail (ANN) Humeral Nail are as follows: | |
| | • Proximal and distal locking screws diameter: subject device screws have<br>smaller diameters than the predicate device | |
| | • Nail-screw configuration, including screw orientation: subject device has a<br>different in range of the screw orientation (screw hole pattern) to the<br>predicate device. Adequate primary stability within all 3 planes can be<br>achieved with both respective nails based on the screw insertion options<br>provided. | |
| | • Axial compression: subject device provides an axial compression<br>mechanism in the long device variants (Proximal Long and AR), allowing a<br>controlled axial compression. The AR variant of the predicate device does<br>not include such an option, and axial compression must be performed free-<br>hand. However, the free-hand option can also be applied with the subject<br>device. | |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions: | |
| | Results of non-clinical performance testing and analyses demonstrate that the<br>Affixus Humeral Nails is as safe, as effective, and performs as well as or better<br>than the legally market devices predicate. Performance analyses included: | |
| | 1. Bioburden study and technical review of Affixus Humeral Nails New<br>Product Introduction (NPI) project | |
| | 2. Biocompatibility of Affixus® Natural Nail® (ANN) Humeral Nail | |
| | 3. Measurement of Torsional Stiffness of Affixus Humeral Nails | |
| | 4. Static four point bend test of Affixus Humeral Nails | |
| | 5. Fatigue strength evaluation of the Affixus blunt tip screw and cortical<br>screws using a three-point bending test setup | |
| | 6. Self-tapping performance of Affixus proximal blunt tip screw 4 mm and<br>cortical bone screw 4 mm | |
| | 7. Four point bending fatigue properties of Affixus Humeral Nails | |
| | 8. Determining the driving torque of Affixus Proximal blunt tip screw 4 mm<br>and cortical bone screw 4 mm | |
| | | |
510(K) SUMMARY
SAP Title: 510(k) Summary
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Zimmer Biomet. The logo consists of a blue circle with a white "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. Below the company name is the tagline "Your progress. Our promise." in a smaller font size.
510(K) SUMMARY
SAP Title: 510(k) Summary
WT-FO 386809 Revision 01 Page 3 of 3
SAP Document: RA2-380737-E00-03
9. Determining the axial pull-out strength of Affixus Proximal blunt tip screw 4 mm and cortical bone screw 4 mm
10. Torsional properties of the Affixus Proximal blunt tip screw 4 mm and cortical bone screw 4 mm
11. Bending fatigue strength and CoreLock Evaluation of the proximal segment (including the mid-shaft segment) of the Affixus Humeral Nail system
12. Axial push-out strength evaluation of the Affixus Humeral Nail locking screw fixed by the CoreLock mechanis
13. Cleaning, autoclave, targeting functionality and impact testing of the Affixus Humerus targeting guides
The tests were performed according to the applicable Standards: ISO 11137-2, ISO 10993-1, ASTM 136-13, ASTM F543-17, ASTM F1264-16, and ISO 6475.
Clinical Performance and Conclusions:
The results of non-clinical data demonstrate that the subject device is substantially equivalent with the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.