The Taurus Clip is indicated for placement within the Gastrointestinal (GI) tract of adult patients for the purpose of: 1. Endoscopic marking 2. Hemostasis for: - Mucosal/sub-mucosal defects < 3 cm - Bleeding ulcers - Arteries < 2 mm - Polyps < 1.5 cm in diameter - Diverticula in the colon - Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel and anchoring to affix fully covered esophageal self-expanding metal stents to the wall of the esophagus 4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively.
Device Story
Sterile, single-use endoscopic clip; consists of titanium clip implant and flexible delivery system (applier). Used in GI tract; compatible with forward-viewing endoscopes (≥2.8mm working channel). Physician operates device via endoscope to manipulate mucosal tissue; clip is deformed and released to provide mechanical closure, hemostasis, marking, or anchoring. Output is physical tissue approximation/fixation. Benefits include reduced bleeding risk, secure tube/stent placement, and perforation management. Device is MR Conditional.
Clinical Evidence
Bench testing only. Performance evaluated via visual assessment, dimensional measurements, compression force, retention, retroflexion angles, rotation, simulated use, back bending, buckling, firing force, opening force, and design validation. All tests met predetermined acceptance criteria. MR conditionality confirmed via testing.
Technological Characteristics
Titanium clip implant; flexible delivery system. Single-use. Sterilized via Ethylene Oxide (ISO 11135-1). Biocompatible per ISO 10993-1/7. MR Conditional. Compatible with ≥2.8mm endoscope working channels.
Indications for Use
Indicated for adult patients requiring endoscopic marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula, post-resection prophylaxis), anchoring of jejunal feeding tubes or esophageal stents, and supplemental closure of luminal perforations <20mm.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 12, 2026
Taurus Endoscopy
Mariel Bolhouse
Co-founder and CEO
1 Place de l'Hôpital, Bat. IHU
67000 Strasbourg France
Re: K253734
Trade/Device Name: Taurus Clip
Regulation Number: 21 CFR 876.4400
Regulation Name: Hemorrhoidal Ligator
Regulatory Class: Class II
Product Code: PKL
Dated: February 9, 2026
Received: February 9, 2026
Dear Mariel Bolhouse:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253734 - Mariel Bolhouse
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K253734 - Mariel Bolhouse
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
SIVAKAMI VENKATAChALAM -S
for
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253734 | ? |
| Please provide the device trade name(s). | | ? |
| Taurus Clip | | |
| Please provide your Indications for Use below. | | ? |
| The Taurus Clip is indicated for placement within the Gastrointestinal (GI) tract of adult patients for the
purpose of:
1. Endoscopic marking
2. Hemostasis for:
- Mucosal/sub-mucosal defects < 3 cm
- Bleeding ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the colon
- Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection
3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel and anchoring to affix fully covered
esophageal self-expanding metal stents to the wall of the esophagus
4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
TAURUS endoscopy
K253734
Page 1 of 3
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR §807.92.
I. SUBMITTER
Taurus Endoscopy
1 Place de l'Hôpital, Bat. IHU
67000 Strasbourg, France
Contact Person:
Mariel Bolhouse
Co-founder and CEO
Taurus Endoscopy
Mariel@taurusendo.com
Date Summary Prepared: November 24, 2025
II. DEVICE
Trade Name: Taurus® Clip
Classification Name: Hemorrhoidal Ligator
Device Name: Hemostatic Metal Clip for the GI Tract
Classification Number: 21 CFR §876.4400
Product Code: PKL
Device Class: Class II
III. LEGALLY MARKETED PREDICATE DEVICE
Predicate Device:
Trade Name: Resolution Clip Platform
Manufacturer: Boston Scientific
510(k) Number: K222503
Classification Number: 21 CFR §876.4400
Product Code: PKL
Device Class: Class II
Reference Device:
Trade Name: Padlock Clip
Manufacturer: Steris
510(k) Number: K192722
Classification Number: 21 CFR §876.4400
Product Code: PKL
Device Class: Class II
{6}
TAURUS endoscopy
K253734
Page 2 of 3
2
# IV. DEVICE DESCRIPTION
The Taurus Clip is a sterile, single-use endoscopic clip designed for closure, hemostasis, endoscopic marking, and anchoring. The Taurus Clip is designed to be compatible with forward-viewing endoscopes with working channels equal to or greater than 2.8mm.
The Taurus Clip Device consists of two main components: one preloaded titanium Clip Implant and a flexible delivery system, referred to as the "Applier".
The Clip Implant can be used to manipulate the mucosal tissue of the GI tract as desired and then applied (deformed and released) with the handle once in place. The Applier is removed after clip deployment.
# V. INTENDED USE/INDICATIONS FOR USE
The Taurus Clip is indicated for placement within the Gastrointestinal (GI) tract of adult patients for the purpose of:
1. Endoscopic marking
2. Hemostasis for:
- Mucosal/sub-mucosal defects < 3 cm
- Bleeding ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the colon
- Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection
3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel; and
Anchoring to affix fully covered esophageal self-expanding metal stents to the wall of the esophagus
4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively
# VI. TECHNICAL CHARACTERISTICS
The proposed Taurus Clip has identical Intended Use, Indications for Use, working length, principle of operation, patient population, system configuration, and utilizes the same endoscopic working channel as the predicate Resolution Clip Platform cleared under K222503. The proposed Taurus Clip has a jaw opening width that falls within the range of the predicate device, and both the subject and predicate devices are single-use and sterilized with Ethylene Oxide. The proposed Taurus Clip has a different clip geometry and longer teeth as compared to the predicate device; however, the length of the teeth of the subject device is shorter than the reference device, Padlock Clip Defect Closure System (K192722).
{7}
TAURUS endoscopy
K253734
Page 3 of 3
3
# VII. SUBSTANTIAL EQUIVALENCE
The proposed Taurus Clip is substantially equivalent to the currently cleared predicate Resolution Clip Platform (K222503). The technological differences in clip geometry do not raise new or different questions of safety and effectiveness.
# VIII. PERFORMANCE DATA
The proposed Taurus Clip meets the requirements of ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing, ISO 11135-1 Sterilization of Health Care products – Ethylene Oxide – Part 1: Requirements for Development, Validation, and Routing Control of Sterilization Process for Medical Devices, and ISO 10993-7 Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals.
The following testing was conducted on the proposed Taurus Clip device:
| Test | Results |
| --- | --- |
| Visual Assessment (device) | Meets Predetermined acceptance criteria |
| Dimensional Measurements | Meets Predetermined acceptance criteria |
| Clip Compression Force | Meets Predetermined acceptance criteria |
| Clip Retention | Meets Predetermined acceptance criteria |
| Measurement of Retroflexion Angles | Meets Predetermined acceptance criteria |
| Clip Rotation | Meets Predetermined acceptance criteria |
| Simulated Use Assessment | Meets Predetermined acceptance criteria |
| Clip Back Bending | Meets Predetermined acceptance criteria |
| Applier Buckling Test | Meets Predetermined acceptance criteria |
| Clip Approach Vertical Oblique | Meets Predetermined acceptance criteria |
| Full Length Firing Force | Meets Predetermined acceptance criteria |
| Clip Opening Force | Meets Predetermined acceptance criteria |
| Design Validation Assessment | Meets Predetermined acceptance criteria |
In addition, the proposed Taurus Clip was evaluated for Magnetic Resonance (MR) to support MR Conditionality. The results from the test data demonstrate that the proposed Taurus Clip is an MR Conditional device.
# IX. CONCLUSION
Based on the comparison of intended use, indications for use, technological characteristics, non-clinical performance data, and principle of operation, the subject device is as safe, as effective, and performs at least as well as the legally marketed predicate device K222503, Class II (21 CFR 876.4400), product code PKL.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.