Unscented Menstrual Long Applicator Tampon; Unscented Menstrual Cardboard Applicator Tampon; Unscented Menstrual Digital Tampons
K253719 · Unibeauty (Hubei) Technology Co., Ltd. · HEB · Dec 17, 2025 · Obstetrics/Gynecology
Device Facts
Record ID
K253719
Device Name
Unscented Menstrual Long Applicator Tampon; Unscented Menstrual Cardboard Applicator Tampon; Unscented Menstrual Digital Tampons
Applicant
Unibeauty (Hubei) Technology Co., Ltd.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Dec 17, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge.
Device Story
Cylindrical menstrual tampons designed for vaginal insertion to absorb menstrual discharge. Consist of an absorbent pledget (cotton, organic cotton, or viscose), overwrap (cotton, organic cotton, or nonwoven fabric), and removal string. Available in digital style (manual insertion) or with cardboard/plastic applicators (for insertion assistance). Provided non-sterile for single-use. Used by patients in a home setting. Absorbs menstrual fluid; aids in hygiene during menstruation.
Clinical Evidence
Bench testing only. No clinical data. Performance validated through non-clinical testing of all product variants, including absorbency testing, biocompatibility, microbiology (Staphylococcus aureus growth/TSST-1 production), and chemical residue assessment, leveraging data from predicate and referenced devices.
Technological Characteristics
Materials: Cotton, organic cotton, viscose (pledget); cotton, organic cotton, spunbond/thermal-bonded nonwoven (overwrap); cardboard, PE (applicator). Dimensions: Pledget length 35-50mm, diameter 11.5-15.2mm. Absorbencies: Light (≤6g), Regular (6-9g), Super (9-12g), Super Plus (12-15g). Non-sterile, single-use, manual or applicator-assisted insertion.
Indications for Use
Indicated for use by women to absorb menstrual discharge during their period.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
Unscented Menstrual Long Applicator Tampon (K251033)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 17, 2025
Unibeauty (Hubei) Technology Co., Ltd.
Sandra Jiang
RA Manager
Bldg. A8, Nonwoven Small And Medium-Sized Enterprise
Industrial Park, Pengchang Town
Xiantao, Hubei
CHINA
Re: K253719
Trade/Device Name: Unscented Menstrual Long Applicator Tampon; Unscented Menstrual Cardboard Applicator Tampon; Unscented Menstrual Digital Tampons
Regulation Number: 21 CFR 884.5470
Regulation Name: Unscented Menstrual Tampon
Regulatory Class: II
Product Code: HEB
Dated: November 24, 2025
Received: November 24, 2025
Dear Sandra Jiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253719 - Sandra Jiang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K253719 - Sandra Jiang
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason Roberts -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253719 | |
| Device Name 1. Unsected menstrual long applicator tampon 2. Unsected menstrual Cardboard Applicator Tampon 3. Unsected menstrual Digital Tampons | |
| Indications for Use (Describe) The Unsected Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Sponsor: Unibeauty (Hubei) Technology Co.,Ltd.
Subject Device: Unscented Menstrual Tampon
K253719
Page 1 of 9
# 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
Summary Prepared Date: October 28, 2025
# I. Submitter's Identification
Submitter Name: Unibeauty (Hubei) Technology Co.,Ltd.
Address: Building A8, nonwoven small and medium-sized enterprise Industrial Park,
PENGCHANG Town, Xiantao City, Hubei Province, China
Contacted Person: Guan Zhou
Telephone Number: +86 13560898147
E-mail: guanzhou@uni-beauty.com
Date of Preparation: October 7, 2025
# II. Identification of the device
## Subject Device
Common Name: Unscented Menstrual Tampon
Trade Name:
- Unscented Menstrual Long Applicator Tampon
- Unscented Menstrual Cardboard Applicator Tampon
- Unscented Menstrual Digital Tampons
Classification Name: Tampon, Menstrual, Unscented
Product Code: HEB
Regulation: 21 CFR 884.5470
Device Class: II
## Predicate Device
Sponsor: Unibeauty (Hubei) Technology Co.,Ltd.
Common Name: Unscented Menstrual Tampon
Trade Name:
- Unscented Menstrual Long Applicator Tampon
- Unscented Menstrual Cardboard Applicator Tampon
- Unscented Menstrual Digital Tampons
Classification Name: Tampon, Menstrual, Unscented
Product Code: HEB
Regulation: 21 CFR 884.5470
Device Class: II
510k Number: K251033
# III. Reason of submission
1) Device size change;
2) Device design change
1 / 9
{5}
K253719
Page 2 of 9
Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
Subject Device: Unscented Menstrual Tampon
## IV. Device Description
The proposed unscented menstrual tampons are cylindrical devices designed to absorb menstrual blood during a women's period. They consist of a tampon, including an absorbent pledget ("absorbent core") completely surrounded by an overwrap ("security veil") and a removal string ("withdrawal cord"), and an applicator (except for the digital style). The tampon design is cylindrical, bullet-like shape. The applicator has a smooth, rounded tip for ease of insertion. According to different composition materials and design styles, the unscented menstrual tampons of this submission are divided into 3 categories (Unscented menstrual long applicator tampon, Unscented menstrual Cardboard Applicator Tampon, Unscented menstrual Digital Tampons) and 7 sub-categories (Cotton + cardboard tube, Organic cotton + cardboard tube, Cotton (digital style), Organic cotton(digital style), viscose (digital style), Cotton + plastic tube and Viscose + cardboard tube). For applicator-style tampons, they feature a built-in applicator made of plastic or cardboard that helps the users insert the tampons into the vagina by pushing the tampon out of the applicator. The digital style tampons consist of a plain tampon designed to be inserted manually using the fingers.
The tampons are provided in 4 absorbencies: Light(L) (≤6g), Regular(R) (6-9g), Super (S)(9-12g) and Super Plus (SP)(12-15g). For Cotton + cardboard tube and Organic cotton + cardboard tube, L and SP are applicable for this application. For Cotton (digital style), Organic cotton (digital style) and Viscose (digital style), only L is applicable. For Cotton + plastic tube and Viscose + cardboard tube, all those four absorbencies are applicable for this application.
Each device is individually wrapped and then packaged in sealed multi-unit containers for retail sale. All tampons and applicators are provided non-sterile and for single use only.
| Category | Subcategory | Size | | | |
| --- | --- | --- | --- | --- | --- |
| | | Light | Regular | Super | Super Plus |
| Unscented Menstrual Cardboard Applicator Tampon | Cotton + cardboard tube | ✓ | | | ✓ |
| | Organic cotton + cardboard tube | ✓ | | | ✓ |
| | Viscose + cardboard tube | ✓ | ✓ | ✓ | ✓ |
| Unscented Menstrual Digital Tampons | Cotton (digital style) | ✓ | | | |
| | Organic cotton (digital style) | ✓ | | | |
| | Viscose (digital style) | ✓ | | | |
| Unscented Menstrual Long Applicator Tampon | Cotton + plastic tube | ✓ | ✓ | ✓ | ✓ |
## V. Indication for use
The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge.
## VI. Technological Characteristics Comparison
{6}
Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
Subject Device: Unscented Menstrual Tampon
K253719
Page 3 of 9
# Technological Characteristics Comparison Table
| Elements of Comparison | Subject Device | | | Predicate Device | | | Verdict |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Device Trade Name | Unscented Menstrual Long Applicator Tampon | Unscented Menstrual Cardboard Applicator Tampon | Unscented Menstrual Digital Tampons | Unscented Menstrual Long Applicator Tampon | Unscented Menstrual Cardboard Applicator Tampon | Unscented Menstrual Tampons | Same |
| Manufacturer | Unibeauty (Hubei) Technology Co.,Ltd. | | | Unibeauty (Hubei) Technology Co.,Ltd. | | | Same |
| K Number | K253719 | | | K251033 | | | -- |
| Product Code | HEB | | | HEB | | | Same |
| Regulation Number | 21 CFR 884.5470 | | | 21 CFR 884.5470 | | | Same |
| Classification | Class II | | | Class II | | | Same |
| Indications for Use | The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | | | The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | | | Same |
| Sterile | No | | | No | | | Same |
| Single-Use | Yes | | | Yes | | | Same |
| Design | Tampon with cylindrical shape and bullet-like tip. Applicator with smooth, rounded tip. | | Tampon with cylindrical shape and bullet-like tip. Digital tampon without applicator. | Tampon with cylindrical shape and bullet-like tip. Applicator with smooth, rounded tip. | Tampon with cylindrical shape and bullet-like tip. Digital tampon without applicator. | | Same |
| Unscented/ Scented | Unscented | | | Unscented | | | Same |
3 / 9
{7}
K253719
Page 4 of 9
Sponsor: Unibeauty (Hubei) Technology Co.,Ltd.
Subject Device: Unscented Menstrual Tampon
| Dimensions
Specifications
(mm) | Pledget length (Dry):
35±5.0--50±5.0mm
Pledget diameter(Dry):
11.5±2.0--15.2±2.0mm
Total length of the product:
120.0±4.0--122.2±4.0mm
Length of the push rod on product:
70±2.0mm
Outside diameter (outer tube):
14.5±1.0--16.2±1.0mm | Pledget length (Dry):
35±5.0--50±5.0mm
Pledget diameter(Dry):
11.5±2.0--15.2±2.0mm
Total length of the product:
122.5±4.0mm
Length of the push rod on product:
67±2.0mm
Outside diameter (outer tube):
14.5±1.0--16.1±1.0mm | Pledget length (Dry):
35±5.0mm
Pledget diameter(Dry):
11.5±2.0mm | Pledget length (Dry):
37±3.0--50±3.0
Pledget diameter(Dry):
11.5±1.0--15.2±1.0
Total length of the product:
118.0±2.0
Length of the push rod on product:
70±1
Outside diameter (outer tube):
14.5±0.2--16.2±0.2 | Pledget length (Dry):
37±3.0--50±3.0
Pledget diameter(Dry):
11.5±1.0--15.2±1.0
Total length of the product:
122.5±2.0
Length of the push rod on product:
67±1
Outside diameter (outer tube):
14.1±0.2--16.1±0.2 | Pledget length (Dry):
Viscose core:
43±5.0mm- 50±5.0
organic cotton core
and cotton core:
38.5±5.0mm-50±5.0
Pledget diameter(Dry):
12.5±0.2--15.2±0.2 | Different |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Weight
Specifications(g)
(without single wrapper) | L: 4.00-4.60g
R: 4.60-5.40g
S: 5.40-6.70g
SP: 6.60-7.60g | L: 3.00-3.90g
R: 3.90-4.60g
S: 4.60-5.70g
SP: 5.60-6.70g | Viscose core, organic cotton core and cotton core:
L: 1.00-1.70g | L: 3.90-4.40g
R: 4.45-5.15g
S: 5.65-6.45g
SP: 6.45-7.35g | R: 3.90-4.60g
S: 4.90-5.70g | viscose core:
R: 1.70-2.30g
S: 2.40-3.10g
SP: 3.00-3.60g
organic cotton core & cotton core:
R: 1.95-2.55g
S: 2.50-3.30g
SP: 3.30-4.10g | Different |
4 / 9
{8}
Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
Subject Device: Unscented Menstrual Tampon
K253719
Page 5 of 9
| Absorbent Pledget | Cotton | Organic cotton, Cotton and Viscose | Organic cotton, Cotton and Viscose | Organic cotton | Organic cotton and Cotton | Organic cotton, Cotton, and Viscose | Similar |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Overwrap | Cotton | Viscose: Spunbond/thermal-bonded nonwoven fabric/cotton
Organic cotton type: Organic cotton
Cotton type: Cotton | Viscose: Spunbond/thermal-bonded nonwoven fabric/cotton
Organic cotton type: Organic cotton
Cotton type: Cotton | Organic cotton | Organic cotton and Cotton | Organic cotton, Cotton, and Viscose | Different |
| Removal String | Cotton | Organic cotton and Cotton | Organic cotton and Cotton | Organic cotton | Organic cotton and Cotton | Organic cotton and Cotton | Similar |
| Applicator | PE | Cardboard | None | LDPE | Cardboard | None | Similar |
| Absorbencies | L:≤6g
R: 6-9g
S: 9-12g
SP: 12-15g | L:≤6g
R: 6-9g
S: 9-12g
SP: 12-15g | L:≤6g | L:≤6g
R: 6-9g
S: 9-12g
SP: 12-15g | R: 6-9g
S: 9-12g | R: 6-9g
S: 9-12g
SP: 12-15g | Same |
| Primary Packaging | Paper/PE/PP | Paper/PE/PP | PET+CPP/paper | Paper/PE | Paper/PE | PET+CPP | Similar |
5 / 9
{9}
K253719
Page 6 of 9
Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
Subject Device: Unscented Menstrual Tampon
The subject device has the similar indication for use as the predicate device as well as comparable technological characteristics. The differences include the material composition and dimensions; however, the differences do not raise different questions of safety and effectiveness.
6 / 9
{10}
K253719
Page 7 of 9
Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
Subject Device: Unscented Menstrual Tampon
The subject device and predicate device are similar in the pledget composition and components (tampon and applicator). The subject and predicate device differ in the combination method of pledge and applicator, and available absorbencies.
For Unscented Menstrual Long Applicator Tampon: pledget material was changed from organic cotton to cotton; All the other materials remain the same.
For Unscented Menstrual Cardboard Applicator Tampon: Absorbency options L (Light) and SP (Super Plus) have been added to the Cotton + cardboard tube and Organic cotton + cardboard tube subcategories. Additionally, viscose has been added as a pledget material with all absorbencies (L, R, S, and SP) offered.
For Unscented Menstrual Digital Tampons: Absorbency option L (Light) has been added to the Cotton (digital style), Organic cotton (digital style), and Viscose (digital style) subcategories.
The unscented menstrual tampon (including versions with long applicators, cardboard applicators, and digital tampons) has been updated with new sizes and pledget materials, resulting in slight changes to dimensional and weight specifications compared to the predicate device.
While the method of combining the pledget and applicator differs slightly from that of the predicate device, the materials used remain consistent. The absorbent pledget is made of cotton, organic cotton, or viscose; the overwrap consists of cotton (for cotton type) or organic cotton (for organic cotton type), or Spunbond/thermal-bonded nonwoven fabric/cotton (for Viscose type); the removal string is made of cotton or organic cotton; and the applicator is composed of cardboard or PE.
In summary, the device is substantially equivalent to the predicate devices in terms of features, design, specifications, mode of operation, and intended use. The noted differences do not raise any new safety or effectiveness concerns.
## VII. Substantial Equivalence Discussion:
## Technical Equivalence
Technical equivalence is determined by means of the condition for use, specification, design, deployment method and principle of operation.
Subject devices differ slightly from predicate devices in technical characteristics due to changes in the combinations of pledget and applicators, and the addition of new absorbencies that impact the weight and dimensional specifications. However, these differences do not raise different questions of safety and effectiveness.
## Biocompatibility Assessment
{11}
K253719
Page 8 of 9
Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
Subject Device: Unscented Menstrual Tampon
Biocompatibility studies were performed in accordance with the FDA guidance document "Guidance for industry and Food and Drug Administration Staff - Use of International Standard 1S0 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process" issued on September 8, 2023.
The material of the proposed device is identical to the predicate device (K251033), and K232598 and K252613 (i.e. the absorbent pledget is made of organic cotton, cotton, and viscose, the applicator is made of cardboard and PE, overwrap and removal string are made of cotton, organic cotton and Spunbond/thermal-bonded nonwoven fabric/cotton). Therefore, biocompatibility test data from the predicate device, K232598 and K252613 were leveraged to demonstrate the safety of the material of the proposed device.
## Microbiology Assessment
Per the FDA guidance “Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s) - Guidance for Industry and FDA Staff”, menstrual tampons should demonstrate that the tampon, in its final manufactured form, does not:
- Enhance the growth of Staphylococcus aureus;
- Increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1);
- Alter the growth of normal vaginal microflora.
The proposed Unscented Menstrual Tampon (Unscented Menstrual Long Applicator Tampon, Unscented Menstrual Cardboard Applicator Tampon, Unscented Menstrual Digital Tampons) are composed of the same raw material as devices cleared in K232598, K252613 and the predicate (K251033). The microbiology test data from the tampons cleared in the predicate device and K232598 were leveraged to demonstrate the microbiological safety of the tampons used with the applicator introduced in the current submission
## Chemical Residue Assessment
Chemical residue assessment of tampons is conducted in accordance with the recommendations outlined in the guidance document “Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s) - Guidance for Industry and FDA Staff” dated July 2005. The tampons listed as compatible with the subject device applicator are identical to the predicate device (K251033), and K232598. Chemical residue testing of the predicate device and K232598 were leveraged to demonstrate the absence of harmful chemicals in the cotton and viscose versions of this device.
## VIII. Summary of Verification and Validation:
In light of the new absorbency levels introduced, non-clinical performance testing was conducted across all product variants: Unscented Menstrual Long Applicator Tampon, Unscented Menstrual Cardboard Applicator Tampon, and Unscented Menstrual Digital Tampons. Test results confirmed that all devices meet the required performance standards. In accordance with Quality System procedures, all necessary testing was performed to validate
{12}
K253719
Page 9 of 9
Sponsor: Unibeauty (Hubei) Technology Co., Ltd.
Subject Device: Unscented Menstrual Tampon
the cumulative modifications to the subject devices. A risk analysis was carried out to identify the verification and validation activities required under 21 CFR 820.30.
## IX. Conclusion
The nonclinical tests demonstrate that the subject device, Unscented Menstrual Tampon, is as safe and effective as the legally marketed predicate device (K251033). Therefore, the subject device is substantially equivalent to the predicate device.
9 / 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.