Unscented Menstrual Long Applicator Tampon; Unscented Menstrual Cardboard Applicator Tampon; Unscented Menstrual Digital Tampons

K253719 · Unibeauty (Hubei) Technology Co., Ltd. · HEB · Dec 17, 2025 · Obstetrics/Gynecology

Device Facts

Record IDK253719
Device NameUnscented Menstrual Long Applicator Tampon; Unscented Menstrual Cardboard Applicator Tampon; Unscented Menstrual Digital Tampons
ApplicantUnibeauty (Hubei) Technology Co., Ltd.
Product CodeHEB · Obstetrics/Gynecology
Decision DateDec 17, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 884.5470
Device ClassClass 2

Indications for Use

The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge.

Device Story

Cylindrical menstrual tampons designed for vaginal insertion to absorb menstrual discharge. Consist of an absorbent pledget (cotton, organic cotton, or viscose), overwrap (cotton, organic cotton, or nonwoven fabric), and removal string. Available in digital style (manual insertion) or with cardboard/plastic applicators (for insertion assistance). Provided non-sterile for single-use. Used by patients in a home setting. Absorbs menstrual fluid; aids in hygiene during menstruation.

Clinical Evidence

Bench testing only. No clinical data. Performance validated through non-clinical testing of all product variants, including absorbency testing, biocompatibility, microbiology (Staphylococcus aureus growth/TSST-1 production), and chemical residue assessment, leveraging data from predicate and referenced devices.

Technological Characteristics

Materials: Cotton, organic cotton, viscose (pledget); cotton, organic cotton, spunbond/thermal-bonded nonwoven (overwrap); cardboard, PE (applicator). Dimensions: Pledget length 35-50mm, diameter 11.5-15.2mm. Absorbencies: Light (≤6g), Regular (6-9g), Super (9-12g), Super Plus (12-15g). Non-sterile, single-use, manual or applicator-assisted insertion.

Indications for Use

Indicated for use by women to absorb menstrual discharge during their period.

Regulatory Classification

Identification

An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION December 17, 2025 Unibeauty (Hubei) Technology Co., Ltd. Sandra Jiang RA Manager Bldg. A8, Nonwoven Small And Medium-Sized Enterprise Industrial Park, Pengchang Town Xiantao, Hubei CHINA Re: K253719 Trade/Device Name: Unscented Menstrual Long Applicator Tampon; Unscented Menstrual Cardboard Applicator Tampon; Unscented Menstrual Digital Tampons Regulation Number: 21 CFR 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: II Product Code: HEB Dated: November 24, 2025 Received: November 24, 2025 Dear Sandra Jiang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253719 - Sandra Jiang Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K253719 - Sandra Jiang Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jason Roberts -S Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K253719 | | | Device Name 1. Unsected menstrual long applicator tampon 2. Unsected menstrual Cardboard Applicator Tampon 3. Unsected menstrual Digital Tampons | | | Indications for Use (Describe) The Unsected Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Sponsor: Unibeauty (Hubei) Technology Co.,Ltd. Subject Device: Unscented Menstrual Tampon K253719 Page 1 of 9 # 510(K) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. Summary Prepared Date: October 28, 2025 # I. Submitter's Identification Submitter Name: Unibeauty (Hubei) Technology Co.,Ltd. Address: Building A8, nonwoven small and medium-sized enterprise Industrial Park, PENGCHANG Town, Xiantao City, Hubei Province, China Contacted Person: Guan Zhou Telephone Number: +86 13560898147 E-mail: guanzhou@uni-beauty.com Date of Preparation: October 7, 2025 # II. Identification of the device ## Subject Device Common Name: Unscented Menstrual Tampon Trade Name: - Unscented Menstrual Long Applicator Tampon - Unscented Menstrual Cardboard Applicator Tampon - Unscented Menstrual Digital Tampons Classification Name: Tampon, Menstrual, Unscented Product Code: HEB Regulation: 21 CFR 884.5470 Device Class: II ## Predicate Device Sponsor: Unibeauty (Hubei) Technology Co.,Ltd. Common Name: Unscented Menstrual Tampon Trade Name: - Unscented Menstrual Long Applicator Tampon - Unscented Menstrual Cardboard Applicator Tampon - Unscented Menstrual Digital Tampons Classification Name: Tampon, Menstrual, Unscented Product Code: HEB Regulation: 21 CFR 884.5470 Device Class: II 510k Number: K251033 # III. Reason of submission 1) Device size change; 2) Device design change 1 / 9 {5} K253719 Page 2 of 9 Sponsor: Unibeauty (Hubei) Technology Co., Ltd. Subject Device: Unscented Menstrual Tampon ## IV. Device Description The proposed unscented menstrual tampons are cylindrical devices designed to absorb menstrual blood during a women's period. They consist of a tampon, including an absorbent pledget ("absorbent core") completely surrounded by an overwrap ("security veil") and a removal string ("withdrawal cord"), and an applicator (except for the digital style). The tampon design is cylindrical, bullet-like shape. The applicator has a smooth, rounded tip for ease of insertion. According to different composition materials and design styles, the unscented menstrual tampons of this submission are divided into 3 categories (Unscented menstrual long applicator tampon, Unscented menstrual Cardboard Applicator Tampon, Unscented menstrual Digital Tampons) and 7 sub-categories (Cotton + cardboard tube, Organic cotton + cardboard tube, Cotton (digital style), Organic cotton(digital style), viscose (digital style), Cotton + plastic tube and Viscose + cardboard tube). For applicator-style tampons, they feature a built-in applicator made of plastic or cardboard that helps the users insert the tampons into the vagina by pushing the tampon out of the applicator. The digital style tampons consist of a plain tampon designed to be inserted manually using the fingers. The tampons are provided in 4 absorbencies: Light(L) (≤6g), Regular(R) (6-9g), Super (S)(9-12g) and Super Plus (SP)(12-15g). For Cotton + cardboard tube and Organic cotton + cardboard tube, L and SP are applicable for this application. For Cotton (digital style), Organic cotton (digital style) and Viscose (digital style), only L is applicable. For Cotton + plastic tube and Viscose + cardboard tube, all those four absorbencies are applicable for this application. Each device is individually wrapped and then packaged in sealed multi-unit containers for retail sale. All tampons and applicators are provided non-sterile and for single use only. | Category | Subcategory | Size | | | | | --- | --- | --- | --- | --- | --- | | | | Light | Regular | Super | Super Plus | | Unscented Menstrual Cardboard Applicator Tampon | Cotton + cardboard tube | ✓ | | | ✓ | | | Organic cotton + cardboard tube | ✓ | | | ✓ | | | Viscose + cardboard tube | ✓ | ✓ | ✓ | ✓ | | Unscented Menstrual Digital Tampons | Cotton (digital style) | ✓ | | | | | | Organic cotton (digital style) | ✓ | | | | | | Viscose (digital style) | ✓ | | | | | Unscented Menstrual Long Applicator Tampon | Cotton + plastic tube | ✓ | ✓ | ✓ | ✓ | ## V. Indication for use The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. ## VI. Technological Characteristics Comparison {6} Sponsor: Unibeauty (Hubei) Technology Co., Ltd. Subject Device: Unscented Menstrual Tampon K253719 Page 3 of 9 # Technological Characteristics Comparison Table | Elements of Comparison | Subject Device | | | Predicate Device | | | Verdict | | --- | --- | --- | --- | --- | --- | --- | --- | | Device Trade Name | Unscented Menstrual Long Applicator Tampon | Unscented Menstrual Cardboard Applicator Tampon | Unscented Menstrual Digital Tampons | Unscented Menstrual Long Applicator Tampon | Unscented Menstrual Cardboard Applicator Tampon | Unscented Menstrual Tampons | Same | | Manufacturer | Unibeauty (Hubei) Technology Co.,Ltd. | | | Unibeauty (Hubei) Technology Co.,Ltd. | | | Same | | K Number | K253719 | | | K251033 | | | -- | | Product Code | HEB | | | HEB | | | Same | | Regulation Number | 21 CFR 884.5470 | | | 21 CFR 884.5470 | | | Same | | Classification | Class II | | | Class II | | | Same | | Indications for Use | The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | | | The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge. | | | Same | | Sterile | No | | | No | | | Same | | Single-Use | Yes | | | Yes | | | Same | | Design | Tampon with cylindrical shape and bullet-like tip. Applicator with smooth, rounded tip. | | Tampon with cylindrical shape and bullet-like tip. Digital tampon without applicator. | Tampon with cylindrical shape and bullet-like tip. Applicator with smooth, rounded tip. | Tampon with cylindrical shape and bullet-like tip. Digital tampon without applicator. | | Same | | Unscented/ Scented | Unscented | | | Unscented | | | Same | 3 / 9 {7} K253719 Page 4 of 9 Sponsor: Unibeauty (Hubei) Technology Co.,Ltd. Subject Device: Unscented Menstrual Tampon | Dimensions Specifications (mm) | Pledget length (Dry): 35±5.0--50±5.0mm Pledget diameter(Dry): 11.5±2.0--15.2±2.0mm Total length of the product: 120.0±4.0--122.2±4.0mm Length of the push rod on product: 70±2.0mm Outside diameter (outer tube): 14.5±1.0--16.2±1.0mm | Pledget length (Dry): 35±5.0--50±5.0mm Pledget diameter(Dry): 11.5±2.0--15.2±2.0mm Total length of the product: 122.5±4.0mm Length of the push rod on product: 67±2.0mm Outside diameter (outer tube): 14.5±1.0--16.1±1.0mm | Pledget length (Dry): 35±5.0mm Pledget diameter(Dry): 11.5±2.0mm | Pledget length (Dry): 37±3.0--50±3.0 Pledget diameter(Dry): 11.5±1.0--15.2±1.0 Total length of the product: 118.0±2.0 Length of the push rod on product: 70±1 Outside diameter (outer tube): 14.5±0.2--16.2±0.2 | Pledget length (Dry): 37±3.0--50±3.0 Pledget diameter(Dry): 11.5±1.0--15.2±1.0 Total length of the product: 122.5±2.0 Length of the push rod on product: 67±1 Outside diameter (outer tube): 14.1±0.2--16.1±0.2 | Pledget length (Dry): Viscose core: 43±5.0mm- 50±5.0 organic cotton core and cotton core: 38.5±5.0mm-50±5.0 Pledget diameter(Dry): 12.5±0.2--15.2±0.2 | Different | | --- | --- | --- | --- | --- | --- | --- | --- | | Weight Specifications(g) (without single wrapper) | L: 4.00-4.60g R: 4.60-5.40g S: 5.40-6.70g SP: 6.60-7.60g | L: 3.00-3.90g R: 3.90-4.60g S: 4.60-5.70g SP: 5.60-6.70g | Viscose core, organic cotton core and cotton core: L: 1.00-1.70g | L: 3.90-4.40g R: 4.45-5.15g S: 5.65-6.45g SP: 6.45-7.35g | R: 3.90-4.60g S: 4.90-5.70g | viscose core: R: 1.70-2.30g S: 2.40-3.10g SP: 3.00-3.60g organic cotton core & cotton core: R: 1.95-2.55g S: 2.50-3.30g SP: 3.30-4.10g | Different | 4 / 9 {8} Sponsor: Unibeauty (Hubei) Technology Co., Ltd. Subject Device: Unscented Menstrual Tampon K253719 Page 5 of 9 | Absorbent Pledget | Cotton | Organic cotton, Cotton and Viscose | Organic cotton, Cotton and Viscose | Organic cotton | Organic cotton and Cotton | Organic cotton, Cotton, and Viscose | Similar | | --- | --- | --- | --- | --- | --- | --- | --- | | Overwrap | Cotton | Viscose: Spunbond/thermal-bonded nonwoven fabric/cotton Organic cotton type: Organic cotton Cotton type: Cotton | Viscose: Spunbond/thermal-bonded nonwoven fabric/cotton Organic cotton type: Organic cotton Cotton type: Cotton | Organic cotton | Organic cotton and Cotton | Organic cotton, Cotton, and Viscose | Different | | Removal String | Cotton | Organic cotton and Cotton | Organic cotton and Cotton | Organic cotton | Organic cotton and Cotton | Organic cotton and Cotton | Similar | | Applicator | PE | Cardboard | None | LDPE | Cardboard | None | Similar | | Absorbencies | L:≤6g R: 6-9g S: 9-12g SP: 12-15g | L:≤6g R: 6-9g S: 9-12g SP: 12-15g | L:≤6g | L:≤6g R: 6-9g S: 9-12g SP: 12-15g | R: 6-9g S: 9-12g | R: 6-9g S: 9-12g SP: 12-15g | Same | | Primary Packaging | Paper/PE/PP | Paper/PE/PP | PET+CPP/paper | Paper/PE | Paper/PE | PET+CPP | Similar | 5 / 9 {9} K253719 Page 6 of 9 Sponsor: Unibeauty (Hubei) Technology Co., Ltd. Subject Device: Unscented Menstrual Tampon The subject device has the similar indication for use as the predicate device as well as comparable technological characteristics. The differences include the material composition and dimensions; however, the differences do not raise different questions of safety and effectiveness. 6 / 9 {10} K253719 Page 7 of 9 Sponsor: Unibeauty (Hubei) Technology Co., Ltd. Subject Device: Unscented Menstrual Tampon The subject device and predicate device are similar in the pledget composition and components (tampon and applicator). The subject and predicate device differ in the combination method of pledge and applicator, and available absorbencies. For Unscented Menstrual Long Applicator Tampon: pledget material was changed from organic cotton to cotton; All the other materials remain the same. For Unscented Menstrual Cardboard Applicator Tampon: Absorbency options L (Light) and SP (Super Plus) have been added to the Cotton + cardboard tube and Organic cotton + cardboard tube subcategories. Additionally, viscose has been added as a pledget material with all absorbencies (L, R, S, and SP) offered. For Unscented Menstrual Digital Tampons: Absorbency option L (Light) has been added to the Cotton (digital style), Organic cotton (digital style), and Viscose (digital style) subcategories. The unscented menstrual tampon (including versions with long applicators, cardboard applicators, and digital tampons) has been updated with new sizes and pledget materials, resulting in slight changes to dimensional and weight specifications compared to the predicate device. While the method of combining the pledget and applicator differs slightly from that of the predicate device, the materials used remain consistent. The absorbent pledget is made of cotton, organic cotton, or viscose; the overwrap consists of cotton (for cotton type) or organic cotton (for organic cotton type), or Spunbond/thermal-bonded nonwoven fabric/cotton (for Viscose type); the removal string is made of cotton or organic cotton; and the applicator is composed of cardboard or PE. In summary, the device is substantially equivalent to the predicate devices in terms of features, design, specifications, mode of operation, and intended use. The noted differences do not raise any new safety or effectiveness concerns. ## VII. Substantial Equivalence Discussion: ## Technical Equivalence Technical equivalence is determined by means of the condition for use, specification, design, deployment method and principle of operation. Subject devices differ slightly from predicate devices in technical characteristics due to changes in the combinations of pledget and applicators, and the addition of new absorbencies that impact the weight and dimensional specifications. However, these differences do not raise different questions of safety and effectiveness. ## Biocompatibility Assessment {11} K253719 Page 8 of 9 Sponsor: Unibeauty (Hubei) Technology Co., Ltd. Subject Device: Unscented Menstrual Tampon Biocompatibility studies were performed in accordance with the FDA guidance document "Guidance for industry and Food and Drug Administration Staff - Use of International Standard 1S0 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process" issued on September 8, 2023. The material of the proposed device is identical to the predicate device (K251033), and K232598 and K252613 (i.e. the absorbent pledget is made of organic cotton, cotton, and viscose, the applicator is made of cardboard and PE, overwrap and removal string are made of cotton, organic cotton and Spunbond/thermal-bonded nonwoven fabric/cotton). Therefore, biocompatibility test data from the predicate device, K232598 and K252613 were leveraged to demonstrate the safety of the material of the proposed device. ## Microbiology Assessment Per the FDA guidance “Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s) - Guidance for Industry and FDA Staff”, menstrual tampons should demonstrate that the tampon, in its final manufactured form, does not: - Enhance the growth of Staphylococcus aureus; - Increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1); - Alter the growth of normal vaginal microflora. The proposed Unscented Menstrual Tampon (Unscented Menstrual Long Applicator Tampon, Unscented Menstrual Cardboard Applicator Tampon, Unscented Menstrual Digital Tampons) are composed of the same raw material as devices cleared in K232598, K252613 and the predicate (K251033). The microbiology test data from the tampons cleared in the predicate device and K232598 were leveraged to demonstrate the microbiological safety of the tampons used with the applicator introduced in the current submission ## Chemical Residue Assessment Chemical residue assessment of tampons is conducted in accordance with the recommendations outlined in the guidance document “Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s) - Guidance for Industry and FDA Staff” dated July 2005. The tampons listed as compatible with the subject device applicator are identical to the predicate device (K251033), and K232598. Chemical residue testing of the predicate device and K232598 were leveraged to demonstrate the absence of harmful chemicals in the cotton and viscose versions of this device. ## VIII. Summary of Verification and Validation: In light of the new absorbency levels introduced, non-clinical performance testing was conducted across all product variants: Unscented Menstrual Long Applicator Tampon, Unscented Menstrual Cardboard Applicator Tampon, and Unscented Menstrual Digital Tampons. Test results confirmed that all devices meet the required performance standards. In accordance with Quality System procedures, all necessary testing was performed to validate {12} K253719 Page 9 of 9 Sponsor: Unibeauty (Hubei) Technology Co., Ltd. Subject Device: Unscented Menstrual Tampon the cumulative modifications to the subject devices. A risk analysis was carried out to identify the verification and validation activities required under 21 CFR 820.30. ## IX. Conclusion The nonclinical tests demonstrate that the subject device, Unscented Menstrual Tampon, is as safe and effective as the legally marketed predicate device (K251033). Therefore, the subject device is substantially equivalent to the predicate device. 9 / 9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...