Milex™ Incontinence Dish Pessaries; Milex™ Incontinence Dish Pessaries with Support; Milex™ Cube Pessaries; Milex™ Cube Pessaries with Drainage Holes
K253698 · Cooper Surgical · HHW · Jan 20, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K253698
Device Name
Milex™ Incontinence Dish Pessaries; Milex™ Incontinence Dish Pessaries with Support; Milex™ Cube Pessaries; Milex™ Cube Pessaries with Drainage Holes
Applicant
Cooper Surgical
Product Code
HHW · Obstetrics/Gynecology
Decision Date
Jan 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.3575
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CooperSurgical Milex™ pessaries are intended to provide support to pelvic organs when inserted in the vagina. The following Indications for Use are associated with each of the following pessary styles: Milex™ Incontinence Dish Pessaries and Incontinence Dish Pessaries with Support: Incontinence dish pessary is indicated for use as removable structures placed in the vagina to treat uterine prolapse, including cystocele and rectocele, as well as stress urinary incontinence in women. Milex™ Cube Pessaries and Cube Pessaries with Drainage Holes: Cube pessary is indicated for nonsurgical management of pelvic organ prolapse in Stage III prolapse including rectocele and/or cystocele.
Device Story
Milex™ Pessaries are removable, silicone-based vaginal inserts designed to provide mechanical support to pelvic organs. Available in multiple styles (Incontinence Dish, Incontinence Dish with Support, Cube, Cube with Drainage Holes) and sizes, they are intended for nonsurgical management of uterine prolapse, cystocele, rectocele, and stress urinary incontinence. The devices are inserted into the vagina by a clinician or patient to provide structural support, alleviating symptoms of prolapse. They are distributed non-sterile for single-patient, multiple-use, requiring cleaning with mild soap and water. By physically supporting the pelvic floor and organs, the pessaries aim to improve patient quality of life and manage prolapse symptoms without surgical intervention.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation, and chronic toxicity) and shelf-life/use-life testing per ASTM F1980 supporting a 5-year life.
Technological Characteristics
Materials: Liquid silicone rubber and colorant. Form factor: Multiple shapes (Dish, Cube) and sizes. Principle: Mechanical support of pelvic organs. Connectivity: None. Sterilization: Non-sterile. Standards: ASTM F1980 (aging), ISO 10993 series (biocompatibility).
Indications for Use
Indicated for women with uterine prolapse, cystocele, rectocele, or stress urinary incontinence (Incontinence Dish styles) or Stage III pelvic organ prolapse including rectocele/cystocele (Cube styles).
Regulatory Classification
Identification
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse (falling down of uterus), uterine retroposition (backward displacement), or gynecologic hernia.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 20, 2026
Cooper Surgical
Timothy Good
Regulatory Affairs Supervisor
95 Corporate Dr.
Trumbull, Connecticut 06611
Re: K253698
Trade/Device Name: Milex™ Incontinence Dish Pessaries; Milex™ Incontinence Dish Pessaries with Support; Milex™ Cube Pessaries; Milex™ Cube Pessaries with Drainage Holes
Regulation Number: 21 CFR 884.3575
Regulation Name: Vaginal Pessary
Regulatory Class: II
Product Code: HHW
Dated: November 20, 2025
Received: November 24, 2025
Dear Timothy Good:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253698 - Timothy Good
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K253698 - Timothy Good
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason Roberts -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253698 | |
| Device Name CooperSurgical Milex™ Pessaries | |
| Indications for Use (Describe) The CooperSurgical Milex™ pessaries are intended to provide support to pelvic organs when inserted in the vagina. The following Indications for Use are associated with each of the following pessary styles: Milex™ Incontinence Dish Pessaries and Incontinence Dish Pessaries with Support: Incontinence dish pessary is indicated for use as removable structures placed in the vagina to treat uterine prolapse, including cystocele and rectocele, as well as stress urinary incontinence in women. Milex™ Cube Pessaries and Cube Pessaries with Drainage Holes: Cube pessary is indicated for nonsurgical management of pelvic organ prolapse in Stage III prolapse including rectocele and/or cystocele. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
CooperSurgical
95 Corporate Drive
Trumbull, CT 06611
T 203 601 5200
www.coopersurgical.com
# 510(k) Summary
# CooperSurgical Milex™ Pessaries
## I. Submitter Information
Company Name: CooperSurgical Inc.
Company Address: 95 Corporate Drive
Trumbull, CT 06611
Telephone: 203-601-5200
Contact Person: Timothy Good
Regulatory Affairs Supervisor
Date Prepared: January 16, 2026
## II. Device Information:
Trade Name: CooperSurgical Milex™ Pessaries
Common Name: Vaginal pessary
Classification Name: Vaginal pessary
Regulatory Class: Class II
Regulation Medical Specialty: Obstetrics/Gynecology
Regulation Number: 884.3575
Product Code: HHW (pessary, vaginal)
## III. Predicate Devices:
| Name | Manufacturer | 510(k) |
| --- | --- | --- |
| EIS Vaginal Pessaries | EIS Corporation | K132313 |
| Mentor Evacare Vaginal Pessaries | Mentor Corporation | K993308 |
## Reference Device:
| Name | Manufacturer | 510(k) |
| --- | --- | --- |
| CooperSurgical Milex™ Pessaries | CooperSurgical Inc. | K250438 |
The predicate devices have not been subject to a design related recall.
510(k) Summary
Page 1 of 6
{5}
510(k) Summary
Page 2 of 6
# IV. Device Description:
Milex™ Incontinence Dish Pessaries, Incontinence Dish Pessaries with Support, Cube Pessaries, and Cube Pessaries with Drainage Holes are intended for support to pelvic organs when inserted into the vagina. Milex™ Pessaries are made of silicone and distributed in a non-sterile condition. The Milex™ Pessaries are available in a variety of styles, each having a range of sizes. All pessaries are produced in a like fashion, utilizing injection molding of liquid silicone rubber. The Milex™ Pessaries are manufactured in pink for single patient use.
This submission includes the following designs of Milex™ Pessaries: Incontinence Dish Pessary (Dish without Support, Dish with Support) and Milex™ Cube Pessary (Cube without Drainage Holes, Cube with Drainage Holes).
# V. Indications for Use:
The CooperSurgical Milex™ Pessaries are intended to provide support to pelvic organs when inserted in the vagina.
The following Indications for Use are associated with each of the following pessary styles:
## Milex™ Incontinence Dish Pessaries and Incontinence Dish Pessaries with Support:
Incontinence dish pessary is indicated for use as removable structures placed in the vagina to treat uterine prolapse, including cystocele and rectocele, as well as stress urinary incontinence in women.
## Milex™ Cube Pessaries and Cube Pessaries with Drainage Holes:
Cube pessary is indicated for nonsurgical management of pelvic organ prolapse in Stage III prolapse including rectocele and/or cystocele.
{6}
VI. Comparison of Technological Characteristics with the Predicate Device
The subject Milex™ Pessary Devices have identical intended use, principles of operation, similar fundamental scientific technology and equivalent or identical Indications for Use with the legally marketed predicate devices.
CooperSurgical Milex™ Pessaries are vaginal pessaries used to function as a supportive structure of the uterus, bladder and/or rectum. Similar to the identified predicate devices, CooperSurgical Milex™ Pessaries are available in multiple shapes and sizes and are manufactured from liquid silicone.
Table 1. Indications for Use Comparison
| Device | Indications for Use |
| --- | --- |
| Milex™ Pessaries | The CooperSurgical Milex™ pessaries are intended to provide support to pelvic organs when inserted in the vagina.
The following Indications for Use are associated with each of the following pessary styles:
• Milex™ Incontinence Dish Pessaries:
Incontinence dish pessary is indicated for use as removable structures placed in the vagina to treat uterine prolapse, including cystocele and rectocele, as well as stress urinary incontinence in women. |
| Predicate Device (K132313)
EIS
Vaginal Pessaries | EIS Vaginal Pessaries are indicated for the use as removable structures placed in the vagina to treat uterine prolapse, including cystocele and rectocele, as well as stress urinary incontinence in women. |
| Milex™ Pessaries | The CooperSurgical Milex™ pessaries are intended to provide support to pelvic organs when inserted in the vagina.
The following Indications for Use are associated with each of the following pessary styles:
• Milex™ Cube Pessaries:
Cube pessary is indicated for nonsurgical management of pelvic organ prolapse in Stage III prolapse including rectocele and/or cystocele. |
| Predicate Device (K993308)
Mentor EvaCare™ Vaginal Pessaries | A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse.
The specific indications for use by pessary style is as follows:
• Cube Pessary: Support of third degree prolapse, procidentia, cystocele, and rectocele. |
510(k) Summary
Page 3 of 6
{7}
510(k) Summary
Page 4 of 6
| Reference Device (K250438) | The CooperSurgical Milex pessaries are intended to provide support to pelvic organs when inserted in the vagina.
The following Indications for Use are associated with each of the following pessary styles:
**Milex® Shaatz Folding Pessaries**
o Shaatz pessary is indicated for temporary, nonsurgical management of pelvic organ prolapse in Stage I and Stage II prolapse, complicated by a mild cystocele.
**Milex® Gellhorn Pessaries**
o Gellhorn pessary is indicated for temporary, nonsurgical management of pelvic organ prolapse in Stage III prolapse or procidentia.
**Milex® Ring Folding Pessaries**
o Milex Ring Pessary is indicated for use as removable structures placed in the vagina to treat uterine prolapse, including cystocele and rectocele, as well as stress urinary incontinence in women. |
| --- | --- |
As shown in the table above, the subject Milex™ pessaries and the predicate devices, EIS Vaginal Pessaries and Mentor EvaCare™ Vaginal Pessaries, cleared under K132313 and K993308 share the same intended use: both are removable structures placed in the vagina to provide nonsurgical support to pelvic organs. This intended use is consistent across all pessary styles included in the subject device and aligns with the general use of vaginal pessaries included in the subject device, to manage conditions such as uterine prolapse, cystocele, rectocele, and stress urinary incontinence.
The reference device submission (K250438) describes pessary materials identical to those in this submission. These reference devices support that the inclusion of other shapes and the broader range of indications included in this CooperSurgical Milex™ Pessaries submission do not represent a new intended use and that the bundled shapes and indications included in this submission are consistent with 510k cleared pessaries of the same type.
Therefore, there are no intended use concerns.
Table 2: Technological Differences
| Technological characteristic | Differences Compared to Predicates |
| --- | --- |
| Dimensions and sizing | |
| Milex™ Dish Pessaries | Milex™ - 2.125 in. (55mm) – 3.375 in. (85mm) |
{8}
510(k) Summary
Page 5 of 6
The size range for the Milex™ Dish Pessaries is within the dimensional envelope of the identified predicate device. The CooperSurgical sizes do not raise different questions of safety or effectiveness.
Milex™ Cube Pessaries
Milex™ - 1.00 in. (25mm) – 2.25 in. (57mm)
The size range for the Milex™ Cube Pessaries is within the dimensional envelope of the identified predicate device. The CooperSurgical sizes do not raise different questions of safety or effectiveness.
Materials
The subject device is made of liquid silicone and colorant. The predicate devices are also made of silicone and colorant, but silicone and additive formulations may differ. The differences do not raise different questions of safety and effectiveness.
VII. Performance Data
Evidence was submitted to support the following verification activities for the Milex™ Pessary Devices:
- Shelf Life/Use Life of Device
- Accelerated aging testing was performed to support a 5-year use life for the subject device per ASTM F1980.
- Cleaning
- The subject device is provided non-sterile and is intended as a single patient multiple use device. The patient labeling includes instructions for cleaning, which specify that the pessary should be washed using mild soap and warm water. These cleaning instructions are consistent with those cleared for the identified predicate devices and are identical to the instructions previously reviewed and cleared in K250438 for other Milex pessary devices.
- Biocompatibility
- Biocompatibility for the Milex™ Pessary Devices was performed in accordance with 2023 FDA biocompatibility guidance Use of International Standard ISO 10993-1, “Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process” and the testing performed by CooperSurgical is provided in Table 3.
Table 3: Summary of biological testing
| Test Description | Results | ISO Standard |
| --- | --- | --- |
| Cytotoxicity | Non-cytotoxic | 10993-5:2009 |
| Sensitization | Non-sensitizer | 10993-10:2021 |
{9}
| Test Description | Results | ISO Standard |
| --- | --- | --- |
| Irritation or Intracutaneous Reactivity | Non-Irritant | 10993-23:2023 |
| Subacute/Subchronic Toxicity | Non-Subacute/Subchronic Toxic | 10993-11:2017, and 10993-6:2016 |
| Material Mediated Pyrogenicity | Non-Material Mediated Pyrogenic | 10993-11:2017. |
| Acute System Toxicity | Non-Acute Systemic Toxic | 10993-11:2017 |
| Genotoxicity | Non-Mutagenic/Non-Genotoxic | 10993-3:2014, 10993-33:2015, 10993-11:2017 |
| Implantation | Device induced no local response | 10993-6:2016 |
| Chronic Toxicity | Device induced no systemic toxicity | 10993-6:2016, and 10993-11:2017 |
VIII. Conclusion
The subject and predicate devices share the same intended use, fundamental scientific technology, principles of operation, and identical or similar Indications for Use. Differences in design sizes or materials between the subject and predicate devices do not raise any new questions of safety and effectiveness. Based on the verification evidence activities provided in this pre-market notification application, the subject Milex™ Pessaries are substantially equivalent to the legally marketed predicate devices.
510(k) Summary
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.