A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse.
Device Story
Mentor EvaCare™ Vaginal Pessaries are removable medical-grade silicone elastomer devices designed to support pelvic organs. Available in multiple styles (e.g., Ring, Cube, Gellhorn, Hodge, Dish, Donut, Oval, Shaatz, Mar-Land, Gehrung), they are inserted into the vagina by a clinician to treat conditions including uterine prolapse, cystocele, rectocele, and stress urinary incontinence. The device acts as a mechanical support structure. Clinicians use a fitting set to determine the appropriate size for the patient. By providing structural support to the pelvic floor, the pessary helps alleviate symptoms associated with pelvic organ prolapse and incontinence, potentially improving patient quality of life and avoiding or delaying surgical intervention.
Clinical Evidence
Bench testing only. Biocompatibility testing was performed on the silicone elastomer materials.
Technological Characteristics
Medical-grade silicone elastomers. Multiple form factors (Ring, Cube, Gellhorn, Hodge, Dish, Donut, Oval, Shaatz, Mar-Land, Gehrung). Mechanical support device. Non-powered. No software or electronics.
Indications for Use
Indicated for patients with uterine prolapse, cystocele, rectocele, stress urinary incontinence, uterine retroversion, or incompetent cervix. Specific indications vary by style: Donut (3rd degree prolapse, cystocele, rectocele); Ring (1st/2nd degree prolapse, cystocele); Dish (stress incontinence, minor prolapse); Oval (1st/2nd degree prolapse, cystocele); Shaatz (1st/2nd degree prolapse, cystocele); Mar-Land (stress incontinence, minor prolapse); Hodge (1st/2nd degree prolapse, cystocele, stress incontinence, incompetent cervix, retroversion); Gehrung (cystocele, rectocele, 2nd/3rd degree prolapse); Gellhorn (2nd/3rd degree prolapse, procidentia); Cube (3rd degree prolapse, procidentia, cystocele, rectocele). Fitting set used for sizing.
Regulatory Classification
Identification
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse (falling down of uterus), uterine retroposition (backward displacement), or gynecologic hernia.
DesChutes Medical Products, Inc. Pelvx Donut (K974117)
DesChutes Medical Products, Inc. Pelvx Cube (K974117)
DesChutes Medical Products, Inc. Pelvx Incontinence Ring (K974116)
DesChutes Medical Products, Inc. Pelvx Ring (K974115)
DesChutes Medical Products, Inc. Pelvx Ring with Support (K974115)
Bioteque America, Inc. Pessary Flexible Silicone Donut Ring (K920747)
Bioteque America, Inc. Flexible Silicone Gellhorn (K920187)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 1 3 1999
# Mentor EvaCare™ Pessary 510(k) Notification
# 510(k) SUMMARY Mentor EvaCare™ Vaginal Pessaries
# This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92
The assigned 510(k) number is: K993308
Donna A. Crawford Contact Person: Manager, Corporate Regulatory Affairs Mentor Corporation 201 Mentor Drive Santa Barbara, CA 93111
- Telephone: (805) 879-6304 (805) 681-6004 FAX:
Date Prepared: September 30, 1999
## Device Name and Classification:
| Proprietary Name: | Mentor EvaCare™ Vaginal Pessaries |
|----------------------|-----------------------------------|
| Common Name: | Vaginal Pessary |
| Classification Name: | Vaginal Pessary |
| Classification: | Class II (21 CFR § 884.3575) |
| Product Code: | 85 HHW |
#### Manufacturer:
Mentor Urology 1615 West River Road North Minneapolis, MN 55411
{1}------------------------------------------------
#### 510(k) SUMMARY Mentor EvaCare™ Vaginal Pessaries
#### Substantial Equivalence Claim:
Mentor EvaCare™ Vaginal Pessaries are substantially equivalent to preamendment devices and pessaries manufactured by:
- Milex Products, Inc. ● Inflat-o-Ball K904026
- . DesChutes Medical Products, Inc.
- Pelvx Donut, Pelvx Cube K974117 Pelvx Incontinence Ring K974116
- Pelvx Ring, Pelvx Ring with Support K974115
- Bioteque America, Inc. . Pessary Flexible Silicone Donut Ring K920747 Flexible Silicone Gellhorn K920187
#### Device Description:
Mentor EvaCare™ Vaginal Pessaries are manufactured from medical grade silicone elastomers and are available in the following styles:
- Cube Pessarv ●
- Dish Pessary ●
- Gellhorn Pessary ●
- Ring Pessary . ●
- Mar-Land Pessary .
- . Hodge Pessary
- Oval Pessary .
- Gehrung Pessary ●
- Fitting Set .
- . Cube Pessary with Drainage
- Dish Pessary with Support ●
- Gellhorn Pessary with Drainage ●
- Ring Pessary with Support ●
- Mar-Land Pessary with Support t
- . Hodge Pessary with Support
- Shaatz Pessary -.
- Donut Pessary ، .
#### Indications For Use:
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse.
The specific indications for use by pessary style are as follows:
Donut Pessary: Support of third degree prolapse, cystocele and rectocele.
{2}------------------------------------------------
- Support of first or mild degree prolapse. Ring Pessary with Ring Pessary: support can also be used on an accompanying cystocele.
- Dish Pessary: Control of stress urinary incontinence and minor degrees of prolapse.
- Oval Pessary: Support of first or second degree prolapse and cystocyle.
- Shaatz Pessary: Support of first or mild second degree prolapse and cystocele.
- Mar-Land Pessary: Control of stress urinary incontinence and minor degrees of prolapse.
- Hodge Pessary: Support of first to second degree prolapse, uterine retroversion or incompetent cervix, stress urinary incontinence.
- Gehrung Pessary: Support of cystocele and rectocele, support of second to third degree prolapse.
- Gellhorn Pessary: Support of second to third degree prolapse or procidentia.
- Cube Pessary: Support of third degree prolapse, procidentia, cystocele, and rectocele.
- Fitting Set: Used to determine the proper size of pessary for each patient
## Summary of Testing:
Mentor has previously performed biocompatibility testing on the pessary component materials.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 3 1999
Ms. Donna A. Crawford Manager, Corporate Requlatory Affairs MENTOR Corporation 201 Mentor Drive Santa Barbara, CA 93111
Re: K993308 Mentor EvaCare™ Vaginal Pessaries Dated: October 25, 1999 Received: November 23, 1999 Regulatory Class: II 21 CFR 884.3575/Procode: 85 HHW
Dear Ms. Crawford:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls of the Act include requirements for annual registration, ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requiation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K 9g 3308
Mentor EvaCare™ Vaginal Pessary (various styles) Device Name:
Indications For Use:
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse.
Please refer to the attached page for a listing of indications for use for each pessary style.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-------------------------------------------|--------------|
| Division of Reproductive, Abdominal, ENT, | |
| and Radiological Devices | |
| 510(k) Number | K993308/5001 |
Prescription Use (Per 21 CFR 801.109)
OR Over the Counter Use
(Optional Format 1-2-96)
{5}------------------------------------------------
# INDICATIONS FOR USE
- Donut Pessary: Support of third degree prolapse, cystocele and rectocele.
- Ring Pessary: Support of first or second degree prolapse. Ring with support can also be used on an accompanying cystocele.
- Dish Pessary: Control of stress urinary incontinence and support of minor degrees of prolapse.
- Oval Pessary: Support of first or second degree prolapse and accompanying cystocyle.
- Shaatz Pessary: Support of first or second degree prolapse and accompanying cystocyle.
- Mar-Land Pessary: Control of stress urinary incontinence and support of minor degrees of prolapse.
- Hodge Pessary: Support of first to second degree prolapse, a cystocele, stress urinary incontinence and an incompetent cervix or uterine retroversion..
- Gehrung Pessary: Support of cystocele and rectocele, as well as support of second to third degree prolapse.
- Gellhorn Pessary: Support of second to third degree prolapse or procidentia.
- Cube Pessary: Support of third degree prolapse, procidentia, cystocele, and rectocele.
- Fitting Set: Used to determine the proper size of pessary for each patient.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.