K253629 · Shenzhen Root Innovation Technology Co., Ltd. · BTA · May 26, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253629
Device Name
Momcozy Nasal Aspirator (BN007)
Applicant
Shenzhen Root Innovation Technology Co., Ltd.
Product Code
BTA · General, Plastic Surgery
Decision Date
May 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The device is designed for using suction to remove nasal secretion and mucus in children (age 2~12 years old) at home environment. The additional spray function helps moisturize and soften nasal secretions in children (over 3 years old), making them easier to aspirate.
Device Story
Momcozy Nasal Aspirator (BN007) is a handheld, battery-operated device for home use by parents/caregivers to clear nasal mucus in children. Device features a pump generating negative pressure (54-66 kPa) for suction and a sprayer module for moisturizing nasal passages. User operates two buttons to activate suction or spray functions. A soft silicone nasal mask guides spray mist; interchangeable silicone tips (gourd-shaped or long) contact nostrils to collect mucus into a 6 mL collection cup. Device is powered by a 1200 mAh rechargeable battery. By softening secretions with the spray function, the device facilitates more effective suction-based removal, potentially improving patient comfort and breathing. Safety is managed through biocompatible materials and adherence to home-use electrical standards.
Clinical Evidence
Bench testing only. Performance evaluations included suction vacuum range (54-66 kPa), spray capacity (0.24-0.48 mL/6s), and biocompatibility (ISO 10993). Electrical safety and electromagnetic compatibility were verified per IEC 60601-1 and IEC 60601-1-2. No clinical data provided.
Technological Characteristics
Materials: ABS, PCTG, silicone, stainless steel, PET. Sensing/Actuation: Electric pump for negative pressure. Energy: 1200 mAh rechargeable battery (DC 5V 2A). Form factor: Handheld (77.4 x 50 x 148.8 mm, 200g). Connectivity: None. Sterilization: Not specified (user-cleanable). Software: Not applicable (mechanical/electrical control). Standards: ISO 10993, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62304, IEC 62366-1.
Indications for Use
Indicated for children aged 2-12 years for the removal of nasal secretions and mucus via suction in a home environment. The spray function is indicated for children over 3 years old to moisturize and soften nasal secretions to facilitate aspiration.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 26, 2026
Shenzhen Root Innovation Technology Co., Ltd.
Rex Hsieh
Regulatory Affairs
#2-201, Floor 2, Hasee Computer Bldg., # 2 Beier Rd.
Bantian St., Longgang District
Shenzhen, 518129
China
Re: K253629
Trade/Device Name: Momcozy Nasal Aspirator (BN007)
Regulation Number: 21 CFR 878.4780
Regulation Name: Powered Suction Pump
Regulatory Class: Class II
Product Code: BTA, KMA
Dated: April 26, 2026
Received: April 27, 2026
Dear Rex Hsieh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253629 - Rex Hsieh
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253629 - Rex Hsieh
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOYCE C.
LIN-S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
Submission Number (if known)
K253629
Device Name
Momcozy Nasal Aspirator (BN007)
Indications for Use (Describe)
The device is designed for using suction to remove nasal secretion and mucus in children (age 2~12 years old) at home environment. The additional spray function helps moisturize and soften nasal secretions in children (over 3 years old), making them easier to aspirate.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
AViTA
Traditional 510(k)
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92.
# 1 APPLICANT INFORMATION
Applicant: Shenzhen Root Innovation Technology Co., Ltd.
#2-201, Floor 2 Hasee Computer Building, No. 2 Beier Rd,
Bantian Street, Longgang Shenzhen 518129 CHN
Applicant Contact: Rex hsieh
Email: rex_hsieh@avita.com.tw
Phone Number: +886-2-8512-1568 ext.5306
Fax Number: +886-2-8512-1347
Date prepared: May 25, 2026
# 2 SUBJECT DEVICE
Trade/Proprietary Name: Momcozy
Common Name: Momcozy Nasal Aspirator BN007
Review Panel: General & Plastic Surgery
Classification Product Code: BTA, KMA
Regulation Number: 878.4780
Device Class: II
# 3 PREDICATE DEVICE
510(k) Number: K243138
Trade Name: Geon (S2) Nasal Aspirator
Manufacturer: Geon Corporation
510(k) Number: K241852
Trade Name: Nasal Aspirator (NS 13)
Manufacturer: AViTA Corporation
{5}
AVITA
Traditional 510(k)
# 4 DEVICE DESCRIPTION
Momcozy Nasal Aspirator BN007 consists of a pump that can generate negative pressure suction, a silicone tip for contacting the nostrils, a collection cup for collecting nasal mucus, and a sprayer module including a water tank and a soft silicone sprayer nasal mask that guides the spray mist toward the nostrils to moisturize and soften nasal secretions prior to suction. The device also includes two buttons that activate the suction and sprayer functions.
-Specifications of device.
| Property | Specification |
| --- | --- |
| Model | BN007 |
| Target Population | Children aged 2–12 years |
| Use Environment | Home use |
| Aspiration Tips | Soft silicone (ISO 10993 tested) |
| Suction Pressure Range | 54Kpa~66Kpa (tolerance: 3Kpa) |
| Spray Capacity | 0.24 ml-0.48ml/6sec |
| Sound Pressure Level | ≤60dB |
| Water Tank Capacity | Approx. 5 mL |
| Collection Cup Capacity | Approx. 6 mL |
| Operating Environment | Temperature: 16 °C – 35 °C (60.8 °F – 95 °F) Relative Humidity: 15 % – 85 % (non-condensing) |
| Storage / Transportation Environment | Temperature: –25 °C – +55 °C (–13 °F – +131 °F) Relative Humidity: 15 % – 85 % (non-condensing) |
| Battery Capacity | 1200 mAh (rechargeable battery, DC 5 V 2 A) |
| Main Materials | ABS ,PCTG ,silicone, Stain steel, PET |
| Dimensions | Approx. 77.4 mm (W) × 50 mm (D) × 148.8 mm (H) |
| Weight | Approx. 200 g |
{6}
AVITA
Traditional 510(k)
| Tip type | Picture | Suggestions for use: | Dept h: | Size: OD x D x L (mm) |
| --- | --- | --- | --- | --- |
| Gourd shape tip | | With a stopper that does not go deep into. | Shall ow | Φ5xΦ2.6x35.68 |
| Long tip | | Recommended for use when mucus is thinner. | Deep | Φ5xΦ3x35.68 |
# 5 INDICATIONS FOR USE
The device is designed for using suction to remove nasal secretion and mucus in children (age 2~12 years old) at home environment. The additional spray function helps moisturize and soften nasal secretions in children (over 3 years old), making them easier to aspirate.
Page 3 of 11
{7}
AViTA
Traditional 510(k)
# 6 IDENTIFICATION OF THE PREDICATE DEVICE
| NO. | Device Name | Manufacture | Classification and Code | Classification regulation | 510K number |
| --- | --- | --- | --- | --- | --- |
| Subject Device | Momcozy Nasal Aspirator BN007 | AViTA Corporation | Class II BTA, KMA | 878.4780 | K253629 |
| Predicate Device | Geon (S2) Nasal Aspirator | Geon Corporation | Class II BTA | 878.4780 | K243138 |
| Predicate Device | Nasal Aspirator (NS 13) | AViTA Corporation | Class II BTA, KMA | 878.4780 | K241852 |
# 7 NON-CLINICAL TESTING
A series of safety, essential performance, and biocompatibility tests were conducted to evaluate the safety and effectiveness of Momcozy Nasal Aspirator BN007. The tests listed below were conducted in accordance with:
- ISO 10993-1: Biological Evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
- ISO 10993-5: Biological Evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for Sensitization.
- ISO 10993-23: Biological evaluation of medical devices – Part 23: Tests for irritation.
- IEC 60601-1 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2- Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC 60601-1-11-Medical electrical equipment - Part 1-11: General requirements for basic safety and
{8}
AViTA
Traditional 510(k)
essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 62304 - Medical device software - Software life-cycle processes
- IEC 62366-1 - Medical devices - Application of usability engineering to medical devices
- IEC 60601-1-6 - Medical electrical equipment General requirements for basic safety and essential performance. Collateral standard: Usability.
The evaluation of suction vacuum and spray capacity were completed during verification testing of the Momcozy BN007 (NS24) Nasal Aspirator. The test results demonstrated that the device passed all performance evaluations. For Model BN007, no measured values exceeded the specified tolerance limits after repeated operation, cleaning, and disinfection cycles. As part of demonstrating acceptable performance of the BN007 Nasal Aspirator in establishing substantial equivalence to the predicate device, the following key functional characteristics were assessed:
1. Suction Vacuum
BN007 (NS24) Measured Range:
- 54 kPa ~ 66 kPa
- Tolerance: ±3 kPa
Predicate Device:
- Geon S2 Nasal Aspirator (K243138): Claimed suction vacuum range 54 kPa ~ 66 kPa
- AViTA NS13 Nasal Aspirator (K241852): Claimed vacuum pressure 100–120 mmHg
Note: No head-to-head testing of the predicate devices was performed. Substantial equivalence is supported by comparison to the claimed specifications of the cleared predicate devices.
→ Comparison Conclusion:
The suction vacuum of BN007 (54–66 kPa, tolerance ±3 kPa) is identical to the claimed range of the Geon S2 predicate device (K243138). Verification testing demonstrated that BN007 remained within the specified tolerance after repeated operation, cleaning, and disinfection cycles.
2. Spray Capacity
BN007 (NS24) Measured Value:
- 0.24 mL-0.48mL / 6 seconds
Predicate Device (AViTA NS13):
- Comparable spray output (non-clinical performance benchmark)
(NS13 is used as a performance benchmark due to same device type and same functional operating
Page 5 of 11
{9}
AViTA
Traditional 510(k)
principle.)
→ Comparison Conclusion:
The spray capacity of BN007 demonstrates consistent spray output and meets the expected functional performance of a nasal aspirator. Results confirm that BN007 achieves a comparable spray delivery rate to the benchmark device and does not introduce new safety or effectiveness concerns.
Overall Performance Conclusion
The BN007 Nasal Aspirator meets all manufacturer-defined specifications for suction vacuum and spray capacity. The verification data demonstrate that BN007 performs as well as the predicate devices and does not raise new questions of safety or effectiveness.
Page 6 of 11
{10}
AVITA
Traditional 510(k)
# 8 DEVICE COMPARISON TABLE
- Substantial Equivalence Differences & Justifications (Momcozy BN007)
| Item | Subject (Momcozy Nasal Aspirator BN007) Classification: 21CFR 878.4780 Product Code: BTA, KMA FDA Class II | Predicate (Geon (S2) Nasal Aspirator) Classification: 21CFR 878.4780 Product Code: BTA FDA Class II | vs. Geon S2 (K243138) | Predicate ((AViTA Nasal Aspirator NS13) Classification: 21CFR 878.4780 Product Code: BTA, KMA FDA Class II | vs. NS13 (K241852) |
| --- | --- | --- | --- | --- | --- |
| Intended use | The device is designed for using suction to remove nasal secretion and mucus in children (age 2~12 years old) at home environment. The additional spray function helps moisturize and soften nasal secretions in children (over 3 years old), making them easier to aspirate. | Geon (S2) Nasal Aspirator is a hand-held, non-invasive, electronic medical device that provides negative pressure to remove nasal secretions and mucus from children (age 2~12 years old) at home environment. | Different | The device is designed for using suction to remove nasal secretion and mucus in Children (age 2~12 years old) at home environment. The additional spray/irrigation function helps moisten and loosen/soften nasal secretions in Children (over 3 years old), making them easier to aspirate. | Same |
| Effective Vacuum Pressure | 54kPa~66kPa(tolerance:3kPa) | 54kPa~66kPa (tolerance:3kPa) | Same | Vacuum Pressure: 100-120 mmHg | Different |
| Sound Pressure Level | ≤60dB | 55~60dB(tolerance:5dB) | Different | ≤60dB | Same |
| Spray Capacity | 0.24-0.48ml/6sec | N/A | Different | 0.4~0.5 ml/10sec (2.4~3 ml/min) | Different |
| Operating Temperature Range | 16°C ~ 35°C (60.8°F ~ 95°F) 15%~85% RH | 15°C ~ 40°C (41°F~104°F) / RH15%~90% | Different | 16°C ~ 35°C (60.8°F ~ 95°F) 15%~85% RH | Same |
| Storage Temperature Range | -25°C ~ 55°C (-13°F ~ 131°F) 15%~85% RH | -25°C ~ 55°C (-13°F~158°F) / RH up to 90% | Different | -25°C ~ 55°C (-13°F ~ 131°F) 15%~85% RH | Same |
| Power Requirements | 1200mAh,5V2A | 2 x 1.5V AAA batteries | Different | 2 x 1.5V AAA batteries | Different |
| Housing | ABS | ABS | Same | ABS | Same |
| Storage pouch | No | No | Same | No | Same |
| Unit Weight | Approx.200g (Exclude batteries) | approx.120g (Exclude batteries) | Different | Approx.158g (Exclude batteries) | Different |
| Sprayer Nasal Mask | Soft silicone interface that guides spray mist toward nostrils | NA | Different | Spray/irrigation nozzle used to deliver moisturizing spray | Different |
Page 7 of 11
{11}
AVITA
Traditional 510(k)
| Tips Information | | | | | |
| --- | --- | --- | --- | --- | --- |
| | Tip type | Picture | Suggestions for use: | Depth: | Size: OD x D x L (mm) |
| | Gourd shape tip | | With a stopper that does not go deep into. | Shallow | Φ5xΦ2.6x35.6 8 |
| | Long tip | | Recommended for use when mucus is thinner. | Deep | Φ5xΦ3x35.68 8 |
| | OD: outside diameter / D: diameter Momcozy Nasal Aspirator BN007 tip type based on the anticipated use. These three types of tips are suitable for children aged 2 to 12 years. | | | | |
| Different | | | | | |
| --- | --- | --- | --- | --- | --- |
| | Tip type | Picture | Suggestions for use: | Depth: | Size: OD x D x L (mm) |
| | Gourd shape tip | | With a stopper that does not go deep into. | Shallow | Φ5xΦ2.6x35.6 8 |
| | Long tip | | Recommended for use when mucus is thinner. | Deep | Φ5xΦ3x35.68 8 |
| | | | | | |
Same
Page 8 of 11
{12}
AViTA
Traditional 510(k)
- Difference in Spray Capacity (BN007 vs. AViTA NS13)
Difference:
- BN007 spray volume: 0.24-0.48 ml / 6 sec (2.4-4.8 ml/min)
- NS13 spray volume: 0.4–0.5 ml / 10 sec (≈2.4–3 ml/min)
Assessment:
The spray capacity of BN007 partially overlaps with the claimed spray output range of AViTA NS13. Although BN007 has a wider spray output range, both devices provide a spray function intended to moisturize and soften nasal secretions prior to suction.
Justification for SE:
This difference does not impact substantial equivalence because:
1. Both devices use the spray function to support nasal moisturizing, which assists mucus removal.
2. The spray function of BN007 has the same intended functional purpose as the spray/irrigation function of AViTA NS13.
3. The difference in spray output reflects an operational design difference and does not alter the fundamental suction-based operating principle of the device.
4. Verification testing demonstrated that BN007 met its predefined spray capacity acceptance criteria.
5. No new questions of safety or effectiveness are introduced.
Conclusion:
The spray capacity difference represents an operational design difference and does not raise new questions of safety or effectiveness.
- Power Source Difference (Rechargeable vs. AAA Batteries)
Difference:
- BN007: Rechargeable 1200 mAh battery (5V/2A)
- Geon S2 / NS13: 2 × AAA batteries
Assessment:
BN007 uses a built-in rechargeable battery, while predicates use replaceable AAA batteries.
Page 9 of 11
{13}
AViTA
Traditional 510(k)
## Justification for SE:
The change in power source does not affect safety or performance because:
1. Output suction and spray functions are independently verified and within the same range as predicates.
2. Rechargeable power does not alter intended use or raise new EMC risks (EMC fully tested).
3. Device remains non-life-supporting and non-critical; power method does not influence clinical outcome.
## Conclusion:
This difference is technological but does not change intended use or introduce new risks.
## Difference in Spray Delivery Interface (BN007 vs. Predicate Devices)
### Nasal Spray Mask
#### Difference:
- BN007 includes a soft silicone sprayer nasal mask that guides the spray mist toward the nostrils.
- The Geon S2 predicate device does not include a spray function.
- The AViTA NS13 predicate device uses a sprayer head/nozzle to deliver moisturizing spray toward the nasal entrance.
#### Assessment:
The sprayer nasal mask in BN007 functions as a soft silicone interface that guides the spray mist toward the nasal entrance during the spray operation. This component does not independently generate pressure, suction, or aerosol output. Instead, it serves only as a physical interface that directs the spray mist toward the nostrils.
## Justification for SE:
The design difference does not impact substantial equivalence because:
1. The sprayer nasal mask performs the same functional role as the spray delivery interface used in the
Page 10 of 11
{14}
AViTA
Traditional 510(k)
predicate device NS13.
2. The component only guides the moisturizing spray and does not alter the fundamental operating principle of the device, which is suction-based removal of nasal secretions.
3. The spray function is intended only to moisturize and soften nasal secretions prior to suction and does not introduce new therapeutic functions or additional risks.
**Conclusion:**
The addition of the sprayer nasal mask represents a minor design variation in the spray delivery interface and does not raise new questions of safety or effectiveness.
# 9 SUBSTANTIAL EQUIVALENCE CONCLUSIONS
The differences between the BN007 and the predicate devices—including suction pressure range (compared with Geon S2), spray capacity (compared with AViTA NS13), power source, device weight, and accessory variations—do not introduce new questions of safety or effectiveness. The fundamental operating principle remains the same: both the subject and predicate devices generate controlled negative pressure to remove nasal secretions in children in a home environment. The additional spray function serves only to moisturize and soften nasal secretions prior to suction and does not alter the clinical purpose of the device. Therefore, the BN007 is substantially equivalent to the identified predicate devices.
Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.