Clarius Ejection Fraction AI

K253593 · Clarius Mobile Health Corp. · QIH · Mar 2, 2026 · Radiology

Device Facts

Record IDK253593
Device NameClarius Ejection Fraction AI
ApplicantClarius Mobile Health Corp.
Product CodeQIH · Radiology
Decision DateMar 2, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, PCCP

Intended Use

Clarius Ejection Fraction AI is intended for semi-automatic non-invasive measurement of the left ventricular ejection fraction on ultrasound data acquired by the Clarius Ultrasound Scanner (i.e., phased array and curvilinear scanners). The user shall be a healthcare professional trained and qualified in ultrasound. The user retains the responsibility of confirming the validity of the measurements based on standard practices and clinical judgment. Clarius Ejection Fraction AI is intended for use in adult patients only.

Device Story

Clarius Ejection Fraction AI is a machine learning-based software integrated into the Clarius Ultrasound App. It processes cardiac ultrasound images (PLAX, PSAX, AP4, AP2 views) acquired by Clarius phased array or curvilinear scanners. The device uses a deep learning image segmentation algorithm to identify the left ventricle in End Diastolic (ED) and End Systolic (ES) phases. It automatically captures frames, segments anatomy, and calculates ejection fraction. Users can manually adjust measurements via caliper crosshairs. Used in professional healthcare settings (hospitals, clinics) by trained clinicians. The output serves as an assistive tool to inform clinical management; clinicians retain responsibility for final assessment. Benefits include standardized, semi-automated cardiac measurements, potentially improving efficiency and consistency in cardiac evaluations.

Clinical Evidence

Retrospective verification study (n=279 exams) compared AI measurements to human expert mean. Primary endpoint: non-inferiority of AI to human reviewers (equivalence margin 10%). Results showed non-inferiority with statistically significant p-values (e.g., 5.57e-21 for Apical view). ICC values (0.78) indicated moderate-to-good reliability between AI and human reviewers. Clinical validation study confirmed usability and performance in simulated environments.

Technological Characteristics

Ultrasound image processing software; deep learning neural network (DNN) architecture; non-adaptive ML algorithm; integrated into Clarius App (iOS/Android); compatible with Clarius PA HD3, PAL HD3, C3 HD3 transducers. Complies with IEC 62304 (software lifecycle), ISO 14971 (risk management), and IEC 62366-1 (usability).

Indications for Use

Indicated for semi-automatic non-invasive measurement of left ventricular ejection fraction in adult patients using ultrasound data from Clarius phased array or curvilinear scanners. Intended for use by trained healthcare professionals as an assistive tool.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 2, 2026 Clarius Mobile Health Corp. Agatha Szeliga Director, Regulatory Affairs 205-2980 Virtual Way Vancouver, BC V5M 4X3 Canada Re: K253593 Trade/Device Name: Clarius Ejection Fraction AI Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: January 30, 2026 Received: January 30, 2026 Dear Agatha Szeliga: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253593 - Agatha Szeliga Page 2 established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. {2} K253593 - Agatha Szeliga Page 3 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K253593 | | | Device Name Clarius Ejection Fraction AI | | | Indications for Use (Describe) | | | Clarius Ejection Fraction AI is intended for semi-automatic non-invasive measurement of the left ventricular ejection fraction on ultrasound data acquired by the Clarius Ultrasound Scanner (i.e., phased array and curvilinear scanners). The user shall be a healthcare professional trained and qualified in ultrasound. The user retains the responsibility of confirming the validity of the measurements based on standard practices and clinical judgment. Clarius Ejection Fraction AI is intended for use in adult patients only. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K253593 # 510(k) Summary This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR § 807.92. **Subject Device Trade Name:** Clarius Ejection Fraction AI **Device Classification Name:** Automated Radiological Image Processing Software **Regulation Number, Name and Product Code:** | Regulation Number | Regulation Name | Product Code | | --- | --- | --- | | 21 CFR § 892.2050 | Medical Image Management and Processing System | QIH | **FDA 510(k) Review Panel:** Radiology **Classification:** Class II **Manufacturer:** Clarius Mobile Health Corp. 205-2980 Virtual Way Vancouver, BC V5M 4X3 Canada **Contact Name:** Agatha Szeliga Director, Regulatory Affairs agatha.szeliga@clarius.com **Date 510(k) Summary Prepared:** February 28, 2026 **Predicate Device Information:** | Device Trade Name: | Caption Interpretation Automated Ejection Fraction Software | | --- | --- | | 510(k) Reference: | K210747 | | Manufacturer Name: | Caption Health, Inc. | | Regulation Name: | Medical Image Management and Processing System | | Device Classification Name: | Automated Radiological Image Processing Software | | Primary Product Code: | QIH | | Regulation Number: | 21 CFR § 892.2050 | | Regulatory Class: | Class II | Note: The predicate device has not been subject to a design-related recall. **Device Description** Clarius Ejection Fraction AI is a machine learning algorithm that is integrated into the Clarius App software as part of the complete Clarius Ultrasound Scanner system for use in cardiac ultrasound applications, specifically intended for use by trained healthcare practitioners for semi-automatic real-time Page 1 of 15 {5} C clarius ultrasound anywhere K253593 measurement of the left ventricular (LV) ejection fraction (EF) on ultrasound image data acquired by the Clarius Ultrasound Scanner system (i.e., phased array and curvilinear scanners) using a deep learning image segmentation algorithm. During the ultrasound imaging procedure, the anatomical site is selected through a preset software selection (i.e., Cardiac Basic, Cardiac Advanced) from the Clarius App in which Clarius Ejection Fraction AI will engage when enabled by the user to place a segmentation mask or landmark markers on the ultrasound image to identify the left ventricle (LV) in both End Diastolic (ED) and End Systolic (ES) phases. Using the segmentation volume or landmark markers in both phases, Clarius Ejection Fraction AI will calculate the EF of the cardiac images obtained in Parasternal Long Axis (PLAX), Parasternal Short Axis (PSAX), and Apical (AP4, AP2) views. Clarius Ejection Fraction AI operates by performing the following tasks: Automatic capture of the ED and ES frames used to create the EF measurement Automatic calculations and measurements for the left ventricular ejection fraction. The user has the option to manually adjust the measurements made by Clarius Ejection Fraction AI by moving the caliper crosshairs. Clarius Ejection Fraction AI does not perform any functions that could not be accomplished manually by a trained and qualified user. Clarius Ejection Fraction AI is an assistive tool intended to inform clinical management and is not intended to replace clinical decision-making. The clinician retains the ultimate responsibility of ascertaining the measurements based on standard practices and clinical judgment. Clarius Ejection Fraction AI is indicated for use only in adult patients. Clarius Ejection Fraction AI is integrated into the Clarius App software, which is compatible with iOS and Android operating systems two versions prior to the latest iOS or Android stable release build and is intended for use with the following Clarius Ultrasound Scanner system transducers (previously 510(k)-cleared in K213436 and K232704). Clarius Ejection Fraction AI is not a stand-alone software device. | Clarius Ultrasound Transducers | PA HD3; PAL HD3; C3 HD3 | | --- | --- | | Clarius App Software | Clarius Ultrasound App (Clarius App) for iOS; Clarius Ultrasound App (Clarius App) for Android | # Indications for Use for Clarius Ejection Fraction AI Clarius Ejection Fraction AI is intended for semi-automatic non-invasive measurement of the left ventricular ejection fraction on ultrasound data acquired by the Clarius Ultrasound Scanner (i.e., phased array and curvilinear scanners). The user shall be a healthcare professional trained and qualified in ultrasound. The user retains the responsibility of confirming the validity of the measurements based on standard practices and clinical judgment. Clarius Ejection Fraction AI is intended for use in adult patients only. {6} C clarius ultrasound anywhere K253593 # Comparison of the Subject Device and Legally Marketed Device for Demonstration of Substantial Equivalence The following table provides a comparison of the subject device, Clarius Ejection Fraction AI, to the predicate device, Caption Interpretation Automated Ejection Fraction Software. The comparison of the subject device to the legally marketed device shows that the subject device has the same intended use, similar indications for use, the same principle of operation, and is based on a similar AI/ML algorithm for measurement of left ventricular ejection fraction, comparable to the legally marketed device referenced herein. Page 3 of 15 {7} C clarius ultrasound anywhere K253593 Table 1 - Comparison of the Subject Device to the Legally Marketed Device | Criteria | SUBJECT DEVICE | PREDICATE DEVICE | RATIONALE (if subject device differs from predicate device) | | --- | --- | --- | --- | | Device Trade Name | Clarius Ejection Fraction AI | Caption Interpretation Automated Ejection Fraction Software | | | 510(k) Holder/ Manufacturer | Clarius Mobile Health Corp. | Caption Health, Inc. | Not applicable | | Submission Reference | Current Submission | K210747 | Not applicable | | Primary Product Code | QIH | QIH | Same as predicate device. | | Device Classification Name | Automated Radiological Image Processing Software | Automated Radiological Image Processing Software | Same as predicate device. | | Regulation Name | Medical Image Management and Processing System | Medical Image Management and Processing System | Same as predicate device. | | Regulation Number | 21 CFR § 892.2050 | 21 CFR § 892.2050 | Same as predicate device. | | Intended Use | Intended for use as an assistive tool utilizing an artificial intelligence/machine learning-based algorithm for semi-automated measurement of cardiac ultrasound images for determination of left ventricular ejection fraction. | Intended for use as an assistive tool utilizing an artificial intelligence/machine learning-based algorithm for semi-automated measurement of cardiac ultrasound images for determination of left ventricular ejection fraction. | Same as predicate device. | | Indications for Use | Clarius Ejection Fraction AI is intended for semi-automatic non-invasive measurement of the left ventricular ejection fraction on ultrasound data acquired by the Clarius Ultrasound Scanner (i.e., phased array and curvilinear scanners). The user shall be a healthcare professional trained and qualified in ultrasound. The user retains the responsibility of confirming the validity of the | The Caption Interpretation Automated Ejection Fraction software is used to process previously acquired transthoracic cardiac ultrasound images, to store images, and to manipulate and make measurements on images using an ultrasound device, personal computer, or a compatible DICOM-compliant PACS system in order to provide automated estimation of left ventricular | Equivalent to the predicate device. Both the subject device and the predicate device are indicated for automated/semi-automated measurement of the left ventricular ejection fraction on ultrasound data acquired using ultrasound devices. Both devices are intended for use as assistive “tools” to aid the clinician in performing cardiac assessments. The minor differences in the indications for use between the subject device and the predicate | {8} K253593 | Criteria | SUBJECT DEVICE | PREDICATE DEVICE | RATIONALE (if subject device differs from predicate device) | | --- | --- | --- | --- | | Device Trade Name | Clarius Ejection Fraction AI | Caption Interpretation Automated Ejection Fraction Software | | | | measurements based on standard practices and clinical judgment. Clarius Ejection Fraction AI is intended for use in adult patients only. | ejection fraction. This measurement can be used to assist the clinician in a cardiac evaluation. The Caption Interpretation Automated Ejection Fraction Software is indicated for use in adult patients. | device do not impact the safety and effectiveness of the subject device relative to the predicate device. | | Radiological application/Supported modality | Ultrasound | Ultrasound | Same as predicate device. | | Principle of Operation/Technology | Ultrasound image processing software implementing artificial intelligence utilizing non-adaptive machine learning algorithms trained with clinical and/or artificial data intended for measurements of cardiac ultrasound data. | Ultrasound image processing software implementing artificial intelligence utilizing non-adaptive machine learning algorithms trained with clinical and/or artificial data intended for measurements of cardiac ultrasound data. | Same as predicate device. | | Quantitative and/or Qualitative Analysis | LV EF measurement | LV EF measurement | Same as predicate device. | | Segmentation | Yes – Segmentation of anatomical structures (cardiac anatomy/heart) | Yes – Segmentation of anatomical structures (cardiac anatomy/heart) | Same as predicate device. | | Measurement | Yes – Measurement of LV ejection fraction | Yes – Measurement of LV ejection fraction | Same as predicate device. | | Algorithm Methodology | Artificial Intelligence (AI)/Machine Learning (ML) | Artificial Intelligence (AI)/Machine Learning (ML) | Same as predicate device. | | Automation (Yes or No) | Yes | Yes | Same as predicate device. | | Manual adjustment/Manual editing capability (Yes or No) | Yes | Yes | Same as predicate device. | Page 5 of 15 {9} K253593 | Criteria | SUBJECT DEVICE | PREDICATE DEVICE | RATIONALE (if subject device differs from predicate device) | | --- | --- | --- | --- | | Device Trade Name | Clarius Ejection Fraction AI | Caption Interpretation Automated Ejection Fraction Software | | | Environment of Use | Professional healthcare setting (e.g., hospital, clinic) | Professional healthcare setting (e.g., hospital, clinic) | Same as predicate device. | | Anatomical Site | Heart | Heart | Same as predicate device. | | Intended Users | Licensed healthcare professionals | Licensed healthcare professionals | Same as predicate device. | | Patient Population | Adults | Adults | Same as reference device. | Page 6 of 15 {10} C clarius ultrasound anywhere K253593 # Non-Clinical Performance Testing Summary Clarius Ejection Fraction AI was designed and developed by Clarius Mobile Health Corp. in accordance with the applicable requirements, design controls, and standards to establish safety and effectiveness of the device. Non-clinical performance testing has demonstrated that Clarius Ejection Fraction AI complies with the following FDA-recognized consensus standards: | Standard Recognition Number | Title of Standard | | --- | --- | | 13-79 | IEC 62304:2006 + A1:2015 - Medical device software — Software life cycle processes | | 5-125 | ISO 14971:2019 Medical devices — Application of risk management to medical devices | | 5-129 | IEC 62366-1:2015 + A1:2020 Medical devices — Part 1: Application of usability engineering to medical devices | | 5-134 | ISO 15223-1:2021 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied | Safety and performance of Clarius Ejection Fraction AI have been evaluated through verification and validation testing in accordance with applicable specifications, acceptance criteria, and performance standards. The traceability analysis provides traceability between the requirement specifications, design specifications, risks, and verification testing of the subject device. All requirements and risk controls have been successfully verified and traced. A comprehensive risk analysis was performed for the subject device and appropriate risk controls have been implemented to mitigate hazards. Software verification and validation activities were conducted in accordance with IEC 62304:2006 + AMD1:2015 – Medical device software – Software lifecycle processes and ISO 14971:2019 Medical devices – Application of risk management to medical devices, and in accordance with relevant FDA guidance documents, General Principles of Software Validation, Final Guidance for Industry and FDA Staff (issued January 11, 2002), Guidance for the Content of Premarket Submissions for Device Software Functions (issued June 14, 2023), and Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (issued September 27, 2023). Cybersecurity and vulnerability analyses were conducted, and it has been determined that Clarius conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient. The following processes were followed and applied during the design and development of Clarius Ejection Fraction AI: - Risk Analysis - Design Reviews - Integration Testing - System Testing - Performance Testing - Usability Engineering - Software Verification & Validation {11} K253593 - Cybersecurity Analysis Clarius Ejection Fraction AI was tested and was found to be safe and effective for the intended use, intended users, intended patient population, and use environments, as demonstrated through verification and validation testing evaluating its clinical usage and performance. Validation testing was performed to ensure that the final product meets the requirements for the specified clinical application and performs as intended to meet users' needs, while demonstrating substantial equivalence to the predicate device. ## Clinical Performance Evaluation Summary Following the completion of Clarius Ejection Fraction AI model algorithmic development (i.e., training, tuning/ validation, and internal testing), which was intended to create a documented baseline of the AI model, clinical verification testing and clinical design validation were performed to evaluate its clinical performance. Data used for model development was collected from the Clarius Cloud and/or partner clinics and was partitioned by unique anonymous patient identifiers to ensure there was no data overlap between the training, internal testing, and clinical verification datasets. As part of the truthing process, Clarius only included data from the institutions/clinical sites that were not represented in the data used for algorithmic development of the Clarius Ejection Fraction AI model (i.e., training, tuning, and internal testing data) to prevent data leakage. The exclusion criteria used were that images of inadequate quality were not added to the sample size (non-diagnostic images with artifacts obstructing specific anatomy) and images with incomplete anatomy and views. In measurement comparisons, Clarius excluded the subjects where the Ejection Fraction AI model failed to generate a measurement since there was no value to compare. To aggregate measurements from different truthers, the mean of the three values was taken and was treated as one reviewer mean. No clinical information was provided to the clinicians regarding patients utilized in the clinical truthing process. The clinicians only had access to the ultrasound image for identifying the cardiac anatomy, segmenting the left ventricle, and performing measurements. The lighting and monitor size/resolution were operator-dependent using their clinical judgement. The truthing process was not based on any follow-up medical examination. The clinical performance of Clarius Ejection Fraction AI was evaluated through a retrospective analysis of anonymized ultrasound images obtained from multiple clinical sites predominantly from the United States, representing different ethnic groups, genders, and ages. The clinical verification data to evaluate the clinical performance of Clarius Ejection Fraction AI was entirely independent from the training, tuning (validation) and internal testing datasets used in the development of the AI model. The Clarius Ejection Fraction AI Deep Neural Network (DNN) model was developed and trained using three data sets: training, tuning, and internal testing. The DNN parameters and weights were updated on the training data and evaluated on the validation (tuning) data at each epoch. Once the AI model was fully trained, its generalizability was tested by evaluating it on the internal testing dataset (internal testing prior to clinical (external) verification). The internal test data was fully independent of the training/tuning dataset and was labelled by experts. Then, following internal testing, a single model was selected, and a completely separate test dataset was used for performance testing of the AI model (clinical verification). This verification dataset was independent of the training/tuning, and internal testing datasets, in order to ensure robust results. ## Summary of the Clinical Verification Study Ultrasound images were randomly obtained from an anonymized multi-center database of images from the United States, Canada, Germany, Turkey, United Kingdom, Philippines, Australia, Italy, Sweden, Page 8 of 15 {12} C clarius ultrasound anywhere K253593 Mexico, Belgium, Singapore, El Salvador, Lithuania, Norway, Venezuela, Malaysia, Switzerland, South Africa, Indonesia, Greece, Nigeria, New Zealand, Austria, Morocco, Iraq, South Korea, Jamaica, Israel, Taiwan, The Netherlands, Dominican Republic, Uganda, Ireland, Bahrain, and Vatican, representing various ethnicities, genders, and ages of the subjects. The verification study was conducted using de-identified ultrasound data previously collected and stored on a cloud platform. No clinical or sociodemographic information—such as age, gender, or clinical diagnosis—was available or accessible at any point during the study. This data was fully anonymized prior to Clarius' access and use, in accordance with applicable privacy laws and ethical guidelines. Institutions included in the Clarius Ejection Fraction AI model development (i.e., training, tuning, and internal testing datasets) were excluded from this study. Images of the cardiac anatomy were collected and the total sample size included in the study was 279 exams, with the majority representing patients from the United States. The geographic distribution of data collected is shown in Table 1: | Table 1: Geographic Data | | | --- | --- | | Location | Number of Images | | United States | 72 | | Canada | 44 | | Germany | 22 | | Unknown | 21 | | Turkey | 18 | | United Kingdom | 10 | | Philippines | 9 | | Australia | 8 | | Italy | 7 | | Sweden | 7 | | Mexico | 6 | | Belgium | 5 | | Singapore | 5 | | El Salvador | 4 | | Lithuania | 4 | | Norway | 3 | | Venezuela | 3 | | Malaysia | 2 | | Switzerland | 2 | | South Africa | 2 | | Indonesia | 2 | | Greece | 2 | | Nigeria | 2 | | New Zealand | 2 | | Austria | 2 | | Morocco | 2 | | Iraq | 2 | | South Korea | 1 | {13} K253593 | Jamaica | 1 | | --- | --- | | Israel | 1 | | Taiwan | 1 | | The Netherlands | 2 | | Dominican Republic | 1 | | Uganda | 1 | | Ireland | 1 | | Bahrain | 1 | | Vatican | 1 | | Total | 279 | The primary objective of the retrospective verification study was to determine whether Clarius Ejection Fraction AI measurements are non-inferior to those obtained manually by human experts/qualified ultrasound users by determining if the magnitude of the mean absolute difference between Clarius Ejection Fraction AI and mean reviewer measurements is greater than the magnitude of the mean absolute difference among reviewers themselves. The significance level was set to 0.025, and the equivalence margin was set at 10% (0.10). The secondary objective was to determine the correlation between Clarius Ejection Fraction AI predictions and those of human experts among the different Clarius scanner models (i.e., C3 HD3, PA HD3). Each reviewer was blinded to the Clarius Ejection Fraction AI output and the other reviewers' annotations. All ultrasound exams were captured using Clarius' 510(k)-cleared curvilinear and phased array ultrasound scanners. An assessment of the magnitude of the difference between Clarius Ejection Fraction AI and human experts' ejection fraction measurement data was performed to ascertain whether Clarius Ejection Fraction AI measurement is non-inferior to those of human experts/ qualified ultrasound users. The mean absolute difference between reviewer pairs was calculated and compared to the mean absolute difference between the Clarius Ejection Fraction AI measurement and mean reviewer measurement using a one-sided t-test and an equivalence/err or margin of 10%. The automatic LV EF measurement was found to be non-inferior to that of experienced ultrasound users as shown by statistically significant p-values of 5.57e-21 (97.5%CI: -inf, -3.00), 1.57e-36 (97.5%CI: -inf, -2.1) and 1.12e-18 (97.5%CI: -inf, -2.38) for the Apical, PSAX and PLAX views respectively. The automatic EF measurement was found to be non-inferior with statistically significant p-values for the various views/measurement methods. The non-inferiority performance testing summary is shown in Table 2: | Table 2: Non-Inferiority (T-Test) Result Summary for Clinical Performance of Clarius Ejection Fraction AI | | | | | | | --- | --- | --- | --- | --- | --- | | EF measurement method | Scanning view | p-value | t-value | Equivalence Margin | Mean Difference | | Simpsons single plane | Apical | 5.57e-21 (97.5%CI: -inf, -3.00) | -11 | 10 | -6.27 | clarius ultrasound anywhere Page 10 of 15 {14} C clarius ultrasound anywhere K253593 The Intraclass Correlation Coefficient (ICC) was calculated to show reliability among reviewers and Clarius EF AI for the various views/measurement methods, as shown in Table 3 below: Table 3: ICC values of Reviewers and Clarius Ejection Fraction AI | Comparison Pair | ICC | 95% CI | | --- | --- | --- | | Reviewer1 vs. Reviewer2 | 0.67 | [0.59 0.74] | | Reviewer1 vs. Reviewer3 | 0.64 | [0.53 0.71] | | Reviewer2 vs. Reviewer3 | 0.54 | [0.41 0.64] | | AI_EF vs. Mean_Reviewers | 0.78 | [0.71 0.83] | The results of the clinical verification study (retrospective analysis) evaluating the performance of Clarius Ejection Fraction AI have demonstrated that Clarius Ejection Fraction AI's performance is non-inferior to that of experienced ultrasound reviewers/clinicians for measurement of the left ventricular ejection fraction, thus meeting the primary objective of the study. Furthermore, the study validated that there is moderate to good correlation between human experts and Clarius EF AI across different Clarius scanners (C3 HD3 and PA HD3). Therefore, the clinical performance of Clarius Ejection Fraction AI has been adequately verified for automated left ventricular ejection fraction measurements and has been determined to be as reliable and accurate as compared to human clinical experts. # Summary of the Clinical Validation Study A clinical validation study was conducted to evaluate the design and clinical usage of Clarius Ejection Fraction AI, as it is integrated into the Clarius App software, to determine if it performs as intended in a representative user environment, meets the product requirements, is clinically usable, and meets users' needs for use in semi-automated measurements of the left ventricular ejection fraction. Testing was performed using production equivalent units in a simulated use environment. The results of the clinical validation study showed consistent results among all users, meeting the predefined acceptance criteria. The users were able to activate Clarius Ejection Fraction AI using Clarius' curvilinear and phased array ultrasound scanners (i.e., C3 HD3, PA HD3), image the cardiac anatomy to identify the left ventricle, perform live segmentation with the segmentation mask or landmark markers, perform automated measurements of the ejection fraction of the cardiac images obtained in PLAX, PSAX, and/or Apical (AP2 and/or AP4) chamber views, visualize the ES and ED frames that the AI uses in calculating the EF, manually adjust the measurements, change the segmentation mask opacity, and display and save the LV EF measurement with each exam. Therefore, based on the results of the clinical validation study it has been determined that Clarius Ejection Fraction AI performs as intended and meets user needs for use in semi-automated left ventricular ejection fraction measurements in cardiac ultrasound applications. {15} C clarius ultrasound anywhere K253593 # Predetermined Change Control Plan (PCCP) Clarius Ejection Fraction AI uses a machine learning (ML) algorithm for automated measurement of the left ventricular ejection fraction on ultrasound image data acquired by the Clarius Ultrasound Scanner. Modifications to Clarius Ejection Fraction AI will be made in accordance with its Predetermined Change Control Plan (PCCP). The PCCP provides a description of the device's planned modifications, a modification protocol to test, verify, validate, and implement the modifications in a manner that ensures the continued safety and effectiveness of the device, while mitigating any risks associated with changes to the Ejection Fraction AI model to not adversely impact the device's performance, safety, or effectiveness associated with its indications for use, and an impact assessment of the planned modifications. The modifications outlined in the PCCP are summarized in the table below. In accordance with the PCCP, the modified Clarius Ejection Fraction AI model will be adequately trained, tuned, tested, and validated before release of the modified Ejection Fraction AI model. Implemented modifications to the Clarius Ejection Fraction AI algorithm will be communicated to users via the Clarius App software update notification and through updated labelling. Summary of planned modifications to Clarius Ejection Fraction AI per the PCCP: | Modification | Rationale | Testing Methods | Impact Assessment | | --- | --- | --- | --- | | Modification of data input sources (Clarius ultrasound scanners) | To add data from current Clarius scanners and future 510(k) cleared Clarius scanners to the Clarius Ejection Fraction AI model so the model can be deployed on more scanners. | Internal testing, clinical design validation and usability validation to assess the model's performance and ensure it performs as intended to meet users' needs. | By accommodating a wider array of image geometries and characteristics with the use of new 510(k)-cleared Clarius ultrasound scanners, the updated Ejection Fraction AI model will be better equipped to handle different transducer models of the Clarius Ultrasound Scanner used in varying clinical scenarios. Benefit-Risk Analysis: Benefits: Enhanced compatibility; Flexibility for diverse clinical settings. Risks: Data skewing and concept drift. | {16} K253593 | Modification | Rationale | Testing Methods | Impact Assessment | | --- | --- | --- | --- | | | | | **Risk Mitigation:** Internal testing to ensure that data skewing and concept drift are mitigated. | | Modification of training hyperparameters (initial learning rate, width multiplier, dropout rate) | Improvement and optimization of Clarius Ejection Fraction AI’s performance | Re-training of the Ejection Fraction AI model with modified hyperparameters to optimize its performance followed by internal testing and a comparison of the original Ejection Fraction AI model to the modified Ejection Fraction AI model (using performance metrics) followed with clinical performance testing (verification and validation). | Improved performance metrics of modified Ejection Fraction AI model with increased accuracy and more robust measurements displayed to users. **Benefit-Risk Analysis:** Benefits: Improved performance; generalization. Risks: Overfitting; unintended bias. **Risk Mitigation:** Proper regularization techniques and cross-validation and dropout will be employed to mitigate overfitting. Internal testing and verification will be conducted to mitigate unintended biases. | | Modification of post-processing steps | Improvement and optimization of Clarius Ejection Fraction AI’s performance and robustness | Internal testing and a comparison of the original Ejection Fraction AI model to the modified Ejection Fraction AI model (using performance metrics) and clinical performance | Improved performance metrics of modified Ejection Fraction AI model. **Benefit-Risk Analysis:** | clarius ultrasound anywhere Page 13 of 15 {17} K253593 | Modification | Rationale | Testing Methods | Impact Assessment | | --- | --- | --- | --- | | | | testing (verification and validation). | Benefits: Improved performance; generalization. Risks: Overfitting; unintended bias. Risk Mitigation: Proper regularization techniques and cross-validation and dropout will be employed to mitigate overfitting. Internal testing and verification will be conducted to mitigate unintended biases. | | Modification of masked autoencoder architecture and training | Optimization of model robustness, accuracy, and generalizability across diverse patient populations and scanning conditions | Internal testing and a comparison of the original Ejection Fraction AI model to the modified Ejection Fraction AI model (using performance metrics) and clinical performance testing (verification and validation). | Improved performance metrics of modified Ejection Fraction AI model with increased accuracy, improved generalizability, and more robust measurements displayed to users. Benefit-Risk Analysis: Benefits: Improved performance; generalization. Risks: Overfitting; unintended bias. Risk Mitigation: Proper regularization techniques and cross-validation and dropout will be | Page 14 of 15 {18} K253593 | Modification | Rationale | Testing Methods | Impact Assessment | | --- | --- | --- | --- | | | | | employed to mitigate overfitting. Internal testing and verification will be conducted to mitigate unintended biases. | # Conclusion & Summary of Substantial Equivalence Based on the information presented in this Traditional 510(k) premarket notification and based on the fundamental scientific technology utilizing artificial intelligence/machine learning algorithms, technological characteristics, principle of operation, intended use, intended patient population, and environment of use, Clarius Ejection Fraction AI has been determined to be substantially equivalent in terms of safety and effectiveness to the legally marketed predicate device, Caption Interpretation Automated Ejection Fraction Software (K210747). The subject device and the predicate device employ radiological (ultrasound) image processing software applications which implement artificial intelligence/machine learning algorithms trained with clinical and/or artificial data intended for analysis of ultrasound data, utilizing very similar machine-learning algorithms for detection, segmentation, and measurement of left ventricular ejection fraction. Performance testing of Clarius Ejection Fraction AI, including the results from clinical verification and validation studies, has demonstrated that Clarius Ejection Fraction AI automated measurement adequately aligns with expert clinicians' manual measurements, and thereby performs as intended for use in semi-automated cardiac ultrasound measurements of the left ventricular ejection fraction. Any differences in the indications for use or technological characteristics between the subject device and the legally marketed predicate device do not raise any issues related to safety or effectiveness. Therefore, Clarius Ejection Fraction AI is as safe and effective as the predicate device, Caption Interpretation Automated Ejection Fraction Software (K210747), and therefore substantially equivalent. Page 15 of 15
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%