K253574 · Siemens Medical Solutions USA, Inc. · JAK · Mar 20, 2026 · Radiology
Device Facts
Record ID
K253574
Device Name
SOMATOM X.cite; SOMATOM X.ceed
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Mar 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K253574 · Mar 20, 2026
SOMATOM X.cite; SOMATOM X.ceed
Siemens Medical Solutions USA, Inc.
Retrospective patient cases of respiratory 4D CT examinations
Retrospective blinded rater studies were performed by board-approved radio-oncologists and medical physicists to evaluate the performance of the HD FoV 5.0 and ZeeFree RT algorithms on clinical patient data.
Performance of ZeeFree RT reconstruction algorithm
Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of X-ray transmission data. The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions. This CT system can be used for low dose lung cancer screening in high risk populations*. * As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Device Story
SOMATOM X. Platform (X.cite, X.ceed) are single-source whole-body CT scanners; utilize continuously rotating tube-detector system (Stellar detector, Vectron tube) to acquire X-ray transmission data; reconstruct cross-sectional images in DICOM format. Operated by trained staff (physicians, radiologists, technologists) in professional healthcare facilities. Software (SOMARIS/10 syngo CT VB20) manages patient data, scan control, and image reconstruction. Features include FAST/CARE/GO technologies for automated positioning, planning, and dose management; supports optional post-processing plugins (e.g., myAblation Guide). Output images aid in diagnosis, treatment preparation, and therapy planning (e.g., radiation therapy, surgery). Benefits include optimized workflow, reduced radiation dose, and enhanced image quality for clinical decision-making.
Clinical Evidence
Bench testing only. Performance validated via phantom testing (homogeneous water, tissue-equivalent, static/dynamic torso) and retrospective blinded rater studies by board-approved radio-oncologists/physicists. Evaluated accuracy of FAST 3D camera, FAST Planning success rates (>90% without editing), and image quality of HD FoV 5.0 and ZeeFree RT algorithms. No clinical prospective/retrospective patient outcome studies reported.
Indicated for patients requiring cross-sectional imaging via CT reconstruction for diagnosis, treatment planning, and therapeutic interventions, including low-dose lung cancer screening in high-risk populations as defined by medical societies.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SOMATOM X. Platform CT Scanner Systems with SOMARIS/10 syngo CT VB10 (K233650)
SOMATOM go. Platform CT Scanner Systems with SOMARIS/10 syngo CT VB20 (K250822)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 20, 2026
Siemens Medical Solutions USA, Inc.
% Kenny M Bello
Regulatory Affairs Professional
810 Innovation Dr.
KNOXVILLE, TN 37932
Re: K253574
Trade/Device Name: SOMATOM X.cite; SOMATOM X.ceed
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: February 17, 2026
Received: February 17, 2026
Dear Kenny M Bello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253574 - Kenny M Bello
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253574 - Kenny M Bello
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253574 | ? |
| Please provide the device trade name(s). | | ? |
| SOMATOM X.cite;
SOMATOM X.ceed | | |
| Please provide your Indications for Use below. | | ? |
| This computed tomography system is intended to generate and process cross-sectional images of patients
by computer reconstruction of X-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and
radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature,
including the results of the National Lung Screening Trial (N Engl J Med 2011;
365:395-409) and subsequent literature, for further information. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
SIEMENS Healthineers
K253574
510(k) Summary
for
SOMATOM X. Platform CT Scanner Systems
with software version SOMARIS/10 syngo CT VB20
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
I. Contact Details
Submitter
Siemens Medical Solutions USA, Inc.
810 Innovation Drive
Knoxville, TN 37932
Establishment Registration Number: 1034973
Importer/Distributor
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355
Establishment Registration Number: 2240869
Location of Manufacturing Site (1)
Siemens Healthineers AG
Siemensstr. 1 -OR- Rittigfeld 1
D-91301 Forchheim, Germany
Establishment Registration Number: 3004977335
Location of Manufacturing Site (2)
SIEMENS SHANGHAI, MEDICAL EQUIPMENT LTD
278 Zhou Zhu Rd
Shanghai, CHINA, 201318
Establishment Registration Number: 3003202425
Note: Descriptions in this submission use the short company name Siemens. It covers both manufacturing locations and names as listed above. Brand name on all products is Siemens Healthineers.
Siemens Healthineers AG
1 of 32
{5}
SIEMENS Healthineers
# Submitter Contact Person:
Kenny M Bello
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
Molecular Imaging
810 Innovation Drive
Knoxville, TN 37932
Phone: (205) 856-6099
monsuru.bello@siemens-healthineers.com
# II. Device Name and Classification
Table 1: Subject Device Names
| Product Name | Trade Name |
| --- | --- |
| SOMATOM X.cite | SOMATOM X.cite |
| SOMATOM X.ceed | SOMATOM X.ceed |
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
# III. Predicate Device
## Primary predicate device:
Trade Name: SOMATOM X. Platform CT Scanner Systems with SOMARIS/10 syngo CT VB10 software
510(k) Number: K233650
Clearance Date: March 26, 2024
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
Recall Information: All predicate device recalls have been considered in the subject device design.
## Secondary predicate device:
Trade Name: SOMATOM go. Platform CT Scanner Systems with SOMARIS/10 syngo CT VB20 software
510(k) Number: K250822
Clearance Date: July 3, 2025
Siemens Healthineers AG
{6}
SIEMENS Healthineers
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
Recall Information: All predicate device recalls have been considered in the subject device design.
Note:
K233650 was a bundle submission with various Siemens CT Scanner Systems, including the primary predicate devices SOMATOM X.cite, SOMATOM X.ceed, and the CT scanner systems SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro.
K250822 was a bundle submission with various Siemens CT Scanner Systems, including the CT scanner systems SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, and SOMATOM Pro.Pulse.
IV. Device Description
Siemens intends to market a new software version, SOMARIS/10 syngo CT VB20 for the following SOMATOM Computed Tomography (CT) Scanner Systems:
SOMATOM X. Platform CT scanner systems:
- SOMATOM X.cite
- SOMATOM X.ceed
In this submission, the above listed CT scanner systems are jointly referred to as subject devices by "SOMATOM X. Platform" CT scanner systems.
The subject devices SOMATOM X. Platform CT scanner systems with SOMARIS/10 syngo CT VB20 are Computed Tomography X-ray Systems which feature one continuously rotating tube-detector system and function according to the fan beam principle (single source). The SOMATOM X. Platform CT scanner systems with software SOMARIS/10 syngo CT VB20 produces CT images in DICOM format, which can be used by trained staff for software applications, e.g. post-processing applications, commercially distributed by Siemens Healthcare GmbH and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.
Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.
The platform software for SOMATOM X. Platform, SOMARIS/10 syngo CT VB20, is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation.
Siemens Healthineers AG
{7}
SIEMENS Healthineers
The software platform provides plugin software interfaces that allow for the use of specific commercially available post processing software algorithms in an unmodified form from the cleared stand-alone post processing version.
New software version syngo CT VB20 (SOMARIS/10 syngo CT VB20) is a modified software version based on syngo CT VB10 (SOMARIS/10 syngo CT VB10) which was cleared for the predicate devices in K233650.
Software version SOMARIS/10 syngo CT VB20 will be offered ex-factory and as an optional upgrade for the applicable existing SOMATOM X. Platform CT Scanner Systems.
The bundle approach is feasible for this submission since the subject devices have similar technological characteristics, software operating platform, and supported software characteristics. The supporting data are similar, primarily one review division/group will be involved, and the indications for use is the same between the devices. All subject devices will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use remains unchanged compared to the predicate devices.
# V. Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
# VI. Indications for Use Comparison
## Subject Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Siemens Healthineers AG
{8}
SIEMENS Healthineers
## Primary Predicate Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
## Secondary Predicate Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
## Comparison:
The subject device Indications for Use is the exact same as the primary and the secondary predicate's Indications for Use (K233650). As such, the Indications for Use remains unchanged from the predicate devices and is substantially equivalent to the predicate devices.
## VII. Comparison of Technological Characteristics with the Predicate Device
Supported by the subject device, SOMARIS/10 syngo CT VB20 software version is a further development of the SOMARIS/10 syngo CT VB10 software version which is cleared in K233650.
The SOMATOM X. Platform with SOMARIS/10 syngo CT VB20 software version provides the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the predicate devices. The software features of the SOMATOM X. Platform have been modified or improved in comparison to the predicate devices to support enhanced device functionality compared to the predicate devices.
The new syngo CT VB20 software reuses all unmodified software features of the legacy software syngo CT VB10 cleared in K233650. Additionally, no features present in the predicate device are descoped.
Software version SOMARIS/10 syngo CT VB20 is designed to reuse hardware independent extended functionalities and GO technologies provided by Siemens cleared software applications.
Siemens Healthineers AG
{9}
SIEMENS Healthineers
The intended use and fundamental scientific technology for the SOMATOM X. Platform remain unchanged from the predicate devices.
At a high level, the subject and predicate devices are based on the same subset of technological elements:
- Scanner Principle – Whole body X-Ray Computed Tomography Scanner
- System Acquisition – Continuously rotating tube detector system
- Iterative Reconstruction – Support of various iterative reconstruction principles
- Workplaces – Support of workplaces that include reconstruction and image evaluation software
- Patient table
- Patient table foot switch for movement
- Stellar Detector technology
- Tin filtration technology
- Vectron X-ray Tube
- Power Generator
- Scan&GO
- Mobile workflow (Tablet)
- Optional injector arm
- Optional support of CT guided intervention workflow (myNeedle Companion – myNeedle Guide/myNeedle Laser)
- Optional support of FAST 3D Camera for patient positioning workflow
- Scanner display and control functionality
- Remote Scan Control
- Interfaces for Respiratory Gating Devices
- Standard CARE technologies
- Standard FAST technologies
- Standard GO technologies
- Precision matrix (large image matrices)
- HD FoV
- DirectDensity
- Stack artefact reduced reconstruction
The terms in the following Table 2 are used to indicate various technological characteristics in comparison to the primary and secondary predicate devices information:
Table 2: Overview of term definition.
| Term | Definition |
| --- | --- |
| Modified | This feature is modified from the predicate devices |
| Enabled | This feature is currently supported by other cleared Siemens CT systems or cleared Siemens stand-alone software applications. This feature will be supported for the subject device with software version SOMARIS/10 syngo CT VB20 and is substantially equivalent compared to the cleared version of the predicate devices. |
The subject devices SOMATOM X. Platform CT scanner systems with SOMARIS/10 syngo CT VB20 will support hardware and software modifications/further developments in comparison to the predicate
Siemens Healthineers AG
{10}
SIEMENS Healthineers
devices as listed in Table 3. The column “HW/ SW change” in the table contains reference to a hardware (HW) or/and software (SW) change. The columns under “Predicate devices” of the Table 3 indicate the clearance of the technological properties in the corresponding predicate device.
Table 3: Summary of modifications (hardware and software) of the subject devices SOMATOM X.cite and SOMATOM X.ceed (SOMATOM X. Platform) with software version SOMARIS/10 syngo CT VB20 compared to the predicate devices.
| | Technological property | HW/SW change | Subject devices | Predicate devices | |
| --- | --- | --- | --- | --- | --- |
| | | | SOMATOM X. Platform with syngo CT VB20 | SOMATOM X. Platform with syngo CT VB10 (K233650) | SOMATOM go. Platform with syngo CT VB20 (K250822) |
| 1. | CARE Moodlight | HW/SW | enabled (same as in K250822) | n.a. | cleared |
| 2. | Interfaces for respiratory gating systems – Open Online Interface | HW | enabled (same as in K250822) | n.a.¹ | cleared |
| 3. | Eco Power Mode | HW/SW | enabled (same as in K250822) | n.a. | cleared |
| 4. | FAST 3D Camera/ FAST Integrated Workflow | SW | enabled² (same as in K250822) | cleared | cleared |
| 5. | Oncology Exchange | SW | enabled (same as in K250822) | n.a. | cleared |
| 6. | DirectSetup Notes | SW | enabled | n.a. | cleared |
¹ The primary predicate devices SOMATOM X. Platform with software version syngo CT VB10 (K233650) support interfaces for respiratory gating with exception of Open Online interface. Open Online interface is now introduced with the subject devices, and it is same as cleared in the secondary predicate devices SOMATOM go. Platform with software version syngo CT VB20 (K250822).
² The subject devices SOMATOM X. Platform with software version syngo CT VB20 support FAST 3D Camera/ FAST Integrated workflow, which was cleared with the primary predicate devices with software version syngo CT VB20 (K233650). With software version syngo CT VB20, retrained FAST 3D Camera algorithms, collision indication, and Centerline/Grid Overlay are introduced. This modification is same as already cleared in the secondary predicate devices SOMATOM go. Platform with software version syngo CT VB20 (K250822).
Siemens Healthineers AG
{11}
SIEMENS Healthineers
| | Technological property | HW/SW change | Subject devices | | Predicate devices | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SOMATOM X. Platform | SOMATOM X. Platform | SOMATOM go. Platform | |
| | | | with syngo CT VB20 | with syngo CT VB10 (K233650) | with syngo CT VB20 (K250822) | |
| | | | (same as in K250822) | | | |
| 7. | FAST Planning | SW | enabled^{3}
(same as in K250822) | cleared | cleared | |
| 8. | myExam Companion – myExam Compass/myExam Cockpit | SW | enabled^{4}
(same as in K250822) | cleared | cleared | |
| 9. | HD FoV 5.0 | SW | enabled
(same as in K250822) | n.a. | cleared | |
| 10. | CT guided intervention – myAblation Guide interface | SW | enabled
(same as in K250822) | n.a. | cleared | |
| 11. | Flex 4D Spiral | SW | modified
(same as in K233650, with exception of the modifications regarding scan range and cycle time which are | cleared | cleared | |
Siemens Healthineers AG
3 The subject devices SOMATOM X. Platform with software version syngo CT VB20 support FAST Planning, which was cleared with the primary predicate devices with software version syngo CT VB20 (K233650). With software version syngo CT VB20, detection of additional body regions is introduced. This modification is same as already cleared in the secondary predicate devices SOMATOM go. Platform with software version syngo CT VB20 (K250822).
4 The subject devices SOMATOM X. Platform with software version syngo CT VB20 support FAST Planning, which was cleared with the primary predicate devices with software version syngo CT VB20 (K233650). With software version syngo CT VB20, the clinical decision trees are available not only for adults but also for children. This modification is same as already cleared in the secondary predicate devices SOMATOM go. Platform with software version syngo CT VB20 (K250822).
8 of 32
{12}
SIEMENS Healthineers
| | Technological property | HW/SW change | Subject devices | | Predicate devices | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SOMATOM X. Platform | SOMATOM X. Platform | SOMATOM go. Platform | |
| | | | with syngo CT VB20 | with syngo CT VB10 (K233650) | with syngo CT VB20 (K250822) | |
| | | | introduced with syngo CT VB20) | | | |
| 12. | ZeeFree RT | SW | enabled
(same as in K250822) | n.a. | cleared | |
| 13. | Direct Density | SW | enabled^{5}
(same as in K250822) | cleared | cleared | |
| 14. | myExam Contrast | SW | enabled
(same as in K250822) | n.a. | cleared | |
A tabular summary of the comparable hardware properties between the subject devices SOMATOM X. Platform with software version syngo CT VB20 and the predicate devices are listed in Table 4 below (modifications are in gray shaded sections).
A tabular summary of the comparable software properties between the subject devices SOMATOM X. Platform with software version syngo CT VB20 and the predicate devices are listed in Table 5 later in this document (modifications are in gray shaded sections).
Siemens Healthineers AG
9 of 32
5 The subject devices SOMATOM X. Platform with software version syngo CT VB20 support DirectDensity, which was cleared with the primary predicate devices with software version syngo CT VB20 (K233650). With software version syngo CT VB20, calculation of stopping-power ratio (Kernel St) is introduced. This modification is same as already cleared in the secondary predicate devices SOMATOM go. Platform with software version syngo CT VB20 (K250822).
{13}
SIEMENS Healthineers
Table 4: Comparison of technical hardware characteristics between subject devices SOMATOM X.cite and SOMATOM X.ceed with software version syngo CT VB20 and the primary predicate devices with software version syngo CT VB10 (K233650).
| Hardware property | Subject device SOMATOM X. Platform with SOMARIS/10 syngo CT VB20 | | Primary predicate device SOMATOM X. Platform with SOMARIS/10 syngo CT VB10 (K233650) | | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- | --- |
| | SOMATOM X.cite | SOMATOM X.ceed | SOMATOM X.cite | SOMATOM X.ceed | |
| scanner | Whole body Computed Tomography (CT) Scanner System | | Whole body Computed Tomography (CT) Scanner System | | Same as the primary predicate devices |
| System Configuration | Single Source | | Single Source | | Same as the primary predicate devices |
| Environment of Use | Professional Healthcare Facility | | Professional Healthcare Facility | | Same as the primary predicate devices |
| Generator Max. power (kW) | 90 kW or 105 kW | 105 kW or 120 kW | 90 kW or 105 kW | 105 kW or 120 kW | Same as the primary predicate devices |
| Detector technology | Stellar\(^{infinity}\) | | Stellar\(^{infinity}\) | | Same as the primary predicate devices |
| Detector volume coverage | 38.4 mm | | 38.4 mm | | Same as the primary predicate devices |
| Detector physical rows | 64 | | 64 | | Same as the primary predicate devices |
| Detector Slice width | 0.6 mm | | 0.6 mm | | Same as the primary predicate devices |
| Detector DAS channel no. | 840 | 920 | 840 | 920 | Same as the primary predicate devices |
| Detector Image slices | 128 | | 128 | | Same as the primary predicate devices |
| Tube technology | Vectron | | Vectron | | Same as the primary predicate devices |
| Tube kV steps | 70–140 kV in 10 kV steps | | 70–140 kV in 10 kV steps | | Same as the primary predicate devices |
| Tube Max. current | 1100 mA (for 90 kW) 1200 mA (for 105 kW) | 1200 mA (for 90 kW) 1300 mA (for 105 kW) | 1100 mA (for 90 kW) 1200 mA (for 105 kW) | 1200 mA (for 90 kW) 1300 mA (for 105 kW) | Same as the primary predicate devices |
| Tube | • 0.6 x 0.7 / 8" | • 0.4 x 0.5 / 8" | • 0.6 x 0.7 / 8" | • 0.4 x 0.5 / 8" | Same as the primary predicate devices |
Siemens Healthineers AG
{14}
SIEMENS
Healthineers
| Hardware property | Subject device
SOMATOM X. Platform
with SOMARIS/10 syngo CT
VB20 | | Primary predicate device
SOMATOM X. Platform
with SOMARIS/10 syngo CT
VB10
(K233650) | | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- | --- |
| | SOMATOM X.cite | SOMATOM X.ceed | SOMATOM X.cite | SOMATOM X.ceed | |
| focus | • 0.8 x 1.1 / 8" | • 0.6 x 0.7 / 8"
• 0.8 x 1.1 / 8" | • 0.8 x 1.1 / 8" | • 0.6 x 0.7 / 8"
• 0.8 x 1.1 / 8" | |
| Tube
Heat storage capacity | higher than 30 MHU | | higher than 30 MHU | | Same as the primary predicate devices |
| Tube
Cooling rate | 2.7 MHU / min | 2.7 MHU / min | 2.7 MHU / min | 2.7 MHU / min | Same as the primary predicate devices |
| Gantry
Bore size | 82 cm | | 82 cm | | Same as the primary predicate devices |
| Gantry
Scan FoV | 50 cm | | 50 cm | | Same as the primary predicate devices |
| Gantry
Extended FoV | 81.5 cm | | 81.5 cm | | Same as the primary predicate devices |
| Gantry
Rot. time | 0.3, 0.5, 1.0 sec | 0,25, 0.3, 0.5, 1.0 sec | 0.3, 0.5, 1.0 sec | 0,25, 0.3, 0.5, 1.0 sec | Same as the primary predicate devices |
| Gantry
Tilt (degree) | +/- 25"
+/-30 " (only with Vitus) | | +/- 25"
+/-30 " (only with Vitus) | | Same as the primary predicate devices |
| Patient table
Type | Vario RT (1600 mm)
Vitus (2000 mm)
Vario 2 (2000 mm) | Vario RT (1600 mm)
Vitus (2000 mm)
Vario 2.D (2000 mm) | Vario RT (1600 mm)
Vitus (2000 mm)
Vario 2 (2000 mm) | Vario RT (1600 mm)
Vitus (2000 mm)
Vario 2.D (2000 mm) | Same as the primary predicate devices |
| Patient table
Installation option | Regular installation (Vario RT, Vario 2 and Vitus):
474 mm
Installation option with extended distance (Vitus): | Regular installation (Vario RT, Vario 2 and Vitus):
474 mm | Regular installation (Vario RT, Vario 2 and Vitus):
474 mm
Installation option with extended distance (Vitus): | Regular installation (Vario RT, Vario 2 and Vitus):
474 mm | Same as the primary predicate devices |
Siemens Healthineers AG
11 of 32
{15}
SIEMENS
Healthineers
| Hardware property | Subject device
SOMATOM X. Platform
with SOMARIS/10 syngo CT
VB20 | | Primary predicate device
SOMATOM X. Platform
with SOMARIS/10 syngo CT
VB10
(K233650) | | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- | --- |
| | SOMATOM X.cite | SOMATOM X.ceed | SOMATOM X.cite | SOMATOM X.ceed | |
| | 674 mm (474 mm + 200 mm) | | 674 mm (474 mm + 200 mm) | | |
| Patient table
Max. weight capacity (kg) | 227 (Vario RT)
340 (Vitus)
307 (Vario 2) | 227 (Vario RT)
340 (Vitus)
307 (Vario 2.D) | 227 (Vario RT)
340 (Vitus)
307 (Vario 2) | 227 (Vario RT)
340 (Vitus)
307 (Vario 2.D) | Same as the primary predicate devices |
| Max. scan length
topogram | 1680 mm (Vario RT)
2080 mm (Vitus)
2080 mm with table extension (Vario 2) | 1680 mm (Vario RT)
2080 mm (Vitus)
2080 mm with table extension (Vario 2.D) | 1680 mm (Vario RT)
2080 mm (Vitus)
2080 mm with table extension (Vario 2) | 1680 mm (Vario RT)
2080 mm (Vitus)
2080 mm with table extension (Vario 2.D) | Same as the primary predicate devices |
| Max. scan length
Image acquisition | 1600 mm (Vario RT)
2000 mm (Vitus)
2000 mm (Vario 2)
with table extension | 1600 mm (Vario RT)
2000 mm (Vitus)
2000 mm (Vario 2.D)
with table extension | 1600 mm (Vario RT)
2000 mm (Vitus)
2000 mm (Vario 2)
with table extension | 1600 mm (Vario RT)
2000 mm (Vitus)
2000 mm (Vario 2.D)
with table extension | Same as the primary predicate devices |
| Spectral filtration | Combined Tin Filter / Split Filter, and extra Tin Filter | Combined Tin Filter / Split Filter, and extra Tin Filter | Combined Tin Filter / Split Filter, and extra Tin Filter | Combined Tin Filter / Split Filter, and extra Tin Filter | Same as in the primary predicate devices |
| FAST 3D Camera | option for patient positioning with FAST 3D Camera
installation option: ceiling mounted | | option for patient positioning with FAST 3D Camera:
installation option: ceiling mounted | | Same as the primary predicate devices |
| High Power 70/ High Power 80 | Generator 90 kW:
up to 1100 mA (@70 kV)
up to 1100 mA (@80 kV) | Generator 105 kW:
up to 1200 mA (@70 kV)
up to 1200 mA (@80 kV) | Generator 90 kW:
up to 1100 mA (@70 kV)
up to 1100 mA (@80 kV) | Generator 105 kW:
up to 1200 mA (@70 kV)
up to 1200 mA (@80 kV) | Same as the primary predicate devices |
Siemens Healthineers AG
12 of 32
{16}
SIEMENS
Healthineers
| Hardware property | Subject device
SOMATOM X. Platform
with SOMARIS/10 syngo CT
VB20 | | Primary predicate device
SOMATOM X. Platform
with SOMARIS/10 syngo CT
VB10
(K233650) | | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- | --- |
| | SOMATOM X.cite | SOMATOM X.ceed | SOMATOM X.cite | SOMATOM X.ceed | |
| | Generator 105 kW: | Generator 120 kW: | Generator 105 kW: | Generator 120 kW: | |
| | up to 1200 mA (@70 kV)
up to 1200 mA (@80 kV) | up to 1300 mA (@70 kV)
up to 1300 mA (@80 kV) | up to 1200 mA (@70 kV)
up to 1200 mA (@80 kV) | up to 1300 mA (@70 kV)
up to 1300 mA (@80 kV) | |
| x-ray foot switch | Option to trigger hands-free scanning | | Option to trigger hands-free scanning | | Same as the primary predicate devices |
| Table foot switch | Option for table patient movement | | Option for table patient movement | | Same as the primary predicate devices |
| i-joystick | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | | Same as the primary predicate devices |
| Tablet dock | Option for mounting of the tablet on the patient table. | | Option for mounting of the tablet on the patient table. | | Same as the primary predicate devices |
| CARE 2D camera | Patient observation camera integrated on gantry front | | Patient observation camera integrated on gantry front | | Same as the primary predicate devices |
| CARE Moodlight | LED lights integrated on the gantry front to indicate the system status:
• System startup
• Scan countdown
• Radiation on | | n.a. | | CARE Moodlight is cleared in K233650 for the secondary predicate devices SOMATOM go.
Platform CT scanner systems. |
| Respiratory Motion management - Interfaces for respiratory gating | acquisition of respiration-correlated scans supported
Interface connectors for:
• Anzai respiratory gating system
• Varian RGSC respiratory gating system
• Open interface compatible respiratory gating systems | | acquisition of respiration-correlated scans supported
Interface connectors for:
• Anzai respiratory gating system
• Varian RGSC respiratory gating system
• Open interface compatible respiratory gating systems | | Same as the primary predicate devices, the subject devices SOMATOM X.
Platform systems with software version syngo CT VB20 support acquisition of respiration-correlated scans. For measurement of the |
Siemens Healthineers AG
13 of 32
{17}
SIEMENS Healthineers
| Hardware property | Subject device
SOMATOM X. Platform
with SOMARIS/10 syngo CT VB20 | | Primary predicate device
SOMATOM X. Platform
with SOMARIS/10 syngo CT VB10
(K233650) | | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- | --- |
| | SOMATOM X.cite | SOMATOM X.ceed | SOMATOM X.cite | SOMATOM X.ceed | |
| | • Open Online interface | | | | breathing signal of the patient, interfaces for connecting of 3^{rd} party respiratory gating devices are offered.
Introduction of new Open Online interface.
Open Online interface is cleared for the secondary predicate devices SOMATOM go.
Platform with syngo CT VB20 in K250822. |
Table 5: Comparison of software characteristics between the subject devices SOMATOM X. Platform CT scanner systems with software version syngo CT VB20 and the primary predicate devices with software version syngo CT VB10 (K233650).
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| Operating System | Windows based
SOMARIS/10 syngo CT VB20
Note: the short version syngo CT VB20 is also used as labeling information | Windows based
SOMARIS/10 syngo CT VB10
Note: the short version syngo CT VB10 is also used as labeling information | Windows based
SOMARIS/10 syngo CT VB20
Note: the short version syngo CT VB20 is also used as labeling information | Software version upgraded due to new functionalities.
Same as in the secondary predicate device |
Siemens Healthineers AG
{18}
SIEMENS Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| Workplace | syngo Acquisition Workplace (ICS) named as “myExam Console”
Image Reconstruction System (IRS)
2^{nd} workplace option named as “myExam Satellite” with Remote Recon function | syngo Acquisition Workplace (ICS) named as “myExam Console”
Image Reconstruction System (IRS)
2^{nd} workplace option named as “myExam Satellite” with Remote Recon function | syngo Acquisition Workplace (ICS) named as “myExam Console”
Image Reconstruction System (IRS)
2^{nd} workplace option named as “myExam Satellite” with Remote Recon function | Same as the primary and secondary predicate devices |
| Standard system software | • syngo Examination
• syngo Viewing
• syngo Filming
• syngo Archiving & Network | • syngo Examination
• syngo Viewing
• syngo Filming
• syngo Archiving & Network | • syngo Examination
• syngo Viewing
• syngo Filming
• syngo Archiving & Network | Same as the primary and secondary predicate devices |
| Detector firmware | Stellar detector firmware supported | Stellar detector firmware supported | Stellar detector firmware supported | Same as the primary and secondary predicate devices |
| Teamplay | Support of Teamplay protocols | Support of Teamplay protocols | Support of Teamplay protocols | Same as the primary and secondary predicate devices |
| Protocols | Support of:
• Protocols for Radiation Therapy Planning support patient marking
• Protocols that allow scanning with support of an external respiratory gating system | Support of:
• Protocols for Radiation Therapy Planning support patient marking
• Protocols that allow scanning with support of an external respiratory gating system | Support of:
• Protocols for Radiation Therapy Planning support patient marking
• Protocols that allow scanning with support of an external | Same as the primary and secondary predicate devices |
Siemens Healthineers AG
15 of 32
{19}
SIEMENS Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| | • Protocol supporting contrast bolus-triggered data acquisition
• Contrast media protocols (including coronary CTA)
• Pediatric Protocols
• Flex Dose Profile
• Turbo Flash Spiral
• Dual Energy acquisition (TwinBeam DE and TwinSpiral DE)
• Dynamic imaging (Flex 4D Spiral)
• Protocols supporting CT Intervention (scan modes: 2D i-sequence, 3D i-sequence, i-spiral, i-Fluoro)
• Protocols supporting Cardiac Scanning
• Protocols for DirectBreathhold | • Protocol supporting contrast bolus-triggered data acquisition
• Contrast media protocols (including coronary CTA)
• Pediatric Protocols
• Flex Dose Profile
• Turbo Flash Spiral
• Dual Energy acquisition (TwinBeam DE and TwinSpiral DE)
• Dynamic imaging (Flex 4D Spiral)
• Protocols supporting CT Intervention (scan modes: 2D i-sequence, 3D i-sequence, i-spiral, i-Fluoro)
• Protocols supporting Cardiac Scanning
• Protocols for DirectBreathhold | respiratory gating system
• Protocol supporting contrast bolus-triggered data acquisition
• Contrast media protocols (including coronary CTA)
• Pediatric Protocols
• Flex Dose Profile
• Turbo Flash Spiral
• Dual Energy acquisition (TwinBeam DE and TwinSpiral DE)
• Dynamic imaging (Flex 4D Spiral)
• Protocols supporting CT Intervention (scan modes: 2D i-sequence, 3D i-sequence, i-spiral, i-Fluoro)
• Protocols supporting Cardiac Scanning
• Protocols for DirectBreathhold | |
| Advanced Reconstruction | Recon&GO:
- Spectral Recon (Dual Energy Reconstruction including Virtual Unenhanced, Monoenergetic plus | Recon&GO:
- Spectral Recon (Dual Energy Reconstruction including Virtual Unenhanced, Monoenergetic plus | Recon&GO:
- Spectral Recon (Dual Energy Reconstruction including Virtual Unenhanced, Monoenergetic plus | Same as the primary and secondary predicate devices |
Siemens Healthineers AG
16 of 32
{20}
SIEMENS
Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite SOMATOM X.ceed with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite SOMATOM X.ceed with SOMARIS/10 syngo CT VB10 (K233650) | SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 (K250822) | |
| | - Inline Results DE SPP (Spectral Post-Processing with photon-counting image data) - Inline Anatomical ranges (Parallel/Radial) incl. Virtual Unenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removal | - Inline Results DE SPP (Spectral Post-Processing with photon-counting image data) - Inline Anatomical ranges (Parallel/Radial) incl. Virtual Unenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removal | - Inline Results DE SPP (Spectral Post-Processing with photon-counting image data) - Inline Anatomical ranges (Parallel/Radial) incl. Virtual Unenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removal | |
| Image viewing | CT View&GO offers: - basic post-processing viewer (CT View&GO) - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI) - Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average | CT View&GO offers: - basic post-processing viewer (CT View&GO) - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI) - Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average | CT View&GO offers: - basic post-processing viewer (CT View&GO) - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI) - Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average | Same as the primary and secondary predicate devices |
| Software interface | • Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) | • Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical | • Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical | Same as the secondary predicate devices In software version syngo CT VB20, the subject devices |
Siemens Healthineers AG
17 of 32
{21}
SIEMENS Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| | cleared as medical devices in their own right.
• CT View&GO software interfaces (plug-ins and AV apps) for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose.
Note: The clearance of standalone post-processing functionalities software is mandatory precondition.
These advanced visualization tools are designed to support the technician & physician in the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners
• CT View&GO plug-in interface for myAblation Guide | devices in their own right.
• CT View&GO software interfaces (plug-ins and AV apps) for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose.
Note: The clearance of standalone post-processing functionalities software is mandatory precondition.
These advanced visualization tools are designed to support the technician & physician in the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners. | cleared as medical devices in their own right.
• CT View&GO software interfaces (plug-ins and AV apps) for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose.
Note: The clearance of standalone post-processing functionalities software is mandatory precondition.
These advanced visualization tools are designed to support the technician & physician in the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners | introduce plug-in interface for the stand alone software application myAblation Guide. |
Siemens Healthineers AG
18 of 32
{22}
SIEMENS Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| Cybersecurity | IT Hardening | IT Hardening | IT Hardening | Same as the primary and secondary predicate devices |
| HD FoV | HD FoV 4.0
HD FoV 5.0 | HD FoV 4.0 | HD FoV 4.0
HD FoV 5.0 | Same as the secondary predicate devices
In addition to HD FoV 4.0, the subject devices SOMATOM X.
Platform with syngo CT VB20 support a new generation of extended field of view reconstruction algorithm: HD FoV 5.0. |
| Standard technologies | FAST technologies
CARE technologies
GO technologies | FAST technologies
CARE technologies
GO technologies | FAST technologies
CARE technologies
GO technologies | Same as the primary and secondary predicate devices |
| Iterative Reconstruction Methods | ADMIRE
iMAR | ADMIRE
iMAR | ADMIRE
iMAR
SAFIRE | Same as the primary predicate devices |
| Matrix sizes | 256 x 256 pixels
512 x 512 pixels
768 x 768 pixels
1024 x 1024 pixels
(Precision Matrix) | 256 x 256 pixels
512 x 512 pixels
768 x 768 pixels
1024 x 1024 pixels
(Precision Matrix) | 256 x 256 pixels
512 x 512 pixels
768 x 768 pixels
1024 x 1024 pixels
(Precision Matrix) | Same as the primary and secondary predicate devices |
Siemens Healthineers AG
19 of 32
{23}
SIEMENS Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| DirectDensity | including relative electron density,
relative mass density,
and stopping power ratio | including relative electron density and relative mass density | including relative electron density,
relative mass density,
and stopping power ratio | Same as the secondary predicate devices
Compared to the predicate devices, the subject devices SOMATOM X.cite and SOMATOM X.ceed with syngo CT VB20 support a modified DirectDensity functionality. |
| Stack artefact reduced reconstruction | Provides images with reduced stack artifacts after the reconstruction:
ZeeFree: StackRecon types in cardiac-related examinations
ZeeFree RT: StackRecon types in respiratory-related examinations | Provides images with reduced stack artifacts after the reconstruction:
ZeeFree: StackRecon types in cardiac-related examinations | Provides images with reduced stack artifacts after the reconstruction:
ZeeFree: StackRecon types in cardiac-related examinations
ZeeFree RT: StackRecon types in respiratory-related examinations | Same as the secondary predicate devices
ZeeFree RT is a new functionality introduced with syngo CT VB20 software version of the subject devices. |
| Oncology Exchange | transfer of prescription information from the ARIA Oncology Information System (OIS) to the CT scanner | n.a. | transfer of prescription information from the ARIA Oncology Information System (OIS) to the CT scanner | Same as the secondary predicate devices
New functionality introduced with syngo CT VB20 software |
Siemens Healthineers AG
20 of 32
{24}
SIEMENS
Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| | | | | version of the subject devices. |
| DirectSetup Notes | patient and accessory setup documentation in the Scan&GO workflow | n.a. | patient and accessory setup documentation in the Scan&GO workflow | Same as the secondary predicate devices
The subject devices SOMATOM X.
Platform CT scanner systems are enabled to support DirectSetup Notes. |
| FAST Integrated Workflow (FAST 3D Camera) | FAST integrated workflow:
• FAST Range
• FAST Isocentering
• FAST Direction
• Collision Indication
• Centerline/ Grid Overlay | FAST integrated workflow:
• FAST Range
• FAST Isocentering
• FAST Direction | FAST integrated workflow:
• FAST Range
• FAST Isocentering
• FAST Direction
• Collision Indication
Centerline/ Grid Overlay | Same as the secondary predicate devices
Compared to the predicate devices, the FAST 3D Camera of the subject devices supports in syngo CT VB20 a Collision indication and Centerline/Grid Overlay function. |
| FAST Planning | detects globally used scan and recon ranges from the topogram image
additional body regions: | detects globally used scan and recon ranges from the topogram image | detects globally used scan and recon ranges from the topogram image
additional body regions: | Same as the secondary predicate devices
The subject devices SOMATOM X. |
Siemens Healthineers AG
21 of 32
{25}
SIEMENS
Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| | - Head/Neck (Vascular), Lateral/APPA
- Vascular – Runoff Knee
- Neck and chest
- Neck and Lung
- Extremities – Both knees | | - Head/Neck (Vascular), Lateral/APPA
- Vascular – Runoff Knee
- Neck and chest
- Neck and Lung
- Extremities – Both knees | Platform CT scanner systems with syngo VB20 support FAST Planning functionality with wide range of body regions. In syngo CT VB20, FAST Planning has been extended with additional body regions. |
| Eco Power Mode | reduce the energy consumption during idle times. | n.a. | reduce the energy consumption during idle times. | Same as the secondary predicate devices
New feature introduced with syngo CT VB20 of the subject devices SOMATOM X.cite and SOMATOM X.ceed. |
| Breathhold technique | Respiratory Motion Management support breath hold triggered spiral scans with manual breath hold triggered examinations.
DirectBreathold: externally triggered workflow to start a spiral scan by receiving a trigger signal from an | Respiratory Motion Management support breath hold triggered spiral scans with manual breath hold triggered examinations.
DirectBreathold: externally triggered workflow to start a spiral scan by receiving a trigger signal from an | Respiratory Motion Management support breath hold triggered spiral scans with manual breath hold triggered examinations.
DirectBreathold: externally triggered workflow to start a spiral scan by receiving a trigger signal from an | Same as the primary and the secondary predicate devices |
Siemens Healthineers AG
22 of 32
{26}
SIEMENS Healthineers
| Software property | Subject device | Primary predicate device | Secondary predicate device | Assessment of the substantial equivalency (SE) |
| --- | --- | --- | --- | --- |
| | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB20 | SOMATOM X.cite
SOMATOM X.ceed
with SOMARIS/10 syngo CT VB10
(K233650) | SOMATOM go.
Platform
with SOMARIS/10 syngo CT VB20
(K250822) | |
| | external gating device. | external gating device. | external gating device. | |
| myExam Contrast | Bi-directional exchange of patient, scan, and injection parameters for a seamless synchronization of the CT scanner | n.a. | Bi-directional exchange of patient, scan, and injection parameters for a seamless synchronization of the CT scanner | Same as the secondary predicate devices
New feature introduced with syngo CT VB20 of the subject devices
SOMATOM X.cite and X.ceed. |
Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation is completed. Test results show that the subject devices SOMATOM X. Platform with syngo CT VB20 are comparable to the primary predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the primary predicate devices.
# VIII. Performance Data
## Non-Clinical Testing
Non-clinical testing, (integration and functional) including phantom tests were conducted for the SOMATOM X. Platform during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.
The general purpose of each test is to verify and validate the functionality of the subject device modifications.
Testing will cover all related subsystems that contribute to the device modifications. Test levels are defined. For each test level several test activities are performed. The test specification and acceptance criteria are related to the corresponding requirements. Various test activities are performed to specific modifications on different test levels to ensure safe and effective integration in the system. Three test levels are defined:
System Validation test:
Siemens Healthineers AG
{27}
SIEMENS Healthineers
- Acceptance test (workflow and user manual test)
- Legal and Regulatory test
System Verification test:
- System Integration Test (functional)
- Functionality verification
- Image Quality (IQ) Evaluation
Tests are conducted for all software components developed in product development and for the complete product itself. Several activities are considered for this process, including creation of test specifications that relate to software/hardware requirements including tests to address risk mitigations that are identified, documented, and traced by hazard keys.
Bench testing:
Additional evaluation tests are performed as bench tests to support the new device or device modification on Non-Clinical Performance Testing as listed in Table 6 below.
Table 6: Non-clinical performance testing (bench testing).
| Feature/Non-clinical supportive testing | Bench Testing performed |
| --- | --- |
| FAST 3D Camera/ FAST Integrated Workflow | The FAST 3D camera sub-features FAST Isocentering, FAST Range, and FAST Direction have been optimized using additional data from adults and adolescence patients. The bench test evaluates and compares the accuracy of the three sub-features in software version syngo CT VB20 to the accuracy of the predicate devices with syngo CT VB10.
The objectives of the bench tests are to demonstrate that the FAST 3D camera achieves comparable or better results for both, adults and adolescents, as the predicate device for adults.
Overall, the subject devices with syngo CT VB20 delivers comparable or improved accuracy to the predicate devices with syngo CT VB10 predicate device for adults and extends the support to adolescents. |
| FAST Planning | The purpose of the test is to provide a clear reporting on the applied algorithm, its product development, validation, and verification on patient data, which enable the claims.
Objective of the test is to assess the fraction (percentage) of ranges calculated by the FAST Planning algorithm that are correct and can be applied without change. Additionally, calculation time was measured to check whether it meets interactive requirements.
The test results show that the editing actions for the scanner technician can be reduced to a minimum and that the calculation time is fast enough for interactive speed during scanning. For more than 90% of the ranges no editing action was necessary to cover standard ranges. For more than 95%, the speed of the algorithm was sufficient. |
| HD FoV 5.0 | The bench test contains a detailed description and evaluation of the new HD FoV 5.0 algorithm for extended field of view reconstruction. Results |
Siemens Healthineers AG
{28}
SIEMENS Healthineers
| Feature/Non-clinical supportive testing | Bench Testing performed |
| --- | --- |
| | obtained with the new HD FoV 5.0 algorithm are compared with its predecessor, the HD FoV 4.0 algorithm, based on physical and anthropomorphic phantoms.
In addition to the bench test, the performance of the algorithm was evaluated by board-approved radio-oncologists and medical physicists by means of retrospective blinded rater study.
This comparison is conducted to demonstrate that the HD FoV 5.0 algorithm is as safe and effective as the HD FoV 4.0 algorithm. |
| Flex 4D Spiral | The performed bench test report describes the technical background of Flex 4D Spiral and its functionalities with SOMATOM CT scanners, demonstrate the proper function of those, and assess the image quality of Flex 4D Spiral. |
| ZeeFree RT | The bench test evaluates the performance of the ZeeFree RT reconstruction. The objectives of the tests are to demonstrate that compared to the Standard reconstruction, ZeeFree RT
• introduces no relevant errors in terms of CT values and noise levels measured in a homogeneous water phantom, and
• introduces no relevant errors in terms of CT values measured in a phantom with tissue-equivalent inserts, even in the presence of metals and in combination with the iMAR algorithm, and
• introduces no relevant geometrical distortions in a static torso phantom, and
• introduces no relevant deteriorations of the position or shape of a dynamic thorax phantom when moving a spherical shape according to regular, irregular, and patient breathing motion.
In addition, the performance of the algorithm was evaluated by board-approved radio-oncologists by means of a retrospective blinded rater study on 30 patient cases of respiratory 4D CT examinations.
The test results show that the ZeeFree RT reconstruction
• can successfully be applied to 4D respiratory-gated sequence images (Direct i4D), and
• enables the optional reconstruction of stack artefact corrected images, which reduce the strength of misalignment artefacts, if such stack alignment artefacts are identified in non-corrected standard images, and
• does not introduce relevant new artefacts, which were previously not present in the non-corrected standard reconstruction, and |
Siemens Healthineers AG
25 of 32
{29}
SIEMENS Healthineers
A list of recognized and general consensus standards considered for the subject devices is provided as Table 7 and Table 8 below.
Table 7: Recognized Consensus Standards.
| Date of Entry | Recognition Number | Standard Developing Organization | Standard Designation Number and Date | Title of Standard |
| --- | --- | --- | --- | --- |
| 12/18/2023 | 12-352 | NEMA | PS 3.1 - 3.20 2023e | Digital Imaging and Communications in Medicine (DICOM) Set |
| 07/06/2020 | 12-325 | NEMA | XR 25-2019 | Computed Tomography Dose Check |
| 07/06/2020 | 12-330 | NEMA | XR 28-2018 | Supplemental Requirements for User Information and System Function Related to Dose in CT |
Siemens Healthineers AG
{30}
SIEMENS Healthineers
| Date of Entry | Recognition Number | Standard Developing Organization | Standard Designation Number and Date | Title of Standard |
| --- | --- | --- | --- | --- |
| 12/23/2019 | 12-328 | IEC | 61223-3-5 Edition 2.0 2019-09 | Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests and constancy tests - Imaging performance of computed tomography X-ray equipment [Including: Technical Corrigendum 1 (2006)] |
| 03/14/2011 | 12-226 | IEC | 61223-2-6 Second Edition 2006-11 | Evaluation and routine testing in medical imaging departments - Part 2-6: Constancy tests - Imaging performance of computed tomography X-ray equipment |
| 06/07/2021 | 12-336 | IEC | 60601-1-3 Edition 2.2 2021-01 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment |
| 06/27/2016 | 12-302 | IEC | 60601-2-44 Edition 3.2: 2016 | Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography |
| 12/23/2019 | 5-125 | ANSI AAMI ISO | 14971: 2019 | Medical devices - Applications of risk management to medical devices |
| | | ISO | 14971 Third Edition 2019-12 | Medical devices - Application of risk management to medical devices |
| 01/14/2019 | 13-79 | ANSI AAMI IEC | 62304:2006/A1:2016 | Medical device software - Software life cycle |
Siemens Healthineers AG
27 of 32
{31}
SIEMENS Healthineers
| Date of Entry | Recognition Number | Standard Developing Organization | Standard Designation Number and Date | Title of Standard |
| --- | --- | --- | --- | --- |
| | | | | processes [Including Amendment 1 (2016)] |
| | | IEC | 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION | Medical device software - Software life cycle processes |
| 05/30/2022 | 19-46 | ANSI AAMI | ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)] |
| 12/21/2020 | 19-36 | ANSI AAMI IEC | 60601-1-2:2014 [Including AMD 1:2021] | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| | | IEC | 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| 07/06/2020 | 5-129 | ANSI AAMI IEC | 62366-1:2015+AMD1:2020 (Consolidated Text) | Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1 |
| | | IEC | 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION | Medical devices - Part 1: Application of usability engineering to medical devices |
| 07/09/2014 | 12-273 | IEC | 60825-1 Edition 2.0 2007-03 | Safety of laser products - Part 1: Equipment classification, and requirements |
Siemens Healthineers AG
28 of 32
{32}
SIEMENS Healthineers
| Date of Entry | Recognition Number | Standard Developing Organization | Standard Designation Number and Date | Title of Standard |
| --- | --- | --- | --- | --- |
| 12/21/2020 | 5-132 | IEC | 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
| 12/23/2019 | 12-309 | IEC | 60601-2-28 Edition 3.0 2017-06 | Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis |
| 12/20/2021 | 12-341 | IEC | 62563-1 Edition 1.2 2021-07 CONSOLIDATED VERSION | Medical electrical equipment - Medical image display systems - Part 1: Evaluation methods |
In addition to the recognized consensus standards above, the general use standards NOT CURRENTLY RECOGNIZED by FDA, but considered for the subject devices are listed in Table 8.
Table 8: General Use Consensus Standards.
| Standard Developing Organization | Standard Designation Number and Date | Title of Standard | How was Standard Used |
| --- | --- | --- | --- |
| IEC | 60601-1:2005+A1:2012+A2:2020 | Medical electrical equipment - part 1: general requirements for basic safety and essential performance | ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] |
| IEC/ISO | 17050-1 | Conformity Assessment – Supplier's declaration of conformity – Part 1: General requirements | Declaration of conformance to FDA recognized consensus standards. |
| IEC/ISO | 17050-2 | Conformity assessment – Supplier's declaration of conformity – Part 2: Supporting documentation. | General consensus standards not currently recognized by FDA. |
Siemens Healthineers AG
{33}
SIEMENS Healthineers
A list of applicable guidance documents considered for this submission is provided as Table 9 below.
Table 9: FDA Guidance Document and Effective Date
| FDA Guidance Document | Issue date |
| --- | --- |
| User Fees and Refunds for Premarket Notification Submissions (510(k)s): Guidance for Industry and Food and Drug Administration Staff | 10/05/2022 |
| Refuse to Accept Policy for 510(k)s: Guidance for Industry and Food and Drug Administration Staff | 04/21/2022 |
| Electronic Submission Template for Medical Device 510(k) Submissions | 10/2/2023 |
| Deciding When to Submit a 510(k) for a Change to an Existing Device | 10/25/2017 |
| The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] | 07/28/2014 |
| Content of Premarket Submissions for Software Contained in Medical Devices | 06/14/2023 |
| Off-The-Shelf Software Use in Medical Devices | 08/11/2023 |
| Applying Human Factors and Usability Engineering to Medical Devices | 02/03/2016 |
| Pediatric Information for X-ray Imaging Device Premarket Notifications | 11/28/2017 |
| Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions | 06/27/2025 |
| Electromagnetic Compatibility (EMC) of Medical Devices | 06/06/2022 |
| Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices | 09/06/2017 |
| Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices | 09/14/2018 |
## Verification and Validation
The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claims of substantial equivalence.
## Cybersecurity
Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.
## Wireless Coexistence Testing
Siemens Healthineers AG
{34}
SIEMENS Healthineers
Additionally, Siemens conforms to the requirements for Radio Frequency Wireless Technology as defined in FDA guidance document “Radio Frequency Wireless Technology in Medical Devices, Guidance for Industry and Food and Drug Administration Staff” issued on August 14, 2013 by adhering to the EMC and risk based verification and validation requirements in design, testing, and labeling of the wireless remote control components of the subject devices.
The Radio Frequency Wireless Technology of the optional Remote Scan Control and supporting Control Device iPad for Scan&GO complies to 47 CFR part 15 subpart c – Intentional Radiators. All Radio device labels will show an FCC ID code to show compliance. Shielding requirement applicable to the SOMATOM X. Platform and respective Scatter Radiation diagrams for typical room installations are provided in the User Documentation and Planning Guide of the intended Scanners in accordance with IEC60601-2-44.
Siemens has considered several measures to address wireless coexistence by design to ensure the safe operation of the wireless components in combination with the applicable system supported functionality. Wireless technology in the system setup to perform a task in a given shared environment where other systems have an ability to perform their tasks and may or may not be using the same set of rules has been considered. According to FDA guidance “Radio Frequency Wireless Technology in Medical Devices” Siemens has addressed the safety, effectiveness, and high likelihood of coexistence with other devices of this technology in our product design by our Risk Management Process, Failure Mode and Effects Analysis (FMEA) Process, and Requirement Engineering Process. As part of the risk management process, hazardous situations associated with the Scan&GO and its connection to the host system via Wi-Fi were addressed as part of the Risk Management process.
Testing for co-existence considered for following scenarios:
- Co-Channel Testing
- Adjacent Channel Testing
- RF Interference Testing
- Separation Distance/Location Testing
Scan&GO is designed to allow dynamic frequency selection and transmission power control by default in accordance with IEEE 802.11h. Adjacent channel testing is addressed by the fact that Scan&GO does not support shared medium access to Siemens Wi-Fi network. RF interference was tested by successfully ensuring that wireless communications were actively transmitting in situations where possible interference may exist. Recommended distance and router locations requirements are documented in the user documentation.
## Summary
The features described in this premarket notification are supported with verification and validation testing, dosimetry and imaging performance, and analysis of phantom images to assess device and feature performance during product development. The risk analysis was completed, and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria. Verification and validation testing of the device was found acceptable to support the claim of substantial equivalence.
## General Safety and Effectiveness Concerns
Siemens Healthineers AG
{35}
SIEMENS Healthineers
The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a system related risk analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing according to the risk management process. In order to minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
IX. Conclusions
Verification and validation and phantom testing were performed. The non-clinical data supports the safety of the device and the hardware and software verification and validation demonstrates that the subject devices SOMATOM X.cite and SOMATOM X.ceed with software version syngo CT VB20 perform as intended in the specified use conditions. The data included in this submission demonstrates that the SOMATOM X.cite and SOMATOM X.ceed with software version syngo CT VB20 with the described modifications performs comparably to the predicate devices currently marketed for the same intended use. The conclusion drawn from the non-clinical tests demonstrates that the subject devices are as safe, as effective, and perform as well as or better than the predicate devices.
Siemens Healthineers AG
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.