SOMATOM DEFINITION

K122471 · Siemens Medical Systems, Inc. · JAK · Sep 11, 2012 · Radiology

Device Facts

Record IDK122471
Device NameSOMATOM DEFINITION
ApplicantSiemens Medical Systems, Inc.
Product CodeJAK · Radiology
Decision DateSep 11, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1750
Device ClassClass 2

Indications for Use

The SOMATOM P45 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

Device Story

Whole body X-ray Computed Tomography (CT) system; utilizes two continuously rotating tube-detector systems; operates on fan beam principle. Inputs: X-ray transmission data. Processing: SOMARIS/7 VA44 software performs patient/data management, scan control, image reconstruction, and archiving. Features CARE (Combined Application to Reduce Exposure) and FAST (Fully Assisted Scanner Technology) for scan optimization; supports single-click 3D reconstruction of Dual Energy Scans. Outputs: DICOM format CT images. Used in clinical settings by trained medical professionals. Output used by clinicians for diagnostic evaluation. Benefits: Provides cross-sectional anatomical visualization for clinical decision-making.

Clinical Evidence

Bench testing only. Verification and validation testing, including phantom testing, were conducted to support software modifications. Risk analysis and hazard mitigation were performed. Results confirmed software specifications met acceptance criteria.

Technological Characteristics

Whole body CT scanner; dual-source (two rotating tube-detector systems); fan beam geometry. Software: SOMARIS/7 VA44 running on Windows 7. Features: CARE (dose reduction), FAST (automated scanning), and 3D reconstruction. Connectivity: DICOM compliant. Energy source: X-ray.

Indications for Use

Indicated for patients requiring cross-sectional imaging of the body via X-ray computed tomography. No specific age or gender restrictions stated.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ SIEMENS Special 510(k) Submission: SOMATOM Definition CT System SEP 1 1 2012 510(K) SUMMARY FOR SOMATOM P45 Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Date Prepared: August 10, 2012 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. #### 1. Contact Person: Mrs. Kimberly Manqum Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions, Inc. USA 51 Vallev Stream Parkwav D02 Malvern, PA 19355-1406 Phone: (610) 448-1772 Fax: (610) 448-1778 Email: kimberly.mangum@siemens.com #### 2. Device Name and Classification SOMATOM P45 Product Name: Propriety Trade Name: SOMATOM Definition Classification Name: Computed Tomography X-ray System Classification Panel: Radiology 21 CFR §892.1750 CFR Section: Device Class: Class II Product Code: 90JAK 20 {2}------------------------------------------------ ## 3. Substantial Equivalence: SOMATOM P45 Computed Tomography Siemens X-ray systems. configured with software version SOMARIS/7 VA44 is substantially equivalent to the following medical devices in commercial distribution: | Predicate Device Name | FDA Clearance<br>Number | FDA Clearance Date | |------------------------------------|-------------------------|--------------------| | Siemens SOMATOM P45 | K052216 | 09/08/2005 | | Siemens SOMATOM Definition<br>Edge | K120579 | 05/23/2012 | ### 4. Device Description: . The Siemens SOMATOM P45 is a whole body X-ray Computed Tomography System, which features two continuously rotating tube-detector systems and functions according to the fan beam principle. The SOMATOM P45 produces CT images in DICOM format, which can be used by postprocessing applications commercially distributed by Siemens and other vendors. The system software is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation. The new version of system software, SOMARIS/7 VA44, supports a Windows 7 operating system, additional scanning and evaluation techniques CARE (Combined Application to Reduce Exposure) and FAST (Fully Assisted Scanner Technology), and single click 3D reconstruction of Dual Energy Scans. The computer system delivered with the CT scanner is able to run the post processing applications optionally. ### 5. Summary of Technical Characteristics of the Subject Device as Compared with the Predicate Device: SOMATOM P45 configured with software version SOMARIS/7 VA44 does not have significant changes in materials, energy source, or technological characteristics when compared to the predicate devices. The intended use and fundamental scientific technology are similar to the predicate devices; {3}------------------------------------------------ SIEMENS therefore Siemens believes that they are substantially equivalent to the predicate devices. #### 6. Nonclinical Testing: The modifications described in this premarket modification were supported with verification/validation as well as phantom testing. Non clinical tests are conducted for SOMATOM P45 configured with software version SOMARIS/7 VA 44 during product development. The risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence. # 7. Indications for Use: The SOMATOM P45 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) ### 8. General Safety and Effectiveness Concerns: The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards. {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of an eagle or bird-like figure, with three curved lines forming the body and wings. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Ms. Kimberly Mangum Technical Specialist, Regulatory Affairs Siemens Medical Systems, Inc. 51 Valley Stream Parkway MALVERN PA 19355 Re: K122471 Trade/Device Name: SOMATOM P45 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: August 10, 2012 Received: August 14, 2012 Dear Ms. Mangum: We have reviewed your Section 510(k) premarket notification of intent to market the device We nave reviewed your becaon of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the citerosule) to regarry manager date of the Medical Device American Comments, or to commerce prior to May 20, 1970, the chaoinen with the provisions of the Federal Food, Drug, devices may nave occh recuired in assess approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require apple of the general controls provisions of the Act. The r ou may, dicierore, market the act include requirements for annual registration, listing of general controls provisions of the tice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such If your device is classified (see above) into exass ff (Specifice (an be found in Title 21, additional controls. Dations (CFR), Parts 800 to 895. In addition, FDA may publish further Out of Peachar Regarning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please or advised that I DA s issuation of a state complies with other requirements of the Act that FDA has made a decemmanon that your contribution we rederal agencies. You must or any Federal statutes and regulations adminities, but not limited to: registration and listing (21) comply with all the Act's requirements, including, but not limited to: r Comply with an the Act s requirements, and 809); medical device reporting (reporting of SEP 11 2012 {5}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice mouitements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirements as bet form in are qu your device as described in your Section 510(k) premarket while a to began mading of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you dosire specific ad reserved In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket 5 150. Finds, prease note are of 97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the r ou may other other general and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 510(k) Number (if known): Device Name: SOMATOM P45 Indications for Use: The SOMATOM P45 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulting from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) Use Prescription - AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In vitro Diagnostic Device (OIVD) Andrew D. Oja of In Vitro Diagnostic Device
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...