Capenergy C Equipments RF C200 and C400 Drakarian

K253480 · Capenergy Medical S.L. · PBX · Jul 10, 2026 · General, Plastic Surgery

Device Facts

Record IDK253480
Device NameCapenergy C Equipments RF C200 and C400 Drakarian
ApplicantCapenergy Medical S.L.
Product CodePBX · General, Plastic Surgery
Decision DateJul 10, 2026
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Capenergy C Equipments RF C200 and C400 Drakarian are intended to provide topical heating for the purpose of elevating tissue temperature for dermatological procedures. The massage device is intended to provide a temporary reduction in the appearance of Cellulite. They are indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. The Capenergy C Equipments RF C200 and C400 Drakarian are indicated for using for non-invasive reduction of abdominal circumference associated with subcutaneous fat induced by lipolytic processes.

Device Story

System generates radiofrequency (RF) energy for tissue heating; utilizes AC/DC power supply, RF generator, controller, and user interface. Handpiece with internal/external electrodes connects to console. Device monitors tissue temperature and impedance via thermocouple feedback loop; automatically adjusts energy delivery to maintain therapeutic range (40-45°C). Used in clinical settings by professionals for dermatological/surgical procedures and non-invasive fat reduction. Output includes controlled RF energy delivery; provider monitors via screen/buttons. Benefits include localized tissue heating for hemostasis, coagulation, and aesthetic fat reduction.

Clinical Evidence

Single-center, prospective, paired case-control study. Evaluated efficacy in reducing abdominal circumference and subcutaneous fat over 12 weeks. Primary endpoints: statistically significant reduction in abdominal circumference and subcutaneous fat at 12 weeks vs. baseline. Secondary endpoints: reduction at intermediate visits, patient satisfaction, and discomfort (NSR). Results showed improvement; adverse events were consistent with known profiles.

Technological Characteristics

RF generator (0.448-1.2 MHz); bipolar/monopolar modes. Materials: biocompatible per ISO 10993-1. Feedback: thermocouple-based temperature/impedance monitoring. Connectivity: standalone console with screen/buttons. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 62304. Sterilization: reusable electrodes. Software: firmware-controlled.

Indications for Use

Indicated for dermatological procedures requiring topical heating, temporary reduction of cellulite appearance, electrocoagulation and hemostasis in general surgery, and non-invasive reduction of abdominal circumference associated with subcutaneous fat via lipolytic processes.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** U.S. FOOD & DRUG ADMINISTRATION July 10, 2026 Capenergy Medical S.L. Pilar Sanchez General Manager Av. Mare De Deu De Montserrat, 41 Bis Pje 1º Derecha Sant Joan Despí, Barcelona 08970 Spain Re: K253480 Trade/Device Name: Capenergy C Equipments RF C200 and C400 Drakarian Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX, GEI Dated: May 11, 2026 Received: May 27, 2026 Dear Pilar Sanchez: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253480 - Pilar Sanchez Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K253480 - Pilar Sanchez Page 3 Sincerely, JAMES H. JAMES H. JANG -S Digitally signed by JAMES H. JANG -S Date: 2026.07.10 14:46:10 -04'00' For Colin Kejing Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K253480 Device Name Capenergy C Equipments RF C200 and C400 Drakarian Indications for Use (Describe) The Capenergy C Equipments RF C200 and C400 Drakarian are intended to provide topical heating for the purpose of elevating tissue temperature for dermatological procedures. The massage device is intended to provide a temporary reduction in the appearance of Cellulite. They are indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. The Capenergy C Equipments RF C200 and C400 Drakarian are indicated for using for non-invasive reduction of abdominal circumference associated with subcutaneous fat induced by lipolytic processes. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} K253480 Page 1 of 9 SUBMISSION DATE: 07/10/2025 SENDER NAME: Capenergy Medical SL PRESENTER'S ADDRESS: Av. Mare de Déu de Montserrat, 41 bis Pje 1º derecha 08970 Sant Joan Despí Barcelona Spain CONTACT: Pilar Sánchez, General Director TELEPHONE: +34 93 477 43 48 Email: pilar@capenergy.com COMMERCIAL NAME OF THE DEVICE: Capenergy C Equipments RF C200 and C400 Drakarian COMMON NAME: Massager, Vacuum, Radiofrequency-induced heat DESCRIPTION OF THE REGULATIONS: General and Plastic Surgery CLASS: Class II REGULATION NUMBER: 21 CFR 878.4400 PRODUCT CODE: PBX, GEI # PREDICATE DEVICES Primary predicate: K192224 BTL-899 from BTL Industries, Inc. Secondary Predicate(s): K163415 Syneron Medical Ltd. SlimShape System K222260 Capenergy C Equipment RF System C25, C50, C100, C200, C300, C400, C500 by Capenergy Medical SL # DEVICE DESCRIPTION: The Capenergy C Drakarian RF C200 and C400 system, including its accessories, generates radiofrequency energy with temperature and impedance feedback for procedures requiring tissue heating. The Capenergy system constantly monitors the temperature and {5} K253480 Page 2 of 9 impedance of the tissue being treated, automatically adjusting energy delivery to maintain effective and safe tissue heating. The Capenergy C Drakarian RF C200 and C400 systems consist of an AC/DC power supply, an RF generator, a controller, and a user interface. The RF handpiece connects to the console via a cable, and a switch activates power management. The handpiece comprises internal and external electrodes. #### **INTENDED USE / INDICATIONS FOR USE:** The Capenergy C Equipments RF C200 and C400 Drakarian are intended to provide topical heating for the purpose of elevating tissue temperature for dermatological procedures. The massage device is intended to provide a temporary reduction in the appearance of Cellulite. They are indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. The Capenergy C Equipments RF C200 and C400 Drakarian are indicated for using for non-invasive reduction of abdominal circumference associated with subcutaneous fat induced by lipolytic processes. #### **SUMMARY OF COMPARISON WITH THE PREDICATE DEVICE:** To establish substantial equivalence, Capenergy C Equipments' RF systems are compared to the following previously approved devices: ##### **Primary predicate:** K192224 BTL-899 from BTL Industries, Inc. ##### **Secondary Predicate:** K163415 SlimShape System from Syneron Medical Ltd. K222260 Capenergy C Equipment RF System C25, C50, C100, C200, C300, C400, C500 by Capenergy Medical SL The comparison of the subject device with the primary predicate device is summarized in the following table: {6} K253480 Page 3 of 9 | | Subject Device | Primary Predicate Device | First predicate of reference Device | Second predicate of reference Device | | --- | --- | --- | --- | --- | | **Device name** | Capenergy C RF C200 and C400 Drakarian equipment, includes Capenergy Non-Invasive Monopolar RF Reusable Electrode Set and Automatic Bipolar Belt | BTL-899 | SlimShape System | The Capenergy C RF System devices - C25, C50, C100, C200, C300, C400, C500 include a set of Capenergy reusable non-invasive RF monopolar electrodes and automatic plates | | **Company Name** | Capenergy Medical SL | BTL Industries, Inc. | Syneron Medical Ltd. | Capenergy Medical SL | | Number 510(k) | Not Assigned | K192224 | K163415 | K222260 | | Regulation number | 21 CFR 878.4400 | 21 CFR 878.4400 | 21 CFR 878.4400 | 21 CFR 878.4400 | | Product Code, Class | PBX, GEI, Class II | GEI, Class II | GEI, Class II | PBX, GEI, Class II | | Instructions for use | The Capenergy C Equipments RF C200 and C400 Drakarian are intended to provide topical heating for the purpose of elevating tissue temperature for dermatological procedures. The massage device is intended to provide a temporary reduction in the appearance of Cellulite. They are indicated for use | BTL-899 is indicated for use in: • Non-invasive lipolysis (fat breakdown) of the abdomen. • Reduction of abdominal circumference - BTL-899 is designed for use with skin types I to III. | The SlimShape system is indicated for non-invasive lipolysis (fat breakdown) of the abdomen. The device is indicated for reducing abdominal circumference. | The Capenergy C RF System devices -C25, C50, C100, C200, C300, C400, C500- are designed to provide topical heating in order to raise tissue temperature for the treatment of selected medical conditions, such as: pain relief, muscle spasms, increased local circulation, and dermatological procedures. | {7} K253480 Page 4 of 9 | | in dermatological and general surgical procedures for electrocoagulation and hemostasis. The Capenergy C Equipments RF C200 and C400 Drakarian are indicated for using for non-invasive reduction of abdominal circumference associated with subcutaneous fat induced by lipolytic processes. | | | The massage device is designed to provide a temporary reduction in the appearance of cellulite. The Capenergy C RF System equipment -C25, C50, C100, C200, C300, C400, C500- are indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. | | --- | --- | --- | --- | --- | | Device Description | The Capenergy RF generator produces an oscillating electric field in the antenna (electrode). This field is transmitted to the surrounding soft tissue, causing it to heat up. A thermocouple in the electrode measures this temperature increase and maintains a feedback loop to ensure a reference temperature in the tissue. The system uses bipolar radio frequency. | The system combines bipolar radiofrequency with electromagnetic stimulation. | The system combines bipolar radiofrequency with mechanical suction. | The Capenergy RF generator produces an oscillating electric field in the antenna (electrode). This field is transmitted to the surrounding soft tissue, causing it to heat up. A thermocouple in the electrode measures this temperature increase and maintains a feedback loop to ensure a reference temperature in the tissue. The system uses monopolar radio frequency. | {8} K253480 Page 5 of 9 | Clinical use | Prescription only | Prescription only | Prescription only | Prescription only | | --- | --- | --- | --- | --- | | RF Power | Max. 180 W (45 W +/-10 per channel) | Max. 60 W (30 W per channel) | 80 W (1 channel) | Max. 180 W (45 W +/-10 per channel) | | Internal Cut | 40-45°C | 40 –43°C | 40 –43°C | 40-45°C | | Treatment time | Up to 30 min | Up to 30 min | Up to 30 min | Up to 30 min | | RF Frequency | Fixed values are set for treatment time, output power percentage, and operating frequency . Three options are available: 0.8 MHz +/-25%, 1.0 MHz +/-25%, and 1.2 MHz +/-25%. | 27.12 MHz | 1 MHz | Fixed values are set for treatment time, output power percentage, and operating frequency . Four options are available: 0.448 MHz +/-25%, 0.8 MHz +/-25%, 1.0 MHz +/-25%, and 1.2 MHz +/-25%. | | RF type | Bipolar | Bipolar | Bipolar | Monopolar | | Waveform | Sinusoidal | Sinusoidal | Sinusoidal | Sinusoidal | | Safety class Protection | Class I – Type BF | Class II – Type BF | Class I – Type BF | Class I – Type BF | | Use the interface | Screen and buttons | Touch screen | Touch screen | Screen and buttons | | Supply voltage and frequency | 100–120/200–240 V ± 10%, 50/60 Hz | 100 – 240 V AC, 50–60 Hz | 100 – 230 V AC, 50–60 Hz | 100–120/200–240 V ± 10%, 50/60 Hz | | Firmware controlled | Yes | Yes | Yes | Yes | | Temperature sensor and control | Yes | Yes | Yes | Yes | {9} K253480 Page 6 of 9 | Parameter selection (Intensity, Time) | Yes | Yes | Yes | Yes | | --- | --- | --- | --- | --- | | Application | Automatic, manual, and hands-free applicators Automatic hands-free applicator, secured by a fastening belt. | Hands-free applicator, secured by a fastening belt. | Hands-free applicator, secured by a fastening belt. | Automatic, manual, and hands-free applicators | | Impedance fabric | 50-300 ohms | 50-300 ohms | 50-300 ohms | 50-300 ohms | | Dimensions (Width x Height x Depth) | 170 x 220 mmx250 mm 345x 220 mmx420 mm 562 mmx 220 mmx420 mm | 23 x 39 x29 in (592 x 985 x 730 mm) | 19.6 x 40.7 x20.7 in | 170 x 220 mmx 250 mm 345 x 220 mmx220 mm 562 mmx 420 mmx220 mm | | Weight | 13 -26.5 Kg | 85 Kg | 53 Kg | 13 -26.5 Kg | | Compliance with voluntary standards / laboratory tests performed | IEC 60601-1:2005 + /A1:2012 IEC 60601-1-2:2015 IEC 60601-1-6:2010 IEC 60601-2-2:2017 IEC 62304:2006 ISO10993-1:2009 ISO10993-5 ISO10993-10 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 60601-1-2 ISO10993-1:2009 ISO10993-5 ISO10993-10 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 60601-1-2 ISO10993-1:2009 ISO10993-5 ISO10993-10 | IEC 60601-1:2005 + /A1:2012 I CE 60601-1-2:2015 IEC 60601-1-6:2010 IEC 60601-2-2:2017 IEC 62304:2006 ISO10993-1:2009 ISO10993-5 ISO10993-10 | | Environmental conditions | Temperature: 10° to 40°C(+/- 20C) Relative humidity: less than 80%. For indoor use only Unsafe RM | Temperature: 10° to 30°C Relative humidity: 30% to 75%. For indoor use only Unsafe RM | Temperature: 10° to 30°C Relative humidity: 30% to 80%. For indoor use only Unsafe RM | Temperature: 10° to 40°C(+/- 20C) Relative humidity: less than 80%. For indoor use only Unsafe RM | {10} K253480 Page 7 of 9 # **SUMMARY DISCUSSION OF NONCLINICAL TESTS (21 CFR 807.92(b)(1))** Preclinical functional tests were performed to evaluate the safety and performance of the Capenergy C RF C200 and C400 Drakarian devices in relation to their intended use and to address potential risks associated with radiofrequency tissue heating. *and* The patient contact materials were tested. Testing included verifying the device's ability to reach and maintain therapeutic tissue temperatures (40-45 °C) during treatments of up to 30 minutes under foreseeable operating conditions, as well as its efficacy in treatments. These tests were performed using the C200 device in 2019, which is equivalent to the C200 Drakarian, as it uses the same active and passive plates, the same as those that make up the CBelt belt, and the same temperature control, common to the entire Capenergy (K222260) device family. Additional non-clinical testing addressed electrical safety and electromagnetic compatibility in accordance with IEC 60601-1 and IEC 60601-1-2 and confirmed that the device meets applicable safety and EMC requirements. The software verification and validation were performed in accordance with IEC 62304 and the FDA software validation guide, demonstrating that the software functions reliably and as intended. Materials in contact with the patient were evaluated in accordance with ISO 10993-1, including applicable biocompatibility endpoints, and were determined to be biocompatible for the intended duration and type of contact. The results of these non-clinical tests demonstrate that the device in question works as intended and is as safe and effective as the identified precursor devices. Although the device in question, Drakarian, has a higher maximum output power than the original device, the integrated temperature monitoring and automatic power adjustment ensure that tissue temperatures remain within the same therapeutic range as the original devices. In addition to electrical safety testing, the software was verified and validated according to IEC 62304:2006 (Software for Medical Devices), Software Lifecycle Processes, and FDA guidelines on software validation. The results of these tests conclude that the software meets these requirements. The materials in contact with the patient have been evaluated in accordance with the requirements of ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing, and were confirmed to be biocompatible for the intended use. {11} K253480 Page 8 of 9 The results of these non-clinical tests demonstrate that the device in question works as intended and is as safe and effective as the original devices. # **SUMMARY DISCUSSION OF CLINICAL TRIALS (21 CFR 807.92(b)(2))** The conducted clinical study was a single-center, prospective cohort, a paired case-control study where each individual served as their own control. Objectives: Primary Objectives - Demonstrate the effectiveness in reducing abdominal circumference and subcutaneous fat in 12 weeks after treatments with the Capenergy bipolar radiofrequency belt. - Demonstrate the safety of treatment with the Capenergy bipolar radiofrequency belt. Secondary Objectives - Evaluate the efficacy in reducing abdominal circumference at all visits (starting from visit 2). - Evaluate the effectiveness in reducing abdominal fat at all visits (starting from visit 3). - Evaluate the level of comfort during treatments, at all visits (starting with visit 2), reported by patients. - Evaluate the satisfaction reported by the subjects, and the improvement reported by the researcher Endpoints: Primary Efficacy Endpoints - Statistically significant reduction in abdominal circumference after Drakarian device treatments at 12 weeks of follow-up versus baseline. - Statistically significant reduction in subcutaneous fat after Drakarian equipment treatments at 12 weeks of follow-up versus baseline. Secondary Efficacy Endpoints - Statistically significant reduction in abdominal circumference after treatment with Drakarian equipment at each visit (beginning with the second treatment) compared to baseline, as measured by tape measures and a caliper. {12} K253480 Page 9 of 9 - Statistically significant reduction in abdominal fat after treatment with Drakarian equipment at each visit (beginning with the third treatment) compared to baseline, using ultrasound imaging measurements. - Patient self-assessment for satisfaction and improvement using questionnaires at each visit (starting with the second treatment). - Patient discomfort (pain) level after each treatment (Tx.1, Tx.2 and Tx.3), using a 10-point visual analog scale (Numeric Scale Response (NSR)). # Primary Safety End Point - The number, severity, and type of any adverse events recorded throughout the study. - Immediate response in each treatment (Tx.1, Tx.2 & Tx.3). The results showed improvement and the AEs are consistent with known events. # SUMMARY OF THE DISCUSSION ON SUBSTANTIAL EQUIVALENCE. Based on technological comparison to the predicate devices and performance testing, the Capenergy C RF C200 and C400 Drakarian devices with their Capenergy reusable non-invasive RF monopolar electrode set and their CBelt bipolar belt , are substantially equivalent in terms of safety and effectiveness for the requested indications for use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...