K163415 · Syneron Medical, Ltd. · GEI · Apr 20, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163415
Device Name
SlimShape System
Applicant
Syneron Medical, Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SlimShape System is indicated for non-invasive lipolysis (breakdown of fat) of the abdomen. The device is indicated for reduction in circumference of the abdomen.
Device Story
SlimShape System delivers 1 MHz bipolar sinusoidal radiofrequency (RF) energy and mechanical vacuum to abdominal fat tissue; designed for hands-free treatment based on predefined user protocols. System includes console (RF generator, vacuum pump, computer, touch-screen interface) and applicator with integrated temperature sensors for safety control. Used in clinical settings by trained operators. Output consists of thermal and mechanical energy application to subcutaneous tissue; healthcare providers monitor treatment via touch-screen. Device induces fat necrosis in subcutaneous tissue while sparing dermis/epidermis; clinical benefit includes reduction in abdominal circumference and fat thickness. System is non-invasive; intended to provide aesthetic body contouring.
Clinical Evidence
Prospective, single-arm, self-controlled study (n=52) evaluated safety and efficacy. Primary endpoints: abdominal circumference and fat thickness reduction at 12 weeks. Results: mean circumference reduction 1.98±2.53 cm; mean fat thickness reduction 6.03±3.09 mm (p<0.05). Secondary: 86% correct photo identification by blinded reviewers; 64% of subjects achieved ≥1.5 cm reduction. Safety: 8 mild adverse events in 144 sessions; no serious adverse events. Histology study (n=11) confirmed discrete fat necrosis in subcutaneous tissue with intact dermis/epidermis.
Technological Characteristics
Console-based system with applicator; 1 MHz bipolar sinusoidal RF energy; mechanical vacuum; integrated temperature sensors. Biocompatibility per ISO 10993-5/10. Electrical safety per IEC 60601-1, IEC 60601-2-2; EMC per IEC 60601-1-2. Software-controlled user interface.
Indications for Use
Indicated for non-invasive lipolysis and reduction in abdominal circumference in healthy male and female adults (18-65 years).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle with three heads. The eagle is facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 20, 2017
Syneron Medical Ltd. % Ms. Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K163415
Trade/Device Name: SlimShape System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 30, 2017 Received: March 30, 2017
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K163415
Device Name
SlimShape System
Indications for Use (Describe)
The SlimShape System is indicated for non-invasive lipolysis (breakdown of fat) of the abdomen. The device is indicated for reduction in circumference of the abdomen.
Type of Use (*Select one or both, as applicable*)
| <span style="font-family: sans-serif;"> </span> <span style="text-decoration: overline;"><b> </b></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
{3}------------------------------------------------
## 510(k) SUMMARY
## Syneron Medical Ltd.'s SlimShape System
# Submitted by:
Syneron Medical Ltd. P.O.B. 550 Industrial Zone Tavor Building Yokneam Illit, 20692 Israel
# Contact Person:
Ruthie Amir, MD, Global Vice President of Clinical, Regulatory and Professional Education Syneron Medical Ltd. P.O.B. 550 Industrial Zone Tavor Building Yokneam Illit, 20692 Israel Phone: 972-73-244-2200 Facsimile: 972-73-244-2202
# Submission Correspondent:
Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103 Phone: 267 675 4600 Facsimile: 267 675 4601
Date Prepared: April 19, 2017
#### Name of Device
Syneron SlimShape System
## Common or Usual Name
Electrosurgical, Cutting & Coagulation & Accessories
# Classification
Electrosurgical, Cutting & Coagulation & Accessories
21 CFR 878.4400, Class II, product code GEI
## Predicate Device
Syneron Medical Ltd.'s Transcend System (K120510)
#### Intended Use / Indications for Use
The SlimShape System is indicated for non-invasive lipolysis (breakdown of fat) of the abdomen. The device is indicated for reduction in circumference of the abdomen.
{4}------------------------------------------------
# Device Description
The SlimShape System is comprised of a console and an applicator that delivers radiofrequency energy to the fat tissue and also applies mechanical vacuum to the treatment area. The SlimShape System is designed to enable hands-free abdominal treatment that is based on a predefined user protocol. The system also incorporates a skin temperature control mechanism as a safety feature during treatment. Users interface with and control the SlimShape System via the touch-screen panel.
# Technological Characteristics
The SlimShape System has very similar technological characteristics compared to its predicate. The SlimShape and the predicate are both comprised of a system console and applicator(s). The console in both devices consists of the RF generator, vacuum pump, computer, and the touch-screen control panel. Although the SlimShape System does not include infrared light, this difference compared to the predicate does not alter the intended therapeutic effect of the device or raise new types of safety or effectiveness questions, as demonstrated in clinical and nonclinical performance testing. Further, technological differences in the shape and dimensions of the applicators of the SlimShape and predicate device also do not raise any different safety or effectiveness questions, given that the device applicators have the same or similar energy parameters with similar treatment areas. In addition, the applicators of both systems similarly apply the vacuum and RF energy to the treated area. Both devices use the 1 MHz bipolar sinusoidal RF signal. In addition, the power intensity level for SlimShape is within the maximum output power cleared for the predicate device. Both devices include a temperature control mechanism through temperature sensors in the applicator. Notably, the SlimShape System has been evaluated in clinical and bench testing, and the technological differences do not raise new types of safety or effectiveness questions compared to the predicate.
Therefore, the SlimShape System has similar technological characteristics compared to its predicate device.
# Performance Data
Nonclinical Performance Testing: The following nonclinical performance testing was conducted to support the substantial equivalence of the SlimShape System to the predicate device. In all instances, the SlimShape System functioned as intended.
- Biocompatibility of the patient-contacting components of the device was established. Cytotoxicity testing per ISO 10993-5, sensitization testing per ISO 10993-10, and irritation testing per ISO 10993-10 were performed to further confirm the biocompatibility of the device.
- Software verification and validation was performed, and demonstrated that the software performs as intended.
- . Electrical safety (IEC 60601-1, IEC 60601-2-2), electromagnetic compatibility (IEC 60601-1-2) and electromagnetic immunity testing was conducted and results were passing.
{5}------------------------------------------------
Clinical Data: In addition, clinical evaluation of the intended population was performed in several separate prospective studies, including a single-arm, prospective, selfcontrolled study of the SlimShape device for non-invasive lipolysis and circumference reduction of the abdomen. The study enrolled 52 subjects who were treated with the SlimShape device. The study included Caucasian females (87%) and males (13%) across a range of ages and skin types. Mean age was 45 years. The baseline mean weight was 72±11 kg and mean BMI was 26.29±2.62 kg/m².
Forty subjects underwent 3 biweekly treatments using the Temperature Control mode on the abdomen. Subject follow up was conducted at 4 weeks, and 12 weeks after the last treatment. The study included circumference and fat thickness measurements of the post treatment abdomen compared to baseline measurements, as well as visual assessment of the before and after photographs by 2 blinded reviewers.
The study met its primary efficacy endpoint, with an average of 1.98±2.53 cm reduction in abdominal circumference and an average of 6.03±3.09 mm reduction (-24.5%±11.3%) in fat thickness measured by ultrasound at 12 weeks follow-up compared to baseline.
Secondary effectiveness endpoints demonstrated that pre- and post-treatment photographs of the subjects were correctly identified in 86% of the photographs by blinded reviewers. Additional analysis demonstrated reductions in abdomen circumference and fat thickness over time and reductions in circumference at upper and lower abdomen. Responder analysis showed that 64% of the subjects at 12 week follow up achieved at least 1.5 cm of circumference reduction. In addition, 59% of subjects were very satisfied or satisfied with the treatments at 12 week follow up, and 77% of subjects would recommend the procedure. Furthermore, 91% of the subjects reported none to moderate pain.
Out of a total of 144 treatment sessions, 6 subjects reported on 8 mild adverse events. There were no reports of serious or unanticipated adverse events. No subjects discontinued from the study due to adverse events. The anticipated immediate responses after treatment (e.q., erythema, edema) were consistent with those observed with other similar devices. The immediate responses were transient and resolved completely within several hours without any intervention.
| Study Design | Prospective, single-arm, self-controlled, single-site clinical study |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sample size | 52 subjects at one site were enrolled and treated in the study |
| Principal<br>Eligibility<br>Criteria | • Healthy male or female, ≥ 18 and ≤ 65 years of age at the time of<br>enrollment.<br>• Subject agrees to maintain their weight by not making any major<br>changes in their diet or lifestyle during the course of the study.<br>• General good health confirmed by medical history and skin examination<br>of the treated area. |
| Follow up<br>intervals | 3 treatment visits and follow up visits at 4 weeks, 8 weeks, and 12 weeks. |
The study design and results are further summarized in the table below.
{6}------------------------------------------------
| Endpoints | Primary: Statistically significant abdominal circumference and fat reduction post<br>SlimShape treatment at the final follow-up compared to baseline, based on tape measurements and ultrasound imaging.<br> |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Secondary:<br>- Correct identification of the pre- and post-treatment photos as<br>demonstrated in at least 70% or greater of treated subjects.<br>- Abdominal circumference reduction at each visit, based on tape measurements.<br>- Abdominal fat reduction at each visit, measured by Ultrasound.<br>- Subject's self-assessment for satisfaction and recommendation.<br>- Subject satisfaction self-assessment.<br>- Subject's discomfort (pain) level after each treatment. |
| Effectiveness<br>Results | Primary: Met endpoint; Both the midline reduction in abdominal circumference and the reduction in abdominal fat layer were statistically significant at the final follow-up compared to baseline. |
| | Secondary:<br>- The blinded reviewers assessed correctly 86% of the photographs.<br>- The amount of circumference reduction was statistically significant compared to baseline at each visit, as well as at the upper and lower abdomen.<br>- The majority of the subjects were satisfied (53% at treatment 3, 62% at 4 week follow up, 59% at 8 week follow up and 59% at 12 week follow up).<br>- Most of the subjects reported they will recommend this procedure (75%, 76%, 76% and 77% reported they recommended at treatment 3, 4, 8 and 12 weeks follow up, respectively).<br>- Most subjects (91%) reported none to moderate pain for each treatment. |
| Safety<br>Results | There were no reports of serious or unanticipated adverse events, with 8 mild adverse events after 144 treatments. No subjects discontinued from the study due to adverse events. The immediate responses included erythema and edema, which resolved completely within several hours without any intervention. |
Histology data from a study of 11 healthy subjects further confirm the safety profile of the device. The study results demonstrated that treatment consistently demonstrated discrete areas of fat necrosis within the subcutaneous tissue. The effect is present solely in the subcutaneous tissue, while the dermal and epidermal layers are intact with no adverse skin events.
Therefore, clinical evaluation of the SlimShape device demonstrated the favorable performance and safety profile of the device for lipolysis of the abdomen and non-invasive reduction of the circumference. Results thus further support substantial equivalence of the device compared to the predicate.
# Conclusions
The SlimShape has the same intended use and similar indications for use, technological characteristics and principles of operation as its predicate device. Performance data demonstrate that the differences between the SlimShape and the predicate do not raise new
{7}------------------------------------------------
types of safety or effectiveness questions. Clinical studies of the SlimShape have demonstrated the safety and effectiveness profile of the device in the intended population.
Thus, the SlimShape is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.