Sol-Glide Safety Needle; Sol-Glide Safety Needle with Syringe

K253110 · Sol-Millennium Medical, Inc. · FMF · Jun 18, 2026 · General Hospital

Device Facts

Record IDK253110
Device NameSol-Glide Safety Needle; Sol-Glide Safety Needle with Syringe
ApplicantSol-Millennium Medical, Inc.
Product CodeFMF · General Hospital
Decision DateJun 18, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sol-Glide™ Safety Needle is used in conjunction with a standard syringe. The Sol-Glide™ Safety Needle with Syringe is used to inject fluids into, or withdraw fluids from the body. The safety mechanism encapsulates the needle after use. In the activated position, the protective shield guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination.

Device Story

The Sol-Glide Safety Needle is a sterile, single-use hypodermic needle featuring a manually activated sliding arm; it is used either alone with a standard syringe or pre-attached to a Luer Lock/Slip syringe. The device is operated by a clinician who, after completing an injection or fluid withdrawal, manually slides the protective shield over the needle to encapsulate it. This action guards against accidental needlesticks during handling and disposal. The device is used in clinical settings for standard fluid administration or withdrawal. The output is the physical delivery or extraction of fluids; the safety mechanism provides a protective barrier to prevent injury. The device benefits patients and clinicians by reducing the risk of sharps-related injuries.

Clinical Evidence

No clinical data. Bench testing only, including performance testing per ISO 7864, ISO 9626, ISO 80369-7, ISO 7886-1, and ISO 23908, as well as biocompatibility (ISO 10993-1), sterility (ISO 11135), and packaging integrity (ASTM F88, ASTM F1929, ISTA 3A).

Technological Characteristics

Single-use hypodermic needle (18G-25G) and syringe (1mL-10mL). Materials: Polypropylene (hub, cap, syringe barrel/plunger), Stainless Steel SUS 304 (cannula), IR rubber (gasket), Silicone oil (lubricant), UV-cured adhesive. Manual safety mechanism. EO sterilization. Complies with ISO 7864, ISO 9626, ISO 80369-7, ISO 7886-1, ISO 23908.

Indications for Use

Indicated for injection of fluids into or withdrawal of fluids from the body using a safety needle and syringe combination. Intended for use by clinicians to reduce risk of accidental needlestick injury via a manually activated protective shield.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{1} **FDA** **U.S. FOOD & DRUG** ADMINISTRATION June 18, 2026 Sol-Millennium Medical, Inc. Mike Xie VP, QA/RA 311 S Wacker Drive Suite 4100 Chicago, Illinois 60606 Re: K253110 Trade/Device Name: Sol-Glide Safety Needle; Sol-Glide Safety Needle with Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, QNQ, FMI Dated: May 18, 2026 Received: May 18, 2026 Dear Mike Xie: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K253110 - Mike Xie Page 2 Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {3} K253110 - Mike Xie Page 3 Sincerely, **SHRUTI N. MISTRY -S** Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. # Indications for Use 510(k) Number (if known) K253110 Device Name Sol-Glide Safety Needle; Sol-Glide Safety Needle with Syringe Indications for Use (Describe) The Sol-Glide™ Safety Needle is used in conjunction with a standard syringe. The Sol-Glide™ Safety Needle with Syringe is used to inject fluids into, or withdraw fluids from the body. The safety mechanism encapsulates the needle after use. In the activated position, the protective shield guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} Sol M Building a Healthier Tomorrow # 510(k) Summary - K253110 Date Prepared: 06/18/2026 1. SUBMITTER INFORMATION: Company Name: Sol-Millennium Medical, Inc. Address: 311 South Wacker Drive, Suite 4100, Chicago, Illinois, USA 60606 Company Person: Mike Xie Telephone: 847-400-4821 Email: yxie@solm.com 2. DEVICE IDENTIFICATION Trade Name: Sol-Glide Safety Needle; Sol-Glide Safety Needle with Syringe Common Name: Safety Hypodermic Needle, Safety Hypodermic Needle with Syringe Classification Name: Needle, hypodermic, single lumen Classification: II Product Code: FMF Subsequent Product codes: QNQ, FMI Regulation Number: 21 CFR 880.5570, 21 CFR 880.5860 Review Panel: General Hospital 3. PREDICATE DEVICES: Primary Predicate Device : K241821 - Sol-M Luer Lock Syringe with Safety Needle; Luer Lock Syringe with Exchangeable Needle; Luer Lock Syringe with Blunt Fill Needle Secondary Predicate Device: K242291 - Luer Lock/Slip Tip Syringe (Low Dead Space) without Needle or with Exchangeable Needle; Luer Lock Syringe (Low Dead Space) with Safety Needle Third Predicate Device: K012736 - Monoject Safety Needle 4. DEVICE DESCRIPTION: The SOL-Glide Safety Needle is a sterile, single-use, hypodermic needle with safety feature to be initiated by the end user by activating the sliding arm over the needle after injection completes, that is intended to administer medication in a conjunction with standard syringe. Needle gauge ranging from 18G to 25G, needle length including ¼", 1" and 1½". The SOL-Glide™ safety needle is sterilized by Ethylene Oxide Gas and supplied sterile in individual blister pack. The Sol-Glide Safety Needle with Syringe is the same Sol-Glide Safety Needle mentioned above but pre-attached to a Luer Lock or Luer Slip Slip syringe offered in 1ml (standard and low dead space), 3ml, 5ml or 10mL configurations in a sterile individual blister pack. 5. INDICATIONS FOR USE The Sol-Glide™ Safety Needle is used in conjunction with a standard syringe. The Sol-Glide™ Safety Needle with Syringe is used to inject fluids into, or withdraw fluids from the body. The safety mechanism encapsulates the needle after use. In the activated position, the protective shield guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. 6. TECHNOLOGICAL COMPARISONS Table 1. Comparison of Sol-Glide Safety Needle with syringe (needle with Luer lock syringe combination configuration) with the primary predicate | Item | Subject Device Sol-Glide Safety Needle with syringe (needle with Luer lock syringe) | Primary predicate Sol-M Luer Lock Syringe with Safety Needle; Luer Lock Syringe with Exchangeable Needle; Luer Lock Syringe with Blunt Fill Needle | Remark | | --- | --- | --- | --- | | K number | K253110 | K241821 | | {6} **Sol M**Building a Healthier Tomorrow | Classification | Class II | Class II | Same | | --- | --- | --- | --- | | Product Code and regulation number | FMF, FMI, QNQ 21 CFR 880.5570 21 CFR 880.5860 | FMF, FMI, MEG 21 CFR 880.5570 21 CFR 880.5860 | Different - Note 1 | | Indications for use | The Sol-Glide™ Safety Needle is used in conjunction with a standard syringe. The Sol-Glide™ Safety Needle with Syringe is used to inject fluids into, or withdraw fluids from the body. The safety mechanism encapsulates the needle after use. In the activated position, the protective shield guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. | The Luer Lock Syringe with Safety Needle is used to inject fluids into, or withdraw fluids from, the body. The safety mechanism covers the needle after use. In the activated position, the protective arm guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. | Different - Note 2 | | Needle Gauge Range | 18G - 25G | 18G - 30G | Different - Note 3 | | Needle Length Range | 5/8 inch to 1 ½ inch | 5/8 inch to 1 ½ inch | Same | | Needle Performance | Complied with ISO 7864, ISO 9626 | Complied with ISO 7864, ISO 9626 | Same | | Syringe Volume Range | 1mL, 3mL, 5mL and 10mL | 1mL, 3mL, 5mL and 10 mL | Same | | Luer Connector Performance | Luer Lock Complied with ISO 80369-7 | Luer Lock Complied with ISO 80369-7 | Same | | Material (Syringe) | Barrel | Polypropylene(PP) | Polypropylene(PP) | | | Plunger | Polypropylene(PP) | Polypropylene(PP) | | | Gasket | IR rubber | IR rubber | | | Lubricant | Silicone oil | Silicone oil | | Material (Needle) | Needle hub | Polypropylene(PP) | Polypropylene(PP) | | | Needle cap | Polypropylene(PP) | Polypropylene(PP) | | | Needle | Stainless Steel SUS 304 | Stainless Steel SUS 304 | | | Adhesive | UV Glue | UV Glue | | Biocompatibility | Complied with ISO 10993-1 | Complied with ISO 10993-1 | Same | | Sterilization Method | EO Sterilized | EO Sterilized | Same | #### Discussion in details: ##### Note 1: Product Code and regulation number The subject device includes product codes FMF, FMI and QNQ, whereas the predicate has similar codes of FMF, FMI and MEG, which correspond to the syringe piston component and the hypodermic single lumen needle component, respectively. The absence of MEG from the subject device listing in this comparison does not reflect a difference in intended use, operating principle, or safety performance as the subject device uses only the syringe (FMF) component from this predicate. Additionally, the subject device has QNQ which this predicate does not, however the QNQ refers to a low dead space syringe that will be applicable in the second predicate and does not apply to this primary predicate. Therefore, the difference in product code assignment is administrative/classification-related and does not raise new or different questions of safety or effectiveness when compared to predicate. ##### Note 2: Indications for use The subject device and predicate have the similar indications for use. Both describe the functions of a safety needle and syringe combination that when used together can inject and withdraw fluids from the body. Both describe a safety mechanism that guards against accidental needle stick injuries. Although the wording differs slightly between the devices, the functional intent and clinical use are equivalent. Therefore, these differences do not raise new or different questions of safety or effectiveness when compared to predicate. {7} **Sol M**Building a Healthier Tomorrow # **Note 3: Needle gauge Range** The predicate device has a greater range of gauges than the subject device. The subject device only intends to use a subset of needle gauges within range in combination with the Sol-Glide Safety needle and within the cleared range of the reference device. Therefore, these differences do not raise new or different questions of safety or effectiveness when compared to predicate. **Table 2. Comparison of Sol-Glide Safety Needle with syringe (needle with luer lock and luer slip Low Dead Space syringe combination configuration) with secondary predicate** | Item | Proposed Device Sol-Glide Safety Needle with syringe (needle with luer lock and luer slip Low Dead Space syringe) | Secondary predicate device Luer Lock/Slip Tip Syringe (Low Dead Space) without needle or with Exchangeable Needle Luer Lock Syringe (Low Dead Space) with Safety needle | Remark | | --- | --- | --- | --- | | K number | K253110 | K242291 | | | Classification | Class II | Class II | Same | | Product Code and regulation number | FMF, FMI, QNQ 21 CFR 880.5570 21 CFR 880.5860 | QNQ 21 CFR 880.5860 | Different - Note 1 | | Indications for use | The Sol-Glide™ Safety Needle is used in conjunction with a standard syringe. The Sol-Glide™ Safety Needle with Syringe is used to inject fluids into, or withdraw fluids from the body. The safety mechanism encapsulates the needle after use. In the activated position, the protective shield guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. | The Luer Lock/Slip Tip Syringe (Low Dead Space) without needle is to inject fluids into, or withdraw fluids from, the body. The Luer Lock/Slip Tip Syringe (Low Dead Space) with exchangeable needle is used to inject fluids into, or withdraw fluids from, the body. The Luer Lock Syringe (Low Dead Space) with Safety needle is used to inject fluids into, or withdraw fluids from, the body. The safety mechanism covers the needle after use. In the activated position, the protective arm guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. | Different - Note 2 | | Needle Gauge Range | 23G - 25G | 23G - 30G | Different - Note 3 | | Needle Length Range | 5/8 inch - 1 inch | 1/2 inch to 1 ½ inch | Different - Note 4 | | Needle Performance | Complied with ISO 7864, ISO 9626 | Complied with ISO 7864, ISO 9626 | Same | | Syringe Volume Range | 1mL | 1mL | Same | | Dead Space Specification | Syringe: Low Dead Space Max: 0.02mL Needle: Dead Space not calculated | Syringe: Low Dead Space Max: 0.02mL Needle: Dead Space not calculated | Same | | Luer Connector Performance | Luer lock and Luer Slip Complied with ISO 80369-7 | Luer lock and Luer Slip Complied with ISO 80369-7 | Same | | Syringe performance | Conforming with ISO 7886-1 | Conforming with ISO 7886-1 | Same | | Material (Syringe) | Barrel | Polypropylene(PP) | Same | | | Plunger | Polypropylene(PP) | | | | Gasket | IR rubber | | | | Lubricant | Silicone oil | | {8} **Sol M**Building a Healthier Tomorrow | Material (Needle) | Needle hub | Polypropylene(PP) | Polypropylene(PP) | Same | | --- | --- | --- | --- | --- | | | Needle cap | Polypropylene(PP) | Polypropylene(PP) | | | | Needle | Stainless Steel SUS 304 | Stainless Steel SUS 304 | | | | Adhesive | UV Glue | UV Glue | | | Biocompatibility | Complied with ISO 10993-1 | | Complied with ISO 10993-1 | Same | | Sterilization Method | EO Sterilized | | EO Sterilized | Same | # **Discussion in details:** # **Note 1: Product Code and regulation number** This predicate device is used only for the syringe component of the clearance. Both subject device and this predicate device are combinations of needles pre-attached to syringes however the needle component is different between the subject device and predicate device while the syringe is the same. The predicate device is QNQ which is encompassed in the subject device's list of product codes and regulation. The intent of this predicate device is to be included with the subject device so the subject device including the needle's product codes (FMI) in addition to the syringe's (QNQ) does not raise new or different questions of safety or effectiveness when compared to predicate. Additionally, the subject device's FMF product code corresponds to other non low dead space syringes that are bundled with the needle and accounted for in the primary predicate. # **Note 2: Indications for use** The subject device and predicate device have similar indications for use. Both describe the functions of a safety needle and syringe combination that when used together can inject and withdraw fluids from the body. Both describe a safety mechanism that guards against accidental needle stick injuries. Although the wording differs slightly between the devices, the functional intent and clinical use are equivalent. Therefore, these differences do not raise new or different questions of safety or effectiveness when compared to predicate. # **Note 3: Needle Gauge Range** The predicate device has a greater range of gauges than the subject device. The subject device only intends to use a subset of needle gauges within range in combination with the Sol-Glide Safety needle and within the cleared range of the reference device. Therefore, these differences do not raise new or different questions of safety or effectiveness when compared to predicate. # **Note 4: Needle Length Range** The predicate device has a greater range of needle length than the subject device. The subject device only intends to use a subset of needle lengths within range in combination with the Sol-Glide Safety needle and within the cleared range of the reference device. Therefore, these differences do not raise new or different questions of safety or effectiveness when compared to predicate. **Table 3. Substantial equivalent comparison of Sol-Glide Safety Needle (needle only) to the third predicate.** | Item | Subject Device Sol-Glide Safety Needle | Third Predicate device Monoject Safety Needle | Remark | | --- | --- | --- | --- | | K number | K253110 | K012736 | | | Classification | Class II | Class II | Same | | Product Code and regulation number | FMI | FMI | Same | | Regulation Number | 21 CFR 880.5570 | 21 CFR 880.5570 | Same | | Indications for use | The Sol-Glide™ Safety Needle is used in conjunction with a standard syringe. The Sol-Glide™ Safety Needle with Syringe is used to inject fluids into, or withdraw fluids from the body. The safety mechanism encapsulates the needle after use. In the activated position, the protective shield guards against accidental needle stick during normal handling and disposal of the used | The primary function of the device is to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The needle stick prevention feature of the device helps prevent accidental needle sticks by shielding the needle after use. | Different - Note 1 | {9} Sol M Building a Healthier Tomorrow | | | needle/syringe combination. | | | | --- | --- | --- | --- | --- | | Configuration | | Needle cap, needle tube, needle hub with safety arm | Needle cap, needle tube, needle hub with safety arm | Same | | Safety feature | | Active, manual | Active, manual | Same | | Shielding Function | | Protective shield covers needle after use to reduce accidental needlestick injury | Protective shield covers needle after use to reduce accidental needlestick injury | Same | | Operation Mode | | For manual use only | For manual use only | Same | | Sterilized | | Yes | Yes | Same | | Single use | | Single use | Single use | Same | | Needle | size | 18G, 19G, 20G, 21G, 22G, 23G, 25G | 18G, 19G, 20G, 21G, 22G, 23G, 25G. | Same | | | Length | 5/8", 1", 1 ½" | 5/8", 1", 1 1/2" | | | Cannula wall | | Thin Wall | Thin Wall | Same | | Needle Performance | | Complied with ISO 7864, ISO 9626 | Complied with ISO 7864, ISO 9626 | Same | | Luer Connector Performance | | Complied with ISO 80369-7 | Complied with ISO 80369-7 | Same | | Materials | Needle cap | Polypropylene | Information not available in public record | Different - Note 2 | | | Needle hub | Polypropylene | Information not available in public record | | | | Needle tube | Stainless Steel SUS 304 | Information not available in public record | | | | Lubricants | Silicone oil | Information not available in public record | | | | Adhesive | UV adhesive | Information not available in public record | | | Biocompatibility | | Complied with ISO 10993-1 | Complied with ISO 10993-1 | Same | | Sterilization Method | | EO Sterilized | Radiation Sterilized | Different - Note 3 | # Discussion in details: # Note 1: Indications for use The subject device and predicate device have the same intended use and similar operating principles. Both devices are intended for the injection and withdrawal of fluids and incorporate manually activated safety mechanisms designed to reduce accidental needlestick injuries after use. Although the wording differs slightly between the devices, the functional intent and clinical use are equivalent. Therefore, these differences do not raise new or different questions of safety or effectiveness when compared to predicate. # Note 2: Materials The specific material composition for the predicate device (K012736) is not fully available in the public 510(k) summary or other public records. The proposed device utilizes standard medical-grade materials commonly used in hypodermic safety needle devices, including polypropylene, stainless steel, silicone oil lubricant, and a UV-cured medical-grade adhesive. All patient-contacting and externally communicating components of the proposed device have been evaluated for biocompatibility in accordance with ISO 10993 standards and demonstrated to be safe for the intended use. The adhesive is fully encapsulated within the device and does not directly contact the patient. Therefore, any differences in material specification do not raise new questions of safety or effectiveness when compared to predicate. # Note 3: Sterilization method The predicate device is sterilized using radiation sterilization, while the proposed device is sterilized using ethylene oxide (EO). Both radiation and EO are FDA-recognized, validated terminal sterilization methods widely used for single-use medical devices. The EO sterilization process for the proposed device has been validated in accordance with ISO 11135 and demonstrates a Sterility Assurance Level (SAL) of 10⁻⁶, equivalent to the predicate device. Therefore, the difference in sterilization method does not raise new questions of safety or effectiveness when compared to predicate. {10} **Sol M**Building a Healthier Tomorrow ## 7. Non clinical tests | Document Reference | Title | | --- | --- | | ISO 23908:2024 | Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling | | ISO 9626:2016-08-01 | Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods | | ISO 80369-7:2021-05 | Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications | | ISO 11607-1:2019 | Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems | | ASTM F88/F88M-21 | Standard Test Method for Seal Strength of Flexible Barrier Materials | | ASTM F1929-15 | Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration | | ISO 10993-1: 2018 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process | | ISTA 3A 2018 | Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less | | ISO 7886-1: 2017 | Sterile hypodermic syringes for single use - Part 1: Syringes for manual use | | ISO 7864: 2016-08-01 | Sterile hypodermic needles for single use - Requirements and test methods | | USP 788 | Particulate matter in injections | | USP 151 | Pyrogenicity test | | USP 85 | Bacterial Endotoxin | ## 8. Clinical test This submission does not contain clinical tests. ## Conclusion The Sol-Glide™ Safety Needle and Sol-Glide™ Safety Needle with Syringe are substantially equivalent to the predicate devices K241821, K242291, and K012736 with respect to intended use and technological characteristics. The subject device incorporates syringes previously cleared under K241821 (standard syringe configurations) and K242291 (Low Dead Space syringe configuration) and these components are unchanged from their legally marketed predicate devices with respect to design, materials, intended use, and performance specifications. The subject device meets the relevant ISO, ASTM and USP standards which demonstrate that any differences between subject device and predicates do not raise new or different questions of safety or effectiveness when compared to the predicate. Therefore, the Sol-Glide™ Safety Needle and Sol-Glide™ Safety Needle with Syringe are substantially equivalent to the predicate devices K241821, K242291, and K012736.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...