Luer Lock Syringe with Safety Needle; Luer Lock Syringe with Exchangeable Needle; Luer Lock Syringe with Blunt Fill Needle

K241821 · Sol-Millennium Medical, Inc. · FMF · Sep 20, 2024 · General Hospital

Device Facts

Record IDK241821
Device NameLuer Lock Syringe with Safety Needle; Luer Lock Syringe with Exchangeable Needle; Luer Lock Syringe with Blunt Fill Needle
ApplicantSol-Millennium Medical, Inc.
Product CodeFMF · General Hospital
Decision DateSep 20, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Luer Lock Syringe with Exchangeable Needle is used to inject medicines and vaccines into, or withdraw fluids from. the body. The Luer Lock Syringe with Safety Needle is used to inject fluids into, or withdraw fluids from, the body. The safety mechanism covers the needle after use. In the activated position, the protective arm guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. The Luer Lock Syringe with Blunt Fill Needle is used for aspiration from multi-dose medicine vials.

Device Story

Manual, single-use, 3-piece piston syringes with Luer lock connectors. Variants include safety needles with protective arms to prevent accidental needle sticks, exchangeable needles, and blunt fill needles for vial aspiration. Operated by medical professionals in clinical settings. Device components include polypropylene barrels/plungers, IR rubber gaskets, and stainless steel (SUS 304) needle tubes. Safety mechanism manually activated post-use to cover needle point. Benefits include accurate fluid delivery/aspiration and reduced risk of sharps injury for healthcare workers.

Clinical Evidence

No clinical data included. Substantial equivalence supported by bench testing, including syringe/needle performance (ISO 7886-1, ISO 7864, ISO 9626), biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis), and sterilization/packaging validation (USP <85>, <151>, <788>, ASTM F88, F1929, ISTA 3A).

Technological Characteristics

Materials: Polypropylene (barrel, plunger, needle hub, cap), IR rubber (gasket), Stainless Steel SUS 304 (needle tube). Lubricant: Silicone oil. Adhesive: Epoxy glue. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 7886-1, ISO 7864, ISO 9626, ISO 80369-7, ISO 10993 series, ASTM F756, F88, F1929, USP <85>, <151>, <788>.

Indications for Use

Indicated for medical professionals for the injection of medicines/vaccines or withdrawal of fluids from the body, and for aspiration from multi-dose vials. Applicable to patients requiring fluid injection or withdrawal.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 20, 2024 Sol-Millennium Medical Inc. % Alice Huang, RA manager Shanghai Mind-link Consulting Co., Ltd. 377 Tianzhu Road, Jiading Shanghai, 201821, China #### Re: K241821 Trade/Device Name: Luer Lock Syringe with Safety Needle: Luer Lock Syringe with Exchangeable Needle; Luer Lock Syringe with Blunt Fill Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, MEG, FMI Dated: June 18, 2024 Received: September 4, 2024 #### Dear Alice Huang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {3}------------------------------------------------ Sincerely, # Shruti N. Mistry -S Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K241821 #### Device Name Luer Lock Syringe with Safety Needle Luer Lock Syringe with Exchangeable Needle Luer Lock Syringe with Blunt Fill Needle #### Indications for Use (Describe) The Luer Lock Syringe with Exchangeable Needle is used to inject medicines and vaccines into, or withdraw fluids from. the body. The Luer Lock Syringe with Safety Needle is used to inject fluids into, or withdraw fluids from, the body. The safety mechanism covers the needle after use. In the activated position, the protective arm guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. The Luer Lock Syringe with Blunt Fill Needle is used for aspiration from multi-dose medicine vials. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Sol M, which is a company focused on building a healthier tomorrow. The logo features a green circle and the text "Sol M" in a dark green font. To the right of the company name is the text "Building a Healthier Tomorrow" in a smaller, lighter font. # K241821 510(K) Summary ## 1. Preparation date: September 20, 2024 ## 2. Submitter Address: 311 S Wacker Drive, Suite 4100, Chicago, Illinois, 60606, United States Contact person: Manu Kalia, Global QARA Director, +1 847-313-9577, mkalia@solm.com Submission correspondent: Alice Alice Huang, RA Manager, +86-15618536177, alice.huang@mind-link.net ## 3. Device Common name: Luer Lock Syringe with Safety Needle Luer Lock Syringe with Exchangeable Needle Luer Lock Syringe with Blunt Fill Needle Trade name: Sol-Care Luer Lock Syringe with Safety Needle Sol-M Luer Lock Syringe with Exchangeable Needle Sol-M Luer Lock Syringe with Blunt Fill Needle #### Regulatory Information: Classification Name: Syringe, Piston Classification: II Product Code: FMF Regulation Number: 21CFR 880.5860 Review Panel: General Hospital; Classification Name: Syringe, Antistick Classification: II Product Code: MEG {6}------------------------------------------------ Image /page/5/Picture/0 description: The image is a logo for Sol M. The logo features an orange circle on the left, followed by the word "Sol" in green, and then the letter "M" in green. To the right of the "M" are the words "Building a Healthier Tomorrow" in a smaller font. The logo is simple and modern, and the colors are bright and inviting. Regulation Number: 21CFR 880.5860 Review Panel: General Hospital; Classification Name: Needle, Hypodermic, Single Lumen Classification: II Product Code: FMI Regulation Number: 21CFR 880.5570 Review Panel: General Hospital #### 4. Predicate device K221247 Sterile Disposable Syringe with Safety Needle, Sterile Disposable Syringe with Needle, Sterile Disposable Safety Needle, Sterile Disposable Needle #### 5. Device description The Luer Lock Syringe with Exchangeable Needle is a sterile, single-use, standard 3-piece piston syringe with a detachable pre-attached needle. The Luer Lock with Safety Needle is a syringe and needle combination with the safety needle. It contains a mechanism that covers the needle point after use. In the activated position, the needle cover guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. The Safety needle can be pre-attached or aside the syringe. The Luer Lock Syringe with Blunt Fill Needle is a sterile, single-use, standard 3-piece piston syringe with a detachable pre-attached needle. It is used for aspiration from multi-dose medicine vials. The proposed devices are offered in various gauge sizes and length. The proposed devices are available in EO sterilized to achieve a Sterility Assurance Level (SAL) of 10-6. The device is for medical professionals use only and for prescription use only. - Luer Lock Syringe with Safety Needle (Syringe range: 1ml, 3ml, 5ml,10ml; Needle range: 20G, 21G, 23G, 25G; Needle length from 5/8" to 1 1/2"). The Safety needle can be pre-attached or aside the syringe. - Luer Lock Syringe with Exchangeable Needle (Syringe range: 3ml, 5ml, 10ml, Needle range: 20G, 21G, 22G, 23G, 25G; Needle length from 5/8" to 1 1/2"). - Luer Lock Syringe with Blunt Fill Needle (Syringe range: 3ml, 5ml, 10ml, Needle: 18G, {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Sol M. The logo consists of an orange circle, followed by the text "Sol M" in green. To the right of the "M" is the text "Building a Healthier Tomorrow" in a smaller font size. The logo is simple and modern, and the colors are bright and inviting. Needle length:1 1/2"). #### 6. Indications for use/Intended use The Luer Lock Syringe with Exchangeable Needle is used to inject medicines and vaccines into, or withdraw fluids from, the body. The Luer Lock Syringe with Safety Needle is used to inject fluids into, or withdraw fluids from, the body. The safety mechanism covers the needle after use. In the activated position, the protective arm guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination. The Luer Lock Syringe with Blunt Fill Needle is used for aspiration from multi-dose medicine vials. #### 7. Summary of Technology Characteristics Table 1. Substantial equivalent comparison of Standard Luer Lock Syringe with Safety Needle | Item | Proposed Device<br>Luer Lock Syringe with<br>Safety Needle | Predicate Device<br>Sterile Disposable Syringe<br>with Safety Needle | Remark | | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|---------------------| | K number | K241821 | K221247 | | | | Classification | Class II | Class II | Same | | | Product Code | FMF<br>FMI<br>MEG | FMF<br>FMI<br>MEG | Same | | | Regulation<br>Number | 21 CRF 880.5860<br>21 CRF 880.5570 | 21 CRF 880.5860<br>21 CRF 880.5570 | Same | | | Indications for use | The Luer Lock Syringe with<br>Safety Needle is used to inject<br>fluids into, or withdraw fluids<br>from, the body. The safety<br>mechanism covers the needle<br>after use. In the activated<br>position, the protective arm<br>guards against accidental<br>needle stick during normal<br>handling and disposal of the<br>used needle/syringe<br>combination. | The Sterile Disposable<br>Syringe with Safety Needle<br>is intended for use in the<br>aspiration and injection of<br>fluids for medical purpose.<br>After withdrawal of the<br>needle from the body, the<br>attached needle safety shield<br>can be manually activated to<br>cover the needle<br>immediately after use to<br>minimize risk of accidental<br>needle sticks. | Different<br>Note 1 | | | Configuration | Barrel, plunger, Gasket, needle<br>cap, needle tube, needle hub<br>with safety arm | Barrel, plunger, piston,<br>needle hub, needle tube,<br>needle cap, safety<br>mechanism | Different<br>Note 2 | | | Operation Mode | For manual use only | For manual use only | Same | | | Sterilized | Yes | Yes | Same | | | Single use | Single use | Single use | Same | | | Label/Labeling | Complied with 21CFR part 801 | Complied with 21 CFR part 801 | Same | | | Syringe | Volume | 1ml, 3ml, 5ml, 10ml | 0.5ml, 1ml, 2ml, 3ml, 5ml,<br>10ml, 20ml, 30ml, 50ml,<br>60ml | Different<br>Note 3 | | | Connect | 1ml: Luer lock, Fixed;<br>Others: Luer lock | 0.5ml,1ml: Luer Lock, Luer<br>Slip, Fixed needle;<br>Others: Luer Lock and<br>Luer Slip | | | Needle | Size | 20G, 21G, 22G, 23G, 25G | 18G, 20G, 21G, 22G, 25G,<br>27G | | | | Length | 1", 5/8", 11/2" | 1/2", 5/8", 3/4", 1", 1-1/4",<br>11/2" | | | Syringe<br>Performance | | Complied with ISO 7886-1 | Complied with ISO 7886-1 | Same | | | | | | | | Needle<br>Performance | | Complied with ISO 7864,<br>ISO 9626 | Complied with ISO 7864,<br>ISO 9626 | Same | | Luer Connector<br>Performance | | Complied with ISO 80369-7 | Complied with ISO 80369-7 | Same | | Barrel | | Polypropylene(PP) | Polypropylene(PP) | Same | | Plunger | | Polypropylene(PP) | Polypropylene(PP) | Same | | Gasket/Piston | | IR rubber | Polyisoprene | Different note 4 | | Needle hub | | Polypropylene(PP) | Polypropylene(PP) | Same | | Needle tube | | Stainless Steel SUS 304 | Stainless Steel SUS 304 | Same | | Lubricants | | Silicone oil | Silicone oil | Same | | Adhesive | | Epoxy glue | UV adhesive | Different note 4 | | Cytotoxicity | | No cytotoxicity | No cytotoxicity | Same | | Irritation | | No intracutaneous<br>reactivity | No intracutaneous<br>reactivity | Same | | Sensitization | | No sensitization | No sensitization | Same | | Systemic Toxicity | | No systemic toxicity | No systemic toxicity | Same | | Hemolysis | | No Hemolysis | No Hemolysis | Same | | Pyrogen | | No Pyrogen | No Pyrogen | Same | {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Sol M, a company focused on building a healthier tomorrow. The logo features a blurred green and orange circle followed by the text "Sol M" in a dark green sans-serif font. To the right of "Sol M" is the text "Building a Healthier Tomorrow" in a smaller, lighter font. {9}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for SolM. The logo consists of an orange circle, followed by the text "SolM" in dark green. To the right of the "M" are the words "Building a Healthier Tomorrow" in a smaller, lighter green font. The logo is simple and modern, with a focus on health and the future. | Sterilization | | | |-----------------|------------------|------| | Method | EO Sterilized | Same | | SAL | $10^{-6}$ | Same | | Endotoxin Limit | 20 EU per device | Same | ## Discussion in details: #### Note 1: Intended Use and Indications for use Standard Luer Lock Syringe with Safety Needle and Sterile Disposable Syringe with Safety Needle have similar intended use, but are described differently. Both are intended for use in the withdrawal and injection of fluids, and the protective arm can be manually activated to cover the needle after use to minimize risk of accidental needle sticks. These differences do not raise new or different questions of safety and effectiveness when compared to the predicate device. ## Note 2: Configuration The component names of the proposed device is different to that of the predicate device; however, the components share the same configuration and function. Therefore, this difference in name will not affect the Substantial Equivalence (SE) between the proposed and predicate device. #### Note 3: Syringe volume, Needle gauge and Needle length The syringe volume, needle gauge and needle length of the proposed device is less than that of predicate devices. This difference in syringe and needle size will not affect the performance of the syringe and needle. In addition, the syringe and needle size of the proposed device has been tested. The test results comply with ISO 7864 and ISO 9626 standards requirements. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device. #### Note 4: Patient-contacting Material Although the material of the proposed and the predicate device is different, biocompatibility of the proposed device complies to the ISO 10993 series of standards. Therefore, this difference will not affect the Substantial Equivalence (SE) between the proposed and predicate device. Table 2. Substantial equivalence comparison of Luer Lock Syringe with Exchangeable Needle | Item | Proposed Device | Predicate Device | Remark | | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------| | | Luer Lock Syringe with | Sterile Disposable Syringe | | | | | Exchangeable Needle | with Needle | | | | K number | K241821 | K221247 | | | | Classification | Class II | Class II | Same | | | Product Code | FMF<br>FMI | FMF<br>FMI | Same | | | Regulation<br>Number | 21 CRF 880.5860<br>21 CRF 880.5570 | 21 CRF 880.5860<br>21 CRF 880.5570 | Same | | | Indications for use | The Luer Lock Syringe with<br>Exchangeable Needle is used<br>to inject medicines and<br>vaccines into, or withdraw<br>fluids from, the body. | The Sterile Disposable<br>Syringe with Needle is<br>intended for use in the<br>aspiration and injection of<br>fluids for medical purpose. | Different<br>Note 1 | | | Configuration | Barrel, plunger, Gasket, needle<br>cap, needle tube, needle hub | Barrel (Luer lock/Luer<br>slip/fixed needle), plunger,<br>piston, needle hub, needle<br>tube, needle cap | Different<br>Note 2 | | | Operation Mode | For manual use only | For manual use only | Same | | | Sterilized | Yes | Yes | Same | | | Single use | Single use | Single use | Same | | | Label/Labeling | Complied with 21CFR part 801 | Complied with 21 CFR<br>part 801 | Same | | | Syringe Volume | 1ml, 3ml, 5ml, 10ml | 0.5ml, 1ml, 2ml, 3ml, 5ml,<br>10ml, 20ml, 30ml, 50ml,<br>60ml | Different<br>Note 3 | | | Needle size | 20G, 21G, 22G, 23G, 25G | 18G, 20G, 21G, 22G, 25G,<br>27G | Different<br>Note 3 | | | Length | 1", 5/8", 11/2" | 1/2", 5/8", 3/4", 1", 1-1/4",<br>11/2" | | | | Syringe<br>Performance | Complied with ISO 7886-1 | Complied with ISO 7886-1 | Same | | | Needle<br>Performance | Complied with ISO 7864,<br>ISO 9626 | Complied with ISO 7864,<br>ISO 9626 | Same | | | Luer Connector<br>Performance | Complied with ISO 80369-7 | Complied with ISO 80369-7 | Same | | | Barrel | Polypropylene(PP) | Polypropylene(PP) | Same | | | Plunger | Polypropylene(PP) | Polypropylene(PP) | Same | | | Gasket/Piston | IR rubber | Polyisoprene | Different Note 4 | | | Needle hub | Polypropylene(PP) | Polypropylene(PP) | Same | | | Needle tube | Stainless Steel SUS 304 | Stainless Steel SUS 304 | Same | | | Needle cap | Polypropylene(PP) | Polypropylene(PP) | Same | | | Lubricants | Silicone oil | Silicone oil | Same | | | Adhesive | Epoxy glue | UV adhesive | Different Note 4 | | | Cytotoxicity | No cytotoxicity | No cytotoxicity | Same | | | Irritation | No intracutaneous<br>reactivity | No intracutaneous<br>reactivity | Same | | | Sensitization | No sensitization | No sensitization | Same | | | Systemic Toxicity | No systemic toxicity | No systemic toxicity | Same | | | Hemolysis | No Hemolysis | No Hemolysis | Same | | | Pyrogen | No Pyrogen | No Pyrogen | Same | | | Sterilization | | | | | | Method | EO Sterilized | EO Sterilized | Same | | | SAL | 10-6 | 10-6 | Same | | | Endotoxin Limit | 20 EU per device | 20 EU per device | Same | | | Item | | Proposed Device<br>Luer Lock Syringe with Blunt<br>Fill Needle | Predicate Device<br>Sterile Disposable Syringe<br>with Needle | Remark | | K number | | K241821 | K221247 | | | Classification | | Class II | Class II | Same | | Product Code | | FMF<br>FMI | FMF<br>FMI | Same | | Regulation<br>Number | | 21 CRF 880.5860<br>21 CRF 880.5570 | 21 CRF 880.5860<br>21 CRF 880.5570 | Same | | Indications for use | | The Luer Lock Syringe with<br>Blunt Fill Needle is used for<br>aspiration from multi-dose<br>medicine vials. | The Sterile Disposable<br>Syringe with Needle is<br>intended for use in the<br>aspiration and injection of<br>fluids for medical purpose. | Different<br>Note 1 | | Configuration | | Barrel, plunger, Gasket, needle<br>cap, needle tube, needle hub | Barrel (Luer lock/Luer<br>slip/fixed needle), plunger,<br>piston, needle hub, needle<br>tube, needle cap | Different<br>Note 2 | | Operation Mode | | For manual use only | For manual use only | Same | | Sterilized | | Yes | Yes | Same | | Single use | | Single use | Single use | Same | | Label/Labeling | | Complied with 21CFR part 801 | Complied with 21 CFR<br>part 801 | Same | | Syringe | Volume | 3ml, 5ml, 10ml | 0.5ml, 1ml, 2ml, 3ml, 5ml,<br>10ml, 20ml, 30ml, 50ml,<br>60ml | Different<br>Note 3 | | Needle | size | 18G | 18G, 20G, 21G, 22G, 25G,<br>27G | Different<br>Note 3 | | | Length | 11/2" | 1/2", 5/8", 3/4", 1", 1-1/4",<br>11/2" | | | Syringe<br>Performance | | Complied with ISO 7886-1 | Complied with ISO 7886-1 | Same | | Needle<br>Performance | | Complied with ISO 7864,<br>ISO 9626 | Complied with ISO 7864,<br>ISO 9626 | Same | | Luer Connector<br>Performance | | Complied with ISO 80369-7 | Complied with ISO 80369-7 | Same | | | | | | | | Barrel | Polypropylene(PP) | Polypropylene(PP) | Same | | | Plunger | Polypropylene(PP) | Polypropylene(PP) | Same | | | Gasket/Piston | IR rubber | Polyisoprene | Different Note 4 | | | Needle hub | Polypropylene(PP) | Polypropylene(PP) | Same | | | Needle tube | Stainless Steel SUS 304 | Stainless Steel SUS 304 | Same | | | Needle cap | Polypropylene(PP) | Polypropylene(PP) | Same | | | Lubricants | Silicone oil | Silicone oil | Same | | | Adhesive | Epoxy glue | UV adhesive | Different Note 4 | | | Cytotoxicity | No cytotoxicity | No cytotoxicity | Same | | | Irritation | No intracutaneous<br>reactivity | No intracutaneous<br>reactivity | Same | | | Sensitization | No sensitization | No sensitization | Same | | | Systemic Toxicity | No systemic toxicity | No systemic toxicity | Same | | | Hemolysis | No Hemolysis | No Hemolysis | Same | | | Pyrogen | No Pyrogen | No Pyrogen | Same | | | Sterilization | | | | | | Method | EO Sterilized | EO Sterilized | Same | | | SAL | 10-6 | 10-6 | Same…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...