Sterile Disposable Syringe with Safety Needle, Sterile Disposable Syringe with Needle, Sterile Disposable Safety Needle, Sterile Disposable Needle
K221247 · Jiangsu Kangbao Medical Equipment Co., Ltd. · MEG · Nov 1, 2022 · General Hospital
Device Facts
Record ID
K221247
Device Name
Sterile Disposable Syringe with Safety Needle, Sterile Disposable Syringe with Needle, Sterile Disposable Safety Needle, Sterile Disposable Needle
Applicant
Jiangsu Kangbao Medical Equipment Co., Ltd.
Product Code
MEG · General Hospital
Decision Date
Nov 1, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile Disposable Syringe with Safety Needle is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks. The Sterile Disposable Syringe with Needle is intended for use in the aspiration and injection of fluids for medical purpose. The Sterile Disposable Safety Needle is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks. The Sterile Disposable Needle is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose.
Device Story
Device consists of sterile, single-use syringes and hypodermic needles, with or without integrated safety mechanisms. Safety versions feature a manually activated shield to cover the needle post-withdrawal, reducing accidental needlestick risk. Syringes are available in volumes from 0.5ml to 60ml with luer lock or luer slip connectors; needles range from 18G to 27G. Used in clinical settings for fluid aspiration and injection. Operated manually by clinicians. Safety mechanism activation is a manual step performed by the user after needle withdrawal. Benefits include accurate fluid delivery/aspiration and enhanced sharps injury protection for the operator.
Clinical Evidence
No clinical study included. Bench testing only. Includes biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, complement activation, thromboresistance, pyrogen), mechanical/physical testing (ISO 7886-1, ISO 7864, ISO 9626, ISO 80369-7), and simulated clinical study for safety mechanism performance per ISO 23908:2011.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, hub), Polyisoprene (piston), Stainless Steel SUS 304 (needle tube), Silicone oil (lubricant), UV adhesive. Energy: Manual. Connectivity: None. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ISO 7886-1, ISO 7864, ISO 9626, ISO 80369-7, ISO 10993 series, ASTM F88, F1886, F1929.
Indications for Use
Indicated for aspiration and injection of fluids for medical purposes in patients requiring such procedures. Intended for manual, single-use by healthcare professionals.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 1, 2022
Jiangsu Kangbao Medical Equipment Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O.box 120-119 Shanghai, 200120 China
Re: K221247
Trade/Device Name: Sterile Disposable Syringe with Safety Needle, Sterile Disposable Syringe with Needle, Sterile Disposable Safety Needle, Sterile Disposable Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: MEG. FMF. FMI Dated: September 30, 2022 Received: September 30, 2022
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K221247
Device Name
Sterile Disposable Syringe with Safety Needle; Sterile Disposable Syringe with Needle; Sterile Disposable Safety Needle; Sterile Disposable Needle
# Indications for Use (Describe)
The Sterile Disposable Syringe with Safety Needle is intended for use in the aspiration of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.
The Sterile Disposable Syringe with Needle is intended for use in the aspiration of fluids for medical purpose.
The Sterile Disposable Safety Needle is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.
The Sterile Disposable Needle is intended to be used with a luer lock syringe for aspiration and injection of fluids for medical purpose.
Type of Use *(Select one or both, as applicable)*
| <span> <span style="font-size: 14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: K221247
- 1. Date of Preparation: November 1, 2022
- 2. Sponsor Identification
## Jiangsu Kangbao Medical Equipment Co., Ltd.
78#, North Suzhong Road Baoying 225800 Yangzhou PEOPLE'S REPUBLIC OF CHINA
Establishment Registration Number: 3009742443
Contact Person: Rujun Tang
Position: Management Representative
Tel: +86-514-88223540
Fax: +86-514-88232089
Email: 76823131@qq.com
- Designated Submission Correspondent 3.
Ms. Diana Hong (Primary Contact Person)
Ms. Tingting Su (Alternative Contact Person)
## Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Sterile Disposable Syringe with Safety Needle Sterile Disposable Syringe with Needle Sterile Disposable Safety Needle Sterile Disposable Needle
Common Name: Syringes with Needle
Regulatory Information Classification Name: Syringe, Piston Classification: II; Product Code: FMF; Regulation Number: 21CFR 880.5860; Review Panel: General Hospital;
Classification Name: Needle, Hypodermic, Single Lumen Classification: II Product Code: FMI; Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital;
Classification Name: Piston Syringe Classification: II; Product Code: MEG; Regulation Number: 21 CFR 880. 5860; Review Panel: General Hospital
### Indications for use:
The Sterile Disposable Syringe with Safety Needle is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.
The Sterile Disposable Syringe with Needle is intended for use in the aspiration and injection of fluids for medical purpose.
The Sterile Disposable Safety Needle is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to
{5}------------------------------------------------
minimize risk of accidental needle sticks.
The Sterile Disposable Needle is intended to be used with a luer slip or luer lock syringe for aspiration and injection of fluids for medical purpose.
### Device Description
The Sterile Disposable Safety Needle is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of needle tube, needle hub and safety mechanism. The proposed device is available in variety of needle gauges and lengths. The safety needle is compatible for use with a luer slip and luer lock syringe. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks. Needle specification is same as the needle size of Sterile Disposable Syringe with Safety Needle.
| Needle Gauge | Needle length |
|--------------|--------------------------------------|
| 18G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 20G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 21G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 22G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 25G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 27G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
Needle specification:
Compared with Sterile Disposable Safety Needle, Sterile Disposable Needle has the same components and specifications except without safety mechanism.
The Sterile Disposable Syringe with Safety Needle is intended for manual and single use only to aspirate and inject of fluids for medical purpose. There are two kinds of sterile disposable syringe with safety needle: syringe with fixed needles and syringe without fixed needles. Sterile disposable syringe with safety needle of 0.5ml and 1ml are available syringe with fixed needles. Syringe without fixing needle are consists of syringe (piston, barrel, plunger) and hypodermic needle with a safety mechanism. The proposed device is available in a variety combination of syringe volume and needle size.
Syringe volume: 0.5ml,1ml,2ml,3ml,5ml,10ml,20ml,30ml,50ml and 60ml Needle specification:
| Needle Gauge | Needle length |
|--------------|--------------------------------------|
| 18G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 20G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 21G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 22G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 25G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
| 27G | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" |
{6}------------------------------------------------
Compared with Sterile Disposable Syringe with Safety Needle, Sterile Disposable Syringe with Needle has the same components and specifications except without safety mechanism. The Sterile Disposable Syringe with Needle do not have a type of the syringe with fixed needle.
The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of 5 years.
- રું. Identification of Predicate Device
### Predicate Device
510(k) Number: K170651 Product Name: Sterile Disposable Syringe with Safety Needle Sterile Disposable Syringe with Needle Sterile Disposable Syringe Sterile Disposable Safety Needle Sterile Disposable Needle
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ISO 10993-5:2009 Biological evaluation of medical Devices-Part 5: Tests for in Vitro Cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization.
- > ISO 10993-11:2017 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity
- > ISO 10993-4:2017 Biological Evaluation of Medical Devices--Part 4: Selection of Tests for Interactions with Blood
- > ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
- > ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- > ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)
{7}------------------------------------------------
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- > ISO 7864:2016 Sterile hypodermic needles for single use — Requirements and test methods
- ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices >
- > ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications
- > ISO 7886-1:2017 Sterile hypodermic syringes for single use- Part 1: Syringes for manual use.
- > ISO 10993-7:2008 Biological Evaluation of Medical Device-Part 7: Ethylene Oxide Sterilization Residuals
- > ISO 23908:2011 Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
- > USP<85> Bacterial Endotoxins Test
- > USP<151> Pyrogen Test
- > USP<788> Particulate Matter in Injections
- > ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems (DC-13, Level II)
Physical, Mechanical, Chemical testing listed in following table were performed on the proposed device. The test results show that the device conforms with the requirements of related standards.
| General requirements | Clause 5 of ISO 7886-1:2017 |
|----------------------------------|------------------------------|
| Extraneous matter | Clause 6 of ISO 7886-1:2017 |
| Lubricant | Clause 7 of ISO 7886-1;2017 |
| Tolerance on graduated capacity | Clause 8 of ISO 7886-1:2017 |
| Graduated scale | Clause 9 of ISO 7886-1:2017 |
| Barrel | Clause 10 of ISO 7886-1:2017 |
| Piston/ plunger assembly | Clause 11 of ISO 7886-1:2017 |
| Nozzle | Clause 12 of ISO 7886-1:2017 |
| Performance | Clause 13 of ISO 7886-1:2017 |
| Cleanliness | Clause 4.3 of ISO 7864:2016 |
| Limits for acidity or alkalinity | Clause 4.4 of ISO 7864:2016 |
| Limits for extractable metals | Clause 4.5 of ISO 7864:2016 |
| Size designation | Clause 4.6 of ISO 7864:2016 |
| Colour coding | Clause 4.7 of ISO 7864:2016 |
| Needle hub | Clause 4.8 of ISO 7864:2016 |
| Needle Cap | Clause 4.9 of ISO 7864:2016 |
| Needle tube | Clause 4.10 of ISO 7864:2016 |
| Needle point | Clause 4.11 of ISO 7864:2016 |
| Bond between hub and needle tube | Clause 4.12 of ISO 7864:2016 |
{8}------------------------------------------------
| Patency of lumen | Clause 4.13 of ISO 7864:2016 |
|------------------------------------------|--------------------------------|
| Surface finish and appearance | Clause 5.2 of ISO 9626:2016 |
| Cleanliness | Clause 5.3 of ISO 9626:2016 |
| Limits for acidity and alkalinity | Clause 5.4 of ISO 9626:2016 |
| Size designation | Clause 5.5 of ISO 9626:2016 |
| Dimensions | Clause 5.6 of ISO 9626:2016 |
| Stiffness | Clause 5.8 of ISO 9626:2016 |
| Resistance to breakage | Clause 5.9 of ISO 9626:2016 |
| Resistance to corrosion | Clause 5.10 of ISO 9626:2016 |
| Item | Standard |
| Fluid leakage | Clause 6.1 of ISO 80369-7:2016 |
| Sub-atmospheric pressure air leakage | Clause 6.2 of ISO 80369-7:2016 |
| Stress cracking | Clause 6.3 of ISO 80369-7:2016 |
| Resistance to separation form axial load | Clause 6.4 of ISO 80369-7:2016 |
| Resistance to separation form unscrewing | Clause 6.5 of ISO 80369-7:2016 |
| Resistance to overriding | Clause 6.6 of ISO 80369-7:2016 |
| Particulate testing | USP <788> |
Sterile barrier packaging testing were performed on the proposed device, which include visual inspection (ASTM F1886/F1886M-16), seal strength (ASTM F88/F88-15) and dye penetration test (ASTM F1929-15). The test result showed that the device package can maintain its integrity.
Sterilization and shelf life testing listed in following table were performed on the proposed device. EO ECH residue did not exceed the limit of ISO 10993-7. Endotoxin limit did not exceed 20EU/device. Shelf life test result showed that the device can maintain its performance during the claimed shelf life.
| EO residue | ISO 10993-7:2008 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| ECH residue | ISO 10993-7:2008 |
| Bacteria Endotoxin Limit | USP <85> |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package Test<br>were performed on aging samples to verify the<br>claimed shelf life of the device |
### Biocompatibility testing
The contact level of the proposed device is blood path, indirect, and the contact duration is limited contact (<24 hours). The proposed device was evaluated for the following tests. The results for the
{9}------------------------------------------------
biocompatibility testing showed that there are no negative impacts from the materials that are used in the proposed device.
- > Cytotoxicity,
- > Sensitization,
- > Irritation,
- Acute Systemic Toxicity, >
- > Hemolysis,
- > Complement activation,
- > Thromboresistance study
- > Pyrogen
### Simulated Clinical Study
A simulated clinical study was performed on proposed device according to FDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 and ISO 23908:2011 to evaluate the safety mechanism of the proposed device. The results demonstrated that the proposed device met the pre-established criteria.
Safety Feature Test
The safety feature test was performed on both proposed device and predicate device to determine its safety feature. The results demonstrated that both the proposed device and predicate device meet the acceptance criteria.
#### 7. Clinical Test Conclusion
No clinical study is included in this submission.
{10}------------------------------------------------
#### Summary of Technology Characteristics 8.
| ITEM | Proposed Device | Predicate Device<br>K170651 | Remark |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product | Sterile Disposable Syringe with<br>Safety Needle | Sterile Disposable Syringe with<br>Safety Needle | / |
| Product Code | FMF<br>FMI<br>MEG | FMF<br>FMI<br>MEG | Same |
| Regulation<br>Number | 21 CRF 880.5860<br>21 CRF 880.5570 | 21 CRF 880.5860<br>21 CRF 880.5570 | Same |
| Class | Class II | Class II | Same |
| Indications for<br>Use | The Sterile Disposable Syringe<br>with Safety Needle is intended<br>for use in the aspiration and<br>injection of fluids for medical<br>purpose. After withdrawal of the<br>needle from the body, the<br>attached needle safety shield can<br>be manually activated to cover<br>the needle immediately after use<br>to minimize risk of<br>accidental needle sticks. | The Sterile Disposable Syringe<br>with Safety Needle is intended for<br>use in the aspiration and injection<br>of fluids for medical purpose.<br>After withdrawal of the needle<br>from the body, the attached needle<br>safety shield can be manually<br>activated to cover the needle<br>immediately after use to<br>minimize risk of<br>accidental needle sticks. | Same |
| Configuration | Syringe | Barrel (luer lock/ luer slip) | Different |
| | | Plunger<br>Piston<br>Needle hub<br>Needle tube<br>Needle cap<br>Safety sheath | |
| Operation Mode | For manual use only | For manual use only | Same |
| Sterilized | Yes | Yes | Same |
{11}------------------------------------------------
| | Single Use | Single Use | Single Use | Same |
|----------------------------|-------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------|-----------|
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | | Same |
| Syringe | Volume | 0.5ml, 1ml, 2ml, 3ml, 5ml, 10ml, 20ml, 30ml, 50ml, 60ml | 1ml, 2ml 3ml, 5ml, 10ml, 20ml, 30ml, 50ml,60ml | Different |
| | Connector Type | 0.5ml,1ml: Luer Lock, Luer Slip, Fixed needle; Others: Luer Lock and Luer Slip | Luer Lock/ Luer slip | |
| Needle | Size | 18G, 20G, 21G, 22G, 25G, 27G | 16G, 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G | Different |
| | Length | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" | 5/16", 1/2", 5/8", 1", 1-1/4", 1-1/2" 3/4", | |
| Syringe Performance | | Complied with ISO 7886-1 | Complied with ISO 7886-1 | Same |
| Needle Performance | | Complied with ISO 7864, ISO 9626 | Complied with ISO 7864, ISO 9626 | Same |
| Luer Connector Performance | | Complied with ISO 80369-7 | Complied with ISO 594-2 | Different |
| Barrel | | Polypropylene (PP) | Polypropylene (PP) | |
| Plunger | | Polypropylene (PP) | Polypropylene (PP) | |
| Piston | | Polyisoprene | Polyisoprene | |
| Needle hub | | Polypropylene (PP) | Polypropylene (PP) | Same |
| Needle tube | | Stainless Steel SUS 304 | Stainless Steel SUS 304 | Same |
| Lubricants | | Silicone oil | Silicone Oil | |
| Adhesive | | UV adhesive | UV glue | |
| Cytotoxicity | | No cytotoxicity | No cytotoxicity | |
| Irritation | | No intracutaneous reactivity | No intracutaneous reactivity | |
| Sensitization | | No sensitization | No skin sensitization | Same |
| Systemic Toxicity | | No systemic toxicity | No systemic toxicity | |
{12}------------------------------------------------
| Hemolysis | No Hemolysis | No Hemolysis | No Hemolysis | |
|----------------------------|----------------------------------------------------|----------------------------------------------------|-------------------------------------------------------|------|
| Pyrogen | No Pyrogen | No Pyrogen | No Pyrogen | |
| Complement<br>Activation | Not show potentials to<br>activate complete system | Not show potentials to<br>activate complete system | Not show potentials to<br>activate complete<br>system | |
| In-vivo<br>Thrombogenicity | No thrombogenicity | No thrombogenicity | No thrombogenicity | |
| Sterilization | | | | |
| Method | EO Sterilized | EO Sterilized | EO Sterilized | Same |
| SAL | 10-6 | 10-6 | 10-6 | Same |
| Endotoxin<br>Limit | 20 EU per device | 20 EU per device | 20 EU per device | Same |
# Different - Configuration
The configuration of proposed device is similar to the configurations of predicate device. For 0.5ml and 1ml syringe, the proposed device has luer lock connector and fixed needle, the other volumes have luer lock and luer slip connector. The predicate device only has luer slip connector. The proposed device has luer lock connector, luer slip connector and fixed needle have been tested and meet the requirements of relevant standards. Based on above analysis, the difference on configuration will not raise new questions on safety and effectiveness of the proposed device.
# Different -Syringe Volume
The syringe volume for proposed device is similar to the predicate devices. The predicate device does not have a 0.5ml syringe. This difference will not raise new questions on safety and effectiveness of the proposed device.
# Different - Syringe Connector Type
The syringe connector type of proposed device is similar as the predicate device. For 0.5ml and 1ml syringe, the proposed device has luer lock connector, luer slip connector and fixed needle, the other volumes have luer lock connector and luer slip. The predicate device has luer lock or luer slip connector.
{13}------------------------------------------------
In addition, the proposed device has luer lock connector, luer slip connector and fixed needle have been tested and meet the requirements of relevant standards. Based on above analysis, the difference on connector type will not raise new questions on safety and effectiveness of the proposed device.
## Different -Needle Size and Length
The needle size and length for proposed device is different from the predicate device. The needle length is very close to that of the predicate device. This difference will not raise new questions on safety and effectiveness of the proposed device.
# Different -Luer Connector Performance
Although the proposed device and the predicate device follow different luer connector standards - this is because ISO 594-1,594-2 is replaced by ISO 80369-7. The test results of the proposed device show that the connector performance meet the requirements of ISO 80369-7. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.
{14}------------------------------------------------
| ITEM | Proposed Device | Predicate Device<br>K170651 | Remark |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product | Sterile Disposable Syringe with<br>Safety Needle | Sterile Disposable Syringe with<br>Safety Needle | / |
| Product Code | FMF<br>FMI | FMF<br>FMI | Same |
| Regulation Number | 21 CRF 880.5860<br>21 CRF 880.5570 | 21 CRF 880.5860<br>21 CRF 880.5570 | Same |
| Class | Class II | Class II | Same |
| Indication for Use | The Sterile Disposable Syringe with<br>Needle is intended for use in the<br>aspiration and injection of fluids<br>for medical purpose. | The Sterile Disposable Syringe<br>with Needle is intended for use in<br>the aspiration and injection of<br>fluids for medical purpose. | Same |
| Configuration | Syringe | Barrel (luer lock/luer<br>slip)<br>Plunger<br>Piston | Different |
| | Syringe | Barrel (luer lock/ luer<br>slip/ fixed needle)<br>Plunger<br>Piston | |
| | Needle | Needle hub<br>Needle tube<br>Needle cap | |
| | Needle | Needle hub<br>Needle tube<br>Needle cap | |
| Operation Mode | For manual use only | | Same |
| Sterilized | Yes | | Same |
| Single Use | Single Use | | Same |
| Label/Labeling | Complied with 21 CFR part 801 | | Same |
| Syringe | Volume | 0.5ml, 1ml, 2ml, 3ml, 5ml, 10ml,<br>20ml, 30ml, 50ml, 60ml | Different |
| | Volume | 1ml, 2ml, 3ml, 5ml, 10ml, 20ml,<br>30ml, 50ml, 60ml | |
| Syringe | Connector<br>Type | Luer Lock and Luer Slip | Same |
| Needle | Size | 18G, 20G, 21G, 22G, 25G, 27G | Different |
| Needle | Size | 16G, 18G, 19G, 20G, 21G, 22G,<br>23G, 24G, 25G, 26G, 27G,<br>28G, 29G, 30G | |
| | Length | 1/2", 5/8", 3/4", 1", 1-1/4", 1-1/2" | |
| | Length | 5/16", 1/2", 5/8", 3/4", 1", 1-1/4",<br>1-1/2" | |
| Syringe<br>Performance | Complied with<br>ISO 7886-1 | | Same |
| Needle Performance | Complied with ISO 7864 | | Same |
| | ISO 9626 | ISO 9626 | |
| Luer<br>Connector<br>Performance | Complied with<br>ISO 80369-7 | Complied with<br>ISO 594-2 | Different |
| Patient-contact materials | | | |
| Barrel | Polypropylene (PP) | Polypropylene (PP) | |
| Plunger | Polypropylene (PP) | Polypropylene (PP) | |
| Piston | Polyisoprene | Polyisoprene | |
| Needle hub | Polypropylene (PP) | Polypropylene (PP) | Same |
| Needle tube | Stainless Steel SUS 304 | Stainless Steel SUS 304 | |
| Lubricants | Silicone oil | Silicone Oil | |
| Adhesive | UV adhesive | UV glue | |
| Biocompatibility | | | |
| Cytotoxicity | No cytotoxicity | No cytotoxicity | |
| Irritation | No intracutaneous reactivity | No intracutaneous reactivity | |
| Sensitization | No sensitization | No skin sensitization | Same |
| Systemic Toxicity | No systemic toxicity | No systemic toxicity | |
| Hemolysis | No Pyrogen | No Pyrogen | |
| Pyrogen | Not show potentials to activate<br>complete system | Not show potentials to activate<br>complete system | |
| Complement<br>Activation | No thrombogenicity | No thrombogenicity | |
| In-vivo<br>Thrombogenicity | No thrombogenicity | No thrombogenicity | |
| Sterilization | | | |
| Method | EO Sterilized | EO Sterilized | Same |
| SAL | 10-6 | 10-6 | |
| Endotoxin<br>Limit | 20 EU per device | 20 EU per device | |
Table 2. Comparison of Sterile Disposable Syringe with Needle
{15}------------------------------------------------
# 510(k) Summary
{16}------------------------------------------------
### Different -Syringe Volume
The syringe volume for proposed device is similar to the predicate devices. The predicate device does not have a 0.5ml syringe. This difference will not raise new questions on safety and effectiveness of the proposed device.
### Different -Needle Size and Length
The needle size and length for proposed device is different from the predicate device. This difference will not raise new questions on safety and effectiveness of the proposed device.
## Different - Luer Connector Performance
The luer connector performance of proposed device is complied with ISO 80369-7, the predicate device is complied with complied with ISO 594-2. They are all the test standards of luer connector. At present, ISO 594-2 has been replaced by ISO 80369-7. The predicate device was tested according to ISO 594-2 because the application time was earlier. The proposed devices are tested according to the latest version of the standard ISO 80369-7. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.
{17}------------------------------------------------
| ITEM | | Proposed Device | Predicate Device K170651 | Remark |
|---------------------------|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product | | Sterile Disposable Safety<br>Needle | Sterile Disposable Safety<br>Needle | / |
| | | | | |
| Product Code | | FMI | FMI | Same |
| Regulation Number | | 21 CRF 880.5860 | 21 CRF 880.5860 | Same |
| Class | | Class II | Class II | Same |
| Indication for Use | | The Sterile Disposable Safety<br>Needle is intended to be used<br>with a luer slip or luer lock<br>syringe for aspiration and<br>injection of fluids for medical<br>purpose. After withdrawal of<br>the needle from the body, the<br>attached needle safety shield<br>can be manually activated to<br>cover the needle immediately<br>after use to minimize risk of<br>accidental needlestick. | The Sterile Disposable Safety<br>Needle is intended to be used<br>with a luer slip or luer lock<br>syringe for aspiration and<br>injection of fluids for medical<br>purpose. After withdrawal of<br>the needle from the body, the<br>attached needle safety shield<br>can be manually activated to<br>cover the needle immediately<br>after use to minimize risk of<br>accidental needlestick. | Same |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| Configuration | | Needle hub | Needle hub | Same |
| | | Needle tube | Needle tube | |
| | | Needle cap | Needle cap | |
| | | Safety machine | Safety sheath | |
| Operation Mode | | For manual use only | For manual use only | Same |
| Sterilized | | Yes | Yes | Same |
| Single Use | | Single Use | Single Use | Same |
| Label/Labeling | | Complied with 21 CFR part<br>801 | Complied with 21 CFR part<br>801 | Same |
| | | | | |
| Needle | Size | 18G, 20G, 21G, 22G, 25G,<br>27G | 16G, 18G, 19G, 20G, 21G,<br>22G, 23G, 24G, 25G, 26G,<br>27G, 28G, 29G, 30G | Different |
| | Length | 1/2", 5/8", 3/4", 1", 1-1/4",<br>1-1/2" | 5/16", 1/2", 5/8", 3/4", 1",<br>1-1/4", 1-1/2" | |
| Needle Performance | | Complied with ISO 7864,<br>ISO 9626 | Complied with ISO 7864,<br>ISO 9626 | Same |
| Patient-contact Materials | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.