K253100 · Asclepion Laser Technologies GmbH · GEX · Oct 23, 2025 · General, Plastic Surgery
Device Facts
Record ID
K253100
Device Name
MultiPulse TFL
Applicant
Asclepion Laser Technologies GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 23, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MultiPulse TFL Laser system and its fibre optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy.
Device Story
MultiPulse TFL is a Class IV laser system (1940 nm wavelength) for soft tissue surgery. It delivers up to 200W power via optical fibers in CW or pulsed modes. Operated by a physician via touchscreen and footswitch, the system includes an optical bench, power supply, and water cooling. It functions by delivering laser energy to target tissue to perform incision, excision, resection, ablation, vaporization, coagulation, or hemostasis. The device is used in clinical settings (OR, endoscopic suites) to treat conditions ranging from urological calculi and BPH to ENT and gynecological lesions. Output is controlled by the surgeon to achieve precise tissue effects, facilitating minimally invasive procedures and potentially reducing patient recovery time compared to traditional surgery.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 62304, IEC 62366-1, and ISO 14971. Software verification and validation conducted per FDA guidance. Biocompatibility established via predicate devices.
Technological Characteristics
Class IV laser; 1940 nm wavelength; 200W max power; CW or pulsed (up to 12 ms) emission. Delivery via single-use/reusable optical fibers. Includes power supply, water cooling, and control electronics. Interface: touchscreen and footswitch. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22.
Indications for Use
Indicated for surgical soft tissue procedures (incision, excision, resection, ablation, vaporization, coagulation, hemostasis) in Urology, Gastroenterology, Thoracic/Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery, and Arthroscopy. Includes specific applications for urological calculi, BPH, spinal surgery, and various gastrointestinal/ENT/gynecological lesions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 23, 2025
Asclepion Laser Technologies GmbH
Francesco Dell'antonio
RA Consultant
Bruesseler Strasse 10
Jena, 07745
Germany
Re: K253100
Trade/Device Name: MultiPulse TFL
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: September 8, 2025
Received: September 24, 2025
Dear Francesco Dell'antonio:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
2K253100 - Francesco Dell'antonio
Page
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
3K253100 - Francesco Dell'antonio
Page
Sincerely,
YAN FU -S
Digitally signed by YAN FU
Date: 2025.10.23 10:33:09
-04'00'
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K253100
Device Name
MultiPulse TFL
Indications for Use (Describe)
MultiPulse TFL Laser system and its fibre optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy.
Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Urethral Strictures
- Bladder Neck Incisions (BNI)
- Ablation and resection of Bladder Tumors, Urethral Tumors and Ureteral Tumors Ablation of Benign Prostatic Hypertrophy (BPH)
- Transurethral incision of the prostate (TUIP)
- Laser Resection of the Prostate Laser Enucleation of the Prostate
- Laser Ablation of the Prostate
Condyloma
Lesions of external genitalia
- Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- Treatment of distal impacted fragments remaining in the ureters following lithotripsy
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of renal calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Appendectomy, Polyps, Biopsy, Gall Bladder calculi, Biliary/Bile duct calculi, Ulcers, Gastric ulcers, Duodenal ulcers, Non Bleeding Ulcers, Pancreatitis, Hemorrhoids, Cholecystectomy, Benign and Malignant Neoplasma, Angiodysplasia, Colorectal cancer, Telangiectasia, Telangiectasia of the Osler-WeberRenu disease, Vascular Malformation, Gastritis, Esophagitis, Esophageal ulcers, Varices, Colitis, Mallory-Weiss tear, Gastric Erosions.
Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) including: Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including: Percutaneous Laser Disc Decompression/Discectomy, Foraminoplasty, Ablation and coagulation of soft vascular and nonvascular tissue in minimally invasive spinal surgery.
Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization,
{4}
coagulation and hemostasis) of soft tissue: Intra-uterine treatment of submucous fibroids, benign endometrial polyps and uterine septum by incision, excision, ablation and/or vessel coagulation, Soft tissue excision procedures such as excisional conization of the cervix.
## ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including: Endonasal/sinus Surgery, Partial turbinectomy, Polypectomy, Dacryocystorhinostomy, Frontal Sinusotomy, Ethmoidectomy, Maxillary antrostomy, Functional endoscopic sinus surgery, Lesions or tumors (oral, nasal, glossal, pharyngeal and laryngeal), Tonsillectomy, Adenoidectomy.
## General surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Cholecystectomy, Lysis of adhesion, Appendectomy, Biopsy, Skin incision, Tissue dissection, Excision of external tumors and lesions, Complete or partial resection of internal organs, tumors and lesions, Mastectomy, Hepatectomy, Pancreatectomy, Splenectomy, Thyroidectomy, Parathyroidectomy, Herniorrhaphy, Tonsillectomy, Lymphadenectomy, Partial Nephrectomy, Pilonidal Cystectomy, Resection of lipoma, Debridement of Decubitus Ulcer, Hemorrhoids, Debridement of Statis Ulcer.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
K253100
Applicant / Management: Asclepion Laser Technologies GmbH
Manufacturer: Asclepion Laser Technologies GmbH
Name and Address: Bruesseler Strasse 10, 07745 Jena - Germany
510(k) Contact Person: Mr. Francesco Dell’Antonio
Regulatory Consultant
E-Mail: francesco.dellantonio@asclepion.com
Date Prepared: 8th September 2025
Trade Name: MultiPulse TFL
Common Name: MultiPulse TFL
Classification: Class II
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Regulation Number: 21 CFR 878.4810
Product Code: GEX
Main Predicate Device: MULTIPULSE TM+1470 (K133891), Asclepion Laser Technologies GmbH
Reference Device: Fiber Dust PRO (K220426), Quanta System Spa CYBER TM 200 (K131081), Quanta System Spa
Performance Standards:
There are no mandatory performance standards for this device.
Description of the device:
The MultiPulse TFL laser system and its fiber optic delivery system is a laser Class IV, operating in CW or pulsed mode at a wavelength of 1940 nm. The laser power up to 200W is transmitted through different optical fibers. Besides of the optical bench the device consists of a power supply, a water cooling unit and a control electronic. The device is operated by a touch screen and a foot switch.
Description of the modifications:
The subject device is a modification to previously cleared MultiPulse TM+1470 due to some technical modifications concerning laser emission parameters and the related controlling software and hardware.
The subject device and the predicate device have the differences shown in the table below:
| | Main Predicate Device | Subject device |
| --- | --- | --- |
| Device name | MULTIPULSE TM+1470 | MultiPulse TFL |
| 510k number | K133891 | - |
{6}
| Wavelength | 1940 um and 1470um | 1940 um |
| --- | --- | --- |
| Power max. | 150W | 200W |
| Pulse Duration | CW or up to 750ms | CW or up to 12 ms |
Additional reference devices were considered in the substantial equivalence discussion, to justify why the differences do not raise any concern about safety and efficacy.
The subject device has the same intended use of the unmodified device. Moreover, the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.
Based on the nature of the changes implemented, the device underwent and successfully passed performance testing and software verifications and validation according to the relevant standards.
## Intended Use:
MultiPulse TFL Laser system and its fibre optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy.
## Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Urethral Strictures
- Bladder Neck Incisions (BNI)
- Ablation and resection of Bladder Tumors, Urethral Tumors and Ureteral Tumors Ablation of Benign Prostatic Hypertrophy (BPH)
- Transurethral incision of the prostate (TUIP)
- Laser Resection of the Prostate Laser Enucleation of the Prostate
- Laser Ablation of the Prostate
- Condyloma
- Lesions of external genitalia
- Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- Treatment of distal impacted fragments remaining in the ureters following lithotripsy
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of renal calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
## Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Appendectomy, Polyps, Biopsy, Gall Bladder calculi, Biliary/Bile
{7}
duct calculi, Ulcers, Gastric ulcers, Duodenal ulcers, Non Bleeding Ulcers, Pancreatitis, Hemorrhoids, Cholecystectomy, Benign and Malignant Neoplasma, Angiodysplasia, Colorectal cancer, Telangiectasia, Telangiectasia of the Osler-WeberRenu disease, Vascular Malformation, Gastritis, Esophagitis, Esophageal ulcers, Varices, Colitis, Mallory-Weiss tear, Gastric Erosions.
## Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) including: Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including: Percutaneous Laser Disc Decompression/Discectomy, Foraminoplasty, Ablation and coagulation of soft vascular and nonvascular tissue in minimally invasive spinal surgery.
## Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue: Intra-uterine treatment of submucous fibroids, benign endometrial polyps and uterine septum by incision, excision, ablation and/or vessel coagulation, Soft tissue excision procedures such as excisional conization of the cervix.
## ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including: Endonasal/sinus Surgery, Partial turbinectomy, Polypectomy, Dacryocystorhinostomy, Frontal Sinusotomy, Ethmoidectomy, Maxillary antrostomy, Functional endoscopic sinus surgery, Lesions or tumors (oral, nasal, glossal, pharyngeal and laryngeal), Tonsillectomy, Adenoidectomy.
## General surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Cholecystectomy, Lysis of adhesion, Appendectomy, Biopsy, Skin incision, Tissue dissection, Excision of external tumors and lesions, Complete or partial resection of internal organs, tumors and lesions, Mastectomy, Hepatectomy, Pancreatectomy, Splenectomy, Thyroidectomy, Parathyroidectomy, Herniorrhaphy, Tonsillectomy, Lymphadenectomy, Partial Nephrectomy, Pilonidal Cystectomy, Resection of lipoma, Debridement of Decubitus Ulcer, Hemorrhoids, Debridement of Statis Ulcer.
## Technological Characteristics Comparison
{8}
| Specification | Main Predicate device | Additional predicate device | Additional predicate device | Subject device |
| --- | --- | --- | --- | --- |
| Trade/Device Name | MULTIPULSE TM+1470 | Fiber Dust PRO | CYBER TM 200 | MultiPulse TFL |
| Submitter | Asclepion Laser Technologies Gmbh | Quanta System Spa | Quanta System Spa | Asclepion Laser Technologies Gmbh |
| 510(k) number | K133891 | K220426 | K131081 | - |
| Wavelength | 1940 um and 1470um | 1940 um | 2010 um | 1940 um |
| Power, max. | 150W | 60W | 200W | 200W |
| Pulse Duration | CW or up to 750ms | CW or up to 15ms | CW or up to 75ms | CW or up to 12 ms |
| Repetition rate, max | 2500 Hz | 2500 Hz | 100 Hz | 2500 Hz |
| Delivery system | Optical fiber (single use and reusable) | Optical fiber (single use and reusable) | Optical fiber (single use and reusable) | Optical fiber (single use and reusable) |
## Non clinical Performance Data:
The following performance data were applied in support of the substantial equivalence determination:
- IEC 60601-1:2005/A1:2012/AMD:2020: Medical electrical equipment – Part 1: General requirements for safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests
- IEC 62304: Medical Device Software – Software life cycle processes
- IEC 62366-1: Medical devices – Application of usability engineering to medical devices
- IEC 60601-2-22:2019 Medical electrical equipment – Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment
- ISO 14971:2019 Medical devices – Application of risk management to medical devices
- Software verification and validation testing were conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions”.
MultiPulse TFL passed all the required testing and is in compliance with all applicable sections of the abovementioned performance standards.
## Biocompatibility:
The biocompatibility of MultiPulse TFL is established based on the predicate devices.
## Comparison with predicate device:
The subject and predicate devices have the same intended use and the same fundamental scientific technology. Any minor difference does not raise concern about safety and effectiveness.
{9}
Conclusions
The differences between the subject and predicate device do not raise new types of questions regarding safety and effectiveness, and the subject device is as safe, as effective, and performs as well as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.