K131081 · Quanta System Spa · GEX · Jul 3, 2013 · General, Plastic Surgery
Device Facts
Record ID
K131081
Device Name
CYBER TM FAMILY
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 3, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cyber Tm Family (that includes Cyber Tm 150, Cyber Tm 150, Cyber Tm 180 and Cyber Tm 200) and its accessories are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy. Note: the Cyber Tm 180 and Cyber Tm 200 are only approved for the treatment of BPH when used at power levels greater than 150W.
Device Story
Cyber Tm Family is a continuous wave (CW) Thulium:YAG laser surgical system (2.01 µm wavelength). Laser energy is delivered via fiber optics to target tissues for incision, excision, ablation, vaporization, coagulation, and hemostasis. Operated by physicians in OR, clinic, or endoscopic settings. Models range from 120W to 200W; Cyber Tm 180 and 200 are specifically indicated for BPH treatment at power levels >150W. Device provides precise thermal tissue effects; comparable to predicate laser systems. Benefits include controlled tissue ablation and hemostasis in various surgical specialties.
Clinical Evidence
Bench testing only. Comparative studies measured depth of thermal injury, coagulation, and ablation depth against unmodified predicate devices. Results showed no significant differences in tissue effects, including carbonization, when maintaining comparable power density via fiber diameter adjustments. No clinical human trials were required or presented.
Technological Characteristics
CW Thulium:YAG laser source; 2.01 µm wavelength; adjustable power 5W–200W. Delivery via fiber optics. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. Software-controlled power output. Standalone surgical laser system.
Indications for Use
Indicated for surgical procedures (incision, excision, ablation, vaporization, coagulation, hemostasis) on soft, mucosal, fatty, and cartilaginous tissue across urology, gastroenterology, thoracic/pulmonary, gynecology, ENT, dermatology, plastic surgery, general surgery, and arthroscopy. Specific applications include BPH treatment (Cyber Tm 180/200 >150W), urethral/bladder tumors, GI polyps/ulcers/neoplasms, gynecological fibroids/polypectomy, ENT sinus/laryngeal procedures, and orthopedic spinal decompression. Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# K131081
# JUL 0 3 2013
# 7. 510(K) SUMMARY - Device Modification
# Introduction:
This document contains the 510(k) Summary for the Cyber Tm Family. The content of this summary is based on the requirements of 21 CFR 807.92(c). This is the first submission for the Cyber Tm Family
| Applicant / | Quanta System SPA |
|-------------------|----------------------|
| Manufacturer | Via IV Novembre, 116 |
| Name and Address: | Solbiate Olona (VA) |
| | Italy, 21058 |
510(k) Contact Person:
Maurizio Bianchi Regulatory Affairs Manager Quanta System SPA
Email: maurizio.bianchi@guantasystem.com Phone: +39 0331 376797 +39 0331 367815 Fax:
Date Prepared: Device Name:
November 20th, 2012 Cyber Tm Family
21 CFR 878.4810
Cyber
Classification: Class II
Classification Name:
Laser surgical instrument for use in general and plastic surgery and in dermatology.
Regulation Number:
Product Code: GEX
Basis for Submission: Technology, Engineering, and Performance Changes
### Legally Marketed Device
The Cyber Tm Family is claimed to be substantially equivalent to these legally marketed devices:
- Cyber Tm 120 of the Cyber Surgical Laser Family (K090962) Quanta System SpA
- Cyber Tm 150 (K102749) Quanta System SpA
- RevoLix Family (K110941) - AllMed System Inc.
#### Performance Standards:
There are no mandatory performance standards for this device.
{1}------------------------------------------------
# General Device Description (unmodified predicate devices vs new modified devices):
The Cyber Tm Family is laser surgical instrument for use in general and plastic surgery and in dermatology (GEX).
In the Cyber Tm Family is installed a single Tm:YAG Laser Source with CW emission at 2.01um with adjustable power from 5 W to the maximum output power (as given below). The laser radiation is delivered to the tissue to be treated through fiber optics.
The Cyber Tm Family (that includes Cyber Tm 150, Cyber Tm 150, Cyber Tm 180 and Cyber Tm 200) and its accessories are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including:
Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy.
Note: the Cyber Tm 180 and Cyber Tm 200 are only approved for the treatment of BPH when used at power levels greater than 150W.
| Models | Type of laser | Wavelength | Min Power | Max Power |
|--------------------------------------------|----------------|------------|-----------|-----------|
| Cyber Tm 120 (unmodified) predicate device | CW Tulium: YAG | 2.010 μm | 5 W | 120 W |
| Cyber Tm 150 (unmodified) predicate device | CW Tulium: YAG | 2.010 μm | 5 W | 150W |
| Cyber Tm 180 (new) modified device | CW Tulium:YAG | 2.010 μm | 5 W | 180 W |
| Cyber Tm 200 (new) modified device | CW Tulium: YAG | 2.010 μm | 5 W | 200 W |
The Cyber Tm Family has the following models (and related main characteristics):
The first two models of the Cyber Tm Family, the Cyber Tm 120 and Cyber Tm 150. have been already cleared and represent the unmodified devices (Legally Marketed Devices). The other two, the Cyber Tm 180 and Cyber Tm 200, are the derived from the previous ones.
All models share the same structure, the same components and the same software. The differentation of the models derives only from the factory set of the maximum output power, as detailed in the next section.
# Description of the modification:
The first two models of the Cyber Tm Family, the Cyber Tm 120 and Cyber Tm 150, have been already cleared and represent the unmodified devices (Legally Marketed Devices). The other two, the Cyber Tm 180 and Cyber Tm 200, are the derived from the previous ones.
The modification of the devices is limited to the Tm:YAG Laser Source that is installed inside Cyber Tm Family.
So all models of the Cyber Tm Family share the same structure, the same components and the same software. The differentation of the models derives only from the factory set of the maximum output power of the laser resonator.
7-2
{2}------------------------------------------------
# Summary of Performance and basic Safety testing:
The Cyber Tm Family (that includes Cyber Tm 120, Cyber Tm 150, Cyber Tm 180 and Cyber Tm 200) and its accessories have been tested in accordance with:
- IEC 60601-1:2005
ﺎ
- IEC 60601-1-2:2005 ı
- IEC 60601-1-4: 1996 (First Ed.) -+ Am.l: 1999 (Consolidated 1.1 Ed.) -
- for use with IEO 60601-1 (1988), Amts 1 (1991) and 2 (1995) -
- । IEC 60601-2-22:2005
- ، IEC 60825-1:2007
The Cyber Tm Family (that includes Cyber Tm 120, Cyber Tm 150, Cyber Tm 180 and Cyber Tm 200) and its accessories comply with European Medical Device 93/42/EEC.
Moreover a specific bench testing was conducted on the proposed Cyber Tm Family devices to support a determination of substantial equivalence to the predicate device. The depth of thermal injury and coaqulation were measured and compared to the unmodified predicate devices and did not show significant differences. The ablation depth was also measured and found to be of comparable depth, keeping the same power density through the use of different fiber diameter. Other tissue effects such as carbonization were also found to be comparable.
#### Indications for Use
The Cyber Tm Family is laser surgical instrument for use in general and plastic surgery and in dermatology (GEX).
In the Cyber Tm Family is installed a single Tm:YAG Laser Source with CW emission at 2010 nm with adjustable power from 5 W to the maximum output power (as qiven below). The laser radiation is delivered to the tissue to be treated through fiber optics.
The Cyber Tm Family (that includes Cyber Tm 150, Cyber Tm 150, Cyber Tm 180 and Cyber Tm 200) and its accessories are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including:
Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy.
Note: The Cyber Tm 180 and Cyber Tm 200 are only approved for the treatment of BPH when used at power levels greater than 150W
The Cyber Tm Family has the following models:
| Models | Type of laser | Wavelength | Min Power | Max Power |
|--------------------------------------------|---------------|-------------------------------------|-----------|-----------|
| Cyber Tm 120 (unmodified) predicate device | CW Tulium:YAG | 2.010 $ \mu $ m | 5 W | 120 W |
| Cyber Tm 150 (unmodified) predicate device | CW Tulium:YAG | 2.010 $ \mu $ m | 5 W | 150W |
| Cyber Tm 180 (new) modified device | CW Tulium:YAG | 2.010 $ \mu $ m | 5 W | 180 W |
| Cyber Tm 200 (new) modified device | CW Tulium:YAG | 2.010 $ \mu $ m | 5 W | 200 W |
The first two models of the Cyber Tm Family, the Cyber Tm 120 and Cyber Tm 150, have been
{3}------------------------------------------------
already cleared and represent the unmodified predicate devices (Legally Marketed Devices). The Cyber Tm 180 and Cyber Tm 200, are the derived from the previous ones. The RevoLix Family (K110941) has been already cleared and is a secondary predicate device (Legally Marketed Devices).
The intended use of the predicate devices (Legally Marketed Devices) are the same as the modified devices.
# Substantial Equivalence:
The first two models of the Cyber Tm Family, the Cyber Tm 120 and Cyber Tm 150, have been already cleared and represent the unmodified predicate devices (Legally Marketed Devices). The Cyber Tm 180 and Cyber Tm 200. are the derived from the previous ones.
The RevoLix Family (K110941) has been already cleared and is a secondary predicate device (Legally Marketed Devices).
All models of the Cyber Tm Family share the same structure, the same components and the same software. The modified devices have the same intended use and the same technological characteristics and principles of operation as the unmodified devices. It has been demonstrated that the modified devices of the Cyber Tm Family are safe and effective as the unmodified devices and the predicate devices.
Thus the modified devices included in the Quanta System Cyber Tm Family are substantially equivalent to the previously legally marketed devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a human figure in motion, composed of three curved lines.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
# July 3, 2013
Quanta System Spa % Mr. Maurizio Bianchi Via IV Novembre Nº 116 21058 Solbiate Olona (VA) Italy
Re: K131081
Trade/Device Name: Cyber Tm Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general surgery and
plastic surgery and in dermatology
Regulatory Class: Class II Product Code: GEX Dated: June 12, 2013 Received: June 13, 2013
Dear Mr. Bianchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
# Page 2 - Mr. Maurizio Bianchi
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
- Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510{k) Number (if known): K131081
Device Name: Cyber Tm Family
# Indications for Use:
The Cyber Tm Family (that Includes Cyber Tm 150, Cyber Tm 180 and Cyber Tm 200) and its accessories are intended for use in surgical procedures using open, laparoscopic incision, resection, abletton, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urclogy, Gastroentarology, Thoracio Gynecslogy, ENT, Demainly, Plastic Surgery, General Surgary and Arthroscopy.
#### Urology:
Open and endoscopic surgery (incision, resection, eblation, vaporization, coagulation and hemostasis) including:
| | Urethral Strictures | |
|--|---------------------|--|
Bladder Neck Incisions (8NI)
Ablation and resection of Bladder Tumors, Uretheral Turnors and Ureteral Turnors.
- Ablation of Benign Prostatic Hypertrophy (BHP),
- Transurethral incision of the prostate (TUIP)
- Laser Resection of the Prostrate
- Laser Enucleation of the Prostate
- Laser Ablation of the Prostate
- Condylomas
- Lesions of external genitalia
Note; The Cyber Tm 180 and Cyber Tm 200 are only approved for the treatment of BPH when used at power levels greater than 150V.
Prescription Use (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) C-ncurrence of CDRH, Office of Device Evaluation (ODE) Neil R Ogden … 2013.06.27 115:35:22 -04'00'
(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number K131081
Pag 1 of 5
6-2
{7}------------------------------------------------
510(k) Number (if known): K131081
Device Name: Cyber Tm Family Indications for Use: continued
Open and endoscopic gastroenterslogy surgery (incision, essecion, ablation, vaporization, coagulation, coagulation and hemostasis) including:
| • | Appendectomy |
|---|-------------------------------------------------|
| • | Polyps |
| • | Biopsy |
| • | Gall Bladder calculi |
| • | Biliary/Bile duct calculi |
| • | Ulcers |
| • | Gastric ulcers |
| • | Duodenal ulcers |
| • | Non Bleeding Ulcers |
| • | Pancreatitas |
| • | Hemorrhoids |
| • | Cholecystectomy |
| • | Benign and Malignant Neoplasm |
| • | Angiodysplasia |
| • | Colorectal cancer |
| • | Telangiectasias |
| • | Telangiectasias of the Osler-Weber-Renu disease |
| • | Vascular Malformation |
| • | Gastritis |
| • | Esophagitis |
| • | Esophageal ulcers |
| • | Varices |
| • | Colitis |
- Gastric Erosions
Prescription Use (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number K131081
Pag_2 of 5
{8}------------------------------------------------
510(k) Number (if known): K131081
Device Name: Cyber Tm Family
# Indications for Use: continued
#### Gynecology:
Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagnation and homostasis):
- Intra-uterine treatment of submucous fibroids
- benian endometrial polype: and uterine septum by incision, ablation and or vessel
- coagulation
- Soft tissue excision procedures such as excisional conization of the cervix
#### ENT:
Endoscopic endonasal surgery (incision, resection, shiption, vaporization, coagulation and hemostasis of soft tissus) including: Endonasal/sinus Surgery
- Partial turbinectomy
- Polypectomy
- Dacryocystor hinostomy
- Frontal Sinusotomy
- Ethmoidectorny
- Maxillary antrostomy
- Functional endoscopic sinus surgery
- Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal
- Tonsillectomy
- Adenoidoctomy
Dermatology and Plastic Surgery:
Incision, excision, resection, abletton, vegatiation and hemostasisof soft, mucosal, fatty and cartlaginous thaus, In therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- Basal Cell Carcinomas
- Lesion of skin and subcutaneous tissue
- Skin tags
- Plantar warls
Prescription Use (Part 21 C.F.R. 801 Subpart D)
#### AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number K131081
Pag 3 of 5
e-4
{9}------------------------------------------------
510(k) Number (if known): K131081
Device Name: Cyber Tm Family
# Indications for Use: continued
#### Gonoral Surgery
Open laparoscopic and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Cholecystectorny
- Lysis of adhesion
- Appendactorny Blopsy
- Skin incision
- Tissue dissection
- Excision of external turnors and lesions
- Complete or partial resection of internal organs, tumors and lesions
- Mastectorny
- Hepatectomy
- Pancreatectomy Splenectomy
- Thyroidectomy
- Parathyroidectomy
- Herniorrhaphy
- Tonsillectorny
- Lymphadenectomy
- Partial Nephrectorny
- Pilonidal Cystectorny
- Resection of lipoma
- Debridement of Decubitus Ulcer Hemorrhoids
- Debridement of Statis Ulcer
Prescription Use (Part 21 C.F.R. 801 Subpart D)
#### AND/OR
# Over-The-Counter Use (21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) for MXM
Division of Surgical Devices
510{k} Number K131081
### Pag 4 of 5
{10}------------------------------------------------
510(k) Number (if known): K131081
Device Name: Cyber Tm Family
#### Indications for Use: continued
#### Thoracic and Pulmonary
Open and endbecopic thoracic and pulmon, excision, excision, resection, ablation, vaportzation, coagulation and hemostasis) of soft tissue
- Laryngeal lesions
- Airway obstructions including carcinoma
- Polyps and granuloma
- Palliation of obstructing carcinoma of the tracheobroncial tree
#### Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and cosgulation of soft and cartileginous (issue);
- Ablation of soft and cartilaginous tissue in Minimal Invasive Spinal Surgery including
- Percutaneous Laser Disc Decompression/Discectorny
- Foreminoplasty
- Ablation and coagulation of soft vascular and non vascular tissue in minimally invesive spinal surgery.
Prescription Use (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number K131081
Pag_5_of_5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.