K102749 · Quanta System Spa · GEX · Dec 10, 2010 · General, Plastic Surgery
Device Facts
Record ID
K102749
Device Name
QUANTA CYBER TM 150W
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 10, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Cyber Tm 150W and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery.and Arthroscopy
Device Story
Cyber Tm 150W is a surgical laser system for soft tissue procedures. Input: electrical power; Output: 2.01μm continuous wave (CW) laser light (5-150W) delivered via optical fibers (Sidefire or bare). System comprises rack-mounted laser source, cooling chillers, power supply, and beam delivery optics. Operated by physicians in OR/clinic settings. Controlled via touch-screen PC on orientable arm. Laser energy facilitates incision, excision, resection, ablation, vaporization, coagulation, and hemostasis. Clinical benefit: precise tissue removal and hemostasis across multiple surgical specialties. Output viewed directly by surgeon during procedure; clinical decision-making based on real-time visualization of tissue interaction.
Clinical Evidence
Bench testing and animal studies performed. Animal study compared 120W vs 150W application on tissue. Results showed no significant difference in lateral coagulation width or axial coagulation depth. Ablation crater depth showed no significant difference at 5 and 10mm/s, with increased depth at 1mm/s for 150W. Carbonization patterns were similar between power levels.
Technological Characteristics
Tm:YAG laser source; 2.01μm wavelength; 5-150W power output. Rack-mounted system with dual chillers, power supply, and optical bench. Delivery via SMA-connected optical fibers (Sidefire 600μm; bare 200-1000μm). Includes 650nm aiming beam. Controlled by touch-screen PC. Safety features: emergency stop, key-switch, interlock connector, circuit breaker. Class II device.
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Quanta System Cyber Surgical Laser Family (K090962)
Submission Summary (Full Text)
{1}------------------------------------------------
| Title: | 510(k) SUMMARY<br>Quanta System Cyber Tm 150W | K102749<br>DEC 10 2010 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Submitter: | Quanta System SpA<br>via IV Novembre,116<br>21058 Solbiate<br>Olona VA / Italy | |
| Contact: | Dr. Isabella Carrer<br>Medical Division Manager | |
| Date Prepared: | October 5, 2009 | |
| Device Trade Name: | Quanta System Cyber Tm 150W | |
| Common Name: | Laser surgical instrument for use in general surgery and<br>dermatology | |
| Classification Name: | Instrument, surgical, powered, laser | |
| Predicate Devices: | - Quanta System Cyber Surgical Laser Family<br>(K090962); | |
| Intended Use / | 2.01μm Applications: | |
| Indications for Use: | The Cyber Tm 150W and its fiber optic delivery system<br>are intended for use in surgical procedures using open,<br>laparoscopic and endoscopic incision, excision, resection,<br>ablation, vaporization, coagulation and hemostasis of<br>soft tissue in use in medical specialties including:<br>Urology, Gastroenterology, Thoracic and Pulmonary,<br>Gynecology, ENT, Dermatology, Plastic Surgery,<br>General Surgery.and Arthroscopy | |
| | <b>Urology</b> | |
| | Open and endoscopic surgery (incision, excision, | |
resection, ablation, vaporization, coagulation and hemostasis) including:
- Urethral Strictures
- -Bladder Neck Incisions (BNI)
{2}------------------------------------------------
- Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral
- Tumors. -
- Ablation of Benign Prostatic -Hypertrophy (BHP),
- Transurethral incision of the prostate (TUIP)
- Laser Resection of the Prostrate -(HoLRP)
- Laser Enuculeation of the Prostate -(HoLEP)
- Laser Ablation of the -Prostate(HoLAP)
- Condylomas a
- Lesions of external genitalia -
## Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy .
- Polyps -
- Biopsy -
- Gall Bladder calculi -
- Biliary/Bile duct calculi -
- -Ulcers
- Gastric ulcers -
- Duodenal ulcers -
- Non Bleeding Ulcers -
- Pancreatitas -
- Hemorrhoids -
- Cholecystectomy -
- Benign and Malignant Neoplasm -
- Angiodysplasia =
- Colorectal cancer -
- Telangiectasias -
- Telangiectasias of the Osler-Weber--Renu disease
- Vascular Malformation -
- Gastritis -
- Esophagitis -
- Esophageal ulcers -
- Varices -
- Colitis
- Mallory-Weiss tear -
- Gastric Erosions
{3}------------------------------------------------
#### Gynecology
Open and laparoscopic gynecological surgery (incision. excision, resection.
ablation, vaporization, coagulation and hemostasis)
- Intra-uterine treatment of submucous fibroids, benign endometrial polyps,
- and uterine septum by incision. excision, ablation and or vessel
- coagulation -
- -Soft tissue excision procedures such as excisional conization of the cervix
### ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation,
vaporization, coagulation and hemostasis of soft tissue) including:
- Endonasal/sinus Surgery -
- Partial turbinectomy -
- Polypectomy -
- Dacryocystorhinostomy -
- Frontal Sinusotomy -
- Ethmoidectomy -
- Maxillary antrostomy -
- Functional endoscopic sinus surgery
- Lesions or tumors of the oral, nasal. -
- glossal, pharyngeal and laryngeal
- Tonsillectomy -
- -Adenoidectomy
#### General Surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Cholecystectomy -
- Lysis of adhesion -
- Appendectomy 1
- Biopsv -
- Skin incision -
- Tissue dissection -
{4}------------------------------------------------
- Excision of external tumors and lesions
- Complete or partial resection of internal organs, tumors and lesions
- Mastectomy -
- Hepatectomy -
- Pancreatectomy -
- Splenectomy -
- Thyroidectomy -
- Parathyroidectomy -
- Herniorrhaphv -
- Tonsillectomy .
- Lymphadenectomy -
- Partial Nephrectomy -
- Pilonidal Cystectomy ﺳ
- Resection of lipoma -
- Debridement of Decubitus Ulcer -
- Hemorrhoids
- Debridement of Statis Ulcer -
- Biopsy -
#### Arthroscopy
-
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)
- Ablation of soft and cartilaginous tissue in Minimal Invasive Spinal
- Surgery including -
- Percutaneous Laser Disc -Decompression/Discectomy
- Foraminoplasty -
- Ablation and coagulation of soft vascular and non vascular tissue in
- minimally invasive spinal surgery.
The Cyber Tm 150W is a surgical laser instrument for use in general surgery and dermatology. The Cyber Tm 150W is:
| Technological |
|------------------|
| Characteristics: |
| Models | Wavelength | Laser Power |
|----------|------------|-------------|
| Cyber Tm | 2.01μm | 150W |
- Cyber Tm is intended for use in surgical procedures using open, laparoscopic and
{5}------------------------------------------------
endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery and Arthroscopy
The device consists of a rack that houses the cooling units (2 chillers), the laser source and the power supply. Above the rack, into a cover-proof light, the laser head and optical bench with the beam delivery optics and the power meter are housed.
The Cyber Tm includes a Tm:YAG Laser Source that emit a Cw laser light at 2.01pm wavelength with power adjustable from 5 to 150W.
An SMA connector allows the connection of an optical fiber in which the main beam and the 650nm aiming beam are launched. Sidefire fibers (600pm) and bare fibers (200, 365, 400, 550, 600, 800 and 1000um) are available for both laser source.
The emergency red push button , the key-switch and the operation status led are housed in the front part of the system.
The footswitch connector is housed in the back side of the system. A metal door closes the lower rack.
On the back panel are housed the magneto-thermal switch (circuit braker), the line cable with IEC309 /32A plug and a safety interlock connector. The rear panel also contains the power supply and chiller outlet grids cooling .
The device is controlled by a touch-screen PC mounted on a orientable arm.
Animal tests were performed to demonstrate the Performance Data consistent and predictable ablation and coagulation of different tissues used in the study.
> Quanta System Cyber Tm 150W is as safe and effective as the predicate devices. The Cyber Tm 150W has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Cyber Tm 150W and its predicate devices raise no new issues of safety or effectiveness. From a clinical point of view, comparing 120W vs 150W application on animal tissue it has been showed that:
> > lateral coagulation width and axial coagulation depth do not show any
Substantial Equivalence:
{6}------------------------------------------------
significant difference – therefore the remaining tissue is affected in the same amount
- depth of the ablation crater: no significant differences when velocities of 5 and 10mm/s were used, while a significant increase of ablation depth at 1mm/s at 150W compared to 120W
- carbonisation of Tm:YAG laser power occurred at 150W in the same manner as could be observed at 120W
Thus, the Cyber Tm 150W is substantially equivalent.
{7}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Quanta System SPA % Regulatory Technology Services, LLC Mr. Mark Job 1394 25th Street, NW Buffalo, Minnesota 55313
DEC 1 0 2010
Re: K102749
Trade/Device Name: Quanta System Cyber Tm 150W Regulation Number: 21 CFFR 878.4810 Regulation Name: Lase surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 27, 2010 Received: December 01, 2010
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act (Act to not require approval of the Federal Food, Drug,
You may, therefore, market the device, subject to the secure approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of PMA).
general controls provisions of the Act include recorders of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against mistranding and adulteration. Please note: CDRH does not evaluate information related to contract ing and
warranties. We remind vou: bowever, that design is hat warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 – Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that substantial equivalence detemination does not mean
that FDA has made a determination administers leavith other requirements of the Act or any Federal statutes and regulations administered by other requirements of for comply with all the Act's requirements, including, but not limited to: registration and listing (21)
CFR Part 807); labeling (21 CFR Part 801); modical 10: registration and l CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting and listing);
CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as and forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the set
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, e the election of the of Schild (QS) regulation (21 CFR Part 820); and if applicable, the elect
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 10
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm/11/2/2/2/2
the Center for Devices and Radiological Health CDRH/CDRH/Offices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Offices/ucm 115809.htm for
note the regulation entitled "Misbranding by references of Compliance. Also, please note the regulation entitled than is (CDRH s) Office of Compliance. Also, please
807.97). For questions regarding the reporting of cl 807.97). For question reporting of ference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation ( CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the manner (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
A
B
reta
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
| | Indications for Use Statement | |
|---------------------------|-------------------------------|--|
| 510(k) Number (if known): | K102749 | |
Device Name: Cyber Tm 150W
DEC 1 0 2010
Indications for Use:
## 2.01um Applications:
The Cyber Tm 150W and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Gastroenterology, Thoracic and Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, General Surgery.and Arthroscopy
## Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Urethral Strictures -
- Bladder Neck Incisions (BNI) ー
- Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral
- Tumors. -
- Ablation of Benign Prostatic Hypertrophy (BHP). -
- Transurethral incision of the prostate (TUIP)
- Laser Resection of the Prostrate (HoLRP) -
- Laser Enuculeation of the Prostate (HoLEP) -
- Laser Ablation of the Prostate (HoLAP)
- Condylomas -
- Lesions of external genitalia -
## Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including
- Appendectomy -
- Polyps -
- Biopsy -
- Gall Bladder calculi -
- Biliary/Bile duct calculi -
(Division Sign-Off) (Division Sign-Off)
Division of Surgicel, Orthopedic, Orthopedic,
Devices and retive Devices Division of Surgers
and Restorative Devices and Restorative Devices
510(k) Number K1027
{10}------------------------------------------------
- Ulcers
- Gastric ulcers -
- Duodenal ulcers -
- Non Bleeding Ulcers -
- Pancreatitas -
- Hemorrhoids -
- Cholecystectomy -
- Benign and Malignant Neoplasm -
- Angiodysplasia -
- Colorectal cancer -
- Telangiectasias -
- Telangiectasias of the Osler-Weber-Renu disease -
- Vascular Malformation -
- Gastritis -
- Esophagitis -
- -Esophageal ulcers
- Varices I
- Colitis
- Mallory-Weiss tear
- Gastric Erosions
## Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis)
- Intra-uterine treatment of submucous fibroids, benign . endometrial polyps,
- and uterine septum by incision, excision, ablation and or vessel -
- coagulation
- Soft tissue excision procedures such as excisional conization of the cervix
#### ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including
- Endonasal/sinus Surgery -
- Partial turbinectomy -
- Polypectomy -
tissue) including:
(Division Sign-Off)
(Division Sign of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102749
{11}------------------------------------------------
A.
- Dacrvocystorhinostomy
- Frontal Sinusotomy
- Ethmoidectomy
- Maxillary antrostomy -
- Functional endoscopic sinus surgery -
- Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal
- Tonsillectomy
- Adenoidectomy
## General Surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Cholecystectomy -
- Lysis of adhesion -
- Appendectomy
- Biopsy -
- Skin incision
- Tissue dissection
- Excision of external tumors and lesions
- Complete or partial resection of internal organs, tumors and lesions
- -Mastectomy
- Hepatectomy -
- Pancreatectomy -
- Splenectomy -
- Thyroidectomy ﮯ
- Parathvroidectomy a
- Herniorrhaphy .
- Tonsillectomy -
- Lymphadenectomy -
- Partial Nephrectomy ﮯ
- Pilonidal Cystectomy
- Resection of lipoma
- Debridement of Decubitus Ulcer
- Hemorrhoids
- Debridement of Statis Ulcer
- Biopsy
Mark A. Milburn
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102749
Arthroscopy
20
{12}------------------------------------------------
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)
- Ablation of soft and cartilaginous tissue in Minimal Invasive -Spinal
- Surgery including -
- Percutaneous Laser Disc Decompression/Discectomy -
- -Foraminoplasty
- Ablation and coagulation of soft vascular and non vascular tissue in
- minimally invasive spinal surgery. -
Prescription Use X Use (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter
(21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mula A. Millican
(Division Sign-Off)
Division of Surgical, Orthopedic,
Division of Surgices Division of Surges
and Restorative Devices
K102749
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.