Portable DR Imaging System

K253058 · Shantou Institute of Ultrasonic Instruments Co., Ltd. · IZL · Jun 18, 2026 · Radiology

Device Facts

Record IDK253058
Device NamePortable DR Imaging System
ApplicantShantou Institute of Ultrasonic Instruments Co., Ltd.
Product CodeIZL · Radiology
Decision DateJun 18, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1720
Device ClassClass 2

Indications for Use

The Portable DR Imaging System is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's extremities. This device is not intended for mammography. This device is only intended for adults, not for paediatric application. This device is only suitable for stand-mounted use in hospital shielded rooms.

Device Story

Portable X-ray system for extremity imaging; consists of X-ray unit, flat panel detector, and PIE-5 workstation. System control software (ARM platform, C language) manages high-voltage generation, X-ray tube excitation, collimator, and battery monitoring. PIE-5 workstation (Windows-based) handles patient registration, image acquisition, processing, review, diagnosis, and PACS transmission. Operated by qualified physicians or technicians in hospital shielded rooms. Output consists of digital radiographic images used for clinical diagnosis. Benefits include portable, high-quality extremity imaging for adult patients.

Clinical Evidence

Comparative clinical image quality assessment performed. Images of extremities acquired using both subject and predicate devices were evaluated by an ABR Board-Certified Radiologist. Results demonstrated substantial equivalence in image quality and clinical utility for diagnostic requirements.

Technological Characteristics

Portable X-ray system; 150-560W generator; 40-90kV range. Detector: TFT-amorphous Silicon, 150µm pixel pitch, 14"x17" max size. Connectivity: Wired/Wireless IEEE 802.11a/g/n (2.4/5GHz) with WEP/WPA/WPA2. Software: Embedded system control (ARM/C) and PIE-5 workstation (Windows/C). Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, ISO 10993, and IEC 62304.

Indications for Use

Indicated for adult patients requiring X-ray imaging of extremities. Contraindicated for pediatric patients and mammography. Must be used in hospital shielded rooms with stand-mounted configuration.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** **U.S. FOOD & DRUG** ADMINISTRATION June 18, 2026 Shantou Institute of Ultrasonic Instruments Co., Ltd. % Guang Chen Primary Correspondent #77, Jinsha Road Shantou, CN 515041 CHINA Re: K253058 Trade/Device Name: Portable DR Imaging System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL, MQB, LLZ Dated: May 20, 2026 Received: September 22, 2025 Dear Guang Chen: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253058 - Guang Chen Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253058 - Guang Chen Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Lu Jiang, Ph.D. Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. # Indications for Use 510(k) Number (if known) K253058 Device Name Portable DR Imaging System Indications for Use (Describe) The Portable DR Imaging System is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's extremities. This device is not intended for mammography. This device is only intended for adults, not for paediatric application. This device is only suitable for stand-mounted use in hospital shielded rooms. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} SIUI Portable DR Imaging System--- # 510(k) Summary This summary of 510(k) safety and effectiveness information is provided in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c). The assigned 510(k) number is: K253058 ## 1. Submitter Shantou Institute of Ultrasonic Instruments Co., Ltd. (SIUI) 77 Jinsha Road, Shantou, Guangdong 515041, China Tel: 86-754-88250151 Fax: 86-754-86603664 Contact Person: Chen Guang Shantou Institute of Ultrasonic Instruments Co., Ltd. 77 Jinsha Road, Shantou, Guangdong 515041, China Date Prepared: June 17, 2026 ## 2. Subject Device Name of Device: Portable DR Imaging System 510(k) submitter: SIUI Regulation Name: Mobile x-ray system (21 CFR 892.1720) Regulatory Class: II Product Code: IZL, MQB, LLZ ## 3. Predicate Device Name of Device: SR-2300 Portable DR Imaging System (K202353) 510(k) holder: SIUI Regulation Name: Mobile x-ray system (21 CFR 892.1720) Regulatory Class: II 1/7 {5} SIUI Portable DR Imaging System--- Product Code: IZL, MQB, LLZ Classification Name: Mobile x-ray system IZL (per 21 CFR 892.1720) Stationary x-ray system MQB (per 21 CFR 892.1680) Medical image management and processing system LLZ (per 21 CFR 892.2050) #### 4. Device Description The Portable DR Imaging System is a portable digital device developed, designed and manufactured by SIUI. The device consists of the following major components: a portable X-ray unit, a flat panel detector, and workstation. This system is a Track 3 device. The Portable DR Imaging System contains two pieces of software, one is the embedded software in the X-ray unit: system control software, and the other is PIE-5 DR Workstation (hereinafter referred to as “PIE-5”). The system control software of the Portable DR Imaging System is software designed by SIUI. The software adopts development languages of Keil, the program language is C language, and the hardware platform is ARM. The system control software is for real-time interaction and control with various circuit modules inside the Portable DR Imaging System. The software responds to user operations on the control panel. The user can adjust and control the kV and mAs parameters, and the software will display the parameters or directly load the APR parameters. The software loads the control data from X-ray output into the high-voltage generation control circuit of the system control board, control the high-voltage tank to generate high-voltage and to excite the X-ray tube inside to emit X-rays, control the switch of the collimator indicator, monitor the working status of the device and the battery power status, and control the display of the status indicators. PIE-5 uses the C programming language, and the operating platform is Microsoft Windows. The software is a medical image processing system, consisting of image acquisition, image review and diagnosis, image processing and digital storage functions, which mainly address digital image workflow in radiographic department of hospitals. The functions of the software include patient registration, image acquisition and processing, image review and diagnosis, image printing, report editing and PACS transmission, digital image storage and recovery. 2/7 {6} SIUI Portable DR Imaging System # **5. Comparisons of SR-300Pro, SR-200Pro, SR-100Pro, SR-300, SR-200, SR-100, SR-2300Pro, SR-1100Pro, SR-1000Pro, SR-2300, SR-2000, SR-1100, and SR-1000 devices** The Portable DR Imaging System includes the models SR-300Pro /SR-2300 / SR-1000 / SR-2000 / SR-2300Pro / SR-1000Pro / SR-1100 /SR-1100Pro / SR-300 / SR-200 / SR-100 /SR-200Pro /SR-100Pro. These models used the same safe critical components, with the same circuit modules and working principles, and the same appearance. Except for the differences in functional configuration, these models are completely the same. The mechanical structure is designed for market requirements of the devices according to international general safety standard (IEC 60601-1). The hardware design is identical to the following SR series models: SR-300Pro, SR-200Pro, SR-100Pro, SR-300, SR-200, SR-100, SR-2300Pro, SR-1100Pro, SR-1000Pro, SR-2300, SR-2000, SR-1100, and SR-1000. A list of the differences between various models intended for sale in the market is shown below, which are identical except for the differences in settings: Table 1 | Model | DICOM Setting | Report Setting | Process Setting | Custom Setting | | --- | --- | --- | --- | --- | | SR-300Pro | Standard | Standard | Standard | Standard | | SR-300 | Standard | Standard | Standard | Optional | | SR-200 | Standard | Standard | Optional | Standard | | SR-100 | Standard | Optional | Standard | Standard | | SR-200Pro | Optional | Standard | Standard | Standard | | SR-100Pro | Standard | Standard | Optional | Optional | | SR-2300 | Standard | Optional | Standard | Optional | | SR-1000 | Standard | Optional | Optional | Standard | | SR-2000 | Optional | Standard | Standard | Optional | | SR-2300Pro | Optional | Standard | Optional | Standard | | SR-1000Pro | Optional | Optional | Standard | Standard | | SR-1100 | Standard | Optional | Optional | Optional | | SR-1100Pro | Optional | Standard | Optional | Optional | 3/7 {7} SIUI Portable DR Imaging System ## 6. Indications for Use The Portable DR Imaging System is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient’s extremities. This device is not intended for mammography. This device is only intended for adults, not for pediatric application. This device is only suitable for stand-mounted use in hospital shielded rooms. ## 7. Comparison of Technological Characteristics with the Predicate Device The comparison between the overall specifications of the predicate device (K202353) and the subject device is shown in Table 2. Any differences between the predicate and the subject device have no impact on safety or efficacy of the subject device and do not raise any new potential or increased safety risks, and the subject device is equivalent in performance to existing legally marketed devices. ### Subject Device : Portable DR Imaging System (K253058) #### Predicate Device: SIUI SR-2300 Portable DR Imaging System (K202353) | Description | | Subject Device : Portable DR Imaging System (K253058) | Predicate Device: SIUI SR-2300 Portable DR Imaging System (K202353) | | --- | --- | --- | --- | | **Indications for use** | | The Portable DR Imaging System is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient’s extremities. This device is not intended for mammography. This device is only intended for adults, not for pediatric application. This device is only suitable for stand-mounted use in hospital shielded rooms. | Intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient’s anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities). This device is not intended for mammography. | | **Basic Genera-** | **Peak generator** | 150-560W | 5kW | 4/7 {8} SIUI Portable DR Imaging System | **tor Charac- teristics** | **Tube current** | 1.6-8mA | 10-100mA | | --- | --- | --- | --- | | | **Tube voltage adjustable range** | 40-90kV, step value 1kV | 40-125kV, step value 1kV | | | **mAs range** | 0.16mAs-8mAs | 0.4mAs-200mAs | | **Collimator** | | Built in | Built in | | **X-ray Generator** | | One model | One model | | **Operator console** | | Touch Control or Touch Screen | Button Control or Touch Screen | | **Digital X-Ray Detectors** | | SFD-1X/ SFD-2X/ SFD-3X | SFD-1X | | **Panel Shape** | | Rectangular Panel | Rectangular Panel | | **Detector Size** | | Max: 14” X 17” | 14” X 17” | | **Pixel Pitch** | | 150µm | 150µm | | **Materials Scintillator** | | TFT-amorphous Silicon | TFT-amorphous Silicon | | **Communication Method** | | Wire Wireless IEEE 802.11a//g/n(2.4GHz/ 5GHz) Security: WEP/WPA/WPA2 | Wire Wireless IEEE 802.11a//g/n(2.4GHz/ 5GHz) Security: WEP/WPA/WPA2 | | **Acquisition Software** | | PIE-5 | PIE-5 | | **Software Function** | | Image viewing Image search Image storage Image annotation Image measurement Image processing Image stitch | Image viewing Image search Image storage Image annotation Image measurement Image processing Image stitch | | **DICOM 3.0 Compatibility** | | Yes | Yes | | **Power Source** | | AC Line or rechargeable batteries (Generator only) | AC Line or rechargeable batteries | ## 8. Non-clinical Testing Summary The Portable DR Imaging System complies with and/or were tested in accordance with the following FDA guidance and International Standards: - ● IEC 60601-1:2005+ AMD1:2012+AMD2:2020 Medical electrical equipment - 5/7 {9} SIUI Portable DR Imaging System--- Part 1: General requirements for basic safety and essential performance - ● IEC 60601-1-2:2014+AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests. - ● IEC 60601-1-3:2021 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment - ● IEC 60601-2-54:2022 Medical electrical equipment - Part 2-54:Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy - ● ISO 14971:2019 Medical devices - Application of risk management to medical devices - ● ISO 10993-1:2018 Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process - ● ISO 10993-5:2009 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicity - ● ISO 10993-10:2021 Biological evaluation of medical devices - Part 10:Tests for skin sensitization - ● ISO 10993-23:2021 Biological evaluation of medical devices - Part 23:Tests for irritation - ● IEC 62304:2006+A1:2015 Medical device software – Software life cycle processes - ● IEC 60601-1-6:2010+ AMD1:2013+AMD2:2020 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability - ● IEC 62366-1:2015+AMD1:2020 Medical devices – Part 1: Application of usability engineering to medical devices - ● ISO 15223-1:2021 Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements - ● ISO 13485:2016 Medical devices - Quality management systems - Requirements for regulatory purposes - ● Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - ● Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices 6/7 {10} SIUI Portable DR Imaging System - Content of Premarket Submissions for Device Software Functions - Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions - Electromagnetic Compatibility (EMC) of Medical Devices The subject device and the predicate device are comparable in terms of technical features, general functions, applications and intended uses. The test results showed compliance with the above standards. Non-clinical Detector Performance testing as prescribed in the standard of IEC 60601-2-54: 2022. 9. Clinical Testing SIUI performed Image Quality and Performance Testing by acquiring images of the extremities using both the Portable DR Imaging System (subject device) and the predicate device. The images were assessed for image quality and clinical utility by an ABR Board-Certified Radiologist. Comparative analysis demonstrated that the image quality of the subject device was substantially equivalent to that of the predicate device. The system exhibited sufficient clinical utility and fulfilled diagnostic requirements for extremities. 10. Conclusion The subject device Portable DR Imaging System and the predicate device SIUI SR-2300 Portable DR Imaging System (K202353) are comparable in terms of technical features, general functions, applications and indications for use. 7/7
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...