K252835 · Guangdong Prestige Technology Co., Ltd. · INI · Dec 23, 2025 · Physical Medicine
Device Facts
Record ID
K252835
Device Name
Electric Scooter (S202572J, S2060, M2089, S2050)
Applicant
Guangdong Prestige Technology Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Dec 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
It is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Electric scooter (models S202572J, S2060, M2089, S2050) provides indoor mobility for disabled/elderly users. Battery-powered (25.2V Li-ion) rear-wheel drive vehicle; features rigid frame (carbon fiber or aluminum alloy), seat, control panel, electric motor, and electromagnetic braking. User operates via acceleration push rod on control panel to move forward/backward; releasing rod engages electromagnetic brake. Models include anti-tip wheels. S202572J/S2060 include NFC unlocking. Device status displayed on screen. Used in indoor environments; operated by patient. Benefits include improved mobility for individuals with limited walking ability.
Clinical Evidence
No clinical study included. Safety and performance established via bench testing, including biocompatibility, EMC (IEC 60601-1-2, ISO 7176-21, IEEE ANSI C63.18-2014), and performance testing per ISO 7176 series (stability, braking, dimensions, speed, obstacle climbing, power/control systems). Resistance to ignition tested per ISO 16840.
Indicated for disabled or elderly persons limited to a seated position requiring indoor mobility assistance.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 23, 2025
Guangdong Prestige Technology Co., Ltd.
% Cassie Lee
Official Correspondent
Share Info (Guangzhou) Medical Consultant Ltd.
No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road,
Huangpu District
Guangzhou, Guangdong 510000
China
Re: K252835
Trade/Device Name: Electric Scooter (S202572J, S2060, M2089, S2050)
Regulation Number: 21 CFR 890.3800
Regulation Name: Motorized Three-Wheeled Vehicle
Regulatory Class: Class II
Product Code: INI
Dated: December 1, 2025
Received: December 1, 2025
Dear Cassie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252835 - Cassie Lee
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252835 - Cassie Lee
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252835 | ? |
| Please provide the device trade name(s). | | ? |
| Electric Scooter (S202572J, S2060, M2089, S2050) | | |
| Please provide your Indications for Use below. | | ? |
| It is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary for K252835
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 890.92.
1. Submitter's Information
Company Name: Guangdong Prestige Technology Co., Ltd.
Address: No.8, Road.1, 3rd Ind. District, Houjie Qiaotou, Houjie Town, Dongguan City, Guangdong Province, China
Establishment Registration Number: 3008841035
Postal Code: 523950
Phone: +86 0769-85922501
Fax: +86 0769-85922505
Contact Person (including title): Zhang Zhao (General Manager)
E-mail: leon@solaxtech.com
Application Correspondent:
Contact Person: Ms. Cassie Lee
Company: Share Info (Guangzhou) Medical Consultant Ltd.
Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China
Tel: +86 20 8200 6973
Email: 382198657@qq.com
2. Subject Device Information
Type of 510(k) Submission: Traditional
Common Name: Electric Scooter
Trade Name: Electric Scooter
Model: S202572J, S2060, M2089, S2050
Classification Name: Vehicle, Motorized 3-Wheeled
Regulation Description: Motorized three-wheeled vehicle.
Review Panel: Physical Medicine
Product Code: INI
Regulation Class: 2
Regulation Number: 21 CFR 890.3800
3. Predicate Device Information
Predicate Device
510k number: K190737
Trade Name: Solax Electric Scooter
Model: S204311M, S204161, S204143
{5}
Classification name: Vehicle, Motorized 3-Wheeled
Regulation Description: Motorized three-wheeled vehicle.
Review Panel: Physical Medicine
Product Code: INI
Regulation Class: 2
Regulation Number: 21 CFR 890.3800
## 4. Device Description
The Electric Scooter (model: S202572J, S2060, M2089, S2050) is an indoor electric scooter that is battery operated.
The Electric Scooter (Model: S202572J, S2060, M2089, S2050) is with 125 kg weight capacity. It is basic conventional rear wheels drive, rigid frame vehicle that are battery powered. It consists primarily of a foldable frames (S202572J: carbon fiber frame, S2060, M2089, S2050: aluminum alloy frame), a sealed trans-axle motors drive system, electromagnetic braking system, electric motor controller and a Li-ion batteries with an off-board battery charger. It is powered by 25.2 V/10 AH rechargeable Lithium batteries with 15 km which maximum speed up-to 6 km/hr. The direction control lever to control the direction of the Electric Scooter.
For model S202572J, when pushing the right acceleration push rod forward with the right finger, the electric scooter moves forward, when pushing the left acceleration push rod forward with the left finger, the electric scooter moves backward, the electromagnetic brake will be actuated when the acceleration push rod is released and the electric scooter will stop. It consists of four parts which are chair part, control part, folding part and drive part. Overall, it mainly has a basic carbon fiber frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, a control panel, an electric motor, an electromagnetic brake, two armrests and a Lithium battery with an off-board charger. The control panel has eight buttons, the acceleration push rod, a display screen showing the device status and NFC unlocking sensing area.
For model S2060, when pushing the right acceleration push rod backward with the right finger, the electric scooter moves forward, when pushing the left acceleration push rod backward with the left finger, the electric scooter moves backward, the electromagnetic brake will be actuated when the acceleration push rod is released and the electric scooter will stop. It consists of four parts which are chair part, control part, folding part and drive part. Overall, it mainly has a basic aluminum alloy frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, a control panel, an electric motor, an electromagnetic brake and a Lithium battery with an off-board charger. The control panel has five buttons, the acceleration push rod, a display screen showing the device status and NFC unlocking sensing area.
For model M2089, when pushing the right acceleration push rod forward with the right finger, the electric scooter moves forward, when pushing the left acceleration push rod forward with the left finger, the electric scooter moves backward, the electromagnetic brake will be actuated when the acceleration push rod is released and the electric scooter will stop. It consists of four parts which are chair part, control part, folding part and drive part. Overall, it mainly has a basic aluminum alloy frame, two front wheels, two rear wheels, four backline anti-tip wheels, a seat, a control panel, an electric motor, an electromagnetic brake, two armrests and a Lithium battery with an off-board charger. The control panel has five buttons, the acceleration push rod, a display screen showing the device status.
For model S2050, when pushing the right acceleration push rod forward with the right finger, the electric scooter moves forward, when pushing the left acceleration push rod forward with the left finger, the electric scooter moves backward, the electromagnetic brake will be actuated when the acceleration push rod is released and the electric scooter will stop. It consists of four parts which are chair part, control part, folding part and drive part. Overall, it mainly has a basic aluminum alloy frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, a control panel, an electric motor, an electromagnetic brake and a Lithium battery with an off-board charger. The control panel has a battery level indication area, a speed control knob, a horn button, a power control keyhole and the acceleration push rod.
{6}
The model S202572J, S2060, M2089, S2050 are identical in circuit structure and key components (such as battery, charger, controller, motor) except for some differences shown in the appearance and function.
About appearance of the four models, the model S202572J and model S2060 are different in the material of the frame, the control panel, the size of tires, the overall size and the backrest shape. The model S2050 and model S2060 are different in the shape of seat bracket, the control panel and the backrest shape. The model M2089 and model S2060 are different in the shape of seat bracket, the pedal, the overall size and the backrest shape. The model S2060, S2050 have no armrest, while the model S202572J, M2089 have two armrests.
About function of the four models, the folding method of model M2089 are different from the model S202572J, S2060, S2050. The tiller height of model S202572J, S2060, S2050 is adjustable, while the tiller height of model M2089 is non-adjustable.
## 5. Intended Use / Indications for Use
It is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
## 6. Tests Summary
### 6.1 Summary of Non-Clinical Tests
Electric Scooter (Model: S202572J, S2060, M2089, S2050) has been evaluated the safety and performance by lab bench testing as following:
- Biocompatibility Testing
The patient-contacting materials of Electric Scooter (model: S202572J, S2060, M2089, S2050) have the same nature of tissue contact and contact duration as the Mobile scooter (K122749).
- Electromagnetic compatibility (EMC)
EMC testing was conducted on the proposed device. The system was shown to comply with IEC 60601-1-2:2020, ISO 7176-21, and IEEE ANSI C63.18-2014 for electromagnetic compatibility.
- Performance testing
Performance test and functional tests were conducted on the proposed device in accordance with product design specifications. Data generated from the test met the predetermined acceptance criteria. The following includes:
The device complied with requirements for ISO 7176-1 static stability; ISO 7176-2 dynamic stability; ISO 7176-3 effectiveness of brakes; ISO 7176-4 theoretical distance rang; ISO 7176-5 dimensions, mass and manoeuvring space; ISO 7176-6 maximum speed, acceleration and deceleration; ISO 7176-7 static dimensions of the device; ISO 7176-8 static, impact and fatigue strengths; ISO 7176-9 climatic tests; ISO 7176-10 obstacle-climbing ability; ISO 7176-11 dummies; ISO 7176-13 coefficient of friction of test surfaces; ISO 7176-14 power and control systems; ISO 7176-15 information disclosure, ISO 7176-21 documentation and labelling; ISO 7176-22 set-up procedures.
- Resistance to Ignition Testing
The seat cushion, backrest and armrests of this device were tested for resistance to ignition, and the test results all met the criteria requirements of ISO 16840.
- Software Verification and Validation
The Software verification and validation is in compliance with FDA Guidance-Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
{7}
> Cybersecurity
The subject device no any external interfaces, according to FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", no need cybersecurity evaluation.
# 6.2 Summary of Clinical Performance Test
No clinical study is included in this submission.
# 7. Comparison to predicate device and conclusion
Compare with the predicate devices:
| Elements of Comparison | Subject Device | Predicate Device | Remark |
| --- | --- | --- | --- |
| Company | Guangdong Prestige Technology Co., Ltd. | Dongguan Prestige Technology Co., Ltd. | -- |
| Trade Name | Electric Scooter | Solax Electric Scooter | -- |
| Model | S202572J, S2060, M2089, S2050 | S204311M, S204161, S204143 | -- |
| Classification Name | Vehicle, Motorized 3- Wheeled | Vehicle, Motorized 3- Wheeled | Same |
| 510K Number | K252835 | K190737 | -- |
| Common or Usual name | Electric Scooter | Solax Electric Scooter | |
| Product Code | INI | INI | Same |
| Regulation Number | 21 CFR 890.3800 | 21 CFR 890.3800 | Same |
| Regulation Description | Motorized three-wheeled vehicle. | Motorized three-wheeled vehicle. | Same |
| Intended use | It is a motor driven, indoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | Similar. The difference between the subject device and the predicate device is that the subject device can only be used indoors, while the predicate device can be used indoors and outdoors. Therefore, the expected use of the predicate device includes the content of the subject device. So, the difference will not raise any safety or effectiveness issue. |
| Use condition | Indoor use | Indoor and outdoor use | Similar. The predicate device contains the conditions of use of the subject device and is within the scope of the predicate device, so it does not affect the safety and effectiveness of the device. |
| Intended user | Disabled or elderly person limited to a seated position. | Disabled or elderly person limited to a seated position | Same |
{8}
| Maximum capacity | 125 kg | 125 kg Approx. | Same |
| --- | --- | --- | --- |
| Remote control | No | No | Same |
| Drive system | PG 45A / Rear wheel drive | PG 45A / Rear wheel drive | Same |
| Brake system | Electromagnetic | Electromagnetic | Same |
| Motor | 24 V/150 W | 24V 120W | Similar. Although the “motor” of the subject device is slightly different from that of the predicate device, they both comply with EMC standards and ISO 7176-14 standards, so this difference does not affect the safety and effectiveness. |
| Battery | Lithium Battery 25.2 V, 10 Ah | Lithium battery 24V/10AH | Similar. Although the "Battery" is slightly difference between the subject device and predicate device, it is complied with the standards IEC 62133-2, ISO 7176-4, and ISO 7176-14, it will not affect the safety and effectiveness. |
| Battery weight | 1.7 kg | 1.84 kg | Similar. Although the "Battery weight" is slightly difference between the subject device and predicate device, it is complied with the standards IEC 62133-2 ISO7176-5, and ISO 7176-14, the difference will not affect the safety and effectiveness. |
| Battery charger | Off-board charger 29.4 VDC, 2 A | DC24V/2A | Similar. Although the "Battery charger " is slightly difference between the subject device and predicate device, it is complied with the standards IEC 60601-1, it will not affect the safety and effectiveness. |
| Frame design | Rigid frame | Rigid frame | Same |
| Frame design - Material | S202572J: Carbon fiber S2060, M2089, S2050: Aluminum alloy | Aluminum alloy | Different. The frame material for model S2060, M2089 and S2050 is consistent with that of the predicate device. The frame material for model S202572J is different |
{9}
| | | | from that of the predicate device, but both comply with the requirements of the ISO 7176 series standards and have the same durability level as the predicate device. Therefore, the use of different frame materials does not raise new safety and efficacy issues. |
| --- | --- | --- | --- |
| Folding mechanism | Manually fold or unfold | Manual folding for S204311M, S204161,S204143 | Same |
| Tires | S202572J: 7 inches for front wheel (solid tire) 7 inches for rear wheel (solid tire) 66.67 mm for anti-tip wheel S2060, M2089, S2050: 8 inches for front wheel (solid tire) 8 inches for rear wheel (solid tire) 66.67 mm for anti-tip wheel | 6 inches for front wheel (solid wheel) 7 inches for rear wheel (solid wheel) | Similar. Although the “Tires” is difference between the subject device and predicate device, it is complied with the standard ISO 7176 series, the difference will not affect the safety and effectiveness. |
| Base weight (not including battery) | S202572J: 13.3 kg S2060: 18.2 kg M2089: 14.3 kg S2050: 18.2 kg | S204311M: 24 kg S204161, S204143: 24.7 kg | Different. Although the “Base weight” is difference between the subject device and predicate device, the subject device is lighter than the predicate device and offers better portability. Furthermore, the subject device complies with the requirements of the standard ISO 7176-5. So, the difference in this parameter does not affect the safety and effectiveness. |
| Overall dimension (L × W × H) | S202572J: 955 x 520 x 950 mm S2060: 1030 x 460 x 968 mm M2089: 1008 mm x 545 mm x 935 mm S2050: 1030 mm x 460 mm x 968 mm | S204311M: 980 x 450 x 940 mm S204161: 930 x 450 x 860 mm S204143: 980 x 450 x 880 mm | Different. These parameters differ between the subject device and the predicate device, but the subject device complies with the requirements of standard ISO 7176-5, and. In addition, minor variations in physical characteristics do not affect critical functions or normal use, so these parameter |
| Turning Diameter | S202572J: 2960 mm S2060, S2050: 2800 mm M2089: 2730 mm | S204311M: 1.55 m (with a speed less than 3 km/h is recommend) S204161, S204143: 1.35 m (with a speed less than | |
{10}
| | | 3 km/h is recommend) | differences do not affect the safety and effectiveness. |
| --- | --- | --- | --- |
| Ground clearance | S202572J: 40 mm
S2060, S2050: 50 mm
M2089: 65 mm | S204311M: 38 mm
S204161, S204143: 58 mm | |
| Cruising Range | 15 km (9.32 mile) | 15 km (9.32 mile) | Same |
| Max Speed forward | 6 km/h (3.7mph) | 6 km/h (3.7mph) | Same |
| Maximum obstacle climbing | 25 mm | S204311M: 40 mm
S204161, S204143: 60 mm | Different.
Although this parameter is difference between the subject device and predicate device, it is complied with the standard ISO 7176-10.
The difference in this parameter does not affect the safety and effectiveness. |
| Maximum safe operational incline degree | 12° | S204311M: 0-13°
S204161, S204143: 0-15° | Similar.
This parameter is slightly different but similar to that of the predicate device, and it complies with the standards ISO 7176-1, ISO 7176-2 and ISO 7176-3, so this difference does not affect the safety and effectiveness. |
**Final Conclusion:**
The subject device is as safe, as effective, and performs as well as the legally marketed predicated devices K190737.
8. Date of the summary prepared: December 19, 2025
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.