Solax Electric Scooter (Models: S204311M, S204161, S204143)
K190737 · Dongguan Prestige Sporting Goods Co., Ltd. · INI · May 10, 2021 · Physical Medicine
Device Facts
Record ID
K190737
Device Name
Solax Electric Scooter (Models: S204311M, S204161, S204143)
Applicant
Dongguan Prestige Sporting Goods Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
May 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Solax Electric Scooter is a motor-driven, three-wheeled mobility vehicle for disabled or elderly users. Input: user operation via steering handles and control panel. Operation: battery-powered (Li-ion) DC motor drives rear wheels; electromagnetic brake engages automatically when not in use. Output: physical mobility for the user. Used in indoor/outdoor environments; operated by the patient. Features include manual/remote folding mechanisms, aluminum alloy frame, and anti-tip tires. Provides mobility assistance to users with limited walking ability.
Clinical Evidence
Bench testing only. Performance verified against ISO 7176 series (stability, braking, energy consumption, dimensions, speed, strength, climatic tests, obstacle climbing, and EMC). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Aluminum alloy frame; 24V 120W motor; 24V/10AH Li-ion battery; electromagnetic braking; 6-inch front/7-inch rear solid tires. Connectivity: remote control functionality. Standards: ISO 7176 series (wheelchair performance), ISO 10993 (biocompatibility), IEC 62304 (software), IEC 60601-1-6/62366 (usability).
Indications for Use
Indicated for disabled or elderly persons with mobility limitations who are able to walk but require a seated position for transportation, for indoor and outdoor use.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 10, 2021
Dongguan Prestige Sporting Goods Co., Ltd. Ms. SK Manager SK Medical Device International Corp. Suite 52, Floor 16, No. 119 Xing Guang Ying Jing, Shui Ying Road Guangzhou. 510663 China
Re: K190737
Trade/Device Name: Solax Electric Scooter (Models: S204311M, S204161, S204143) Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: February 3, 2021 Received: February 9, 2021
Dear Ms. SK:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director. Acute Injurv Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190737
Device Name
Solax Electric Scooter (Models: S204311M, S204161, S204143)
Indications for Use (Describe)
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary for K190737
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## 1. Submitter's Information
- 510(k) Owner's Name: Dongguan Prestige Sporting Goods Co., Ltd. �
- � Establishment Registration Number: 3008841035
- Address: 3rd industrial ,Qiaotou Area, Houjie Town, Dongguan City, Guangdong province, �
China
- � Tel: 13763128800
- � Fax: 86-769-85922505
- � Contact Person: Zhang Zhao (General Manager)
- � E-mail: leon@wisefame.com
## 2. Application Correspondent:
- � Contact Person: Ms. SK
- � SK Medical Device International Corp.
- � Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou,
Guangdong, China
- � Tel: +86-18620793542
- � Email: medical-device@qq.com
## 3. Subject Device Information
| Trade Name: | Solax Electric Scooter |
|-------------|------------------------|
|-------------|------------------------|
{4}------------------------------------------------
| | Model: S204311M, S204161, S204143 |
|----------------------|-----------------------------------|
| Common Name: | Scooter |
| Classification name: | Vehicle, Motorized 3-Wheeled |
| Review Panel: | Physical Medicine |
| Product Code: | INI |
| Regulation Class: | 2 |
| Regulation Number: | 890.3800 |
| | 4. Predicate Device Information | |
|--|----------------------------------|--|
| | | |
| Sponsor | Dongguan Prestige Sporting Goods Co., Ltd. |
|----------------------|---------------------------------------------------------------------|
| Device Name | Solax Electric Scooter<br>Model: S302121, S302131, S302141, S302151 |
| Common Name: | Scooter |
| Classification name: | Vehicle, Motorized 3-Wheeled |
| Review Panel: | Physical Medicine |
| 510(k) Number | K172440 |
| Product Code | INI |
| Regulation Number | 890.3800 |
| Regulation Class | 2 |
## 5. Device Description
The Solax Electric Scooter (Models: S204311M, S204161, S204143) is an indoor/outdoor electric scooter that is intended to be used by individuals that are able to walk, but suffer from mobility limitations. It has the following main components: two 6 inch solid front tires, two 7 inch solid rear tires, two anti-tip tires, control panel, steering handles, seat folding lever, backrest, arm rest, seat, steering column, seat frame, front/ rear covers, folding release lever, angle adjustment lever, height adjustment lock, carry handle, an off-board charger and aluminum alloy made frame, It is powered by two Li-ion DC rechargeable batteries with 18 km (on level surface) which maximum speed up to 6 km/hr.
{5}------------------------------------------------
The movement of the scooter is controlled by the steering handles and control panel. The device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually.
#### 6. Intended Use / Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
#### 7. Performance Summary
The following performance, safety and biocompatibility tests were conducted with Solax Electric Scooter:
- · ISO 7176-1: 2014, Wheelchairs Part 1: Determination of static stability
- ISO 7176-2:2017, Wheelchairs Part 2: Determination of dynamic stability of Powered Wheelchairs
- ISO 7176-3: 2012, Wheelchairs Part 3: Determination of effectiveness of brakes
- · ISO 7176-4, Third edition 2008-10-01, Wheelchairs Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
- · ISO 7176-5, Second edition 2008-06-01, Wheelchairs Part 5: Determination of overall dimensions, mass and manoeuvring space
- ISO 7176-6: 2018, Wheelchairs Part 6: Determination of maximum speed, acceleration and deceleration of Powered Wheelchairs
- ISO 7176-7, First Edition 1998-05-15, Wheelchairs Part 7: Measurement of seating and wheel dimensions
- ISO 7176-8:2014, Wheelchairs Part 8: Requirements and test methods for static, impact and fatigue strengths
{6}------------------------------------------------
- ISO 7176-9:2009, Wheelchairs Part 9: Climatic tests for Powered Wheelchairs
- · ISO 7176-10:2008, Wheelchairs Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
- ISO 7176-11, Second edition 2012-12-01, Wheelchairs Part 11: Test dummies
- · ISO 7176-13, First edition 1989-08-01, Wheelchairs Part 13: Determination of coefficient of friction of test surfaces
- · ISO 7176-14:2008. Wheelchairs Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods
- · ISO 7176-15:1996, Wheelchairs Part 15: Requirements for information disclosure, documentation and labeling
- · ISO 7176-16, Second edition 2012-12-01, Wheelchairs Part 16: Resistance to ignition of postural support devices
- ISO 7176-21 Second edition 2009-04-01 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
- · IEC 62304: 2006 (First Edition), Medical device software, Software life- cycle processes.
- · IEC 60601-1-6 Medical electrical equipment- Part1-6: General requirements for safety-Collateral Standard: Usability Edition 3.1, 2013
- · IEC 62366 Medical devices Application of usability engineering to medical devices Edition 1.0.2007
The materials and manufacturing used for the Solax Electric Scooter are identical to those of the predicate device, which were demonstrated to conform with the following biocompatibility standards:
- · ISO 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for In Vitro
{7}------------------------------------------------
cytotoxicity
- ISO 10993-10: 2009, Biological evaluation of medical devices Part 10: Tests for irritation and delayed-type hypersensitivity
## 8. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Solax Electric Scooter is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Predicate Device | Subject Device | Remark |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Proprietary<br>name | Solax Electric Scooter | Solax Electric Scooter | -- |
| Model | S302121, S302131, S302141,<br>S302151 | S204311M, S204161, S204143 | -- |
| 510K Number | K172440 | K190737 | -- |
| Common or<br>Usual name | Electrical scooter | Electrical scooter | -- |
| Intended use | It is a motor driven, indoor and<br>outdoor transportation vehicle with<br>the intended use to provide mobility<br>to a disabled or elderly person<br>limited to a seated position. | It is a motor driven, indoor and<br>outdoor transportation vehicle with<br>the intended use to provide mobility<br>to a disabled or elderly person<br>limited to a seated position. | SE |
| Indications<br>for use | It is a motor driven, indoor and<br>outdoor transportation vehicle with<br>the intended use to provide mobility<br>to a disabled or elderly person<br>limited to a seated position. | It is a motor driven, indoor and<br>outdoor transportation vehicle with<br>the intended use to provide mobility<br>to a disabled or elderly person<br>limited to a seated position. | SE |
| Rx or OTC? | OTC | OTC | SE |
| Size (unfold) | 930*450*865mm | S204311M: 980*450*940mm<br>S204161: 930*450*860mm | SE<br>Note 1 |
| | | S204143: 980*450*880mm | |
| Tires | 6 inches for front wheel (solid wheel)<br>7 inches for rear wheel (solid wheel) | 6 inches for front wheel (solid wheel)<br>7 inches for rear wheel (solid wheel) | SE |
| Speed | 6 km/h (3.7mph) | 6 km/h (3.7mph) | SE |
| Safe Gradient / Maximum Gradient | 0-12° | S204311M: 0-13°<br>S204161, S204143: 0-15° | SE<br>Note 1 |
| Range | 15 km (9.32mile) | 15 km (9.32mile) | SE |
| Turning circle | 1.55 m | S204311M: 1.55m (with a speed less than 3km/h is recommend)<br>S204161, S204143: 1.35 m (with a speed less than 3km/h is recommend) | SE<br>Note 1 |
| Base weight(not including battery) | 24 kg | S204311M: 24kg<br>S204161, S204143: 24.7 kg | SE<br>Note 1 |
| Battery weight | 1.84 kg | 1.84 kg | SE |
| Brake | Electromagnetic | Electromagnetic | SE |
| Drive system | PG 45A / Rear wheel drive | PG 45A / Rear wheel drive | SE |
| Maximum capacity | 125 kg Approx. | 125 kg Approx. | SE<br>Note 1 |
| Ground clearance | 36 mm | S204311M: 38mm<br>S204161, S204143: 58mm | SE<br>Note 1 |
| Obstacle Climbing Ability | 38 mm | S204311M: 40mm<br>S204161, S204143: 60mm | SE<br>Note 1 |
| Frame design - Construction | X type | X type | SE |
| Frame design - Materials | aluminum alloy frame | aluminum alloy frame | SE |
| Folding mechanism | Manual folding and remote folding | Manual folding for S204311M,<br>S204161, S204143 | SE<br>Note 2 |
| Remote<br>control | None | Yes | SE<br>Note 2 |
| Battery | Lithium battery<br>24V/10AH | Lithium battery<br>24V/10AH | SE |
| Motor | 24V 120W | 24V 120W | SE |
| Battery<br>charger | DC24V/2A | DC24V/2A | SE |
| Biocompatibility | All user directly contacting materials<br>are compliance with ISO10993-5<br>and ISO10993-10 requirements. | All user directly contacting materials<br>are compliance with ISO10993-5<br>and ISO10993-10 requirements. | SE |
{8}------------------------------------------------
{9}------------------------------------------------
## K190737
## Comparison in Detail(s):
#### Note 1:
The design and technological characteristics of the Solax Electric Scooter is basically similar to the predicate device chosen. There are minor differences between the devices including overall dimensions, Safe Gradient, Base weight, Maximum capacity, Ground clearance, Obstacle Climbing Ability. There is no deleterious affection of safety and effectiveness about the differences and these minor differences do not influence the intended use function and use of the device. Moreover, the non-clinical tests and the predicate comparisons demonstrate that these differences in their technological characteristics do not raise any questions as to the safety and effectiveness. Therefore the Solax Electric Scooter (models: S204311M, S204143) is substantially equivalent to the Solax Electric Scooter (models: S302121, S302141, S302151) (K172440).
## Note 2:
Although the "Folding mechanism" and "Remote control" is a little different with - predicate device,
but these difference will not affect the basic safety and indications for use of subject device, they
are not necessary function which will not raise any safety issue and effectiveness issue.
## 9. Date of the summary prepared: May 6, 2021
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.