High Tone Power therapy (HiToP) (Model: HiToP PNP)

K252816 · Mohammadali Nezakati · IPF · Jun 2, 2026 · Physical Medicine

Device Facts

Record IDK252816
Device NameHigh Tone Power therapy (HiToP) (Model: HiToP PNP)
ApplicantMohammadali Nezakati
Product CodeIPF · Physical Medicine
Decision DateJun 2, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

HiToP® PNP is intended to be used as: Powered Muscle Stimulator (EMS): -Relaxation of muscle spasms -Increase local blood circulation -Muscle re-education -Maintaining or increasing range of motion Transcutaneous Electrical Nerve Stimulator (TENS): adjunctive use in adults to provide temporary relief of pain associated with mild to moderate diabetic peripheral neuropathy of the lower extremities. HiToP® PNP should be applied to normal, healthy, dry, and clean skin.

Device Story

HiToP® PNP is a non-invasive electrotherapeutic device for pain management and muscle stimulation. It delivers medium-frequency sinusoidal electrical waves (4096–32768 Hz) to muscle groups via skin-applied electrodes. The device utilizes SimulFAM® (Simultaneous Frequency Amplitude Modulation) technology, where amplitude and frequency are modulated simultaneously. Operated in healthcare facilities/hospitals, the user controls parameters via a knob and graphic display. The device provides electrical impulses to induce muscle contraction or nerve stimulation. Clinical benefits include temporary relief of diabetic peripheral neuropathy pain and improved muscle function. The device includes safety features like automatic over-current, overload, no-load, and shut-off trips, plus a patient override mechanism (connector removal).

Clinical Evidence

A randomized controlled trial (Awad et al., 2026) compared HiToP® PNP to standard TENS in 60 male patients with type 2 diabetes and mild to moderate peripheral neuropathy. Patients received 30-minute sessions, 3x/week for 3 months. Primary endpoints: DN4 questionnaire and pain pressure threshold (PPT). HiToP® PNP showed superior results: DN4 scores decreased 26% (HiToP) vs 14% (TENS) (p=0.001); PPT increased 20.9% (HiToP) vs 10.9% (TENS) (p=0.001). No serious adverse events reported.

Technological Characteristics

Medium-frequency sinusoidal current (4096–32768 Hz); SimulFAM® (Simultaneous Frequency Amplitude Modulation); Biphasic waveform; 1 output channel; Regulated Voltage (CV) output; ABS plastic housing; 100-240V AC power; 2 MOPP isolation. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, IEC 62304, and IEC 62366. Non-sterile.

Indications for Use

Indicated for adults with mild to moderate diabetic peripheral neuropathy of the lower extremities for temporary pain relief, and for general muscle stimulation (spasm relaxation, circulation, re-education, range of motion). Contraindicated for use on non-healthy, wet, or unclean skin.

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** **U.S. FOOD & DRUG** ADMINISTRATION June 2, 2026 Mohammadali Nezakati 65 Enterprise Suite 400 Aliso Viejo, CA 92656 Re: K252816 Trade/Device Name: High Tone Power therapy (HiToP®) (Model: HiToP® PNP) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ Dated: September 3, 2025 Received: September 4, 2025 Dear Mohammadali Nezakati: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252816 - Mohammadali Nezakati Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K252816 - Mohammadali Nezakati Page 3 Sincerely, **AMBER T. BALLARD -S** Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. # Indications for Use 510(k) Number (if known) K252816 Device Name High Tone Power therapy (HiToP®) (Model: HiToP® PNP) Indications for Use (Describe) HiToP® PNP is intended to be used as: Powered Muscle Stimulator (EMS): -Relaxation of muscle spasms -Increase local blood circulation -Muscle re-education -Maintaining or increasing range of motion Transcutaneous Electrical Nerve Stimulator (TENS): adjunctive use in adults to provide temporary relief of pain associated with mild to moderate diabetic peripheral neuropathy of the lower extremities. HiToP® PNP should be applied to normal, healthy, dry, and clean skin. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} HillTek LLC HiToP® 510(k) Submission # 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92: # I. SUBMITTER Mohammadali Nezakati 65 Enterprise Suite 400 Aliso Viejo, CA 92656 +1.949.620.1408 Date Prepared: May 8, 2026 # II. DEVICE Name of Device: High Tone Power therapy (HiToP®) (Model: HiToP® PNP) Classification Name: Powered muscle stimulator Regulation: 21 CFR 890.5850 Regulatory Class: Class II Product Classification Code: IPF, GZJ # III. PRIMARY PREDICATE DEVICE - Predicate Manufacturer: gbo Medizintechnik AG - Predicate Trade Name: HiToP® (Model: HiToP® 1touch) - Predicate 510(k): K213655 # IV. DEVICE DESCRIPTION The High Tone Power therapy (HiToP® PNP) is a non-invasive electrotherapeutic device used as an Electrical muscle stimulator/ powered muscle stimulator and Transcutaneous Electrical Nerve Stimulator (TENS). The device is designed for pain management, muscle strengthening, training, muscle relaxation, and re-education. The device stimulates the underlying muscle groups through electrical impulses transmitted via electrodes applied on the skin. The device can deliver medium frequency sinusoidal electrical waves with different programmable parameters such as frequency, duration of contraction, duration of rest, and total session duration to underlying muscle groups. In High Tone power therapy, the amplitude and the frequency are modulated simultaneously. 510(k) Summary Page 1 of 10 {5} HillTek LLC HiToP® 510(k) Submission The High Tone Power Therapy devices (HiToP® PNP) provide a therapy with medium frequency sine current waves ranging from 4096 – 32768 Hz. The amplitude (A) and frequency (f) of Sine waves can be simultaneously modulated, so this method is called SimulFAM® which stands for Simultaneous Frequency Amplitude Modulation. HiToP® PNP is used at healthcare facilities, and hospitals. The HiToP® PNP parameters are controlled by the driver (control knob), and a graphic display that shows therapy parameters. # SimulFAM®X A frequency scan of three octaves is realized. The frequency scan is realized with a fixed speed of 20 Hz. # V. INDICATIONS FOR USE HiToP® PNP is intended to be used as: # Powered muscle stimulator (PMS) - Relaxation of muscle spasms - Increase local blood circulation - Muscle re-education - Maintaining or increasing range of motion # Transcutaneous Electrical Nerve Stimulator (TENS) - Adjunctive use in adults to provide temporary relief of pain associated with mild to moderate diabetic peripheral neuropathy of the lower extremities. HiToP® PNP should be applied to normal, healthy, dry, and clean skin. 510(k) Summary Page 2 of 10 {6} HillTek LLC HiToP® 510(k) Submission # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The following characteristics were compared between the subject device and the predicate device to demonstrate substantial equivalence: | Device | Subject Device HiToP® PNP | Predicate device HiToP®1 touch | SE comparison | | --- | --- | --- | --- | | Submitter | M Nezakati | HillTek LLC | - | | 510K Number | K252816 | K213655 | - | | Common or Usual Name | Transcutaneous Electrical Nerve Stimulator for pain relief | Transcutaneous Electrical Nerve Stimulator for pain relief | Same | | | Powered Muscle Stimulator | Powered Muscle Stimulator | Same | | Product Code | IPF, GZJ, | IPF, LIH, GZJ | Similar. The subject device does not include a mode for interferential current therapy, therefore, product code LIH is not applicable to the subject device. | | Product Classification | Class II | Class II | Same | | Classification Name | Stimulator, muscle, powered (IPF) | Stimulator, muscle, powered (IPF) | Same | | | Stimulator, nerve, transcutaneous, for pain relief (GZJ) | Stimulator, nerve, transcutaneous, for pain relief (GZJ) | Same | | | N/A | Interferential Current Therapy (LIH) | The subject device does not include a mode for interferential current therapy | | Regulation Number | 21 CFR 890.5850 | 21 CFR 890.5850 | Same for applicable regulation | | Indications for Use | **HiToP® PNP** is intended to be used as: **Powered muscle stimulator (PMS)** | Powered muscle stimulator • Relaxation of muscle spasms • Increase local blood circulation • Muscle re-education | **Similar:** The TENS mode of the subject device is indicated for temporary relief of pain associated with mild to moderate diabetic | 510(k) Summary Page 3 of 10 {7} HillTek LLC HiToP® 510(k) Submission | | - Relaxation of muscle spasms - Increase local blood circulation - Muscle re-education - Maintaining or increasing range of motion**Transcutaneous Electrical Nerve Stimulator (TENS)** - Adjunctive use in adults to provide temporary relief of pain associated with mild to moderate diabetic peripheral neuropathy of the lower extremities.**HiToP® PNP** should be applied to normal, healthy, dry, and clean skin. | - Maintaining or increasing range of motion**TENS** - Symptomatic relief and management of chronic, intractable pain | peripheral neuropathy of the lower extremities, while the TENS mode of the predicate device is indicated for chronic, intractable pain. However, these differences do not alter the intended use of the device, nor do they raise different questions of safety and effectiveness relative to the predicate device. | | --- | --- | --- | --- | | Type of Use | Prescription (Rx Only) | Prescription (Rx Only) | Same | | Power Source(s) | 100-240V~, 50-60 Hz, 50VA / AC Line | 100-240V~, 50-60 Hz, 45VA / AC Line | Same | | Method of Line Current Isolation | 2 MOPP | 2 MOPP | Same | | Patient leakage current – normal condition (μA) | 16.0μA | 8.0 μA | **Different.** However these differences do not raise different questions of safety or effectiveness. Additionally, the subject device has been tested for patient leakage current under Normal and Single fault conditions according to standard requirement IEC IEC60601-1. | | Patient leakage current – single fault condition (μA) | 25μA AC | 16.0μA (AC SFC 264V 60 Hz) | | 510(k) Summary Page 4 of 10 {8} HillTek LLC HiToP® 510(k) Submission | Number of Output Channels | | 1 | 1 | Same | | --- | --- | --- | --- | --- | | Number of Output Modes | | 1 | 2 | Different: The predicate device has two output modes, SimulFAM®X and SimulFAM®I. The subject device has a fixed output mode, SimulFAM®X. These differences do not raise different questions of safety or effectiveness. | | Synchronous or alternating? | | Alternating | N/A (1 channel only) | Different. These differences do not raise different questions of safety or effectiveness. Additionally, both the subject device and the predicate device produce alternating output. | | Method of Channel Isolation | | Transformer isolated | Transformer isolated | Same | | Regulated Current or Regulated Voltage? | | Regulated Voltage (CV) | Regulated Voltage (CV) | Same | | Software/Firmware/Microprocessor Control? | | yes | yes | Same | | Automatic Over current Trip? Yes/no | | yes | yes | Same | | Automatic Overload Trip? Yes/no | | yes | yes | Same | | Automatic No-Load Trip? Yes/no | | yes | yes | Same | | Automatic Shut Off? Yes/no | | yes | yes | Same | | Patient Override Control? Yes/no | | yes | yes | Same | | Patient Override Control method | | Removal of patient connector at the device. | Removal of patient connector at the device. | Same | | Indicator Display | On/off status? Yes/No | yes | yes | Same | | | Voltage/Current Level? Yes/No | yes | yes | Same | | Timer Range | | 30 – 60 min | 1 – 90 min | Different. However these differences do not raise different questions of safety or effectiveness. Additionally, the | 510(k) Summary Page 5 of 10 {9} HillTek LLC HiToP® 510(k) Submission | | | | treatment time is adjusted by the user or dependent on selected programs. | | --- | --- | --- | --- | | Compliance with 21 CFR 898? | Yes | Yes | Same | | Compliance with voluntary standards • IEC60601-1 • IEC60601-1-2 • IEC60601-2-10 • ISO14971 • IEC62304 • IEC62366 | Yes | Yes | Same | | Weight (lbs., oz.) | 3.08 lbs. without accessories | 5.5 lbs. without accessories | **Different.** However these differences do not raise different questions of safety or effectiveness. There are minor differences between the subject device and the predicate devices in device weight and dimensions. | | Dimensions (in.) [W x H x D] | 9.05" x 5.11" x 9.05" | 10.6" x 9.64" x 2.36" | **Different.** However these differences do not raise different questions of safety or effectiveness. There are minor differences between the subject device and the predicate devices in device weight and dimensions. | | Environment of use | Healthcare facilities and Hospitals | Healthcare facilities and Hospitals | **Same** | | Electrode shape (square, round, rectangular, oval) in cm² | Rectangular 96 cm² | Rectangular 12, 48, 96, 200 cm² | **Similar:** The subject device using only one of the four sizes of the predicate device | | Housing Material and Construction | ABS Plastic | Aluminum | **Different.** However this difference does not raise different questions of safety or effectiveness. | 510(k) Summary Page 6 of 10 {10} HillTek LLC HiToP® 510(k) Submission OUTPUT SPECIFICATIONS -Comparison of Subject Device Models with Predicate Rx Device | | Subject device HiToP® PNP | Predicate device HiToP® 1touch | SE Comparison | | --- | --- | --- | --- | | 510K Number | K252816 | K213655 | - | | Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Biphasic | Same | | Maximum output Current (±20%) | 100mA@500 Ω load 38mA@2 k Ω 7.6m@10 k Ω | 100mA@500 Ω load 38mA@2 k Ω 7.6m@10 k Ω | Same | | Maximum output voltage (±20%) | 50V per channel @500 Ω 50V per channel @2k Ω 50V per channel @10k Ω | 76V per channel @500 Ω 76V per channel @2k Ω 76V per channel @10k Ω | Different. However these differences do not raise different questions of safety or effectiveness. | | Carrier Frequency | 4096 – 32768 Hz | 4096 – 32768 Hz | Same | | Modulation Frequency | 20 Hz fixed | 0.1 to 200 Hz adjustable | Different. However these differences do not raise different questions of safety or effectiveness. The subject device has a single fixed program. The subject device Modulation Frequency is within the range of the predicate device Modulation Frequency. | | Phase Duration | 15.26 to 122.07 μs | 15.26 to 122.07 μs | Same | | Pulse Width | Not applicable | 1.5 s – 120 s | Different. However these differences do not raise different questions of safety or effectiveness. Pulse width is not applicable to HiToP devices output. | | Maximum Current Density, (mA/cm²) | 100mA @ 500 Ω /96cm² = 1.04 | 100mA @ 500 Ω /12cm² = 8.3 | Different: The subject device Max Current is lower than the predicate device. Due to the different sizes of the smallest electrode surface. The differences would not affect the safety and effectiveness of the subject device. | | Maximum Power Density, (W/CM²), (using smallest | 5W / 96cm²=0.052 | 5W / 12cm²=0.42 | Different: The subject device Max Power is lower than the predicate device, due to the different sizes of the | 510(k) Summary Page 7 of 10 {11} HillTek LLC HiToP® 510(k) Submission | electrode conductive surface area) | | | smallest electrode surface. The differences would not affect the safety and effectiveness of the subject device. | | --- | --- | --- | --- | | Maximum Power Intensity | 5W | 5W | Same | | Maximum Phase Charge @500 Ω | 8.32 μC | 10.99 μC | Different: The subject device Maximum Phase Charge is lower than the predicate device, due to the different sizes of the smallest electrode surface. The differences would not affect the safety and effectiveness of the subject device. | The following performance data were provided in support of the substantial equivalence determination. ### Sterilization & Shelf-life Testing Not Applicable- Device is non-sterile. ### Biocompatibility Testing Biocompatibility assessment of conductive rubber electrodes in the subject device was performed in conformance with iso 10993-1, "biological evaluation of medical devices - part 1: evaluation and testing within a risk management process" as recognized by FDA. The following endpoints were evaluated for conductive rubber electrodes. - Cytotoxicity – ISO 10993-5 - Skin Irritation – ISO 10993-10 - Sensitization – ISO 10993-10 ### NOTE: Elastic straps have not been evaluated for biocompatibility. Precautionary statements and the measures necessary to mitigate the risks in the event of an adverse event associated with the use of the elastic straps are included within the device labeling. ### Electrical safety and electromagnetic compatibility (EMC) The following electrical safety and EMC testing reports were provided: - Electrical Safety Testing IEC 60601-1 510(k) Summary Page 8 of 10 {12} HillTek LLC HiToP® 510(k) Submission - EMC Testing IEC 60601-1-2 - Home Use Environment IEC 60601-1-11 - Nerve and Muscle Stimulators IEC 60601-2-10 ### Software Verification and Validation Testing Software verification and validation testing was provided in accordance with IEC 62304 and FDA guidance documents. ### Benchtop Performance Testing The following benchtop performance testing was provided: - Output waveforms - Basic unit characteristics - Output specifications - Uniform current distribution ### Animal Study Animal performance testing was not required to demonstrate safety and effectiveness of the device. ### Clinical Testing Human clinical performance testing was provided to support the safety and effectiveness of the subject device as an adjunctive use in adults to provide temporary relief of pain associated with mild to moderate diabetic peripheral neuropathy of the lower extremities. The primary supporting study is: Awad, Z.M., Badr, N.M., Ismail, L.A., & El-Moatasem, A.M. (2026). High-tone power therapy versus transcutaneous electrical nerve stimulation on polyneuropathic pain in diabetic patients. Physiotherapy Research International. 31:e70163. doi:10.1002/pri.70163 This randomized controlled trial enrolled 60 male patients with type 2 diabetes and mild to moderate diabetic peripheral neuropathy pain. Patients were randomized to receive either HiToP® PNP or Enraf-Nonius TENS applied to the lower extremities for 30 minutes per session, three times per week, over approximately three months. The primary outcomes were the Douleur Neuropathique 4 (French for "Neuropathic Pain 4", DN4) questionnaire and pain pressure threshold (PPT). The study demonstrated statistically significant improvements in both primary outcomes, with larger improvements observed in the HiToP® PNP group compared to the TENS group. DN4 scores decreased 26% in the HiToP group versus 14% in the TENS group, with a statistically significant difference between-group difference (mean difference: -0.87, 95% CI: -1.17 to -0.56, p=0.001). Pain pressure threshold increased 20.9% in the HiToP group versus 10.9% in the TENS group (between-group mean difference: 0.3, 95% CI: 0.1 to 0.5, p=0.001). Only minor adverse effects were reported, including mild skin irritation and transient fatigue, with no serious adverse events observed. 510(k) Summary Page 9 of 10 {13} HillTek LLC HiToP® 510(k) Submission--- ## VII. CONCLUSIONS Based on the comparison to the predicate device (K213655, HiToP® 1touch) and on the outcomes of non-clinical and clinical performance tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device and any differences between the devices do not pose any new questions of safety and effectiveness. Thus, the HiToP® PNP is substantially equivalent to the predicate device. ---510(k) Summary Page 10 of 10
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...