HiToP®4touch, HiToP®2touch, and HiToP®1touch is intended to be used as Powered muscle stimulator - Relaxation of muscle spasms - Increase local blood circulation - Muscle re-education - Maintaining or increasing range of motion TENS - Symptomatic relief and management of chronic, intractable pain
Device Story
HiToP® (HiToP®4touch, HiToP®2touch, HiToP®1touch) is a non-invasive electrical muscle stimulator and TENS device; used in healthcare facilities and hospitals by clinicians. Device delivers medium frequency sinusoidal electrical waves (4096–32768 Hz) to muscle groups via skin-applied electrodes. Operates using SimulFAM® (Simultaneous Frequency Amplitude Modulation) where amplitude and frequency are modulated simultaneously; frequency scan of three octaves realized at speeds of 0.1–200 Hz. Controlled via graphical user interface and driver knob. Output parameters include frequency, rest duration, and session duration. Device provides therapeutic stimulation for pain management, muscle strengthening, training, relaxation, and re-education. Benefits include non-invasive symptomatic pain relief and improved muscle function.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by benchtop performance testing, including output waveform analysis, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility testing (ISO 10993-5, ISO 10993-10) for conductive rubber electrodes.
Technological Characteristics
Medium frequency sinusoidal electrical stimulator (4096–32768 Hz). Materials: Aluminum housing, conductive rubber electrodes. Connectivity: Standalone. Power: 100-240V AC. Standards: IEC 60601-1 (Safety), IEC 60601-1-2 (EMC), IEC 60601-2-10 (Nerve/Muscle Stimulators), ISO 14971 (Risk Management), IEC 62304 (Software), IEC 62366 (Usability). Software: Microprocessor-controlled with automatic over-current, overload, no-load, and shut-off safety features.
Indications for Use
Indicated for patients requiring relaxation of muscle spasms, increased local blood circulation, muscle re-education, maintenance or increase of range of motion, and symptomatic relief/management of chronic, intractable pain.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2023
HillTek LLC % Bhoomika Joyappa Medical Device Academy Inc, 345 Lincoln Hill Road Shrewsbury, Vermont 05738
### Re: K213655
Trade/Device Name: HiTop® (Models: HiToP®4touch, HiToP®2touch, HiToP®1touch) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ, LIH Dated: December 19, 2022 Received: December 20, 2022
### Dear Bhoomika Joyappa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
### 510(k) Number (if known) K213655
### Device Name
HiTop® (Models: HiToP®4touch, HiToP®2touch, HiToP®1touch)
Indications for Use (Describe) HiToP®4touch, HiToP®2touch, and HiToP®1touch is intended to be used as
Powered muscle stimulator
- Relaxation of muscle spasms
- Increase local blood circulation
- · Muscle re-education
- · Maintaining or increasing range of motion
### TENS
· Symptomatic relief and management of chronic, intractable pain
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY K213655
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
#### l. SUBMITTER
HillTek LLC 421 S Brookhurst Street; Ste#142 Anaheim, CA 92804, USA +1.206.406.9390 Contact Person: Mohammadali Nezakati Date Prepared: November 18, 2021
#### II. DEVICE
| Name of Device: | HiToP® (Models: HiToP®4touch, HiToP®2touch, HiToP®1touch) |
|------------------------------|-----------------------------------------------------------|
| Classification Name: | Physical Medicine Devices and Neurology |
| Regulation: | 21 CFR §890.5850 |
| Regulatory Class: | Class II |
| Product Classification Code: | IPF, GZJ, LIH |
#### III. PRIMARY PREDICATE DEVICE
- Predicate Manufacturer: Mettler Electronics Corp.
- Predicate Trade Name: Sys*Stim 240 ●
- . Predicate 510(k): K113017
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The High Tone Power therapy (HiToP®) is a non-invasive electrical muscle stimulator/ powered muscle stimulator and Transcutaneous Electrical Nerve Stimulator (TENS). The device is designed for pain management, muscle strengthening, training, muscle relaxation, and re-education. The device stimulates the underlying muscle groups through electrical impulses transmitted via electrodes applied on the skin. The device can deliver medium frequency sinusoidal electrical waves with different programmable
{4}------------------------------------------------
parameters such as frequency, duration of rest, and total session duration to underlying muscle groups. In High Tone power therapy, the amplitude and the frequency are modulated simultaneously.
The High Tone Power Therapy devices (HiToP®) provide a therapy with medium frequency sine current waves ranging from 4096 – 32768 Hz. The amplitude (A) and frequency (f) of Sine waves can be simultaneously modulated, so this method is called SimulFAM® which stands for Simultaneous Frequency Amplitude Modulation.
### SimulFAM®X
A frequency scan of three octaves is realized. The frequency scan is realized with different speed (0.1 - 200 Hz).
There are three models of HiToP®, which are HiToP® touch, HiToP® 1 touch. Each model's intended use, working principle, product structure, and significant parameters are the slight difference in product appearance and number of programs. HiToP®4 touch, HiToP®2 touch, HiToP®1 touch is used at healthcare facilities and hospitals.
The device's parameters are controlled by the driver (control knob), and the user can select the desired therapy using a graphical user interface.
#### V. INDICATIONS FOR USE
HiToP®4 touch, HiToP®2 touch, and HiToP®1 touch is intended to be used as Powered muscle stimulator
- . Relaxation of muscle spasms
- Increase local blood circulation .
- Muscle re-education .
- Maintaining or increasing range of motion ●
### TENS
- Symptomatic relief and management of chronic, intractable pain ●
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PRIMARY PREDICATE DEVCE
The following characteristics were compared between the predicate device to demonstrate substantial equivalence:
| Device | Subject device Model<br>HiToP®4 touch | Subject device Model<br>HiToP®2 touch | Subject device Model<br>HiToP®1 touch | Primary Predicate<br>Device<br>Sys*Stim 240 | SE comparison |
|--------|---------------------------------------|---------------------------------------|---------------------------------------|---------------------------------------------|---------------|
|--------|---------------------------------------|---------------------------------------|---------------------------------------|---------------------------------------------|---------------|
{5}------------------------------------------------
| Submitter | HillTek LLC | HillTek LLC | HillTek LLC | Mettler Electronics<br>Corp | - | | Interferential Current<br>Therapy LIH | Interferential Current<br>Therapy LIH | Interferential Current<br>Therapy LIH | Interferential Current<br>Therapy LIH | Same |
|------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K Number | K213655 | K213655 | K213655 | K113017 | - | | NA | NA | NA | Lamp, Infrared,<br>Therapeutic Heating<br>ILY | Different:<br>The subject device is<br>not a Lamp, Infrared,<br>Therapeutic Heating<br>device. Therefore,<br>product code ILY is not<br>applicable to the<br>subject device. This<br>eliminates the risks<br>associated with that<br>function |
| Common or Usual Name | Transcutaneous<br>Electrical Nerve<br>Stimulator for pain<br>relief<br>Powered Muscle<br>Stimulator | Transcutaneous<br>Electrical Nerve<br>Stimulator for pain<br>relief<br>Powered Muscle<br>Stimulator | Transcutaneous<br>Electrical Nerve<br>Stimulator for pain<br>relief<br>Powered Muscle<br>Stimulator | Transcutaneous<br>Electrical Nerve<br>Stimulator for pain<br>relief<br>Powered Muscle<br>Stimulator | Same | Regulation Number | 21 CFR 890.5850 | 21 CFR 890.5850 | 21 CFR 890.5850 | 21 CFR 890.5850<br>21 CFR 890.5500 | Similar<br>21 CFR 890.5500 is<br>used for Lamp,<br>Infrared, Therapeutic<br>Heating device that is<br>not applicable to the<br>subject device. This<br>eliminates the risks<br>associated with that<br>function |
| | NA | NA | NA | Infrared lamp | - | Indications for Use | Powered muscle<br>stimulator<br>• Relaxation of<br>muscle spasms<br>• Increase local<br>blood circulation<br>• Muscle re-<br>education<br>• Maintaining or<br>increasing range of<br>motion<br>TENS | Powered muscle<br>stimulator<br>• Relaxation of<br>muscle spasms<br>• Increase local<br>blood circulation<br>• Muscle re-<br>education<br>• Maintaining or<br>increasing range<br>of motion<br>TENS | Powered muscle<br>stimulator<br>• Relaxation of<br>muscle spasms<br>• Increase local<br>blood circulation<br>• Muscle re-<br>education<br>• Maintaining or<br>increasing range of<br>motion | Indications for<br>Medium Frequency<br>(Russian), Biphasic,<br>High Volt Pulsed<br>Current (HVPC),<br>Interferential (4P) and<br>Premodulated (2P3)<br>waveforms:<br>Relaxation of muscle<br>spasms<br>• Prevention or<br>retardation of<br>disuse atrophy | Similar:<br>The indication of use<br>is identical in both<br>subject device and<br>predicate devices. The<br>additional indications<br>for use i.e Indications<br>for DC (Direct Current)<br>Mode and the laser<br>and cluster applicators<br>in the predicate<br>devices is not<br>applicable for the |
| Product Code | IPF, LIH, GZJ | IPF, LIH, GZJ | IPF, LIH, GZJ | IPF, ILY, GZJ, LIH | Similar<br>The subject device is<br>used for electrical<br>stimulation. It is not a<br>Lamp, Infrared,<br>Therapeutic Heating<br>device. Therefore,<br>product code ILY is not<br>applicable to the<br>subject device. This<br>eliminates the risks<br>associated with that<br>function | | | | | | |
| Product Classification | Class II | Class II | Class II | Class II | Same | | | | | | |
| Classification Name | Stimulator, nerve,<br>transcutaneous, for<br>pain relief (GZJ)<br>Stimulator, muscle,<br>powered (IPF) | Stimulator, nerve,<br>transcutaneous, for<br>pain relief (GZJ)<br>Stimulator, muscle,<br>powered (IPF) | Stimulator, nerve,<br>transcutaneous, for<br>pain relief (GZJ)<br>Stimulator, muscle,<br>powered (IPF) | Stimulator, nerve,<br>transcutaneous, for<br>pain relief (GZJ)<br>Stimulator, muscle,<br>powered (IPF) | Same | | | | | | |
{6}------------------------------------------------
{7}------------------------------------------------
| Symptomatic relief and management of chronic, intractable pain | Symptomatic relief and management of chronic, intractable pain | TENS Symptomatic relief and management of chronic, intractable pain | Increase local blood circulation Muscle re-education Maintaining or increasing range of motion Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Indications for Microcurrent, Interferential (4P3), Premodulated (2P), Biphasic, and TENS waveforms: Symptomatic relief and management of chronic, intractable pain Post-traumatic acute pain Post-surgical acute pain Indications for DC (Direct Current) Mode Relaxation of muscle spasm The laser and cluster | subject device as the subject device is not an Infrared, Therapeutic Heating device and does not work in DC mode |
|----------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
{8}------------------------------------------------
| Type of Use | Prescription (Rx Only) | Prescription (Rx Only) | Prescription (Rx Only) | Prescription (Rx Only) | Same<br>Different: | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|---------------------------------------------------|---------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Power Source(s) | 100-240V~, 50-60 Hz,<br>200VA / AC Line | 100-240V~, 50-60 Hz,<br>120VA / AC Line | 100-240V~, 50-60 Hz,<br>45VA / AC Line | AC line or optional<br>battery pack 10.8 V AC<br>line or optional<br>battery pack Lithium<br>Ion | The subject device is<br>limited to use with AC<br>line as the power<br>source. The subject<br>device does not<br>include an optional<br>battery pack like the<br>predicate device, and<br>this does not<br>introduce any new<br>risks. The subject<br>device complies with<br>IEC 60601-1<br>requirements for<br>evaluation of safety.<br>These differences do | |
| Sys*Stim 240 emit infrared energy for:<br>Temporary increase in local blood circulation Temporary relief of minor muscle and joint aches, pains and stiffness Relaxation of muscles Temporary relief of muscle spasms Temporary relief of minor pain and stiffness associated with arthritis | | | | | | |
| | | | | | question of safety and<br>effectiveness | |
| Method of Line Current Isolation | 2 MOPP | 2 MOPP | 2 MOPP | Yes, value unknown | Similar:<br>The subject device has<br>been tested for<br>electrical isolation<br>protection against<br>electric shock<br>according to standard<br>requirement IEC<br>IEC60601-1. Does not<br>raise any safety and<br>effectiveness concerns | |
| Patient leakage current - normal<br>condition (μΑ) | 4.7 μΑ | 4.7 μΑ | 8.0 μΑ | 78 μΑ (Less than 100 μΑ) | Similar:<br>The subject device<br>demonstrated<br>electrical safety by<br>passing ANSI AAMI<br>ES60601-1 and IEC<br>60601-2-10 tests. The<br>difference does not<br>raise the issue of<br>product's safety and<br>effectiveness. | |
| Patient leakage current - single<br>fault condition (μΑ) | 34μΑ<br>(AC SFC 264V 60 Hz) | 34μΑ<br>(AC SFC 264V 60 Hz) | 16.0μΑ<br>(AC SFC 264V 60 Hz) | 78 µA (Less than 500 μΑ) | | |
| Number of Output Channels | 4 | 2 | 1 | 2 | The number of<br>channels of output<br>current do not affect<br>the treatment, and<br>the design of subject<br>device comply with<br>IEC60601-2-10 for<br>performance<br>requirement | |
| Number of Output Modes | 2 | 2 | 2 | 9 (one less output) | Similar:<br>Two output modes of<br>subject device are<br>basically like two<br>outputs of predicate | |
| | | | | | | |
| | | | | | device (Russian and<br>Interferential). The<br>minor difference does<br>not raise different<br>question of safety and<br>effectiveness | |
| Synchronous or alternating? | Alternating | Alternating | N/A (1 channel only) | Synchronous | Different:<br>The subject device will<br>not raise any safety<br>concerns of output<br>current by alternating<br>stimulation mode, and<br>the subject device<br>complies with<br>IEC60601-2-10 for this<br>performance<br>requirement. The<br>difference does not<br>raise different<br>questions of safety<br>and effectiveness | |
| Method of Channel Isolation | Transformer isolated | Transformer isolated | Transformer isolated | Not Available | - | |
| Regulated Current or Regulated<br>Voltage? | Regulated Voltage (CV) | Regulated Voltage<br>(CV) | Regulated Voltage (CV) | Not Available | The subject device<br>complies with<br>IEC60601-2-10 for<br>performance<br>requirement. So, such<br>a minor difference<br>does not raise<br>different question of<br>safety and<br>effectiveness. | |
| Software/Firmware/Microprocesso<br>r Control? | yes | yes | yes | yes | Same | |
| Automatic Over current Trip?<br>Yes/no | yes | yes | yes | yes | Same | |
| Automatic Overload Trip? Yes/no | yes | yes | yes | yes | Same | |
| Automatic No-Load Trip? Yes/no | yes | yes | yes | yes | Same | |
| Automatic Shut Off? Yes/no | | yes | yes | yes | yes | Same |
| Patient Override Control? Yes/no | | yes | yes | yes | yes | Same |
| Patient Override Control method | | Removal of patient<br>connector at the<br>device. | Removal of patient<br>connector at the<br>device. | Removal of patient<br>connector at the<br>device. | Patient interrupt<br>switch | Different:<br>The subject device<br>does not have a<br>patient interrupt<br>switch like the<br>predicate device. The<br>subject device has an<br>automatic shutdown<br>function and an<br>overload protection<br>function that can<br>monitor the<br>unintended change in<br>the intensity of<br>current that can<br>prevent the patient<br>from unexpected<br>conditions. the<br>difference would not<br>affect the safety and<br>effectiveness of the<br>subject device. |
| Indicator<br>Display | On/Off Status?<br>Yes/ No | Yes | Yes | Yes | Same | |
| | Low Battery<br>Yes/No | No | No | No | Different:<br>The subject device<br>does not use battery<br>for its operation. This<br>eliminates the risks<br>associated with that<br>function. Therefore,<br>the subject device will<br>not have any safety or<br>effectiveness concerns | |
| | Voltage/Current<br>Level? | Yes | Yes | Yes | associated with this<br>function | |
| Timer Range | | 1 – 90 min | 1 – 90 min | 1 – 90 min | 0-60 min | Similar:<br>The timer range in the<br>subject device is<br>within the range or<br>slightly above the<br>range of the predicate<br>device. The design of<br>the timer range is<br>based on the intended<br>use. the operating<br>time is adjustable by<br>the operator<br>according to<br>physician's direction.<br>So, the difference in<br>timer setting range<br>would not impact its<br>safety and<br>effectiveness<br>compared to the<br>predicate devices |
| Compliance with 21 CFR 898? | | Yes | Yes | Yes | yes | Same |
| Compliance with voluntary<br>standards<br>• IEC60601-1<br>• IEC60601-1-2<br>• IEC60601-2-10<br>• ISO14971<br>• IEC62304<br>• IEC62366 | | Yes | Yes | Yes | Information not<br>available | Same |
| Weight (lbs., oz,) | | 24.2508 lbs. without<br>accessories | 24.2508 lbs. without<br>accessories | 5.5 lbs. without<br>accessories | 4.5 lb. | Different<br>There are minor<br>differences between |
| | | | | | | |
| Dimensions (in.) [W x H x D] | 13.77" x 11.41" x 13.77" | 13.77" x 11.41" x<br>13.77" | 10.6" x 9.64" x 2.36" | 8" x 8" x13" | the subject device and<br>the predicate devices<br>in device weight. But<br>the subject device<br>complies with<br>IEC60601-1, and<br>IEC60601-2-10<br>Standard<br>requirement. Different<br>design criteria will<br>have no effect on its<br>function and does not<br>raise safety and<br>performance concerns<br>Different<br>There are minor<br>differences between<br>the subject device and<br>the predicate devices<br>in device weight. But<br>the subject device<br>complies with<br>IEC60601-1, and<br>IEC60601-2-10<br>Standard<br>requirement. Different<br>dimensions will have<br>no effect on its<br>function and does not<br>raise safety and<br>performance concerns | |
| Environment of use…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.