EnCor EnCompass™ Breast Biopsy and Tissue Removal System
K252681 · Bard Peripheral Vascular, Inc. · KNW · Dec 12, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K252681
Device Name
EnCor EnCompass™ Breast Biopsy and Tissue Removal System
Applicant
Bard Peripheral Vascular, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Dec 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The EnCor EnCompass™ Breast Biopsy and Tissue Removal System is indicated to acquire breast tissue for histologic examination with partial or complete removal of the abnormality.
Device Story
Console-based vacuum-assisted breast biopsy system; used for diagnostic/histologic tissue sampling. System comprises console, handheld driver, foot switch, and disposable probes. Operates via vacuum-assisted suction to draw tissue into probe cutter; saline lavage optional. Used in clinical settings (ultrasound, stereotactic, or MRI suites) by physicians. Console provides user interface/touchscreen; driver/foot switch control procedure. Output is retrieved tissue samples for pathology. Benefits include minimally invasive tissue removal and diagnostic accuracy. Wireless connectivity included for foot switch. System supports marker placement via adapter.
Clinical Evidence
Bench testing only. No clinical data presented. Performance validated via non-clinical testing including mechanical/tensile testing, tissue penetration, sample mass/contiguity, needle throw, imaging mount tests, lavage, biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601), human factors (IEC 62366-1), software/cybersecurity (IEC 62304), and MR safety (ASTM F2503).
Indicated for patients requiring breast tissue acquisition for histologic examination and partial or complete removal of breast abnormalities. Compatible with ultrasound, stereotactic, and MR imaging guidance.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Marquee™ Disposable Core Biopsy Instrument and Instrument Kit (K250032)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 12, 2025
Bard Peripheral Vascular, Inc.
Jessica Meyer
Regulatory Affairs Specialist
1625 West 3rd Street
Tempe, Arizona 85281
Re: K252681
Trade/Device Name: EnCor EnCompass™ Breast Biopsy and Tissue Removal System
Regulation Number: 21 CFR 876.1075
Regulation Name: Gastroenterology-Urology Biopsy Instrument
Regulatory Class: Class II
Product Code: KNW
Dated: August 25, 2025
Received: August 25, 2025
Dear Jessica Meyer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K252681 - Jessica Meyer
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K252681 - Jessica Meyer
Page 3
Sincerely,
JESSICA CARR -S
Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
EnCor EnCompass™ Breast Biopsy and Tissue Removal System
Page 13 of 89
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252681 | ? |
| Please provide the device trade name(s). | | ? |
| EnCor EnCompass™ Breast Biopsy and Tissue Removal System | | |
| Please provide your Indications for Use below. | | ? |
| The EnCor EnCompass™ Breast Biopsy and Tissue Removal System is indicated to acquire breast tissue
for histologic examination with partial or complete removal of the abnormality. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
BD
BAIRD
BERKELEY
EnCor EnCompass™ Breast Biopsy and Tissue Removal System
510(k) Summary - K252681
21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
## Submitter Information:
Applicant: Bard Peripheral Vascular, Inc.
1625 West 3rd Street
Tempe, Arizona 85281
Phone: (602) 830-5070
Fax: (312) 949-0436
Contact Person: Jessica Meyer, Regulatory Affairs Specialist
Date of Submission: August 25, 2025
## Subject Device:
Name of Device: EnCor EnCompass™ Breast Biopsy and Tissue Removal System
Common or Usual Name: Instrument, Biopsy (Product Code: KNW)
Classification Panel: General & Plastic Surgery
Regulatory Class: II
Regulation Number: 876.1075
## Predicate Device:
EnCor Enspire™ Breast Biopsy System (K233220, cleared on October 27, 2023)
## Reference Device:
Brevera® Breast Biopsy System with Corlumina Imaging Technology (K163052)
ATEC® Breast Biopsy System (K042290)
Marquee™ Disposable Core Biopsy Instrument and Instrument Kit (K250032)
## Device Description:
The subject device, the EnCor EnCompass™ Breast Biopsy and Tissue Removal System, is a console-based vacuum-assisted breast biopsy device. It is intended to obtain breast tissue samples for diagnostic or histologic analysis and may be used under Ultrasound, Stereotactic, or MR imaging guidance. The system consists of a console, handheld driver, foot switch, biopsy probes, canister liner bags, MRI extension tubing, a stereotactic adapter, MRI introducer sets,
Bard Peripheral Vascular, Inc.
Page 1 of 5
{6}
BD
BAORD
BAYRID
D
D
D
D
and spare sample containers. Replacement parts and components of the console, driver, and foot switch that are also sold separately include a reusable vacuum canister, exhaust filter, driver extension cable, foot switch cable, and power cable.
The EnCor EnCompass™ Breast Biopsy and Tissue Removal System consists of multiple devices that are intended to be used together to obtain breast tissue samples. This includes the EnCor EnCompass™ Console, EnCor EnCompass™ Driver, and EnCor EnCompass™ Foot Switch. The console, driver, and foot switch are designed to provide a user interface and to accept inputs used for specific procedures. The console contains a vacuum system and touchscreen monitor. The driver contains multiple buttons to control procedure functionality and is compatible for use with Ultrasound, Stereotactic, and Magnetic Resonance Imaging (MRI) procedures. The foot switch also contains buttons and a pedal to control procedure functionality and has both wired and wireless connectivity options. They are intended to be used with EnCor EnCompass™ Probes, Canister Liner Bags, MRI Extension Tubing, VisiLoc™ MRI Introducer Sets, and Stereotactic Adapter, which are sold separately. The Probes are the Applied Part of the EnCor EnCompass™ Breast Biopsy and Tissue Removal System.
The EnCor EnCompass™ Probe contains a tubular cutter for tissue acquisition and either a sharp trocar tip at the distal end for insertion or a blunt tip at the distal end for use in areas with sensitive structures. A detachable tissue collection chamber is located on the proximal end of the device. The tissue collection chamber incorporates both a suction line for acquisition and transport of tissue samples and a saline line for optional lavage functionality. A Y-valve luer lock is provided for introducing anesthetic to the biopsy site. Compatible breast tissue markers which are sold and packaged separately may be inserted through the proximal end of the device.
The integrated tubing cassette provides vacuum to an EnCor EnCompass™ Probe during a biopsy procedure.
The EnCor EnCompass™ Probe sample container collects samples as they are acquired and transported during the biopsy procedure. The basket allows for drainage of fluid that is acquired during sampling, vacuuming, and lavaging of the cavity. The basket allows visualization of the samples during and after the procedure and is detachable for sample retrieval post procedure.
The marker adapter allows for marker placement using a marker deployment device through the back of the probe.
The EnCor EnCompass™ Canister Liner Bag collects fluid waste generated during the procedure.
The EnCor EnCompass™ Stereotactic Adapter provides a mechanical interface between the combined EnCor EnCompass™ Driver and EnCor EnCompass™ Probe and the stereotactic biopsy table. The EnCor EnCompass™ Stereotactic Adapter is reusable and affixes directly to the EnCor EnCompass™ Probe with a T-shape fastener and two pins and the EnCor EnCompass™ Driver with rear sliding latches.
**Indications for Use:**
The EnCor EnCompass™ Breast Biopsy and Tissue Removal System is indicated to acquire breast tissue for histologic examination with partial or complete removal of the abnormality.
Bard Peripheral Vascular, Inc.
Page 2 of 5
{7}
BD
BAORD
BAYRID
D
# Technological Comparison to Predicate Device:
The EnCor EnCompass™ Breast Biopsy and Tissue Removal System has the following similarities to the predicate devices, the EnCor Enspire™ Breast Biopsy System, Drivers, and Probes (K233220, cleared on October 27, 2023).
- Same intended use
- Same indications for use
- Same target patient population
- Same principle of operation
- Same fundamental scientific technology
- Same imaging compatibilities
- Same sterility level and method of sterilization
- Same general packaging configuration
All components of the subject device are of similar design and function as the components of the predicate device. The changes between the subject device and the predicate device(s) include:
- Enhanced user interface
- Enhanced sampling options
- Reduced inventory requirements
- Addition of wireless technology
- Patient contacting materials
# Performance Data:
To demonstrate substantial equivalence of the subject device, the EnCor EnCompass™ Breast Biopsy and Tissue Removal System, to the predicate device, both technological characteristics and performance criteria were evaluated. Testing was conducted based on recommendations in the following FDA Guidance Documents on non-clinical testing of medical devices, recognized consensus standards, and internal risk assessment procedures:
- ISO 10993 Series
- IEC 60601 Series
- IEC 62366-1
- IEC 62304
- ASTM F2096
- ASTM F2503
Bard Peripheral Vascular, Inc.
Page 3 of 5
{8}
BD
BAARD
BAYRID
D
- AAMI TIR12
- ANSI AAMI ST98
- “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process” (September 2023)
- “Applying Human Factors and Usability Engineering to Medical Devices” (February 3, 2016)
- “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling” (March 17, 2015)
- “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment” (October 10, 2023)
- “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” (September 27, 2023)
- “Electromagnetic Compatibility of Medical Devices” (June 6, 2022)
The following tests were performed:
- Performance Testing
- Reliability
- Mechanical/Tensile Testing
- Tissue Penetration
- Sample Mass and Contiguity
- Needle Throw Length
- Stereotactic Imaging Mount Test
- Lavage
- Design Validation
- Packaging Validation
- Sterilization Validation
- Biocompatibility
- Human Factors Evaluation
- Electrical Safety and EMC
- Reprocessing Validation
- Software Testing
- Cybersecurity Testing
- Wireless Testing
- MR Testing
Bard Peripheral Vascular, Inc.
Page 4 of 5
{9}
BD
BAORD
BBS
The results from these tests demonstrate that the technological characteristics and performance criteria of the EnCor EnCompass™ Breast Biopsy and Tissue Removal System are comparable to the predicate device, that it can perform in a manner equivalent to devices currently on the market for the same intended use, and that there are no concerns of safety or effectiveness related to the changes from the predicate device.
## Conclusions:
The subject device, the EnCor EnCompass™ Breast Biopsy and Tissue Removal System, and the predicate device share the same or similar characteristics: intended use, indications for use, target patient population, principle of operation, fundamental scientific technology, imaging compatibility, sterility level and method of sterilization, and general packaging configuration. Additionally, the EnCor EnCompass™ Breast Biopsy and Tissue Removal System met predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. Therefore, BD concludes that the subject device, the EnCor EnCompass™ Breast Biopsy and Tissue Removal System, is substantially equivalent to the legally marketed predicate device, the EnCor Enspire™ Breast Biopsy System.
Bard Peripheral Vascular, Inc.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.