Brevera Biopsy System with Corlumina Imaging Technology
K163052 · Hologic, Inc. · KNW · Mar 29, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K163052
Device Name
Brevera Biopsy System with Corlumina Imaging Technology
Applicant
Hologic, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Mar 29, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Hologic Brevera breast biopsy system with CorLumina imaging technology is intended to provide breast tissue samples for diagnostic sampling of breast abnormalities. The Brevera breast biopsy system excises and optionally delivers in-line radiographic images of the excised tissue. The Brevera breast biopsy system is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (for example, fibrocystic lesion), the Brevera breast biopsy system may also be used to partially remove such palpable lesions. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of histological abnormality, for example, malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedure.
Device Story
Brevera Biopsy System with CorLumina Imaging Technology combines vacuum-assisted breast biopsy (VAB) and cabinet X-ray imaging. System inputs: breast tissue samples acquired via minimally invasive handpiece. Operation: vacuum-assisted handpiece excises tissue; transports samples through tubing to console collection tray; integrated cabinet X-ray subsystem images tissue in-line. Output: radiographic images of excised tissue displayed on screen for physician review. Used in clinical settings with stereotactic biopsy guidance; operated by physicians. Benefits: enables real-time visual confirmation of calcifications (as small as 240um) at point of care; ensures suspect lesion collection; improves biopsy workflow efficiency.
Clinical Evidence
Bench testing only. Included functional testing of biopsy specimen acquisition, core handling, and X-ray generation. Animal model testing demonstrated successful tissue biopsy and transport. Biocompatibility testing per ISO 10993-1 and G95-1; sterilization testing per ISO 11737-1/2. Usability and software integration testing confirmed system performance met all acceptance criteria, including recognition of calcifications as small as 240um.
Technological Characteristics
Vacuum-assisted biopsy handpiece and console with integrated cabinet X-ray imaging. Materials meet biocompatibility requirements. Connectivity: integrated electro-mechanical system. Sterilization: disposable components validated per ISO 11737. Software: moderate level of concern, manages biopsy cycle timing, vacuum suction, and radiographic image acquisition/display.
Indications for Use
Indicated for diagnostic sampling of breast abnormalities in patients requiring breast tissue for histologic examination; includes partial or complete removal of imaged abnormalities and partial removal of palpable lesions classified as benign.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
ATEC handpiece and ATEC Sapphire console (K042290)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 29, 2017
Hologic, Inc. Ms. Sarah Fairfield Manager, Regulatory Affairs 250 Campus Drive Marlborough, Massachusetts 01752
Re: K163052
Trade/Device Name: Brevera Biopsy System with Corlumina Imaging Technology Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: February 24, 2017 Received: February 27, 2017
Dear Ms. Fairfield:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163052
### Device Name
Brevera Breast Biopsy System with Corlumina Imaging Technology
### Indications for Use (Describe)
The Hologic Brevera breast biopsy system with CorLumina imaging technology is intended to provide breast tissue samples for diagnostic sampling of breast abnormalities. The Brevera breast biopsy system excises and optionally delivers in-line radiographic images of the excised tissue. The Brevera breast biopsy system is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (for example, fibrocystic lesion), the Brevera breast biopsy system may also be used to partially remove such palpable lesions. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of histological abnormality, for example, malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedure.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
#### 1. Date: March 27 2017
#### 2. Submitter:
Hologic, Inc. 250 Campus Dr. Marlborough, MA 01752
Sarah Fairfield, Manager, Regulatory Affairs, RAC Contact: B.S. Biomedical Engineering Email: sarah.fairfield@hologic.com Telephone: 508.263.8857
#### 3. Device:
Trade Name: Brevera Biopsy System with Corlumina Imaging Technology Common Name: Biopsy instrument Classification Name: Gastroenterology-urology biopsy instrument Classification Regulation: 21 CFR 876.1075 Class: II
#### 4. Predicate Device:
Tradename: ATEC handpiece and ATEC Sapphire console Submitter / 510(k) Holder: Hologic, Inc. 510(k) #'s: K042290
Tradename: Trident Specimen Radiography System Submitter / 510(k) Holder: Hologic, Inc. 510(k) #'s: K111508
#### ട്. Device Description:
The Brevera breast biopsy system with CorLumina imaging technology enables a physician to utilize vacuum-assisted breast tissue biopsy technology and cabinet x-ray technology to remove suspicious lesions and confirm the suspect lesion has been collected at the point of care. The vacuum assisted breast biopsy subsystem collects breast tissue and transports it through tubing into a collection tray located in the console. Within the console, the in-line cabinet x-ray subsystem then allows the user to visually confirm calcifications that may be present in the collected breast tissue. The overall system is intended to be used with a stereotactic biopsy guidance system
#### 6. Intended Use:
The Hologic Brevera breast biopsy system with CorLumina imaging technology is intended to provide breast tissue samples for diagnostic
{5}------------------------------------------------
sampling of breast abnormalities. The Brevera breast biopsy system excises targeted tissue and optionally delivers in-line radiographic images of the excised tissue. The Brevera breast biopsy system is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (for example, fibroadenoma, fibrocystic lesion), the Brevera breast biopsy system may also be used to partially remove such palpable lesions. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of histological abnormality, for example, malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedure.
#### 7. Comparison of Characteristics:
The method of use, mechanism of action, mode of operation and functional performance of the Brevera Biopsy System with CorLumina Imaging Technology is similar to the ATEC breast biopsy system, as it relates to vacuum-assisted biopsy, and is similar to the Trident specimen radiography system as it relates to in line specimen imaging of the acquired breast biopsy sample. The overall workflow for the Brevera Biopsy System with CorLumina Imaging Technology is similar to the ATEC breast biopsy system, for the biopsy portion of the workflow, and is similar to the Trident specimen radiography system for the specimen imaging portion of the workflow. The following table contains a comparison summary of the subject device to the predicate devices.
| Characteristic | ATEC Breast<br>Biopsy System<br>(K042290,<br>PREDICATE<br>DEVICE)) | Trident<br>Specimen<br>Radiography<br>System<br>(K111508,<br>PREDICATE<br>DEVICE) | Brevera Biopsy<br>System<br>(SUBJECT<br>DEVICE) | Comparison to<br>Predicates |
|-----------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------|
| Major<br>Components | Minimally<br>invasive<br>biopsy<br>handpiece and<br>console. | Imaging system. | Minimally invasive<br>biopsy handpiece,<br>console, and<br>specimen imaging<br>system. | Substantially<br>Equivalent to<br>predicates |
| Technology:<br>Biopsy | Vacuum-<br>assisted biopsy<br>(VAB) device | N/A | Vacuum-assisted<br>biopsy (VAB)<br>device | Substantially<br>Equivalent to<br>predicates |
{6}------------------------------------------------
| Characteristic | ATEC Breast<br>Biopsy System<br>(K042290,<br>PREDICATE<br>DEVICE)) | Trident<br>Specimen<br>Radiography<br>System<br>(K111508,<br>PREDICATE<br>DEVICE) | Brevera Biopsy<br>System<br>(SUBJECT<br>DEVICE) | Comparison to<br>Predicates |
|------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------|----------------------------------------------|
| Technology:<br>Imaging | N/A | Radiographic | Radiographic | Substantially<br>Equivalent to<br>predicates |
#### 8. Performance Testing:
Performance testing applicable to the Brevera Biopsy system with Corlumina Imaging Technology was determined based on an analysis of risks and hazards posed by the intended use of the device and the proposed modifications that the subject device has over the predicate devices. The scope of the analysis included the principal components that the system are made up of that relate to a vacuum assisted biopsy and x-ray cabinet specimen radiography. Specifically included features are:
- ·Biopsy specimen acquisition and core handling
- ·X-ray generation for imaging
- •System's performance when used in live tissue with an Animal Model
- •User control of system operation
| Performance Testing Overview | | |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Modifications | Test Method | Results |
| Specimen<br>acquisition; two<br>component<br>handpiece | Confirm the device fires<br>and arms when<br>instructed. | The Gen 2 Biopsy System was tested, and it was<br>demonstrated that the device has an inner cannula that<br>rotates and advances to cut tissue.<br>All Acceptance Criteria Met |
| | Confirm that a core is<br>sucked in to the aperture<br>and the inner cannula<br>cuts the tissue. | The Brevera System was tested and it was<br>demonstrated that the average cycle time for the<br>acquisition of each core is less than or equal to 8<br>seconds for all cycles except for the first and last cycle.<br>All Acceptance Criteria Met |
| | To confirm that the core<br>travels the correct<br>pathway to the tissue<br>filter and travels in a<br>consistent time. | The Brevera System was tested and it was<br>demonstrated that the average cycle time for the<br>acquisition of each core is less than or equal to 8<br>seconds for all cycles except for the first and last cycle.<br>All Acceptance Criteria Met |
| Performance Testing Overview | | |
| Modifications | Test Method | Results |
| | To confirm that the core<br>is the intended weight. | The Brevera System was tested against the ATEC<br>Biopsy System and was found that core weight is<br>equivalent or larger than the predicate device (ATEC)<br>with the same size needle and aperture.<br>All Acceptance Criteria Met |
| Integrated<br>Specimen<br>collection | Confirm that the device<br>pulls saline consistently<br>into the cannulas and to<br>the body cavity. | The Brevera Disposable was flow tested and all parts<br>demonstrated that the system delivered 0.9% buffered<br>saline into the aperture as well as not leak fluid when<br>exposed to twice the volume of a saline bag (500 mL).<br>All Acceptance Criteria Met |
| | To confirm that the<br>console supplies suction<br>to pull the core and any<br>excess liquid to the<br>canister. | The Brevera biopsy System was tested, and it was<br>demonstrated that core specimens are transported into<br>a tissue filter via vacuum and that waste fluids are<br>collected in a disposable vacuum canister |
| | Confirm the console | All Acceptance Criteria Met |
| | responds to all<br>commands (Setup and<br>Test Mode) | The Brevera System software performs as intended.<br>All Acceptance Criteria Met |
| Integrated<br>Specimen<br>imaging | Confirm that the x-ray<br>takes an image of the<br>core at the correct time<br>and populates the image<br>to the screen | The Brevera System was tested and it was<br>demonstrated that it allows collection and imaging of<br>up to two cores in each filter chamber.<br>The Brevera System was tested and it was<br>demonstrated that a user can visually recognize<br>calcifications without obstruction by the tissue.<br>All Acceptance Criteria Met |
| | Visually recognize<br>calcifications as small as<br>240um from the<br>radiographic images | The Brevera System was tested and demonstrated the<br>system enables a user to visually recognize<br>calcifications as small as 240um from the radiographic<br>images based on predicate (Trident) imaging<br>capability.<br>All Acceptance Criteria Met |
| Disposable/sterile<br>biopsy handpiece<br>component | Testing per ISO 11737-1 &<br>11737-2 and ASTM F1980-07 | Disposable components maintained sterile barrier<br>integrity after simulated aging conditions.<br>All Acceptance Criteria Met |
| | Testing per ISO-10993-1<br>and G95-1. | Disposable components met required biocompatibility<br>requirements.<br>All Acceptance Criteria Met |
| Performance Testing Overview | | |
| Modifications | Test Method | Results |
| System performance:<br>Software for both<br>biopsy and imaging | Integration testing of<br>electro-mechanical<br>system | The Brevera System provides correct power to the<br>Corlumina Imaging Technology and Vacuum Assisted<br>Biopsy subsystems to allow for successful integration<br>of core collection and imaging of core specimens.<br>All Acceptance Criteria Met |
| | Monitoring and<br>controlling of<br>performance<br>requirements | The Brevera System software performs as intended.<br>All Acceptance Criteria Met |
{7}------------------------------------------------
{8}------------------------------------------------
The Brevera Biopsy System with Corlumina Imaging Technology successfully performed system design control verification and validation tests, which are summarized in accordance with FDA's Guidance for the Content of premarket Submissions for Software Contained in Medical Devices (issued on May 11, 2005) based on a moderate level of concern. Animal testing in a live model was also completed to demonstrate the system's ability to successfully biopsy and transport tissue to the tissue filter. Additional bench testing including functional testing, biocompatibility, sterilization, package testing, and usability testing was also performed on the Brevera system. The comparative and other performance testing showed that the overall system demonstrated equivalent performance, and showed to be as safe and effective as the predicate devices, and met all acceptance criteria.
#### 9. Conclusion:
Based on the intended use, descriptive information and performance testing provided in this submission, the Brevera Biopsy System with Corlumina Imaging Technology has been shown to be equivalent in technology, method of operation, functional performance and intended use to the predicate ATEC Breast Biopsy System(K042290) and Trident Specimen Radiography System(K111508).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.