Vanquish Water Vapor Ablation Device

K252388 · Francis Medical, Inc. · KNS · Nov 26, 2025 · Gastroenterology, Urology

Device Facts

Record IDK252388
Device NameVanquish Water Vapor Ablation Device
ApplicantFrancis Medical, Inc.
Product CodeKNS · Gastroenterology, Urology
Decision DateNov 26, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Vanquish Water Vapor Ablation System is indicated for the thermal ablation of targeted prostate tissue via a transurethral approach.

Device Story

Vanquish Water Vapor Ablation System provides focal thermal ablation of prostate tissue using water vapor energy. Procedure performed in outpatient setting under transrectal ultrasound (TRUS) guidance. System heats sterile water to generate vapor, delivered via transurethral needle with emitter holes into prostate tissue. Includes stabilization and needle guidance systems. Physician operates device to target tissue; thermal energy causes ablation. Output is ablated prostate tissue; clinical benefit is treatment of prostate cancer. Healthcare provider uses TRUS to visualize placement and monitor procedure. Clinical decision-making relies on real-time imaging and device performance to ensure targeted ablation.

Clinical Evidence

Prospective, single-arm pivotal study (VAPOR 2, NCT05683691) of 235 patients; interim analysis of 110 intermediate-risk prostate cancer patients. Primary endpoints: 74% negative biopsy rate at 6 months (95% CI: 64%-82%); 21% average prostate volume reduction at 6 months; 52% PSA reduction at 12 months. Safety: No UADEs or serious device-related AEs; 94% of AEs non-serious; 96% of device/procedure-related AEs mild to moderate (CTCAE 1 or 2).

Technological Characteristics

Transurethral ablation device using water vapor energy. Features include stabilization system and needle guidance system. Bench testing included functional performance, biocompatibility, software V&V, cybersecurity, sterilization validation, electrical safety, and EMC. Complies with ISO 14155:2020.

Indications for Use

Indicated for thermal ablation of targeted prostate tissue via transurethral approach in patients with intermediate-risk prostate cancer.

Regulatory Classification

Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION November 26, 2025 Francis Medical, Inc. % Alexia Haralambous Director, Regulatory Affairs RQM+ 2790 Mosside Blvd, #800 Monroeville, Pennsylvania 15146 Re: K252388 Trade/Device Name: Vanquish Water Vapor Ablation Device Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and accessories Regulatory Class: II Product Code: KNS Dated: July 31, 2025 Received: October 29, 2025 Dear Alexia Haralambous: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K252388 - Alexia Haralambous Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {3} K252388 - Alexia Haralambous Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MARK J. ANTONINO -S Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} Vanquish Water Vapor Ablation Device Page 9 of 53 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252388 | ? | | Please provide the device trade name(s). | | ? | | Vanquish Water Vapor Ablation Device | | | | Please provide your Indications for Use below. | | ? | | The Vanquish Water Vapor Ablation System is indicated for the thermal ablation of targeted prostate tissue via a transurethral approach. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {5} K252388, Page 1 of 2 | 510(k) #: K252388 | 510(k) Summary | Prepared on: 2025-11-26 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | Francis Medical, Inc. | | | Applicant Address | 7351 Kirkwood Ln N, Suite 130 Maple Grove MN 55369 United States | | | Applicant Contact Telephone | 612-910-9790 | | | Applicant Contact | Mr. Mike Kujak | | | Applicant Contact Email | mkujak@francismedical.com | | | Correspondent Name | RQM+ | | | Correspondent Address | 2790 Mosside Blvd, #800 Monroeville PA 15146 United States | | | Correspondent Contact Telephone | 267-778-5114 | | | Correspondent Contact | Ms. Alexia Haralambous | | | Correspondent Contact Email | aharalambous@rqmplus.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | Vanquish Water Vapor Ablation Device | | | Common Name | Endoscopic electrosurgical unit and accessories | | | Classification Name | Unit, electrosurgical, endoscopic (with or without accessories) | | | Regulation Number | 876.4300 | | | Product Code(s) | KNS | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K150786 | Rezūm System | KNS | | Device Description Summary | | 21 CFR 807.92(a)(4) | | The Vanquish Water Vapor Ablation System provides a method of focal thermal ablation that utilizes the high energy stored in water vapor to target prostate tissue. In a minimally invasive transurethral outpatient procedure performed under transrectal ultrasound (TRUS) guidance, sterile water is heated within the system and converted into vapor. This vapor is then delivered through small emitter holes at the distal end of a transurethral needle into the prostate tissue. | | | | Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | | The Vanquish Water Vapor Ablation System is indicated for the thermal ablation of targeted prostate tissue via a transurethral approach. | | | {6} K252388, Page 2 of 2 | Indications for Use Comparison | 21 CFR 807.92(a)(5) | | --- | --- | | Both the subject and predicate devices are intended for transurethral ablation of prostate tissue. Therefore, the devices have the same intended use. | | | Technological Comparison | 21 CFR 807.92(a)(6) | | Both the subject and primary predicate device are transurethral ablation devices for prostate tissue which use water vapor as the source of ablation energy. Both are intended for partial gland ablation using the same delivery mechanism. However, they slightly differ in ablation energy levels, and the subject Vanquish device includes a stabilization system and a needle guidance system which the predicate Rezūm device does not have. Overall, the principle of operation, mechanism of action, and conditions of use between the subject Vanquish and predicate Rezūm devices are identical. | | | Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) | | Francis Medical conducted functional performance bench testing, biocompatibility testing, software verification & validation, cybersecurity testing, sterilization validation, electrical safety and electromagnetic compatibility testing on the Vanquish system and components and provided the results as part of this submission to support the proposed indications for use. The performance of the Vanquish System is supported by the manufacturer sponsored VAPOR 2 (NCT05683691) pivotal study, a prospective, single-arm clinical study of an intermediate risk prostate cancer patient population to evaluate the safety and effectiveness of the Vanquish System to ablate targeted prostate tissue per the FDA Guidance Clinical Investigations for Prostate Tissue Ablation Devices. The VAPOR 2 study is being conducted in compliance with ISO 14155:2020 Clinical investigation of medical devices for human subjects – Good clinical practice. A total of 235 patients were treated with the Vanquish System at 26 sites in the United States from July 2023 to February 2025. An interim analysis of the first 110 patients treated with the Vanquish System followed to one year supports safety and effectiveness performance for this 510(k) submission. The VAPOR 2 study interim effectiveness endpoints per FDA Guidance are reported as follows using an intent-to-treat (ITT) approach in 110 focally treated subjects: • Biopsy Results: with worst case imputation, 74% of subjects (81/110) show negative biopsy at six months in targeted treatment field (two-sided exact 95% confidence interval: 64% to 82%). • Prostate Volume Reduction: average prostate volume reduction at 6-months for 110 subjects as measured by MRI is 21% (±13% standard deviation) • PSA Reduction for 108 subjects at 12-months is 52% (±28% standard deviation) The VAPOR 2 study interim safety endpoints per FDA Guidance are reported in the clinical study report attached to this submission. Safety and effectiveness data were collected at enrollment, Vanquish treatment, and follow-up visits at 7, 30, 90, 180, and 365 days. Subjects are being assessed for adverse events throughout the study. The VAPOR 2 study interim safety results are summarized as follows for 110 focally treated subjects: • No unanticipated adverse device effects (UADE) were reported, • No serious device-related adverse events were reported, • Nearly all (94%) reported adverse events were non-serious, and • Nearly all (96%) device and / or procedure-related adverse events were mild to moderate (CTCAE 1 or 2). The VAPOR 2 study confirms that the Vanquish System is safe and effective for the ablation of targeted prostate tissue. The use of the proposed device does not result in any new concerns about safety or effectiveness compared to the predicate device. Based on the intended use, technological characteristics, and the non-clinical and clinical evidence presented in this submission, the Vanquish System is substantially equivalent to the predicate device. | |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...