Insufflator (OPTO-IFL1000)

K252318 · Guangdong Optomedic Technologies, Inc. · HIF · Apr 16, 2026 · Obstetrics/Gynecology

Device Facts

Record IDK252318
Device NameInsufflator (OPTO-IFL1000)
ApplicantGuangdong Optomedic Technologies, Inc.
Product CodeHIF · Obstetrics/Gynecology
Decision DateApr 16, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1730
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures.

Device Story

The Insufflator (OPTO-IFL1000) is a CO2 insufflation device used in clinical settings to create and maintain pneumoperitoneum for laparoscopic surgery. It connects to a CO2 gas source and delivers gas through specialized insufflation tubes (with or without heating) into the abdominal cavity. The device monitors intra-abdominal pressure and flow rates, automatically adjusting gas delivery to maintain a set pressure and compensate for leaks. It features multiple operational modes (Pediatric, Adult, Obese, Veress) with specific pressure and flow limits. The device provides visual and audible alarms for overpressure, gas supply depletion, overheating, and fluid contamination. Healthcare providers use the device interface to set parameters and monitor real-time pressure/flow data. By maintaining a stable surgical field of view and operating space, the device facilitates laparoscopic access for surgeons, potentially reducing procedural complexity and improving patient outcomes.

Clinical Evidence

No clinical data. Bench testing only. Verification included electromagnetic compatibility (IEC 60601-1-2:2020) and performance testing of gas supply, pressure/flow accuracy, leakage compensation, overpressure/overheating alarms, and heating function. Software verification and validation were performed per FDA guidance for 'Enhanced' documentation level.

Technological Characteristics

CO2 insufflator; 110-240V power; dimensions 370x410x166.5mm; weight 11kg. Features include heating function (OPTO-T1000H tubing), pressure range 1-30 mmHg, flow rate 0.1-50 L/min. Software-controlled with multiple modes (Pediatric, Adult, Obese, Veress). Sterilization: Steam sterilization for insufflation tubes. Connectivity: Standalone device.

Indications for Use

Indicated for patients undergoing diagnostic or therapeutic laparoscopic procedures requiring pneumoperitoneum via abdominal gas distension.

Regulatory Classification

Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD &amp; DRUG ADMINISTRATION April 16, 2026 Guangdong OptoMedic Technologies, Inc. Weijuan Guo Regulatory Affairs Engineer Suite 503, Building A, Golden Valley Intellicreation Community, No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan, Guangdong 528200 CHINA Re: K252318 Trade/Device Name: Insufflator (OPTO-IFL1000) Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF Dated: July 23, 2025 Received: July 25, 2025 Dear Weijuan Guo: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K252318 - Weijuan Guo Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {3} K252318 - Weijuan Guo Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JASON ROBERTS -S Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252318 | ? | | Please provide the device trade name(s). | | ? | | Insufflator (OPTO-IFL1000) | | | | Please provide your Indications for Use below. | | ? | | The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {5} OptoMedic Traditional 510(k) Premarket Notification K252318 Page 1 of 5 # 510(k) Summary Prepared in accordance with the requirements of 21 CFR Part 807.92 Date Prepared: April 15, 2026, ## I. General Information 510(k) Submitter/Owner: Guangdong OptoMedic Technologies, Inc. Suite 503, Building A, Golden Valley Intellicreation Community, No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan, Guangdong, 528200, P.R. China Establishment Registration Number: Not yet registered Primary Contact Person: Weijuan Guo Regulatory Affairs Engineer Tel: +86 (757) 8670 2920 Email: guoweijuan@optomedic.com Alternative Contact Person: Minghua Wu Regulatory Affairs Engineer Tel: +86 (757) 8670 2920 Email: Minghua Wu @optomedic.com ## II. Device Identification Device Trade Name: Insufflator (OPTO-IFL1000) Common or Usual Name: Insufflator Model: OPTO-IFL1000 Regulation Name: Laparoscopic Insufflator Regulation Number: 21 CFR 884.1730 Regulatory Class: Class II Product Code: HIF ## III. Predicate Device 510(k) Number: K222812 Product Name: Insufflator (SL102) {6} OptoMedic Traditional 510(k) Premarket Notification K252318 The predicate device has not been subject to a design-related recall. # IV. Reference Device 510(k) Number: K231342 Product Name: OPTO-IFL-1000 Sponsor: Guangdong OptoMedic Technologies, Inc. # V. Device Description Insufflator (OPTO-IFL1000) is a $\mathrm{CO}_{2}$ insufflation device for creating and maintaining a pneumoperitoneum during laparoscopic examinations and operations. It is capable of establishing the surgical field of view and operating space. $\mathrm{CO}_{2}$ gas can be injected into abdominal cavity by the device, and the gas separates the abdominal wall from the internal organs of the abdominal cavity, forming a space for the operation and visual field. The device is to be used with the following insufflation tubes: 1. OPTO-T1000H (with heating function) 2. OPTO-T1000 (without heating function) # VI. Indications for Use The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures. # VII. Comparison of Technological Characteristics with the Predicate Device Table 1 Technological Characteristics Comparison | Description | Subject Device (K252318) | Predicate Device (K222812) | | --- | --- | --- | | Product | Insufflator (OPTO-IFL1000) | Insufflator (SL102) | | Regulation Number | 21 C.F.R. § 884.1730 | 21 C.F.R. § 884.1730 | | Product Code | HIF | HIF | | Device class | Class II | Class II | | Indication for use | The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures. | The Insufflator is a device intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it. | | Prescription/Over-the-counter use | Prescription | Prescription | {7} OptoMedic Traditional 510(k) Premarket Notification K2523181 Page 3 of 5 | Distension Medium | CO₂ | CO₂ | | --- | --- | --- | | Configuration | Insufflator, heatable insufflation tube set insufflation tube, central intake pipe (optional), gas filter, power supply cord, fuse (F3.15AH250V), cart | Insufflator (housing, power supply, pressure reducers, venting system, fluid sensor, gas heater, various setting keys and display elements), high pressure tube and pneumoperitoneum tube | | Insufflation Mode Selection | Pediatric Mode: 1-15 mmHg, 0.1-15 L/min Adult Mode: 1-20 mmHg, 1-50 L/min Obese Mode: 1-30 mmHg, 1-50 L/min Veress Mode: 1-15 mmHg, 1-5 L/min | Pediatric mode: 1-10mmHg, 1-20 L/min Adult mode: 1-30mmHg, 1-50 L/min Bariatric mode: 1-30mmHg, 1-50 L/min | | Pressure range | 1-30 mmHg | 1-30 mmHg | | Accuracy of the measured pressure | ±2mmHg | ±2mmHg | | Flow Rate Range | 0.1~50L/min | 1-50L/min | | Accuracy of the measured flow | the measured flow<2L/min: ±0.5L/min; 2L/min≤the measured flow≤10L/min: ±2L/min; the measured flow>10L/min: ±20% | the measured flow≤10L/min:±2L/min; the measured flow>10L/min: ±20% | | Overpressure alarm | When the nominal pressure is exceeded by more than 4 mmHg, visual and audible alarms will be issued. When the measured pressure exceeds 30 mmHg for 0.5 seconds, a medium priority overpressure alarm will be triggered | When the actual pressure is > 3 mmHg above the nominal pressure, the overpressure warning is initiated. | | Gas supply warning function | When gas cylinder pressure falls below 15 bar or central gas supply pressure falls below 3 bar, injection will be interrupted, device warning initiated. | When gas cylinder pressure falls below 15 bar or central gas supply pressure falls below 2 bar, injection will be interrupted, device warning initiated. | | Gas Overheating alarm | At >41°C, visual and acoustic alarm; gas injection interrupted | At>41°C, visual and acoustic alarm; gas injection interrupted | | Heating Function | Support heating function (with OPTO-T1000H tubing) | Does not support heating function (SL102) | | Contamination warning | If liquid enters the device through the insufflation connection port, the device warning will be initiated and device will be allowed to complete the current operation. Insufflation cannot be restarted after device is | When the fluid penetrates into the device through the pneumoperitoneum tube joint, the gas injection will be interrupted, device warning initiated. | {8} OptoMedic Traditional 510(k) Premarket Notification K252318 Page 4 of 5 | | turned off. | | | --- | --- | --- | | Reprocessing for insufflation tube | Yes, Steam sterilization | Yes, Steam sterilization | | Dimensions (W*H*D) | 370mm* 410mm *166.5mm | 267mm x 395mm x134mm | | Weight | 11Kg (N.W.) | About 8.2kg | | Power supply | 110-240 V~ | 110-240 V~ | The differences between the subject and predicate device include the mode, flow rate and pressure accuracy, overpressure alarm indications and actions, gas supply warning indication, heating functions, and dimensions/weight. The different technological characteristics do not raise different questions of safety and effectiveness. ## VIII. Performance data Non-clinical tests were conducted to verify that the subject device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards: - IEC 60601-1-2:2020 Edition 4.1 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances The reference device is the previous version of the subject device, without modes. The operation of the subject hardware is unchanged from the reference device. Therefore, electrical safety information from the reference device is leveraged to support the electrical safety of the subject device. Per the FDA guidance document *Content of Premarket Submissions for Device Software Functions*, issued June 2024, the software verification and validation testing were conducted and the test results demonstrated the software function met the requirements. The software for this device was considered a “Enhanced” software documentation level. Performance testing was also conducted and demonstrated that the proposed system performs according to specifications and functions as intended and substantially equivalent to the predicate device. And the test result shows that the preset acceptance criteria are met and are substantially equivalent to the predicate device performance: 1. Gas Supply Indication 2. Accuracy of the Pressure 3. Accuracy of the Pressure-Under Leak Condition (Continuous leakage compensation testing) 4. Overpressure Alarm 5. Overpressure Reduction 6. Under-pressure Replenishment (Transient leakage compensation testing) {9} OptoMedic Traditional 510(k) Premarket Notification K252318 Page 5 of 5 7. Accuracy of the Flow 8. Heating Function 9. Overheating Alarm 10. Accuracy of Gas Consumption Display ## IX. Conclusions The performance testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device to support a substantial equivalence determination.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...