K222812 · Scivita Medical Technology Co., Ltd. · HIF · May 16, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K222812
Device Name
Insufflator
Applicant
Scivita Medical Technology Co., Ltd.
Product Code
HIF · Obstetrics/Gynecology
Decision Date
May 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Insufflator is a device intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Device Story
Microprocessor-controlled CO2 insufflator; inputs include gas from central supply or cylinders; transforms inputs via pressure regulation and heating (optional) to maintain set pressure/flow in peritoneal cavity; outputs gas through pneumoperitoneum tube; used in clinical settings by healthcare professionals; monitors pressure, flow, and temperature; features overpressure, gas supply, contamination, venting, and heater defect warnings; provides automated gas delivery to distend cavity for laparoscopic access; benefits patient by enabling minimally invasive surgical visualization and instrumentation.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via comparative testing against predicate for leakage compensation, flow rate, gas temperature, overpressure release time, overpressure alarm, and set pressure output. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18 confirmed.
Technological Characteristics
Microprocessor-controlled CO2 insufflator. Materials: steel plate housing, resin composite front panel; high pressure tube (synthetic rubber, braided steel); pneumoperitoneum tube (Silicone, SUS304, PSU). Power: 100-240V, 50/60Hz. Features: pressure reducers, venting system, fluid sensor, gas heater. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 17665-1, ISO 14971, AAMI TIR 30/12.
Indications for Use
Indicated for use in laparoscopic procedures to distend the peritoneal cavity with gas to facilitate the use of a laparoscope. Applicable for pediatric, adult, and bariatric patient populations.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below that.
May 16, 2023
Scivita Medical Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China
Re: K222812 Trade/Device Name: Insufflator Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF Dated: April 14, 2023 Received: April 14, 2023
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222812
Device Name Insufflator
Indications for Use (Describe)
The Insufflator is a device intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: _ K222812
- 1. Date of Preparation: 05/12/2023
- 2. Sponsor Identification
## Scivita Medical Technology Co., Ltd. No. 8, Zhong Tian Xiang, Suzhou Industrial Park, 215000 Suzhou, Jiangsu Prov., P.R. CHINA.
Establishment Registration Number: 3020746799.
Contact Person: Ruqin Wu Position: Regulation Manager Tel: +86-512-81877788 Fax: +86-512-85187285 Email: wuruqin@scivitamedical.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)
Mid-Link Consulting Co., Ltd. P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{5}------------------------------------------------
#### 4. Identification of Subject Device
Trade Name: Insufflator Common Name: Laparoscopic Insufflator
Classification Name: Laparoscopic insufflator Regulatory Class: II Product Code: HIF (Insufflator, Laparoscopic) Regulation Number: 21 CFR 884.1730 Review Panel: Obstetrics/Gynecology
Indications for Use:
The Insufflator is a device intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Device Description:
The subject device consists of an insufflator, a high pressure tube and a pneumoperitoneum tube. The insufflator is a microprocessor-controlled CO2 insufflator that consists of the following major components and features: housing, power supply, pressure reducers, venting system, fluid sensor, gas heater, various setting keys and display elements. The insufflator is not intended to enter the field and cannot be sterilized. The pneumoperitoneum tube shall be cleaned, disinfected, and sterilized prior to subsequent use.
| System name | Component name | Model |
|-------------|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Insufflator | Insufflator | SL101, SL102 |
| | Pneumoperitoneum tube | IL-QFH02, IL-QF02 |
| | High pressure tube | High pressure Tube for Central Gas supply:<br>IL-GY(G)02, IL-GY(E)02, IL-GY(A)02<br>High pressure Tube for Gas Cylinder:<br>IL-GY02 |
Table 1. System Configuration
The SL101 insufflator has the gas heating function, while the SL102 insufflator does not have the gas heating function. And the IL-QFH02 pneumoperitoneum tube has the gas heating function, while the IL-QF02 pneumoperitoneum tube does not have the gas heating function. The high pressure tube is provided in two types, connected to Central Gas supply and to Gas Cylinder. The high pressure tube for Central Gas supply can be available in 3 models, IL-GY(G)02, IL-GY(E)02, IL-GY(A)02; while the high pressure tube for Gas Cylinder can only be available in IL-GY02 model.
{6}------------------------------------------------
### 5. Identification of Predicate Devices
Predicate Device 510(k) Number: K030837 Trade/Device Name: 40 L HIGH FLOW INSUFFLATOR F108
The predicate device has not been subject to a design-related recall.
Reference Device 510(k) Number: K153513 Trade/Device Name: Insufflator 50L FM134
#### 6. Summary of Technological characteristics
| ITEM | Subject Device | Predicate Device | Reference Device | Remark |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product | Insufflator | K030837 | K153513 | / |
| | | 40 L HIGH FLOW<br>INSUFFLATOR F108 | Insufflator<br>FM134<br>50L | |
| Product Code | HIF | HIF | HIF, OSV | Same |
| Regulation No. | 21 CFR 884.1730 | 21 CFR 884.1730 | 21 CFR 884.1730 | Same |
| Class | II | II | II | Same |
| Indications for Use | The Insufflator is a<br>device intended to<br>facilitate the use of the<br>laparoscope by filling the<br>peritoneal cavity with<br>gas to distend it. | The 40L High Flow<br>Insufflator F108 with<br>low flow mode is a<br>device intended to<br>facilitate the use of the<br>laparoscope by filling<br>the peritoneal cavity<br>with gas to distend it. | The device Insufflator<br>50L FM134 is a CO2<br>insufflator intended<br>for use during<br>diagnostic and/or<br>therapeutic<br>endoscopic<br>procedures to distend<br>a cavity by filling it<br>with gas. The<br>Standard, Pediatric<br>and Bariatric<br>operating modes of<br>the device are<br>indicated to fill and<br>distend a peritoneal<br>cavity with gas during<br>a laparoscopic | Same |
| | | | | |
| | | | procedure. The<br>Pediatric operating<br>mode is specifically<br>indicated for pediatric<br>laparoscopic<br>procedures. The<br>Vessel Harvesting<br>operating mode is<br>indicated for use<br>during endoscopic<br>vessel harvesting<br>procedures to create a<br>cavity along the<br>saphenous vein or<br>radial artery. | |
| Configuration | Insufflator (housing,<br>power supply, pressure<br>reducers, venting system,<br>fluid sensor, gas heater,<br>various setting keys and<br>display elements), high<br>pressure tube and<br>pneumoperitoneum tube | Housing, power<br>supply, pressure<br>reducers, a venting<br>system, a fluid sensor,<br>a gas heater, various<br>setting keys and<br>display elements | Casing, world power<br>supply, pressure<br>reducers, venting<br>system, redundant<br>pressure pressure<br>measurement, fluid<br>sensor, gas heater,<br>software controlled<br>graphical user<br>interface (GUI) touch<br>screen and various<br>setting keys and<br>display elements | Different |
| Mechanism of<br>Action | Microprocessor<br>controlled CO2<br>insufflator | Microprocessor<br>controlled CO2<br>insufflator | Microprocessor<br>controlled CO2<br>insufflator | Same |
| ITEM | Subject Device | Predicate Device K030837 | Reference Device K153513 | Remark |
| Product | Insufflator | 40 L HIGH FLOW<br>INSUFFLATOR F108 | Insufflator 50L<br>FM134 | / |
| Operating Modes | Pediatric mode, adult mode and bariatric mode | Veress operating mode, High Flow operating mode | Standard, Pediatric and Bariatric operating mode | Different |
| Set Pressure Range | Pediatric mode:<br>1-10mmHg<br>Adult mode and bariatric mode: 1-30mmHg | 1-30mmHg | / | Different |
| Accuracy of Pressure | $\pm$ 2mmHg | $\pm$ 1mmHg | / | Different |
| Set Gas Flow Range | 1-50L/min | 1-40L/min | 1-50L/min | Different |
| Power Supply | 100-240V, 50/60Hz | 100-240V, 50/60Hz | / | Same |
| Weight | About 8.2kg | About 7kg | / | Different |
| Dimension (W x H x D) | 267mm x 395mm x 134mm | 267mm x 138mm x 410mm | / | Different |
| Overpressure warning function | Has the overpressure warning function;<br>When the actual pressure is > 3 mmHg above the nominal pressure, the overpressure warning is initiated. | Has the overpressure warning function;<br>When the actual pressure is > 4 mmHg above the nominal pressure, the overpressure warning is initiated. | / | Different |
| Gas supply warning function | When gas cylinder pressure falls below 15 bar or central gas supply pressure falls below 2 bar, injection will be interrupted, device warning initiated. | When gas cylinder pressure falls below 15 bar or central gas supply pressure is too low, injection will be interrupted, device warning initiated. | / | Different |
| Contamination warning function | When the fluid penetrates into the device through the pneumoperitoneum tube joint, the gas injection will | When the fluid penetrates into the device through the pneumoperitoneum | / | Same |
Table 2. General Comparison
{7}------------------------------------------------
{8}------------------------------------------------
| | Table 3. Performance and Safety Comparison | | |
|--|--------------------------------------------|--|--|
|--|--------------------------------------------|--|--|
{9}------------------------------------------------
| | be interrupted, device warning initiated. | tube joint, the gas injection will be interrupted, device warning initiated. | | |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------|-----------|
| Venting defect warning function | When the device vents defect, the gas injection will be interrupted, device warning initiated. | / | / | Different |
| Heater defect warning function | When the gas temperature is greater than 41°C, the gas injection will be interrupted, device warning initiated. | When the gas temperature is greater than 42°C, the gas injection will be interrupted, device warning initiated. | / | Different |
| Material | Insufflator: The shell material is steel plate, and the front panel is resin composite material.<br>High Pressure Tube: Synthetic rubber, double braided steel wire<br>Pneumoperitoneum Tube: Silicone, SUS304, PSU | Host: The outer surface is steel plate and resin.<br>Tube: Silicone, PSU, Special steel | / | Different |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
The differences between the subject and predicate devices do not raise different questions of safety or effectiveness.
- 7. Summary of Non-Clinical Performance Testing
Non clinical tests were conducted to verify that the subject device met all design specifications as was same/similar to the predicate device and reference device. The test results demonstrated that the subject device complies with the following standards:
- A IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+AM1:2012, Medical Electrical Equipment-Part 1: General requirements for basic safety and essential performance, including the US National Differences
{10}------------------------------------------------
- A IEC 60601-1-2:2014 Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests
- > ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems
- > AAMI TIR 30: 2011/ (R) 2016 A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices.
- > AAMI TIR 12:2020 Designing, testing, and labeling medical devices intended for processing by health care facilities: A guide for device manufacturers.
- > ISO 15883-2:2006 Washer-disinfectors - Part 2: Requirements and tests for washer-disinfectors employing thermal disinfection for surgical instruments, anaesthetic equipment, bowls, dishes, receivers, utensils, glassware, etc.
- A ISO 17665-1: 2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices.
- > ISO 14971: 2019 Medical devices - Application of risk management to medical devices
- > IEC 60601-2-18:2009 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
The software verification and validation for software with a Major Level of Concern were conducted in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005. The test results demonstrated the software function met the requirements.
The following comparative performance testing was also conducted on the subject and predicate devices:
- > Compensation of leakages
- > Flow rate
- > Gas temperature
- > Overpressure release time
- A Overpressure alarm test
- > Set pressure output
- 8. Summary of Clinical Testing
No clinical study is included in this submission.
#### 9. Conclusion
The results of nonclinical testing demonstrate that the subject devices are as safe and effective as the predicate device to support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.