K252197 · Nobel Biocare AB · DZE · Feb 18, 2026 · Dental
Device Facts
Record ID
K252197
Device Name
Nobel Biocare S Series Implants
Applicant
Nobel Biocare AB
Product Code
DZE · Dental
Decision Date
Feb 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. NobelReplace® S TiUltra™, NobelReplace® S TiUnite® NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are indicated for single or multiple unit restorations. The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants can be used in splinted or non-splinted applications. The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied. NobelParallel™ S TiUltra™, NobelParallel™ S TiUnite® NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function. NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are indicated for single or multiple restorations in splinted or non-splinted applications. This can be achieved by 2-stage or 1-stage surgical techniques in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. Implants with <7 mm length are for delayed loading only when appropriate stability has been achieved.
Device Story
Endosseous dental implants (NobelActive S, NobelParallel S, NobelReplace S) surgically placed in maxilla or mandible; anchor/support prosthetic tooth replacements; restore esthetics and chewing function. Implants feature consistent Narrow Platform (NP) conical connection across all diameters; allows use of existing prosthetic workflows, drills, and components. Manufactured from commercially pure titanium; TiUltra (hydrophilic) or TiUnite surface treatments. Used by dental surgeons; 1-stage or 2-stage surgical protocols; immediate, early, or delayed loading. Output is a stable foundation for prosthetic abutments/crowns. Benefits include standardized prosthetic interface, simplified surgical planning, and restoration of oral function.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing, including dynamic loading (ISO 14801), packaging performance (ASTM D4169), and MR compatibility analysis (ASTM F2052, F2213, F2119, F2182). Biocompatibility confirmed per ISO 10993-1.
Technological Characteristics
Commercially pure titanium endosseous implants. Surface: Anodic oxidation (TiUltra hydrophilic or TiUnite). Connection: Narrow Platform (NP) conical connection with hex interface. Dimensions: Diameters 3.5mm–5.5mm; lengths 7mm–18mm. Sterilization: Gamma radiation (SAL 10-6). MR Conditional. Packaging: PET vial/HDPE cap in PETG blister.
Indications for Use
Indicated for edentulous or partially edentulous patients requiring tooth replacement in the maxilla or mandible. Suitable for single or multiple unit restorations in splinted or non-splinted applications. Applicable for 1-stage or 2-stage surgical techniques with immediate, early, or delayed loading protocols, provided sufficient primary stability is achieved.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
NobelProcera HT ML FCZ Implant Bridge and Framework (K160158)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 18, 2026
Nobel Biocare AB
Bernice Jim
Head of RA Development Implant systems, US/CA & PLM, Strategy & EBS
Vastra Hamngatan 1
Goteborg, Vastra Gotaland SE 411 17
SWEDEN
Re: K252197
Trade/Device Name: Nobel Biocare S Series Implants
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE
Dated: July 14, 2025
Received: January 21, 2026
Dear Bernice Jim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252197 - Bernice Jim
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252197 - Bernice Jim
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K252197
Device Name
Nobel Biocare S Series Implants
Indications for Use (Describe)
NobelActive® S TiUltra™, NobelActive® S TiUnite®
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function.
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are indicated for single or multiple unit restorations in splinted or non-splinted applications.
This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique.
NobelReplace® S TiUltra™, NobelReplace® S TiUnite®
NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are indicated for single or multiple unit restorations.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants can be used in splinted or non-splinted applications.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.
NobelParallel™ S TiUltra™, NobelParallel™ S TiUnite®
NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function.
NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are indicated for single or multiple restorations in splinted or non-splinted applications.
This can be achieved by 2-stage or 1-stage surgical techniques in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique.
Implants with <7 mm length are for delayed loading only when appropriate stability has been achieved.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 2 of 2
{5}
N
Nobel Biocare
# 1 – 510(k) Summary
## 1.1 Submitter Information
Submitter: Nobel Biocare AB
Vastra Hamngatan 1
Goteborg 411 17
Sweden
Submitted By: Nobel Biocare Services AG
Balz-Zimmerman-Strasse 7
8302 Kloten
Switzerland
Contact Person: Bernice Jim, PhD
E-Mail: regulatory.affairs.nb@envistaco.com
Telephone: +41 43 211 42 00
Prepared By: Heather Yates
Date Prepared: February 13, 2026
## 1.2 Device Name
Proprietary Name: Nobel Biocare S Series Implants
Common Name: Dental Implant
Classification Name: Endosseous Dental Implant
Regulation Number: 21 CFR 872.3640
Device Class: Class II
Product Code: DZE
## 1.3 Predicate and Reference Devices
### Primary Predicate device
K202344, TiUltra Implants and Xeal Abutments
### Reference Devices:
1. K230108 Straumann BLC and TLC Implants
2. K142260 NobelActive
3. K173418 NobelParallel Conical Connection
4. K073142 NobelReplace Hexagonal Implant
5. K212125 Nobel Biocare Dental Implant Systems Portfolio - MR Conditional
Page 1 of 17
{6}
N
Nobel Biocare
6 K023113 Replace TiUnite Endosseous Implant
7 K071370 NOBELACTIVE INTERNAL CONNECTION IMPLANT
8 K083100 GOLDADAPT NON-ENGAGING NOBELACTIVE NP & RP, GOLDADAPT ENGAGING NOBELACTIVE NP & RP
9 K132746 NOBELPROCERA ANGULATED SCREW CHANNEL ABUTMENT CONICAL CONNECTION
10 K233208 NobelProcera Titanium ASC Abutment, Omnigrip Clinical Screw Titanium
11 K200040 Universal Base Conical Connection (CC)
12 K132749 NOBELPROCERA OVERDENTURE BAR
13 K161416 Multi-unit Abutment Plus
14 K202452 NobelProcera Zirconia Implant Bridge
15 K240346 NobelProcera Zirconia Implant Bridge
16 K041236 PROCERA IMPLANT BRIDGE, MODELS 15-1001, 15-1002, 15-1051, 15-1052
17 K133731 NOBELACTIVE WIDE PLATFORM (WP)
18 K061003 SFB & CFB IMPLANTS
19 K161435 Temporary Snap Abutment
20 K091756 NobelProcera Ti Abutment
21 K062566 NOBELREPLACE TAPERED CONICAL CONNECTION
22 K072570 NOBELACTIVE MULTI UNIT ABUTMENT
23 K160158 NobelProcera HT ML FCZ Implant Bridge and Framework
## 1.4 Device Description Summary
The Subject Device, Nobel Biocare S Series Implants, comprises of six Device Lines NobelActive S TiUltra™, NobelParallel™ S TiUltra™, NobelReplace S TiUltra™, NobelActive S TiUnite, NobelParallel™ S TiUnite, NobelReplace S TiUnite
Nobel Biocare S Series Implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental prostheses to restore chewing function. The Subject Device features a consistent diameter at the implant-abutment interface, regardless of the implant size or type. This means that the prosthetic components, such as abutments and restorative elements, have uniform dimensions and connections across all implant sizes within the system. The connection used is the Narrow Platform (NP) conical connection.
The implant bodies of the Subject Devices and Predicate devices are the same. This allows use of the same drills and drilling protocols, rescue procedures and rescue tools. By using the existing Narrow Platform connection, the same prosthetic workflows, prosthetic components, and lab components can be used.
The implants are available in different sizes as listed in Table 1.
Page 2 of 17
{7}
N
Nobel Biocare
Table 1. Implant specifications
| Implant line | Conical connection size | A - Implant diameter [mm] | | | Lengths [mm] |
| --- | --- | --- | --- | --- | --- |
| | | A - Implant diameter [mm] | B - Platform diameter [mm] | C - Abutment interface [mm] | |
| NobelActive® S TiUltra™/TiUnite® | NP | 3.5 | 3.5 | 3.0 | 8.5, 10, 11.5, 13, 15, 18 |
| | NP | 4.3 | 3.9 | 3.0 | 8.5, 10, 11.5, 13, 15, 18 |
| | NP | 5.0 | 3.9 | 3.0 | 8.5, 10, 11.5, 13, 15, 18 |
| | NP | 5.5 | 3.9 | 3.0 | 7, 8.5, 10, 11.5, 13, 15 |
| NobelParallel™ S TiUltra™/TiUnite® | NP | 3.75 | 3.5 | 3.0 | 7, 8.5, 10, 11.5, 13, 15, 18 |
| | NP | 4.3 | 3.9 | 3.0 | 7, 8.5, 10, 11.5, 13, 15, 18 |
| | NP | 5.0 | 3.9 | 3.0 | 7, 8.5, 10, 11.5, 13, 15, 18 |
| | NP | 5.5 | 3.9 | 3.0 | 7, 8.5, 10, 11.5, 13, 15 |
| NobelReplace® S TiUltra™/TiUnite® | NP | 3.5 | 3.5 | 3.0 | 8, 10, 11.5, 13, 16 |
| | NP | 4.3 | 3.9 | 3.0 | 8, 10, 11.5, 13, 16 |
| | NP | 5.0 | 3.9 | 3.0 | 8, 10, 11.5, 13, 16 |
The Nobel Biocare S Series Implants are manufactured from commercially pure titanium and incorporate a conical connection (Size Narrow Platform (NP)) with a hex interface in the collar region, which combines with Nobel Biocare's existing prosthetic Abutment portfolio cleared in K071370, K083100, K132746, K233208, K202344, K200040, K132749, K161416, K202452, K240346, K041236, K133731, K061003 and K161435.
Table 2 outlines which single-unit restoration abutments with an NP connection can be used with the subject device.
Table 2. Compatible Single-unit Restoration Abutments for NobelBiocare S Series Implants
| Single-Unit Restoration Abutments (Proprietary Name) |
| --- |
| Group 1 – Titanium Abutments |
| Esthetic Abutment Conical Connection (with NP Connection only) (K071370) |
| Snappy Abutments (with NP Connection only) (K071370) |
| Universal Base Conical Connection(with NP connection only) (K200040) |
| NobelProcera Abutments Conical Connection Titanium (with NP Connection only) (K071370) |
| NobelProcera Abutments Conical Connection Titanium (with NP connection only) (K233208) |
| Group 2 – Gold Abutments |
| GoldAdapt Conical Connection (with NP connection only) (K083100) |
| Group 3 – Titanium and Zirconia Abutments |
| NobelProcera Angulated Screw Channel Abutment (with NP connection only) (K132746) |
| NobelProcera FCZ Implant Crown (with NP connection only) (K132746) |
Table 3 outlines which multi-unit restoration abutments and bridges with an NP connection can be used with the subject devices.
Page 3 of 17
{8}
N
Nobel Biocare
Table 3. Compatible Multi-unit Restoration Abutments for Nobel Biocare S Series Implants
| Multi-unit Restoration Abutments and Bridges (Proprietary Name) |
| --- |
| Procera Implant Bridge Titanium (K041236) |
| Multi-Unit Abutment Plus Conical Connection (with NP connection only) (K161416) |
| Multi-Unit Abutment Xeal Conical Connection (wit NP connection only) (K202344) |
| NPr Impl Bar Overdenture Ti (K132749) |
| NobelProcera Implant Bridge Zirconia (K202452; K240346) |
Table 4 outlines the temporary and healing restorations with an NP connection that can be used with the subject devices.
Table 4. Compatible Temporary and Healing Restoration Abutments for Nobel Biocare S Series Implants
| Temporary and Healing Restorations (Proprietary Name) |
| --- |
| Healing Abutment Conical Connection (with NP connection only) (K133731) |
| Temporary Snap Abutment (with NP connection only) (K161435) |
| Temporary Abutment Conical Connection (with NP connection only) (K061003) |
Table 5 lists the corresponding submissions where compatible Class II accessories with an NP connection were previously identified.
Table 5. List of 510(k) Class II Compatible Accessories to Nobel Biocare S Series Implants
| Compatible Accessories |
| --- |
| NobelReplace Tapered Conical Connection, Cover Screw Conical Connection NP (K062566) |
| NobelActive Internal Connection Implant, Clinical Screw Conical Connection NP (K071370) |
| NobelActive Multi Unit Abutment, Screw Multi-Unit Angled Abutment Conical Connection NP (K072570) |
| NobelProcera HT ML FCZ Implant Bridge and Framework, Metal Adapter (NP Only) (K160158) |
| NobelProcera® Titanium ASC Abutment, Omnigrip Clinical Screw Titanium CC NP (K233208) |
| NobelProcera Angulated Screw Channel Abutment Conical Connection, Omnigrip™ Clinical Screw CC NP (K132746) |
The On1™ Base/On1™ Base Xeal must not be used in combination with NobelActive® S, NobelParallel™ S, NobelReplace® S implants.
NobelActive® S, NobelParallel™ S, NobelReplace® S implants in combination with NobelProcera® Angulated Screw Channel Abutment and NobelProcera® FCZ Implant Crown are indicated for maxillary lateral and mandibular central/lateral incisors only.
NobelActive® S, NobelParallel™ S, NobelReplace® S implants may be used in combination with NobelProcera® Angulated Screw Channel Abutment and NobelProcera® FCZ Implant Crown up to a maximum angulation of 20°
NobelActive® S, NobelParallel™ S, NobelReplace® S implants may be used in combination with GoldAdapt Engaging CC NP abutments in straight (0°) configurations only.
Page 4 of 17
{9}
N
Nobel Biocare™
No angulation is allowed for Universal Base Conical Connection Abutments when mated with the NobelActive® S, NobelParallel™ S, NobelReplace® S implants with diameters greater than 3.5mm.
## 1.5 Indications for Use
### NobelActive® S TiUltra™, NobelActive® S TiUnite®
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function.
NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are indicated for single or multiple unit restorations in splinted or non-splinted applications.
This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique.
### NobelReplace® S TiUltra™, NobelReplace® S TiUnite®
NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are indicated for single or multiple unit restorations.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants can be used in splinted or non-splinted applications.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.
Page 5 of 17
{10}
N
Nobel Biocare™
# NobelParallel™ S TiUltra™, NobelParallel™ S TiUnite®
NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function.
NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are indicated for single or multiple restorations in splinted or non-splinted applications.
This can be achieved by 2-stage or 1-stage surgical techniques in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique.
Implants with <7 mm length are for delayed loading only when appropriate stability has been achieved.
## 1.6 Technological Comparison
The Nobel Biocare S Series Implants consists of six device lines; NobelActive S TiUltra™, NobelParallel™ S TiUltra™, NobelReplace S TiUltra™, NobelActive S TiUnite, NobelParallel™ S TiUnite, NobelReplace S TiUnite.
The subject device and the predicate device have the same Indications for Use, Macro design and material surface treatment. They are single-use implants, delivered in the same packaging configuration, provided sterile (gamma sterilization, SAL 10-6) with a shelf-life of 5 years. Both the subject devices and predicate device are labelled as MR Conditional
The design concept of Subject Device Lines combines the existing outer geometry of the implant bodies of the predicate devices but have the same Narrow Platform (NP) conical connection for all implant diameters. This introduces an increased platform shift. This contrasts with the Predicate which have a larger size conical connection (Regular Platform (RP) or Wide Platform (WP)) on implants with diameters larger than 3.75mm. Only Predicate Devices with diameters of 3.75mm or 3.5mm feature the Narrow Platform (NP) connection. The use of the existing NP Platform allows the use of existing prosthetic workflows, rescue procedures, prosthetic components, lab components, and rescue tools.
The differences between the subject device and the predicate do not raise different questions of substantial equivalence. Additional implant diameter to connection size combinations were tested whenever deemed to represent a worst-case. Worst-case
Page 6 of 17
{11}
N Nobel Biocare™
fatigue testing was conducted to demonstrate that the subject device performs at least as well as the defined benchmark reference system.
Page 7 of 17
{12}
N
Nobel Biocare
1.6.1 Substantial Equivalence Table:
| Technological Characteristics | Subject Device - Nobel Biocare S Series Implants | Primary Predicate Device TiUltra implants - K202344 | Reference Device #1 Straumann BLC and TLC Implants - K230108 | Reference Device #2 NobelActive - K142260 | Comparison |
| --- | --- | --- | --- | --- | --- |
| Regulatory Classification | 21 CFR 872.3640 Endosseous dental implant | 21 CFR 872.3640 Endosseous dental implant | 21 CFR 872.3640 Endosseous dental implant | 21 CFR 872.3640 Endosseous dental implant | Same as Predicate |
| Product Code | DZE | DZE | DZE, NHA | NHA, DZE | Same as Predicate |
| Indications for Use | NobelActive® S TiUltra™, NobelActive® S TiUnite® | NobelActive® TiUltra™ and NobelActive® S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. | NobelActive TiUltra™ | Straumann dental implants are indicated for the functional and esthetic oral rehabilitation of the upper or lower jaw of edentulous or partially edentulous patients. They can be used for immediate, early or late implantation following the extraction or loss of natural teeth. The implants can be placed with immediate function for single-tooth and/or multiple tooth restorations when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. | NobelActive implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. |
| NobelActive® S TiUltra™ and NobelActive® S TiUnite® implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. | NobelActive® TiUltra implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. | NobelActive TiUltra™ | NobelActive implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing | NobelActive implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. |
| This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing | This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing | | | |
1 Note that the Subject Devices do not include 3.0 sizes. Therefore indications pertaining to these devices do not feature in the Subject Devices Indications for Use
Page 8 of 17
{13}
N
Nobel Biocare
| Technological Characteristics | Subject Device - | Primary Predicate Device | Reference Device #1 | Reference Device #2 | Comparison |
| --- | --- | --- | --- | --- | --- |
| | Nobel Biocare S Series Implants | TiUltra implants - K202344 | Straumann BLC and TLC Implants - K230108 | NobelActive - K142260 | |
| | sufficient primary stability and appropriate occlusal loading for the selected technique. | sufficient primary stability and appropriate occlusal loading for the selected technique. | | | |
| | | NobelActive TiUltra 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. NobelActive TiUltra 3.0 implants are indicated for single-unit restorations only. | | | |
| | NobelReplace® S TiUltra™, NobelReplace® S TiUnite® NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. | NobelReplace CC TiUltra NobelReplace CC TiUltra™ implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. | | | |
| | The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants are | NobelReplace CC TiUltra™ implants are indicated for single or multiple unit restorations. | | | |
{14}
N
Nobel
Biocare
| Technological Characteristics | Subject Device - | Primary Predicate Device | Reference Device #1 | Reference Device #2 | Comparison |
| --- | --- | --- | --- | --- | --- |
| | Nobel Biocare S Series Implants | TiUltra implants - K202344 | Straumann BLC and TLC Implants - K230108 | NobelActive - K142260 | |
| | indicated for single or multiple unit restorations.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants can be used in splinted or non-splinted applications.
The NobelReplace® S TiUltra™ and NobelReplace® S TiUnite® implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.
NobelParallel™ S TiUltra™, NobelParallel™ S TiUnite® NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function. | The NobelReplace CC TiUltra™ implants can be used in splinted or non-splinted applications.
The NobelReplace CC TiUltra™ implant may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.
NobelParallel CC TiUltra
NobelParallel CC TiUltra implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function. | | | |
Page 10 of 17
{15}
N
Nobel
Biocare
| Technological Characteristics | Subject Device - | Primary Predicate Device | Reference Device #1 | Reference Device #2 | Comparison | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Nobel Biocare S Series Implants | TiUltra implants - K202344 | Straumann BLC and TLC Implants - K230108 | | NobelActive - K142260 |
| | | NobelParallel™ S TiUltra™ and NobelParallel™ S TiUnite® implants are indicated for single or multiple restorations in splinted or non-splinted applications. This can be achieved by 2-stage or 1-stage surgical techniques in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. Implants with <7 mm length are for delayed loading only when appropriate stability has been achieved. | NobelParallel™ CC TiUltra™ implants are indicated for single or multiple restorations in splinted or non-splinted applications. This can be achieved by a 2-stage or 1-stage surgical techniques in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. Implants with <7mm length are for delayed loading only when appropriate stability has been achieved | | | |
| Design Features | Implant Length | NobelActive S implants 7.0mm, 8.5mm, 10.0mm, 11.5mm, 13mm, 15mm, 18mm NobelParallel S implants 7.0mm, 8.5mm, 10.0mm, 11.5mm, 13mm, 15mm, 18mm NobelReplace S implants | NobelActive TiUltra implants 7.0mm, 8.5mm, 10.0mm, 11.5mm, 13mm, 15mm, 18mm NobelParallel CC TiUltra implants 7.0mm, 8.5mm, 10.0mm, 11.5mm, 13mm, 15mm, 18mm NobelReplace CC TiUltra implants | 6mm, 8mm, 10mm, 12mm, 14mm, 16mm, 18mm | 7.0mm, 8.5mm, 10.0mm, 11.5mm, 13mm, 15mm, 18mm | Same as Predicate |
Page 11 of 17
{16}
N
Nobel
Biocare
| Technological Characteristics | Subject Device - | Primary Predicate Device | Reference Device #1 | Reference Device #2 | Comparison | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Nobel Biocare S Series Implants | TiUltra implants - K202344 | Straumann BLC and TLC Implants - K230108 | | NobelActive - K142260 |
| | | 8.0mm, 10.0mm, 11.5mm, 13mm, 16mm | 8.0mm, 10.0mm, 11.5mm, 13mm, 16mm | | | |
| | Implant Diameter | NobelActive S implants 3.5mm, 4.3mm, 5.0mm, 5.5mm | NobelActive TiUltra implants 3.5mm, 4.3mm, 5.0mm, 5.5mm | Ø 3.3, 3.75, 4.5, 5.5, 6.5 mm | 3.5mm, 4.3mm, 5.0mm, 5.5mm | Same as Predicate |
| | | NobelParallel S implants 3.75mm, 4.3mm, 5.0mm, 5.5mm | NobelParallel CC TiUltra Implants 3.75mm, 4.3mm, 5.0mm, 5.5mm | | | |
| | | NobelReplace S implants 3.5mm, 4.3mm, 5.0mm | NobelReplace CC TiUltra Implants 3.5mm, 4.3mm, 5.0mm | | | |
| | Platform Compatibility | Conical Connection NP | Conical Connection NP, RP, WP | n/a | Conical Connection NP, RP, WP | The Subject Device has the same implant/abutment connection as the NP of the Predicate. |
| | Platform shift | NobelActive S implants 0.295-1.175mm | NobelActive TiUltra implants 0.295-0.45mm | 0.3-1.9mm | 0.295-0.45mm | Within the range of Reference Device #1. |
| | | NobelParallel S implants 0.388-1.275mm | NobelParallel CC TiUltra Implants 0.388-0.778mm | | | |
| NobelReplace S implants 0.295-1.028mm | | NobelReplace CC TiUltra Implants 0.295-0.828mm | | | | |
| Macro Design | NobelActive S implants Tapered implant with back-tapered coronal design, | NobelActive TiUltra implants Tapered implant with back-tapered coronal design, | n/a | Tapered implant with back-tapered coronal design, | Same as Predicate | |
{17}
N
Nobel
Biocare
| Technological Characteristics | Subject Device - | Primary Predicate Device | Reference Device #1 | Reference Device #2 | Comparison | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Nobel Biocare S Series Implants | TiUltra implants - K202344 | Straumann BLC and TLC Implants - K230108 | | NobelActive - K142260 |
| | | reverse-cutting flutes and double lead threads | reverse-cutting flutes and double lead threads | | reverse-cutting flutes and double lead threads | |
| | NobelParallel S implants Double lead thread with groove; Tapered apex with bone cutting flutes, slightly tapered body | NobelParallel CC TiUltra implants Double lead thread with groove; Tapered apex with bone cutting flutes, slightly tapered body | | | | |
| | NobelReplace S implants Tapered implant with single lead threads | NobelReplace CC TiUltra implants Tapered implant with single lead threads | | | | |
| | Material | Commercially pure titanium | Commercially pure titanium | n/a | Commercially pure titanium | Same as Predicate |
| | Surface Treatment | Anodic oxidation | Anodic oxidation | n/a | Anodic oxidation | Same as Predicate |
| Implant/Abutment Connection | Anodic oxidation on collar and inside the connection | Anodic oxidation on collar and inside the connection | n/a | Machined on collar and inside the connection (NP and RP platforms) Anodic oxidation on collar and inside the connection (WP platform) | Same as Predicate for TiUltra and as Reference #2 for TiUnite surface. | |
| Surface Topography | TiUltra - Three-level surface [area roughness (Sa) thickness (Thk)]:
• Level 0 (collar):
o Sa = 0.5–1.5 μm
o Thk = 0.166 ± 0.008μm
• Level 1 (transition): | TiUltra - Three-level surface [area roughness (Sa) thickness (Thk)]:
• Level 0 (collar):
o Sa = 0.5–1.5 μm
o Thk = 0.166 ± 0.008μm
• Level 1 (transition): | n/a | | | Same as Predicate for TiUltra and as Reference #2 for TiUnite |
{18}
N
Nobel
Biocare
| Technological Characteristics | Subject Device - | Primary Predicate Device | Reference Device #1 | Reference Device #2 | Comparison | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Nobel Biocare S Series Implants | TiUltra implants - K202344 | Straumann BLC and TLC Implants - K230108 | | NobelActive - K142260 |
| | | o Sa = 1.4 ± 0.4μm
o Thk = 12.0 ± 1.2μm
• Level 2 (body):
o Sa = 1.5 ± 0.4μm
o Thk = 12.0 ± 1.2μm
TiUnite – Single level surface:
• Sa = 1.2 ± 0.5μm
• Thk = 12.5±2.5μm | o Sa = 1.4 ± 0.4μm
o Thk = 12.0 ± 1.2μm
• Level 2 (body):
o Sa = 1.5 ± 0.4μm
o Thk = 12.0 ± 1.2μm | | TiUnite – Single level surface:
• Sa = 1.2 ± 0.5μm
• Thk = 12.5±2.5μm | |
| | Color Coding | No color coding on implants.
Packaging is color coded | No color coding on implants.
Packaging is color coded | n/a | Color Coding only on the product for the 5.5mm diameter implant (blue). Packaging of all products are color coded | Same as Predicate for TiUltra and as Reference #2 for TiUnite |
| | Surface preservation | TiUltra – Hydrophilic surface
Sodium dihydrogen phosphate dihydrate and magnesium chloride hexahydrate salt
TiUnite – n/a | Hydrophilic surface
Sodium dihydrogen phosphate dihydrate and magnesium chloride hexahydrate salt | n/a | n/a | Same as Predicate for TiUltra and as Reference #2 for TiUnite |
| Packaging | | Polyethylene terephthalate (PET) vial with high density polyethylene (HDPE) cap (1st sterile barrier) placed in a polyethylene terephthalate glycol (PETG) blister sealed with (Tyvek) lid (2nd sterile barrier) inside a cardboard box (protective packaging). | Polyethylene terephthalate (PET) vial with high density polyethylene (HDPE) cap (1st sterile barrier) placed in a polyethylene terephthalate glycol (PETG) blister (2nd sterile barrier) inside a cardboard box (protective packaging). | n/a | Polyethylene terephthalate (PET) vial with high density polyethylene (HDPE) cap (1st sterile barrier) placed in a polyethylene terephthalate glycol (PETG) blister (2nd sterile barrier) inside a cardboard box (protective packaging). | Same as Predicate |
| Sterilization (SAL) | | Gamma Radiation (SAL 10-6) | Gamma Radiation (SAL 10-6) | n/a | Gamma Radiation (SAL 10-6) | Same as Predicate |
Page 14 of 17
{19}
N
Nobel
Biocare
| Technological Characteristics | Subject Device - | Primary Predicate Device | Reference Device #1 | Reference Device #2 | Comparison |
| --- | --- | --- | --- | --- | --- |
| | Nobel Biocare S Series Implants | TiUltra implants - K202344 | Straumann BLC and TLC Implants - K230108 | NobelActive - K142260 | |
| MR Compatibility | MR Conditional | MR Conditional (as per K212125) | n/a | MR Conditional (as per K212125) | Same as K212125 |
| Biocompatibility Testing | The final product, as shipped is biocompatible according to ISO 10993-1 | The final product, as shipped is biocompatible according to ISO 10993-1 | n/a | The final product, as shipped is biocompatible according to ISO 10993-1 | Same as Predicate |
| Performance Testing – Bench | Dynamic loading test for endosseous dental implants. | Dynamic loading test for endosseous dental implants, insertion test. | n/a | Dynamic loading test for endosseous dental implants, insertion test. | Same as Predicate |
| Clinical Testing | None | None | n/a | None | Same as Predicate |
Page 15 of 17
{20}
Nobel Biocare™
## 1.7 Non-Clinical Tests Summary & Conclusions:
Non-clinical testing was performed on the Subject devices Nobel Biocare S Series Implants:
- Packaging system performance testing per ASTM D4169
- Dynamic loading testing performed according to ISO 14801 was conducted according to ISO 14801 and the FDA Guidance Document entitled, "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" (May 12, 2004)
- Modified surface treatment information according to the FDA Guidance Document entitled, 'Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutment' (May 12, 2004) referenced from K202344 for TiUltra Surface and K173418 for TiUnite Surface.
- Referenced from K212125, non-clinical analysis and testing to evaluate the metallic Subject Devices in the MR environment was conducted according to ASTM F2052 (magnetically induced displacement force), ASTM F2213 (magnetically induced torque), ASTM F2119 (image artifact), and ASTM F2182 (RF induced heating), and the FDA guidance document Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (issued May 2021) Referenced from K202344, K142260, K173418, K073142, biological assessments were performed according to ISO 10993-1 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process' and to the FDA Guidance document 'Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016' for each subject device. The subject devices are equivalent in material and manufacturing processes to the primary predicate and reference devices, therefore, no new issues regarding biocompatibility were raised and no additional biocompatibility testing was required.
- Referenced from K202344, Endotoxin testing was completed in accordance with:
- USP 42-NF37:2019, Medical Devices—Bacterial Endotoxin and Pyrogen Tests
- ANSI/AAMI ST72:2011/ (R)2016, Bacterial endotoxins — Test methods, routine monitoring, and alternatives to batch testing.
- Referenced from K202344, packaging performance testing and sterile barrier shelf-life data of 5 years conducted according to ISO 11607-1:2019, Sterilization Validation in accordance with:
Page 16 of 17
{21}
N
Nobel Biocare
- ISO 11137-1:2006, Sterilization of health care products — Radiation — Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11137-2: 2013, Sterilization of health care products — Radiation — Part 2: Establishing the sterilization dose
Clinical performance data is not required to establish substantial equivalence for the subject devices.
Based on a comparison of intended use, indications, material composition, technological characteristics, principle of operation, features and performance data, the Subject Devices; Nobel Biocare S Series Implants are deemed to be substantially equivalent to the predicate devices.
Page 17 of 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.