Barbed PDO Suture

K252171 · Sutura Medical Technology, Inc. · NEW · Jun 12, 2026 · General, Plastic Surgery

Device Facts

Record IDK252171
Device NameBarbed PDO Suture
ApplicantSutura Medical Technology, Inc.
Product CodeNEW · General, Plastic Surgery
Decision DateJun 12, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4840
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Barbed PDO Suture is indicated for placement in the subcutaneous plane for temporary fixation of the cheek adipose layer in midface suspension surgery.

Device Story

Barbed PDO Suture is an absorbable, sterile, monofilament surgical suture made of polydioxanone (PDO). It features spiral bidirectional barbs along its length, allowing for tissue approximation and fixation without surgical knots. The device is supplied pre-loaded in a cannula. During midface suspension surgery, a physician places the suture in the subcutaneous plane. The barbs engage the surrounding soft tissue, anchoring and maintaining the cheek adipose layer in an elevated, repositioned state. This mechanical fixation provides temporary support to facial soft tissue. The device is intended for use in a clinical setting by trained healthcare providers. The cannula-based delivery method is designed to facilitate accurate placement while potentially reducing the risk of vascular or nerve injury compared to traditional needle-based techniques.

Clinical Evidence

Prospective, multi-center clinical study conducted at four U.S. sites with 57 enrolled subjects (48 completed). Evaluated safety and effectiveness for midface suspension surgery. Results showed a safety profile consistent with suture-based procedures; no serious adverse events or unanticipated adverse device effects reported. Most adverse events were mild and resolved without sequelae. Bench testing included USP <861> (diameter), USP <881> (tensile strength), sterility, endotoxin testing, shelf-life stability, and barb pull-out strength testing.

Technological Characteristics

Synthetic absorbable monofilament polydioxanone (PDO) suture; bidirectional barbs; stainless steel (STS-304) components; cannula-based delivery; ethylene oxide sterilization; implantable (>30 days contact); USP sizes 5-0 to 2; dyed with D&C Violet No. 2.

Indications for Use

Indicated for temporary fixation of the cheek adipose layer in the subcutaneous plane during midface suspension surgery in patients requiring soft tissue elevation.

Regulatory Classification

Identification

An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{1} June 12, 2026 Sutura Medical Technology, Inc. Annie Kim Regulatory Affairs Manager 3 Allied Drive, Suite 210 Dedham, Massachusetts 02026 Re: K252171 Trade/Device Name: Barbed PDO Suture Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable Polydioxanone Surgical Suture Regulatory Class: Class II Product Code: NEW Dated: May 8, 2026 Received: May 11, 2026 Dear Annie Kim: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K252171 - Annie Kim Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K252171 - Annie Kim Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TEK N. LAMICHHANE -S Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. # Indications for Use 510(k) Number (if known) K252171 Device Name Barbed PDO Suture Indications for Use (Describe) The Barbed PDO Suture is indicated for placement in the subcutaneous plane for temporary fixation of the cheek adipose layer in midface suspension surgery. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} ## 510(k) SUMMARY--- | 510(k) #: K252171 | | **510(k) Summary** | Prepared on: 2026-0605 | | --- | --- | --- | --- | | **Contact Details** | | | | | **Applicant Name** | Sutura Medical Technology Inc. | | | | **Applicant Address** | 2121 SW 3rd Avenue, Suite 500 Miami FL 33129 United States | | | | **Applicant Contact Telephone** | (570) 550-0700 | | | | **Applicant Contact** | Ms. Annie Kim | | | | **Applicant Contact Email** | akim@healthpolicyassociates.com | | | | **Device Name** | | | | | **Device Trade Name** | Barbed PDO Suture | | | | **Common Name** | Barbed PDO Suture | | | | **Classification Name** | Suture, surgical, absorbable, polydioxanone | | | | **Regulation Number** | 21 CFR 878.4840 | | | | **Product Code** | NEW | | | | **Legally Marketed Predicate Devices** | | | | | **Predicate #** | **Predicate Trade Name (Primary Predicate is listed first)** | **Product Code** | | | K192423 | Mint Product Family | NEW | | | K191361 | DemeDIOX Barbed Absorbable Surgical Suture | NEW | | | **Device Description Summary** | | | | | The Barbed PDO Suture is an absorbable, sterile surgical monofilament suture composed of polyester, poly (p-dioxanone). The Barbed PDO Suture consists of an absorbable monofilament strand thread with spiral bidirectional barbs and is supplied pre-loaded with a | | | | K252171 1 510(k) Summary {6} cannula, available in suture size USP 5-0 to 2. The material is dyed with D&C Violet No.2. and contains no additives. The barbing allows for tissue approximation without the use of surgical knots. The mechanism of action is based on tissue fixation in an elevated position. The device consists of an absorbable PDO monofilament with bidirectional barbs. When placed in the subcutaneous plane, the barbs engage surrounding soft tissue, allowing the suture to anchor and maintain tissue in a repositioned (elevated) state. ## Intended Use/Indications for Use The Barbed PDO Suture is indicated for placement in the subcutaneous plane for temporary fixation of the cheek adipose layer in midface suspension surgery. ## Indications for Use Comparison The subject and predicate device (Mint Product Family, K192423) share the same intended use of temporary fixation and suspension of facial soft tissue in an elevated position. Both devices are intended for subcutaneous placement to elevate and support cheek tissue, and therefore operate within the same intended use and mechanism of action. ## Technological Comparison The subject device and the predicate device (Mint Product Family, K192423) have shared technological characteristics, including: synthetic absorbable monofilament PDO suture material, bi-directional barbs along the length of the suture, implantable use with long-term (>30 days) tissue contact, ethylene oxide sterilization, and the same principle of action. Differences in technique of deployment and suture size range do not alter the fundamental mechanism of action or intended use. The cannula-based delivery method represents an alternative, well-established technique for subcutaneous placement in aesthetic procedures as it may reduce the risks of puncturing blood vessels and nerves as it travels beneath the skin. Additionally, the subject device is consistent with the reference device, DemeDIOX Barbed Absorbable Surgical Suture (K191361), in terms of available size range. Neither of these differences raise new questions of safety and effectiveness. | Characteristics | Subject Device Sutura Barbed PDO Suture | Predicate Device Mint Product Family (K192423) | Reference Device DemeDIOX Barbed Absorbable Surgical Suture (K191361) | Comparison | | --- | --- | --- | --- | --- | | FDA Classification | Class II | Class II | Class II | Same | | Product Code | NEW | NEW | NEW | Same | | Regulation Number | 21 CFR 878.4840 | 21 CFR 878.4840 | 21 CFR 878.4840 | Same | K252171 2 510(k) Summary {7} | Regulation Name | Suture, Surgical, Absorbable, Polydioxanone | Suture, Surgical, Absorbable, Polydioxanone | Suture, Surgical, Absorbable, Polydioxanone | Same | | --- | --- | --- | --- | --- | | Indication for Use | The Barbed PDO Suture is indicated for placement in the subcutaneous plane for temporary fixation of the cheek adipose layer in midface suspension surgery. | Indicated for use in mid-face suspension surgery to temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds. | Indicated for use in soft tissue approximation where use of absorbable sutures is appropriate. | The subject device's indication for use is similar to that of the predicate device in that they both intend to temporarily fixate and elevate the cheek sub dermis. The reference device's indication is broader, as it is intended for general soft tissue approximation. | | Material Composition | Polydioxanone (PDO) suture; Stainless steel (STS-304) | Polydioxanone (PDO) suture; Stainless steel (SUS 304) | Polydioxanone (PDO) suture; Stainless steel (STS-304) | The subject and predicate devices utilize the same PDO suture material and equivalent stainless steel materials (STS-304 and SUS-304), which are recognized designations for Type 304 stainless steel. No differences that would impact safety or performance. | | Suture Characteristic | Synthetic Absorbable Monofilament | Synthetic Absorbable Monofilament | Synthetic Absorbable Monofilament | Same | | Technique of Deployment | Cannula | Needle | Needle | Different; both are established techniques for subcutaneous placement in aesthetic procedures and enable accurate positioning of the suture within the intended tissue plane. This difference does not alter the fundamental mechanism of action or intended use and does not raise new questions | K252171 3 510(k) Summary {8} | | | | | of safety or effectiveness. | | --- | --- | --- | --- | --- | | **Technological Characteristics** | Bi-directional barbs along the long axis of the suture monofilament | Bi-directional barbs along the long axis of the suture monofilament | Uni-directional barbs along the long axis of the suture monofilament | Same as predicate device | | **Sterilization** | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Same | | **Absorbable** | Absorbable | Absorbable | Absorbable | Same | | **Size (USP)** | Available in sizes 5-0 to 2. | Size 1-0 | Available in sizes 5-0 to 2. | Different from predicate device; Same as reference device | | **Patient Contact** | Implant | Implant | Implant | Same | | **Duration of Contact** | Over 30 Days | Over 30 Days | Over 30 Days | Same | ## Non-Clinical and/or Clinical Tests Summary & Conclusions In accordance with the Class II Special Controls Guidance: Surgical Sutures; Guidance for Industry and FDA (June 2003), the Barbed PDO Suture was evaluated through nonclinical testing and supporting information to demonstrate safety and performance. These evaluations included conformity to applicable USP standards (USP <861> Suture Diameter and USP <881> Tensile Strength), sterility and bacterial endotoxin testing, shelf life supported by stability testing, biocompatibility assessment, and performance testing, including barb holding strength (pull-out testing). A prospective, multi-center clinical study was conducted to evaluate the safety and effectiveness of the Barbed PDO Suture for midface suspension surgery. A total of 57 subjects were enrolled across four U.S. sites, with 48 subjects completing the study. The safety profile observed in the clinical study was consistent with that expected for suture-based procedures. No serious adverse events or unanticipated adverse device effects were reported. Most adverse events were mild in severity and resolved without sequelae. Based on the totality of nonclinical and clinical evidence, the Barbed PDO Suture is safe and effective for its intended use and performs as well as the identified legally marketed predicate device. K252171 4 510(k) Summary
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...