K192423 · Hans Biomed Corporation · NEW · Sep 9, 2020 · General, Plastic Surgery
Device Facts
Record ID
K192423
Device Name
MINT Product Family
Applicant
Hans Biomed Corporation
Product Code
NEW · General, Plastic Surgery
Decision Date
Sep 9, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4840
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MINT™ is indicated for use in mid-face suspension surgery to temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.
Device Story
MINT™ is a sterile, synthetic, absorbable polydioxanone (PDO) monofilament surgical suture featuring bi-directional barbs along its long axis. It is designed for minimally invasive facial rejuvenation. The device is used by surgeons in a clinical setting to approximate and elevate soft tissues (cheek subcutaneous fat and SMAS layers) without requiring surgical knots. The barbs embed into the tissue upon placement, providing mechanical fixation. Some variants include stainless steel needles for insertion and threading within the dermis. The device provides a mechanical lift to reduce nasolabial fold depth. It is absorbed by the body over time, with tensile strength retention of approximately 50% at 6 weeks post-implantation. The device benefits patients by providing a temporary, minimally invasive alternative to traditional surgical mid-face lifts.
Clinical Evidence
Prospective clinical study of 62 subjects (male/female) evaluated nasolabial fold improvement using the Wrinkle Severity Rating Scale (WSRS) and Global Aesthetic Improvement Scale (GAIS). Primary endpoint: WSRS improvement (below -1 point) at 12 weeks. Results: 96.72% (59/62) of subjects showed improvement at 12 weeks. Mean WSRS reduction from baseline was statistically significant at all time points (4, 8, 12, 24 weeks; p<0.001). GAIS scores indicated high evaluator and subject satisfaction, with 86.89% of evaluators rating improvement as 'very much improved' at 12 weeks.
Technological Characteristics
Synthetic absorbable monofilament suture made of polydioxanone (PDO). Features bi-directional barbs (cog shape). Supplied with or without stainless steel (SUS 304) needles. Sterilized via Ethylene Oxide. Compliant with USP <861> (diameter), USP <871> (suture-needle attachment), and USP <881> (tensile strength).
Indications for Use
Indicated for mid-face suspension surgery in patients requiring temporary fixation of the cheek subcutaneous fat and SMAS layers to treat moderate to severe nasolabial folds.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 9, 2020
Hans Biomed Corporation % Michael Dun Senior Consultant, Quality and Regulatory Affair Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, Texas 78746
Re: K192423
Trade/Device Name: MINT Product Family Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable Polydioxanone Surgical Suture Regulatory Class: Class II Product Code: NEW Dated: June 18, 2020 Received: August 10, 2020
Dear Michael Dun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192423
Device Name MINTTM
Indications for Use (Describe)
MINT™ is indicated for use in mid-face suspension surgery to temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
## MINT™ Product Family
#### 1. Submission Sponsor
Hans Biomed Corporation
7, Jeongui-ro 8-gil, Songpa-gu
Seoul, 05836
Republic of Korea
Contact: Lucy Choi
Title: General Manager
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327.9997
Contact: Michael Dun
Title: Senior Consultant, Quality and Regulatory Affairs
#### 3. Date Prepared
9 September2020
#### 4. Device Identification
| Trade/Proprietary Name: | MINT™ Product Family |
|-------------------------|------------------------------------------|
| Common/Usual Name: | Absorbable surgical suture |
| Classification Name: | Absorbable polydioxanone surgical suture |
| Regulation Number: | 21 CFR §878.4840 |
{5}------------------------------------------------
| Product Code: | NEW |
|-----------------------|---------------------------|
| Device Class: | Class II |
| Classification Panel: | General & Plastic Surgery |
#### 5. Legally Marketed Predicate Device(s)
The primary predicate is the Silhouette InstaLift™ device cleared under K163676 by Silhouette Lift, Inc. The reference device is the original MINT™ device cleared under K130191 by Hans Biomed Corporation. Neither the predicate nor reference devices have been subject to a design-related recall.
#### 6. Indication for Use Statement
MINT™ is indicated for use in mid-face suspension surgery to temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.
#### 7. Device Description
The MINT™ device is a sterile synthetic absorbable surgical suture comprised of polydioxanone, (GHoOJ)n. Polydioxanone has been found to be nonantigenic and to elicit only a slight tissue reaction during absorption. The pigment of the violet dye is D&C Violet No.2 (21CFR §74.3602). MINT™ is available in a range of gauge sizes and lengths.
Each suture has bi-directional barbs along the long axis of the suture monofilament. The MINT™ Synthetic Absorbable PDO suture approximates tissues, without the need to tie surgical knots, by using the opposing barbs on the suture surface to embed in the tissues after the surgeon precisely places the suture within the tissues. Barbed suture lifting is a minimally invasive surgical technique for facial rejuvenation
While the formation of barbs in the MINT™ reduces the tensile strength relative to non-barbed suture material of the same size, tying knots in non-barbed suture materials also reduce their effective strength. For this reason, the strength of the MINT™ can be compared with USP knot strength of nonbarbed sutures and the USP size of MINT™ is 1 while its tensile strength is equivalent to that of USP 2-0, as demonstrated under the original 510(k) for the MINT™ device (K130191).
The MINT™ Product Family includes models which are supplied with needles. For those models supplied with a needle, designated as "Lifting Thread Combined with needle" in the MINT™ Product Family instructions for use, the needle is used to make the insertion point and for threading of the suture in the patient's dermis.
#### 8. Substantial Equivalence Discussion
{6}------------------------------------------------
The MINT™ Product Family is currently indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate (K130191). The purpose of this submission is to obtain clearance of a new indication for the MINT™ Product Family for use in "lifting" procedures which includes variants supplied with needles. Due to the difference in intended use, the cleared MINT™ device family (K130191) is presented as a reference device in this submission while the cleared Silhouette InstaLift™ device family (K163676) is presented as the predicate as it has an identical intended use to that for the subject MINT™ device and includes variants supplied with identical composition to the needles supplied with certain variants of the subject device. Additionally, while the predicate device is comprised of a different material (poly glycolide/L-lactide) and has a different number of barbs to that used in the subject device (polydioxanone (PDO)) questions of biocompatibility and clinical effectiveness are identical. Therefore, different questions of safety and effectiveness are not raised with the subject device in relation to these difference in technological characteristics.
While there are differences in the indications for use between the subject MINT™ Product Family and the predicate Silhouette InstaLift™ device family (K163676), these differences do not constitute differences in intended use or raise different questions of safety and effectiveness. Both devices are absorbable surgical sutures for use in soft tissue indicated for mid-face suspensions surgery to temporarily fixate the cheek in an elevated position. Both devices achieve nasolabial fold depth reduction as it is a clinical outcome of mid-face suspension surgery, as reported in the following published literature, copies of which accompany this submission:
1. Noone, R.B. (2006) Suture Suspension Malarplasty with SMAS Plication and Modified SMASectomy: A Simplified Approach to Midface Lifting. Plast. Reconstr. Surg. 117: 792.
2. Benito, J., et al. (2011) Facial Rejuvination and Improvement of Malar Projection Using Sutures with Absorbable Cones: Surgical Technique and Case Series. Aesth. Plast. Surg. 35: 248-253.
3. Paul, M. D. et al. (2006) The Evolution of the Midface Lift in Aesthetic Plastic Surgery. Plast. Reconstr. Surg. 117: 1809-1827.
The following table compares the MINT™ Product Family to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device and shown by clinical and other scientific data presented in this submission.
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Hans Biomed Corporation | SILHOUETTE LIFT, INC | |
| Trade Name | MINT™ | Silhouette InstaLift™ | |
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
| | Hans Biomed Corporation | SILHOUETTE LIFT, INC | |
| Trade Name | MINT™ | Silhouette InstaLift™ | |
| 510(k)<br>Number | K192423 | K163676 | Not applicable |
| Product Code | NEW | GAM | Different;<br>The difference in Product<br>Code does not raise new<br>questions of safety and<br>effectiveness. Both product<br>codes cover absorbable<br>surgical sutures. |
| Regulation<br>Number | 21 CFR 878.4840 | 21 CFR 878.4493 | Different;<br>The difference in<br>Regulation Number does<br>not raise new questions of<br>safety and effectiveness.<br>Both regulations cover<br>absorbable, sterile flexible<br>surgical sutures that are<br>intended for use in soft<br>tissue approximation,<br>coated or uncoated, and<br>with or without a standard<br>needle attached. |
| Regulation<br>Name | Suture, Surgical,<br>Absorbable, Polydioxanone | Suture, Absorbable,<br>Synthetic, Polyglycolic Acid | Different;<br>The difference in<br>Regulation Name does not<br>raise new questions of<br>safety and effectiveness.<br>Both regulations cover<br>absorbable, sterile flexible<br>surgical sutures that are<br>intended for use in soft<br>tissue approximation,<br>coated or uncoated, and<br>with or without a standard<br>needle attached. |
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
| | Hans Biomed Corporation | SILHOUETTE LIFT, INC | |
| Trade Name | MINT™ | Silhouette InstaLift™ | |
| | | | needle attached. |
| Intended Use | Absorbable surgical sutures<br>for use in soft tissue | Absorbable surgical sutures<br>for use in soft tissue | Identical |
| Indication for<br>use | MINT™ is indicated for use<br>in mid-face suspension<br>surgery to temporarily<br>fixate the cheek<br>subcutaneous fat layer and<br>SMAS layer in an elevated<br>position for the treatment<br>of moderate to severe<br>nasolabial folds. | The Silhouette InstaLift™<br>device is indicated for use in<br>mid-face suspension surgery<br>to temporarily fixate the<br>cheek subdermis in an<br>elevated position. | Similar. While there are<br>differences, these do not<br>constitute differences in<br>intended use or raise<br>different questions of<br>safety and effectiveness.<br>Both devices are<br>absorbable surgical sutures<br>for use in soft tissue<br>indicated for mid-face<br>suspension surgery to<br>temporarily fixate the<br>cheek in an elevated<br>position. Both devices<br>achieve nasolabial fold<br>depth reduction as it is a<br>clinical outcome of mid-<br>face suspension surgery, as<br>reported in published<br>literature referenced in this<br>510(k). |
| Raw Material | Polydioxanone suture<br>Stainless steel (SUS 304)<br>needle | Poly glycolide/L-lactide<br>suture<br>Stainless steel (SUS 304)<br>needle | Different; While there are<br>differences in suture<br>material to the predicate<br>device K163676, these do<br>not raise new questions of<br>safety and effectiveness.<br>Biocompatibility data on file<br>for the reference MINT™<br>device family K130191<br>demonstrate the device's |
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
| | Hans Biomed Corporation | SILHOUETTE LIFT, INC | |
| Trade Name | MINT™ | Silhouette InstaLift™ | biocompatibility while<br>clinical data presented in<br>this submission also<br>demonstrate equivalent<br>safety and performance. |
| Suture<br>Characteristic | Synthetic Absorbable<br>Monofilament | Synthetic Absorbable<br>Monofilament | Identical |
| Technique of<br>Deployment | Needle-based deployment | Needle-based deployment | Identical;<br>Both include variants<br>supplied with needles. |
| Technological<br>Characteristic | Bi-directional barbs along<br>the long axis of the suture<br>monofilament | Bi-directional barbs along the<br>long axis of the suture<br>monofilament | Identical |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Identical |
| Size (USP) | 1-0 | 1-0 | Identical |
| Absorbable | Absorbable | Absorbable | Identical |
| Patient<br>contact | Implant | Implant | Identical |
| Duration of<br>contact | Over 30 Days | Over 30 Days | Identical |
| Number of<br>barbs per<br>linear length<br>of suture | 20 barbs per 10cm in all<br>sutures. | 8 barbs in 30cm sutures.<br>12 barbs in 27.5cm sutures.<br>16 barbs in 26.8cm sutures. | Different; While there are<br>differences in the number<br>of barbs to the predicate<br>device K163676, these<br>devices are substantially<br>equivalent as no new<br>questions of safety and<br>effectiveness are raised, as<br>demonstrated by barb<br>holding strength testing<br>and clinical data presented<br>in this submission. |
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
| | Hans Biomed Corporation | SILHOUETTE LIFT, INC | |
| Trade Name | MINTTM | Silhouette InstaLiftTM | |
| Barb Holding Strength<br>(Average ± Standard Deviation) | $20.94 ± 3.93 N$ | $9.90 ± 0.93 N$ | Different; While there are differences in the average barb holding strength to predicate device K163676, the differences do not raise new questions of safety and effectiveness as the average barb holding strength for the subject MINTTM device is superior to that of the predicate. Clinical data presented in this submission also demonstrate equivalent safety and performance. |
| Barb shape | Cog shape | Cog shape | Identical |
| Barb size | 0.601mm | 0.601mm | Identical |
| Barb direction | A section and B section is opposite direction | A section and B section is opposite direction | Identical |
| Pattern of the barbs | Bi-directional barbs along the long axis of the suture | Bi-directional barbs along the long axis of the suture | Identical |
| Suture diameter | Compliant with USP <861> requirements | Compliant with USP <861> requirements | Identical |
| Suture tensile strength | Compliant with USP <881> requirements | Compliant with USP <881> requirements | Identical |
| Suture-Needle attachment | Compliant with USP <871> requirements | Compliant with USP <871> requirements | Identical |
#### Table 5A - Comparison of Characteristics
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### 9. Non-Clinical Performance Data
Additional testing was conducted and relevant scientific data collated for this 510(k). Previous tests performed on the MINT™ reference device (K130191) included those for overall design, sterilization, biocompatibility, and the physical and performance testing as described in the guidance entitled Class II Special Controls Guidance Document: Surgical Sutures, confirming that the design output meets the
{11}------------------------------------------------
design inputs and specifications for the sutures. These data also apply to the subject device, in all aspects except the expanded indications for use and provision of variants with needles.
The non-clinical performance testing performed on the subject device MINT™ Product Family variants including needles included physical and performance and real-time stability testing was focused on generating scientific data to support the inclusion of the needles in addition to generating scientific data to support the expansion of suture variants in order to demonstrate compliance with USP <861> Suture diameter, USP <871> Suture-Needle attachment, and USP <881> Suture tensile strength requirements.
Additional comparative barb holding strength testing was performed to demonstrate that the MINT™ Product Family displays superior barb holding strength to that for the Silhouette InstaLift™ predicate device (K163676).
The results of non-clinical performance testing demonstrate substantial equivalency between the MINT™ Product Family (K192423) and the predicate Silhouette InstaLift™ device (K163676) as they both comply with all the requirements of USP <861> Suture diameter, USP <871> Suture-Needle attachment, and USP <881> Suture tensile strength, and that the MINT™ Product Family displays superior barb holding strength to that for the Silhouette InstaLift™.
#### 10. Animal Performance Data
In vivo studies in animal models to demonstrate absorption and mechanical strength of the subject device were performed.
The in vivo absorption study was completed on 30 rats prepared and anesthetized in accordance with the procedure and was performed on Monosorb, identical to the sutures used in the MINT™ Product Family. The results of this study indicated in vivo absorption of the sutures occurs between 180°220 days post-implantation.
The in vivo mechanical strength study was completed on rats prepared and anesthetized in accordance with the procedure and performed on MINT™ barbed suture. The results of this study indicated that tensile strength retention of MINT™ (USP 1) was 44.5% at 6 weeks and was unable to be measured at 10 and 12 weeks post-implantation due to severe degradation. Based on the results of this study, the breaking strength retention of MINT™ is considered to be approximately 50% at 6 weeks.
#### 11. Clinical Performance Data
A prospective clinical study was conducted to evaluate safety and effectiveness of the MINT™ Product Family to support mid-face suspension surgery to fixate the cheek sub dermis in an elevated position. The aim of this study was to evaluate the surgical outcomes associated with using MINT™ for improving nasolabial folds.
{12}------------------------------------------------
In this study, 62 male and female subjects were assessed pretreatment and compared to 4, 8, 12, and 24 weeks following treatment using the previously validated and published 5-grade Wrinkle Severity Rating Scale (WSRS) score and rating by independent, blinded assessors. The correction of nasolabial folds using MINT™ was evaluated using blinded evaluator Global Aesthetic Improvement Scale (GAIS) ratings.
The primary effectiveness evaluation was carried out at 12 weeks after surgery by comparing the photos of nasolabial folds of the patients in the study group by blinded evaluators. Twenty-four weeks were set for final safety and effectiveness evaluation period. The primary effectiveness evaluation was carried out at 12 weeks after surgery by comparing the photos of the patients in the study group.
The ratio of subjects in the study group who showed wrinkle improvement (WSRS: below -1 point) by independent evaluation at 12 weeks after the application of test device, the primary effectiveness endpoint, was analyzed. The results showed 59 subjects (96.72%) in FA set showed improvement (WSRS: below -1 point) and the lowest confidence level of 20.95% was greater than 0, indicating superiority.
The results of secondary effectiveness evaluation showed the ratio of the subjects in FA who showed improvement (WSRS: below -1 point) by the evaluation of independent evaluators after the application of test device was 100% (61 subjects) at week 4, 96.72% (59 subjects) at week 8 and 90.16% (55 subjects) at week 24. The ratio of the subjects in FA who showed improvement (WSRS: below -1 point) by the evaluation of testers was 100% (61 subjects) at week 4, 8, 12 and 24. This tendency was represented in the results of an analysis of PP test.
When comparing the WSRS evaluated by independent evaluators at 4, 8, 12 and 24 weeks after the application of test device with baseline of FA set was 3.30 ± 0.45 points; the difference in the mean of WSRS in comparison with baseline at each point was -1.56 ± 0.43 points at week 4, 1.41 ± 0.48 points at week 8, -1.23 ± 0.41 points at week 12, and -1.20 ± 0.45 points at week 24, indicating a decrease over time and statistically significant difference in comparison with baseline at all points of time (p<0.001). When comparing WSRS evaluated by testers at 4, 8, 12 and 24 weeks after the application of test device with the baseline by point of time, the mean of WSRS in FA set before the application of medical device was 3.21 ± 0.41 points; the difference in the mean of WSRS in comparison with the baseline at each point of time was -1.23 ± 0.42 points at week 4 and -1.25 ± 0.43 points at 8 weeks, indicating an increase in wrinkle improvement. It was -1.18 ± 0.39 points at week 12 and -1.16 ± 0.37 points at week 24 after the application of test device, indicating that the difference in the mean of WSRS after 12 weeks became smaller than that at week 8. Also, there was the statistically significant difference in comparison with the baseline at all points of time (p<0.0001).
The mean of GAS in FA set assessed by evaluators at 12 weeks after the application of test device in comparison with baseline was 1.93 ± 0.36 points at week 12 and 1.85 ± 0.44 points at week 24, indicating the high satisfaction level of evaluators. When analyzing the distribution of GAIS point evaluated by evaluators at each point of time, 2 points (very much improved) took the greatest
{13}------------------------------------------------
proportion- 53 subjects (86.89%) at week 12, and 48 subjects (78.69%) at week 24. The mean of GAIS evaluated by the subject at 12 and 24 weeks after the application of test device was 1.92 ± 0.63 points in FA set, indicating a high satisfaction level of subjects. When analyzing the distribution of GAIS point evaluated by evaluators at each point of time, 2 points took the greatest proportion - 35 subjects (57.38%) at week 12 and 32 subjects (52.46%) at week 24.
Results of the clinical investigation support the indications for the MINT™ Product Family to support mid-face suspension surgery to fixate the cheek sub dermis in an elevated position and is effective for the improvement of facial wrinkles. Clinical study conclusion confirms that the device is safe and effective as used according to the instructions for use.
#### 12. Statement of Substantial Equivalence
The subject device is identical to the predicate in relations for use. The indications for use for the subject device were evaluated in a clinical study which demonstrated acceptable performance and safety, thereby supporting a finding of substantial equivalence. This clinical data also demonstrates that no different questions of safety and effectiveness were raised for differences in the number of barbs between the subject and predicate devices. Differences in raw material are addressed by the data presented in the 510(k) for the reference device along with the presented non-clinical, animal, and clinical performance data. The needles supplied with the subject device were evaluated in pre-clinical (bench) performance testing which demonstrate acceptable performance and safety.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.