Sleepnet Arie Full Face Vented Mask; Sleepnet Arie Nasal Vented Mask
Applicant
Sleepnet Corporation
Product Code
BZD · Anesthesiology
Decision Date
Jan 15, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sleepnet Arie Full Face Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 4 cm H2O. The mask is to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. The mask is intended for single-patient multi-use in the home, hospital, or institutional environment. The Sleepnet Arie Nasal Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 4 cm H2O. The mask is to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. The mask is intended for single-patient multi-use in the home, hospital, or institutional environment.
Device Story
Interface for CPAP/BiPAP machines; delivers pressurized air to patient airway. Composed of injection-molded thermoplastics and liquid silicone rubber (LSR) cushion; secured to face via adjustable headgear. Vented frame/tubing assembly includes anti-asphyxia valve (AAV), swivel connector, and tubing. Used in home, hospital, or institutional settings by patients. Provides non-invasive ventilation support; helps maintain airway patency during sleep. Clinicians use device output (pressure/flow) to manage therapy settings. Benefits include improved patient compliance and effective delivery of prescribed pressure.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 17510, ISO 3744, ISO 5356, IEC 60068-2-31, and ISO 18562. Testing included CO2 rebreathing (met ISO criteria), exhaust flow, pressure drop, dead space, AAV pressure/resistance, and sound levels. Biocompatibility confirmed via ISO 10993 series (cytotoxicity, sensitization, irritation, chemical characterization).
Technological Characteristics
Materials: Polycarbonate, Liquid Silicone Rubber (LSR), Polypropylene, TPE. Interface: Face-sealing mask with vented frame/tubing. Energy: Passive (powered by external PAP device). Connectivity: None. Sterilization: Provided non-sterile. Standards: ISO 17510 (sleep apnea therapy), ISO 10993 (biocompatibility), ISO 18562 (gas pathway biocompatibility).
Indications for Use
Indicated for adult patients (>30kg) requiring positive airway pressure therapy (CPAP or bi-level) at or above 4 cm H2O.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 15, 2026
Sleepnet Corporation
% Todd Courtney
Vice President, Anesthesia, Respiratory, Sleep, and ENT Regulatory Affairs
Mcra, LLC
803 7th Street, NW
3rd Floor
Washington, District of Columbia 20001
Re: K251847
Trade/Device Name: Sleepnet Arie Full Face Vented Mask; Sleepnet Arie Nasal Vented Mask
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous Ventilator (IPPB)
Regulatory Class: Class II
Product Code: BZD
Dated: December 11, 2025
Received: December 15, 2025
Dear Todd Courtney:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251847 - Todd Courtney
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251847 - Todd Courtney
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Binoy J. Mathews -S
Digitally signed by
Binoy J. Mathews -S
Date: 2026.01.15
11:23:52 -05'00'
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251847 | |
| Device Name Sleepnet Arie Full Face Vented Mask; Sleepnet Arie Nasal Vented Mask | |
| Indications for Use (Describe) The Sleepnet Arie Full Face Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 4 cm H2O. The mask is to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. The mask is intended for single-patient multi-use in the home, hospital, or institutional environment. The Sleepnet Arie Nasal Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 4 cm H2O. The mask is to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. The mask is intended for single-patient multi-use in the home, hospital, or institutional environment. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
Device Trade Name: Arie Full Face Vented Mask and Arie Nasal Vented Mask
Manufacturer: Sleepnet Corporation
5 Merrill Industrial Drive
Hampton, NH 03842
Contact: Todd Courtney
Vice President, Anesthesia, Respiratory, Sleep, and ENT Regulatory Affairs
MCRA, LLC
Office: (202) 552-5800
Prepared by: MCRA, LLC
803 7th Street, NW, 3rd Floor
Washington, DC 20001
Office: 202.552.5800
Date Prepared: January 15, 2026
Classifications: 21 CFR 868.5905; Noncontinuous ventilator (IPPB)
Class: II
Product Codes: BZD
510(k) Number: K251847
Primary Predicate: AirFit F20 Full Face Mask, K153563 (for Arie Full Face Vented Mask)
AirFit N20 Nasal Mask, K171212 (for Arie Nasal Vented Mask)
Reference Device: F&P Evora Full Face Mask, K212371 (for Arie Full Face Vented Mask only)
Indications For Use:
Arie Full Face Vented Mask
The Sleepnet Arie Full Face Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 4 cm H₂O. The mask is to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. The mask is intended for single-patient multi-use in the home, hospital, or institutional environment.
{5}
# Arie Nasal Vented Mask
The Sleepnet Arie Nasal Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 4 cm H2O. The mask is to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. The mask is intended for single-patient multi-use in the home, hospital, or institutional environment.
## Device Description:
The Sleepnet Arie Full Face Vented and Nasal Mask are designed based on the human facial shape, and the operating characteristics during application. The Sleepnet mask utilizes a Liquid Silicone Rubber (LSR) cushion design. The device is composed of medical-grade injection molded thermoplastics & clear liquid silicone rubber. The device is used as the interface between a CPAP/BiPAP machine and the end user. The masks are held in place with adjustable headgear that straps the mask to the face.
The Arie Full Mask is accompanied by a vented frame/tubing assembly which includes the Arie Full Face Vented Mask Headgear connector, valve, swivel ball, tubing, swivel connector assembly (swivel connector and clear swivel), and the JustFit Full Face Headgear with clips.
## Predicate Device:
Sleepnet submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, Arie Full Face Vented and Nasal Vented Mask is substantially equivalent in indications, design principles, and performance to the following predicate devices, which have been determined by FDA to be substantially equivalent to pre-amendment devices:
Primary Predicate: AirFit F20 Full Face Mask, K153563 (for Arie Full Face Vented Mask)
AirFit N20 Nasal Mask, K171212 (for Arie Nasal Vented Mask)
Reference Device: F&P Evora Full Face Mask, K212371 (for Arie Full Face Vented Mask only)
## Performance Testing Summary:
The following tests have been performed on the Arie Full Face Vented and Nasal Vented Masks:
- CO₂ Rebreathing
- Exhaust Flow Test
- Pressure Drop Test
- Dead Space Measurement
- Anti-Asphyxia Valve Pressure Test
- Anti-Asphyxia Valve Breathing in Resistance Test
- Anti-Asphyxia Valve Life Test
- Sound Test
{6}
Tests were conducted according to recognized standards:
- ISO 17510: Medical devices — Sleep apnoea breathing therapy — Masks and application accessories
- ISO 3744: Acoustics — Determination of sound power levels and sound energy levels of noise sources using sound pressure — Engineering methods for an essentially free field over a reflecting plane
- ISO 5356: Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and sockets
- IEC 60068-2-31: Environmental testing — Part 2-31: Tests — Test Ec: Rough handling shocks, primarily for equipment-type specimens
- ISO 17664: Processing of health care products — Information to be provided by the medical device manufacturer for the processing of medical devices
- ISO 10993-1: Biological evaluation of medical devices. Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process
- ISO 10993-5: Biological evaluation of medical devices. Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: Biological evaluation of medical devices. Part 10: Tests for skin sensitization
- ISO 10993-12: Biological evaluation of medical devices. Part 12: Sample preparation and reference materials
- ISO 10993-18:2020/Amd1:2022: Biological evaluation of medical devices. Part 18: Chemical characterization of medical device materials within a risk management process
- ISO 10993-23: Biological evaluation of medical devices. Part 23: Tests for irritation
- ISO 18562-1:2017: Biocompatibility evaluation of breathing gas pathways in healthcare applications. Part 1: Evaluation and testing within a risk management process
- ISO 18562-2: Biocompatibility evaluation of breathing gas pathways in healthcare applications. Part 2: Tests for emissions of particulate matter
- ISO18562-3: Biocompatibility evaluation of breathing gas pathways in healthcare applications. Part 3: Tests for emissions of volatile organic substances
## Substantial Equivalence:
The Arie Full Face Vented and Nasal Vented Masks are substantially equivalent to AirFit F20 Full Face Mask (K153563) and AirFit N20 Nasal Mask (K171212) with respect to intended use, patient population, exhaust flow, AAV pressure, pressure drop, and therapy pressure range.
{7}
Table 1: Substantial Equivalence -Arie Full Face Vented Mask
| | Subject Device | Primary Predicate | Reference Device | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- |
| Device Name | Arie Full Face Vented Vented Mask | Resmed AirFit F20 Full Face Mask | F&P Evora Full Face Mask | N/A |
| Company | Sleepnet Corporation | Resmed Ltd | Fisher & Paykel Healthcare Ltd | N/A |
| 510(k) Number | K251847 | K153563 | K212371 | N/A |
| Classification | 21 CFR 868.5905 | 21 CFR 868.5905 | 21 CFR 868.5905 | Identical |
| Product Code | BZD | BZD | BZD | Identical |
| Common/Usual Name | CPAP/BiPAP Mask | Full Face Mask | Full Face Mask | Identical |
| Principle of Operation | Seal over the nose and mouth to allow delivery of pressurized air from a PAP device. Interface for CPAP device or Bi-level/Auto-PAP devices | Interface for CPAP device or Bi-level/Auto-PAP devices | Interface for CPAP device or Bi-level/Auto-PAP devices | Identical |
| Patient Use Type | Adult Subjects >30kg | Adult Subjects >30kg | Adult Subjects >30kg | Identical |
| Indications for Use | Arie Full Face Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 4 cm H2O. The mask is to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. The mask is intended for single-patient multi-use in the home, hospital, or institutional environment. | The AirFit F20 is a non-invasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel system. The AirFit F20 is: • to be used by patients weighing more than 66 lb (30 kg) for whom positive airway pressure therapy has been prescribed • intended for single-patient reuse in the home environment and multi-patient reuse in the hospital/institutional environment. | The F&P Evora Full Face mask is intended to be used by adults weighing ≥ 66lbs (30kgs) who have been diagnosed by a physician as requiring CPAP or Bi-Level therapy. The F&P Evora Full Face mask is intended for single patient use in the home and for multiple patient use in the hospital or other clinical setting where proper disinfection of the device can occur between patient uses | Comparable. The proposed devices's indications are a subset of the proposed predicates, as the subject device is not intended to be multi-patient use, which reduces the overall risk of the subject device compared to the predicate. |
| Environment | Home, hospital or institutional environment | Home, hospital or institutional environment | Home, hospital or institutional environment | Identical |
{8}
| | Subject Device | Primary Predicate | Reference Device | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- |
| Operating and Storage Conditions | Operating Temperature: 5°C to 40°C
Storage Temperature: -20°C to 60°C | Not Publicly Available | Operating Temperature: 5°C to 40°C
Storage Temperature: -20°C to 50°C | Identical to reference |
| Reuse | Single patient, multi-use | Single patient, Multi-Use
Multiple patient use | Single patient, Multi-Use
Multiple patient use | Comparable. The subject device is not intended to be used on multiple patients, only single patient, multi-use which reduces the overall risk of the subject device compared to the predicate. |
| Useful Life | 12 months | Not Publicly Available | 24 months | N/A |
| Anatomical site | Face (seals around the nose and mouth) | Face (seals around the nose and mouth) | Face (seals around the nose and mouth) | Identical |
| Mask Size | Three sizes available – Small, Medium, Large | Offered in various sizes | Three sizes available – Extra Small, Small-Medium, Large | Similar |
| Mask Weight | Small – 68 grams
Medium – 75 grams
Large – 85 grams | Not Publicly Available | Not Publicly Available | N/A |
| Exhaust flow characteristics | Pressure (cm H2O) Flow (lpm)
4 28.5
10 39.1
20 54.9
30 67.2
40 81.2 | Not Publicly Available | Not Publicly Available | The exhaust flow plays a critical role in washing out the exhaled CO2. The CO2 washout test was successfully completed on the Large size of the subject device, thus proving that the exhaust flow is acceptable. |
| Mask Dead Space | Small: 140 mL
Medium: 225 mL
Large: 240 mL | Not Publicly Available | XS: 165.2 cm³
S-M: 162.6 cm³
L: 164.1 cm³ | Comparable. The dead space also plays a critical role in washing |
{9}
| | Subject Device | Primary Predicate | Reference Device | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- |
| | | | | out the exhaled CO₂. The CO₂ washout test was successfully completed on the Large size of the subject device, thus proving that the dead space is acceptable. |
| Sound Power Level | • A-weighted Sound Power Level: 32.1 dBA
• A-weighted Sound Pressure Level at 1m distance: 24.1 dBA | Not Publicly Available | • A-weighted Sound Power Level of the Mask: 28.2 dBA, with uncertainty 2.5 dBA
• Sound Pressure Level of the Mask: 20.2 dBA, with uncertainty 2.5 dBA | Similar |
| AAV pressure | • AAV opening pressure: 1.16 cm H₂O
• AAV closing pressure: 1.82 cm H₂O | Not Publicly Available | Not Publicly Available | ISO 17510 requires that the AAV opening pressure be less than the minimum rated therapy pressure. As seen, the opening pressure is less than 4 cm H₂O for the subject device. |
| Sterility | Provided Non-Sterile | Provided Non-Sterile | Provided Non-Sterile | Identical |
| Validated Cleaning (Single-Patient Use) | Mild soap and warm water. | Not Publicly Available | Not Publicly Available | Cleaning instructions are validated and include mild soap/detergent and warm water. |
| Validated Disinfection | N/A | Not Publicly Available | Thermal Disinfection: 90°C for 1 min | Different. The subject device is not intended to be used on multiple patients so disinfection is not required. |
{10}
| | Subject Device | | Primary Predicate | Reference Device | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- | --- |
| Therapy Pressure Range | 4 to 40 cm H₂O. | | Not Publicly Available | 4 to 30 cmH₂O | Similar |
| Resistance to Flow (Pressure drop) | • Pressure drop at 50 L/min: 1.05 cm H₂O
• Pressure drop at 100 L/min: 3.09 cm H₂O | | Not Publicly Available | • Pressure drop at 50 L/min: 0.9 ± 0.3 cm H₂O
• Pressure drop at 100 L/min: 2.2 ± 0.3cm H₂O | Comparable. The pressure drop is slightly higher for the subject device than for the reference predicate. There are no acceptance criteria in ISO 17510 for pressure drop. The pressure drop values are typically disclosed in the IFU’s as a reference for the clinicians so that the therapy pressure setting can be compensated accordingly.
The F&P Evora has a more comparable pressure drop to the subject device and is included as a reference device. |
| Materials | Polycarbonate, Liquid Silicone Rubber, Polypropylene, TPE | | Not Publicly Available | Not Publicly Available | Biocompatibility was tested via FDA recognized standards |
| CO₂ Rebreathing | Condition | ETCO₂: Percentage Relative Increase over Baseline | Not Publicly Available | Not Publicly Available | The subject device meets ISO 17510 acceptance criteria for CO₂ rebreathing. Maximum allowable increase in ETCO₂ in normal condition is 20% and in single fault condition is 60%. |
| | Normal: 4 cm H₂O
5 cm H₂O
10 cm H₂O | 16%
14%
12% | | | |
| | Single Fault: SF1 – Open end
SF2 - Occluded | 35%
36% | | | |
{11}
Table 2: Substantial Equivalence -Arie Nasal Vented Mask
| | Subject Device | Proposed Predicate | Comparison |
| --- | --- | --- | --- |
| Device Name | Arie nasal vented mask | AirFit N20 Nasal Mask | N/A |
| Company | Sleepnet Corporation | Resmed Ltd | N/A |
| 510(k) Number | K251847 | K171212 | N/A |
| Classification | 21 CFR 868.5905 | 21 CFR 868.5905 | Identical. |
| Product Code | BZD | BZD | Identical. |
| Common/Usual Name | CPAP/BiPAP Mask | Vented Nasal Mask | Similar |
| Principle of Operation | Seal over the nose to allow delivery of pressurized air from a PAP device. Interface for CPAP or Bi-level/Auto-PAP devices | Seal over the nose to allow delivery of pressurized air from a PAP device. Interface for CPAP or Bi-level/Auto-PAP devices | Identical. |
| Patient Use Type | Adult Subjects >30kg | Adult Subjects >30kg | Identical. |
| Indications for Use | Arie Nasal Vented Mask is intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 4 cm H2O. The mask is to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. The mask is intended for single-patient multi-use in the home, hospital, or institutional environment. | The AirFit N20 channels airflow non-invasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel device. The AirFit N20 is: • to be used by patients weighing more than 66 lb (30 kg) for whom positive airway pressure has been prescribed • intended for single patient re-use in the home environment and multi-patient re-use in the hospital / institutional environment | Comparable. The proposed devices' indications are a subset of the proposed predicates, as the subject device is not intended to be multi-patient use. |
| Environment | Home, hospital or institutional environment | Home, hospital or institutional environment | Identical. |
| Operating and Storage Conditions | Operating Temperature: 5°C to 40°C Storage Temperature: -20°C to 60°C | Not Publicly Available | Identical. |
{12}
| | Subject Device | Proposed Predicate | Comparison |
| --- | --- | --- | --- |
| Reuse | Single patient, multi-use | • Single patient, Multi-Use • Multiple patient use | Comparable. The subject device is not intended to be used on multiple patients, only single patient, multi-use. |
| Useful Life | 12 months | Not Publicly Available | N/A |
| Anatomical site | Face (seals around the nose) | Face (seals around the nose) | Identical. |
| Mask Size | Three size available – Small, Medium, Large | Offered in various sizes | Identical. |
| Mask Weight | Small – 41 grams Medium – 44 grams Large – 46 grams | Not Publicly Available | N/A |
| Exhaust flow characteristics | Pressure (cm H2O) Flow (lpm) 4 23.0 10 35.9 20 50.5 30 63.8 | Not Publicly Available | The exhaust flow plays a critical role in washing out the exhaled CO2. The CO2 washout test was successfully completed on the Large size of the subject device, thus proving that the exhaust flow is acceptable. |
| Mask Dead Space | Small: 60 mL Medium: 75 mL Large: 95 mL | Not Publicly Available | The dead space also plays a critical role in washing out the exhaled CO2. The CO2 washout test was successfully completed on the Large size of the subject device, thus proving that the dead space is acceptable. |
| Sound Power Level | A-weighted sound power level: 28.4 dBA. A-weighted sound pressure level at 1m distance: 20.4 dBA. | Not Publicly Available | Comparable. The sound power levels of the machines are within SIO 17510 limits |
| Sterility | Provided Non-Sterile | Provided Non-Sterile | Identical. |
| Validated Cleaning (Single-Patient Use) | Mild soap and warm water; | Not Publicly Available | Validated cleaning instructions include mild soap/detergent and warm water. |
| Validated Disinfection | N/A | Not Publicly Available | The subject device is not intended to be used on multiple patients so disinfection is not required. |
| Therapy Pressure Range | 4 to 30 cm H2O. | Not Publicly Available | Subject device performance has been evaluated per applicable ISO 17510 requirements across the stated pressure range. |
{13}
| | Subject Device | | Proposed Predicate | Comparison |
| --- | --- | --- | --- | --- |
| Resistance to Flow (Pressure drop) | • Pressure drop at 50 L/min: 0.43 cm H₂O
• Pressure drop at 100 L/min: 1.98 cm H₂O | | Not Publicly Available | There are no acceptance criteria in ISO 17510 for pressure drop. The pressure drop values are typically disclosed in the IFU’s as a reference for the clinicians and RT’s so that the therapy pressure setting can be compensated accordingly. |
| Materials | Polycarbonate, Liquid Silicone Rubber, Polypropylene, TPE | | Not Publicly Available | Materials were validated via FDA recognized biocompatibility testing. |
| CO₂ Rebreathing | Condition | ETCO₂: Percentage Relative Increase over Baseline | Not Publicly Available | The subject device meets ISO 17510 acceptance criteria was CO₂ rebreathing. Maximum allowable increase in ETCO₂ in normal condition is 20%.
Note – As per ISO 17510, CO₂ rebreathing testing in single fault condition is not required for a nasal mask. |
| | Normal:
4 cm H₂O
5 cm H₂O
10 cm H₂O | 17%
11%
3% | | |
{14}
Conclusion:
The subject device and the predicate devices have the same intended use, have similar technological characteristics, and are made of similar materials. The subject and predicate devices are packaged in similar materials and are sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.