The iFuse TORQ Navigation instruments are intended to be used with the iFuse TORQ Implant System and the Medtronic StealthStation System to assist the surgeon in precisely locating anatomical structures in iFuse TORQ Implant System procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy. The TORQ iGPS instruments and iGPS Drill Bits are compatible with Globus ExcelsiusGPS® Instrument Trackers and intended to be used with the iFuse TORQ Implant System and the Globus ExcelsiusGPS® Robotic Navigation System (including the Globus Excelsius3D® Imaging System), which is intended for use an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. Use of the iGPS instruments is limited to use only with 10.0 mm and 11.5 mm fully threaded iFuse TORQ implants. The iFuse Bedrock Granite Navigation instruments are intended to be used with the iFuse Bedrock Granite Implant System and the Medtronic StealthStation System to assist the surgeon in precisely locating anatomical structures in iFuse Bedrock Granite Implant System procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data-based model of the anatomy. The Granite iGPS instruments and iGPS Drill Bits are compatible with Globus ExcelsiusGPS® Instrument Trackers and intended to be used with the iFuse Bedrock Granite Implant System and the Globus ExcelsiusGPS® Robotic Navigation System (including the Globus Excelsius3D® Imaging System), which is intended for use an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. Use of the iGPS instruments is limited to use only with the 9.5 mm and 10.5 mm iFuse Bedrock Granite implants.
Device Story
iGPS Navigation Instruments are reusable surgical tools designed for integration with the Globus ExcelsiusGPS Robotic Navigation System; they facilitate precise placement of iFuse TORQ and iFuse Bedrock Granite implants. The instruments incorporate trackers compatible with the robotic system to enable real-time spatial positioning and orientation of guide tubes or instruments relative to patient anatomy. Surgeons use these tools in open or percutaneous procedures; the system relies on pre-acquired CT, MR, or fluoroscopic images to map rigid anatomical structures. By providing visual and spatial guidance, the device assists surgeons in navigating complex pelvic and spinal anatomy, potentially increasing accuracy during implant placement and reducing procedural risks. The system is intended for use by trained surgeons in clinical settings.
Clinical Evidence
Bench testing only. Evidence includes dimensional comparison with predicate instruments and simulated use testing to verify compatibility and performance with the Globus ExcelsiusGPS system.
Technological Characteristics
Reusable navigated surgical instrumentation; compatible with Globus ExcelsiusGPS Robotic Navigation System and Instrument Trackers; designed for use with specific iFuse TORQ and iFuse Bedrock Granite implants.
Indications for Use
Indicated for use in skeletally mature and immature patients undergoing sacroiliac joint fusion, sacropelvic fixation, or pelvic/spinal fracture fixation, where stereotactic navigation is appropriate for precise anatomical localization and instrument guidance.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 15, 2025
SI-BONE Inc.
Jayasri Prabakaran
Associate Director, Regulatory Affairs
471 El Camino Real, Suite 101
Santa Clara, California 95070
Re: K251780
Trade/Device Name: iGPS Navigation Instruments
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: August 15, 2025
Received: August 15, 2025
Dear Jayasri Prabakaran:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251780 - Jayasri Prabakaran
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}
K251780 - Jayasri Prabakaran
Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
iGPS Navigation Instruments
Page 8 of 32
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251780 | ? |
| Please provide the device trade name(s). | | ? |
| iFuse TORQ® Implant System | | |
| Please provide your Indications for Use below. | | ? |
| The iFuse TORQ Implant System is indicated for sacroiliac joint fusion for:
• Sacroiliac joint dysfunction including sacroiliac joint disruption and degenerative sacroiliitis.
• Augmenting immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion.
The iFuse TORQ Implant System is also indicated for fracture fixation of the pelvis, including acute, non-acute and non-traumatic fractures.
The iFuse TORQ Navigation instruments are intended to be used with the iFuse TORQ Implant System and the Medtronic StealthStation System to assist the surgeon in precisely locating anatomical structures in iFuse TORQ Implant System procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy.
The TORQ iGPS instruments and iGPS Drill Bits are compatible with Globus ExcelsiusGPS® Instrument Trackers and intended to be used with the iFuse TORQ Implant System and the Globus ExcelsiusGPS® Robotic Navigation System (including the Globus Excelsius3D® Imaging System), which is intended for use an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.
Use of the iGPS instruments is limited to use only with 10.0 mm and 11.5 mm fully threaded iFuse TORQ implants. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
iGPS Navigation Instruments
Page 8 of 33
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251780 | ? |
| Please provide the device trade name(s). | | ? |
| iFuse Bedrock Granite® Implant System | | |
| Please provide your Indications for Use below. | | ? |
| The iFuse Bedrock Granite Implant System is intended for sacroiliac joint fusion in skeletally mature patients for the following conditions:
• Sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.
• To augment immobilization and stabilization of the sacroiliac joint in patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion.
• Acute, non-acute, and non-traumatic fractures involving the sacroiliac joint | | |
| When connected to compatible pedicle screw systems with 5.5- or 6.0-mm posterior rods made from either titanium alloy or cobalt chrome alloys, the iFuse Bedrock Granite Implant System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to thoracolumbosacral fusion for the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:
• Degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc
confirmed by patient history and radiographic studies
• Spondylolisthesis
• Trauma (i.e., fracture or dislocation)
• Spinal stenosis
• Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
• Spinal tumor
• Pseudarthrosis
• Failed previous fusion | | |
| When connected to compatible pedicle screw systems with 5.5- or 6.0-mm posterior rods made from either titanium alloy or cobalt chrome alloys, the iFuse Bedrock Granite Implant System is intended to provide immobilization and stabilization of spinal segments in skeletally immature patients as an adjunct to thoracolumbar fusion for the treatment of progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis, as well as the following conditions: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach. | | |
| Please refer to the additional information section in the Instructions for Use on compatible pedicle screw system rods. | | |
| The iFuse Bedrock Granite Navigation instruments are intended to be used with the iFuse Bedrock Granite Implant System and the Medtronic StealthStation System to assist the surgeon in precisely locating anatomical structures in iFuse Bedrock Granite Implant System procedures, in which the use of stereotactic | | |
{5}
surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data-based model of the anatomy.
The Granite iGPS instruments and iGPS Drill Bits are compatible with Globus ExcelsiusGPS® Instrument Trackers and intended to be used with the iFuse Bedrock Granite Implant System and the Globus ExcelsiusGPS® Robotic Navigation System (including the Globus Excelsius3D® Imaging System), which is intended for use an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.
Use of the iGPS instruments is limited to use only with the 9.5 mm and 10.5 mm iFuse Bedrock Granite implants.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
iGPS Navigation Instruments
Page 9 of 33
{6}
SI-BONE, Inc.
Traditional 510(k) Submission
iGPS instruments
# 510(k) SUMMARY
# iGPS Navigation Instruments
I. 510(k) SUBMITTER
SI-BONE, Inc.
471 El Camino Real, Suite 101,
Santa Clara, CA 95050
Phone: 408-207-0700
Fax: 408-557-8312
Contact Person: Jayasri Prabakaran, Associate Director, Regulatory Affairs
FDA Establishment Registration No.: 3007700286
Date Prepared: September 12, 2025
II. DEVICE
Trade Name of Device: iGPS Navigation Instruments
Classification Name: Orthopedic Stereotaxic Instrument
Classification: II
Regulation Number: 21 CFR 882.4560; Stereotaxic instruments
Product Code: OLO
III. PREDICATE DEVICES
Primary Predicate:
| Predicate Device | Manufacturer | 510(k)# | Clearance Date |
| --- | --- | --- | --- |
| ExcelsiusGPS instruments | Globus Medical Inc. | K171651 | 16 August 2017 |
Additional Predicate:
| Predicate Device | Manufacturer | 510(k)# | Clearance Date |
| --- | --- | --- | --- |
| E-GPS Navigated instruments | Precision Spine, Inc. | K220862 | 14 July 2022 |
K251780
{7}
SI-BONE, Inc.
Traditional 510(k) Submission
iGPS instruments
Reference devices:
| Predicate Device | Manufacturer | 510(k)# | Clearance Date |
| --- | --- | --- | --- |
| iFuse TORQ instruments | SI-BONE, Inc. | K203247 | 25 February 2021 |
| iFuse Bedrock Granite instruments | SI-BONE, Inc. | K220195 | 26 May 2022 |
## IV. DEVICE DESCRIPTION
SI-BONE’s iGPS Navigation Instruments are navigated and reusable instrumentation compatible with the Globus ExcelsiusGPS Robotic Navigation System. The iGPS Navigation Instruments are designed to allow surgeons to place iFuse TORQ and iFuse Bedrock Granite implants using the Globus ExcelsiusGPS® Robotic Navigation System (including the Globus Excelsius3D® Imaging System) along with Globus ExcelsiusGPS® Instrument Trackers.
## V. INDICATIONS FOR USE (iFuse TORQ Implant System)
The iFuse TORQ Implant System is indicated for sacroiliac joint fusion for:
- Sacroiliac joint dysfunction including sacroiliac joint disruption and degenerative sacroiliitis.
- Augmenting immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion.
The iFuse TORQ Implant System is also indicated for fracture fixation of the pelvis, including acute, non-acute and non-traumatic fractures.
The iFuse TORQ Navigation instruments are intended to be used with the iFuse TORQ Implant System and the Medtronic StealthStation System to assist the surgeon in precisely locating anatomical structures in iFuse TORQ Implant System procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy.
The TORQ iGPS instruments and iGPS Drill Bits are compatible with Globus ExcelsiusGPS® Instrument Trackers and intended to be used with the iFuse TORQ Implant System and the Globus ExcelsiusGPS® Robotic Navigation System (including the Globus Excelsius3D® Imaging System), which is intended for use an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.
K251780
Page 2 of 4
{8}
SI-BONE, Inc.
Traditional 510(k) Submission
iGPS instruments
Use of the iGPS instruments is limited to use only with 10.0 mm and 11.5 mm fully threaded iFuse TORQ implants.
VI. INDICATIONS FOR USE (iFuse Granite Implant System)
The iFuse Bedrock Granite Implant System is intended for sacroiliac joint fusion in skeletally mature patients for the following conditions:
- Sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.
- To augment immobilization and stabilization of the sacroiliac joint in patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion.
- Acute, non-acute, and non-traumatic fractures involving the sacroiliac joint
When connected to compatible pedicle screw systems with 5.5- or 6.0-mm posterior rods made from either titanium alloy or cobalt chrome alloys, the iFuse Bedrock Granite Implant System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to thoracolumbosacral fusion for the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:
- Degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies
- Spondylolisthesis
- Trauma (i.e., fracture or dislocation)
- Spinal stenosis
- Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
- Spinal tumor
- Pseudarthrosis
- Failed previous fusion
When connected to compatible pedicle screw systems with 5.5- or 6.0-mm posterior rods made from either titanium alloy or cobalt chrome alloys, the iFuse Bedrock Granite Implant System is intended to provide immobilization and stabilization of spinal segments in skeletally immature patients as an adjunct to thoracolumbar fusion for the treatment of progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis, as well as the following conditions:
spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Please refer to the additional information section in the Instructions for Use on compatible pedicle screw system rods.
The iFuse Bedrock Granite Navigation instruments are intended to be used with the iFuse Bedrock Granite Implant System and the Medtronic StealthStation System to assist the surgeon in precisely locating anatomical structures in iFuse Bedrock Granite Implant System procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra, can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data-based model of the anatomy.
K251780
Page 3 of 4
{9}
SI-BONE, Inc.
Traditional 510(k) Submission
iGPS instruments
The Granite iGPS instruments and iGPS Drill Bits are compatible with Globus ExcelsiusGPS® Instrument Trackers and intended to be used with the iFuse Bedrock Granite Implant System and the Globus ExcelsiusGPS® Robotic Navigation System (including the Globus Excelsius3D® Imaging System), which is intended for use an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.
Use of the iGPS instruments is limited to use only with 9.5 mm and 10.5 mm iFuse Bedrock Granite implants.
## VII. SUMMARY OF SUBSTANTIAL EQUIVALENCE
The subject devices are substantially equivalent to the primary predicate device in terms of intended use, technological characteristics, materials, and principles of operation. The differences in the indications for use statement from the previously cleared, legally marketed predicate are not critical to the intended use of the devices.
## VIII. PERFORMANCE DATA
SI-BONE performed the following performance tests:
- Dimensional comparison with Globus ExcelsiusGPS robotic navigation instruments
- Bench testing
- Simulated use testing
The test results demonstrate that the device is substantially equivalent to the predicate device.
## IX. CONCLUSION
The iGPS Navigation Instruments are substantially equivalent to its predicate in terms of intended use, technological characteristics, materials, and principles of operation. The difference in the indication for use statement from the previously cleared, legally marketed predicates are not critical to the intended of the device. Test results demonstrate that the device is substantially equivalent to the legally marketed device and do not raise different questions of safety and effectiveness than the predicate device.
K251780
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.