The Serafin is intended for the orthodontic treatment of malocclusion in patients with mixed and permanent dentition.
Device Story
Serafin® is a clear aligner system for orthodontic treatment of malocclusion. Input: 3D scanned image file of patient's dentition. Process: Dental practitioner uses 3D software to create a treatment plan; software generates incremental 3D models of teeth; models are 3D printed; thermoplastic polyurethane (Zendura FLX) is thermoformed over models and trimmed. Output: Series of transparent, removable appliances worn by patient (at least 20 hours/day) to apply continuous gentle force for tooth realignment. Context: Used in dental clinics; prescribed by dental practitioners. Benefit: Corrects malocclusion (overbite, underbite, crossbite, gaps, crowding) over 6-24 months. Output affects clinical decision-making by providing a staged, customized treatment path for tooth movement.
Clinical Evidence
Bench testing only. Physical/chemical testing included appearance, dimension, water sorption/solubility, 3-point flexural strength, and hardness per ISO 20795-2. Biological evaluation included cytotoxicity, sensitization, irritation, systemic toxicity, and chemical characterization per ISO 10993 series (Parts 1, 5, 10, 11, 17, 18, 23) and ISO 7405. All results met preset criteria.
Technological Characteristics
Material: Zendura FLX (thermoplastic urethane/copolyester, 3-layer structure). Principle: Thermoforming on 3D-printed models. Form factor: Transparent removable orthodontic appliances. Connectivity: None. Sterilization: Not stated. Software: 3D modeling/treatment planning software. Standards: ISO 20795-2 (orthodontic base polymers), ISO 10993 (biocompatibility).
Indications for Use
Indicated for orthodontic treatment of malocclusion in patients with mixed and permanent dentition.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 11, 2025
TNS Co., Ltd.
Byung Ju Kang
Regulatory Affairs Manager
19F, Unit 1907, Daeryung Post Tower 6th
298 Beotkkot-ro, Geumcheon-gu
Seoul, 08510
REPUBLIC OF KOREA
Re: K251758
Trade/Device Name: Serafin®
Regulation Number: 21 CFR 872.5470
Regulation Name: Orthodontic Plastic Bracket
Regulatory Class: Class II
Product Code: NXC
Dated: September 8, 2025
Received: September 8, 2025
Dear Byung Ju Kang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K251758 - Byung Ju Kang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K251758 - Byung Ju Kang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251758 | ? |
| Please provide the device trade name(s). | | ? |
| Serafin® | | |
| Please provide your Indications for Use below. | | ? |
| The Serafin is intended for the orthodontic treatment of malocclusion in patients with mixed and permanent dentition. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
510(k) SUMMARY
K251758
Date: September 8, 2025
1. SUBMITTER
TNS Co., Ltd.
19F, Unit 1907, Daeryung Post Tower 6th
298 Beotkkot-ro, Geumcheon-gu, Seoul 08510
TEL: +82-2-785-2804
FAX: +82-2-785-2803
Contact Name: Won-gun Chang
Email: tns0109@tnsmile.kr
2. DEVICE
· Trade Name: Serafin®
· Common Name: Clear Aligner System
· Regulation Number 872.5470
· Class: 2
· Classification Product Code: NXC
· Classification Name: Orthodontic plastic bracket
3. PREDICATE DEVICE
Primary Predicate: Serafin® (K232074)
Additional Predicate: Invisalign System (K220287, Align Technology Inc.)
4. DEVICE DESCRIPTION
Serafin® is composed of series of transparent removable orthodontic appliances manufactured using thermoforming technology based on a 3D model of the patient's teeth, customized according to a Dental Practitioner's specific prescription, indicated for
{6}
treatment of dental malocclusion by applying continuous gentle force to realign teeth.
The time for each stage of aligner usage is set in the personal treatment plan, and this varies for each patient and the Dental Practitioner's choice. Each set of aligners needs to be worn for at least 2 weeks. Patients the total treatment time lasts about 6 to 24 months.
## 5. INDICATIONS FOR USE
Serafin®
- The Serafin is intended for the orthodontic treatment of malocclusion in patients with mixed and permanent dentition.
## 6. NON-CLINICAL TESTING
The following test articles were tested based on the referenced standard. All the test results met the preset test criteria.
1) Physical and chemical test
- Guideline on Review and Approval for appliance, Orthodontic, Resin (Sheet Type)(2018.12.) (MFDS)
- Appearance, Dimension, Water sorption & solubility, 3-point flexural strength, hardness
- Appearance, pH, Potassium permanganate reducing substances residue, Residue on evaporation, Heavy metals, Ultraviolet-visible spectrum
- ISO 20795-2 Dentistry – Base polymers – Part 2: Orthodontic base polymers
2) Biological evaluation
- ISO 10993-1 – Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 7405 – Dentistry - Evaluation of biocompatibility of medical devices used in dentistry
- ISO 10993-17 – Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substances
- ISO 10993-18 – Biological evaluation of medical devices - Part 18: Chemical characterization
{7}
- ISO 10993-5 – Biological evaluation of medical devices - Part 5: Cytotoxicity
- ISO 10993-10 - Biological evaluation of medical devices - Part 10: Skin sensitization, Acute oral mucosa irritation, Intracutaneous reactivity
- ISO 10993-11 - Biological evaluation of medical devices - Part 11: Acute systemic toxicity
- ISO 10993-23 - Biological evaluation of medical devices - Part 23: Tests for irritation
- ADA ANSI 41-2020 Evaluation of Biocompatibility of Medical Devices Used in Dentistry
7. SUBSTANTIAL EQUIVALENCE
| | Subject device | Predicate Device | Discuss/Justify the Differences |
| --- | --- | --- | --- |
| 510(k) Number | New | K220287 | - |
| Trade Name | Serafin® | Invisalign | - |
| Manufacturer | TNS Co., Ltd. | Align Technology Inc. | - |
| Common Name | Clear Aligner System | Aligner, Sequential | Same |
| Device Class | 2 | 2 | Same |
| Product Code | NXC | NXC | Same |
| Device Description | Serafin® is composed of series of transparent removable orthodontic appliances manufactured using thermoforming technology based on a 3D model of the patient's teeth, customized according to a Dental Practitioner's specific prescription, indicated for treatment of dental malocclusion by applying continuous gentle force to | Invisalign is a series of transparent removable orthodontic appliances, indicated for treatment of dental malocclusion. Invisalign consists of upper and lower parts in variable sizes and lengths customized for a patient. It is produced by using 3D printer based on 3D scanned image file sent by a Dental Practitioner. The aligners and attachment templates are made | Equivalent |
{8}
| | realign teeth.
The time for each stage of aligner usage is set in the personal treatment plan, and this varies for each patient and the Dental Practitioner's choice. Each set of aligners needs to be worn for at least 2 weeks. For patients, the total treatment time lasts about 6 to 24 months. | of thermoplastic polymer, with/without PFA system made of photopolymerizable methacrylate based resin. | |
| --- | --- | --- | --- |
| **Indications for use** | **The Serafin is intended for the orthodontic treatment of malocclusion in patients with mixed and permanent dentition.** | **The Invisalign System is intended for the orthodontic treatment of malocclusion in patients with primary, mixed, and permanent dentition.** | **Same intended use; dentition range aligned via additional predicate (K220287).** |
| Clinical condition | Overbite, underbite, crossbite, gap teeth, open bite, crowded teeth, baby and permanent teeth.
Limitations include severe cases where large movement of a tooth is necessary such as a tooth rotated over 20 degrees, large gap between teeth, teeth which require intrusion or extrusion | Overbite, underbite, crossbite, gap teeth, open bite, crowded teeth, baby and permanent teeth.
Limitations include severe cases where large movement of a tooth is necessary such as a tooth rotated over 20 degrees, large gap between teeth, teeth which require intrusion or extrusion | Same |
| **Patient Population** | **Children (mixed dentition); Adolescents/Adults (permanent dentition).** | **Children (primary/mixed); Adolescents/Adults (permanent).** | **Difference addressed by identifying Invisalign (K220287) as an additional predicate; no new questions of safety or effectiveness.** |
| Use Location | Dental intraoral Device | Dental intraoral Device | Same |
| OTC or | Rx only | Rx only | Same |
{9}
| Prescription (Rx) Device | | | |
| --- | --- | --- | --- |
| Design | | | |
| Principle of Operation | Serafin® is series of transparent removable orthodontic appliances, indicated for treatment of dental malocclusion. It is produced by using 3D printer based on 3D scanned image file sent by a Dental Practitioner along with a prescription of a specific patient. Series of clear plastic appliances (Aligners) move the patients’ teeth in small increments from their original state to a final, treated state by applying continuous gentle force to teeth while the patient is wearing it. | The Invisalign System consists of a series of doctor-prescribed, thin, clear plastic removable orthodontic appliances (aligners) and proprietary 3-D software. The aligners gently move the patient’s teeth in small increments from their original state to a more optimal, treated state. | Equivalent |
| Appearance | | | Equivalent |
| 3D Software | 3D Software: Produces a 3D model file of the digital scan. Create a treatment plan, which is reviewed by the dental practitioner to reject or request modifications. Approved treatment plan model is 3D printed. | 3D Software: Produces 3D model file of the PVS impression or the digital scan. Create a treatment plan which is reviewed by dental practitioner to reject or request modifications. Approved treatment plan model is 3D printed. | Equivalent |
{10}
| Materials | Thermoplastic polymer : Zendura FLX : Thermoplastic Urethane, Copolyester (3-layer structure) | Thermoplastic polymer SmartTrack™ (Multilayer structure) | Equivalent |
| --- | --- | --- | --- |
| Physical State | Transparent removable orthodontic appliances consist of upper and lower parts | Transparent removable orthodontic appliances consist of upper and lower parts | Same |
| Material property | Zendura FLX Demonstrates sufficient Ultimate Flexural Strength, Water Sorption, Water Solubility as per ISO 20795-2 : 2013. Intended, marketed to be used as dental aligner | SmartTrack™ is material engineered for use with Invisalign orthodontic treatment (Polyurethane) Intended to be used as dental aligner | Equivalent |
| Manufacturing process | Thermoforming on models, Trimming, Marking | Thermoforming on models, Trimming, Marking | Same |
| Biological | | | |
| Tissue Contact | Surface device, permanent use (>30 days), Mucosal membrane | Surface device, permanent use (>30 days), Mucosal membrane | Same |
| Anatomical site | Intact teeth, gingival mucosa, interior vermilion (inner lip), and tongue | Intact teeth, gingival mucosa, interior vermilion (inner lip), and tongue | Same |
| Contact duration | at least 20 hrs or longer a day, up to 30 days | at least 20 hrs or longer a day, up to 30 days | Same |
| Biocompatibility | Biological risk assessment, Toxicological risk assessment as per ISO 10993-1, ISO 7405 | Biocompatibility per the ISO 10993 series standards | Equivalent |
We identified an additional predicate (Invisalign System, K220287) that covers primary, mixed, and permanent dentition. The subject and predicates share the same intended use and technological characteristics; the absence of MAF in the subject device does not raise new questions of safety or effectiveness.
{11}
# 8. SUBSTANTIAL EQUIVALENCE DISCUSSION
The subject device shares the same indications for use and principle of operation as the predicate devices. They are intended for the same clinical conditions, patient populations, and use locations.
Predicate device includes an optional mandibular advancement feature (MAF), but subject device does not provide a MAF feature. The MAF option is a necessary function for 'Skeletal malocclusion', so it does not correspond to the indications for subject device.
Both devices comprise a series of transparent, removable orthodontic appliances, used for at least 20 hours a day over a period of several weeks. The manufacturing process begins with a 3D image file of a patient's dentition, which serves as an input to each manufacturer's software. Each software identifies the individual teeth requiring treatment. The software then generates a 3D model of the patient's dentition, incrementally modifying it towards alignment. This 3D model is then printed out using a 3D printer. Subsequently, a thermoplastic polyurethane material is heated and moulded onto the 3D printed model, known as a thermoforming process. Any excess material is eliminated by trimming process. The resulting plastic appliances, both for Serafin® and Invisalign, are designed to apply a gentle, continuous force on the malocclusion site to realign the teeth.
Correspondingly, the principles of operation, intended use, manufacturing process, and 3D modeling process for Serafin® and Invisalign are identical. Both devices utilize a similar thermoforming polymer as a material, which demonstrates chemical, biological, and functional equivalence. The materials are thermoplastic polymers with a biocompatible polyurethane copolymer and are intended for use as clear aligner material.
However, since the materials are not exactly the same, further biological assessments and pre-clinical performance tests have been carried. For biocompatibility, Serafin® was tested as per ISO 10993-1 in Biological Risk Assessment. In addition, Toxicological
{12}
Risk Assessment was conducted to monitor and mitigate the biological risk from unintended process residuals and degradation material under the light-curing process in 3D printing and the thermoforming process.
Regarding physical properties and clinical performance, Zendura FLX was tested according to “ISO 20795-2 Dentistry - Base polymers - Part 2: Orthodontic base polymers,” under a 37°C wet environment. Since orthodontic base polymers are categorized as Type 3 (thermoplastic materials), testing was conducted for Flexural Modulus, Flexural Strength, Tear Strength, Water Sorption, and Water Solubility. The Flexural Modulus did not meet standard requirement of 1500 MPa. However, this requirement is relevant to highly rigid or inflexible materials, which does not reflect the current requirements for thermoplastic aligners. Considering that Serafin® and Invisalign are engineered to apply a subtle force for realignment, this standard does not apply, and the value primarily serves as an informative detail about the device's physical properties. The other results conformed to the standard, indicating good toughness, an effective ability to move teeth, low water uptake, and low water solubility.
Consequently, it can be concluded that the subject devices are substantially equivalent to the predicate devices in terms of similarity, risk acceptability, and efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.