The Invisalign System is intended for the orthodontic treatment of malocclusion.
Device Story
Invisalign System consists of removable, customized, clear thermoplastic aligners; proprietary 3D ClinCheck software; and attachment templates. Input: dental scans or PVS impressions. Process: ClinCheck software creates 3D treatment plans; dental practitioners review/approve plans; Align personnel (Treat software) design incremental tooth movements. Output: series of aligners moving teeth gradually; optional mandibular advancement feature (precision wings) for skeletal malocclusion. Used in dental clinics; operated by dental practitioners. Attachments bonded to teeth via composite using disposable templates to assist retention/force. Benefits: non-invasive, aesthetic orthodontic correction.
Clinical Evidence
No clinical or animal testing performed. Bench testing not required as submission involves only labeling modifications.
Technological Characteristics
Thermoformed polyurethane thermoplastic polymer aligners; 3D CAD software for treatment planning; optional precision wings (mandibular advancement); attachment templates. System is prescription-only, intraoral, and non-sterile.
Indications for Use
Indicated for orthodontic treatment of malocclusion in patients with primary, mixed, and permanent dentition. Addresses Class I, II, and III malocclusions, including antero-posterior, vertical (open bite), transverse (narrow arch/jaw, crossbite), and inter-arch (spacing/crowding) corrections at all severity levels.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
Predicate Devices
Invisalign System with Mandibular Advancement Feature (MAF) (K181739)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 7, 2022
Align Technology, Inc. Shweta Daga Director, Regulatory Affairs 2820 Orchard Parkway San Jose, California 95134
Re: K220287
Trade/Device Name: Invisalign System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: March 10, 2022 Received: March 11, 2022
Dear Shweta Daga:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220287
Device Name
Invisalign System
Indications for Use (Describe)
The Invisalign System is intended for the orthodontic treatment of malocclusion.
**Type of Use (Select one or both, as applicable)**
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart G) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
# Align Technology's Invisalign System
# General Information
| 510(k) Sponsor | Align Technology, Inc. |
|-------------------------|-----------------------------------------------------------------------------|
| Address | 2820 Orchard Parkway<br>San Jose, CA 95134 |
| FDA Registration Number | 2953749 |
| Contact Person | Shweta Daga<br>Director, Regulatory Affairs<br>Align Technology, Inc. |
| Contact Information | Email: sdaga@aligntech.com<br>Phone: +1 408-470-1000<br>Fax: +1408-470-1010 |
| Date Prepared | Jan 31, 2022 |
## Name of Device and Name/Address of Sponsor
| Name of Device | Invisalign System |
|-------------------------|----------------------------------------------------------------------|
| Name/Address of Sponsor | Align Technology, Inc.<br>2820 Orchard Parkway<br>San Jose, CA 95134 |
| Trade/Proprietary Name | Invisalign System |
| Common Name | Aligner, Sequential |
| Classification Name | Orthodontic Plastic Bracket |
# Predicate Device
| Name of Device | Invisalign System with Mandibular Advancement Feature (MAF) |
|-------------------------|----------------------------------------------------------------------|
| Name/Address of Sponsor | Align Technology, Inc.<br>2820 Orchard Parkway<br>San Jose, CA 95134 |
| Trade/Proprietary Name | Invisalign System with Mandibular Advancement Feature |
| Common Name | Aligner, Sequential |
| Classification Name | Orthodontic Plastic Bracket |
{4}------------------------------------------------
## Purpose of the Special 510(k) notice.
The purpose of this Special 510(k) notice is to request clearance for proposed changes related to labeling material which include the Doctor's Instructions for Use (IFU), and Patient's IFU, with minor changes and clarifications made based on post-market data. Additionally, this Special 510(k) premarket notification describes all non-significant changes implemented through documentation since the last 510(k) clearance.
#### Intended Use
The Invisalign System is indicated for the orthodontic treatment of malocclusion.
## Device Description
The Invisalign System (subject device) consists of removable orthodontic appliances (aligners), proprietary 3D software and attachment template. The system consists of a series of doctor prescribed, customized, thin, clear plastic aligners that gently move the patient's teeth in small increments from their original state to a more optimal, treated state to address malocclusion. The optional mandibular advancement feature positions patients iaw forward to address skeletal malocclusion. The system is used in patients with primary, mixed, and permanent dentition. The Invisalign System is intended for treating dental and skeletal malocclusion in patients with Class I, II and III including Antoro-Posterior (A-P), vertical (open bite, transverse (narrow arch or jaw, crossbite), or inter-arch (spacing and crowding) correction at all severity levels. In some cases, surgical intervention, or addition of other appliances in addition treatment may be needed at the doctor's discretion.
The Invisalign System aligners consist of a series of customized orthodontic devices made from proprietary thermoformed polyurethane sheeting material. The first device in the series matches the patient's dentition in its current state and then each subsequent aligner stage has the shape of the dentition shifted gradually toward the final desired position. The aligners can accommodate use of tooth attachments and elastics (through use of precision cut outs or buttons or power arms). The system can also be ordered with additional aligner features such as precision cut outs, bite ramps, compliance indicators, and mandibular advancement features (MAF). MAF is also referred to as precision wings and is prescribed by the dental practitioner to position the lower jaw forward during correction of skeletal malocclusion. The subject device system of this Special 510(k), referred to as the Invisalign System, is a modification to the predicate device the Invisalign System with Mandibular Advancement Feature (MAF) (K181739, cleared on Oct 26, 2018) and is inclusive of both standard aligners and aligners with mandibular advancement feature.
The proprietary ClinCheck (CC) 3-D software is inclusive of both dental practitioner-facing functions and internal Align personnel-facing functions. The Align personnel-facing functions are termed 'Treat' internally. This internal facing software enables Align's computer-aided design (CAD) designers to create a treatment plan based on a doctor's prescription and inputs (e.g., dental scans). The treatment plans created using the Align personnel-facing software are then reviewed and approved by doctors before the aligners are manufactured. The doctors' reviews and approvals are exchanged via ClinCheck software (dental practitioner-facing software). Once the treatment plan is approved by the dental practitioner, the information is sent to the manufacturing facility for creation of the aligners.
When attachments are prescribed as part of the treatment plan, a disposable template is provided to assist the dental practitioner in positioning and forming the attachments from dental composite (sold separately). Attachment templates are disposable polyurethane appliances which match the patient's existing dentition and include wells for the placement of dental composite. During the first
{5}------------------------------------------------
visit, the dental practitioner uses the template in bonding dental composite to the tooth surface to create attachments on the teeth. These attachments help create forces on the tooth which can assist in aligner retention or optimized aligner force system for tooth movement.
All these components and features were available with the predicate device.
#### Comparison with Predicate Device
Proposed modified device and predicate device have the same indications for use, intended use, technological characteristics, and principle of operation. The material of the Modified Device is identical to the Predicate Device.
The differences between the modified device and the predicate device are minor labeling updates which do not raise any questions regarding safety and effectiveness. In accordance with 21 CFR 807.92(a)(6) a summary of the technological characteristics' comparison of the proposed modified device to the predicate Device is provided below.
#### Technological Characteristics comparison with the Predicate Device
The Invisalign System (subject device) has no proposed changes to its technological characteristics. The proposed labeling changes do not alter any technological characteristics and do not impact the current safety and effectiveness profile of the Invisalign System.
The Invisalign System with minor modifications implemented to date through documentation to file has similar technological characteristics as the predicate device - Invisalign System with Mandibular Advancement Feature (cleared in K181739). The material and mode of action are unchanged and no new aligner features have been introduced.
| Comparison<br>Criteria | Subject/ Modified Device | Predicate/ Existing Device<br>(K181739) | Comparison<br>Assessment |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Indications for<br>Use | The Invisalign System is<br>intended for the orthodontic<br>treatment of malocclusion. | The Invisalign System is<br>intended for the orthodontic<br>treatment of malocclusion. | Same as predicate<br>device |
| In Use Duration | Aligners are worn for<br>approximately 1-2 weeks of 20-<br>22 hours of wear per day, after<br>which it is replaced by the next<br>stage aligners. This is repeated<br>for duration as prescribed by<br>the Dental Practitioner. | Aligners are worn for<br>approximately 1-2 weeks of 20-<br>22 hours of wear per day, after<br>which it is replaced by the next<br>stage aligners. This is repeated<br>for duration as prescribed by<br>the Dental Practitioner. | Same as predicate<br>device |
| Patient<br>Population | Children, Adolescents and<br>Adults | Pediatrics and Adults | Same as predicate<br>device |
| Use Location | Dental intraoral Devices | Dental intraoral Devices | Same as predicate<br>device |
| OTC or<br>Prescription (Rx)<br>Device | Rx only | Rx only | Same as predicate<br>device |
{6}------------------------------------------------
| Comparison<br>Criteria | Subject/ Modified Device | Predicate/ Existing Device<br>(K181739) | Comparison<br>Assessment |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Design | |
| Operating<br>Principle | Aligners: Sequential aligners<br>apply continuous gentle force to<br>the teeth and/or position<br>mandible forward.<br>3D Software: Produces 3D-<br>model file of the PVS<br>impression or the digital scan.<br>Identifies the individual teeth<br>that require treatment (i.e.,<br>repositioning). Creates a<br>treatment plan (i.e., 3-D models<br>that represent the treatment<br>plan) which is reviewed by the<br>treating dental practitioner using<br>ClinCheck's doctor facing<br>function to reject or request<br>modifications to the set-up prior<br>to approval. | Aligners: Sequential aligners<br>apply continuous gentle force to<br>the teeth and position mandible<br>forward<br>3D Software: Produces 3D-<br>model file of the PVS<br>impression or the digital scan.<br>Identifies the individual teeth<br>that require treatment (i.e.,<br>repositioning). Creates a<br>treatment plan (i.e., 3-D models<br>that represent the treatment<br>plan) which is reviewed by the<br>treating dental practitioner using<br>ClinCheck's doctor facing<br>function to reject or request<br>modifications to the set-up prior<br>to approval. | Same as predicate<br>device<br>Note, mandible<br>advancement feature<br>such as precision<br>wings is an optional<br>feature offered<br>through the predicate<br>device. |
| Materials | Thermoplastic polymer | Thermoplastic polymer | Same as predicate<br>device |
| Lower Jaw<br>Adjustment<br>Mechanism | Invisalign System aligners with<br>mandibular advancement<br>feature (MAF) such as<br>enhanced precision wings with<br>a curved design.<br>For Invisalign system with<br>standard aligners this is not<br>applicable. | Invisalign System Mandibular<br>Advancement Feature wings<br>such as enhanced precision<br>wings with a curved design | Same as predicate<br>device<br>Note, mandible<br>advancement feature<br>such as precision<br>wings is an optional<br>feature offered<br>through predicate<br>device. |
# Performance Data
There are no proposed design changes in the scope of this pre-market notification. The proposed labelling changes do not trigger any performance testing as the changes are minor clarifications and additions that do not impact the current safety and effectiveness profile of the Invisalign System. No animal or clinical testing was required to validate these labeling modifications.
#### Substantial Equivalence
The Invisalign System (Subject Device) and the previously cleared predicate Invisalign System with Mandibular Advancement are similar in that they have:
- same intended use,
- . same indications for use,
- . same principles of operation, and
- same technological characteristics.
{7}------------------------------------------------
The primary change introduced to the Invisalign System through the Special 510(k) submission is proposed labeling updates to the doctor and patient instruction for use. There are no proposed design changes as part of this submission.
# Conclusion
Align Technology's Invisalign System has the same intended use and indications for use as the previously cleared Invisalign System with Mandibular Advancement Feature (MAF) (K181739). In addition, the Invisalign System has the same technological characteristics, and principles of operation as its predicate. The minor labeling differences between the Invisalign System and its predicate device do not raise new questions of safety or efficacy. Thus, the Invisalign System is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.