Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
K251716 · Basic Medical Technology, Inc. · LZA · Aug 27, 2025 · General Hospital
Device Facts
Record ID
K251716
Device Name
Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
Applicant
Basic Medical Technology, Inc.
Product Code
LZA · General Hospital
Decision Date
Aug 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The glove is a disposable device intended for medical purposes that is worn on the examiner’s hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
Device Story
Disposable, non-sterile, powder-free nitrile examination glove; pink color; fingertip textured; ambidextrous; beaded cuff. Used by healthcare professionals in clinical settings to prevent cross-contamination. Provides barrier protection against specific chemotherapy agents, Fentanyl Citrate, simulated gastric acid, and Xylazine. Performance verified via ASTM D6978-05 permeation testing. Healthcare providers use the glove as a standard personal protective equipment (PPE) barrier; permeation data informs selection based on specific drug exposure risks. Benefits include reduced risk of dermal exposure to hazardous substances during patient examination or drug handling.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM D6319-19 (dimensions, physical properties, freedom from holes, powder residue), ISO 10993-23 (irritation), ISO 10993-10 (sensitization), ISO 10993-11 (acute systemic toxicity), and ASTM D6978-05 (chemotherapy drug/Fentanyl/Xylazine permeation).
Technological Characteristics
Synthetic nitrile latex; powder-free; pink; fingertip textured; beaded cuff; ambidextrous. Dimensions and physical properties comply with ASTM D6319-19 and ASTM D412-16. Biocompatibility per ISO 10993 series. Permeation resistance tested per ASTM D6978-05.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Suitable for use with specific chemotherapy drugs, Fentanyl Citrate, simulated gastric acid, and Xylazine in Fentanyl Citrate. Contraindicated for use with Carmustine and Thiotepa due to low permeation times.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs, Simulated Gastric Acid, and Fentanyl Citrate) (K212408)
Reference Devices
Nitrile Powder-Free Exam Gloves with Hyaluronic Acid, and tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate (K250630)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 27, 2025
Basic Medical Technology Inc.
John Zhao
General Manager
5300 Concours Street
Ontario, California 91764
Re: K251716
Trade/Device Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
Regulation Number: 21 CFR 880.6250
Regulation Name: Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved
Product Code: LZC, LZA, OPJ, QDO
Dated: June 4, 2025
Received: June 4, 2025
Dear John Zhao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251716 - John Zhao
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251716 - John Zhao
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ALLAN GUAN -S
For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection
Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K251716
Device Name
Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
Indications for Use (Describe)
The glove is a disposable device intended for medical purposes that is worn on the examiner’s hand to prevent contamination between patient and examiner.
Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
| Chemotherapy Drugs | Minimum Breakthrough Detection Time |
| --- | --- |
| Carboplatin, 10 mg/ml | >240 min. |
| Carmustine, 3.3 mg/ml | 15.3min. |
| Cisplatin, 1 mg/ml | >240 min. |
| Cyclophosphamide, 20 mg/ml | >240 min. |
| Cytarabine HCl 100 mg/ml | >240 min. |
| Dacarbazine, 10 mg/ml | >240 min. |
| Daunorubicin HCl, 5 mg/ml | >240 min. |
| Doxorubicin HCl, 2 mg/ml | >240 min. |
| Etoposide, 20 mg/ml | >240 min. |
| 5-Fluorouracil, 50 mg/ml | >240 min. |
| Gemcitabine HCl, 38 mg/ml | >240 min. |
| Ifosfamide, 50 mg/ml | >240 min. |
| Irinotecan HCl, 20 mg/ml | >240 min. |
| Mechlorethamine HCl, 1 mg/ml | >240 min. |
| Melphalan HCl, 5 mg/ml | >240 min. |
| Methotrexate, 25 mg/ml | >240 min. |
| Mitomycin-C, 0.5 mg/ml | >240 min. |
| Mitoxantrone HCl., 2 mg/ml | >240 min. |
| Pacilitaxel, 6 mg/ml | >240 min. |
| Thiotepa, 10 mg/ml | 28.2 min. |
Fentanyl, Gastric Acid and Xylazine
Minimum Breakthrough Detection Time
Fentanyl Citrate Injection, 50mcg/ml
>240 min.
Simulated Gastric Acid
>240 min.
Fentanyl Citrate (50mcg/ml): Xylazine HCl(100mg/ml), 50:50
>240 min.
*Please note that the following drugs have extremely low permeation times:
Carmustine: 15.3 minutes, Thiotepa: 28.2 minutes
Warning: Do not use with Carmustine and Thiotepa.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 2 of 2
{5}
Basic Medical Technology Inc.
510(K) Summary
K251716
1. Submission Information
Date Prepared: August 27, 2025
Submission Format: Traditional 510(k)
510(K)#: K251716
2. Submitter Information
Applicant Name: Basic Medical Technology Inc.
Address: 5300 Concours Ontario, CA 91764
Contact Person: John Zhao
Tel: (909) 980-1678
Email: hotmailtojohn@yahoo.com
3. Device Information
Trade Name: Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
Common Name: Nitrile Powder-Free Exam Gloves
Classification Name: Non-Powdered Patient Examination Glove
Regulation: 21 CFR 880.6250
Product Code: LZA, LZC, QDO, OPJ
Classification Panel: General Hospital
Device Class: Class I
4. Predicate Device Information
Primary Predicate Device:
Applicant: Pt. Maja Agung Latexindo
510(k) #: K212408
Trade Name: Powder Free Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs, Simulated Gastric Acid, and Fentanyl Citrate)
Reference Device :
Applicant: Basic Medical Technology Inc.
510(k) #: K250630
1 / 7
{6}
2 / 7
# Basic Medical Technology Inc.
Trade Name: Nitrile Powder-Free Exam Gloves with Hyaluronic Acid, and tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate
## 5. Device Description:
Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate is a Class I patient examination glove and specialty chemotherapy gloves, that made from synthetic nitrile latex. They are pink color, non-sterile, powder-free, fingertip textured, ambidextrous with beaded cuff, and single use only, and come in different sizes- XS, S, M, L, XL and XXL.
The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs, Fentanyl Citrate, Gastric Acid and Xylazine in Fentanyl Citrate.
## 6. Indications for Use:
The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
| NO. | Chemotherapy Drugs | Minimum Breakthrough Detection Time |
| --- | --- | --- |
| 1 | Carboplatin, 10 mg/ml | >240 min. |
| 2 | Carmustine, 3.3 mg/ml | 15.3 min. |
| 3 | Cisplatin, 1 mg/ml | >240 min. |
| 4 | Cyclophosphamide, 20 mg/ml | >240 min. |
| 5 | Cytarabine HCl, 100 mg/ml | >240 min. |
| 6 | Dacarbazine, 10 mg/ml | >240 min. |
| 7 | Daunorubicin HCl, 5 mg/ml | >240 min. |
| 8 | Doxorubicin HCl, 2 mg/ml | >240 min. |
| 9 | Etoposide, 20 mg/ml | >240 min. |
| 10 | 5-Fluorouracil, 50 mg/ml | >240 min. |
| 11 | Gemcitabine HCl, 38 mg/ml | >240 min. |
| 12 | Ifosfamide, 50 mg/ml | >240 min. |
| 13 | Irinotecan HCl, 20 mg/ml | >240 min. |
| 14 | Mechlorethamine HCl, 1 mg/ml | >240 min. |
| 15 | Melphalan HCl, 5 mg/ml | >240 min. |
| 16 | Methotrexate, 25 mg/ml | >240 min. |
| 17 | Mitomycin-C, 0.5 mg/ml | >240 min. |
| 18 | Mitoxantrone HCl, 2 mg/ml | >240 min. |
| 19 | Pacilitaxel, 6 mg/ml | >240 min. |
| 20 | Thiotepa, 10 mg/ml | 28.2 min. |
| NO. Fentanyl, Gastric Acid and Xylazine | | Minimum Breakthrough Detection Time |
{7}
# Basic Medical Technology Inc.
| 1 | Fentanyl Citrate Injection,50mcg/ml | >240 min. |
| --- | --- | --- |
| 2 | Gastric Acid | >240 min. |
| 3 | Fentanyl Citrate (50mcg/ml): Xylazine HCl(100mg/ml), 50:50 | >240 min. |
*Please note that the following drugs have extremely low permeation times:
Carmustine: 15.3 minutes, Thiotepa: 28.2 minutes
Warning: Do not use with Carmustine and Thiotepa.
# 7. Comparison of Subject Device and Predicate Device:
General Comparison Table:
| Device | Proposed Device | Predicate Device | Reference Device | Result |
| --- | --- | --- | --- | --- |
| 510K # | K251716 | K212408 | K250630 | Different |
| Applicant Name | Basic Medical Technology Inc. | Pt. Maja Agung Latexindo | Basic Medical Technology Inc. | Same as K250630 |
| Product Name | Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate | Powder Free Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs, Simulated Gastric Acid, and Fentanyl Citrate) | Nitrile Powder-Free Exam Gloves with Hyaluronic Acid, and tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate | Different |
| Product Code | LZA, LZC, QDO, OPJ | LZA, LZC, QDO, OPJ | LZA, LZC, QDO, OPJ | Same |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Device Classification | Class I | Class I | Class I | Same |
| Indications for use | The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate, Simulated Gastric acid and Xylazine in Fentanyl Citrate in accordance with ASTM D6978-05 Standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs. | Powder Free Nitrile Examination Glove is a disposable device intended for medical purpose that is worn on examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy drugs, Simulated Gastric Acid, and Fentanyl Citrate as per ASTM D6978-05(2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. | Nitrile Powder-Free Exam Gloves with Hyaluronic Acid, and tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl, Gastric acid and Xylazine in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs. | Same |
| Powder free | Yes | Yes | Yes | Same |
{8}
# Basic Medical Technology Inc.
| Material | Nitrile | Nitrile | Nitrile | Same |
| --- | --- | --- | --- | --- |
| Color | Pink | Blue | Blue | Different |
| Size | XS, S, M, L, XL, XXL | XS, S, M, L, XL, XXL | XS, S, M, L, XL, XXL | Same |
| Sterile | Non-Sterile | Non-Sterile | Non-Sterile | Same |
| Single Use | Single use | Single use | Single use | Same |
| Design feature | Ambidextrous | Ambidextrous | Ambidextrous | Same |
| Other | N/A | N/A | Hyaluronic Acid coated on inside | Different |
| Chemo Drugs Claim | See below comparison table | See below comparison table | See below comparison table | Different |
Analysis: * The finished subject device has been tested with performance and Biocompatibility, all the test results meet the requirements, so the difference of color and other does not raise questions of safety and effectiveness.
Technological Characteristic Comparison Table:
| Characteristics | | Proposed device | Predicate device K212408 | Reference Device K250630 | Result |
| --- | --- | --- | --- | --- | --- |
| Dimension Reference standard ASTM D6319-19 | Length | Min 220mm for XS, S Min 230mm for M, L, XL, XXL | Min 220mm for XS, S Min 230mm for M, L, XL, XXL | Min 220mm for XS, S Min 230mm for M, L, XL, XXL | Same |
| | Width | XS: 70±10 mm S: 80±10 mm M: 95±10 mm L: 110±10 mm XL: 120±10 mm XXL: 130±10 mm | XS: 70±10 mm S: 80±10 mm M: 95±10 mm L: 110±10 mm XL: 120±10 mm XXL: 130±10 mm | XS: 70±10 mm S: 80±10 mm M: 95±10 mm L: 110±10 mm XL: 120±10 mm XXL: 130±10 mm | |
| | Thickness | Palm: Min 0.05 mm Finger: Min 0.05 mm | Palm: Min 0.05 mm Finger: Min 0.05 mm | Palm: Min 0.05 mm Finger: Min 0.05 mm | |
| Physical properties ASTM D6319-19 ASTM D412-16 | Before aging | | | | Same |
| | Tensile strength | Min 14MPa | Min 14MPa | Min 14MPa | |
| | Ultimate elongation | Min 500% | Min 500% | Min 500% | |
| | After aging | | | | |
| | Tensile strength | Min 14MPa | Min 14MPa | Min 14MPa | |
| | Ultimate elongation | Min 400% | Min 400% | Min 400% | |
| Freedom from holes ASTM D6319-19 ASTM D5151-19 | | G-I, AQL2.5 | G-I, AQL2.5 | G-I, AQL2.5 | Same |
| Residual Powder ASTM D6319-19 ASTM D6124-06 | | ≤ 2 mg per glove | ≤ 2 mg per glove | ≤ 2 mg per glove | Same |
| Biocompatibility | Irritation | Not a primary skin | Not a primary skin | Not a primary skin | Same |
{9}
Basic Medical Technology Inc.
| | | irritant | irritant | irritant | |
| --- | --- | --- | --- | --- | --- |
| | Sensitization | Not a primary skin sensitizer | Not a primary skin sensitizer | Not a primary skin sensitizer | Same |
| | Acute Systemic Toxicity Test | Not toxic | Not toxic | Not toxic | Same |
Chemotherapy Drugs Comparison Claim:
| Chemotherapy Drugs | Minimum Breakthrough Detection Time | | | Result of comparison |
| --- | --- | --- | --- | --- |
| | Proposed device | Predicate device K212408 | Reference Device K250630 | |
| Carboplatin, 10 mg/ml | >240 min. | >240 min | >240 min | Same |
| Carmustine, 3.3 mg/ml | 15.3 min. | 58.3 min | 11.6 min | Different |
| Cisplatin, 1 mg/ml | >240 min. | >240 min | >240 min | Same |
| Cyclophosphamide, 20 mg/ml | >240 min. | >240 min | >240 min | Same |
| Cytarabine HCl, 100 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Dacarbazine, 10 mg/ml | >240 min. | >240 min | >240 min | Same |
| Daunorubicin HCl, 5 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Docetaxel, 10 mg/ml | / | >240 min | / | Different |
| Doxorubicin HCl, 2 mg/ml | >240 min. | >240 min | >240 min | Same |
| Etoposide, 20 mg/ml | >240 min. | >240 min | >240 min | Same |
| 5-Fluorouracil, 50 mg/ml | >240 min. | >240 min | >240 min | Same |
| Gemcitabine HCl, 38 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Idarubicin HCl, 1 mg/ml | / | / | >240 min | Different |
| Ifosfamide, 50 mg/ml | >240 min. | >240 min | >240 min | Same |
| Irinotecan HCl, 20 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Mechlorethamine HCl, 1 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Melphalan HCl, 5 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Methotrexate, 25 mg/ml | >240 min. | >240 min | >240 min | Same |
| Mitomycin-C, 0.5 mg/ml | >240 min. | >240 min | >240 min | Same |
| Mitoxantrone HCl, 2 mg/ml | >240 min. | / | >240 min | Same as K250630 |
| Pacilitaxel, 6 mg/ml | >240 min. | >240 min | >240 min | Same |
| Thiotepa, 10 mg/ml | 28.2 min. | 119min | 25.3 min | Different |
| Vincristine Sulfate,1 mg/ml | / | >240 min | >240 min | Different |
| Fentanyl, Gastric Acid and Xylazine | Minimum Breakthrough Detection Time | | | Result of comparison |
| | Proposed device | Predicate device K212408 | Reference Device K250630 | |
| Fentanyl Citrate Injection,50mg/ml | >240 min. | >240 min | >240 min. | Same |
| Simulated Gastric Acid | >240 min. | >240 min | >240 min. | Same |
| Fentanyl Citrate (50mcg/ml): Xylazine HCl(100mg/ml), 50:50 | >240 min. | / | >240 min. | Same as K250630 |
* Chemotherapy drugs and the minimum breakthrough time of subject device will be listed on labeling, so this difference does not raise questions of safety and effectiveness.
8. Summary of Non-Clinical Performance Data
{10}
# Basic Medical Technology Inc.
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:
| Test performed | Methodology | Acceptance Criteria | Result |
| --- | --- | --- | --- |
| Freedom From Holes | ASTM D6319-19
ASTM D5151-19 | Meet requirement inspection level G- 1, AQL 2.5 (ISO2859-1) | Pass |
| Dimension-Length | ASTM D6319-19 | Minimum 220mm for size XS-S
Minimum 230mm for size M-XXL | Pass |
| Dimension-Width | ASTM D6319-19 | XS: 70±10mm
S: 80±10mm
M: 95±10mm
L:110±10mm
XL: 120±10mm
XXL: 130±10mm | Pass |
| Dimension-Thickness | ASTM D6319-19 | Finger: 0.05mm (min)
Palm: 0.05mm (min) | Pass |
| Physical properties | ASTM D6319-19
ASTM D412-16 | Tensile Strength (Min14 Mpa)
Elongation (Before Aging 500% and after aging 400%) Min | Pass |
| Powder Residue | ASTM D6319-19
ASTM D6124-06 | Not more than 2 mg per glove | Pass |
| Skin Irritation | ISO10993-23:2021 | Under the conditions of the study, not a primary skin irritant | Pass |
| Skin Sensitization | ISO10993-10:2021 | Under the conditions of the study, not a contact sensitizer | Pass |
| Acute Systemic Toxicity | ISO 10993-11:2017 | Under the conditions of the study, no signs of acute systemic toxicity were observed | Pass |
| Chemotherapy Drug | ASTM D6978-05 | Refer to the above table | Pass |
- ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
- ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
- ASTM D6124-06 (2017), Standard Test Method for Residual Powder on Medical Gloves
- ASTM D412-16 (2021), Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension
- ASTM D6978-05(2023), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
{11}
7 / 7
# Basic Medical Technology Inc.
- ISO 10993-10: 2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization.
- ISO10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation.
- ISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity
## 9. Summary of Clinical Testing
Not applicable.
## 10. Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the subject device, Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid and Xylazine in Fentanyl Citrate, is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.