AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K251590 · Aug 20, 2025
Methinks CTA Stroke
Methinks Software, S.L
Retrospective clinical CT angiogram (CTA) images
A retrospective, multicenter, multinational study was conducted to evaluate the software's performance in identifying Large Vessel Occlusions (LVO) in 336 clinical cases.
Retrospective study; LVO detection; Clinical imaging data
Methinks CTA Stroke is a radiological computer aided triage and notification software, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow. Methinks CTA Stroke uses an artificial intelligence algorithm to analyze images for findings suggestive of a pre-specified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation. Identification of suspected findings is not for diagnostic use beyond notification. Specifically, the device analyzes CT angiogram images of the brain acquired in the acute setting and sends to PACS and/or notifications to a neurovascular specialist that a suspected large vessel occlusion has been identified and recommends review of those images. Images can be previewed through an image viewer. Methinks CTA Stroke is intended to analyze terminal ICA, MCA-M1 and MCA-M2 vessels for LVOs. Images that are previewed are for informational purposes only and not intended for diagnostic use beyond notification. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. Methinks CTA Stroke is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis.
Device Story
Software-only device; analyzes brain CT angiography (CTA) images in DICOM format; uses AI/ML algorithm to detect suspected large vessel occlusions (LVO) in terminal ICA, MCA-M1, and MCA-M2 vessels; operates in parallel to standard-of-care workflow; sends notifications to neurovascular specialists and/or PACS; provides image preview functionality; used in emergency settings or stroke care departments; operated by trained physicians; output assists in triage/prioritization; clinicians must verify findings on diagnostic-quality non-compressed images; benefits include faster identification and communication of potential LVO cases to specialists, potentially reducing time to treatment.
Clinical Evidence
Retrospective, blinded, multicenter, multinational study (N=336; 110 LVO positive, 226 LVO negative). Ground truth established by two board-certified neuroradiologists (majority vote). Sensitivity: 98.2% (95% CI: 93.6% - 99.8%); Specificity: 91.6% (95% CI: 87.2% - 94.9%). Mean time to notification: 3.30 minutes. Subgroup analyses performed for gender, age, vendor machine, LVO type, and slice thickness.
Technological Characteristics
Software-only; AI/ML/Neural Network algorithm; inputs DICOM CTA images; outputs notifications/PACS integration; no hardware components; cybersecurity controls include authentication, authorization, encryption, and secure updates; complies with FDA cybersecurity guidance and Section 524B of the FD&C Act.
Indications for Use
Indicated for adult patients (age >21) with suspected acute stroke. Used by hospital networks and trained clinicians (emergency physicians, neurologists, radiologists, neurovascular interventionists) to identify and communicate suspected large vessel occlusions (LVO) in terminal ICA, MCA-M1, and MCA-M2 vessels via CT angiogram images.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 20, 2025
Methinks Software S.L.
Claudia Carbonell
Senior QA/RA Manager
Pier 01, Plaça de Pau Vila, Office 2D2
Barcelona, 08039
Spain
Re: K251590
Trade/Device Name: Methinks CTA Stroke
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS
Dated: July 25, 2025
Received: July 25, 2025
Dear Claudia Carbonell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251590 - Claudia Carbonell
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251590 - Claudia Carbonell
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K251590
Device Name
Methinks CTA Stroke
Indications for Use (Describe)
Methinks CTA Stroke is a radiological computer aided triage and notification software, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow.
Methinks CTA Stroke uses an artificial intelligence algorithm to analyze images for findings suggestive of a pre-specified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation.
Identification of suspected findings is not for diagnostic use beyond notification.
Specifically, the device analyzes CT angiogram images of the brain acquired in the acute setting and sends to PACS and/or notifications to a neurovascular specialist that a suspected large vessel occlusion has been identified and recommends review of those images. Images can be previewed through an image viewer. Methinks CTA Stroke is intended to analyze terminal ICA, MCA-M1 and MCA-M2 vessels for LVOs.
Images that are previewed are for informational purposes only and not intended for diagnostic use beyond notification.
Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. Methinks CTA Stroke is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
N
K251590
# 510(k) Summary
The content included in the 510(k) Summary of the Methinks CTA Stroke device is based on the 21 CFR 807.92 requirements.
# I. SUBMITTER
Name: Methinks Software, SL
Address: Pier 01, Office 2D2, Plaça de Pau Vila
08039 Barcelona, Spain
Contact Person: Claudia Carbonell
Date Prepared: August 13, 2025
# II. DEVICE
Name of Device: Methinks CTA Stroke
Common or Usual Name: Methinks CTA Stroke
Classification Name: Radiological computer aided triage and notification software
Regulatory Class: Class II
Product Code: QAS
Regulation Number: 21 CFR §892.2080
# III. PREDICATE DEVICE
The Methinks CTA Stroke device is claimed to be substantially equivalent to the following legally marketed predicate device: Viz.AI, Inc, ContaCT, DEN170073.
Name of the device: ContaCT
Classification Name: Radiological computer aided triage and notification software
Regulatory Class: Class II Special Control
Product Code: QAS
Regulation Number: 21 CFR §892.2080
# IV. DEVICE DESCRIPTION
Methinks CTA Stroke is a software-only device which is intended to be used by trained physicians involved in the management of Acute Stroke (AS) patients at emergency settings or other departments across the stroke care pyramid model. They include trained physicians such as emergency physicians, neurologists, general radiologists, neurovascular interventionists, neuroradiologists and any trained stroke professionals.
The target patients (intended patient population) are male and female in the adult population (above 21 years old) with suspected Acute Stroke.
The Methinks CTA Stroke device analyzes Computed Tomography Angiography (CTA) images from the intended patient population to identify suspected Large Vessel Occlusions (LVO). This information is to be used in conjunction with other patient information by a professional to assist with triage/prioritization of medical images.
{5}
2
The input of the software is Computed Tomography Angiography (CTA) in DICOM format from patients suspected of Acute Stroke. The outputs of the software are notifications sent to the trained physicians intended to be used in conjunction with other patient information for professional judgment to assist with triage/prioritization.
## V. INDICATIONS FOR USE
Methinks CTA Stroke is a radiological computer aided triage and notification software, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow.
Methinks CTA Stroke uses an artificial intelligence algorithm to analyze images for findings suggestive of a pre-specified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation.
Identification of suspected findings is not for diagnostic use beyond notification.
Specifically, the device analyzes CT angiogram images of the brain acquired in the acute setting and sends to PACS and/or notifications to a neurovascular specialist that a suspected large vessel occlusion has been identified and recommends review of those images. Images can be previewed through an image viewer. Methinks CTA Stroke is intended to analyze terminal ICA, MCA-M1 and MCA-M2 vessels for LVOs.
Images that are previewed are for informational purposes only and not intended for diagnostic use beyond notification.
Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. Methinks CTA Stroke is limited to analysis of imaging data and should not be used in lieu of full patient evaluation or relied upon to make or confirm diagnosis.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Methinks CTA Stroke does not raise new questions of safety or effectiveness compared to the previously cleared ContaCT (DEN170073). Both devices are radiological computer-aided triage and notification software applications for determining suspicion of Large Vessel Occlusion (LVO). Both devices use CT angiogram images for the identification of suspected LVO cases.
Compared to the predicate device, the subject device identifies an additional type of occlusion MCA-M2 (the predicate identifies ICA and MCA-M1 only) which does not raise a substantial difference in the indications for use nor different questions of safety and effectiveness.
Identification of suspected findings for both devices are not for diagnostic use beyond notification.
{6}
3
The following table summarizes and compares Methinks CTA Stroke and the ContaCT device (predicate device). A table comparing the key features of the subject and the predicate devices is provided below:
Table 1. Substantial Equivalence Table.
| Substantial Equivalence Table | | |
| --- | --- | --- |
| Comparison Feature | Predicate Device: ContaCT | Subject Device: Methinks CTA Stroke |
| Product Code | QAS | QAS |
| Regulation | 21 CFR §892.2080 | 21 CFR §892.2080 |
| Indications for Use | ContaCT is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, 1 independent of standard of care workflow.
ContaCT uses an artificial intelligence algorithm to analyze images for findings suggestive of a pre-specified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation. Identification of suspected findings is not for diagnostic use beyond notification. Specifically, the device analyzes CT angiogram images of the brain acquired in the acute setting, and sends notifications to a neurovascular specialist that a suspected large vessel occlusion has been identified and recommends review of those images. Images can be previewed through a mobile application.
Images that are previewed through the mobile application are compressed and are for informational purposes only and not intended for diagnostic use beyond notification. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. ContaCT is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. | Methinks CTA Stroke is a radiological computer aided triage and notification software, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow.
Methinks CTA Stroke uses an artificial intelligence algorithm to analyze images for findings suggestive of a pre-specified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation.
Identification of suspected findings is not for diagnostic use beyond notification.
Specifically, the device analyzes CT angiogram images of the brain acquired in the acute setting, and sends to PACS and/or notifications to a neurovascular specialist that a suspected large vessel occlusion has been identified and recommends review of those images. Images can be previewed through an image viewer. Methinks CTA Stroke is intended to analyze terminal ICA, MCA-M1 and MCA-M2 vessels for LVOs.
Images that are previewed are for informational purposes only and not intended for diagnostic use beyond notification.
Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. Methinks CTA Stroke is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. |
| User | Trained Clinicians | Trained Physicians |
| Anatomy | Head | Head |
| Input Data | CT Angiography images | CT Angiography images |
| Technology | Artificial intelligence/ Neural Network | AI/ML/Neural Network |
| Segmentation of ROI | The device does not highlight or direct a user's attention to a specific location in the image file. | The device does not highlight or direct a user's attention to a specific location in the image file. |
| Preview Images | Images that are previewed are for informational purposes only and not intended for diagnostic use beyond notification. | Images that are previewed are for informational purposes only and not intended for diagnostic use beyond notification. |
| Annotation/Localization | The device does not mark, highlight, or direct users' attention to a specific location in the | The device does not mark, highlight, or direct users' attention to a specific location in the |
{7}
4
| | original image. | original image. |
| --- | --- | --- |
| Prioritization Notification | Yes | Yes |
| Clinical SoC Workflow | In parallel to | In parallel to |
| Technical Pipeline | Large Vessel Occlusion (LVO) Notification (one output) | Large Vessel Occlusion (LVO) Notification (one output) |
| Removal of Cases from SoC review | No | No |
VII. PERFORMANCE STANDARDS AND DATA
Performance Data
Cybersecurity has been integrated into the design and development of the Methinks CTA Stroke device in accordance with FDA's cybersecurity guidance and Section 524B of the FD&C Act. The device underwent rigorous non-clinical testing under clinically relevant conditions. Design verification and validation per 21 CFR Part 820.30 confirmed that all design requirements and specifications were met.
Methinks employed a risk-based cybersecurity strategy that includes secure design principles, threat modeling, vulnerability assessment, third-party penetration testing and a Software Bill of Materials (SBOM). Key security controls such as authentication, authorization, secure update mechanisms, and system logging have been implemented. Cybersecurity was addressed as part of the software verification process. Controls including authentication, authorization, cryptography, code, data and execution integrity, confidentiality, monitoring, resiliency and recovery, software updates, patch management, and coordinated vulnerability disclosure are in place to ensure ongoing safety, effectiveness and resilience against cybersecurity threats throughout the device lifecycle.
Methinks conducted a retrospective, blinded, multicenter, multinational study with the Methinks CTA Stroke device with the primary endpoint to evaluate the software's performance in identifying CTA images containing Large Vessel Occlusion (LVO) findings in 336 cases for LVO (LVO Positive: 110, LVO Negative: 226). Institutions included in the validation study were different from institutions included in training, to ensure separation and better representativity of the validation results. This was ensured by checking the countries, states and ZIP codes, when necessary, of the institutions involved.
Ground truthing was established by two US board certified neuroradiologists that read the cases and a third ground truther in case the two first readers were in disagreement regarding LVO findings. The final ground truth was established based on the majority vote.
Sensitivity and specificity exceed the pre-specified performance goals for LVO. Specifically, LVO was observed at Sensitivity: 98.2% (95% CI: 93.6% - 99.8%) and Specificity: 91.6% (95% CI: 87.2% - 94.9%).
The Methinks CTA Stroke time to notification analysis includes the time to get the DICOM exam, de-identify it (if required), analyze and send a notification back to the PACS and notification clients. Methinks CTA Stroke time to notification has been documented for all cases with results shown in the following table.
{8}
5
Table 2. Time to notification of Methinks CTA Stroke.
| Parameter | Mean | 95% CI Lower | 95% CI Upper |
| --- | --- | --- | --- |
| Time to notification of CTA-LVO (minutes) | 3.30 | 3.23 | 3.36 |
Performance across subgroups
Subgroup analyses were conducted for the LVO indication, to support the generalizability of the device across intended patient population and device input.
Performance metrics by subgroups for Methinks CTA Stroke:
Table 3. Performance metrics CTA-LVO by Gender.
| Gender | Measure | N | Estimate | Lower 95% CI | Upper 95% CI |
| --- | --- | --- | --- | --- | --- |
| Male | Sensitivity | 53 | 96.2% | 87.0% | 99.5% |
| | Specificity | 115 | 92.2% | 85.7% | 96.4% |
| Female | Sensitivity | 57 | 100.0% | 93.7% | 100.0% |
| | Specificity | 111 | 91.0% | 84.1% | 95.6% |
Table 4. Performance metrics CTA-LVO by Age.
| Age | Measure | N | Estimate | Lower 95% CI | Upper 95% CI |
| --- | --- | --- | --- | --- | --- |
| < 50 years | Sensitivity | 8 | 100.0% | 63.1% | 100.0% |
| | Specificity | 40 | 97.5% | 86.8% | 99.9% |
| 50 - 70 years | Sensitivity | 41 | 97.6% | 87.1% | 99.9% |
| | Specificity | 106 | 92.5% | 85.7% | 96.7% |
| > 70 years | Sensitivity | 61 | 98.4% | 91.2% | 100.0% |
| | Specificity | 80 | 87.5% | 78.2% | 93.8% |
Table 5. Performance metrics CTA-LVO by vendor machine.
| Vendor machine | Measure | N | Estimate | Lower 95% CI | Upper 95% CI |
| --- | --- | --- | --- | --- | --- |
| Siemens | Sensitivity | 35 | 97.1% | 85.1% | 99.9% |
| | Specificity | 83 | 92.8% | 84.9% | 97.3% |
| General Electric | Sensitivity | 20 | 100.0% | 83.2% | 100.0% |
{9}
6
| | Specificity | 31 | 83.9% | 66.3% | 94.5% |
| --- | --- | --- | --- | --- | --- |
| Philips | Sensitivity | 34 | 97.1% | 84.7% | 99.9% |
| | Specificity | 82 | 91.5% | 83.2% | 96.5% |
| Toshiba / Canon | Sensitivity | 21 | 100.0% | 83.9% | 100.0% |
| | Specificity | 30 | 96.7% | 82.8% | 99.9% |
Table 6. Performance metrics CTA-LVO by LVO subgroup.
| LVO subgroups | Measure | N | Estimate | Lower 95% CI | Upper 95% CI |
| --- | --- | --- | --- | --- | --- |
| ICA | Sensitivity | 35 | 100.0% | 90.0% | 100.0% |
| | Specificity | 0 | - | - | - |
| MCA-M1 | Sensitivity | 44 | 100.0% | 92.0% | 100.0% |
| | Specificity | 0 | - | - | - |
| ICA + MCA-M1 | Sensitivity | 76 | 100.0% | 95.3% | 100.0% |
| | Specificity | 0 | - | - | - |
| MCA-M2 | Sensitivity | 25 | 92.0% | 74.0% | 99.0% |
| | Specificity | 0 | - | - | - |
Table 7. Performance metrics CTA-LVO by slice thickness.
| Slice thickness | Measure | N | Estimate | Lower 95% CI | Upper 95% CI |
| --- | --- | --- | --- | --- | --- |
| 0.3mm ≤ slice thickness ≤ 1mm | Sensitivity | 91 | 97.8% | 92.3% | 99.7% |
| | Specificity | 198 | 92.9% | 88.4% | 96.1% |
| 1mm < slice thickness ≤ 1.3mm | Sensitivity | 19 | 100.0% | 82.4% | 100.0% |
| | Specificity | 28 | 82.1% | 63.1% | 93.9% |
Table 8. Performance metrics CTA-LVO by non-LVO subgroup.
| Non-LVO subgroups | Measure | N | Estimate | Lower 95% CI | Upper 95% CI |
| --- | --- | --- | --- | --- | --- |
| Tumors | Sensitivity | 0 | - | - | - |
{10}
7
| | Specificity | 2 | 100.0% | 15.8% | 100.0% |
| --- | --- | --- | --- | --- | --- |
| Stenosis > 51% | Sensitivity | 0 | - | - | - |
| | Specificity | 27 | 74.1% | 53.7% | 88.9% |
| Intracranial Hemorrhage (ICH) | Sensitivity | 0 | - | - | - |
| | Specificity | 3 | 100.0% | 29.2% | 100.0% |
| Other Acute Ischemic Events | Sensitivity | 0 | - | - | - |
| | Specificity | 61 | 78.7% | 66.3% | 88.1% |
| Chronic Lesions | Sensitivity | 0 | - | - | - |
| | Specificity | 101 | 86.1% | 77.8% | 92.2% |
| No Findings (none of the above) | Sensitivity | 0 | - | - | - |
| | Specificity | 162 | 96.3% | 92.1% | 98.6% |
In summary, the performance validation data demonstrated that the proposed device provides accurate detection of acute LVO under a range of clinically relevant variables associated with the intended use of the device. The notification functionality of the device also met the acceptance criteria.
## Substantial Equivalence Discussion
Both devices have identical intended use and similar technological characteristics/features and roles within a clinical workflow relative to triage and notification. Both devices are intended to automatically process and analyze CT angiogram images to provide a notification to users in case of a suspected LVO being identified.
The subject device does not introduce any new risks when compared to the predicate and both include similar mitigation strategies for reducing the risk of off label use of preview images which are shared via mobile.
## VIII. CONCLUSIONS
In conclusion, Methinks CTA Stroke is substantially equivalent in technological characteristics, safety, and performance characteristics to the legally marketed predicate device, ContaCT (DEN170073).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.