OsteoSight™ Hip (v1)

K251408 · Naitive Technologies, Ltd. · SAO · Sep 2, 2025 · Radiology

Device Facts

Record IDK251408
Device NameOsteoSight™ Hip (v1)
ApplicantNaitive Technologies, Ltd.
Product CodeSAO · Radiology
Decision DateSep 2, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1171
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K251408 · Sep 2, 2025OsteoSight™ Hip (v1)Naitive Technologies, Ltd.Retrospective clinical imaging datasets; Electronic medical records/clinical site databases; Dual-Energy X-ray Absorptiometry (DXA) scan resultsThe sponsor used a retrospective multi-center cohort of routine clinical radiographs and corresponding DXA scans to evaluate the sensitivity, specificity, and AUC of the OsteoSight AI algorithm for identifying low bone mineral density.Retrospective analysis; Clinical performance validation; DXA T-score; Multi-center cohort

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective multi-center validation study; Retrospective analysis of multi-center clinical data; Follow-up/Duration: Not applicablePatients aged 50 years and older who received standard AP radiographs of the hip or pelvis; Sample Size: 3082; Number of Sites: 6Dual-Energy X-ray Absorptiometry (DXA) T-score < -1.0 SDSensitivity, specificity, and AUC for identifying low bone mineral density

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Low bone mineral densitySupervised Machine Learning—Sensitivity 0.370, Specificity 0.951——Retrospective analysis of multi-center studies: 3082 subjects—
Low bone mineral density (subset)Supervised Machine Learning—AUC 0.837, Sensitivity 0.441, Specificity 0.943——Subset of retrospective multi-center study: 2596 subjects—

Indications for Use

OsteoSight™ is a software application intended for use opportunistically with standard anteroposterior (AP) radiographs of the hip or pelvis performed in patients aged 50 years and older. OsteoSight™ provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health. OsteoSight™ should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of an OsteoSight™ report.

Device Story

OsteoSight is an AI-enabled SaMD that processes standard AP hip/pelvis radiographs to estimate femoral neck BMD. It integrates with clinical imaging systems (PACS) via a host integration platform to receive DICOM images. The device uses a supervised machine learning algorithm to analyze images; if estimated BMD falls below a pre-specified threshold, it generates a report for the interpreting physician. The report serves as a notification to prompt clinical assessment of bone health. It does not provide diagnoses or treatment recommendations. Used in clinical settings by radiologists or physician interpreters, the output is reviewed alongside other clinical information to inform patient management. It benefits patients by identifying potential low BMD opportunistically during routine imaging, facilitating timely bone health evaluation.

Clinical Evidence

Retrospective analysis of 3,082 multi-center patient cases (age ≥50). Ground truth defined as DXA T-score < -1.0 SD at the femoral neck within 12 months. In the full intended use population, sensitivity was 0.370 (95% CI 0.349-0.391) and specificity was 0.951 (95% CI 0.936-0.963). In the subset where a result was generated (N=2596), AUC was 0.837 (95% CI 0.821-0.852), sensitivity 0.441 (95% CI 0.386-0.487), and specificity 0.943 (95% CI 0.922-0.961).

Technological Characteristics

Software-only device; no physical materials. Uses supervised machine learning algorithm. Deployed via container orchestration (Docker) on host integration platforms. DICOM-based input. No electrical, mechanical, or radiation safety requirements. Developed per ISO 14971, BS EN 62366, and IEC 62304.

Indications for Use

Indicated for patients aged 50 years and older receiving standard AP radiographs of the hip or pelvis. Not for ruling out low BMD.

Regulatory Classification

Identification

Radiology software for opportunistic evaluation of low bone mineral density. This device is software which opportunistically assesses radiological images to estimate bone mineral density (BMD) intended to assist in a healthcare professional's decision to evaluate patients for possible low BMD within a bone health screening program. The software employs an algorithm that estimates BMD using eligible radiological image data obtained for other clinical purposes.

Special Controls

In combination with the general controls of the FD&C Act, radiology software for opportunistic evaluation of low bone mineral density is subject to the following special controls:

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD &amp; DRUG ADMINISTRATION September 2, 2025 Naitive Technologies Ltd % Ioan Wigley Head of RA/QA c/o Mishcon de Reya Four Station Square CAMBRIDGE, CB1 2GE ENGLAND, UNITED KINGDOM Re: K251408 Trade/Device Name: OsteoSight™ Hip (v1) Regulation Number: 21 CFR 892.1171 Regulation Name: Radiology Software For Opportunistic Evaluation Of Low Bone Mineral Density Regulatory Class: Class II Product Code: SAO Dated: July 29, 2025 Received: July 30, 2025 Dear Ioan Wigley: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K251408 - Ioan Wigley Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K251408 - Ioan Wigley Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251408 | | | Device Name OsteoSight™ Hip (v1) | | | Indications for Use (Describe) OsteoSight™ is a software application intended for use opportunistically with standard anteroposterior (AP) radiographs of the hip or pelvis performed in patients aged 50 years and older. OsteoSight™ provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health. OsteoSight™ should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of an OsteoSight™ report. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {5} 1 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom # 510(k) Summary: K251408 This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92: ## Submitter Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom Tel: +44 020 3321 7000 Contact: Ioan Wigley Date Prepared: July 29, 2025 ## Device Subject Device: OsteoSight™ Hip (v1) Classification: Class II Product Code: SAO Regulation: 21 CFR 892.1171 ## Predicate Device Manufacturer: 16 Bit Inc Device: Rho Predicate Number: DEN230023 ## Indications for Use OsteoSight™ is a software application intended for use opportunistically with standard anteroposterior (AP) radiographs of the hip or pelvis performed in patients aged 50 years and older. OsteoSight™ provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health. OsteoSight™ should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of an OsteoSight™ report. ## Intended Patient Population OsteoSight is intended to be used on patients aged 50 years and older who receive a standard AP radiograph of the hip or pelvis. ## Limitations - OsteoSight is not intended to replace DXA and does not diagnose osteopenia or osteoporosis. Radiologists and referring clinicians should follow recommended practices for screening and assessment {6} 2 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom - OsteoSight cannot be used to rule out low BMD. In certain situations, OsteoSight may not return a report. This does not mean that the patient being assessed is not at risk of low bone mineral density. Radiologists and referring clinicians should follow recommended practices for screening and assessment. - Absence of an OsteoSight report should not be considered as a negative finding # 1 DEVICE DESCRIPTION ## General Description OsteoSight is an AI-enabled software as a Medical Device (SaMD) that processes standard anteroposterior (AP) radiographs of the hip or pelvis in patients aged 50 years and older. The device operates via integration with a host integration platform that facilitates the secure transfer of DICOM images and associated metadata from clinical imaging systems. OsteoSight applies an AI-enabled algorithm pipeline to estimate bone mineral density (BMD) at the femoral neck. If an estimated BMD is below the pre-specified threshold, OsteoSight generates a report indicating that the patient is at risk of having low BMD. This report is delivered to the interpreting physician via the host integration platform. OsteoSight does not provide a diagnosis or treatment recommendation. If the estimated BMD is above or equal to the pre-specified threshold, no report is generated. The OsteoSight report is intended to inform physicians and support further clinical evaluation. Physicians may consider additional assessments, such as a comprehensive fracture risk assessment, in conjunction with other clinical information. {7} 3 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom Table 1. Comparison of Technological Characteristics &amp; Intended Use to Predicate Device | | Subject Device | Predicate Device | Summary | | --- | --- | --- | --- | | 510(k) number | TBC | DEN230023 | | | Legal Manufacturer | Naitive Technologies Limited | 16-Bit Inc | | | Device Name | OsteoSight™ | Rho | | | Anatomical area of Interest | Pelvis or hip | Lumbar spine Thoracic spine Chest Pelvis Knee Hand/wrist | Both predicate and subject device are indicated for quantification of X-ray images of the pelvis. | | Indications for Use | OsteoSight™ is a software application intended for use opportunistically with standard anteroposterior (AP) radiographs of the hip or pelvis performed in patients aged 50 years and older.. OsteoSight™ provides a notification in | Rho is a software application intended for use opportunistically with standard frontal radiographs of the lumbar spine, thoracic spine, chest, pelvis, knee, or hand/wrist performed in patients aged 50 years and | Equivalent intended use | {8} 4 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom | | Subject Device | Predicate Device | Summary | | --- | --- | --- | --- | | | the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health. OsteoSight™ should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of an OsteoSight™ report. | older. Rho provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at L1-L4 or the femoral neck to prompt a clinical assessment of bone health. Rho should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of Rho report. | | | Intended User | Physician | Physician | Same | | Target Patient Population | Adults ≥ 50 years who undergo an X-ray as part of their standard of care | Adults ≥ 50 years who undergo an X-ray as part of their standard of care | Same | | Device Use environment | When reporting the X-ray, the radiologist or physician interpreter can review the OsteoSight | Rho can be installed on-premise or in the cloud, however, it must be within your | Same | {9} 0 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom | | Subject Device | Predicate Device | Summary | | --- | --- | --- | --- | | | Report, and, if in agreement, can simply include this finding in their own words. OsteoSight can be installed in the cloud, via a host integration platform, which may be within the existing secured network of a clinical site. OsteoSight consists of a number of microservices deployed through container orchestration services like Docker. | existing secured network. All traffic to and from Rho can be monitored and controlled by you. Rho consists of a DICOM node, database and a number of microservices deployed through container orchestration services like Docker Compose and Kubernetes. | | | Design: Purpose | To opportunistically analyze standard anteroposterior (AP) radiographs of the hip or pelvis performed in patients aged 50 years and older. The algorithm presents a binary output to indicate whether the patient likely has low BMD at the femoral neck. An OsteoSight Report is | To opportunistically analyze standard anteroposterior (AP) radiographs of the spine, chest, pelvis, knee, hand or wrist performed in patients aged 50 years and older. The algorithm presents a binary output to indicate whether or not the patient likely has low BMD at either the femoral neck or L1- | Same, but fewer anatomical regions. | 5 {10} 6 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom | | Subject Device | Predicate Device | Summary | | --- | --- | --- | --- | | | generated for positive cases that can be sent back to the PACS for physician interpretation. OsteoSight provides a notification in the form of a report to aid radiologists and/or physician interpreters to identify patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health. | L4. A Rho Report is generated for positive cases that can be sent back to the PACS for physician interpretation or viewed through a browser-based interface. Rho provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health. | | | Machine Learning Methodology | Supervised Machine Learning (ML) | Supervised Machine Learning (ML) | Same | | Image source | DICOM Source (e.g., imaging device, intermediate DICOM node, PACS system, etc.) | DICOM Source (e.g., imaging device, intermediate DICOM node, PACS system, etc.) | Same | | Clinical output | Results Report | Results Report | Same | | Clinical Finding | Patients at risk of low BMD | Patients at risk of low BMD | Same | {11} 7 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom | | Subject Device | Predicate Device | Summary | | --- | --- | --- | --- | | Image viewing | Result output through a report with no annotation of the original image | Result output through a report with no annotation of the original image | Same | | Deployment environment | Results will be presented to a physician in a medical setting, typically via a Picture Archiving and Communication System (PACS). | Results will be presented to a physician in a medical setting, typically via a Picture Archiving and Communication System (PACS). | Same | | Human factors | OsteoSight was developed under design controls in compliance with IEC 62366, ‘Application of usability engineering to medical devices’, and FDA guidance ‘Applying Human Factors and Usability Engineering to Medical Devices Guidance for Industry and Food and Drug Administration Staff’. | Rho was developed under design controls and approved based on its compliance to the FD&C Act including validation of its interfaces and risk controls. | Same | | Standards met | OsteoSight was developed in accordance with applicable design standards; ISO 14971, BS EN 62366 and IEC 62304. | No publicly available information. | / | | Materials | N/A – Software device only. | N/A – Software device only. | / | {12} 8 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom | | Subject Device | Predicate Device | Summary | | --- | --- | --- | --- | | Biocompatibility | N/A – Software device only. | N/A – Software device only. | / | | Compatibility with the environment and other devices | OsteoSight processes DICOM images in a software-only configuration and does not require physical integration with imaging equipment. OsteoSight has been validated for its intended use and operational environment. | Rho processes DICOM images in a software-only configuration and does not require physical integration with imaging equipment. | Same | | Sterility | N/A – Software device only. | N/A – Software device only. | / | | Electrical safety | N/A – Software device only. | N/A – Software device only. | / | | Mechanical safety | N/A – Software device only. | N/A – Software device only. | / | | Chemical safety | N/A – Software device only. | N/A – Software device only. | / | | Thermal safety | N/A – Software device only. | N/A – Software device only. | / | | Radiation safety | N/A – Software device only. | N/A – Software device only. | / | {13} naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom # 2 PERFORMANCE DATA ## 2.1 Sterilization and Shelf Life N/A – Not applicable to the subject device. ## 2.2 Biocompatibility Testing N/A – Not applicable to the subject device. ## 2.3 Electrical Safety and Electromagnetic Compatibility (EMC) N/A – Not applicable to the subject device. ## 2.4 Software Verification and Validation OsteoSight software documentation and software verification and validation testing demonstrate that the device meets all requirements for basic documentation level as outlined in the FDA guidance document “Content of Premarket Submissions for Device Software Functions.” ## 2.5 Mechanical Testing N/A – Not applicable to the subject device. ## 2.6 Animal Studies N/A – No animal studies were conducted as part of this application. ## 2.7 Clinical Studies Prospective clinical studies were not conducted as part of this application to demonstrate safety and effectiveness of the device. Clinical performance assessment for the device is based on a retrospective analysis of multi-center studies to compare sensitivity and specificity of the device with the accepted gold standard, Dual-Energy X-ray Absorptiometry (DXA), with pre-specified performance goals. Ground truth low BMD is defined as a T-score of &lt; -1.0 SD as measured at the femoral neck from a DXA scan acquired within 12 months. The datasets used for validation included previously acquired X-rays that included paired DXA T-scores. {14} # native Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom # 2.7.1 Intended Use Population The Intended Use Population, referring to all study subjects who meet the Indications for Use, was made up of de-identified data from six independent cohorts from the US (including the Midwest, Northeast, South, and West). The patient demographics for the Intended Use Population are described in Table 2 with representation across age decades from 50 and above, sex, BMI, and race. Sensitivity and specificity (with $95\%$ confidence intervals) are reported for the full intended use population. In this analysis, "no result" cases are treated as negative device outputs. AUC cannot be calculated for the full population, as the device does not produce a continuous test statistic (eBMD) in cases with no result. Table 2. Demographic information of the Intended Use Population. Interquartile range (IQR), Body mass index (BMI), Dual-energy X-ray absorptiometry (DXA), bone mineral density (BMD). | Data | Total | | --- | --- | | Sample size | 3082 | | Age (Median [IQR]) | 70.0 [64.0 - 76.0] | | (min-max) | (50 - 98) | | Sex (Female:Male:not reported) | 2787:285:10 | | BMI (Mean +/- SD) | 29.3 +/- 6.6 | | (min-max) | (15.1 - 63.8) | | DXA FN BMD (Mean +/- SD) | 0.716 +/- 0.146 | | (min-max) | (0.134 - 1.535) | | Prevalence of Low BMD (%) | 66.7% | | Time Between Scans (Median [IQR] days)(min-max) | 107.5 [31.0 - 207.0] (0 - 365) | | Race | Asian: 81 (2.6%); Black or African American: 171 (5.5%); Hispanic: 574 (18.6%); Other/not known: 855 (27.7%); White: 1401 (45.5%) | | Manufacturer | Agfa: 2 (0.1%); Canon Inc.: 17 (0.6%); Carestream: 77 (2.5%); Eos Imaging: 2 (0.1%); FUJIFILM: 41 | {15} naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom | | (1.3%); GE Healthcare: 22 (0.7%); Iray: 1 (0.0%); Konica Minolta: 1747 (56.7%); Philips: 18 (0.6%); Samsung: 315 (10.2%); Siemens: 655 (21.3%); Swissray: 122 (4.0%); Thales: 2 (0.1%); Varian: 7 (0.2%); Visaris: 54 (1.8%) | | --- | --- | When calculated across the entire intended use population, including cases where no result was produced, sensitivity was 0.370 (95% CI 0.349-0.391) and specificity was 0.951 (95% CI 0.936-0.963), as shown in Table 3 below. These results should be considered within the context that OsteoSight, by design, removes cases where an ROI cannot be placed on the total hip DXA region. Table 3. Performance Characterization of OsteoSight in Intended Use Population. | Population | N | Specificity (95% CI) | Sensitivity (95% CI) | AUC (95% CI) | | --- | --- | --- | --- | --- | | OsteoSight Intended Use Population | 3082 | 0.951 (0.936-0.963) | 0.370 (0.349-0.391) | n/a | # 2.7.2 Intended Use Sub-set Diagnostic accuracy metrics including sensitivity, specificity, and AUC (with $95\%$ confidence intervals), are reported for a subset of the intended use population in which OsteoSight generated a result. Among patients in whom a result was generated, the area under the ROC curve (AUC) was 0.837 (95% CI 0.821-0.852). Sensitivity for detecting low BMD was 0.441 (95% CI 0.386-0.487), while specificity was 0.943 (95% CI 0.922-0.961), as shown in Table 4). {16} # native Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom Table 4. Diagnostic Accuracy of OsteoSight in Subset of Intended Use Population that produced a result. | Population | N | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) | | --- | --- | --- | --- | --- | | Subset of OsteoSight Intended Use Population that produced a result | 2596 | 0.837 (0.821-0.853) | 0.441 (0.386-0.487) | 0.943 (0.922-0.961) | All pre-specified acceptance criteria for the co-primary and secondary endpoints in the clinical studies were met. # 2.7.3 Sub-Group Analysis - Clinical confounders Subgroup analyses were performed in both cohorts, stratified by clinical factors and image characteristics, to evaluate generalizability of performance. Specificity, sensitivity, and accuracy measures were calculated in predefined clinical subgroups, using the population in the intended use population (Table 5a, Table 6a) and in which OsteoSight generated a result (negative or positive) (Table 5b, Table 6b). Table 5a. Diagnostic Accuracy of OsteoSight in predefined clinical subgroups, in the Intended Use Population. | Group | N | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) | | --- | --- | --- | --- | --- | | Sex | | | | | | Females | 2787 | n/a | 0.373 (0.352-0.395) | 0.950 (0.934-0.963) | | Males | 285 | n/a | 0.319 (0.249-0.399) | 0.957 (0.910-0.980) | | Age group (years) | | | | | | 50-59 | 418 | n/a | 0.258 (0.201-0.324) | 0.969 (0.938-0.985) | | 60-69 | 1012 | n/a | 0.347 (0.310-0.385) | 0.940 (0.912-0.960) | | 70-79 | 1045 | n/a | 0.401 (0.366-0.437) | 0.959 (0.930-0.976) | | 80+ | 485 | n/a | 0.489 (0.441-0.537) | 0.912 (0.830-0.957) | {17} # native Native Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom Table 5b. Diagnostic Accuracy of OsteoSight in predefined clinical subgroups, using the population in which OsteoSight generated a result. | Group | N | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) | | --- | --- | --- | --- | --- | | Sex | | | | | | Females | 2347 | 0.840 (0.823-0.856) | 0.446 (0.390-0.501) | 0.942 (0.921-0.960) | | Males | 239 | 0.810 (0.756-0.865) | 0.374 (0.130-0.537) | 0.948 (0.888-0.991) | | Age group (years) | | | | | | Under 18 | 100 | 0.999 (0.996-1.002) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 18-24 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 25-34 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 35-44 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 45-54 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 55-64 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 65-74 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 75-84 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 85-94 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | | 95-104 | 100 | 0.999 (0.996-0.999) | 0.999 (0.997-0.999) | 0.999 (0.998-0.999) | {18} 14 naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom | 50-59 | 352 | 0.838 (0.797-0.879) | 0.314 (0.199-0.462) | 0.964 (0.918-0.995) | | --- | --- | --- | --- | --- | | 60-69 | 893 | 0.809 (0.779-0.838) | 0.391 (0.325-0.467) | 0.932 (0.896-0.961) | | 70-79 | 919 | 0.839 (0.812-0.867) | 0.452 (0.332-0.545) | 0.952 (0.926-0.978) | | 80+ | 428 | 0.834 (0.784-0.885) | 0.552 (0.357-0.721) | 0.899 (0.812-0.957) | | BMI (kg/m²) | | | | | | 18-25 | 333 | 0.846 (0.781-0.910) | 0.543 (0.182-0.750) | 0.906 (0.792-0.981) | | 25-30 | 444 | 0.833 (0.794-0.873) | 0.462 (0.351-0.579) | 0.938 (0.897-0.979) | | >30 | 508 | 0.795 (0.757-0.833) | 0.321 (0.239-0.407) | 0.958 (0.917-0.985) | | Race | | | | | | Asian | 75 | 0.778 (0.629-0.927) | 0.381 (0.095-0.841) | 0.917 (0.750-1.000) | | Black or African American | 133 | 0.830 (0.753-0.908) | 0.344 (0.156-0.562) | 0.960 (0.842-1.000) | | Hispanic | 511 | 0.861 (0.829-0.893) | 0.466 (0.381-0.556) | 0.958 (0.911-0.984) | | White | 1141 | 0.826 (0.798-0.853) | 0.442 (0.281-0.519) | 0.938 (0.906-0.968) | | Other/not known | 736 | 0.828 (0.798-0.859) | 0.437 (0.357-0.555) | 0.931 (0.892-0.965) | | Region | | | | | | Midwest | 308 | 0.849 (0.795-0.902) | 0.396 (0.013-0.600) | 0.971 (0.912-1.000) | | Northeast | 288 | 0.823 (0.761-0.885) | 0.316 (0.250-0.667) | 0.917 (0.850-1.000) | | South | 1846 | 0.845 (0.828-0.863) | 0.476 (0.408-0.527) | 0.944 (0.919-0.964) | | West | 154 | 0.818 (0.721-0.915) | 0.438 (0.281-0.844) | 0.885 (0.769-1.000) | {19} # native Native Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom Table 6a. Diagnostic Accuracy of OsteoSight across X-ray Machine Manufacturer, in the Intended Use Population. | Group | N | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) | | --- | --- | --- | --- | --- | | X-ray Hardware Manufacturer | | | | | | Canon Inc. | 17 | n/a | 0.429 (0.214-0.674) | 1.000 (0.439-1.000) | | Carestream | 77 | n/a | 0.211 (0.125-0.333) | 1.000 (0.839-1.000) | | FUJIFILM | 41 | n/a | 0.444 (0.276-0.627) | 1.000 (0.785-1.000) | | GE Healthcare | 22 | n/a | 0.150 (0.052-0.360) | 1.000 (0.342-1.000) | | Konica Minolta | 1747 | n/a | 0.431 (0.402-0.461) | 0.944 (0.923-0.959) | | Philips | 18 | n/a | 0.143 (0.040-0.399) | 0.750 (0.301-0.954) | | Samsung | 315 | n/a | 0.267 (0.213-0.329) | 0.926 (0.854-0.963) | | Siemens | 655 | n/a | 0.291 (0.250-0.334) | 0.986 (0.959-0.995) | | Swissray | 122 | n/a | 0.398 (0.307-0.497) | 0.875 (0.690-0.957) | | Varian | 7 | n/a | 0.000 (0.000-0.490) | 1.000 (0.439-1.000) | | Visaris | 54 | n/a | 0.426 (0.295-0.567) | 1.000 (0.646-1.000) | | X-ray settings | | | | | | Imager Pixel Spacing =< 0.174 | 1190 | n/a | 0.300 (0.271-0.331) | 0.965 (0.938-0.980) | | Imager Pixel Spacing > 0.174 | 1770 | n/a | 0.428 (0.399-0.457) | 0.944 (0.924-0.960) | | X-ray Voltage =< 75 kVp | 690 | n/a | 0.283 (0.245-0.325) | 0.981 (0.951-0.992) | | X-ray Voltage > 75 kVp | 541 | n/a | 0.309 (0.267-0.355) | 0.930 (0.868-0.964) | | X-ray Exposure =< 22 mAs | 618 | n/a | 0.281 (0.243-0.322) | 0.976 (0.931-0.992) | {20} # native Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom Table 6b. Diagnostic Accuracy of OsteoSight across X-ray Machine Manufacturer, using the population in which OsteoSight generated a result | Group | N | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) | | --- | --- | --- | --- | --- | | X-ray Hardware Manufacturer | | | | | | Canon Inc. | 12 | 0.852 (0.611-1.000) | 0.667 (0.556-1.000) | 1.000 (0.000-1.000) | | Carestream | 48 | 0.899 (0.796-1.000) | 0.343 (0.257-0.886) | 1.000 (0.846-1.000) | | FUJIFILM | 41 | 0.918 (0.835-1.000) | 0.444 (0.481-0.889) | 1.000 (1.000-1.000) | | GE Healthcare | 17 | 0.767 (0.495-1.000) | 0.200 (0.467-1.000) | 1.000 (1.000-1.000) | | Konica Minolta | 1522 | 0.842 (0.822-0.862) | 0.496 (0.415-0.547) | 0.936 (0.907-0.957) | | Philips | 14 | 0.455 (0.000-0.952) | 0.182 (0.000-1.000) | 0.667 (0.000-1.000) | | Samsung | 244 | 0.805 (0.744-0.865) | 0.343 (0.279-0.640) | 0.903 (0.833-1.000) | | Siemens | 513 | 0.861 (0.828-0.893) | 0.375 (0.070-0.552) | 0.982 (0.953-1.000) | | Swissray | 117 | 0.811 (0.702-0.921) | 0.415 (0.255-0.840) | 0.870 (0.739-1.000) | | Varian | 7 | 0.833 (0.456-1.000) | 0.000 (0.000-1.000) | 1.000 (1.000-1.000) | | Visaris | 54 | 0.894 (0.783-1.000) | 0.426 (0.489-0.957) | 1.000 (1.000-1.000) | | X-ray settings | | | | | | Imager Pixel Spacing =< 0.174 | 954 | 0.844 (0.816-0.872) | 0.372 (0.274-0.551) | 0.955 (0.926-0.984) | | Imager Pixel Spacing > 0.174 | 1545 | 0.842 (0.822-0.862) | 0.492 (0.413-0.544) | 0.937 (0.909-0.958) | | X-ray Voltage =< 75 kVp | 547 | 0.849 (0.816-0.883) | 0.358 (0.300-0.527) | 0.976 (0.939-1.000) | {21} naitive Naitive Technologies Ltd c/o Mishcon de Reya, Four Station Square, Cambridge, CB1 2GE England, United Kingdom | X-ray Voltage > 75 kVp | 445 | 0.823 (0.772-0.874) | 0.374 (0.246-0.691) | 0.913 (0.859-0.967) | | --- | --- | --- | --- | --- | | X-ray Exposure =< 22 mAs | 496 | 0.840 (0.799-0.882) | 0.362 (0.057-0.625) | 0.963 (0.925-0.991) | | X-ray Exposure > 22 mAs | 492 | 0.840 (0.802-0.878) | 0.366 (0.294-0.558) | 0.946 (0.899-0.986) | ## 2.8 Post-Market Performance Monitoring Plan Monitoring the real-world performance of OsteoSight in the intended patient population will be performed to (a) Monitor relevant performance characteristics and detecting changes in performance; (b) Identify sources of performance changes between validation and real-world environment over time; and (c) Assess the results from the performance monitoring on safety and effectiveness. Data will be collected from clinical sites and healthcare providers and device logs. Performance metrics will be compared to baseline performance as reported above. Any trends in change in performance over time will be evaluated including root cause analysis. Internal reports and customer notifications will be sent where required. ## 2.9 Special Controls OsteoSight complies with all applicable special controls for the product code SAO, 820.1171; 'Radiology software for opportunistic evaluation of low BMD'. Compliance has been demonstrated through clinical performance testing, software verification and validation, and the provision of a post-market performance management plan. Device labeling also meets the requirements outlined in the special controls. ## 3 CONCLUSION The testing conducted and summarized above demonstrates that OsteoSight performs as intended and supports the determination of substantial equivalence to the predicate device. The testing conducted demonstrates that OsteoSight is as safe and effective as the predicate device and no new questions of safety or effectiveness are raised by OsteoSight.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...