AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN230023 · Apr 9, 2024
Rho
16 Bit, Inc.
Retrospective clinical imaging datasets (TNI, OAI, OMN); Electronic medical records/clinical imaging archives; DXA scan results (standard of care)
Retrospective clinical datasets were used to evaluate the sensitivity, specificity, and AUC of the Rho algorithm against the gold standard of DXA T-scores in a real-world clinical population.
Retrospective study; Clinical validation; Opportunistic screening; Bone mineral density
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective Clinical Validation Study; Retrospective, multi-center study; Follow-up/Duration: Not applicable
Patients aged 50+ undergoing x-rays of the lumbar spine, thoracic spine, chest, pelvis, knee, or hand/wrist; Sample Size: 4842 total cases (3729 TNI, 522 OMN, 591 OAI); Number of Sites: Multiple (6 True North Imaging centers, 6 US sites for OMN, 4 US sites for OAI)
Dual Energy X-ray Absorptiometry (DXA) T-score < -1.0
Rho is a software application intended for use opportunistically with standard frontal radiographs of the lumbar spine, thoracic spine, chest, pelvis, knee, or hand/wrist performed in patients aged 50 years and older. Rho provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at L1-L4 or the femoral neck to prompt a clinical assessment of bone health. Rho should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of Rho report.
Device Story
Software application; analyzes standard frontal radiographs (lumbar/thoracic spine, chest, pelvis, knee, hand/wrist); identifies potential low bone mineral density (BMD) at L1-L4 or femoral neck; provides notification report to radiologists/physicians; intended to prompt clinical assessment of bone health; not for ruling out low BMD; used in clinical settings; supports clinical decision-making by flagging patients for follow-up; does not replace formal screening programs.
Clinical Evidence
Retrospective, multi-center study across three datasets (TNI, OMN, OAI) totaling 4,842 cases. Ground truth established by DXA (T-score < -1.0). Primary endpoints: sensitivity, specificity, and AUC. Results showed high specificity (range 0.71-1.00) but variable sensitivity (range 0.22-0.83) across subgroups. AUC ranged from 0.73 to 0.98. Study demonstrated device met prespecified AUC and specificity goals, though sensitivity was lower in US-based datasets (OMN/OAI) compared to the Canadian dataset (TNI).
Technological Characteristics
Radiology software for opportunistic evaluation of low bone mineral density. Employs an algorithm to estimate BMD from existing radiological image data. Subject to software verification, validation, and hazard analysis. Requires performance characterization regarding hardware acquisition system dependence and measurement reproducibility.
Indications for Use
Indicated for patients aged 50+ undergoing standard frontal radiographs of lumbar spine, thoracic spine, chest, pelvis, knee, or hand/wrist to aid in identifying possible low BMD at L1-L4 or femoral neck. Not for ruling out low BMD.
Regulatory Classification
Identification
Radiology software for opportunistic evaluation of low bone mineral density. This device is software which opportunistically assesses radiological images to estimate bone mineral density (BMD) intended to assist in a healthcare professional's decision to evaluate patients for possible low BMD within a bone health screening program. The software employs an algorithm that estimates BMD using eligible radiological image data obtained for other clinical purposes.
Special Controls
In combination with the general controls of the FD&C Act, radiology software for opportunistic evaluation of low bone mineral density is subject to the following special controls:
Submission Summary (Full Text)
{1}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR RHO
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Radiology software for opportunistic evaluation of low bone mineral density. This device is software which opportunistically assesses radiological images to estimate bone mineral density (BMD) intended to assist in a healthcare professional's decision to evaluate patients for possible low BMD within a bone health screening program. The software employs an algorithm that estimates BMD using eligible radiological image data obtained for other clinical purposes.
NEW REGULATION NUMBER: 21 CFR 892.1171
CLASSIFICATION: Class II
PRODUCT CODE: SAO
#### BACKGROUND
DEVICE NAME: Rho
SUBMISSION NUMBER: DEN230023
DATE DE NOVO RECEIVED: April 3, 2023
#### SPONSOR INFORMATION:
16 Bit Inc 20 Bay Street, 11th Floor, Toronto. Ontario M5J2N8 Canada
#### INDICATIONS FOR USE
Rho is a software application intended for use opportunistically with standard frontal radiographs of the lumbar spine, thoracic spine, chest, pelvis, knee, or hand/wrist performed in patients aged 50 years and older. Rho provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at L1-L4 or the femoral neck to prompt a clinical assessment of bone health. Rho should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of Rho report.
#### LIMITATIONS
{2}------------------------------------------------
- . The sale, distribution, and use of Rho are restricted to prescription use in accordance with 21 CFR 801.109.
- . Rho cannot be used to rule out low BMD.
- . Absence of Rho report should not be considered as a negative finding.
- This device cannot replace DXA screening program: radiologists and referring clinicians . should follow recommended practices for screening and assessment.
# PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
Rho is a machine learning-based software-as-a-medical device that interfaces with institutional Picture Archiving and Communications Systems (PACS) to identify patients 50 years and older undergoing x-ray with possible low bone mineral density (BMD). Eligible x-rays are frontal projections of the lumbar spine, thoracic spine, chest, pelvis, knee or hand/wrist. Rho uses the xray DICOM and DICOM tags of age and sex as inputs into a locked machine learning algorithm. The locked machine learning algorithm is trained on a patient-based dataset (True North Imaging. TNI13). The algorithm presents a binary output to indicate whether or not the patient likely has low BMD at either the femoral neck or L1-L4. A Rho Report is generated for positive cases that can be sent back to the PACS for physician interpretation or viewed through a browser-based interface.
Radiologists can choose to include this finding in their report to the referring physician. Inclusion may trigger a referring physician to conduct a clinical fracture risk assessment related to bone health.
For cases where Rho algorithm outputs a negative result, no Rho Report will be sent, and neither the radiologist nor the referring physician will receive any device output.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
#### SOFTWARE
Rho software documentation and software verification and validation testing provided demonstrate that the device meets all requirements for basic documentation level as outlined in the FDA guidance document, "Content of Premarket Submissions for Device Software Functions".
### SUMMARY OF CLINICAL INFORMATION
{3}------------------------------------------------
## Performance Testing
Clinical performance of the device was provided in the form of a retrospective, multi-center study to compare sensitivity and specificity of the device with the gold standard. Dual Energy Xray Absorptiometry (DXA) with pre-specified performance goals. The primary objective of the study is to demonstrate the ability of Rho to separate two groups (yes/no low BMD by DXA) with a high specificity. Ground truth low BMD is defined as at least one of the femoral necks or L 1-L4 vertebrae having a T-Score < - 1.0 when assessed by DXA. The validation datasets consisted of previously acquired x-rays that include paired DXA T-scores. The performance of the device was demonstrated on three independent datasets, True North Imaging (TNI), Osteoarthritis Initiative (OAI) and OneMedNet (OMN). The TNI dataset was split by geographically separated imaging clinics into a training set (TNI13) and clinical validation set (TNI6). The TNI13 dataset was used exclusively for algorithm development. The TN16 study population consisted of patients with a lumbar spine, thoracic spine, chest, pelvis, knee, or hand x-ray within 1 year of a DXA. The validation datasets included 3729 cases from six True North Imaging centers (TNI6) from Canada, 522 US cases from OneMedNet, and 591 US cases from Osteoarthritis Initiative, a multicenter, longitudinal, prospective observational study of knee osteoarthritis. A comparison of the clinical validation set to US reference population can be found in Table 1.
The study had prespecified coprimary endpoints of sensitivity and specificity and a secondary endpoint of an area under the receiver operating characteristic curve (AUC). Additional endpoints were specified for the following subgroups: age, gender, race, anatomy, x-ray manufacturer, x-ray tube voltage (kVp), exposure and imager pixel spacing.
| | US Reference<br>NHANES, all<br>races1 | TNI6 | Clinical Validation Set<br>OMN | OAI2 |
|-----------------------------------------------|---------------------------------------------------------|-----------------------------------------------------|------------------------------------------------|-------------------------------------------------------------------------------|
| Age (years) | 50-80+ | 50-101 | 50-90 | 50-83 |
| Study population | General population | Imaging clinic<br>patients * | Imaging clinic<br>patients * | Study participants<br>with symptomatic<br>tibiofemoral knee<br>osteoarthritis |
| BMI (kg/m²) | F: 27.4 (0.2) †<br>M: 27.1 (0.2) | F: 26.5 (5.5)<br>M: 26.8 (4.4) | F: 28.2 (5.4)<br>M: 28.0 (4.9) | F: 30.5 (5.7)<br>M: 29.8 (4.3) |
| L1-L4 BMD<br>(Hologic units,<br>g/cm²) | F: 0.95 [0.65 –<br>1.23] ††<br>M: 1.07 [0.80 –<br>1.40] | F: 0.96 [0.71-1.27]<br>M: 1.14 [0.80-<br>1.52] | F: 1.02 [0.74-1.33]<br>M: 1.18 [0.84-1.66] | N/A |
| Femoral neck<br>BMD (Hologic<br>units, g/cm²) | F: 0.72 [0.45 –<br>1.01]<br>M: 0.80 [0.55 –<br>1.06] | F: 0.66 [0.51 –<br>0.84]<br>M: 0.72 [0.54-<br>0.94] | F: 0.71 [0.52-0.96]<br>M: 0.78 [0.56-<br>0.99] | F: 0.75 [0.50-1.15]<br>M: 0.82 [0.51-1.36] |
Table 1. Comparison of the Clinical Validation Set to US reference population: ranges of bone mineral density, prevalence of low bone mineral density, and body mass index
{4}------------------------------------------------
| Prevalence of<br>low BMD by<br>DXA (T<-1) | F: 67% †††<br>M: 40% | F: 89%<br>M: 66% | F: 69%<br>M: 52% | F:51%<br>M:28% |
|-------------------------------------------|-------------------------------------------|----------------------------------|---------------------------|----------------|
| Collection sites | 3 mobile exam<br>centers,<br>44 sites, US | n=6, Toronto,<br>Ontario, Canada | n=6, US | n=4, US |
| DXA<br>Manufacturers | Hologic (100%) | GE (100%) | GE (97%),<br>Hologic (3%) | GE (100%) |
1 https://www.cdc.gov/nchs/data/series/sr 11/sr11 251.pdf
2 DXA bone mineral density (BMD) was only measured as the femoral neck.
https://pubmed.ncbi.nlm.nih.gov/18786841/
*higher possibility of low BMD, but also likely the type of people who would be getting x-rays that Rho would analyze (e.g., knee pain)
+ estimated weighted mean (se), SD is ~4-5. Beydoun MA and Wang J. Gender-ethnic Disparity in BMI and Waist Circumference Distribution Shifts in US Adults. Obesity 2010
+ +mean [lowest 5th percentile - highest 95th percentile] from
https://www.cdc.gov/nchs/data/series/sr 11/sr11 251.pdf using weighted means for age 50+ for L1-4 (Table 1) and Femoral Neck (Table 17).
*** sum of prevalence of low bone mass and osteoporosis from
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4757905/ Table 1
### Device Performance by dataset:
Ground truth is established by DXA acquired from the same patient, and ground truth low BMD is defined as at least one of femoral neck or L1-L4 having a T-Score < - 1.0 when assessed by DXA. DXA data acquired from the same patient is currently used in the patient bone mineral density screening program, and it is the current standard of care and well-established in clinical practice.
Sensitivity, specificity, and AUC performance were assessed within validation datasets and in pre-defined subsets of demographic subgroups. Results are presented in Table 2.
| | Analysis<br>group | n | Sensitivity | Specificity | AUC |
|---------------------|-----------------------------------|------|------------------|------------------|------------------|
| | All data split by dataset | | | | |
| | TNI | 3729 | 0.67 (0.66-0.69) | 0.90 (0.88-0.92) | 0.88 (0.88-0.90) |
| | OMN | 513 | 0.45 (0.41-0.50) | 0.93 (0.90-0.96) | 0.85 (0.82-0.88) |
| | OAI | 591 | 0.36 (0.31-0.41) | 0.94 (0.92-0.96) | 0.82 (0.79-0.85) |
| | Sexes split by dataset | | | | |
| Females | TNI | 3111 | 0.70 (0.68-0.71) | 0.90 (0.88-0.92) | 0.89 (0.88-0.91) |
| | OMN | 274 | 0.48 (0.42-0.54) | 0.94 (0.89-0.98) | 0.87 (0.83-0.90) |
| | OAI | 293 | 0.44 (0.37-0.50) | 0.90 (0.85-0.94) | 0.81 (0.77-0.85) |
| Males | TNI | 618 | 0.49 (0.45-0.53) | 0.90 (0.86-0.93) | 0.82 (0.79-0.85) |
| | OMN | 239 | 0.41 (0.34-0.49) | 0.92 (0.88-0.96) | 0.83 (0.78-0.87) |
| | Analysis<br>group | n | Sensitivity | Specificity | AUC |
| | OAI | 298 | 0.22 (0.15-0.30) | 0.98 (0.96-0.99) | 0.79 (0.75-0.84) |
| | Age decades split by dataset | | | | |
| | TNI | 844 | 0.64 (0.61-0.68) | 0.95 (0.92-0.97) | 0.89 (0.87-0.90) |
| 50-59 | OMN | 106 | 0.43 (0.33-0.53) | 0.95 (0.88-1.00) | 0.88 (0.82-0.93) |
| | OAI | 177 | 0.26 (0.15-0.38) | 0.98 (0.96-0.99) | 0.77 (0.70-0.84) |
| | TNI | 1339 | 0.65 (0.63-0.67) | 0.94 (0.91-0.96) | 0.90 (0.89-0.92) |
| 60-69 | OMN | 181 | 0.48 (0.40-0.55) | 0.91 (0.86-0.97) | 0.83 (0.78-0.88) |
| | OAI | 202 | 0.34 (0.25-0.43) | 0.94 (0.90-0.97) | 0.82 (0.77-0.87) |
| | TNI | 1031 | 0.69 (0.66-0.71) | 0.86 (0.82-0.90) | 0.87 (0.85-0.89) |
| 70-79 | OMN | 176 | 0.48 (0.39-0.56) | 0.93 (0.88-0.97) | 0.88 (0.84-0.92) |
| | OAI | 181 | 0.38 (0.30-0.45) | 0.88 (0.81-0.94) | 0.78 (0.72-0.83) |
| | TNI | 515 | 0.74 (0.71-0.78) | 0.71 (0.62-0.80) | 0.84 (0.80-0.88) |
| 80+ | OMN | 50 | 0.35 (0.21-0.50) | 0.95 (0.85-1.00) | 0.80 (0.69-0.90) |
| | OAI | 31 | 0.56 (0.35-0.75) | 1.00 (1.00-1.00) | 0.95 (0.88-1.00) |
| | X-ray body parts split by dataset | | | | |
| | TNI | 1331 | 0.83 (0.81-0.85) | 0.80 (0.75-0.84) | 0.89 (0.88-0.91) |
| Chest | OMN | 149 | 0.51 (0.39-0.62) | 0.90 (0.80-1.00) | 0.86 (0.81-0.92) |
| | TNI | 791 | 0.57 (0.54-0.60) | 0.95 (0.93-0.98) | 0.91 (0.89-0.93) |
| Lumbar | OMN | 99 | 0.39 (0.29-0.50) | 0.93 (0.86-1.00) | 0.85 (0.79-0.91) |
| | TNI | 328 | 0.75 (0.71-0.80) | 0.92 (0.84-1.00) | 0.93 (0.89-0.95) |
| Thoracic | OMN | 44 | 0.50 (0.35-0.65) | 1.00 (1.00-1.00) | 0.86 (0.76-0.94) |
| | TNI | 457 | 0.72 (0.69-0.76) | 0.99 (0.97-1.00) | 0.96 (0.94-0.97) |
| Pelvis | OMN | 82 | 0.56 (0.44-0.68) | 0.97 (0.92-1.00) | 0.90 (0.84-0.95) |
| | OAI | 197 | 0.55 (0.45-0.64) | 0.93 (0.89-0.97) | 0.91 (0.88-0.94) |
| | TNI | 250 | 0.56 (0.50-0.61) | 0.91 (0.85-0.97) | 0.85 (0.80-0.89) |
| Hand/wrist | OMN | 54 | 0.45 (0.31-0.61) | 0.81 (0.67-0.95) | 0.73 (0.61-0.85) |
| | OAI | 174 | 0.24 (0.15-0.33) | 0.96 (0.92-0.98) | 0.77 (0.70-0.82) |
| | TNI | 572 | 0.41 (0.37-0.44) | 0.96 (0.93-0.99) | 0.87 (0.84-0.90) |
| Knee | OMN | 85 | 0.29 (0.18-0.39) | 0.97 (0.92-1.00) | 0.86 (0.79-0.92) |
| | OAI | 220 | 0.28 (0.21-0.36) | 0.94 (0.91-0.98) | 0.76 (0.71-0.81) |
| | Races split by dataset* | | | | |
| | TNI | 3292 | 0.67 (0.66-0.69) | 0.90 (0.88-0.92) | 0.89 (0.88-0.90) |
| White | OMN | 142 | 0.46 (0.38-0.55) | 0.98 (0.94-1.00) | 0.90 (0.86-0.94) |
| | OAI | 418 | 0.36 (0.30-0.42) | 0.94 (0.91-0.96) | 0.80 (0.77-0.84) |
| | TNI | 24 | 0.33 (0.10-0.60) | 1.00 (1.00-1.00) | 0.77 (0.36-0.91) |
| Black | OMN | 114 | 0.36 (0.25-0.46) | 0.95 (0.90-0.99) | 0.86 (0.80-0.91) |
| | OAI | 147 | 0.40 (0.27-0.53) | 0.95 (0.92-0.98) | 0.86 (0.80-0.91) |
| | TNI | 134 | 0.74 (0.67-0.81) | 0.96 (0.88-1.00) | 0.93 (0.89-0.96) |
| Asian | OMN | 134 | 0.47 (0.38-0.56) | 0.91 (0.84-0.98) | 0.84 (0.78-0.89) |
| | Analysis<br>group | n | Sensitivity | Specificity | AUC |
| | OAI | 5 | 0.00 (0.00-0.00) | 1.00 (1.00-1.00) | 0.83 (0.50-1.00) |
| | OMN | 123 | 0.49 (0.40-0.59) | 0.88 (0.79-0.95) | 0.80 (0.73-0.87) |
| Hispanic | OAI | 10 | 0.20 (0.00-0.50) | 1.00 (1.00-1.00) | 0.98 (0.92-1.00) |
| | BMI category split by dataset** | | | | |
| | TNI | 1584 | 0.75 (0.73-0.77) | 0.87 (0.83-0.91) | 0.90 (0.88-0.91) |
| 18.5 to <25 | OMN | 147 | 0.58 (0.51-0.66) | 0.81 (0.69-0.92) | 0.82 (0.75-0.89) |
| | OAI | 32 | 0.45 (0.27-0.64) | 0.83 (0.62-1.00) | 0.75 (0.58-0.90) |
| | TNI | 1265 | 0.63 (0.61-0.66) | 0.90 (0.87-0.93) | 0.87 (0.85-0.89) |
| 25 to <30 | OMN | 187 | 0.46 (0.38-0.54) | 0.94 (0.89-0.98) | 0.86 (0.81-0.90) |
| | OAI | 80 | 0.33 (0.21-0.45) | 0.98 (0.93-1.00) | 0.83 (0.75-0.90) |
| | TNI | 792 | 0.53 (0.50-0.57) | 0.92 (0.89-0.95) | 0.86 (0.83-0.88) |
| ≥ 30 | OMN | 156 | 0.29 (0.20-0.37) | 0.96 (0.92-0.99) | 0.83 (0.77-0.88) |
| | OAI | 106 | 0.37 (0.22-0.53) | 0.97 (0.94-1.00) | 0.92 (0.87-0.96) |
| | Manufacturer split by dataset | | | | |
| | TNI | 1094 | 0.65 (0.62-0.67) | 0.91 (0.88-0.94) | 0.90 (0.88-0.91) |
| GE | OMN | 34 | 0.77 (0.56-0.86) | 0.76 (0.61-0.82) | 0.87 (0.76-0.96) |
| | OAI | 145 | 0.60 (0.49-0.71) | 0.88 (0.82-0.93) | 0.86 (0.80-0.90) |
| | OMN | 104 | 0.39 (0.30-0.49) | 1.00 (1.00-1.00) | 0.87 (0.81-0.93) |
| FUJIFILM | OAI | 194 | 0.17 (0.11-0.24) | 0.99 (0.97-1.00) | 0.90 (0.86-0.93) |
| AGFA | OAI | 84 | 0.44 (0.29-0.60) | 0.95 (0.89-0.98) | 0.89 (0.82-0.94) |
| Swissray | OAI | 149 | 0.39 (0.28-0.50) | 0.95 (0.91-0.98) | 0.88 (0.83-0.92) |
| Imaging<br>Dynamics | TNI | 2616 | 0.68 (0.67-0.70) | 0.89 (0.87-0.91) | 0.88 (0.87-0.89) |
| Other | TNI | 19 | 0.65 (0.44-0.83) | 1.00 (1.00-1.00) | 0.88 (0.69-1.00) |
| Samsung | OMN | 125 | 0.47 (0.37-0.57) | 0.91 (0.84-0.97) | 0.85 (0.79-0.90) |
| Konica<br>Minolta | OMN | 80 | 0.38 (0.27-0.50) | 0.97 (0.90-1.00) | 0.90 (0.84-0.95) |
| Siemens | OMN | 22 | 0.50 (0.31-0.68) | 1.00 (1.00-1.00) | 0.93 (0.79-1.00) |
| Canon Inc | OMN | 106 | 0.45 (0.34-0.56) | 0.94 (0.88-1.00) | 0.85 (0.78-0.91) |
| Philips | OMN | 21 | 0.56 (0.36-0.76) | 1.00 (1.00-1.00) | 0.96 (0.88-1.00) |
| other | OMN | 21 | 0.55 (0.30-0.80) | 1.00 (1.00-1.00) | 0.95 (0.89-1.00) |
Table 2. Sensitivity and Specificity by analysis group - TNI, OMN and OAI
{5}------------------------------------------------
{6}------------------------------------------------
*Hispanic results are not shown for TNI (n=6). ** BM<18.5 kg/m² as OMN only had 3 patients in that group. Ground truth Low BMD was defined by DXA (at least one of the femoral neck or L1-4 having a T-score <- 1).
The device demonstrated a high specificity across all three datasets and a comparatively low sensitivity; however, the device demonstrated lower sensitivity in the OMN and OAI datasets (e.g. the datasets sourced solely from US healthcare environments) as compared to the TNI6 dataset (the dataset sourced from Canada healthcare environment). A high specificity implies that a high rate of the positive findings of low BMD from the device are real cases of low patient BMD. The risk of false negatives from this device is low because the patients continue to receive
{7}------------------------------------------------
standard of care and still undergo clinical screening assessment of bone health. The device fully met the prespecified AUC performance goal for all subgroups in all three datasets.
Study Limitation: The sponsor originally proposed co-primary endpoints of sensitivity and specificity (performance goals specified as the lower end of the two-sided 95% confidence interval of specificity >0.775 and lower end of the two-sided 95% confidence interval of sensitivity >0.5 for the TN16 and OMN cohorts. However, the final device performance across all analysis groups did not meet those goals. The device met its performance goals of specificity and AUC in all three datasets. More specifically, device sensitivity is lower than 30% for the following subgroups: knee subgroup in both US datasets (OMN and OAL), hand/wrist subgroup in OAI, male subgroup in OAI dataset. There is uncertainty associated with the lower sensitivity in the US datasets and certain subgroups that will be addressed by indicating on the device report that negative results do not necessarily indicate absence of low BMD. A link to the study results will be provided on the device report provided to the radiologist, and performance in all subgroups will be provided in the device labeling. Although the device's low sensitivity suggests that there is a small probable benefit, the risk of false negatives is no different from current standard of care. Uncertainties associated with the device's generalizability across different datasets and different subgroups will also be addressed by postmarket monitoring described in the special controls.
Image /page/6/Figure/2 description: The image is a ROC curve, which plots sensitivity against 1-specificity. There are three curves plotted on the graph, representing the subgroups OAI, OMN, and TNI. The curves show the trade-off between the true positive rate (sensitivity) and the false positive rate (1-specificity) for each subgroup. The ROC curve is a useful tool for evaluating the performance of a binary classification model.
Figure 1: ROC curve for the three clinical datasets, TNI6, OAI and OMN
{8}------------------------------------------------
Reproducibility of Rho Results between x-rays acquired on different days
Rho demonstrated moderate agreement when analyzing x-rays of the same body part that were acquired on the same or different days.
| | x-rays on same day, same time (bilateral images) | x-rays on same day with ≥ 1 minute between | x-rays on same day with ≥ 2 minutes between | x-rays on different day, within 2 years |
|-------------|--------------------------------------------------|--------------------------------------------|---------------------------------------------|-----------------------------------------|
| subjects | 414 | 87 | 32 | 269 |
| % low BMD | 83 | 83 | 88 | 89 |
| % agreement | 85 | 88.5 | 84.4 | 86.2 |
| Kappa | 0.686 | 0.759 | 0.69 | 0.681 |
| z | 14 | 7.11 | 3.97 | 11.2 |
| p | 0 | 1.1E-12 | 7.1E-05 | 0 |
| Table 3. Agreement of Rho Result between 2 x-rays of the same body part | | | | |
|-------------------------------------------------------------------------|--|--|--|--|
| | | | | |
### Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
# LABELING
The labeling meets the requirements of 21 CFR 801.109 for prescription devices and includes information on device inputs and outputs, instructions for use, intended patient population and intended users of the device, adequate warnings and precautions as well as detailed performance testing summaries. Application of the device as an evaluation tool for referral to bone health assessment is clearly stated, and there are warnings that clinical decisions should not be based solely upon the device's output and that the device cannot be used to rule out low BMD. Postmarket data collection and its purpose is acknowledged in labeling.
# POSTMARKET MONITORING PLAN
The De Novo request included specifics of a post-market performance management plan to ensure regular assessment of the generalizability and device performance of Rho in the intended patient population in real-world use. The plan includes collection and review of data on the clinical use of Rho from various sources including healthcare professionals, clinical sites, and
{9}------------------------------------------------
patient registries, analyze collected data to identify any patterns, trends, or adverse events associated with the use of Rho and assess the impact of the frequency and severity of any reported incidents on the risk profile of the device. Through automatic monitoring of the percentage of Rho positive cases at each site, and across sites, the sponsor will initiate root cause investigations in the case of performance drift. In the case of significant performance drift. including significant performance drift of a predefined subgroup, the sponsor will inform customers through a predefined communication plan; this can include adding or modifying. warnings or contraindications in the labeling, or modifying indications for use.
## RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of radiology software for opportunistic evaluation of low bone mineral density and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Incorrect patient management due to misinterpretation<br>of device output or overreliance on device output for<br>radiological image interpretation | Clinical performance testing<br>Labeling<br>Postmarket monitoring plan |
| False positive findings leading to unnecessary radiation<br>exposure to the patient and clinical work-up | Clinical performance testing<br>Postmarket monitoring plan<br>Labeling |
| False negative findings leading to missing or delayed<br>patient assessment | Labeling |
| Device failure leading to the absence or delay of results,<br>leading to missing, inaccurate, or delayed patient<br>assessment | Software verification, validation,<br>and hazard analysis |
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, radiology software for opportunistic evaluation of low bone mineral density is subject to the following special controls:
- (1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must fulfill the following:
- The dataset used for training and development of the advanced algorithm must be (i) distinct from the dataset used for testing to support generalizability of the algorithm.
- (ii) Results from clinical performance testing must characterize the performance of the device compared to a clinically justified ground truth (reference method or clinical comparator).
- (iii) The test dataset must be representative of the intended patient population(s) to support a supplement to a screening program. Test datasets must have adequate representation of cases with clinically relevant confounders. The performance
{10}------------------------------------------------
estimates and confidence intervals of the device for each individual confounding factor must be characterized in the performance testing.
- Clinical performance must characterize the dependence of software output on the (iv) hardware specifications of the acquisition system.
- (v) Clinical performance must characterize the device's reproducibility from repeated measurements on the same patients.
- (vi) The testing report must include a detailed description of pre-specified performance testing protocols (including the study objectives, primary and secondary endpoints, statistical hypotheses, performance goals, sample size calculation, statistical analyses) and dataset(s) used.
- (2) Software verification, validation, and hazard analysis must be performed.
- (3) Labeling must include:
- A description of the intended patient population, the intended user, clinical (i) environment, and context of use, including information on interpretation of outputs within the intended clinical workflow:
- (ii) A summary of the performance testing for each device output, including test methods, dataset characteristics, testing environment, results (with confidence intervals), and a summary of clinical performance for all demographic subgroups from testing dataset(s);
- (iii) A description of measurement reproducibility:
- A description of situations in which the device may fail and clinical (iv) subpopulations or acquisition system characteristics in which device was not evaluated. if any:
- A statement that the device output should not be used to replace a screening (v) program.
- (4) The device manufacturer must develop and implement a post-market performance management plan that ensures regular assessment of the generalizability and device performance in the intended patient population in real-world use. The plan must include:
- (i) Data collection, analysis methods, and procedures for:
- Monitoring relevant performance characteristics and detecting changes in (A) performance:
- (B) Identifying sources of performance changes between validation and realworld environment over time; and
- (C) Assessing the results from the performance monitoring on safety and effectiveness.
- (ii) Procedures for communicating the device's current performance to users.
### BENEFIT-RISK DETERMINATION
The risks of the device are based on clinical studies described above. A false positive result could lead to a clinical assessment of risk, which in turn could lead to unnecessary medical imaging (including DXA). However, a patient undergoing a DXA scan is exposed to a very
{11}------------------------------------------------
small amount of radiation, equivalent to a few days of background radiation. The high specificity of device performance demonstrated across demographic subgroups, implying that a high rate of the positive findings of low BMD from the device are real cases of low patient BMD. further mitigates the risk. In the case of a false negative result the patient continues to receive current standard of care, in that the initiation of a clinical assessment of risk depends on the health care provider.
The probable benefit of the device is that by opportunistically identifying individuals with possibly low BMD, the device facilitates the identification of patients who could benefit from possible subsequent clinical risk assessment of bone health, and DXA screening for the assessment of fracture risk.
The device operates in an opportunistic fashion, analyzing images obtained for other clinical purposes, and the follow-up to being flagged (whether a true positive or false positive) is a clinical risk assessment (questionnaire) that is already recommended in the intended population but often overlooked in the current standard of care. The implication of a negative finding (whether a true negative or false negative) is that the initiation of the clinical risk assessment will not be triggered by the device. As the device is intended to assist in a healthcare professional's decision to refer a patient for subsequent clinical assessment of bone health a false negative outcome is no different than for an individual who had an x-ray that was not analyzed by Rho.
The performance data provided support that the probable benefits outweigh the probable risks, when used as intended.
### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
### Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
Rho is a software application intended for use opportunistically with standard frontal radiographs of the lumbar spine, thoracic spine, chest, pelvis, knee, or hand/wrist performed in patients aged 50 years and older. Rho provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at L1-L4 or the femoral neck to prompt a clinical assessment of bone health. Rho should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of Rho report.
The input to Rho is a standard frontal radiograph of the lumbar spine, thoracic spine, chest, pelvis, knee, or hand/wrist.
The probable benefits outweigh the probable risks for Rho device. The device provides benefits, and the risks can be mitigated by the use of general controls and the identified special controls.
{12}------------------------------------------------
## CONCLUSION
The De Novo request for Rho is granted and the device is classified as follows:
Product Code: SAO Device Type: Radiology software for opportunistic evaluation of low bone mineral density Regulation Number: 21 CFR 892.1171 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.