K251350 · Becton, Dickinson and Company · FMF · Jul 29, 2025 · General Hospital
Device Facts
Record ID
K251350
Device Name
BD Plastipak™ Syringe
Applicant
Becton, Dickinson and Company
Product Code
FMF · General Hospital
Decision Date
Jul 29, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
Device Story
The BD Plastipak™ Syringe is a three-piece, single-use, sterile, hypodermic syringe; available in 20 mL and 50 mL sizes with eccentric luer slip tip configurations. The device consists of a lubricated polypropylene barrel with a graduated scale, a lubricated polyisoprene rubber stopper, and a polypropylene plunger rod. Operated manually by healthcare professionals in clinical settings, the plunger rod is pulled to aspirate fluids or depressed to inject/expel fluids. The male 6% (Luer) connector attaches to compatible female Luer connectors. The device is provided sterile via Ethylene Oxide (ETO) gas. It functions as a standard piston syringe for fluid delivery or aspiration, facilitating medication administration or fluid sampling.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included breakout force, sustaining force, leakage past stopper, volumetric accuracy, dead space, luer leakage, stress cracking, and resistance to separation. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity, and hemocompatibility. All tests met predetermined acceptance criteria.
Technological Characteristics
Three-piece piston syringe. Materials: Polypropylene barrel (HPR35CMD), polyisoprene rubber stopper, silicone lubricant. 6% (Luer) male connector. Sizes: 20 mL, 50 mL. Sterilization: ETO (SAL 10^-6). Standards: ISO 7886-1:2017 (performance), ISO 80369-7:2021 (luer fittings), ISO 10993 series (biocompatibility), ASTM F756-17 (hemocompatibility).
Indications for Use
Indicated for health care professionals for general purpose fluid aspiration/injection.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 29, 2025
Becton, Dickinson and Company
Anamika Tiwari
Senior Regulatory Affairs Specialist
1 Becton Drive
Franklin Lakes, New Jersey 07417
Re: K251350
Trade/Device Name: BD Plastipak™ Syringe
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF
Dated: April 30, 2025
Received: April 30, 2025
Dear Anamika Tiwari:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251350 - Anamika Tiwari
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251350 - Anamika Tiwari
Page 3
Sincerely,
# Shruti N. Mistry -S
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251350 | |
| Device Name BD Plastipak™ Syringe | |
| Indications for Use (Describe) The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
BD
# K251350
BD Plastipak™ Syringe
510(k) Summary (21 CFR §807.92)
| Submitter Information | Submitter Name:
Submitter Address:
Contact Person:
Email Address:
Phone Number:
Fax Number:
Date of Preparation: | Becton, Dickinson, and Company
1 Becton Drive, Franklin Lakes
NJ 07417
Anamika Tiwari
Sr. Regulatory Affairs Specialist
Anamika.tiwari@bd.com
(201) 847-4760
(201)-847-5307
30th June 2025 |
| --- | --- | --- |
| Subject Device | Trade Name:
Common Name:
Regulation Number:
Regulation Name:
Regulatory Class:
Product Code:
Classification Panel: | BD Plastipak™ Syringe
Piston Syringe
21 C.F.R. § 880.5860
Syringe, Piston
Class II device
FMF
General Hospital |
| Predicate Device | Trade Name:
510(k) Reference:
Common Name:
Regulation Number:
Regulation Name:
Regulatory Class:
Product Code:
Classification Panel: | BD Single Use, Hypodermic Syringe
K980987
Piston Syringe
21 C.F.R. § 880.5860
Syringe, Piston
Class II
FMF
General Hospital |
Page 1 of 6
{5}
Page 2 of 6
| Device Description | The BD Plastipak™ Syringe is a three-piece, single use, sterile, hypodermic syringe with a 6% (Luer) male connector in 20 mL and 50 mL eccentric luer slip tip configurations. The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection. The syringe assembly consists of a lubricated polypropylene barrel with a graduated scale, a lubricated synthetic rubber stopper and a polypropylene plunger rod. The plunger rod is pulled back to aspirate fluids or depressed to inject or expel fluids. The syringe barrel incorporates a male 6% (Luer) connector which is connectable to a compatible female 6% (Luer) connector. The BD Plastipak™ Syringe is provided sterile by Ethylene Oxide Gas (ETO) sterilization method. |
| --- | --- |
| Indications for Use | The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection. |
| Technological Characteristics | The following table provides a comparison between the subject and predicate devices. |
{6}
| Attributes | | Subject Devices | Predicate Devices | Comparison |
| --- | --- | --- | --- | --- |
| Intended Use/ Indications for Use | | The BD Plastipak™ Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection. | The BD Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection. | Identical |
| Operating Principle | | Piston Syringe | Piston Syringe | Identical |
| Syringe configuration | | • 20ml
• 50 ml
(Available in Luer-Slip, Eccentric Tip configuration) | • 1ml
• 3 ml
• 5 ml
• 10 ml
• 20 ml
• 30 ml
• 50 ml
(Available in Luer-lok, Luer-Slip, Eccentric Tip configuration) | **Different**
(The syringe sizes covered in the subject submission are within the range of the predicate device sizes) |
| Single Use | | Yes | Yes | Identical |
| Sterile | | Yes | Yes | Identical |
| Use in Sterile field | | Yes | Yes | Identical |
| Device Components | Barrel resin | Polypropylene (HPR35CMD) | Polypropylene (PH-712B) | **Different** |
| | Barrel Lubricant | Silicone Lubricant | Silicone Lubricant | Identical |
| | Barrel Ink | Black Ink | Black Ink | Identical |
| | Plunger Rod | Polypropylene | Polypropylene | Identical |
| | Stopper | Polyisoprene Rubber | Polyisoprene Rubber | Identical |
| | Stopper Lubricant | Silicone Lubricant | Silicone Lubricant | Identical |
| Packaging Configuration | • Blister pack (Primary) | • Blister pack (Primary) | Identical | |
| | | • Shelf Carton (Secondary) | | • Shelf Carton (Secondary) |
| Packaging Material | • Primary- Direct seal paper on Polymer Film | • Primary- Direct seal paper on Polymer Film | Identical | |
| | | • Secondary- Corrugated cardboard | | • Secondary- Corrugated cardboard |
Page 3 of 6
{7}
| Content of syringe package | Sterile: | One syringe per pack (Blister Pack) | One syringe per pack (Blister Pack) | Identical |
| --- | --- | --- | --- | --- |
| Sterilization Method | | ETO | ETO | Identical |
| SAL | | 10^{-6} | 10^{-6} | Identical |
| Shelf Life | | 5 years | 5 years | Identical |
Page 4 of 6
{8}
Page 5 of 6
Discussion
The subject device and predicate device are different with respect to the following item:
1. Change in component material (Barrel resin and supplier)
The purpose of this premarket submission is to introduce new barrel resin material and supplier for the BD Hypodermic Syringe cleared under K980987. Change is being made to allow contingency and reduce supply chain risk of BD Plastipak™ Syringe (subject device). The predicate device covered in Premarket notification K980987 are available in additional volumes i.e., 1ml, 3ml, 5ml, 10ml, 20ml, 30ml and 50ml sizes in Luer- lok, Luer- Slip, & Eccentric tip configuration. The subject device will be available in 20 ml and 50 ml Luer-Slip, Eccentric tip configuration that are covered within the range of the predicate device. Only 20 ml and 50 ml BD Plastipak™ Syringe are affected by the change described in the subject 510(k) Premarket Notification. No change will be made to the product's intended use, technological design, or performance.
The different technological characteristics between the subject and predicate device are evaluated in bench performance testing, and biocompatibility tests demonstrating that there are no new questions or modified risk of safety and effectiveness.
Non-Clinical Testing
BD has performed the following performance tests in accordance with 21 CFR §820.30 to demonstrate that the BD Plastipak™ Syringe performs equivalent to the predicate device.
These tests were performed on the subject device to an internal specification or a Standard:
{9}
| Performance Testing | |
| --- | --- |
| Test | Purpose |
| Breakout Force | ISO 7886-1:2017- Performance evaluation of force to operate the piston. |
| Sustaining Force | ISO 7886-1:2017- Evaluation of force to operate the piston. |
| Leakage Past Stopper | ISO 7886-1:2017- Evaluation of Freedom from air and liquid leakage past plunger stopper. |
| Volumetric Accuracy | ISO 7886-1:2017- Evaluation of Volumetric Accuracy |
| Dead Space | ISO 7886-1:2017- Evaluation of residual volume |
| Luer Leakage | ISO 80369-7:2021 - Evaluation of the luer fittings for leakage |
| Stress Cracking | ISO 80369-7:2021- Evaluation of the luer fittings for stress cracking |
| Resistance to separation from axial load | ISO 80369-7:2021- Evaluation of the luer fitting for separation when subjected to axial force. |
| Biocompatibility | |
| Cytotoxicity | ISO 10993-5:2009- Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity |
| Sensitization | ISO 10993-10:2021- Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization |
| Intracutaneous Reactivity | ISO 10993-23:2021- Biological evaluation of medical devices- Test for Irritation |
| Acute Systemic Toxicity | ISO 10993-11:2017- Biological evaluation of medical devices — Part 11: Tests for systemic toxicity |
| Material Mediated Pyrogenicity | ISO 10993-11:2017- Biological evaluation of medical devices — Part 11: Tests for systemic toxicity USP43-NF38 <151>Pyrogen Test (USP Rabbit Test) |
| Hemocompatibility | ISO 10993-4:2017- Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood ASTM F756-17- Standard Practice for Assessment of Hemolytic Properties of Materials |
The subject device met all the predetermined acceptance criteria for the above listed performance and biocompatibility tests.
| Clinical Testing | Not applicable. |
| --- | --- |
| Summary of Substantial Equivalence | The differences between predicate and the subject device do not raise any new questions or modified risks of safety or effectiveness when compared. Therefore, BD Plastipak™ Syringe is substantially equivalent to the predicate device cleared under K980987. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.