Nova Max Creat eGFR Monitoring System

K251281 · Nova Biomedical Corporation · SHB · Jan 21, 2026 · Clinical Chemistry

Device Facts

Record IDK251281
Device NameNova Max Creat eGFR Monitoring System
ApplicantNova Biomedical Corporation
Product CodeSHB · Clinical Chemistry
Decision DateJan 21, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Intended Use

The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional.

Device Story

System measures creatinine in fingertip capillary whole blood; uses electrochemical amperometry via test strip reagents. Patient applies blood to strip; monitor displays creatinine level and calculates eGFR (CKD-EPI 2021 equation) in 30 seconds. Designed for single-patient home use; operated by patient under physician order. Provides quantitative data to aid kidney function monitoring. Hand-held, battery-powered monitor features 2.8" color touchscreen; stores 400 results. Benefits patient by enabling convenient, prescription-based self-monitoring of renal health.

Clinical Evidence

Lay-user study (n=532, ages 23-93) compared patient-collected capillary fingerstick results to laboratory-measured venous plasma creatinine. Passing-Bablok regression showed creatinine slope 0.9873 (95% CI: 0.9782, 0.9966) and eGFR slope 1.035 (95% CI: 1.001, 1.068). Analytical performance included precision (CV% 3.8-12.17%), linearity (R² > 0.99), and interference testing for numerous substances. Device demonstrated robustness across altitudes up to 12,119 ft and operating conditions (15-40°C, 10-90% RH).

Technological Characteristics

Amperometric enzyme-based creatinine measurement. 3.7V Li-polymer rechargeable battery. 2.8" color touchscreen. Dimensions: 95.25 x 61.98 x 18.80 mm. Weight: 90g. Hematocrit range: 20-70%. Measurement range: 0.30-7.00 mg/dL. Connectivity: None. Traceable to NIST SRM 967b.

Indications for Use

Indicated for adult patients aged 18+ for quantitative measurement of creatinine and eGFR in fresh capillary whole blood from fingertips. Intended for single-patient home use by prescription to monitor kidney function.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION January 21, 2026 Nova Biomedical Corporation Mariya Cesnulevicius Regulatory Affairs Manager 200 Prospect St. Walham, MA 02454 Re: K251281 Trade/Device Name: Nova Max Creat eGFR Monitoring System Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine Test System Regulatory Class: Class II Product Code: SHB Dated: December 22, 2025 Received: December 22, 2025 Dear Mariya Cesnulevicius: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251281 - Mariya Cesnulevicius Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251281 - Mariya Cesnulevicius Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, PAULA V. CAPOSINO -S Paula Caposino, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251281 | | | Device Name Nova Max Creat eGFR Monitoring System | | | Indications for Use (Describe) The Nova Max Creat eGFR Monitoring System is comprised of the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Test Strips. The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} nova # 510(k) Summary 510(k) Owner: Nova Biomedical Corporation Establishment Registration Number: 1219029 510(k) Number: K251281 Address: 200 Prospect Street Waltham, MA, 02454 USA Phone: 781-894-0800 Fax Number 781-899-6259 ## Primary Contact Person Mariya Cesnulevicius Regulatory Affairs Manager ## Secondary Contact Person Robert Zinck, Director of Regulatory and Clinical Affairs ## Date Prepared January 21, 2026 ## Trade Name Nova Max Creat eGFR Monitoring System ## Common Name Creatinine Test System ## Classification Name | Classification Name | Product Code | Regulation Number | Classification | 510(k) Review Panel | | --- | --- | --- | --- | --- | | Creatinine test system | SHB | 862.1225 | Class II | Clinical Chemistry | ## Predicate Device K171059, Nova StatSensor Creatinine Hospital Meter System {5} Page 2 of 7 # Device Description The Nova Max Creat eGFR Monitoring System consists of Nova Max Create eGFR Monitor and Nova Max Create eGFR Creatinine-Test Strips. The Nova Max Creat eGFR test strips, Nova Max Creat eGFR control solution (Level 1 and Level 2), and a single-use, disposable, 21 gauge safety lancet can be purchased separately. # Nova Max Creat eGFR Monitor The monitor is intended to be used in conjunction with the Nova Max Creat eGFR Creatinine-Test Strips to measure creatinine and calculate estimated glomerular filtration rate (eGFR) using capillary whole blood obtained from the fingertip. The hand-held monitor is lightweight, portable, utilizes a 2.8" color touchscreen display and has capability for storage (400 test samples and/or quality control samples) and review of data. The monitor is powered by a rechargeable, 3.7 V Li polymer battery. # Nova Max Creat eGFR Creatinine-Test Strips The Nova Max Creat eGFR Creatinine-Test Strip is designed with an electrode that measures creatinine levels. Creatinine in the capillary whole blood sample mixes with the reagents in the test strip to produce an electric current. The amount of current that is produced depends on how much creatinine is in the sample. A digital readout is displayed on the monitor in 30 seconds, the eGFR is calculated using patient information and the measured creatinine level. The test strips are for single use only. # Nova Max Creat eGFR Creatinine Control Solutions The Nova Max Creat eGFR Creatinine Control Solution is used as a quality control check to ensure that the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Creatinine-Test Strips are working properly as a system. There are two levels of Nova Max Creat eGFR Creatinine Control Solution. Each level of control solution has a known creatinine concentration that reacts with the reagents in the test strip. # Indications for Use The Nova Max Creat eGFR Monitoring System is comprised of the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Test Strips. The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional. # Summary of Performance Testing Analytical bench and clinical testing were conducted to demonstrate that the Nova Max Creat eGFR Monitoring System achieves its intended purpose at the intended use setting. # Linearity Testing Testing was performed to validate the Nova Max Creat eGFR Monitoring System linearity. Ten (10) specimens were tested on the Nova Max Creat eGFR Monitoring System. The same specimens were analyzed on the reference method. The results met the acceptance criteria, validating the linearity of the Nova Max Creat eGFR Monitoring System across the measurement range of 0.30 mg/dL to 7.00 mg/dL. # Analytical Specificity The analytical specificity for the Nova Max Creat eGFR Monitoring System was assessed by testing the effect of various exogenous and endogenous substances on the measurement of creatinine by the system. Whole blood samples were spiked with forty-six (46) potential interferents and the creatinine concentration was measured with the Nova Max Creat eGFR Monitoring System. {6} The test substance was considered to be an interfering substance if the absolute difference between the test value and the control value was greater than 10% for creatinine concentration higher than 2 mg/dL, and greater than ± 0.2 mg/dL if the creatinine concentration is less than 2 mg/dL. Based upon the results, all potential interfering substances tested met the acceptance criteria and were demonstrated not to have clinical interference. No clinical interference was identified up to the test concentrations reported in Table 1. Table 1, Interference Test Result Summary | # | Interferent | Interference concentration | # | Interferent | Interference concentration | | --- | --- | --- | --- | --- | --- | | 1 | Acetaldehyde | 0.2 mg/dL | 24 | Glucose | 1000 mg/dL | | 2 | Acetaminophen | 15.6 mg/dL | 25 | High Hct and Low Hct | 20 – 70 % | | 3 | N-Acetyl-L-Cysteine | 15 mg/dL | 26 | Ibuprofen | 22 mg/dL | | 4 | Ascorbic Acid | 5.25 mg/dL | 27 | Paracetamol-4-acetamidopenol | 30 mg/dL | | 5 | β-Hydroxybutyric Acid | 65 mg/dL | 28 | Pyruvate | 5 mg/dL | | 6 | Potassium bicarbonate | 294 mg/dL | 29 | Lithium Salicylate | 3 mg/dL | | 7 | Bilirubin (unconjugated) | 40 mg/dL | 30 | Potassium thiocyanate | 6 mg/dL | | 8 | Calcium Chloride | 20 mg/dL | 31 | Urea | 120 mg/dL | | 9 | Creatine | 5 mg/dL | 32 | Farxiga (dapagliflozin) | 0.6 mg/dL | | 10 | Dopamine Hydrochloride | 2 mg/dL | 33 | Kerendia (finerenone) | 1.2 mg/dL | | 11 | Formaldehyde | 0.4 mg/dL | 34 | Jardiance (empagliflozin) | 1.5 mg/dL | | 12 | Glycolic Acid | 76 mg/dL | 35 | Lithium Heparin | 1.2 mg/dL | | 13 | Hydroxyurea | 4 mg/dL | 36 | Lithium Bromide | 325.69 mg/dL | | 14 | Lithium Lactate | 90 mg/dL | 37 | Hemoglobin | 2600 mg/dL | | 15 | Methyldopa | 5 mg/dL | 38 | Cholesterol | 1000 mg/dL | | 16 | pH | 8.0 | 39 | L-Dopa | 0.3 mg/dL | | 17 | Sodium Thiosulfate | 265 mg/dL | 40 | Maltose | 360 mg/dL | | 18 | Triglyceride | 1500 mg/dL | 41 | Intralipid | 2000 mg/dL | | 19 | Uric Acid | 23.5 mg/dL | 42 | Lactic Acid | 108.1 mg/dL | | 20 | Acetoacetate | 20 mg/dL | 43 | Salicylic Acid | 60 mg/dL | | 21 | Acetylsalicylic Acid | 3 mg/dL | 44 | Semaglutide (Ozempic or Wegovy) | 0.144 mg/dL | | 22 | Chlorpromazine HCl | 0.4 mg/dL | 45 | Tirzepatide (Mounjaro) | 0.9 mg/dL | | 23 | Ethanol | 600 mg/dL | 46 | Bilirubin (conjugated) | 40 mg/dL | ## Limits of Detection and Quantification Testing was conducted to assess the analytical sensitivity of the Nova Max Creat eGFR Monitoring System. Blank whole blood samples were tested to estimate the limit of the blank (LoB). This value was used to select a low level of creatinine concentration to determine the LoD and LoQ. Low level samples were prepared with creatinine concentration close to the estimated low value for creatinine concentration. These samples were tested repeatedly on the Nova Max Creat eGFR Monitoring System. Samples were also tested on the reference analyzer. The study yielded the following results for creatinine: Limit of Blank (LoB) = 0.078 mg/dL Limit of Detection (LoD) = 0.174 mg/dL Limit of Quantification (LoQ) = 0.174 mg/dL. Page 3 of 7 {7} Page 4 of 7 # Traceability Nova Max Creat eGFR Creatinine Control Materials are traceable to NIST Standard Reference Material 967b # Precision Total Imprecision of the Nova Max Creat eGFR Monitoring System was assessed using methods described in CLSI "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guidelines – Second Edition", CLSI EP5-A3 as guidance. # Day-to-Day Imprecision To investigate day-to-day precision five (5) levels of Nova Max Creatinine Aqueous Solutions were used. Two (2) Nova Max Creat eGFR Monitors and three (3) Nova Max Creat eGFR Creatinine-Test Strip lots were tested in the study. Each level of the aqueous solutions was analyzed in duplicates, twice a day over twenty (20) days. The results (presented in Table 2) met the acceptance criteria indicating that the day-to-day precision of the Creatinine assay on the Nova Max Creat eGFR Monitoring Systems is clinically acceptable. Table 2. Day-to-Day Imprecision Summary | Aqueous Solution | | Test Strip Lot 1 | Test Strip Lot 2 | Test Strip Lot 3 | | --- | --- | --- | --- | --- | | Level 1 | Mean, mg/dL | 0.56 | 0.57 | 0.58 | | | SD | 0.06 | 0.06 | 0.06 | | Level 2 | Mean, mg/dL | 1.33 | 1.35 | 1.37 | | | SD | 0.06 | 0.08 | 0.08 | | Level 3 | Mean, mg/dL | 2.58 | 2.65 | 2.66 | | | CV (%) | 5.00 | 4.64 | 4.93 | | Level 4 | Mean, mg/dL | 4.13 | 4.14 | 4.09 | | | CV (%) | 4.89 | 4.98 | 4.98 | | Level 5 | Mean, mg/dL | 6.03 | 5.97 | 6.04 | | | CV (%) | 4.85 | 4.99 | 4.87 | # Within-run Imprecision To investigate the precision using whole blood, five (5) blood specimens with creatine concentrations spanning the measurement range were tested. Five (5) Nova Max Creat eGFR Monitors and three (3) Nova Max Creat eGFR Creatinine-Test Strip lots were used in the study. The results (presented in Table 3) met the acceptance criteria, indicating that the Nova Max Creat eGFR System report clinically acceptable precision. Table 3. Within-run Imprecision Summary | Whole Blood Specimens | | Test Strip Lot 1 | Test Strip Lot 2 | Test Strip Lot 3 | | --- | --- | --- | --- | --- | | Level 1 | Mean, mg/dL | 0.47 | 0.47 | 0.49 | | | SD | 0.05 | 0.06 | 0.06 | | Level 2 | Mean, mg/dL | 0.82 | 0.83 | 0.83 | | | SD | 0.06 | 0.06 | 0.06 | | Level 3 | Mean, mg/dL | 1.68 | 1.68 | 1.66 | | | SD | 0.09 | 0.10 | 0.10 | | Level 4 | Mean, mg/dL | 2.81 | 2.88 | 2.83 | | | CV (%) | 4.6% | 4.4% | 4.8% | {8} Page 5 of 7 | Level 5 | Mean, mg/dL | 5.83 | 5.88 | 5.97 | | --- | --- | --- | --- | --- | | | CV (%) | 4.7% | 5.0% | 4.9% | ## Flex Studies Nova Biomedical conducted the following flex studies to validate the robustness of the Nova Max Creat eGFR Monitoring System: 1. Short sample 2. Used Nova Max Creat eGFR Creatinine-Test Strips 3. Incorrect test strips 4. Moving during analysis 5. Atypical positions 6. Removing Nova Max Creat eGFR Creatinine-Test Strips during the analysis 7. Double dosing 8. Delayed analysis 9. Flicking 10. Damaged Strip 11. Incorrect Nova Max Creat eGFR Creatinine Control Solutions storage 12. Uncapped Nova Max Creat eGFR Creatinine Control Solutions 13. Extreme Environmental Conditions 14. Improper sampling The results of the Flex Studies were determined to meet the Acceptance Criteria and error codes were correctly generated (where applicable) as specified in the Instructions for Use Manual. ## Clinical Performance Nova Biomedical conducted a clinical study in the hands of lay users. The clinical study was conducted across two (2) sites with a total of 532 subjects. Measurements on fingerstick samples taken by lay users using the Nova Max Creat eGFR Monitoring System were compared to venous plasma creatinine results measured by laboratory operators using an IDMS traceable comparative method. The Pearson correlation coefficient (r) between the fingerstick blood readings obtained by lay users and the corresponding comparative method readings obtained by laboratory operators was calculated using least squares linear regression and is displayed in Table 4. Table 4, Creatinine Linear Regression Analysis Summary | Slope: | 0.9832 | | --- | --- | | Y-Intercept: | 0.0073 | | Correlation Coefficient (r): | 0.9951 | The bias between the capillary whole blood creatinine readings from the Nova Max Creat eGFR Monitoring System and the venous plasma creatinine readings measured by the comparative method was evaluated and shown in Table 5. Table 5, Within Clinical Study Accuracy (Bias/Bias%) | Less than 2 mg/dL SD ≤ 0.2 mg/dL | Greater than 2 mg/dL CV ≤ 10% | | --- | --- | | 97.87% | 95.36% | {9} The results showed that the Nova Max system's measurements were in strong agreement with the comparative method, meeting the clinical accuracy acceptance criteria in the hands of lay users. Table 6: Comparison of Predicate and Proposed Devices | | Predicate Device: Nova StatSensor Creatinine Hospital Meter System | Subject Device: Nova Max Creat GFR Monitoring System | | --- | --- | --- | | Intended Use and Indications for Use | The StatSensor Creatinine Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of creatinine in capillary, venous, and arterial whole blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis. Not for use in neonates. | The Nova Max Creat eGFR Monitoring System is comprised of the Nova Max Creat eGFR Monitor and the Nova Max Creat eGFR Test Strips.The Nova Max Creat eGFR Monitoring System is intended for in vitro diagnostic use for the quantitative measurement of creatinine and estimation of glomerular filtration rate (eGFR) in fresh capillary whole blood obtained from the fingertip of adult patients aged 18 and above. The system is intended for single patient home use by prescription only and should not be shared. It is intended for use by patients as an aid to monitor kidney function on the order of a treating healthcare professional. | | Intended Use Environment | Prescription use, point-of-care settings | Home use by prescription | | Hand-held meter | Yes | Same | | Test Measured/Reported Parameters | Creatinine and eGFR | Creatinine and eGFR | | Operating Principle | Enzyme, Amperometry | Same | | eGFR equations | CKD-EPI (2009), MDRD and Cockcroft-Gault equation | CKD-EPI 2021 equation | | Sample Type | Capillary, Venous and Arterial whole blood specimen | Fingertip capillary whole blood specimen | | Sample Volume | 1.2 μL | Same | | Sample Application | Test strip capillary draw | Same | | Measurement Range for Creatinine | 0.3 - 12.0 mg/dL (1 significant digit) | 0.30 - 7.00 mg/dL (2 significant digits) | | Test Range for eGFR calculated using the CKD-EPI equation | 15 - 90 mL/min/1.73 m2 | Same | | Analysis Time | 30 seconds | Same | | Hematocrit Range | 30% to 60% | 20% to 70% | | Quality Control | Three Levels | Two Levels | | Linearity | Five Levels | None | | Calibration | Automatic, No Calibration Code | Same | {10} | Location of the test strip port | Bottom of the meter | Same | | --- | --- | --- | | Physical dimensions of the meter | 153 mm (6.0 in) x 82.5 mm (3.25 in) x 46 mm (1.8 in) | 95.25 mm (3.75 in) x 61.98 mm (2.44 in) x 18.80 mm (0.74 in) | | Weight | 360 grams | 90 grams | | Menu navigation | Color touch screen display | Same | | Power/Energy Source | 3.7 V Li-Polymer battery (Rechargeable/Replaceable) | 3.7 V Li-Polymer battery (Rechargeable/Non-Replaceable) | | Battery Charging | Charged via a Desk-mount Docking/ Charging Station | Charged like a cell phone, using an external power supply connected to an electric outlet, and a charging cable connecting the power supply to the monitor. | | Meter Memory | 1000 patient results 200 QC results 4000 Operators | 400 Patient and/or QC Results | | Data Transfer | Via Docking Station Ethernet Connectivity | No data transfer is available | | Operating Temperature Range | 59°F to 104°F (15°C to 40°C) | Same | | Operating Humidity Range | 10% to 90% relative humidity | Same | | Operating Altitude Range | Up to 15,000 feet (4572 meters) | Up to 12,000 feet (3658 meters) | | Test Strip Storage Temperature (Closed Vial) | 35.6°F to 46.4°F (2°C to 8°C) | Same | | Test Strip Shelf Life | 24 months | Same | | Control Solutions Storage Temperature (Closed Vial) | 35.6°F to 46.4°F (2°C to 8°C) | Same | | Control Solutions Shelf Life | 24 months | Same | ## Conclusion Based on the performance data provided with this submission, it can be concluded that the Nova Max Creat eGFR Monitoring System has been demonstrated to be safe, and effective for its intended use for monitoring of kidney health in the intended use setting. The similarities in the intended use and technological characteristics between the proposed Nova Max Creat eGFR Monitoring System and the FDA cleared predicate device, StatSensor Creatinine Hospital Meter System, support that the two devices are substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...